Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Meridian M110 Fetal Monitoring System is an intrapartum fetal monitor that externally measures and displays fetal heart rate (FHR), maternal heart rate (MHR), and uterine contractions (UA). The Meridian M110 Fetal Monitoring System acquires and displays the FHR and UA from abdominal surface electrodes that detect the fetal ECG signals, maternal ECG signals, and of uterine muscle contraction signals. Tracings of FHR and UA are displayed onto a primary fetal monitor.

    The Meridian M110 Fetal Monitoring System is indicated for use on women who are at ≥ 37 completed weeks, in labor, with singleton pregnancies using surface electrodes on the maternal abdomen. The Meridian M110 Fetal Monitoring System is intended for use by health care professionals in a clinical setting.

    Device Description

    The Meridian M110 Fetal Monitoring System is an intrapartum fetal monitor that externally measures and displays fetal heart rate (FHR), maternal heart rate (MHR), and uterine contractions (UA). The Meridian M110 Fetal Monitoring System acquires and displays the FHR and UA from abdominal surface electrodes that detect the fetal ECG signals, maternal ECG signals, and of uterine muscle contraction signals. Tracings of FHR and UA are displayed onto a primary fetal monitor.

    AI/ML Overview

    I am sorry, but the provided text does not contain information about the acceptance criteria and study details for the Meridian M110 Fetal Monitoring System. The document is an FDA 510(k) clearance letter confirming that the device is substantially equivalent to legally marketed predicate devices, along with its indications for use. It outlines regulatory requirements but does not include performance acceptance criteria or study results.

    Ask a Question

    Ask a specific question about this device

    K Number
    K142883

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2015-04-17

    (197 days)

    Product Code
    Regulation Number
    884.2740
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MindChild Medical Meridian M100 Fetal Heart Rate Monitor is an intrapartum fetal monitor that externally measures and displays fetal heart rate (FHR). The MindChild Meridian M100 acquires and displays the FHR tracing from abdominal surface electrodes that detect the fetal ECG signal (fECG). FHR tracings are displayed onto a primary fetal monitor. In addition, the M100 synchronizes the TOCO transducer signal which is also displayed to the primary fetal monitor.

    The MindChild Meridian M100 is indicated for use on women who are at >36 completed weeks, in labor, with singleton pregnancies, using surface electrodes on the maternal abdomen. The MindChild Meridian is intended for use by healthcare professionals in a clinical setting.

    Device Description

    The Meridian M100 Fetal Heart Rate Monitor is a modification of the cleared M1000 Fetal Heart Rate Monitor (K120018). The modifications are ones that do not affect the intended use or the fundamental scientific technology. The M100 Fetal Heart Rate Monitor is an AC powered fetal heart rate (FHR) monitor. The product consists of a signal processor containing hardware circuitry and software which acquires and processes analog signals using external, abdominal electrode sensors. Software algorithms extract ECG morphology and monitor fetal heart rate.

    The M100 connects to existing cleared fetal heart rate monitors with DECG (i.e., primary monitor) using the primary monitor's screen for display of fetal heart rate. The patient data is displayed real time on the primary monitor's display. The Meridian M100 is used as an accessory to a main/primary fetal monitor replacing the Doppler sensor on those monitors. The M100 is for use with the GE Corometrics 120 and the GE Corometrics 250 Fetal Monitors.

    AI/ML Overview

    This document is a 510(k) summary for the MindChild Meridian M100 Fetal Heart Rate Monitor, which is a modification of a previously cleared device, the M1000. Therefore, the acceptance criteria and study information provided pertain to demonstrating substantial equivalence to the predicate device, rather than a full de novo validation.

    Here's an analysis of the provided information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of "acceptance criteria" in the format of specific thresholds for clinical performance metrics (e.g., sensitivity, specificity, accuracy for FHR detection). Instead, it focuses on demonstrating that the device meets its design specifications and performs as intended, largely by confirming that the modifications from the predicate M1000 do not adversely affect safety and effectiveness.

    The "Performance Testing" section (page 11) lists various tests performed and their results:

    TestPurposeReported Device Performance
    Dimensional VerificationTo verify products meets specified dimensions and weightMet design specification
    Fetal Heart Rate OutputVerify FHR output waveform for both minimum and maximum amplitude fetal heart output.Met design specification
    TOCO OutputVerify Toco output voltage for pressure applicationMet design specification
    Packaging TestingVerification of packaging and labeling.Met design specification
    Cable Pull StrengthVerify the cable connections are maintained when a 5 ft/lb. load is applied.Met design specification
    Drop TestVerification that unit could sustain dropMet specification
    Push TestVerify mechanical strength of enclosure by applying a 250N force to all sides and bottom of enclosureMet specification
    CleaningVerify that cleaning, disinfecting, and sterilizing the monitor do not create hazards and are compatible with the M100.Met specification
    Software ValidationVerify that all software modifications made performed as intended. Complete re-validation of the software was performed.Software validation demonstrated that the software performed according to design specifications, met requirement specifications and that there were no unintended effects due to any software modifications made.

