Search Filters

Search Results

Found 6 results

510(k) Data Aggregation

    K Number
    K173290

    Validate with FDA (Live)

    Date Cleared
    2017-11-28

    (43 days)

    Product Code
    Regulation Number
    892.1000
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The 1.5T GE and Siemens DuoFLEX® Coil Suites and the 3.0T GE and Siemens DuoFLEX® Coil Suites are indicated for use on the order of a physician in conjunction with a 1.5T GE Healthcare and Siemens Healthcare Magnetic Resonance Scanner system and a 3.0T GE Healthcare and Siemens Healthcare Magnetic Resonance Scanner system to produce 2D and 3D images that when interpreted by a trained physician vield information that may assist in diagnosis.

    Device Description

    The MR Instruments FC 1520G-8R, FC 3000G-8R, FC 1500S-8R, and FC 3000S-8R Coil Suites are identical with the only exceptions:
    FC 1520G-8R (GE MR Systems @ 63.86MHz with GE System Connector Plug)
    FC 3000G-8R (GE MR Systems @ 127.73MHz with GE System Connector Plug)
    FC 1500S-8R (Siemens MR Systems @ 63.6MHz with Siemens System Connector Plug)
    FC 3000S-8R (Siemens MR Systems @ 123.23MHz with Siemens System Connector Plug)
    The coil suites include two different sets of coil paddles and a single, shared preamplifier box (the "Connector Box") with a system connector. One set contains a pair of coil paddles containing four 24cm loops/channels per paddle for a total of eight channels for the two paddles. The second set contains a pair of coil paddles containing four 10cm loops/channels per paddle for a total of eight channels for the two paddles. These paddles connect to the same connector box containing eight preamplifiers; only one pair can be connected at a time. The coil design for all the paddles has the same, simple antennae design and the same system connectivity configuration. The coils can be used in the following configurations:
    24cm by itself
    24cm with 24cm
    24cm with 10cm
    10cm with 10cm
    The aforementioned configurations provide 4 or 8 channels for imaging.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for the "DuoFLEX® Coil Suite," which is an MR diagnostic device (specialty magnetic resonance coil). This document does not describe an AI/ML-driven device or an acceptance study typical for such devices. Instead, it details the substantial equivalence of a new version of an MR coil to a previously cleared predicate device.

    Therefore, many of the requested criteria related to AI/ML model acceptance studies (like sample size for training/test sets, expert adjudication, MRMC studies, ground truth establishment for training, etc.) are not applicable to this document. The "study" described is a set of non-clinical bench tests and performance testing with phantoms and volunteer scans to demonstrate substantial equivalence, not a clinical study to prove diagnostic performance improvement by an AI.

    Here's an analysis of what is available in the document, mapped against the closest applicable aspects of your request, and explicitly stating what is not present:

    Device: DuoFLEX® Coil Suite (Magnetic Resonance Imaging Coil)
    Type of Study Conducted: Non-clinical bench testing, performance testing with phantoms, volunteer scans, and comparison to a predicate device to demonstrate substantial equivalence. This is not an AI/ML model performance study.


    1. Table of Acceptance Criteria and Reported Device Performance:

    The document describes the tests performed to demonstrate substantial equivalence rather than explicit acceptance criteria with numerical performance targets in a table format. The "acceptance" is based on the new device performing equivalently to the predicate and meeting safety standards.

    Acceptance Criteria (Implied from the "Non-Clinical Testing" section)Reported Device Performance / Outcome
    EMC and electrical safety testingTests conducted; presumed to have met standards for safety.
    System safety testingTests conducted; presumed to have met standards for safety.
    Performance testing with phantomsTests conducted; outcomes used to support substantial equivalence.
    Predicate device comparison testsTests conducted; results support substantial equivalence.
    Volunteer scansScans conducted; used to support substantial equivalence.
    IEC 60601-1 Compliance (various tests)Tests conducted; presumed to have met standards for medical electrical equipment safety.
    Biocompatibility for patient contact materialsTests conducted; presumed to have met standards.
    NEMA MS-1 Signal to Noise Ratio (SNR)Test conducted; used for performance evaluation against predicate.
    Image Uniformity ComparisonTest conducted; used for performance evaluation against predicate.
    Evaluation of Sample Clinical Images (Clinical Evaluation Testing)Test conducted; used to support substantial equivalence.
    Design Validation (in actual and simulated use settings)Performed; results support substantial equivalence and safety.

