Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K112203

    Validate with FDA (Live)

    Date Cleared
    2011-12-14

    (135 days)

    Product Code
    Regulation Number
    870.2850
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Compass™ GP disposable pressure transducer with integrated digital display is intended for direct measurement of physiological pressure.

    The Compass™ Thoracentesis disposable pressure transducer with integrated digital display is intended for direct measurement of physiological pressure.

    The Compass™ Paracentesis disposable pressure transducer with integrated digital display is intended for direct measurement of physiological pressure.

    The Compass™ Compartment Pressure disposable pressure transducer with integrated digital display is intended for direct measurement of physiological pressure.

    The Compass™ Epidural Assist disposable pressure transducer with integrated digital display is intended for direct measurement of physiological pressure.

    The Compass™ Arterial Assist disposable pressure transducer with integrated digital display is intended for direct measurement of physiological pressure.

    Device Description

    The Compass devices are disposable, point-of-use pressure measurement and monitoring devices that incorporate a pressure transducer and an integrated pre-programmed diagnostic computer with liquid crystal display (LCD). The Compass devices are designed to attach distally to an inserted needle or catheter, measure the pressure via an embedded pressure sensor, internally convert changes in pressure into electrical currents, and then display the resulting pressure via the integrated LCD.

    AI/ML Overview
    1. Table of Acceptance Criteria and Reported Device Performance:

      The document does not explicitly present a table of acceptance criteria with corresponding performance results. However, it indicates that "Pressure accuracy testing was completed per ANSVAAMI BP22:1994(R)2006." The study concludes that "The results from this in vitro testing demonstrate that the technological and performance characteristics of the Compass devices meet defined design requirements and that they can perform in a manner equivalent to devices currently on the market used for measuring physiological pressure." This implies that the device met the accuracy requirements specified in the ANSVAAMI BP22:1994(R)2006 standard, which served as the acceptance criteria for pressure accuracy. Specific numerical values for the acceptance criteria and the device's performance against them are not provided in this summary.

    2. Sample Size for Test Set and Data Provenance:

      The document describes "in vitro testing" without specifying a particular sample size for a test set in the context of a dataset of medical images or patient records. The testing performed focused on physical device performance, specifically pressure accuracy, rather than evaluating an algorithm on a clinical dataset. The data provenance is related to the physical testing of the device.

    3. Number of Experts and Qualifications for Ground Truth:

      Not applicable. The study is a device performance test (pressure accuracy) and does not involve establishing ground truth by medical experts for a diagnostic or AI-driven task.

    4. Adjudication Method for Test Set:

      Not applicable. The study is a device performance test and does not involve adjudication of expert interpretations for a test set.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

      No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The device is a disposable pressure transducer with an integrated digital display, not an AI system designed to assist human readers in interpreting medical images or data.

    6. Standalone (Algorithm Only) Performance Study:

      The device itself (Compass™) is the standalone device. The testing described is of the device's performance, specifically its pressure accuracy and software functionality. It's not an algorithm in the sense of a software-only diagnostic tool that would typically undergo a standalone performance study in the context of AI. The performance of the integrated pre-programmed diagnostic computer is assessed as part of the overall device.

    7. Type of Ground Truth Used:

      For the pressure accuracy testing, the "ground truth" would have been established by a reference standard pressure measurement device or system, as per the ANSVAAMI BP22:1994(R)2006 standard. This is a technical (metrological) ground truth rather than clinical ground truth (e.g., pathology, outcomes data, or expert consensus on a diagnosis).

    8. Sample Size for Training Set:

      Not applicable. The device is a physical pressure transducer with integrated software, not an AI system requiring a training set in the machine learning sense. The software is described as "integrated pre-programmed diagnostic computer" and "new software," suggesting it was developed through traditional software engineering processes, not by training on a dataset.

    9. How Ground Truth for Training Set Was Established:

      Not applicable, as there is no "training set" in the context of this device's development. The software was likely developed and verified against functional and performance requirements through established software development lifecycles and testing protocols.

    Ask a Question

    Ask a specific question about this device

    K Number
    K103715

    Validate with FDA (Live)

    Date Cleared
    2011-01-20

    (30 days)

    Product Code
    Regulation Number
    870.2850
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K101518

    Validate with FDA (Live)

    Date Cleared
    2010-10-13

    (133 days)

    Product Code
    Regulation Number
    870.2850
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Compass™ Vascular Access is a disposable pressure transducer with integrated digital display intended for direct measurement and monitoring of invasive blood pressure.
    The Compass™ Lumbar Puncture is a disposable pressure transducer with integrated digital display intended for direct measurement and monitoring of intracranial pressure.

    Device Description

    The Compass is a disposable, point-of-use pressure measurement and monitoring device that incorporates a pressure transducer and an integrated pre-programmed diagnostic computer with liquid crystal display (LCD). The Compass is provided as two different models, each corresponding to a specific indication for use:

    • (1) The Compass Vascular Access (VA) is for invasive blood pressure measurement and monitoring during the insertion of vascular access devices.
    • (2) The Compass Lumbar Puncture (LP) is for intracranial pressure measurement and monitoring via lumbar puncture or intraventricular catheter.
      Both models are designed to attach distally to an inserted needle or catheter, measure the pressure via an embedded pressure sensor, internally convert changes in pressure into electrical currents, and then display the resulting pressure via the integrated LCD.
      The Compass VA and Compass LP devices are identical except for two differences: 1) the measurement units in which the pressure is displayed on the LCD (mm Hg for blood pressure measurements with the Compass VA and cm H2O for intracranial pressure measurements with the Compass LP), and 2) the Compass VA blood pressure monitoring device has a proximal port through which commercially available guidewires can be inserted during pressure measurement.
    AI/ML Overview

    The provided 510(k) summary for the Mirador Compass™ device (K101518) describes the device, its indications for use, and the testing performed. However, it does not provide a specific table of acceptance criteria with corresponding reported device performance values in a side-by-side format as typically found in comprehensive performance studies for AI/CAD devices.

