Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MedWaves AveCure™ Ablation System is intended for use in percutaneous, laparoscopic, and intraoperative coagulation-ablation of soft tissue.

    The MedWaves AveCure™ Ablation System is not intended for use in cardiac procedures.

    Device Description

    The MedWaves AveCure® Ablation System is designed to deliver ultrahigh radiofrequency (RF) through an antenna placed next to or embedded into soft-tissue for coagulating a volume of that tissue.

    The MedWaves AveCure® Ablation System is composed of a durable generator-controller, and single-patient-use sterile accessories consisting of four antenna sizes and extension cable sets.

    Each antenna-probe assembly is composed of one of the four antennas mounted at the distal end of one of many shaft configurations that are stiff, flexible or combination of both. The extension cable set is used to connect antennas to the generator controller. The material compositions of antennas and extension cables are nonmagnetic in nature and MRI safe. Therefore, the devices can be used in intraoperative, laparoscopic and percutaneous methods with CT, MR and Ultrasound imaging techniques to positions the antenna for ablation of softtissues.

    The MedWaves AveCure® Ablation System is very efficient in transmitting RF power from the generator-controller to the antenna. With the low-power loss, the system does not need cooling pump, CO- or other mechanism to irrigate the antenna shaft and RF transmission lines for patient and user safety.

    AI/ML Overview

    The provided document does not contain acceptance criteria for device performance based on a study, nor does it describe a study that proves the device meets specific performance criteria.

    Instead, the document is a 510(k) summary for the MedWaves AveCure™ Ablation System, indicating that the FDA has determined the device is substantially equivalent to legally marketed predicate devices. The "Performance Data" section primarily discusses compliance with design specifications, risk analysis, and medical device standards (electrical safety-EMC, biocompatibility, sterility, and packaging shelf-life). It makes a general statement about mechanical and coagulation-ablation performance:

    "The mechanical and coagulation-ablation performance ex-vivo and in-vivo meets the intended design and shows to be similar to the predicated devices."

    This statement does not provide specific acceptance criteria or detailed study results. Therefore, I cannot generate the requested table and information.

    To answer your request, I would need a document that explicitly details:

    • Specific, quantifiable acceptance criteria (e.g., minimum ablation zone diameter, maximum ablation time for a certain volume, specific temperature achieved).
    • The methodology and results of a study designed to test these criteria.
    Ask a Question

    Ask a specific question about this device

    K Number
    K083537

    Validate with FDA (Live)

    Date Cleared
    2009-01-15

    (48 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MedWaves Microwave Coagulation/Ablation System is intended for general surgical use in open procedures for the coagulation and ablation of soft tissues. The system is not intended for use in cardiac procedures.

    Device Description

    The MedWaves Microwave Coagulation/Ablation System consists of a microwave generator and accessories used for the surgical ablation of soft tissue. The accessories include two types of sterile, handheld, single use surgical coagulation/ablation probes and required connector cables. The two versions of coagulation/ablation probes are 15-cm in length, 14gauge in diameter, configured into large or small handle. The large handle configuration is the original blue elongated egg-shaped plastic, and the small handle is white plastic handle with 90-degree cable exit (both are made of the same ASB plastic).

    The MedWaves microwave generator operates at a frequency of 902 to 928 MHz. For general surgery applications, the generator is set by the manufacturer to deliver a maximum of 32 W and is user selectable up to 28W. The system is capable of being run in two primary modes: power control mode and temperature control mode. A dynamic monitoring system within the generator maintains the set point selected by the user (either a power or temperature control point) in order to provide consistent and expected coagulation/ablation performance.

    The system incorporates several monitored parameters that alert the user to potentially unsafe conditions during operation. These parameters and condition alerts include: overtemperature, over-power, inefficient microwave delivery, catheter/probe disconnect and system errors. MedWaves microwave generator utilizes firmware to monitor these processes variables and posts the ablation parameters and progress for the user on an LCD screen.