    2. Sample size used for the test set and the data provenance:

    • Test Set (for performance evaluation):
      • Phantoms: Not specified, but phantom testing was performed.
      • Volunteer Scans: "Volunteer scans" were conducted. The number of volunteers is not specified.
      • Clinical Images: "Sample Clinical Images" were evaluated. The number of images/patients is not specified.
    • Data Provenance: Not explicitly stated for volunteers or clinical images, but it's implied these were internal studies conducted by MR Instruments, Inc. No mention of country of origin or whether prospective/retrospective. Given it's a device clearance, the data would typically be gathered for the purpose of the submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable in the context of an AI/ML ground truth study. The evaluation involved "Evaluation of Sample Clinical Images (Clinical Evaluation Testing)" and comparison to the predicate. The document doesn't detail how "ground truth" (in the sense of a diagnostic label) was established, as the device is a coil for producing images, not interpreting them. The assessment would likely be on image quality, signal consistency, safety, etc., compared to the predicate, rather than diagnostic accuracy.

    4. Adjudication method for the test set:

    • Not applicable. No mention of expert adjudication as this is not a diagnostic AI performance study.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No MRMC study was done. This is not an AI-assisted diagnostic device. The study focused on demonstrating substantial equivalence of the new MR coil to an existing predicate coil. The document explicitly states: "This technology is not new; therefore, a clinical study was not considered necessary prior to release. Additionally, there was no clinical testing required to support the medical devices as the indications for use is equivalent to the predicate device. The substantial equivalence of the devices is supported by the non-clinical testing."

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Partially applicable, but not for an AI algorithm. The "standalone" performance here refers to the performance of the MR coil itself (e.g., SNR, image uniformity, safety) as measured through bench tests, phantoms, and volunteer scans, independent of human interpretation for diagnostic purposes. These non-clinical tests demonstrate the coil's intrinsic characteristics.

    7. The type of ground truth used:

    • Not explicitly defined as "ground truth" in the diagnostic sense. For this device (an MR coil), "ground truth" implicitly relates to physical properties (e.g., magnetic field homogeneity, signal generation), safety parameters, and image quality characteristics (SNR, uniformity) as measured by calibrated equipment and compared against known standards or the predicate device's performance. The "volunteer scans" and "sample clinical images" would be assessed for image quality and usability for diagnostic purposes, not for confirming a specific diagnosis.

    8. The sample size for the training set:

    • Not applicable. This document describes an MR coil, not an AI/ML model that requires a training set.

    9. How the ground truth for the training set was established:

    • Not applicable. As above, this is not an AI/ML model.
    Ask a Question

    Ask a specific question about this device

    K Number
    K130706

    Validate with FDA (Live)

    Date Cleared
    2013-07-25

    (132 days)

    Product Code
    Regulation Number
    892.1000
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DuoFLEX Coil Suite is indicate for use on the order of the physician in conjunction with a 1.5T MRI scanner, as an accessory to product 2D and 3D images. The DuoFLEX Coil Suite is for use with GE Healthcare MR Systems.

    Device Description

    The MR Instruments FC1500G-8R DuoFLEX® Coil Suite includes three different sets of coil paddles and a single, shared preamplifier box (the "Connector Box") with a system connector. One set of coil paddles is a pair of 24cm square containing four loops/channels per paddle for a total of eight channels for the two paddles. The second set of paddles is a pair of 10cm square containing four loops/channels per paddle for a total of eight channels for the two paddles. The third set of paddles is a single rectangular, interventional coil containing a single planar loop. All of these paddles connect to the same connector box containing eight preamplifiers; only one pair can be connected at a time. The coil design for all of the paddles has the same, simple antennae design and the same system connectivity configuration. The coils can be used in several configurations, including any single coil and any two coils, providing 1, 5 or 8 channels for imaging.

    AI/ML Overview

    The DuoFLEX® Coil Suite (1.5T) 1500G-8R is a specialty magnetic resonance coil used in conjunction with a 1.5T MRI scanner to produce 2D and 3D images. The device is intended for use with GE Healthcare MR Systems and is classified as a Class II medical device (21 CFR 892.1000).

