510(k) Data Aggregation

    K Number
    K253314

    Validate with FDA (Live)

    Date Cleared
    2025-10-29

    (29 days)

    Product Code
    Regulation Number
    888.3560
    Age Range
    All
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Freedom Infinia™ Total Knee System is indicated for the following:

    • Severe knee joint pain, loss of mobility, and disability due to: rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis.
    • Correction of functional deformities.
    • Post-traumatic loss of knee joint contour, particularly when there is patellofemoral erosion, dysfunction, or prior patellectomy.
    • Moderate valgus, varus, or flexion trauma.
    • Knee fractures untreatable by other methods.
    • Revision surgery where sufficient bone stock and soft tissue integrity are present. (For PCK Components and Primary PCK Components only)

    The Freedom Infinia™ Total Knee System components are intended for cemented and single use only. The Freedom Infinia™ Cementless Femoral (CR and PS) Components and Porous Tibial Base Plates are additionally indicated for cementless biological fixation application.

    Device Description

    The Freedom® Total Knee System is comprised of a femoral component, all-poly tibial component, patellar component, tibial base plate and tibial articular surface. The Freedom® Total Knee System's Femoral Component is offered as different versions such as stemmed PCK design, primary PCK, cruciate retaining, posterior stabilizing. The Freedom® Total Knee System's Tibial Base Plate is offered with stemmed design and without stemmed design. The Freedom® Total Knee System was originally cleared under the 510(k) number K082019. Later on, several modifications were made and were cleared under 510(k)s K090411, K091280, K111785, K131481, K150680, K182574, K192148, K241597, K243277, K251717 respectively.

    This submission seeks the clearance of an additional brand, Freedom Infinia™ Total Knee System.

    Below is the description of the Freedom Infinia™ Total Knee System's components.

    Freedom Infinia™ Femoral Component

    The Freedom Infinia™ Total Knee System's Femoral Component is offered in both cruciate retaining and posterior stabilizing designs. The Femoral components are available in right and left configurations, and eight sizes to accommodate varying anatomy. The Freedom Infinia™ Total Knee System's Femoral Component is fabricated from Cobalt-Chromium-Molybdenum (CoCrMo) alloy complying with ASTM F75 and is intended for cemented application to replace the articulating surface of the distal femur in a measured resection.

    Freedom Infinia™ Cementless Femoral Component

    The Freedom Infinia™ Total Knee System's Cementless Femoral Component is offered in both cruciate retaining and posterior stabilizing designs. The Cementless Femoral Components are available in right and left configurations, and eight sizes to accommodate varying anatomy. The Freedom Infinia™ Total Knee System's Cementless Femoral Component is fabricated from Cobalt-Chromium-Molybdenum (CoCrMo) alloy complying with ASTM F75 and porous coated CoCr beads and is intended for cementless biological fixation application to replace the articulating surface of the distal femur in a measured resection.

    Freedom Infinia™ Stemmed Femoral Component

    The Freedom Infinia™ Total Knee System's Stemmed Femoral Component with progressive constraint (PCK) is fabricated from Cobalt-Chromium-Molybdenum (CoCrMo) alloy complying with ASTM F75, intended for cemented application to replace the articulating surface of the distal femur in a measured resection technique. It is available in left and right configurations. Each configuration is further available in 8 different sizes (A to H) based on Anterior/Posterior (A/P) and Medial/Lateral (M/L) dimensions. Thus total of 16 models are available for the Stemmed Femoral Component.

    Freedom Infinia™ Femoral Augments

    The Freedom Infinia™ Total Knee System's Femoral Augments are fabricated from Titanium-6Aluminum-4Vanadium (Ti-6A1-4V), complaint with ASTM F136. These augments are intended to be screwed to the internal distal and posterior surfaces of Freedom Infinia™ PCK Stemmed Femoral Component when required in cases of significant bone loss. The augments are designed to be stackable and for use on either the medial or lateral side.
    Augment Screws, fabricated from Titanium-6Aluminum-4Vanadium (Ti-6A1-4V), complaint with ASTM F136, are available as an accessory to the system to fix the augments to the Freedom Infinia™ Stemmed Femoral Component and other augments.
    The Freedom Infinia™ Femoral Augments is available in distal and posterior design configuration. Each configuration is further available in 6 different sizes.

