Search Filters

Search Results

Found 6 results

510(k) Data Aggregation

    K Number
    K103474

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2011-02-07

    (73 days)

    Product Code
    Regulation Number
    872.3770
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ENA TEMP/ ENAMEL PLUS TEMP is a system of liquid and powder resin available in different colours, for aesthetic temporary crown and bridge. It can be used both in laboratory (by indirect method: diagnostic waxing up or vacuum forming matrix) and in dental practice (by direct method in the mouth: silicone impression or preformed crown).

    Device Description

    ENA TEMP/ ENAMEL PLUS TEMP is a system of liquid and powder resin available in different colours, for aesthetic temporary crown and bridge. It can be used both in laboratory (by indirect method: diagnostic waxing up or vacuum forming matrix) and in dental practice (by direct method in the mouth: silicone impression or preformed crown). Please note that the resin is absolutely the same, but they will be sold with two different brand names for commercial reasons. So, in order to simplify, from here on they will be denoted as "Temp"

    AI/ML Overview

    This document describes a 510(k) premarket notification for a dental device, specifically a temporary crown and bridge resin. It focuses on establishing substantial equivalence to a predicate device rather than detailing extensive clinical trials with specific acceptance criteria and outcome measurements usually found in PMAs or clinical studies for novel devices.

    Therefore, the requested information about acceptance criteria, study design, sample sizes for test and training sets, ground truth establishment, expert qualifications, and MRMC studies is largely not present in the provided text. The submission focuses on compliance with performance standards and comparison to a predicate device.

    Here's a breakdown of what can be extracted and what information is missing:


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Stated/Implied)Reported Device Performance
    Compliance with ISO 20795 (Denture Materials)Listed as complying.
    Compliance with ISO 10477 (Crowns and Bridges)Listed as complying.
    Substantial Equivalence to K000894 (C&B RESIN PLUS POWDER/LIQUID, MODELS RETEMP A2, A3, A3.5, CL, RETEMP B2, B3, C2, D3, WH. RE by DENTAL VENTURES OF AMERICA, INC)Stated as substantially equivalent. The specific performance metrics used for this equivalence are not detailed.
    Addressing identified risks:The document lists risks (Allergic reaction, No or incomplete stability, Gaps at tooth side if not tight, Secondary caries, Irritation of respiratory system and skin) but does not provide specific performance data or tests demonstrating how these risks are mitigated or how performance meets criteria related to them.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Not mentioned. No specific "test set" in the context of a clinical performance study is described. The evaluation appears to be based on bench testing (compliance with ISO standards) and comparison to existing products.
    • Data Provenance: Not applicable, as detailed patient data or clinical study data is not presented. The submission is from Micerium SpA, Italy.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    • Number of Experts: Not mentioned.
    • Qualifications: Not mentioned.

    This type of information would be relevant for clinical studies where expert review is used to establish ground truth for a diagnostic or AI-driven device. This submission is for a material used in dentistry, and its performance evaluation likely relies on material science testing rather than expert-adjudicated clinical outcomes.


    4. Adjudication Method for the Test Set

    • Adjudication Method: Not applicable. No clinical test set requiring adjudication is described.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size of AI vs. Without AI Assistance

    • MRMC Study: No, an MRMC study was not done. This device is a dental material, not an AI-driven diagnostic tool or a system involving human readers interpreting outputs.
    • Effect Size of AI: Not applicable.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    • Standalone Performance: Not applicable. This is not an algorithmic device.

    7. The Type of Ground Truth Used

    • Type of Ground Truth: The "ground truth" for this device's performance would primarily be established through material property testing against ISO standards (ISO 20795 for denture materials and ISO 10477 for crowns and bridges) and comparison to the identified predicate device (K000894) based on specifications and intended use. There is no mention of pathology, outcomes data from patients, or expert consensus in the clinical sense for establishing ground truth for this particular submission.

    8. The Sample Size for the Training Set

    • Sample Size: Not applicable. This is not an AI/ML device that requires a training set. The device is a resin for temporary crowns and bridges.

    9. How the Ground Truth for the Training Set Was Established

    • Ground Truth Establishment: Not applicable, as there is no training set for an AI/ML model.
    Ask a Question

    Ask a specific question about this device

    K Number
    K072978

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2008-01-17

    (87 days)

    Product Code
    Regulation Number
    872.3750
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Adhesive for luting ceramic and metal orthodontic brackets

    Device Description

    Adhesive system, light curable both with halogen and LED Light, for brackets cementation. The accessories contact tissue for less than 1 hour and therefore are exempt from 510(k) requirements and are described only generally.