    Instead, the documentation outlines various tests conducted and generally states that the device "meet defined design requirements" and can "perform in a manner equivalent to devices currently on the market."

    Let's break down the information based on your requested points:


    Acceptance Criteria and Reported Device Performance

    The document describes several categories of testing, implying that the acceptance criteria are met if the device performs "per Mirador Biomedical's in-house test methods, protocols, and requirements or in accordance with applicable recognized standards by accredited outside laboratories," and that the results "demonstrate that the technological and performance characteristics... meet defined design requirements and that they can perform in a manner equivalent to devices currently on the market."

    Due to the nature of the provided document, a specific table of acceptance criteria and reported device performance with numerical values cannot be fully constructed as the exact numerical acceptance limits and reported performance values are not explicitly stated for all tests. However, we can infer the categories of performance that were assessed.

    Acceptance Criteria Category (Inferred)Reported Device Performance (as stated in document)
    BiocompatibilityMet ISO 10993-1 standards (cytotoxicity, hemolysis, acute systemic toxicity, intracutaneous reactivity, sensitization, pyrogenicity).
    Functional Testing (General)Met applicable functional testing per ANSI/AAMI BP22:1994(R)2006, ISO 594-1, and ISO 594-2.
    Luer FittingsMet defined design requirements.
    Mechanical Strength and DurabilityMet defined design requirements.
    Power SwitchMet defined design requirements.
    Dimensional and Weight VerificationMet defined design requirements.
    Battery LifeMet defined design requirements.
    Pressure Overrange CapabilityMet defined design requirements.
    Pressure AccuracyMet defined design requirements. (In vivo study showed accurate measurement equivalent to predicate)
    Drift, Temperature Error Band, Temperature Error Band of SensitivityMet defined design requirements.
    Light SensitivityMet defined design requirements.
    Frequency ResponseMet defined design requirements.
    Guidewire Testing (Compass VA)Met defined design requirements.
    SpillageMet defined design requirements.
    SafetyMet defined design requirements.
    Simulated Use Testing (Vascular Access & Lumbar Puncture)Met defined design requirements.
    In vivo PerformanceCompass VA and LP devices accurately measure and monitor labeled physiologic pressures (invasive blood pressure/intracranial pressure) in a manner equivalent to the predicate device.

    Detailed Information on Studies and Ground Truth:

    This device (Mirador Compass™) is a disposable pressure transducer with an integrated digital display and pre-programmed diagnostic computer. It is not an AI/CAD (Computer-Aided Detection/Diagnosis) system in the modern sense. Therefore, many of the typical questions for AI/CAD devices, such as those related to observer performance studies (MRMC), ground truth established by experts, adjudication methods, and training/test sets for AI models, are not applicable to this type of medical device.

    1. Sample sized used for the test set and the data provenance:

    • In vitro (Bench) Testing: The document lists several in vitro tests (Luer Fittings, Mechanical Strength, Power Switch, etc.). The sample sizes for each of these component-level or functional tests are not specified in the summary.
    • In vivo Animal Study: The "in vivo animal study" used a swine model. The specific number of animals or the number of measurements taken is not provided.
    • Data Provenance: The in vitro tests were conducted by "Mirador Biomedical's in-house test methods, protocols, and requirements or in accordance with applicable recognized standards by accredited outside laboratories." The animal study was conducted in a swine model. The country of origin for the animal study is not specified, but the manufacturer is US-based (Seattle, Washington).

    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable for this type of device. The device's performance is measured against physical standards, calibrated equipment, and physiological measurements in an animal model, not against human expert interpretations of images or data that require adjudication for ground truth.

    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable. As this is not an AI/CAD device based on expert interpretation, no adjudication method was used.

    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No MRMC study was done. This device is a measurement tool, not an AI assistance tool for human interpretation.

    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • The device itself is a "standalone" measurement device in that it directly measures and displays pressure. However, it's not an "algorithm-only" standalone performance test in the context of AI/CAD where the algorithm's output is being evaluated for accuracy without human input. Its performance was assessed through direct physical and physiological measurements.

    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • In vitro Tests: Ground truth would be established by reference standards, calibrated measurement equipment, and specified engineering tolerances as defined by design requirements and industry standards (e.g., ANSI/AAMI BP22, ISO 594-1/2).
    • In vivo Animal Study: Ground truth for pressure measurements would be established by direct physiological measurements from the swine model, likely using a highly accurate reference pressure transducer or established invasive pressure monitoring techniques, against which the Compass device's readings are compared. The summary explicitly states that a "predicate Shenzhen DPT pressure sensor was used for relative comparison."

    7. The sample size for the training set:

    • Not applicable. This device is not an AI/Machine Learning diagnostic system that undergoes a "training" phase with a dataset.

    8. How the ground truth for the training set was established:

    • Not applicable, as there is no training set for an AI model.

    In summary, the K101518 submission for the Mirador Compass™ device demonstrates substantial equivalence through extensive bench testing and an in vivo animal study against established standards and a predicate device. The information provided aligns with typical submissions for Class II measurement devices, rather than AI-driven diagnostic tools.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1