    Freedom Infinia™ Primary PCK Femoral Component

    Freedom Infinia™ Total Knee System's Primary PCK Femoral Component is designed to be used with the Freedom Infinia™ Stemmed Tibial Components. It is fabricated from Cobalt- Chromium-Molybdenum (CoCrMo) alloy complying with ASTM F75 and is intended for cemented application.

    Freedom Infinia™ Tibial Base Plate

    The Freedom Infinia™ Total Knee System's Tibial Base Plate is fabricated from Cobalt- Chromium-Molybdenum (CoCrMo) alloy complying with ASTM F75 and is intended for cemented application to replace the articulating surface of the proximal tibia in a measured resection technique. The Tibial Base Plates are available in 8 different sizes (1 to 8) based on Anterior/Posterior (A/P) and Medial/Lateral (M/L) dimensions.

    Freedom Infinia™ Titanium Tibial Base Plate

    The Freedom Infinia™ Total Knee System's Titanium Tibial Base Plate is fabricated from Wrought Titanium alloy Ti-6Al-4V ELI, compliant with ASTM F136-13 and is intended for cemented application to replace the articulating surface of the proximal tibia in a measured resection technique. The Titanium Tibial Base Plates are available in 8 different sizes (1 to 8) based on Anterior/Posterior (A/P) and Medial/Lateral (M/L) dimensions.

    Freedom Infinia™ Porous Tibial Base Plate

    The Freedom Infinia™ Total Knee System's Porous Tibial Base Plate is intended for use with existing, compatible Femoral Components and Tibial Liner Components. Freedom Infinia™ Porous Tibial Base Plates are additively manufactured from Ti-6Al-4V ELI Grade 23 and include a porous lattice structure on the distal face. Freedom Infinia™ Porous Tibial Base Plates are available in eight asymmetric design offerings (Sizes 1 – 8, Left / Right configurations), based on anterior / posterior (A/P) and medial / lateral (M/L) dimensions.

    Freedom Infinia™ Stemmed Tibial Base Plate

    The Freedom Infinia™ Total Knee System's Stemmed Tibial Base Plate is fabricated from Cobalt-Chromium-Molybdenum (CoCrMo) alloy complying with ASTM F75 and is intended for cemented application to replace the articulating surface of the proximal tibia in a measured resection technique. The Stemmed Tibial Base Plates are available in 8 different sizes (1 to 8) based on Anterior/Posterior (A/P) and Medial/Lateral (M/L) dimensions and are supplied with UHMWPE plugs to close off augment screw holes and distal taper when not used and with a set screw to provide additional locking to the stem or offset junction during extraction.

    Freedom Infinia™ Tibial Augments

    The Freedom Infinia™ Total Knee System's Tibial Augments are fabricated from Titanium-Aluminum-Vanadium (Ti-6A1-4V), complaint with ASTM F136. These augments are intended to be screwed to the distal surface Freedom Infinia™ Stemmed Tibial Base Plate when required in cases of significant bone loss. The augments are designed to be stackable, using a system of pins, bosses and screws, and reversible for use on either the medial or lateral side. Augment Screws, fabricated from Titanium-6Aluminum-4Vanadium (Ti-6A1-4V), complaint with ASTM F136, are available as an accessory to the system to fix the augments to the Stemmed Tibial Base Plate and other augments.

    Freedom Infinia™ Stem Extension

    The Freedom Infinia™ Total Knee System's Stem Extensions are fabricated from Titanium-Aluminum-Vanadium (Ti-6A1-4V), complaint with ASTM F136. The Freedom Infinia™ Stem Extensions are intended for use with Freedom Infinia™ Stemmed Tibial Base Plate. The Freedom Infinia™ Stem Extensions are available in a range of diameters and lengths and are fluted distally with a distal slot in the larger lengths and diameters.

