510(k) Data Aggregation

    K Number
    K013725

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2002-01-30

    (82 days)

    Product Code
    Regulation Number
    870.1875
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Intended for use as a diagnostic aid in patient diagnosis, treatment, and monitoring. It amplifies sounds from the body's internal organs, mainly the heart and lungs. Included with the system is a manual visual display of the electrocardiogram (ECG) on the PC monitor to enable the healthcare practitioner to synchronize the phonocardiogram with the beginning of the heart cycle. The ECG is not intended for diagnostic use. The system also includes computer-aided recordings and storage of these sounds along with other patient information.

    Device Description

    The Meditron electronic stethoscope system is intended for use as a diagnostic aid in patient diagnosis, treatment and monitoring. It amplifies, records, plays back, and transmits sounds associated with the heart, arteries, and veins and other internal organs. The modification involves the addition of a 5-lead electrocardiograph to produce an ECG on the PC monitor to enable the healthcare practitioner to synchronize the phonocardiogram with the beginning of the heart cycle.

    AI/ML Overview

    This 510(k) summary for the Meditron II Thestethoscope System (K013725) unfortunately lacks detailed information regarding specific acceptance criteria and the comprehensive study results that would typically prove the device meets these criteria. The document focuses on demonstrating substantial equivalence to predicate devices rather than providing a standalone performance validation report.

    However, based on the provided text, here's what can be extracted and what information is missing:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Acceptance CriteriaReported Device Performance
    Biocompatibility(Not specified)"specifications and testing for biocompatibility... were met"
    Electrical Safety(Not specified)"specifications and testing for... electrical safety... were met"
    Electromagnetic Compatibility (EMC)(Not specified)"specifications and testing for... electromagnetic compatibility (21.20)... were met"
    Product Specifications(Not specified)"specifications and testing for... product specifications were met"
    Software Verification/Validation(Not specified)"software verification/validation reference applicable FDA Consensus started [standards]... were met"
    Clinical EquivalenceComparable performance to legally marketed electronic stethoscopes with an integrated ECG."Clinical study information was gathered for the purpose of demonstrating substantial equivalence to legally marketed electronic stethoscopes with an integrated ECG."

    Missing Information:

    • Specific quantitative or qualitative acceptance criteria for each category (e.g., specific thresholds for biocompatibility tests, electrical leakage limits, EMC emission/immunity levels, software robustness metrics).
    • Detailed results or metrics showing how these criteria were met for each category.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not explicitly stated. The document mentions "Clinical study information was gathered for the purpose of demonstrating substantial equivalence," but it does not specify the number of patients or cases included in this study.
    • Data Provenance: Not explicitly stated. It's unclear if the data was retrospective or prospective, or from which country.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    • Number of Experts: Not specified.
    • Qualifications of Experts: Not specified.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not specified.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    • MRMC Study: No, an MRMC comparative effectiveness study is not noted as being performed. The clinical information gathered was for the purpose of demonstrating substantial equivalence to predicate devices, not for comparing human reader performance with and without AI assistance.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    • Standalone Performance: Not explicitly reported. The device is described as a "diagnostic aid" for a healthcare practitioner, implying a human-in-the-loop context. The summary does not provide any isolated algorithm performance metrics.

    7. The Type of Ground Truth Used

    • Type of Ground Truth: Not explicitly stated in detail for the clinical study. Given the objective of demonstrating substantial equivalence to other electronic stethoscopes with integrated ECGs, the "ground truth" would likely be derived from a comparison to clinical judgment and/or established diagnostic methods, but specifics are absent.

    8. The Sample Size for the Training Set

    • Sample Size for Training Set: Not applicable/Not mentioned. This device does not appear to employ a machine learning algorithm that requires a "training set" in the conventional sense. Its function is to amplify, record, play back, and transmit sounds, and to produce an ECG for synchronization. It's an electronic measurement and display device, not an AI-driven diagnostic system.

    9. How the Ground Truth for the Training Set Was Established

    • Ground Truth for Training Set: Not applicable/Not mentioned, as there is no apparent training set for an AI algorithm.

    Summary of what the document does convey regarding acceptance and studies:

    The primary "study" mentioned is the gathering of "clinical study information" to demonstrate substantial equivalence to existing legally marketed electronic stethoscopes with integrated ECG. The acceptance criteria, while not quantified, seem to be met if:

    1. The device adheres to established standards for biocompatibility, electrical safety, electromagnetic compatibility, and product specifications.
    2. The software is verified and validated.
    3. Its indications for use and technological characteristics are "minor" and do not raise new safety or effectiveness issues compared to predicate devices.

