510(k) Data Aggregation

    K Number
    K111319

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2011-08-05

    (87 days)

    Product Code
    Regulation Number
    892.2050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    IBIS Explorer and Markup Software is a Picture Archiving and Communication System (PACS) intended for acceptance, transfer, display, storage, archive and manipulation of digital medical images, including quantification and report generation.

    Besides general image interpretation and processing tools, IBIS provides specific tool sets for several clinical applications, including:

    • . Oncology, including lesion marking and analysis
    • . Anatomic structure marking, volume measurement and analysis

    Only DICOM for presentation images can be used on an FDA approved monitor for mammography for image diagnosis.

    Only uncompressed or non-lossy compressed images must be used for image diagnosis in mammography.

    Device Description

    The IBIS Explorer and Markup Software is a picture archival and communications system (PACS) providing researchers, radiologists or health care professionals with the tools to accept, display, store, archive and manipulate digital medical images. The IBIS Explorer stores, archives and displays digital medical images, while the Markup Software allows for measurements of specified regions and report generation. Digital images supported by the software include but are not limited to bone scans, chest radiograph (x-ray), ultrasound, computed tomography (CT), and magnetic resonance imaqing (MRI).

    The System is to be used by trained professionals who are responsible for the correct and accurate use of medical images and the subsequent diagnosis. In a typical clinical environment, the system would be used by a trained technician and radiologist.

    The software is installed on an off-the-shelf PC computer system. It is intended to be used with uncompressed digital images that are saved in DICOM format.

    AI/ML Overview

    Here's an analysis of the provided text regarding the IBIS Explorer and Markup Software, focusing on acceptance criteria and the study that proves device performance:

    1. Table of Acceptance Criteria and Reported Device Performance

    Based on the provided text, the acceptance criteria are not explicitly numerical thresholds but rather functional and validation requirements. The device performance is generally stated as meeting these requirements.

    Acceptance Criteria CategorySpecific Criteria (Implied)Reported Device Performance
    Functional VerificationDevice meets all software requirements."Comprehensive design verification and validation (V&V) was performed to ensure that the IBIS Explorer and Markup Software meets its intended use, customer requirements and software requirements."
    User Requirements ValidationDevice meets customer/user requirements."Testing also included verifying the accuracy of the measurements of the Markup Software against the Visage PACS/CS predicate device."
    Measurement AccuracyMarkup Software measurements are accurate (relative to predicate)."Testing also included verifying the accuracy of the measurements of the Markup Software against the Visage PACS/CS predicate device."
    Image Modality SupportSupports various DICOM image modalities from different scanners."Testing was conducted using various DICOM image modalities from different scanners (GE, Toshiba, Philips and Siemens)."
    Clinical Usability/QualityNo anomalies affecting quality or clinical usability."Testing confirmed that no anomalies affect the quality or clinical usability of the system."
    Substantial EquivalenceSame intended use, similar technological characteristics, no new safety/effectiveness questions."The IBIS Explorer and Markup Software is substantially equivalent to the predicate devices. The software has the same intended use and similar technological characteristics, which do not raise any new questions of safety or effectiveness."

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: The document does not explicitly state a sample size for the test set of images used for verification and validation. It mentions "various DICOM image modalities."
    • Data Provenance: The document states that testing was "conducted using various DICOM image modalities from different scanners (GE, Toshiba, Philips and Siemens)." This suggests a diverse set of images, likely retrospective data collected from these different scanner types. The country of origin of this data is not specified, but the testing itself was "conducted at MedQIA."

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not provide information on the number of experts used or their qualifications for establishing ground truth for the test set. The testing primarily focuses on software functionality, compliance with requirements, and comparison of measurement accuracy against a predicate device.

    4. Adjudication Method for the Test Set

    The document does not describe any adjudication method for the test set. Given the nature of the testing described (functional verification and comparison against a predicate device for measurement accuracy), it seems unlikely an adjudication method for case interpretation or diagnosis was employed in the usual sense.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done

    No, a Multi Reader Multi Case (MRMC) comparative effectiveness study was not done. The document describes design verification and validation, including functional testing and measurement accuracy comparison to a predicate device. It does not mention any study involving human readers or comparing performance with and without AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, the testing described appears to be a form of standalone testing for the functionality and measurement accuracy of the software itself. The phrase "verifying the accuracy of the measurements of the Markup Software against the Visage PACS/CS predicate device" implies testing the algorithm's output directly. However, the device is intended for use by trained professionals, implying a human-in-the-loop for interpretation and diagnosis, not the algorithm making diagnoses independently. The clinical usability was also assessed, which would be in the context of human use. So, while foundational testing was standalone, the overall clinical use is human-interfaced.

    7. The Type of Ground Truth Used

    For the "accuracy of the measurements of the Markup Software," the ground truth was essentially the measurements provided by the predicate device (Visage PACS/CS). This means the predicate device served as the reference standard for measurement accuracy. For other functional aspects, the ground truth was the expected output based on software and user requirements. There is no mention of pathology, expert consensus, or outcomes data being used as ground truth for this particular submission.

    8. The Sample Size for the Training Set

    The document does not mention a training set or any machine learning components. The device is described as a PACS system with image manipulation and measurement tools, not an AI-driven diagnostic algorithm that would typically require a training set. The "Performance Data" section describes Verification & Validation activities, not algorithm training.

    9. How the Ground Truth for the Training Set Was Established

    As no training set is mentioned or implied for this device, this question is not applicable based on the provided text.