    Freedom Infinia™ Offset Junction

    The Freedom Infinia™ Total Knee System's Offset Junctions are fabricated from Titanium-Aluminum-Vanadium (Ti-6A1-4V), complaint with ASTM F136, are intended to be used with the Freedom Infinia™ Stemmed Tibial Base Plate to provide an additional 4mm or 6mm offset between the tibial keel and stem. Freedom Infinia™ The Offset Junction is attached to the tibial component and stem through taper junctions. A set-screw is supplied with the Offset Junction to provide additional locking during extraction.

    Freedom Infinia™ Tibial Liner

    The Freedom Infinia™ Total Knee System's Tibial Liner is available in two designs i.e. posterior stabilizing (PS) and cruciate retaining (CR). Based on the Tibial Liner compatibility with Femoral Component configurations, each PS and CR designs are available various sizes with different anterior/posterior (A/P) and medial/lateral (L/P) dimensions, and each size is further bifurcated as on thickness. Tibial liner is fabricated from Ultra High Molecular Weight Polyethylene (UHMWPE) complying with ASTM F648 and is intended for cemented application along Tibial Base Plate to replace the articulating surface of the proximal tibia in a measured resection. Both the CR and PS Tibial Liners have identical locking mechanisms to mate with the Tibial Base Plate.

    Freedom Infinia™ Medial Congruent Liner

    The Freedom Infinia™ Total Knee System's Medial Congruent Liners have been developed with the desire to achieve a higher level of constraint than typically achieved in a traditional CR knee design while also providing the femoral component with a medial-pivoting motion throughout a range of motion. They possess an anterior and buildup and a higher sagittal conformity of the medial compartment that interact with the femoral component, to provide additional constraint. Designed to be used with the CR Femoral Component. Eliminates the need to remove bone for a PS box. Designed to allow high flexion. Includes a deep anterior patellar cut- out to allow for tendon clearance. Facilitates intraoperative flexibility, given compatibility with existing femoral and tibial baseplate options.

    Freedom Infinia™ Ultra-Congruent CR Tibial Liner

    The Freedom Infinia™ Total Knee System's Ultra-Congruent CR Tibial Liner consists of an ultra-high molecular weight polyethylene (UHMWPE) liner complies with ASTM F648 that is designed to be used with the Freedom Infinia™ Total Knee Cruciate Retaining (CR) Femoral Components in the Freedom Infinia™ Total Knee System.

    Freedom Infinia™ PCK Tibial Liner

    The Freedom Infinia™ Total Knee System's PCK Tibial Liner is fabricated from Ultra High Molecular Weight Polyethylene (UHMWPE) complies with ASTM F648. Based on the Tibial liner compatibility with PCK femoral component configurations, each PCK tibial liner design is available different sizes with different anterior/posterior (A/P) and medial/lateral (L/P) dimensions, and each size is further bifurcated on thickness variants.
    The PCK Tibial Liners consist of the same articulating surface of existing Tibial liner components and have a same locking mechanism on the peripheral edge of the distal surface to lock into the Tibial / Stemmed Tibial Base Plate. The PCK tibial liner supplied with a liner securing pin manufactured from Ti-6Al-4V ELI to provide additional locking with stemmed tibial base plate.

    Freedom Infinia™ All Poly Tibial Component

    The Freedom Infinia™ Total Knee System's All Poly Tibial Component is fabricated from ultra high molecular weight polyethylene (UHMWPE) complying with ASTM F648 and is intended for cemented application to replace the articulating surface of the proximal tibia in a measured resection. Both the Total Knee System CR and PS Tibial Inserts have identical locking mechanisms to mate with the Tibial Base Plate.

    Freedom Infinia™ All Poly Patella

    The Freedom Infinia™ Total Knee System's All Poly Patella is available in symmetrical single radius design. Based on different diameter and thickness different models are available. All Poly Patella is fabricated from Ultra High Molecular Weight Polyethylene (UHMWPE) complying with ASTM F648.
    The symmetric design features a central cement recess, and three pegs on the bone interface surface. The symmetric patellar component is intended for cemented application via an onset resurfacing surgical technique.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K253144

    Validate with FDA (Live)

    Date Cleared
    2025-10-22

    (27 days)

    Product Code
    Regulation Number
    888.3560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Freedom® Total Knee System is indicated for patients with severe knee pain and the disability due to:

    • Severe knee joint pain, loss of mobility, and disability due to: rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis.
    • Correction of functional deformities.
    • Post-traumatic loss of knee joint contour, particularly when there is patellofemoral erosion, dysfunction, or prior patellectomy.
    • Moderate valgus, varus, or flexion trauma.
    • Knee fractures untreatable by other methods.
    • Revision surgery where sufficient bone stock and soft tissue integrity are present. (For PCK Components and Primary PCK Components only)

    The Freedom® Total Knee System – Titan PCK Components are intended for cemented and single use only.

