    K063157

    Validate with FDA (Live)

    Date Cleared
    2008-03-10

    (510 days)

    Product Code
    Regulation Number
    888.3045
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Merries K-PHATE is intended to be packed into bony defects of the skeletal system ( extremities, spine, or pelvis) which are not intrinsic to the stability of the bony structure. These defects may be surgically created voids or from traumatic injury to the bone. The device gradually resorbs and is replaced with bone during the healing process. Rigid fixation techniques should be used in conjunction with this device.

    Device Description

    Merries K-PHATE bone graft substitute is a microporous and macroporous biphasic calcium phosphate ceramic consisting of 60% hydroxyapatite and 40% ß-tricalcium phosphate. Merries K-PHATE is supplied sterile in various shapes and sizes. The choice of different form or size of the product depends on the type and size of the recipient site. Blocks and wedges are used for large bony defect while granules are used as bone filler for small area.

    AI/ML Overview

    The provided text describes the 510(k) Premarket Notification Summary for the "Merries K-PHATE Bone graft substitute" (K063157). This document focuses on demonstrating substantial equivalence to predicate devices rather than proving the device meets specific performance criteria through a clinical study. Therefore, much of the requested information regarding acceptance criteria, performance metrics, and study design for a clinical effectiveness study is not explicitly present in the provided text.

    Specifically, this submission relies on:

    1. Material characteristics: The device is a biphasic calcium phosphate ceramic, similar to predicate devices.
    2. Function and intended use: It's designed as a resorbable bone void filler, the same as predicate devices.
    3. Biocompatibility testing: To ensure safety.
    4. Sterilization dose auditing: To ensure sterility.
    5. Chemical requirements: Conformance to ASTM standards for ceramic hydroxyapatite and beta-tricalcium phosphate.

    Here's an attempt to answer your questions based on the provided text, recognizing that a full clinical study with acceptance criteria for effectiveness is not detailed:


    Acceptance Criteria and Device Performance (Based on provided 510(k) Summary)

    Since this is a 510(k) submission for a bone graft substitute, the primary "acceptance criteria" discussed are related to substantial equivalence to legally marketed predicate devices, and the device's safety through biocompatibility and material testing. Performance in terms of clinical outcomes (e.g., rate of bone fusion, strength of new bone) is inferred from the predicate devices rather than directly proven through a new comparative clinical study for this specific device in this submission.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific CriterionReported Device Performance
    Substantial EquivalenceSimilarity in base materials, function, and intended use to listed predicate devices.Stated as "substantially equivalent" in base materials, function, and intended use to MBCP™, Pro Osteon® 500R, Vitoss® Scaffold, and BoneSave™ Bone void Filler.
    Material CompositionConformance to ASTM F1185-88 (ceramic hydroxyapatite)Implicitly stated as conforming to ASTM F1185-88.
    Conformance to ASTM F1088-87 (beta-tricalcium phosphate)Implicitly stated as conforming to ASTM F1088-87.
    Biocompatibility (ISO 10993 equivalent)Non-cytotoxicPass (Cytotoxicity Test)
    Non-irritantPass (Subcutaneous Irritation Test)
    Non-sensitizingPass (Guinea Pig Skin Sensitization Study)
    Compatible with surrounding tissuesPass (Implantation Test)
    Non-mutagenicPass (Sister Chromatid Exchange Test)
    SterilitySterilization dose auditing for sterility.Stated as "Merries K-PHATE has been tested of sterilization dose auditing for sterility."
    Manufacturing & QualityGood Manufacturing Practice (GMP)Required by Act (implied by 510(k) clearance)
    LabelingConformance to labeling regulations (21 CFR Part 801)Required by Act (implied by 510(k) clearance)

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The provided summary does not detail a clinical "test set" for performance evaluation in the way a clinical trial would. The "tests" reported are primarily bench/laboratory tests for material properties and biocompatibility.

    • Sample Size for Bench/Biocompatibility Tests: Not specified in the summary. These tests typically involve a relevant number of samples for each specific test (e.g., cell cultures for cytotoxicity, animal models for irritation/sensitization/implantation).
    • Data Provenance: The manufacturing company is based in Taipei, Taiwan (Merries International Inc.). The specific locations where the individual biocompatibility tests were conducted are not stated, but given the company's location, they were likely performed in a certified lab, potentially in Taiwan or another country as required for international standards. These tests are prospective in nature, conducted specifically for this submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable as there is no human-read clinical test set described in this summary to establish a "ground truth" for diagnostic or prognostic performance. The "ground truth" for the material and biocompatibility tests is based on objective scientific measurements and established standards (e.g., cell viability, tissue response, chemical composition as per ASTM).