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Bench Testing)Reported Device Performance (Summary)
    EMC and electrical safetyTesting conducted. Results support substantial equivalence.
    Electrical and mechanical safetyTesting conducted. Results support substantial equivalence.
    System safetyTesting conducted. Results support substantial equivalence.
    Performance testing with phantomsTesting conducted. Results support substantial equivalence.
    Predicate device comparison testsTesting conducted. Results support substantial equivalence.
    Volunteer scansTesting conducted. Results support substantial equivalence.
    IEC 60601-1 standardsAll specified tests conducted (humidity preconditioning, accessible parts, markings legibility/durability, dielectric strength, ball pressure, creepage distances, surfaces/corners/edges, instability in transport/excluding transport, cleaning/disinfection, mold stress relief, impact test, push test, drop test). Results support substantial equivalence.
    BiocompatibilityTesting conducted for patient contact materials. Results support substantial equivalence.
    NEMA MS-1 Signal to Noise RatioTesting conducted. Results support substantial equivalence.
    Image Uniformity ComparisonTesting conducted. Results support substantial equivalence.
    Clinical comparison for image qualityTesting conducted. Results support substantial equivalence.
    Phantom (cadaver) simulated interventional useTesting conducted for interventional usage. Results support substantial equivalence.
    FMEA/Hazard AnalysisConducted for mechanical and RF designs.
    Performance Requirements TestingIncluding Final Bench Testing, ISO 60601 Testing, Surface Temperature Testing, SNR per NEMA MS-1 and MS-6, and Image Uniformity.

    2. Sample Size Used for the Test Set and Data Provenance

    The documentation provided does not specify a distinct "test set" in the context of a clinical study or a separate dataset for evaluating an AI algorithm. The device is a hardware component (MRI coil), and its performance was primarily evaluated through non-clinical testing including bench tests, phantom studies, and volunteer scans. The exact number of volunteers or phantoms used for these tests is not specified, nor is the country of origin of the data. Given the nature of a 510(k) for a device like an MRI coil, the focus is on engineering and performance validation rather than large-scale clinical trials.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This information is not provided. As per the submission, "This technology is not new, therefore a clinical study was not considered necessary prior to release. Additionally, there was no clinical testing required to support the medical device as the indications for use is equivalent to the predicate device." Therefore, there was no "test set" in the sense of a clinical image dataset requiring expert-established ground truth for performance evaluation against a diagnostic standard. The "clinical comparison for image quality" described in the non-clinical testing refers to an evaluation of the images produced by the coil, not a diagnostic accuracy study.

    4. Adjudication Method for the Test Set

    Not applicable. As explained in point 3, there was no clinical study with a test set requiring adjudication in the context of diagnostic accuracy.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study was done. The submission explicitly states that "This technology is not new, therefore a clinical study was not considered necessary prior to release." The device is an accessory to an MRI scanner, and its equivalence was demonstrated through non-clinical performance and safety testing compared to a predicate device.

    6. Standalone Performance Study (Algorithm Only Without Human-in-the-Loop Performance)

    Not applicable. This device is an MRI coil, which is a hardware component, not an AI algorithm. Therefore, "standalone" performance in the context of an algorithm is not relevant. The performance evaluation focuses on the engineering and image acquisition capabilities of the coil.

    7. Type of Ground Truth Used

    For the non-clinical testing, the "ground truth" was established through:

    • Engineering specifications and regulatory standards: for electrical safety, EMC, mechanical properties, and compliance with IEC 60601-1.
    • Physical measurements: for SNR (NEMA MS-1 and MS-6) and image uniformity.
    • Visual assessment: for image quality (clinical comparison for image quality, volunteer scans, and phantom/cadaver studies), likely made by engineers or experienced MRI operators, comparing the new device's output to established benchmarks or the predicate device's output.

    8. Sample Size for the Training Set

    Not applicable. This device is a hardware component and does not involve AI algorithms that require a "training set" of data.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for an AI algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K110591

    Validate with FDA (Live)

    Date Cleared
    2011-03-29

    (27 days)

    Product Code
    Regulation Number
    892.1000
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This coil is indicated for use in conjunction with 1.5T and 3.0T GE Healthcare Magnetic Resonance Scanner systems to produce images and/or spectra of the head, that when interpreted by a trained physician yield information that may assist in diagnosis.

    Device Description

    The MC 1504G-16R (1.5T) and MC 3004G-16R (3.0T) Head Coils are multi-channel receive only coils consisting of 16 RF elements. Each RF element has active and passive blocking networks, and integrated preamplifiers. The MC 1504G-16R RF elements are tuned to 'H at 1.5T (63.9 MHz), while the MC 3004G-16R RF elements are tuned to 'H at 3.0T (127.8 MHz). The coil housing is constructed out of a rigid plastic with an opening for patient viewing and comfort; the head support/base is included with each coil assembly.