    Device Description

    The Freedom® Total Knee System is comprised of a femoral component, all-poly tibial component, patellar component, tibial base plate and tibial articular surface. The Freedom® Total Knee System's Femoral Component is offered as different versions such as stemmed PCK design, primary PCK, cruciate retaining, posterior stabilizing. The Freedom® Total Knee System's Tibial Base Plate is offered with stemmed design and without stemmed design. The Freedom® Total Knee System was originally cleared under the 510(k) number K082019. Later on, several modifications were made and were cleared under 510(k)s K091280, K192148, K090411, K182574, K131481, K111785, K200912, and K240863 respectively.

    This submission seeks the clearance of Titanium Niobium Nitride (TiNbN) coated version of previously cleared Femoral Augment (non-coated version cleared in K131481), Tibial Augments, Stem Extension & Offset Junction (non-coated versions cleared in K111785). The coated versions are now branded as Freedom® Total Knee System – Titan PCK Components.

    Below is the description of the coated components.

    Freedom® Titan Femoral Augments

    The Freedom® Titan Femoral Augments are fabricated from Titanium-6Aluminum-4Vanadium (Ti-6A1-4V), complaint with ASTM F136, and are coated with Titanium Niobium Nitride (TiNbN). These augments are intended to be screwed to the internal distal and posterior surfaces of the previously cleared PCK Stemmed Femoral Component (K131481) and Titan PCK Stemmed Femoral Component (K240863) when required in cases of significant bone loss. The augments are designed to be stackable and for use on either the medial or lateral side.

    Augment Screws, fabricated from Titanium-6Aluminum-4Vanadium (Ti-6A1-4V), complaint with ASTM F136, and are coated with Titanium Niobium Nitride (TiNbN), are available as an accessory to the system to fix the augments to the Stemmed Femoral Component and other augments.

    The Freedom® Titan Femoral Augments is available in distal and posterior design configuration. Each configuration is further available in 6 different sizes.

    Freedom® Titan Tibial Augments

    The Freedom® Titan Tibial Augments are fabricated from Titanium-Aluminum-Vanadium (Ti-6A1-4V), complaint with ASTM F136, and are coated with Titanium Niobium Nitride (TiNbN). These augments are intended to be screwed to the distal surface of the previously cleared Stemmed Tibial Base Plate (K111785) & Titan Stemmed Tibial Base Plate (K240863) when required in cases of significant bone loss. The augments are designed to be stackable, using a system of pins, bosses and screws, and reversible for use on either the medial or lateral side.

    Augment Screws, fabricated from Titanium-6Aluminum-4Vanadium (Ti-6A1-4V), complaint with ASTM F136, and are coated with Titanium Niobium Nitride (TiNbN), are available as an accessory to the system to fix the augments to the Stemmed Tibial Base Plate and other augments.

    Freedom® Titan Offset Junction

    The Freedom® Titan Offset Junctions are fabricated from Titanium-Aluminum-Vanadium (Ti-6A1-4V), complaint with ASTM F136, and are coated with Titanium Niobium Nitride (TiNbN). Offset Junctions are intended to be used with the previously cleared Stemmed Tibial Base Plate (K111785) & Titan Stemmed Tibial Base Plate (K240863) to provide an additional 4mm or 6mm offset between the tibial keel and stem. The Offset Junction is attached to the tibial component and stem through taper junctions. A set-screw is supplied with the Offset Junction to provide additional locking during extraction.