    Additional performance indicators from the "Device Characteristic Comparison" section (pages 6-10) include:

    • Intended Use: Identical to predicate.
    • Indications for Use: Very similar to predicate, differences don't affect safety or effectiveness.
    • Regulation Number: Identical to predicate (884.2740).
    • Display: Identical information displayed as predicate; M100 uses primary monitor.
    • Controls: Each unit has power on/off; M100 utilizes primary monitor for other controls.
    • Printer: Both allow for downloading/transfer of patient data.
    • Front/Top Connections: Similar functionality for external electrodes; M100 adds TOCO connection, M1000 had optional fetal scalp electrode connection. These differences do not affect substantial equivalence.
    • Rear Connections: M100 has connections to primary monitor; M1000 only power. Functionally identical.
    • Side Connections: No impact.
    • Input Voltage Range: Identical (120 VAC ±10%).
    • Frequency: Identical (60 Hz).
    • Method: Algorithm for fECG acquisition and processing is the same as the M1000. M100 adds TOCO signal synchronization.
    • Measurement Range: Identical (50 - 240 BPM; abdominal sensor mode).
    • Signal Processing: Identical (Analog to digital).
    • Safety: Identical patient protection isolation barrier and compliance to FDA recognized standards.
    • Software: Microsoft Windows 7 embedded operating system; proprietary MindChild ECG algorithm is the same. M100 has software modification for syncing uterine contraction signal.

    2. Sample Size Used for the Test Set and Data Provenance

    The document states: "The algorithm used in the M100 to acquire and process fetal ECG signals through externally placed abdominal electrodes is the same as that used and clinically validated in the cleared M1000 device." (page 5) and "The clinical validation of the software responsible for calculating FHR from fECG signals was performed in support of the cleared 510(k) for the M1000. This clinical validation is directly applicable to the modified design as the signal acquisition components (external electrodes) and the signal processing software and algorithm remains the same in the modified M100 device. Therefore further clinical validation was not necessary for the modified device." (page 13).

    This implies that no new clinical test set was used for the M100 device itself to validate the FHR algorithm. The previous validation for the M1000 is leveraged. The document does not provide details of the M1000's clinical validation, including sample size or data provenance (e.g., country of origin, retrospective/prospective). New hardware and software modifications were assessed through bench testing and software validation (pages 11-12), but these are not clinical studies with a "test set" in the sense of patient data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    As stated above, no new clinical test set was required for the FHR algorithm for the M100 due to its substantial equivalence to the M1000's algorithm. Details about experts used for the M1000's original clinical validation are not provided in this document.

    4. Adjudication Method for the Test Set

    Not applicable, as no new clinical test set was explicitly detailed for the M100 device's FHR algorithm validation.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    Not applicable. This device is a Fetal Heart Rate Monitor, not an AI diagnostic imaging tool or a system designed to assist human readers in interpreting complex data like medical images where MRMC studies are common. It calculates and displays FHR, and its core algorithm was previously validated as part of the M1000.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    The document heavily implies that the FHR algorithm's performance was evaluated independently during the M1000's initial clearance, as it's a "fetal heart rate monitor" that "acquires and processes analog signals using external, abdominal electrode sensors" and "Software algorithms extract ECG morphology and monitor fetal heart rate." (page 4). The consistency of this algorithm between M1000 and M100 suggests a standalone performance evaluation of the algorithm for FHR calculation. However, specific details of this standalone performance, such as metrics (accuracy, precision, recall) or dataset used, are not in this document, as it focuses on demonstrating equivalence.

    7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)

    For the M100's FHR calculation algorithm, the ground truth methodology would have been established during the M1000's original clinical validation. This document does not specify the type of ground truth used for that validation (e.g., comparison against an internal uterine pressure catheter for FHR).

    8. The Sample Size for the Training Set

    The document does not provide information about a training set for the FHR algorithm. This is typical for traditional signal processing algorithms rather than modern deep learning AI models, which explicitly require separate training and testing sets. The algorithm was "clinically validated" for the M1000, suggesting a validation dataset, but not necessarily a "training set" in the current AI sense.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as no specific training set or its ground truth establishment is mentioned in this document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K120018

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-09-19

    (260 days)

    Product Code
    Regulation Number
    884.2740
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MindChild Medical Meridian Fetal Heart Rate Monitor is an intrapartum fetal monitor that externally or internally measures and displays fetal heart rate (FHR). The MindChild Meridian acquires and displays the FHR tracing from abdominal surface electrodes that detect the fetal ECG signal (fECG). The MindChild Meridian may also be used to measure and display fetal heart rate using direct ECG (DECG) with a fetal scalp electrode. The MindChild Meridian is indicated for use on women who are at term (> 36 completed weeks), in labor, with singleton pregnancies, using surface electrodes on the maternal abdomen. The MindChild Meridian is intended for use by healthcare professionals in a clinical setting.