    AI/ML Overview

    Here's an analysis of the provided text to extract the requested information about the device's acceptance criteria and testing:

    This document primarily describes the substantial equivalence of a new head coil (MC 1504G-16R and MC 3004G-16R) to existing predicate devices. The focus is on demonstrating that the new device performs similarly to or better than the predicates, rather than establishing acceptance criteria against a specific, quantitative performance threshold for a diagnostic outcome. Therefore, the information provided isn't in the typical format of acceptance criteria for an AI/CADe device where metrics like sensitivity, specificity, or AUC are assessed against predefined targets.

    However, based on the text, we can infer the performance comparisons made.

    1. Table of Acceptance Criteria and the Reported Device Performance

    Parameter/CriteriaAcceptance Criteria (Inferred from Predicate Equivalence)Reported Device Performance
    Electrical SafetyCompliance with IEC 60601-1Testing conducted in accordance with IEC 60601-1; compliance implied.
    Mechanical SafetyCompliance with IEC 60601-1Testing conducted in accordance with IEC 60601-1; compliance implied.
    Functional Performance (Electrical Bench)Proper assembly, function, and tuning to established specifications (unspecified)Electrical bench measurements performed to ensure proper assembly, function, and tuning to established specifications.
    Performance (Phantom - Signal-to-Noise Ratio (SNR))Equivalent or improved SNR compared to predicate deviceComparative SNR measurements based on NEMA MS-1 performed. (Implied: results were favorable for substantial equivalence). "The use of additional elements may increase coil sensitivity and reduce data acquisition time."
    Image Quality (Volunteer Scans)Equivalence in resulting image quality to predicate deviceVolunteer scans performed; images reviewed in blind fashion by registered technologists to demonstrate equivalence in resulting image quality.
    System CompatibilityCompatible with 1.5T and 3.0T Magnetic Resonance Scanner systemsAll system testing conducted using 1.5T and 3.0T MR Scanner systems running standard clinical applications.

    2. Sample Size Used for the Test Set and the Data Provenance

    • Sample Size (Test Set): Not explicitly stated numbers.
      • Volunteer Scans: "Volunteer scans were performed." The number of volunteers is not specified.
      • Phantoms: "Performance evaluations (phantoms) were completed." The number of phantoms is not specified.
    • Data Provenance: Not specified, but generally, such studies for medical device clearance are conducted internally or at partnering institutions within the country of the manufacturer. The study is prospective in nature, involving new scans/tests specifically for this evaluation.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    • Number of Experts: Not explicitly stated, but "registered technologists in magnetic resonance" were used. This implies more than one.
    • Qualifications of Experts: "Registered technologists in magnetic resonance." No specific experience level (e.g., years of experience) is provided. Their role was to assess "equivalence in resulting image quality," meaning they were evaluating the technical quality of the images rather than making clinical diagnoses.

    4. Adjudication Method for the Test Set

    • Adjudication Method: "Scan images were then reviewed in blind fashion by registered technologists in magnetic resonance." No specific adjudication method (e.g., 2+1, 3+1) is mentioned beyond the blind review. This suggests a qualitative assessment rather than a structured consensus process.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    • No MRMC study was done. This device is a hardware component (MRI head coil), not an AI/CADe system designed to assist human readers. Therefore, there's no mention of human reader improvement with AI assistance.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

    • Not applicable. This is a hardware device (MRI coil), not an algorithm or AI system. Its performance is evaluated through physical and imaging characteristics.

    7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)

    • For image quality evaluation, the "ground truth" was essentially the subjective assessment of "equivalence in resulting image quality" by "registered technologists in magnetic resonance" when comparing images from the proposed device to images from predicate devices. This can be considered a comparative expert assessment of image quality.
    • For physical performance, the "ground truth" was established by engineering specifications and industry standards (e.g., IEC 60601-1 for safety, NEMA MS-1 for SNR measurements).