    Freedom® Titan Stem Extension

    The Freedom® Titan Stem Extensions are fabricated from Titanium-Aluminum-Vanadium (Ti-6A1-4V), complaint with ASTM F136, and are coated with Titanium Niobium Nitride (TiNbN). The Titan Stem Extensions are intended for use with the previously cleared Stemmed Tibial Base Plate (K111785) & Titan Stemmed Tibial Base Plate (K240863). The Stem Extensions are available in a range of diameters and lengths and are fluted distally with a distal slot in the larger lengths and diameters.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K251717

    Validate with FDA (Live)

    Date Cleared
    2025-06-26

    (22 days)

    Product Code
    Regulation Number
    888.3560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Freedom® Total Knee System is indicated for the following:

    • Severe knee joint pain, loss of mobility, and disability due to: rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis.
    • Correction of functional deformities.
    • Post-traumatic loss of knee joint contour, particularly when there is patellofemoral erosion, dysfunction, or prior patellectomy.
    • Moderate valgus, varus, or flexion trauma.
    • Knee fractures untreatable by other methods.
    • Revision surgery where sufficient bone stock and soft tissue integrity are present (For PCK Components and Primary PCK Components only).

    The Freedom® Total Knee System – Titanium Tibial Base Plate is intended for cemented and single use only.

    Device Description

    The Freedom® Total Knee System is comprised of a femoral component, all-poly tibial component, patellar component, tibial base plate and tibial articular surface. The Freedom® Total Knee System's Femoral Component is offered as different versions such as stemmed PCK design, primary PCK, cruciate retaining, posterior stabilizing. The Freedom® Total Knee System's Tibial Base Plate is offered with stemmed design and without stemmed design. The Freedom® Total Knee System was originally cleared under the 510(k) number K082019. Later on, several modifications were made and were cleared under 510(k)s K091280, K192148, K090411, K182574, K131481, K111785 and K200912 respectively.

    The primary purpose of this special 510(k) Device Modification to the Freedom® Metal Backed Tibial Component (K090411) is to notify the FDA of the change in materials used to manufacture the tibial base plate from CoCrMo to Wrought Titanium-6Aluminium-4Vanadium ELI Alloy (Ti6Al4V ELI, ASTM F136-13) as an alternative material option for the tibial base plate.

    The Titanium Tibial Base Plate is fabricated from Titanium alloy Ti-6Al-4V ELI, compliant with ASTM F136 and is intended for cemented application to replace the articulating surface of the proximal tibia in a measured resection technique.

    The Titanium Tibial Base Plates are available in 8 different sizes (1 to 8) based on Anterior/Posterior (A/P) and Medial/Lateral (M/L) dimensions.

    AI/ML Overview

    This 510(k) clearance letter pertains to a Class II medical device, specifically the Freedom® Total Knee System - Titanium Tibial Base Plate (K251717). The submission is a special 510(k) Device Modification, indicating that the changes made to the device are well-defined and do not significantly alter the indications for use or raise new questions of safety and effectiveness.

    The core change is the material of construction for the tibial base plate, shifting from CoCrMo to Wrought Titanium-6Aluminium-4Vanadium ELI Alloy (Ti-6Al-4V ELI). The design, geometry, surfaces, and locking features remain identical to the previously cleared predicate device (K090411).

    Therefore, the acceptance criteria and the study proving the device meets these criteria primarily focus on mechanical performance and biocompatibility related to this material change, rather than extensive clinical efficacy studies or comparative effectiveness studies involving human readers, as would be typical for AI/software devices.


    Acceptance Criteria and Reported Device Performance

    Given the nature of this submission (a material change for an orthopedic implant with identical design), the acceptance criteria are not explicitly numerical thresholds like sensitivity/specificity, but rather involve demonstrating mechanical equivalence to the predicate device and biocompatibility of the new material.

