Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    K Number
    K112323

    Validate with FDA (Live)

    Device Name
    CARDIA-CATH
    Date Cleared
    2011-10-18

    (67 days)

    Product Code
    Regulation Number
    870.2870
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Monitoring combined pressure and electrical impedance (volume) in the cardiovascular system

    Device Description

    The Cath-Cath-Catheter is a disposable/single use, sterile pressure-volume catheter transducer that combines one or two pressure transducers and a series of electrodes mounted at various locations along the distal segment of the catheter body. It terminates in one electrical connection at the proximal end.

    AI/ML Overview

    The medical device in question is the Millar Instruments Cardia-Cath, a Pressure-Volume (PV) Catheter Transducer. The provided text outlines the acceptance criteria and corresponding study results for this device, primarily focusing on verification testing, biocompatibility, sterility, and accelerated aging, demonstrating its substantial equivalence to a legally marketed predicate device (Millar model SPC-550).

    Here's an organized breakdown of the requested information:


    1. Table of Acceptance Criteria and Reported Device Performance

    Test CategoryAcceptance CriteriaReported Device Performance
    Verification Testing (BP-22)- Catheters must meet the same specifications as the legally marketed predicate device as defined in its IFU. - Compliance with BP-22 standard requirements. - Temperature Error Band at Zero Pressure: ≤ 3.24 mmHg (for predicate device). - Sensitivity Error Band: ≤ 3.16% (for predicate device).- All tested catheters met the testing requirements and complied with BP-22 standard requirements. - Deviation 1: Temperature Error Band at Zero Pressure relaxed to 3.5 mmHg. - Deviation 2: Sensitivity Error Band relaxed to 3.5%. - The modified device IFU includes the new (relaxed) values to ensure end-users obtain the same level of data accuracy.
    Transportation Testing- Compliance with ASTM D4169-08 standards for fully packaged finished product. - Samples must pass all tests including bubble leak and packaging integrity. - No impact on device functionality (safety and effectiveness) after transportation testing.- Samples passed all tests including bubble leak test and packaging integrity, as per DDL report No: 1009057 Rev. A. - Functionality testing (Millar reference VRP-2010-0423-01 VAL-180) confirmed no impact on device functionality, safety, and effectiveness.
    Biocompatibility (ISO 10993-1)- Requirements for External communicating device, circulating blood, contact duration A-Limited (<24hrs): Cytotoxicity, Sensitization, Irritation or Intracutaneous reactivity, Systemic toxicity (acute), and Haemocompatibility. - Statistical similarity to the predicate device (Model SPC-550).- Nelson Labs analysis determined that the Cardia-Cath (modified device) was statistically similar, and therefore substantially equivalent, to the predicate (legally marketed) device, Model SPC-550 for all tests: Modified ASTM Hemolysis (Direct Contact Method), Partial Thromboplastin time (PTT) test, MEM Elution, Complement Activation Test, ISO Acute Systemic Injection Test, Rabbit Pyrogen Test (Material Mediated)- ISO, ISO Intracutaneous Reactivity Test, Thrombogenicity Study in Dogs ISO, ISO Guinea Pig Maximization Sensitization Test.
    Sterility- Method: EtO-Overkill, validated per ANSI/AAMI ISO 11135. - Sterility Assurance Level (SAL): ≥ 10⁻⁶. - Pyrogen Free: Limulus Amebocyte Lysate test (LAL) should not show significant interference with lysate gel formation. - Residuals: Ethylene Oxide (EO) < 10mg, Ethylene Chlorohydrin < 5mg (per ANSI/AAMI/ISO 10993-7). - EtO breathable materials (Tyvek) for sterilant penetration and maintenance of sterility.- Sterility method EtO-Overkill validated per ANSI/AAMI ISO 11135. - Actual SAL: 3.33 times more than 10⁻⁶ (meeting the requirement). - LAL test showed no significant interference with lysate gel formation, indicating pyrogen-free status. - Maximum EO residuals: 0.3096mg (<10mg). - Maximum Ethylene Chlorohydrin residuals: 0.1079mg (<5mg). - Packaging uses EtO Breathable materials (Tyvek).
    Accelerated Aging- 2-year accelerated aging test per ASTM F1980-07. - Device must pass test criteria after accelerated aging. - Substantial equivalence to the predicate device (Model SPC-550) after aging.- A 2-year accelerated aging test was performed by DDL. - The tested samples were inspected by Millar, and the Cardia-Cath passed the test criteria and maintained substantial equivalence to the predicate device, Model SPC-550.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the exact sample sizes for each test set (e.g., number of catheters tested for BP-22, transportation, or accelerated aging, or the number of samples for biocompatibility). It mentions:

    • "Selected samples of the modified catheters were tested to the same specification..." for verification testing.
    • "Samples" were tested for transportation testing.
    • "Samples of the legally marketed device and the modified device were sent to Nelson Laboratories" for biocompatibility.
    • The accelerated aging test was performed on "the actual modified device. Cardia-Cath."

    Data Provenance: The studies appear to be prospective verification and validation tests conducted by Millar Instruments, DDL (for transportation and accelerated aging), and Nelson Labs (for biocompatibility). The testing was for a modified device (Cardia-Cath) to establish its substantial equivalence to a legally marketed predicate device (Millar model SPC-550). The country of origin of the data is implied to be primarily the United States, given the involved organizations (Millar Instruments in Houston, TX; Nelson Labs; DDL; FDA submission).


    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    This type of information is generally not applicable to the non-clinical bench and lab testing described in the document. The "ground truth" here is defined by established international standards (ISO 14971, ISO 10993-1, ISO 11135, ASTM D4169-08, ASTM F1980-07) and internal specifications of the manufacturer, rather than expert consensus on clinical findings.

    However, the document states:

    • "The final report [for verification testing] has been reviewed and approved by qualified individuals."
    • Biocompatibility testing was performed "following GLP requirements" by Nelson Labs, an accredited laboratory.

    This indicates that experts with relevant qualifications (e.g., engineers for verification testing, toxicologists/biologists for biocompatibility) would have been involved in the design, execution, and review of these tests. Specific numbers and detailed qualifications are not provided.


    4. Adjudication Method for the Test Set

    Adjudication methods (like 2+1, 3+1) are typically used in clinical studies involving interpretation of medical images or patient data by multiple human readers. This document describes non-clinical performance testing (bench testing, lab analysis, physical testing) against predefined technical specifications and standards. Therefore, an adjudication method for a test set in the conventional sense is not applicable. The "adjudication" is inherent in the objective measurements and pass/fail criteria of the specified tests.


    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. The document describes non-clinical performance and safety testing (bench, lab, and physical tests) of a medical device against established standards and a predicate device. It does not involve human readers interpreting cases, nor does it assess the effect of AI assistance on human reader performance.


    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

    The device described is a physical catheter transducer, not an AI algorithm. Therefore, the concept of "standalone (algorithm only without human-in-the-loop performance)" is not applicable here. The performance described is inherent to the physical device itself.


    7. The Type of Ground Truth Used

    The ground truth for the various tests is defined by:

    • Industry Standards: ISO 14971:2007 (Risk Management), ISO 10993-1 (Biocompatibility), ISO 11135 (Sterilization), ASTM D4169-08 (Transportation), ASTM F1980-07 (Accelerated Aging).
    • Predicate Device Specifications: The "same specification as the legally marketed device as defined in its IFU" for general performance, and specific values for parameters like Temperature Error Band and Sensitivity Error Band of the predicate device (SPC-550).
    • Regulatory Limits: Maximum permissible levels for EO and Ethylene Chlorohydrin residuals per ANSI/AAMI/ISO 10993-7.
    • Objective Measurements: Directly measured parameters against these standards and specifications (e.g., pressure readings, percentage errors, biological responses, SAL values).

    8. The Sample Size for the Training Set

    The concept of a "training set" and "training data" is applicable to machine learning or AI models. Since this document describes a physical medical device and its non-clinical verification and validation testing, there is no training set in the context of AI/ML.


    9. How the Ground Truth for the Training Set Was Established

    As there is no training set for an AI/ML model, this question is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K093111

    Validate with FDA (Live)

    Date Cleared
    2010-01-05

    (96 days)

    Product Code
    Regulation Number
    870.2870
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Mikro-Cath is a single-use cardiovascular catheter intended to be used for medical research and diagnostic purposes. The catheter is used to measure hemodynamic cardiac pressures in the hurnan body to allow physicians to better understand cardiac health. The catheter would be used as a minimally invasive device under short term limited body contact <24 hours. The typical application will be through the femoral artery with the use of an additional guiding catheter.

