510(k) Data Aggregation

    K Number
    K252287

    Validate with FDA (Live)

    Date Cleared
    2025-11-10

    (110 days)

    Product Code
    Regulation Number
    878.4963
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Symani® Surgical System is indicated for soft tissue manipulation in open microsurgical procedures to perform:

    • Anastomosis, suturing, and ligation microsurgery techniques on small blood vessels and lymphatic ducts between 0.1 and 2.5mm in open free-flap surgery of the breast, mouth, scalp, and extremities and in open lymphatic surgery of the extremities.
    • Dissection of soft tissues.
    Device Description

    Symani is designed to provide surgeons with increased precision while leveraging the surgeon's own surgical experiences. The Symani Surgical System consists of three components, the CMM, Console, and Instruments, with selectable scaling factors between 7x to 20x to improve motion control and scale down any physiological tremor of the operator in a similar fashion to conventional robotic-assisted surgical device (RASD) systems, and the instruments with wristed end-effectors to improve dexterity for tissue manipulation and suturing. The subject device has the same intended use as the predicate device, K242358. Subject device includes two new NanoWrist instruments, Scissors and Forceps.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K251510

    Validate with FDA (Live)

    Date Cleared
    2025-09-04

    (111 days)

    Product Code
    Regulation Number
    878.4963
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Symani® Surgical System is intended for soft tissue manipulation to perform anastomosis, suturing, and ligation microsurgery techniques on small blood vessels and lymphatic ducts between 0.1 and 2.5 mm in open free-flap surgery of the breast, mouth, scalp, extremities, and in open lymphatic surgery of the extremities.

    The Symani® Surgical System is indicated for use during microsurgical procedures when use of a motion scaling function is deemed appropriate by the surgeon. The System is indicated for use in adults. It is intended to be used by trained physicians in an appropriate operating environment in accordance with the Instructions for Use.

    Device Description

    Symani is designed to provide surgeons with increased precision while leveraging the surgeon's own surgical experiences. The Symani Surgical System consists of three components, the Cart with Macropositioner and Micromanipulators (CMM), Console, and Instruments, with selectable scaling factors between 7x to 20x to improve motion control and scale down any physiological tremor of the operator in a similar fashion to conventional robotic-assisted surgical device (RASD) systems, and the instruments with wristed end-effectors to improve dexterity for tissue manipulation and suturing. The subject device has the same intended use and indications for use as the predicate device, K242358.

    The Digital Workstation has been added to list of Symani's accessories for displaying surgical images through the 3D Microscopes connected to the Digital Workstation and the overlay information from Symani GUI (when connected to Symani), using Ultra High Definition (UHD) technology to display high-quality resolution images.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K242368

    Validate with FDA (Live)

    Date Cleared
    2025-02-07

    (182 days)

    Product Code
    Regulation Number
    878.4963
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Symani® Surgical System is intended for soft tissue manipulation to perform anastomosis, suturing, and ligation microsurgery techniques on small blood vessels and lymphatic ducts between 0.1 and 2.5 mm in open free-flap surgery of the breast, mouth, scalp, extremities, and in open lymphatic surgery of the extremities. The Symani® Surgical System is indicated for use during microsurgical procedures when use of a motion scaling function is deemed appropriate by the surgeon. The System is indicated for use in adults. It is intended to be used by trained physicians in an appropriate operating environment in accordance with the Instructions for Use.

    Device Description

    Symani is designed to provide surgeons with increased precision while leveraging the surgeon's own surgical experiences. The Symani Surgical System consists of three components, the CMM, Console, and Instruments, with selectable scaling factors (between 7x to 20x) to improve motion control and scale down any physiological tremor of the operator in a similar fashion to conventional robotic-assisted surgical device (RASD) systems, and the instruments with wristed end-effectors to improve dexterity for tissue manipulation and suturing.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study proving the device meets those criteria, based on the provided text.

    Note: The provided document is an FDA 510(k) clearance letter and an accompanying 510(k) summary. It primarily focuses on demonstrating substantial equivalence to a predicate device rather than explicitly defining and meeting pre-defined performance acceptance criteria with numerical targets. However, the document does describe the performance data used to support the claim of substantial equivalence for the new indications (mouth and scalp). Therefore, the "acceptance criteria" inferred below are based on the comparison of the device's performance to the predicate device and general expectations for safety and effectiveness in this medical context.

    1. Table of Acceptance Criteria and Reported Device Performance

    Strict numerical "acceptance criteria" for the new indications are not explicitly stated in the document as a pass/fail threshold. Instead, the performance is compared to the predicate device and general clinical expectations. The implicit acceptance criteria are that the device is "as safe and effective" as the predicate for the added indications.