    MedWaves uses proprietary antenna technology coupled with a delivery system that efficiently transmits the microwave energy to the tip of the probe. The microwave antenna is 20 mm long mounted at the distal end. Efficiency of the system relieves the requirement of having high-powered generator in order to create the desired coagulation/ablation results. The efficiency of the system allows the incorporation of monitoring and controls on the microwave energy delivery in order to deliver the most optimal performance available.

    AI/ML Overview

    The provided text describes the MedWaves Microwave Coagulation/Ablation System. I will extract the requested information based on the content of this document.

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria CategorySpecific Standard/RequirementReported Device Performance/Compliance
    Electrical Safety/EMCEN 60601-1 (Medical electrical equipment – Part 1: General requirements for basic safety and essential performance)Complies with EN 60601-1
    EN 60601-1-2 (Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance – Collateral Standard: Electromagnetic compatibility – Requirements and tests)Complies with EN 60601-1-2
    EN 60601-2-2:2006 STD (Medical electrical equipment – Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and accessories)Complies with EN 60601-2-2:2006 STD
    IEC CISPR11 Class A Group 3 (Industrial, scientific and medical equipment – Radio-frequency disturbance characteristics – Limits and methods of measurement)Conforms to IEC CISPR11 Class A Group 3
    EN61000-3-2 (Electromagnetic compatibility (EMC) – Part 3-2: Limits – Limits for harmonic current emissions)Conforms to EN61000-3-2
    EN61000-3-3 (Electromagnetic compatibility (EMC) – Part 3-3: Limits – Limitation of voltage changes, voltage fluctuations and flicker in public low-voltage supply systems, for equipment with rated current <= 16 A per phase and not subject to conditional connection)Conforms to EN61000-3-3
    EN61000-4-2 (Electromagnetic compatibility (EMC) – Part 4-2: Testing and measurement techniques – Electrostatic discharge immunity test)Conforms to EN61000-4-2
    EN61000-4-3 (Electromagnetic compatibility (EMC) – Part 4-3: Testing and measurement techniques – Radiated, radio-frequency, electromagnetic field immunity test)Conforms to EN61000-4-3
    EN61000-4-4 (Electromagnetic compatibility (EMC) – Part 4-4: Testing and measurement techniques – Electrical fast transient/burst immunity test)Conforms to EN61000-4-4
    EN61000-4-5 (Electromagnetic compatibility (EMC) – Part 4-5: Testing and measurement techniques – Surge immunity test)Conforms to EN61000-4-5
    EN61000-4-6 (Electromagnetic compatibility (EMC) – Part 4-6: Testing and measurement techniques – Immunity to conducted disturbances, induced by radio-frequency fields)Conforms to EN61000-4-6
    EN61000-4-8 (Electromagnetic compatibility (EMC) – Part 4-8: Testing and measurement techniques – Power frequency magnetic field immunity test)Conforms to EN61000-4-8
    EN61000-4-11 (Electromagnetic compatibility (EMC) – Part 4-11: Testing and measurement techniques – Voltage dips, short interruptions and voltage variations immunity tests for equipment with input current up to 16 A per phase)Conforms to EN61000-4-11
    BiocompatibilityISO 10993 (Biological Evaluation of Medical Devices)Complies with ISO 10993
    SterilityANSI/AAMI/ISO 11135:1994 (Medical Devices - Validation and Routine Control of Ethylene Oxide Sterilization)Complies with ANSI/AAMI/ISO 11135:1994
    Packaging, Shelf LifeAAMI ANSI ISO 11607-1:2006 (Packaging for Terminally Sterilized Medical Devices)Complies with AAMI ANSI ISO 11607-1:2006
    Firmware TestingAAMI ANSI IEC 62304 (Medical Device Software Software Lifecycle Processes)Underwent firmware testing in accordance with AAMI ANSI IEC 62304
    Risk ManagementAAMI ANSI ISO 14971 & 14971 Al (Medical Devices - Application of Risk Management to Medical Devices)Underwent firmware testing in accordance with AAMI ANSI ISO 14971 & 14971 Al
    Mechanical & In VitroIntended use and published performance of currently marketed productsCatheter/Probes underwent mechanical and in vitro testing; device performed as expected and comparable to published performance of currently marketed products.
    FunctionalDesign specificationsFunctions as intended and meets design specifications through electrical, mechanical, functional, and biocompatibility tests.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not specify a distinct "test set" in the context of clinical data or image analysis. The performance claims are based on compliance with established engineering and safety standards, as well as mechanical and in vitro testing of the probes. There is no mention of patient data, clinical studies, or country of origin for such data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. The provided summary is for a medical device (microwave coagulation/ablation system) and focuses on engineering, safety, and performance standards. There is no mention of experts establishing ground truth for a test set in the context of diagnostic accuracy or clinical interpretation.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable, as no test set requiring expert adjudication for ground truth (e.g., in an image analysis context) is described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a surgical ablation system, not a diagnostic imaging or AI-assisted diagnostic device. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance is outside the scope of the provided information.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI algorithm. The device, a microwave coagulation/ablation system, operates based on physical principles for tissue ablation. Its "performance" refers to its ability to generate microwave energy and ablate tissue as intended, not an algorithmic output.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    For the mechanical and in vitro testing of the catheter/probes, the "ground truth" was likely established through established scientific methods, physical measurements, and comparison to the "published performance of currently marketed products" and "intended use" specifications. For regulatory compliance, the ground truth is adherence to the specified international and national standards (e.g., ISO, EN, AAMI).