    AI/ML Overview

    The provided document (K072978) is a 510(k) summary for a dental adhesive system (Tender Ortho Bond - Ena Bond - Ena Etch). This type of submission focuses on demonstrating substantial equivalence to a predicate device, primarily through performance testing that adheres to recognized standards. It does not contain details of a clinical study or AI performance evaluation as requested in the prompt.

    Therefore, many of the requested sections about acceptance criteria, device performance, sample sizes, expert involvement, and ground truth establishment, specifically in the context of an AI device and clinical study, cannot be extracted from this document.

    However, I can provide information based on what is available in the document regarding the device's performance evaluation against a standard.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document references "ISO 4049:2000 Dentistry - Polymer - Based Filling, Restorative and Luting materials" as the standard followed for evaluating the device. While the exact acceptance criteria from ISO 4049:2000 are not explicitly detailed in the provided text, the submission states that the standard was followed. The implied "reported device performance" is that the device met the requirements of this standard for its intended use, otherwise, substantial equivalence would not have been granted.

    Acceptance Criterion (from ISO 4049:2000 - Implied)Reported Device Performance
    Adhesion/Bond Strength (for luting orthodontic brackets)Complied with ISO 4049:2000 requirements for polymer-based luting materials.
    Polymerization characteristics (e.g., degree of cure, working time)Complied with ISO 4049:2000 requirements.
    Biocompatibility (e.g., cytotoxicity)Cytotoxicity tests appear in Appendix III (details not provided in this extract), implying satisfactory results.
    Other physical/chemical properties as per ISO 4049:2000Complied with ISO 4049:2000 requirements.

    2. Sample Size Used for the Test Set and Data Provenance

    This document does not describe a "test set" in the context of a clinical study or AI evaluation. The performance evaluation would have been conducted on laboratory samples according to the specifications of ISO 4049:2000. No information on sample sizes for these lab tests or data provenance (country of origin, retrospective/prospective) is provided in the publicly available summary.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    Not applicable. This document does not describe a study involving expert-established ground truth for a test set. The validation is based on adherence to an engineering standard (ISO 4049:2000).

    4. Adjudication Method for the Test Set

    Not applicable. No clinical test set or expert adjudication is described.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No. This document does not mention an MRMC study or any comparison of human readers with or without AI assistance. The device is a dental adhesive, not an AI diagnostic tool.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    No. The device is a dental adhesive. This concept is not applicable.

    7. The Type of Ground Truth Used

    The "ground truth" for this medical device submission is compliance with the ISO 4049:2000 standard for "Dentistry - Polymer-Based Filling, Restorative and Luting materials." This standard dictates the methods and criteria for evaluating the physical, chemical, and biological properties of such materials. The specific tests (e.g., bond strength, cytotoxicity) described within the standard would serve as the "ground truth" for performance.

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI device, so there is no training set in that context. The "training set" for the device's development would refer to the materials and formulations developed and refined prior to final testing, but no specific sample sizes for this are provided in the regulatory summary.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable for an AI training set. For the development of the adhesive material, the "ground truth" would have been established through iterative laboratory testing and material science principles, aiming to meet desired performance characteristics and ultimately comply with ISO 4049:2000. These details are not part of the 510(k) summary.

    Ask a Question

    Ask a specific question about this device

    K Number
    K063170

    Validate with FDA (Live)

    Device Name
    TENDER FIBER
    Manufacturer
    Date Cleared
    2007-01-12

    (86 days)

    Product Code
    Regulation Number
    872.3760
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Tender fiber is a semi-manufactured product made of fibers used to reinforce composite and acrylic prosthodontic appliances and to repair or reinforce denture and rethodontic appliances. It can be use as well to immobilize teeth for periodontic or orthodontic treatments.

    Device Description

    Tender Fiber are silanized glass fibers, impregnated with light curing resin used to reinforce prosthodontic appliances and to repair or reinforce denture and orthodontic appliances.

    AI/ML Overview

    This 510(k) summary provides information for a dental device, "Tender Fiber," and relies on substantial equivalence to predicate devices rather than a direct study demonstrating its performance against specific acceptance criteria. Therefore, much of the requested information regarding specific study details and statistical analysis is not available in this document.