    Here's a table summarizing the implicit acceptance criteria based on standard regulatory requirements for such a device modification, and how the document indicates they were met:

    Acceptance Criteria CategorySpecific Criteria (Implicit from Submission Type)Reported Device Performance (as per 510(k) Summary)
    Mechanical PerformanceThe Titanium Tibial Base Plate must demonstrate equivalent or superior mechanical integrity and fatigue resistance compared to the predicate device to ensure durability and prevent failure under intended physiological loads.Met. "The design, geometry, surface features, locking features and dimensional attributes of the Titanium Tibial Base Plate are identical to those of the previously cleared Freedom® Metal Backed Tibial Base Plate (K090411).""Hence, to evaluate the device function and performance for its intended use, the Freedom® Titanium Tibial Base Plate was subjected to the following mechanical tests:Tibial Tray Fatigue Testing per ASTM F1800Range of Motion & Modular disassembly testing was leveraged from the reference device (K090411) as the subject device uses an identical tibial insert locking mechanism."
    Material BiocompatibilityThe new material (Ti-6Al-4V ELI) must be biocompatible and not elicit adverse biological responses (e.g., toxicity, sensitization, irritation) when implanted in the human body.Met. "Biocompatibility testing is not required for the subject Titanium Tibial Base Plate device, because the material, Ti-6Al-4V ELI Titanium Alloy is compliant with ASTM F136, is a well-established material with a long history of safe use in orthopedic implants.""This material has been used extensively for many years without any major biocompatibility related safety concerns.""Additionally, the material has been used in multiple components of Maxx Orthopedics' previously cleared Freedom® Total Knee System…""It also complies with the biocompatibility requirements outlined in 'Class II Special Controls Guidance Document: Knee Joint Patellofemorotibial and Femorotibial Metal/Polymer Porous-Coated Uncemented Prostheses; Guidance for Industry and FDA'"."Furthermore, Ti-6Al-4V ELI Titanium Alloy (ASTM F136) has also been used in similar tibial base plate components from other legally marketed devices, intended for same anatomical location and patient contact type, as demonstrated in K220737.""A risk-based assessment, following the principles outlined in FDA's 2023 Biocompatibility guidance (Use of ISO 10993-1), confirms that the material change does not introduce new types of patient contact, contact duration, or clinical use conditions that would require additional biocompatibility testing."
    Equivalence in Design/FunctionThe device design must be substantially equivalent to the predicate, with any minor differences not raising new safety or effectiveness concerns.Met. "The design, geometry, surface features, locking features and dimensional attributes of the Titanium Tibial Base Plate are identical to those of the previously cleared Freedom® Metal Backed Tibial Base Plate (K090411).""There are no significant technological differences between the subject and predicate device. The subject device uses a similar design and dimensions, geometry and sizing, and achieves its intended use in an identical manner as the primary predicate and both devices are manufactured using similar subtractive manufacturing techniques."
    Sterilization Method (Minor Change)If the sterilization method changes, its effectiveness must be validated.Addressed. "Minor differences in subject device are that it uses different materials of constructions and sterilization method that are addressed via performance testing and similarity to the secondary predicate devices (K241597)." While not explicitly detailed, the mention implies this was covered.

    Study Details (Relevant to Orthopedic Implants, Not AI)

    The provided text describes a submission for an orthopedic implant and does not involve an AI/software component, nor does it detail a clinical study with human patients for AI performance evaluation. Therefore, many of the requested points regarding sample sizes for test/training sets, expert readers, ground truth for AI, MRMC studies, etc., are not applicable to this type of device clearance and are consequently not found in the document.

    However, based on the information provided, here's what can be inferred/stated:

    1. Sample size used for the test set and data provenance:

      • Test Set: Not applicable in the context of clinical AI performance data. For mechanical testing (ASTM F1800), standard test methods specify the number of samples required (e.g., minimum of 6 samples for fatigue per ASTM F1800-97), but the exact number used in this specific submission is not reported in the summary.
      • Data Provenance: Not applicable for AI performance data. For material and design, the device is manufactured by Maxx Orthopedics Inc. (Norristown, PA, USA) and Meril Healthcare Pvt. Ltd. (Vapi, Gujarat, India).
    2. Number of experts used to establish the ground truth for the test set and qualifications of those experts:

      • Not applicable as this is a mechanical and material modification, not an AI diagnostic device requiring expert interpretation of images for ground truth.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • Not applicable for the type of device/study described.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No. This type of study (MRMC, AI assistance) is not relevant or performed for this device. The submission is for a physical orthopedic implant with a material change.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Not applicable, as there is no algorithm or AI component in this device.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • For the mechanical testing, the "ground truth" would be the engineering specifications and performance of the predicate device, alongside adherence to international standards (e.g., ASTM F1800).
      • For biocompatibility, the "ground truth" is established long-term clinical use data, compliance with ASTM F136 (Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI Alloy for Surgical Implant Applications), and prior FDA clearances of devices using this material.
    7. The sample size for the training set:

      • Not applicable (no AI training set).
    8. How the ground truth for the training set was established:

      • Not applicable (no AI training set).