    Device Description

    The Mikro-Cath is a disposable/single use, sterile cardiovascular pressure catheter transducer for monitoring and measuring pressure anywhere in the cardiovascular system with an appropriately sized guide catheter.

    AI/ML Overview

    Here's an analysis of the provided text regarding the Mikro-Cath device, outlining acceptance criteria and studies:

    I. Acceptance Criteria and Reported Device Performance

    The document primarily focuses on demonstrating substantial equivalence to a predicate device and adherence to recognized standards. It does not explicitly state numerical acceptance criteria for a single performance metric with a corresponding reported device performance value in a table format. Instead, it indicates that the device's performance "maintain[s] the same levels of safety and effectiveness as the predicate devices currently in commercial distribution" by complying with applicable standards.

    Therefore, the table below reflects this approach:

    Acceptance Criterion (Standard)Reported Device Performance (Compliance Claim)
    Performance Testing (AAMI ANSI BP22:1994 (R) 2006)"The performance [tooling] and therefore maintain the same levels of safety and effectiveness as the predicate devices currently in commercial distribution." (Implicitly, the device meets the performance requirements of this standard.)
    Biocompatibility (ISO 10993-1:2003)"Biocompatibility testing in accordance with ISO 10993-1:2003" (Implies compliance with the standard.)
    Sterility Validation (ISO 11135-1:2007)"Sterility validation testing in accordance with ISO 11135-1:2007" (Implies compliance with the standard.)
    Electromagnetic Compatibility (EN 60601-1-2:2007)"EMC testing in accordance with EN 60601-1-2:2007" (Implies compliance with the standard.)
    Packaging Accelerating/Aging (ASTM F 1980 & ISO 11607)"Packaging Accelerating/Aging testing in accordance with ASTM F 1980 (Standard Guide for Accelerated Aging of Sterile Medical Device Packages) and ISO 11607 (Packaging for Terminally Sterilized Medical Devices)." (Implies compliance with the standards.)

    II. Sample Size Used for the Test Set and Data Provenance

    The provided document describes non-clinical tests (e.g., biocompatibility, sterility, EMC, packaging, performance testing to AAMI ANSI BP22). It does not mention specific clinical trials or a test set of patient data in the context of device performance to measure hemodynamic cardiac pressures. The performance testing refers to compliance with safety and effectiveness standards, likely through bench testing, rather than a clinical dataset.

    Therefore, this information is not available in the provided text.

    III. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    Given that no clinical test set of patient data is described, there is no mention of experts establishing ground truth for such a set. The expertise involved would be in conducting and interpreting the non-clinical tests according to the cited standards.

    Therefore, this information is not available in the provided text.

    IV. Adjudication Method for the Test Set

    As no clinical test set requiring adjudication is described, this information is not applicable and not available in the provided text.

    V. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    An MRMC comparative effectiveness study, which typically evaluates human reader performance with and without AI assistance, is not mentioned in the provided document. The device in question is a physical catheter transducer, not an AI or imaging diagnostic tool.

    Therefore, this information is not applicable and not available in the provided text.

    VI. Standalone (Algorithm Only) Performance Study

    The Mikro-Cath is a physical medical device, specifically a catheter tip pressure transducer. It is not an algorithm or AI system. Therefore, a standalone algorithm-only performance study is not applicable to this device.

    Therefore, this information is not applicable and not available in the provided text.

    VII. Type of Ground Truth Used

    For the non-clinical performance testing (e.g., to AAMI ANSI BP22), the ground truth would be the physical measurements and specifications defined by the standard itself, obtained through laboratory or bench testing. For biocompatibility, it would be the results of in vitro and in vivo biological tests as defined by ISO 10993. For sterility, it would be the microbiological test results demonstrating a sterility assurance level as defined by ISO 11135.

    There is no mention of ground truth based on expert consensus, pathology, or outcomes data in the context of clinical performance, as no clinical study is described.