    However, I can extract the relevant information and explicitly state what is not provided.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Biocompatibility:
    Cytotoxicity (ISO 10993-5:1999)"Cytotoxicity tests appears in appendix II.1" (Exact results not provided in this summary, but the report implies compliance.)
    Physical/Chemical Performance:
    ISO 4049:2000 Dentistry - Polymer-Based Filling, Restorative and Luting materialsDevice complies with this standard. (Specific performance metrics like flexural strength, depth of cure, etc., are not presented in this summary.)
    Quality Management Systems:
    ISO 13485:2003 Medical devices -- Quality management systems -- Requirements for regulatory purposesDevice complies with this standard.
    Risk Management:
    ISO 14971:2000/Amd 1:2003 - MEDICAL DEVICES - Application of risk management to medical devicesDevice complies with this standard.
    Overall Safety and Effectiveness:"Test results indicate that there are no hazards presented with the use of Tender Fiber as compared with the predicate devices." (No specific acceptance criteria or quantitative performance given, relies on substantial equivalence.)

    2. Sample size used for the test set and the data provenance

    • Not explicitly stated for performance testing. The document refers to "Cytotoxicity tests" and "Test results," but no details on sample size for these tests are provided.
    • Data Provenance: Not specified for any specific test. However, the manufacturer is Micerium SpA, located in Italy, suggesting the tests would have been performed there or commissioned by them.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not applicable. This document describes a medical device undergoing premarket notification based on substantial equivalence, not a study evaluating human reader performance against ground truth established by experts.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    • Not applicable. See point 3.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No. This is not a study involving AI or human reader comparison. It's a submission for a physical dental device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This is not an algorithmic device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • Not applicable/Not explicitly defined in the context of this submission. The "ground truth" here would relate to the performance standards and biocompatibility tests. For biocompatibility, the chemical and biological reactions serve as the "ground truth" against the ISO standards. For physical properties, the measured values compared against the ISO 4049 standard represent the ground truth.

    8. The sample size for the training set

    • Not applicable. This is not a machine learning or AI device that requires a training set.

    9. How the ground truth for the training set was established

    • Not applicable. See point 8.
    Ask a Question

    Ask a specific question about this device

    K Number
    K062355

    Validate with FDA (Live)

    Device Name
    ENA CEM
    Manufacturer
    Date Cleared
    2006-12-27

    (135 days)

    Product Code
    Regulation Number
    872.3275
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    EnaCem is a dual curing radiopaque fluorescent luting composite, available in several dentine colours, for cementation of posts, ceramic and composite inlay, onlay veneers and crowns. It can be used also for core build-up and as a liner.

    Device Description

    The ENA CEM is a luting material available in a flowable consistency (Ena Cem) and also in a high viscous version (Ena Cem HF) that is available in different colours. Both can be used for cementing prosthesis appliances made in laboratory to the tooth. They can be used also for post-cementation. Ena Cem HF can also be used for core build-up and as a liner. The accessories and try-in materials contact tissue for less than 1 hour and therefore are exempt from 510(k) requirements and are described only generally.

    AI/ML Overview

    The provided text is a 510(k) summary for a dental cement (ENA CEM) and does not contain information about acceptance criteria for a device's performance, nor does it describe a study proving the device meets such criteria. It primarily focuses on the classification, description, and substantial equivalence of the dental cement to existing products.

    Therefore, I cannot fulfill the request to describe acceptance criteria and a study proving their fulfillment based on the provided text. The document is about regulatory approval of a dental cement, not the performance evaluation of an AI/software device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K062367

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2006-12-08

    (116 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This device is intended to provide a fixed anchorage point for attachment of orthodontic appliances to facilitate the orthodontic movement of teeth. It is used temporarily and is removed after orthodontic treatment has been completed. Screws are intended for single use only on adults over the age of 12.

    Device Description

    This device is an endosseous dental implant consisting of the major component, of root-form, screw type, and the accessories, designed to facilitate placing and using the implant in orthodontic procedures. The accessories contact tissue for less than 1 hour and therefore are exempt from 510(k) requirements and are described only generally. The smoothly curved head has a groove in the middle for the attachment of elastics, chains, or coil springs commonly used in orthodontics. The head (groove) has a aperture where a wire or auxiliary can be attached. The rounded head of the implant has an internal hexagon for insertion of the screw driver. The MAS implant is made from Group V titanium.

    AI/ML Overview

    This 510(k) summary describes a dental implant device called the Miniscrews Anchoring System (MASᵀᴹ System). However, it does not contain details about acceptance criteria or a study proving the device meets specific performance metrics.

    The document focuses on establishing substantial equivalence to existing predicate devices based on:

    • Intended Use: Providing a fixed anchorage point for orthodontic appliances.
    • Technological Characteristics: Screw-type endosseous dental implants made of Grade V titanium with similar design concepts (e.g., rounded head with a groove for attachments, self-tapping, roughened screw area, smoothly machined transmucosal neck).
    • Materials: Group V titanium, which is a well-established material in the industry.