    8. The Sample Size for the Training Set

    • Not applicable / No specific training set. As a physical hardware device, there isn't a "training set" in the machine learning sense. The device design and specifications are based on engineering principles and knowledge, not data-driven training.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable. See point 8.
    Ask a Question

    Ask a specific question about this device

    K Number
    K080820

    Validate with FDA (Live)

    Date Cleared
    2008-04-02

    (9 days)

    Product Code
    Regulation Number
    892.1000
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MR Instruments TEM 3002G-HN Dual Tune headcoil is to be used in conjunction with a 3T Magnetic Resonance Imaging system to produce proton ('H) images and spectra of the internal structures of the head. In addition, sodium (20Na) images and spectra may be obtained during the same exam without changing the coil. These images and/or spectral information of the head, when interpreted by a trained physician, may yield information that may assist in diagnosis. This device is for prescription use.

    Device Description

    The TEM 3002G-HN Dual Tune headcoil is a quadrature transmit/receive RF coil tuned to proton ('H) and sodium (22Na) frequencies 3T. The TEM resonant elements and associated matching and quadrature splitting circuitry are enclosed in a rigid plastic housing to prevent any exposure to patient or environment.

    AI/ML Overview

    The provided text describes the TEM 3002G-HN Dual Tune Headcoil, which is a medical device for Magnetic Resonance Imaging (MRI). However, it is a 510(k) summary for a physical medical device (an MRI headcoil), not an AI/ML powered device. As such, it does not include information on acceptance criteria, a study proving the device meets acceptance criteria, sample sizes for test sets, data provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance, type of ground truth, or training set sample sizes, which are all typically associated with the evaluation of AI/ML diagnostic tools.

    Therefore, the prompt cannot be answered with the provided text because it describes a hardware device, not an AI/ML powered diagnostic tool.

    Ask a Question

    Ask a specific question about this device

    K Number
    K072931

    Validate with FDA (Live)

    Date Cleared
    2007-10-30

    (14 days)

    Product Code
    Regulation Number
    892.1000
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MR Instruments MC 3003G-32R Head Coil is to be used in conjunction with a Magnetic Resonance Scanner to produce images and/or spectra of the head, that when interpreted by a trained physician yield information that may assist in diagnosis.

    Device Description

    The MC 3003G-32R is a multi-channel receive only RF head coil. It has 32 RF elements that are tuned to 'H at 3T (127.8 MHz). Each element has active and passive blocking networks and integrated preamplifiers. The housing is constructed out of a rigid plastic with an opening for patient viewing and comfort. A head support/base is included.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device, the MC 3003G-32R Head Coil. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving efficacy through clinical studies with specific acceptance criteria as might be seen for novel devices.

    Therefore, the document does not contain information about specific "acceptance criteria" for diagnostic performance (e.g., sensitivity, specificity, or image quality metrics with thresholds) or a "study that proves the device meets the acceptance criteria" in the way one might expect for a new diagnostic algorithm or AI product.

    Instead, the submission for the MC 3003G-32R Head Coil demonstrates substantial equivalence primarily by comparing its design, intended use, and performance to a predicate device, focusing on safety and basic functional performance.

    Here's a breakdown of the information that is available in the document, acknowledging that it doesn't directly address the structured questions about acceptance criteria for diagnostic performance:


    1. A table of acceptance criteria and the reported device performance

    • No specific diagnostic performance acceptance criteria (e.g., sensitivity, specificity, accuracy thresholds) are provided in the document.
    • The document states that "Design verification and validation testing was performed to ensure that the coil design specifications and customer requirements were met." However, the specific metrics and thresholds for these design specifications and customer requirements are not detailed.
    • Reported Device Performance (as implied by the "Summary of Studies"):
      • Electrical bench measurements
      • Electrical/mechanical safety tests
      • System safety and performance tests with phantoms
      • Predicate device comparison tests
      • Volunteer scans
    • The primary functional difference highlighted is that the MC 3003G-32R Head Coil has a "32 element phased array RF structure as opposed to a 8 element phased array RF structure" of the predicate, which "may increase coil sensitivity and may reduce data acquisition time." This implies improved performance but doesn't quantify it against specific acceptance criteria.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • The document mentions "volunteer scans" as part of the testing activities, but does not specify the sample size of these volunteers.
    • Data provenance is not specified. The studies were "conducted using the GE Healthcare Signa 3.0T MR System running standard clinical applications."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable / Not provided. The "volunteer scans" are mentioned, and the intended use states that "when interpreted by a trained physician yield information that may assist in diagnosis." However, the document does not describe a study involving expert interpretation for establishing ground truth or performance metrics against such ground truth for the device itself. The focus is on safety and technical performance of the coil.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable / Not provided. Since there's no mention of a diagnostic performance study with expert interpretation for ground truth, there's no adjudication method described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No. This device is an MR head coil, a hardware component for image acquisition, not an Artificial Intelligence (AI) diagnostic tool. Therefore, an MRMC study comparing human readers with and without AI assistance is not relevant or described.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This device is hardware; it is not an algorithm, so a standalone algorithm-only performance study would not apply.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • Not explicitly described for diagnostic performance. For the "system safety and performance tests with phantoms," the ground truth would be the known properties and geometry of the phantoms. For "electrical bench measurements" and "electrical/mechanical safety tests," the ground truth would be engineering specifications and safety standards. For "volunteer scans," the purpose seems to be confirming image acquisition capability and safety, rather than a diagnostic performance study against a definitive clinical ground truth.