    VIII. Sample Size for the Training Set

    The device is a physical catheter, not a machine learning model. Therefore, the concept of a "training set" in the context of AI/ML is not applicable. The manufacturing and design process would follow established engineering principles and quality systems rather than data-driven training.

    Therefore, this information is not applicable and not available in the provided text.

    IX. How the Ground Truth for the Training Set Was Established

    As the device is not an AI/ML model, the concept of establishing ground truth for a training set is not applicable.

    Therefore, this information is not applicable and not available in the provided text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K013205

    Validate with FDA (Live)

    Date Cleared
    2002-07-02

    (280 days)

    Product Code
    Regulation Number
    870.2060
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PCU-2000 Pressure Control Unit is a two-channel electrically isolated amplifier that is intended for use with Millar Mikro-Tip Pressure Catheters that have the standard medical sensitivity of 5 microvolts per mmHg of applied pressure and per volt of bridge excitation. The PCU-2000 provides an electrical interface between a physiological pressure transducer and a data acquisition module or medical monitor. It is intended for use in monitoring diagnostic pressures, such as noninvasive or invasive blood pressures, intracranial pressures, gastrointestinal pressures, esophageal pressures, urinary tract pressures, intrauterine pressures, intraocular pressures and other physiological pressures with similar ranges. It is intended for use in critical care areas in a hospital and in diagnostic centers in hospitals or medical clinics. It is intended for use by trained clinicians or research personnel

    Device Description

    The PCU-2000 Pressure Control Unit is a two-channel electrically isolated amplifier that is intended for use with Millar Mikro-Tip Pressure Catheters that have the standard medical sensitivity of 5 microvolts per mmHg of applied pressure and per volt of bridge excitation. The PCU-2000 provides an electrical interface between a physiological pressure transducer and a data acquisition module or medical monitor.

    AI/ML Overview

    The provided document is a 510(k) premarket notification letter from the FDA for a device called "Pressure Control Unit PCU-2000" and an "Indications for Use Statement" attachment.

    This document describes a medical device (transducer signal amplifier and conditioner) and its intended use, rather than an AI/algorithm-driven device that would undergo significant performance evaluation with acceptance criteria, ground truth, and studies like MRMC or standalone performance.

    Therefore, the requested information about acceptance criteria and studies (including sample sizes, ground truth establishment, expert involvement, adjudication methods, and AI-specific studies) cannot be extracted from this document because:

    • The device is a hardware component: The PCU-2000 is described as a "two-channel electrically isolated amplifier" intended to provide an "electrical interface" between a pressure transducer and a data acquisition module or medical monitor. It conditions analog signals, not processes complex image or sensor data with AI.
    • The 510(k) clearance process for this type of device focuses on substantial equivalence: The letter confirms that the FDA has reviewed the device and determined it is "substantially equivalent" to legally marketed predicate devices. This typically involves demonstrating that the new device has the same intended use and technological characteristics as a predicate, or if there are differences, that those differences do not raise new questions of safety or effectiveness. This is different from the extensive performance validation required for AI/ML devices.
    • No mention of AI, algorithms, or complex data analysis: The entire document refers to basic signal amplification and conditioning for physiological pressure measurements. There is no indication of any AI component, predictive modeling, or diagnostic algorithms that would necessitate the type of detailed performance studies you're asking about.

    In summary, the provided text does not contain the information requested because the device described is not an AI/algorithm-driven device that would typically undergo such studies and evaluations.

    Ask a Question

    Ask a specific question about this device

    K Number
    K993694

    Validate with FDA (Live)

    Date Cleared
    2000-03-01

    (120 days)

    Product Code
    Regulation Number
    870.1200
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Millar Angiographic catheter is designed to provide high fidelity pressure readings and deliver radiopaque contrast medium to selected sites in the vascular system.

    Device Description

    Not Found

    AI/ML Overview

    This document is a 510(k) clearance letter for the Mikro-Tip Angiographic Catheter and does not contain the detailed study information requested. The letter states that the device is substantially equivalent to legally marketed predicate devices.

    Therefore, I cannot provide the information requested in the prompt based on the provided document. The document only confirms the FDA clearance for the device and its intended use, but it does not include details on acceptance criteria, device performance studies, sample sizes, ground truth establishment, or expert qualifications.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1