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This is not applicable. No clinical test set with human assessment requiring adjudication is described.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable. This device is a bone graft substitute, not an AI-powered diagnostic or assistive technology. No MRMC study was conducted or is relevant to this type of device.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This is not applicable. This device is a bone graft substitute, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the tests conducted:

    • Material Composition: Ground truth is based on physical and chemical analysis techniques confirming compliance with ASTM standards.
    • Biocompatibility: Ground truth is based on biological assays and histological examination in animal models, evaluated against established criteria for cytotoxicity, irritation, sensitization, and tissue compatibility.
    • Sterility: Ground truth is based on microbiological testing (sterilization dose auditing).

    There is no "expert consensus" or "pathology" in the sense of a clinical diagnostic study. Outcomes data (e.g., long-term bone healing) would typically be part of a full clinical trial, which is not detailed here for this 510(k) submission.

    8. The sample size for the training set

    This is not applicable. This device is a medical implant, not a machine learning model, and therefore does not have a "training set."

    9. How the ground truth for the training set was established

    This is not applicable as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K051698

    Validate with FDA (Live)

    Date Cleared
    2005-09-29

    (97 days)

    Product Code
    Regulation Number
    888.3045
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Merries UNI-OSTEO pellets are used to fill bony defects which may be surgically created voids or from traumatic injury to the bone. The product is indicated to be packed into bony defects of the skeletal system which is not intrinsic to the stability of the bony structure.

    Device Description

    Merries UNI-OSTEO pellets are made of calcium sulfate (CaSO4). These pellets dissolve safely and completely in the body in approximately 30-60 days and resorption rate corresponds with new bone growth. The product is visible on radiographs so that the implant location can be easily identified and monitored. Merries UNI-OSTEO pellets are osteoconductive meaning that these pellets act as a scaffold and facilitate new bone growth.

    AI/ML Overview

    This document describes the premarket notification for the Merries UNI-OSTEO Pellets Bone Graft Substitute. The acceptance criteria and the study proving it are primarily focused on biocompatibility and equivalence to predicate devices, rather than a typical AI/software performance study.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Test/CriteriaAcceptance CriteriaReported Device Performance
    BiocompatibilityNon-cytotoxic, non-sensitizing, non-mutagenic, and compatible with surrounding tissuesPass (for all individual tests listed below)
    Cytotoxicity(Implicitly: non-cytotoxic)Pass
    Intracutaneous Reactivity Test(Implicitly: non-reactive)Pass
    Maximization Sensitization Test(Implicitly: non-sensitizing)Pass
    Genotoxicity(Implicitly: non-mutagenic)Pass
    Implantation(Implicitly: compatible with surrounding tissues upon implantation)Pass
    SterilitySterile (determined by sterilization dose auditing)(Implied: Supports sterility)
    Material ConformanceConforms to USP/NF "Official Monograph for Calcium Sulfate"Conforms
    Mechanical StrengthEquivalent strength to predicate deviceEquivalent strength
    Substantial EquivalenceEquivalence in base materials, function, and intended use to predicate devicesSubstantially Equivalent

    2. Sample Sizes and Data Provenance

    • Test Set Sample Size: Not explicitly stated for each biocompatibility test. The document states "Summaries and reports of all data are contained in the Premarket Notification Submission," which is not provided here.
    • Data Provenance: The studies were conducted by Merries International Inc. (Taiwan, R.O.C.). The type of studies (e.g., animal models for implantation, in vitro for cytotoxicity) implies a mix of experimental data. Whether this data is considered "retrospective" or "prospective" is not applicable in the typical sense for these types of device tests; they are generally tests performed specifically for regulatory submission.

    3. Number of Experts and Qualifications for Ground Truth

    • This information is not applicable as this submission concerns a medical device (bone graft substitute) and its biocompatibility and material properties, not an AI/software device requiring expert interpretation of outputs for ground truth.

    4. Adjudication Method

    • Not applicable for the reasons stated above.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No, an MRMC comparative effectiveness study was not done. This type of study is relevant for AI-powered diagnostic or interpretive tools to assess human reader improvement, which is not the nature of this device.

    6. Standalone (Algorithm Only) Performance Study

    • No, a standalone performance study in the context of an algorithm or AI was not done. This device is a physical bone graft substitute, not a digital algorithm. The "standalone" performance here refers to its physical and biological properties.

    7. Type of Ground Truth Used

    • The "ground truth" for the tests performed can be characterized as:
      • Biocompatibility Standards/Definitions: For tests like cytotoxicity, sensitization, and genotoxicity, the "ground truth" is defined by established scientific protocols and acceptable biological responses.
      • Sterility Standards: For sterility, the "ground truth" is complete absence of viable microorganisms.
      • Material Specifications: For material, the United States Pharmacopoeia (USP) National Formulary (NF) "Official Monograph for Calcium Sulfate" serves as the ground truth.
      • Comparative Performance: For mechanical strength and substantial equivalence, the performance of the predicate devices (Osteoset™ Pellets and Jax™) serves as the "ground truth" or benchmark.

    8. Sample Size for the Training Set

    • Not applicable. This device is not an AI/ML algorithm that requires a "training set."

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable. As above, no training set was used.