    8. The sample size for the training set

    • Not applicable. This device is hardware and does not involve a training set as would an AI algorithm.

    9. How the ground truth for the training set was established

    • Not applicable. As above, no training set for an AI algorithm is involved.

    Summary of the Study Mentioned (from "Summary of Studies"):

    The study described is a design verification and validation process for a hardware device (an MR head coil). It included:

    • Electrical bench measurements: Likely verifying electrical specifications.
    • Electrical/mechanical safety tests: Ensuring the device meets safety standards.
    • System safety and performance tests with phantoms: Using standardized objects with known properties to assess image quality, signal-to-noise ratio, uniformity, etc., under controlled conditions. This would be a technical performance evaluation, not a clinical diagnostic one.
    • Predicate device comparison tests: Direct comparison of the new coil's performance (likely technical and imaging characteristics) against the predicate.
    • Volunteer scans: Scans on human volunteers to confirm safe operation and the ability to acquire images, consistent with its intended use for generating images of the head.

    The study's goal was to demonstrate that the MC 3003G-32R Head Coil is "substantially equivalent" to a legally marketed predicate device in terms of safety and effectiveness for its intended use, rather than to meet specific diagnostic performance metrics through a clinical trial.

    Ask a Question

    Ask a specific question about this device

    K Number
    K040937

    Validate with FDA (Live)

    Date Cleared
    2004-04-23

    (11 days)

    Product Code
    Regulation Number
    892.1000
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TEM 3000 head coil is designed to provide Magnetic Resonance Images of the brain, soft tissues and vasculature of the head. The TEM 3000 head coil is compatible with the 3T Magnetom Trio MR scanner manufactured by Siemens Medical Systems.

    Device Description

    The TEM 3000 Head Coil is a 15-element quadrature transmit/receive coil. The coil elements and associated circuitry are enclosed in a rigid housing to prevent any exposure to patient or environment. The coil housing employs a large open viewing window on the top. The coil design facilitates the scanning of patients with different window on the top: "Ifizes patient comfort and ease of use. Included with the coil housing is a Head Support/Coil Base. It allows the coil to be slid back for better patient access.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the TEM 3000 Head Coil and does not contain information about acceptance criteria or a study proving the device meets them. The document is from 2004 and is a regulatory submission for a medical device rather than a scientific study report.

    Therefore, most of the requested information cannot be extracted from this document, specifically:

    • 1. A table of acceptance criteria and the reported device performance: Not available. The document is a regulatory filing, not a performance study report.
    • 2. Sample sized used for the test set and the data provenance: Not available. No "test set" in the context of a performance study is mentioned.
    • 3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not available.
    • 4. Adjudication method: Not available.
    • 5. If a multi reader multi case (MRMC) comparative effectiveness study was done: Not indicated.
    • 6. If a standalone was done: Not indicated.
    • 7. The type of ground truth used: Not available.
    • 8. The sample size for the training set: Not available.
    • 9. How the ground truth for the training set was established: Not available.

    What the document does indicate:

    The document describes the TEM 3000 Head Coil as a 15-element quadrature transmit/receive coil intended for 'H high-resolution whole-head adult and pediatric imaging, including functional MRI, spectroscopy, and angiography. It is compatible with the Siemens 3T Magnetom Trio MR System.

    The submission focuses on establishing substantial equivalence to a predicate device, the Magnetom Trio Head Coil (K021330), based on similar construction and operation, despite using TEM technology instead of "Birdcage" technology. Substantial equivalence in this context means the new device is as safe and effective as a legally marketed device. This is a regulatory pathway, not a clinical performance study with acceptance criteria as typically understood in a scientific evaluation.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1