510(k) Data Aggregation

    K Number
    DEN200029
    Device Name
    Parallel
    Date Cleared
    2020-11-25

    (209 days)

    Product Code
    Regulation Number
    876.5960
    Type
    Direct
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Parallel is a prescription-only digital therapeutic device intended to provide cognitive behavioral therapy for adults aged 22 years of age and older who have been diagnosed with Irritable Bowel Syndrome (IBS). Parallel is indicated as a 3 month treatment for patients with IBS. Parallel treats IBS by reducing the severity of symptoms and is intended to be used together with other IBS treatments.

    Device Description

    Parallel is a web-based program designed to deliver Cognitive Behavioral Therapy (CBT) to patients with IBS who continue to have symptoms despite other forms of medical therapy. It is a responsive, web-based application that is intended to be a prescription device for use in the home for patients with IBS under the management of a qualified health care professional for the treatment of their IBS. Parallel is comprised of browser-delivered digital therapy that provides the CBT content. A desktop or laptop computer with a web browser and internet connectivity is required for use.

    CBT works by targeting problematic thoughts, feelings, and behaviors, building adaptive coping skills, and interrupting the cycle that is perpetuating the targeted symptoms. CBT can be delivered in-person by a mental health practitioner with adequate training in CBT. A critical component of treatment outcome is the degree to which CBT is administered competently, reliably, and as intended. Parallel treatment uses psychoeducation and teaching behavioral and cognitive skills and techniques to alter patterns of behavior and change unhelpful thoughts.

    Parallel is accessed via a secure website. Initially, the patient will receive a secure email containing an access link for the Parallel web application. Upon receiving the email invitation, the user can register electronically to begin their prescription. Once registered, patients are presented with onboarding material that introduces the program. Upon completion of reading the onboarding material, the patient can start the program.

    Parallel consists of eight CBT sessions, which include interactive components. The interactive components help patients remember the guidance and concepts, reflect, and engage in the therapeutic processes of CBT. In each session, patients review key points from the previous session and review their homework, thereby reinforcing previous learning. The patient must complete the sessions in sequential order. However, a patient may go back to a previously completed session. The prescription is for 90 days and patients should complete the eight sessions within that 90-day period. After the 90 days, patients can still access the sessions that they have completed.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study that proves the Parallel device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (from "Special Controls" and "Benefit-Risk Determination")Reported Device Performance
    1. Clinical Data
    a. Describe a model of therapy for the indicated gastrointestinal conditions.The device implements Cognitive Behavioral Therapy (CBT) for IBS, addressing problematic thoughts, feelings, and behaviors, and building adaptive coping skills. The eight sessions cover understanding IBS, symptom assessment, managing symptoms and eating, exercise and activity, identifying thought patterns, alternative thoughts, relaxation/sleep/stress management, and processing emotions/managing flare-ups.
    b. Validate the model of therapy as implemented by the device using a clinically defined endpoint.The ACTIB Pivotal Trial demonstrated a clinically meaningful benefit of >50 points in IBS-Symptom Severity Score (IBS-SSS) for the Parallel arm compared to the Treatment as Usual (TAU) arm at 3 months post-randomization (actual difference: 53 points).
    c. Evaluate all adverse events.Adverse events (AEs) were collected and analyzed. The proportion of individuals reporting at least one AE was similar across all treatment arms (TAU: 27.3%, telephone: 30.1%, Parallel: 26.5%). Most common SOCs were Psychiatric Disorders, Gastrointestinal Disorders, and Infections and Infestations. The Parallel arm showed slightly higher rates for abdominal pain, generalized pain, diverticulitis, and depression, but only one severe AE (diverticulitis) was reported in the Parallel group. Only one AE in the Parallel arm (abdominal pain) was considered "remotely" related to the device.
    2. Software Documentation
    Described in detail in SRS and SDS.The De Novo request provided appropriate software documentation consistent with a "Minor" level of software concern.
    Software verification, validation, and hazard analysis performed.Performed and documented. (Explicit mention of "hazard analysis" for software is within the "Delayed access to treatment due to device software failure" risk mitigation).
    Documentation demonstrates effective implementation of the behavioral therapy model.The software provides browser-delivered digital therapy that provides the CBT content. The eight sessions are described with their mechanisms of action and session targets. (Implied by the device description and the clinical success at 3 months).
    3. Usability Assessment
    Demonstrate that the intended user(s) can safely and correctly use the device.A usability assessment was conducted per FDA Guidance "Applying Human Factors and Usability Engineering to Medical Devices." It identified no critical tasks associated with the use of the Parallel software, thus a human factors study was not required.
    4. Labeling Requirements
    Instructions for use, including images.Provided.
    Patient and physician labeling listing minimum OS requirements.Provided.
    Patient and physician labeling including a warning that the device is not intended for use in lieu of standard therapeutic intervention or as a substitute for medication.Provided.
    Patient and physician labeling including a warning to seek medical care if having thoughts of harming self or others.Provided.
    Physician and patient labeling summarizing clinical testing.Provided, indicating clinical benefit at 3 months.

    Study Proving Device Meets Acceptance Criteria

    The study that proves the Parallel device meets the acceptance criteria is the Assessing Cognitive behavioral Therapy in Irritable Bowel (ACTIB) Pivotal Trial.