    In summary, the FDA clearance for K251717 as a special 510(k) is based on demonstrating that the new Titanium Tibial Base Plate is mechanically equivalent to its predicate and that the new material is biocompatible and has a well-established history of safe use in similar orthopedic applications. The regulatory review focuses on engineering performance criteria and material biocompatibility rather than clinical performance data from patient studies or AI algorithm validation.

    Ask a Question

    Ask a specific question about this device

    K Number
    K240863

    Validate with FDA (Live)

    Date Cleared
    2024-07-02

    (96 days)

    Product Code
    Regulation Number
    888.3560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Freedom® Total Knee System is indicated for patients with severe knee pain and the disability due to:

    · Severe knee joint pain, loss of mobility, and disability due to: rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis.

    · Correction of functional deformities.

    • · Post-traumatic loss of knee joint contour, particularly when there is patellofemoral erosion,
      dysfunction, or prior patellectomy.

    • · Moderate valgus, varus, or flexion trauma.

    • · Knee fractures untreatable by other methods.

    · Revision surgery where sufficient bone stock and soft tissue integrity are present.

    The Freedom® Total Knee System is intended for cemented and for single use only.

    Device Description

    The Freedom® Total Knee System is comprised of a femoral component, all-poly tibial component, patellar component, tibial base plate and tibial articular surface. The Freedom® Total Knee System's Femoral Component is offered as different versions such as stemmed PCK design, primary PCK, cruciate retaining, posterior stabilizing. The Freedom® Total Knee System's Tibial Base Plate is offered with stemmed design and without stemmed design.

    This submission seeks the clearance of Titanium Nitride (TiNbN) coated version of previously cleared Stemmed PCK Femoral Component (non-coated version cleared in K131481) and Stemmed Tibial Base Plate (non-coated version cleated in K111785). The coated versions are now branded as Titan PCK Stemmed Femoral Component and Titan Stemmed Tibial Base Plate.

    Both these components are manufactured from CoCrMo alloy and are coated completely with Titanium Niobium Nitride (TiNbN).

    AI/ML Overview

    This document is a 510(k) premarket notification for a medical device, the Freedom® Total Knee System, and does not contain information about the development or testing of an AI/ML powered medical device. Therefore, I cannot extract the requested information regarding acceptance criteria or a study proving an AI/ML device meets those criteria.

    The document discusses the substantial equivalence of a new version of a total knee system (with a TiNbN coating) to previously cleared versions. The "performance testing" section refers to mechanical and biocompatibility tests for the physical knee implant components and their coating, not the performance of a software or algorithm. There is no mention of AI, machine learning, or any related technology.

    Ask a Question

    Ask a specific question about this device

    K Number
    K183684

    Validate with FDA (Live)

    Date Cleared
    2019-08-23

    (238 days)

    Product Code
    Regulation Number
    888.3353
    Age Range
    All
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Libertas™ Hip Replacement System is intended for use in total hip arthroplasty. Total hip arthroplasty is intended to provide patient mobility and reduce pain by replacing the damaged hip joint articulation in patients where there is evidence of sufficient sound bone to fix and support the components.

    Total hip replacement is indicated for the following conditions:

    • · Non-inflammatory degenerative joint diseases including osteoarthritis, post traumatic arthritis and avascular necrosis.
    • · Rheumatoid arthritis.
    • · Congenital hip dysplasia.
    • · Acute traumatic fracture of the femoral head or neck.
    • · Certain cases of Ankylosis.
    • Dislocation of the hip.
    • · Correction of functional deformity.
    • · Revision of failed joint reconstruction or treatment.
    • · Treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur.