    The 510(k) pathway relies on demonstrating that a new device is as safe and effective as a legally marketed predicate device, rather than requiring extensive de novo clinical trials to prove efficacy against a specific "gold standard" or to quantify AI performance improvements. The provided text strongly aligns with this approach.

    Ask a Question

    Ask a specific question about this device

    K Number
    K991367

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1999-07-19

    (90 days)

    Product Code
    Regulation Number
    870.1875
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Meditron Stethoscope System is intended for use as a diagnostic aid in patient diagnosis, treatment and monitoring. It amplifies sounds from the bodys internal organs, mainly the heart and respitatory and circulatory organs. It provides computer-aided recordings of these sounds and stores these recordings along with other appropriate patient information.

    Device Description

    The Meditron stethoscope system is intended for use as a diagnostic aid in patient diagnosis, treatment and monitoring. It consists of four major components: electronic stethoscope (thestethoscope), distributor (thedistributor), connector (thecomnector), and CD software (theanalyzer). Thestethoscope amplifies sounds from the body's internal organs, mainly the heart and respiratory and circulatory organs, without introducing signals or energy into the body. Thedistributor is a co-listening device that has been specially developed for teaching purposes as it permits up to six persons to listen to the patient at the same time. Thedistributor can also be used during maternity check-ups to permit the physician and expectant mother to simultaneously listen to the sounds. Theconnector (connection box and cables) links thedistributor to the sound card of a personal computer (PC). Theanalyzer is a Microsoft Windows-based CD software application supplied with theconnector and is designed to provide computeraided recordings with the electronic stethoscope and to store these recordings along with other appropriate patient information.

    AI/ML Overview

    The provided 510(k) summary for the Meditron Stethoscope System primarily focuses on demonstrating substantial equivalence to predicate devices through technical specifications and non-clinical testing. It explicitly states that clinical study information was not submitted for the purpose of demonstrating substantial equivalence.

    Therefore, based on the provided document, the following answers reflect this lack of clinical study data to prove acceptance criteria:

    1. A table of acceptance criteria and the reported device performance:

      Acceptance CriterionReported Device Performance
      Not applicableNo clinical acceptance criteria or performance data are provided, as no clinical study was conducted for this purpose.
    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
      No clinical study data was submitted for demonstrating substantial equivalence. Therefore, no test set size or data provenance is available.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
      No clinical study data was submitted for demonstrating substantial equivalence. Therefore, no information on experts or ground truth establishment is available.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
      No clinical study data was submitted for demonstrating substantial equivalence. Therefore, no adjudication method is described.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
      No clinical study data was submitted for demonstrating substantial equivalence, and the device is an electronic stethoscope system, not explicitly framed as an AI-assisted diagnostic tool for human readers in this document. Therefore, no MRMC study or effect size is reported.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
      No clinical study data was submitted for demonstrating substantial equivalence. While the device includes "computer-aided recordings," the submission does not describe a standalone algorithm performance study.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
      No clinical study data was submitted for demonstrating substantial equivalence. Therefore, no ground truth type is specified.

    8. The sample size for the training set:
      No clinical study data was submitted for demonstrating substantial equivalence, and no machine learning model training is described for a diagnostic purpose requiring such a set.

    9. How the ground truth for the training set was established:
      No clinical study data was submitted for demonstrating substantial equivalence, and no machine learning model training is described for a diagnostic purpose.

    Summary of Clinical Data Absence:

    The Meditron Stethoscope System's 510(k) submission (K991367) explicitly states: "Clinical study information was not submitted for the purpose of demonstrating substantial equivalence to legally marketed electronic stethoscopes."

    Instead, substantial equivalence was primarily demonstrated through:

    • Comparison of Indications for Use: The Meditron system's indications were consistent with predicate devices and the FDA classification regulation (21 CFR 870.1875(b)).
    • Non-Clinical Testing: Biocompatibility, environmental testing (climatic, mechanical, enclosure protection), and electromagnetic compatibility (EMC) tests were conducted according to international standards (ISO-10993, IEC, CISPR, EN). These tests confirmed that product specifications were met and that differences in technological characteristics did not raise new safety or effectiveness concerns.

    Therefore, the acceptance criteria for this specific substantial equivalence determination were met through non-clinical performance and a comparison of device characteristics and intended use to legally marketed predicate devices, rather than through a clinical study demonstrating diagnostic performance.