    Here's the detailed information regarding the study:

    2. Sample Size Used for the Test Set and Data Provenance:

    • Test Set Sample Size (Clinical Study): 558 patients were randomized in total:
      • Telephone arm: 186 patients
      • Parallel arm: 185 patients
      • TAU arm: 187 patients
      • Due to drop-out, complete cases at the end of the trial were:
        • Telephone arm: 136
        • Parallel arm: 124
        • TAU arm: 131
    • Data Provenance: The study was a multicenter clinical trial. While the exact country of origin is not explicitly stated, the reference for the study ("Everitt, H., et al., Therapist telephone-delivered CBT compared with treatment as usual in refractory irritable bowel syndrome: the ACTIB three-arm RCT. Health Technol Assess, 2019. 23(17): p. 1-154.") suggests it was likely conducted in the UK, as "Health Technology Assessment" is a journal associated with the UK's National Institute for Health Research (NIHR). The study design was prospective (Randomized Multicenter Clinical Trial).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:

    • The ground truth for the effectiveness of the therapy was based on the IBS-Symptom Severity Score (IBS-SSS), a validated patient-reported outcome measure. Therefore, there wasn't a separate panel of experts establishing ground truth for individual cases; rather, the "ground truth" of symptom severity was derived from patient self-reporting using a standardized, validated questionnaire.
    • The therapy in the "Telephone" arm was provided by "a therapist with training in CBT." No specific number or qualifications for these therapists (e.g., years of experience, specific certifications) are provided in the document. For the Parallel arm, the therapy is delivered by the software itself, with "three 30-minute telephone support calls" not explicitly stated to be therapeutic in nature but rather for ensuring adequate website resource use.

    4. Adjudication Method for the Test Set:

    • There was no explicit "adjudication method" described for the test set in the conventional sense of expert review of individual cases to determine a definitive outcome. The primary outcome measure (IBS-SSS) is a patient-reported questionnaire scored objectively.
    • The clinical significance of the IBS-SSS change was defined based on a previously published Minimum Clinically Important Difference (MCID) of 50 points, which was agreed upon by the FDA.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done:

    • No, a Multi Reader Multi Case (MRMC) comparative effectiveness study was not explicitly done in the context of comparing human readers' performance with and without AI assistance for tasks like image interpretation. This device is a therapeutic software, not a diagnostic imaging AI.
    • The study was a comparative effectiveness study, comparing the device (Parallel) to telephone-delivered CBT and Treatment As Usual (TAU).
      • Effect Size of Human Readers (Therapists) with vs. without AI assistance: Not applicable in this context, as the "human readers" were therapists delivering CBT (Telephone arm) or providing support (Parallel arm), not interpreting data in an "AI-assisted" diagnostic workflow.
      • However, if we broadly interpret "human readers" as "human therapists," the study does compare purely human-delivered therapy (Telephone CBT) to AI-assisted/delivered therapy (Parallel, which has minimal human support).
        • Telephone (Human-delivered CBT) vs. TAU:
          • 3 months: 69 points
          • 6 months: 58 points
          • 12 months: 62 points
        • Parallel (AI-delivered CBT with minimal human support) vs. TAU:
          • 3 months: 53 points (meets MCID of 50)
          • 6 months: 35.7 points (does not meet MCID)
          • 12 months: 35.5 points (does not meet MCID)
        • Comparison of Human-delivered CBT vs. AI-delivered CBT (Parallel):
          • At 3 months, human-delivered CBT yielded a 69-point improvement, while Parallel yielded a 53-point improvement. This indicates that the purely human-delivered CBT had a larger effect size at 3 months (difference of 16 points). The human-delivered CBT continued to show a significant effect at 6 and 12 months, whereas Parallel's effect diminished below the MCID after 3 months.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Yes, to a significant extent, the Parallel arm represents a standalone (or near-standalone) algorithm-only performance. Patients in the Parallel arm were asked to complete the website program "at home on their own time." While they received "three 30-minute telephone support calls," these were "intended to ensure that participants were adequately using the website resources, and were not equivalent to the therapy sessions provided to the Telephone arm." This suggests the primary therapeutic delivery was via the algorithm without significant human intervention for the CBT itself. The 53-point improvement at 3 months for the Parallel arm is the standalone (or near-standalone) performance.

    7. The Type of Ground Truth Used:

    • The ground truth for the clinical effectiveness outcome (reduction in IBS symptoms) was based on patient-reported outcomes (PROs) using a validated psychometric scale (IBS-SSS). The clinical significance was determined by an agreed-upon Minimum Clinically Important Difference (MCID).

    8. The Sample Size for the Training Set:

    • The document does not explicitly state a separate "training set" sample size for the Parallel device in the context of how the AI/CBT algorithm was initially developed or optimized.
    • The ACTIB trial focuses on the validation of the implemented device. CBT models are typically developed based on established psychological principles and clinical experience, not necessarily through a "training set" in the machine learning sense from a specific clinical trial.
    • The document mentions "access to the Parallel software in an earlier pilot trial" as an exclusion criterion, implying there might have been prior studies or development phases, but details regarding these are not provided.

    9. How the Ground Truth for the Training Set Was Established:

    • As a therapeutic device providing CBT, the "training set" and "ground truth" for developing the CBT content likely came from:
      • Established psychological and medical literature: The principles of CBT for IBS, its mechanisms of action, and session targets detailed in the device description are well-documented clinical practices.
      • Expert consensus/clinical experience: The design of the eight sessions and their content likely reflects the input of clinical experts in CBT and IBS.
      • Prior pilot studies (implied): The exclusion criterion regarding prior access to Parallel suggests there were earlier developmental or pilot phases where efficacy and user experience may have been iteratively improved, thereby establishing a de facto "training" or optimization process. However, specific details on how ground truth was established within these phases are not provided in this document.