    NOTE:

    • · The Modular Shell, Uncemented Stem and Taper uncemented stem are intended for press-fit, uncemented use only.
    • · The Cemented stem is intended for cemented use only.
    Device Description

    The Libertas™ - Taper uncemented femoral stem is fabricated from Ti6Al4V - ELI (Titanium-6Aluminum-4Vanadium Extra Low Interstitial) and it is intended to be used with Libertas Femoral heads, previously cleared under K180973. The proximal portion of stem is coated with commercially pure Titanium with plasma spray method. Libertas™ - Taper uncemented femoral stem is available in two designs (Standard i.e. full profile design and distally reduced design is available in different sizes with three neck angles (132° Standard, 132° Lateral, and 128° Standard).

    AI/ML Overview

    The provided text is a 510(k) summary for the Libertas™ - Taper Uncemented Femoral Stem, a medical device used in hip arthroplasty. This document focuses on demonstrating substantial equivalence to predicate devices through technical comparisons and non-clinical performance testing.

    It does not contain information about acceptance criteria or a study proving the device meets acceptance criteria in the context of an AI/ML medical device, which would typically involve clinical performance metrics like sensitivity, specificity, or reader studies. The document describes mechanical and material testing for an orthopedic implant.

    Therefore, I cannot provide the requested information about acceptance criteria and a study proving those criteria are met, as the provided text pertains to a different type of medical device (a mechanical implant) and a different regulatory pathway (510(k) for substantial equivalence based on physical and material properties, not AI/ML performance).

    To answer your prompt, imagine the prompt was for an AI/ML medical device submission. Here's how I would structure the answer if the provided text contained the relevant information for an AI/ML device:

    Hypothetical Answer (if the provided text were about an AI/ML medical device):

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Metric)Threshold/TargetReported Device PerformanceOutcome
    Standalone Performance
    Sensitivity for Condition X≥ 90%92.5%Met
    Specificity for Condition X≥ 85%88.2%Met
    AUC for Condition Y≥ 0.900.915Met
    Human-in-the-Loop Performance (Reader Study)
    Average Sensitivity (Reader + AI) vs. (Reader Only)Improvement ≥ 5%7.2% improvementMet
    False Positive Rate (Reader + AI) vs. (Reader Only)Reduction ≥ 10%12.1% reductionMet
    Inter-reader Agreement (Kappa) with AI assistanceIncrease of ≥ 0.1Increase of 0.15Met

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: [Number, e.g., 500 cases]
    • Data Provenance: [e.g., Retrospective cohort from hospitals in the USA and Europe]

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    • Number of Experts: [e.g., 3]
    • Qualifications: [e.g., Board-certified radiologists with 10-15 years of experience specializing in [relevant specialty, e.g., musculoskeletal imaging]]

    4. Adjudication Method for the Test Set

    • Method: [e.g., 2+1 adjudication, where two experts independently labeled cases, and a third, senior expert resolved any disagreements. Or, consensus reading by all three experts.]

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and its effect size.

    • Was an MRMC study done? [Yes/No]
    • Effect Size (if Yes): [e.g., Human readers showed an average improvement of 7.2% in sensitivity and a 12.1% reduction in false positives when assisted by the AI compared to reading without AI assistance. This translate to a statistically significant improvement in AUC of 0.05 (p < 0.001) for AI-assisted reading.]

    6. If a Standalone (algorithm only without human-in-the-loop performance) was done.

    • Was a standalone study done? [Yes/No]
    • Details (if Yes): [Metrics as per the table, e.g., The algorithm achieved a standalone sensitivity of 92.5%, specificity of 88.2%, and an AUC of 0.915 for detecting [condition].]

    7. The Type of Ground Truth Used

    • Type: [e.g., Expert consensus (multi-reader consensus), pathological confirmation (biopsy results), long-term clinical outcomes data, or a combination thereof.]

    8. The Sample Size for the Training Set

    • Training Set Sample Size: [Number, e.g., 10,000 cases]

    9. How the Ground Truth for the Training Set was Established

    • Method: [e.g., Ground truth for the training set was established through a combination of expert review (e.g., single expert reading with quality checks), existing hospital reports, and/or automated data extraction from electronic health records, followed by a subset verification by experienced staff.]