    8. The sample size for the training set

    Not applicable. This is not an AI/machine learning device that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K070356

    Validate with FDA (Live)

    Date Cleared
    2007-12-18

    (315 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MedWaves Microwave Coagulation/Ablation System is intended for general surgery use in open procedures for the coagulation and ablation of soft tissues. The system is not intended for use in cardiac procedures.

    Device Description

    The MedWaves Microwave Coagulation/Ablation System consists of a microwave generator and accessories used for the surgical ablation of soft tissue. The accessories include two types of sterile, handheld, single use surgical coagulation/ablation probes and required connector cables. The two versions of coagulation/ablation probes include short and longer length devices. The MedWaves microwave generator operates at a frequency of 902 to 928 MHz. For general surgery applications the generator is set by the manufacturer to deliver a maximum of 32 W and is user selectable up to 28 W. The system is capable of being run in two primary modes: power control mode and temperature control mode. A dynamic monitoring system within the generator maintains the set point selected by the user (either a power or temperature control point) in order to provide consistent and expected coaqulation/ablation performance. The system incorporates several monitored parameters that alert the user to potentially unsafe conditions during operation. These parameters and condition alerts include: high temperature, high-power, inefficient microwave delivery, catheter/probe disconnect and system errors. MedWaves microwave generator utilizes firmware to monitor these processes and post the ablation parameters and progress for the user on an LCD screen. MedWaves uses proprietary antenna technology coupled with a delivery system that efficiently transmits the microwave energy to the tip of the probe. The microwave antenna is 20 mm long mounted at the distal end. Efficiency of the system relieves the requirement of having high powered generator in order to create the desired coagulation/ablation results. The efficiency of the system allows the incorporation of monitoring and controls on the microwave energy delivery in order to deliver the most optimal performance available.

    AI/ML Overview

    MedWaves Microwave Coagulation/Ablation System - Acceptance Criteria and Study Details