    Therefore, I cannot provide the requested information for the following points as they are not present in the provided text:

    • A table of acceptance criteria and the reported device performance.
    • Sample size used for the test set and the data provenance.
    • Number of experts used to establish the ground truth for the test set and their qualifications.
    • Adjudication method for the test set.
    • If a multi-reader multi-case (MRMC) comparative effectiveness study was done, or the effect size of human readers improving with AI vs without AI assistance.
    • If a standalone performance study was done.
    • The type of ground truth used (expert consensus, pathology, outcomes data, etc.).
    • The sample size for the training set.
    • How the ground truth for the training set was established.

    This submission is a traditional 510(k) that relies on demonstrating substantial equivalence to already cleared devices, rather than presenting novel performance data against specific acceptance criteria. The FDA's letter confirms that the device is substantially equivalent to legally marketed predicate devices, allowing its market entry based on this comparison.

    Ask a Question

    Ask a specific question about this device

    K Number
    K062906

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2006-11-17

    (51 days)

    Product Code
    Regulation Number
    872.3690
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ENA HFO is a composite system for direct and indirect aesthetic restorations in anterior and posterior sectors.

    Device Description

    ENA HFO is a system for direct and indirect aesthetic restorations in anterior and posterior sectors, composed by a light curing radiopaque composite, available in different colors, with accessories for application and polishing. The accessories contact tissue for less than 1 hour and therefore are exempt from 510(k) requirements and are described only generally.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study information for the K062906 device:

    It is important to note that the provided text is a 510(k) summary, which focuses on demonstrating substantial equivalence to a predicate device. It does not contain detailed information about specific acceptance criteria, performance studies with sample sizes, expert involvement, or statistical analyses typically found in a full study report or technical documentation. The summary primarily lists the device's properties and claims similarity to existing products.

    Acceptance Criteria and Reported Device Performance

    The 510(k) summary does not explicitly list quantitative acceptance criteria for device performance. Instead, it states that the device has "similar technological characteristics" and "similar physical data" to predicate devices. The "Risk to Health" section implicitly defines areas where the device should not fail, which could be interpreted as a form of acceptance criteria in terms of safety.

    Acceptance Criteria (Implied)Reported Device Performance
    Safety: No allergies or hypersensitivities.Cytotoxicity tests performed (appears in Appendix V) - implies acceptable
    Adhesion: Complete adhesion on teeth.Not explicitly stated as a performance metric, only as a potential adverse effect if not met.
    Curing: Complete and correct curing.Not explicitly stated as a performance metric, only as a potential adverse effect if not met.
    Polymerization: No or minimal polymerization gap/shrinkage.Not explicitly stated as a performance metric, only as a potential adverse effect if not met.
    Filling: Complete filling of the cavity.Not explicitly stated as a performance metric, only as a potential adverse effect if not met.
    General Performance: Similar to predicate devices in composition and physical data."They have similar composition (bis-GMA resin and glass filler) and similar physical data."

    Study Information

    Based on the provided 510(k) summary, the following study information is available:

    1. Sample size used for the test set and the data provenance:

      • Sample Size: Not specified in the provided summary.
      • Data Provenance: Not specified. The summary mentions "cytotoxicity tests" (in Appendix V, which is not provided), but no details on the data origin (e.g., country) or whether it was retrospective or prospective.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This information is not provided in the 510(k) summary. The summary does not describe a clinical study or a study involving expert assessment to establish ground truth for performance metrics.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • This information is not provided as there is no mention of a test set with human adjudication in the summary.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No MRMC study was conducted or reported. This device is a dental composite material, not an AI-assisted diagnostic tool, so such a study would not be applicable.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • This device is a physical material, not an algorithm. Therefore, a standalone algorithm performance study is not applicable and was not performed.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The summary implies that performance (e.g., cytotoxicity, physical data) was assessed against standard laboratory tests and material specifications, likely comparing to predicate devices or industry standards like ISO 4049:2000. It does not refer to clinical outcomes, expert consensus, or pathology for establishing ground truth of inherent device performance.
    7. The sample size for the training set:

      • Not applicable/Not provided. The summary refers to a manufactured product and its testing, not a machine learning model requiring a training set.
    8. How the ground truth for the training set was established:

      • Not applicable/Not provided, as there is no training set for a machine learning model.

    In summary, the provided K062906 510(k) summary is a regulatory document focused on demonstrating substantial equivalence. It highlights the device's intended use, composition, and similarity to existing products, along with general safety considerations (like cytotoxicity tests). It does not offer the detailed efficacy study data, expert involvement, or statistical results that would typically answer many of the questions asked.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1