    Criterion (Inferred from Clinical Outcomes)Device Performance (Mouth)Device Performance (Scalp)Comparison/Narrative towards "Acceptance"
    Effectiveness:
    Intraoperative patency at first attempt (Robotic approach)90.9% (60/66) [95% CI: 81.3%, 96.6%]69.2% (18/26) [95% CI: 48.2%, 85.7%]Mouth: "similar to the predicate device." Scalp: "lower than the predicate device," but "patency of all anastomoses (100%) was confirmed prior to closure." (i.e., issues at first attempt were resolved).
    Operative time (average hours)11.66 hours12.30 hoursProvided for context; no direct comparison to a specific target or predicate.
    Suturing time (average minutes) - Robotic37.4 min41.2 minProvided for context; no direct comparison to a specific target or predicate.
    Rate of intra-operative change from robotic to manual suturing4.2% (3/72) [95% CI: 0.9%, 11.7%]7.7% (2/26) [95% CI: 0.9%, 25.1%]Overall estimated change rate 5.1% (5/98); no explicit target, but implies the system allows for manual fallback.
    Safety:
    Freedom from Serious Adverse Device Effects (SADE)100.0% (0/41) [95% CI: 0.0%, 8.6%]100.0% (0/20) [95% CI: 0.0%, 16.8%]100% freedom from SADE across both sites. This is a strong safety indicator.
    Serious Adverse Events (SAE) Rate (patient-based)12.2% (5/41)10.0% (2/20)"lower than the predicate device" (implied as a positive outcome).
    Adverse Events (AE) Rate (patient-based)41.0% (25/61 patients across both sites)Consistent with literature range (21.5% - 61.3%).Presented as being "consistent with the range identified in the literature search."
    Rate of adverse events resulting in anastomosis revision5% (2/40) for Mouth0% (0/20) for ScalpNo specific target, but provides insight into the clinical impact of AEs.
    3-Day Reoperation Rate (anastomosis specific)5.00% (2/40) [95% CI: 0.61% - 16.92%]0.00% (0/20) [95% CI: 0.00% - 16.84%]Presented as a critical clinical outcome; no direct comparison to a target or predicate given here, but low rates are favorable.

    The study concludes that the device is "as safe and effective as the predicate device" for the new indications, despite the lower initial patency rate for the scalp, due to the 100% confirmed patency post-procedure and favorable safety profile.

    2. Sample Size and Data Provenance

    • Test Set Sample Size: A total of 62 patients were analyzed for the specific free tissue transfer procedures using Symani with mouth or scalp as recipient sites.
      • Mouth: 40 patients
      • Scalp: 20 patients
      • (Note: The patient counts for SADE and SAE are slightly different (e.g., 41 mouth, 20 scalp for SADE) which might indicate minor variations in data availability for specific endpoints or inclusion/exclusion for analysis. The total of 62 patients is the overall 'subset' considered.)
    • Data Provenance: The data comes from a prospective, non-randomized, single-arm, multicenter, post-market clinical follow-up (PMCF) study.
      • Retrospective/Prospective: 55 patients were prospective enrollments, and 7 were retrospective enrollments.
      • Country of Origin: Not explicitly stated in the provided text. The submitter's address is Pisa, Italy, which suggests a European origin for the study, but this is not confirmed for the patients themselves.

    3. Number of Experts and Qualifications for Ground Truth

    The document does not describe the establishment of a "ground truth" using a panel of experts for the test set. Instead, the clinical outcomes (e.g., patency, adverse events, reoperations) are the direct data points collected from patient procedures. These are presumably assessed by the treating surgeons and study staff involved in the PMCF study.

    4. Adjudication Method for the Test Set

    No explicit adjudication method (like 2+1 or 3+1 consensus) for the "test set" (clinical outcomes) is described. The outcomes presented are directly from the clinical study data collection.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC study was performed or described. This type of study is typically used for diagnostic devices to assess how AI affects reader performance. The Symani Surgical System is an electromechanical surgical system, not a diagnostic AI. The study evaluated real-world surgical outcomes, not AI assistance to human readers interpreting images.

    6. Standalone Performance Study (Algorithm Only)

    The Symani Surgical System is a robotic-assisted surgical device with human-in-the-loop control. Therefore, a "standalone" or "algorithm-only" performance study (without human involvement) is not applicable or described, as the device's function is inherently a human-device interaction.