    The provided documentation describes the MedWaves Microwave Coagulation/Ablation System, intended for surgical ablation of soft tissue. The "acceptance criteria" discussed are primarily related to general medical device standards and the device's functional performance in a lab setting, rather than clinical efficacy. The study focuses on demonstrating substantial equivalence to predicate devices through technical performance and compliance with relevant standards.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Standard/TestReported Device Performance
    Electrical Safety/EMCIEC 60601-1-2 Medical Electrical Equipment Part 1-2: General Requirements for Safety: Electromagnetic CompatibilityComplies with standard
    IEC 60601-2-2 2007 Standard Requirement TestsComplies with standard
    BiocompatibilityISO 10993, Biological Evaluation of Medical DevicesComplies with standard
    SterilityANSI/AAMI/ISO 11135:1994 Medical Devices - Validation and Routine Control of Ethylene Oxide SterilizationComplies with standard
    Packaging, Shelf LifeAAMI ANSI ISO 11607-1:2006 Packaging for Terminally Sterilized Medical DevicesComplies with standard
    Mechanical PerformanceMedWaves Catheter/Probes underwent mechanical and in vitro testing in accordance with the intended use.Performed as expected and comparable to published performance of currently marketed products.
    In Vitro PerformanceMedWaves Catheter/Probes underwent mechanical and in vitro testing in accordance with the intended use.Performed as expected and comparable to published performance of currently marketed products.
    Firmware TestingAAMI ANSI IEC 62304 Medical Device Software Software Lifecycle . ProcessesComplies with standard
    AAMI ANSI ISO 14971 & 14971 A1 Medical Devices Application of . Risk Management to Medical DevicesComplies with standard
    Substantial EquivalenceComparison to predicate devices (Microsulis Tissue Ablation (MTA) System K052919, Tyco/Radionics Cool-Tip™ RF System K984552, VivaWave™ Microwave Ablation System K011676, VivaWave™ Microwave Ablation System K053535)Device has the same intended use and general technological characteristics; performed comparably to a predicate device with respect to lesion size and tissue temperatures generated. No new questions of safety or efficacy raised.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not explicitly stated as a number of patients or cases. The "test set" primarily refers to physical units of the device (generator, probes, firmware) undergoing laboratory and in vitro testing.
    • Data Provenance: The testing was conducted by MedWaves Incorporated. The context suggests these are internal company tests, typically performed in a controlled laboratory environment. No country of origin for a clinical data set is mentioned, as this is not a clinical study. The data is retrospective in the sense that it evaluates the manufactured device against pre-defined specifications and standards.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • This document describes engineering and regulatory compliance testing, not a clinical study requiring expert consensus on clinical outcomes. Therefore, no "experts" in the sense of clinicians establishing ground truth for a test set of patient data are mentioned. The ground truth for functional performance and compliance is established by the specifications of the relevant international and national standards (e.g., IEC, ISO, AAMI, ANSI) and the device's design specifications. The "experts" would be the engineers and quality assurance personnel responsible for conducting and evaluating these technical tests. Their specific qualifications are not detailed.

    4. Adjudication Method for the Test Set

    • Not applicable. Adjudication methods (like 2+1, 3+1) are used in clinical studies to resolve discrepancies between readers in assessing ground truth. This document pertains to technical performance and regulatory compliance, where outcomes are typically objectively measured against defined criteria.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    • No, an MRMC comparative effectiveness study was not done. This 510(k) submission focuses on demonstrating substantial equivalence through technical performance and compliance with standards, not on clinical comparative effectiveness with human readers.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) was Done

    • Yes, a standalone performance assessment was effectively done for the device's technical aspects. The testing described (electrical safety, biocompatibility, sterility, mechanical, in vitro, firmware) evaluates the device itself (generator, probes, software) without human-in-the-loop clinical interaction. The "device performed as expected and comparable to the published performance of currently marketed products" based on these tests.

    7. The Type of Ground Truth Used

    • The ground truth for this submission is based on:
      • Engineering specifications and design requirements for the device's components and overall function.
      • International and national medical device standards (e.g., IEC 60601, ISO 10993, ANSI/AAMI/ISO 11135, AAMI ANSI ISO 11607, AAMI ANSI IEC 62304, AAMI ANSI ISO 14971).
      • Published performance data of predicate devices used for comparison of "lesion size and tissue temperatures generated" in in vitro or bench testing.

    8. The Sample Size for the Training Set

    • Not applicable. This is not an AI/machine learning device that requires a "training set" of data. The device's "training" in this context would refer to its design and manufacturing process to meet specifications.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable, as there is no training set mentioned for this type of medical device.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1