    Device Description

    The MindChild Medical Meridian Fetal Heart Rate Monitor is an AC powered fetal heart rate (FHR) monitor. The product consists of a signal processor containing hardware circuitry and software which acquires and processes analog signals using external, abdominal electrode sensors. Software algorithms extract ECG morphology and monitor fetal heart rate. Data is stored locally for retrieval in an off-line mode. The patient data is displayed real time on a touch screen LCD display monitor. Data may be printed out offline for permanent hard copy records. The system also possesses the ability to acquire and analyze signals from a fetal scalp electrode using the device's direct ECG (DECG) capability. The entire system is housed within a medical grade roll-up cart. The device is intended to be used in a hospital environment by trained medical staff.

    The MindChild Medical Meridian Fetal Heart Rate Monitor is an intrapartum fetal monitor that externally or internally measures and displays fetal heart rate (FHR). The MindChild Meridian acquires and displays the FHR tracing from abdominal surface electrodes that detect the fetal ECG signal (fECG). The MindChild Meridian may also be used to measure and display fetal heart rate using direct ECG (DECG) with a fetal scalp electrode.

    AI/ML Overview

    Here's an analysis of the provided text regarding the MindChild Medical Meridian Monitor, structured to answer your questions about acceptance criteria and the supporting study:

    The document (K120018) describes the MindChild Medical Meridian Fetal Heart Rate (FHR) Monitor, an intrapartum fetal monitor. The primary study mentioned is a clinical validation of the Meridian algorithm for calculating FHR from fECG signals.


    1. Table of Acceptance Criteria and Reported Device Performance

    The document doesn't explicitly state "acceptance criteria" with numerical targets in a formal table. However, based on the clinical validation section, we can infer the primary performance metric and its outcome. The comparator device, the HP M1350A Fetal Monitor/FSE, serves as the "Gold Standard" or reference for evaluating performance.

    Acceptance Criteria (Inferred from study goals)Reported Device Performance
    FHR accuracy within a clinically acceptable number of beats per minute (BPM) compared to Gold Standard.Mean RMS error of 5 BPM for the Meridian algorithm (similar to other FDA cleared FHR monitors).
    Percent successful recordings (i.e., # comparator signals / # of Meridian signals x 100)At least 94% successful signal reading for all signals recorded with the comparator FSE device.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document does not explicitly state the total number of patients or recordings used in the clinical validation study. It mentions a "single Bland Altman (BA) difference plot was generated... for each patient," which implies multiple patients were included, but the exact count is not given.
    • Data Provenance: Not specified in the provided text (e.g., country of origin, retrospective or prospective). However, the study involved acquiring and recording fetal ECG signals simultaneously using the Meridian Monitor and the HP M1350A Fetal Monitor/FSE, suggesting it was a prospective data collection in a clinical setting.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document does not mention the use of experts to establish ground truth for the test set. Instead, the HP M1350A Fetal Monitor/FSE itself was considered the "comparator" or "Gold Standard" for FHR measurements.


    4. Adjudication Method for the Test Set

    No adjudication method is described. The comparison was directly between the Meridian algorithm's FHR calculations and those from the HP M1350A Fetal Monitor/FSE.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    No MRMC comparative effectiveness study was done or mentioned. This study focuses on the standalone algorithm's performance against a reference device, not human reader performance or improvement with AI assistance.


    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, a standalone (algorithm-only) performance study was conducted. The clinical validation directly compared the fetal heart rates calculated by the Meridian algorithm to the fetal heart rates calculated by the HP M1350A Fetal Monitor/FSE. There is no mention of human interpretation being part of this primary performance evaluation.


    7. The Type of Ground Truth Used

    The ground truth for the clinical validation was derived from the HP M1350A Fetal Monitor/FSE, which served as the "Gold Standard device" or comparator for calculating fetal heart rate. This is a form of reference device comparison or device-derived ground truth.


    8. The Sample Size for the Training Set

    The document does not provide any information regarding a training set or its sample size. The clinical validation solely describes the testing of the Meridian algorithm.


    9. How the Ground Truth for the Training Set Was Established

    Since no training set is mentioned, there is no information on how its ground truth was established.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1