    7. Type of Ground Truth Used

    The ground truth used is primarily clinical outcomes data obtained directly from patient procedures in a post-market clinical follow-up (PMCF) study. This includes:

    • Intraoperative findings (e.g., patency confirmed prior to closure)
    • Adverse events (SAE, AE)
    • Reoperation rates

    8. Sample Size for the Training Set

    The document does not mention a "training set" in the context of an AI algorithm being trained. The Symani Surgical System is described as having "selectable scaling factors" and "wrested end-effectors" for "motion control" and "dexterity." This suggests a mechanical/software system guided by explicit programming and control algorithms, not a machine learning model that requires a "training set" in the contemporary AI sense (e.g., for image recognition). The "predicate device" is the same Symani Surgical System, implying improvements or expanded indications of an existing (already developed) system rather than a new de novo AI system.

    9. How Ground Truth for the Training Set Was Established

    As there is no mention of an AI "training set" in the provided document for this device, this question is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K250014

    Validate with FDA (Live)

    Date Cleared
    2025-01-27

    (25 days)

    Product Code
    Regulation Number
    878.4963
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Symani® Surgical System is intended for soft tissue manipulation to perform anastomosis, suturing, and ligation microsurgery techniques on small blood vessels and lymphatic ducts between 0.1 and 2.5 mm in open free-flap surgery of the breast and extremities and open lymphatic surgery of the extremities. The Symani® Surgical System is indicated for use during microsurgical procedures when use of a motion scaling function is deemed appropriate by the surgeon. The System is indicated for use in adults. It is intended to be used by trained physicians in an appropriate operating environment in accordance with the Instructions for Use.

    Device Description

    Symani is designed to provide surgeons with increased precision while leveraging the surgeon's own surgical experiences. The Symani Surgical System consists of three components, the CMM, Console, and Instruments, with selectable scaling factors between 7x to 20x to improve motion control and scale down any physiological tremor of the operator in a similar fashion to conventional robotic-assisted surgical device (RASD) systems, and the instruments with wristed end-effectors to improve dexterity for tissue manipulation and suturing. The subject device has the same intended use and indications for use, and same technological characteristics as the predicate device, DEN230032. The Supermicro Needle Holder Suture Cut instrument has been added to the NanoWrist Instruments' portfolio.

    AI/ML Overview

    This document describes a 510(k) submission for the Symani Surgical System, specifically for the addition of a new instrument called the Supermicro Needle Holder Suture Cut.

    Here's an analysis of the acceptance criteria and study information provided, noting what is and isn't available in the document:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly present acceptance criteria in a table format with specific quantitative thresholds. Instead, it describes performance tests conducted to demonstrate that the new instrument maintains the safety and effectiveness of the existing system. The reported performance is an overall conclusion of substantial equivalence.

    Implied Acceptance Criteria and Reported Performance from the Text:

    Acceptance Criteria CategoryReported Device Performance
    Functional EquivalenceThe Supermicro Needle Holder Suture Cut design is the same as the current Supermicro Needle holder with the additional features of a small (micro) blade mounted on the driving tip.
    Simulated UsePerformance tests (including simulated use) demonstrated that use of the Symani Surgical System with the new instrument is as safe and effective as the predicate device.
    Gripping Force & TorquePerformance tests (including gripping force & torque) demonstrated that use of the Symani Surgical System with the new instrument is as safe and effective as the predicate device.
    Structural IntegrityPerformance tests (including structural integrity tests) demonstrated that use of the Symani Surgical System with the new instrument is as safe and effective as the predicate device.
    Overall Safety & EffectivenessThe subject device is deemed to be substantially equivalent to the legally commercialized predicate device (DEN230032).

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size: Not specified. The document only lists "Simulated use test," "Gripping force & torque," and "Structural integrity tests" without detailing the number of test runs, materials used, or specific data points.
    • Data Provenance: Not specified. It's implied that the testing was conducted by Medical Microinstruments, Inc. in Pisa, Italy, as that's their location, but no direct statement on data origin is provided. The study appears to be prospective testing of the new instrument's performance characteristics.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. This device is a surgical instrument and the performance tests described (simulated use, gripping force & torque, structural integrity) are engineering and functional tests, not clinical studies requiring expert interpretation or ground truth establishment in the traditional sense of diagnostic imaging or AI performance. The assessment is against established device performance parameters and comparison to a predicate.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. As described above, this is not a study requiring human adjudication of results in the context of diagnostic interpretation. The performance tests would be evaluated against engineering specifications and comparison to the predicate.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI-powered diagnostic device, but a surgical instrument. Therefore, MRMC studies and AI-assistance effect sizes are not relevant to this submission.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a physical surgical instrument, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Not applicable. The "ground truth" for the performance tests of a surgical instrument involves engineering specifications, material properties, mechanical testing standards, and comparison to the functionally equivalent predicate device's performance. There is no biological or diagnostic "ground truth" in this context.

    8. The sample size for the training set

    Not applicable. As noted, this is a physical medical device, not an AI/ML algorithm requiring a training set.

    9. How the ground truth for the training set was established

    Not applicable. No training set is involved.