Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K120577

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2015-06-24

    (1213 days)

    Product Code
    Regulation Number
    868.1840
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Medikro Spirometer is a device that measures lung air volume and airflow rate for pulmonary disease diagnosis and screening. These measurements provide information about a patient's pulmonary function which may be compared with normal values or the patient's previous values.

    The device is designed for

    • . adult and pediatric patients,
    • hospital and clinic use only.
    Device Description

    Medikro® SpiroStar spirometer runs on a personal computer with Microsoft Windows operating systems.

    Medikro® SpiroStar spirometer unit is connected to PC via USB port (USB model) or via serial port (DX model). SpiroSafe disposable flow transducer is connected to the spirometer unit via pressure tube. A nose clip is used to prevent air flow from nose during measurements. Optional calibration syringe is used for recommended daily volume calibrations.

    Medikro® SpiroStar spirometers are used to measure lung air volume and airflow rate. Medikro® Spirometry Software is used to perform the measurement and calculate the measurement volume based on chosen reference value. The reference value is based on patient's gender, race and age. User can analyze the results in different presentations and create a final report based on the results and patient information.

    The spirometers take all the power that it needs from the USB or serial port, so no other external or internal power supply is needed.

    AI/ML Overview

    The Medikro® SpiroStar device, including models M929 (USB) and M921 (DX), is a diagnostic spirometer intended to measure lung air volume and airflow rate for pulmonary disease diagnosis and screening. The device's performance was evaluated against the American Thoracic Society (ATS) recommendations for spirometers to demonstrate substantial equivalence to predicate devices.

    1. A table of acceptance criteria and the reported device performance

    Performance Criteria (ATS Recommendations)Acceptance Criteria for Spirometers (ATS)Reported Device Performance (Medikro® SpiroStar USB & DX)
    Measuring FVCMeets ATS recommendationsMeets ATS recommendations
    Measuring FEV1Meets ATS recommendationsMeets ATS recommendations
    Measuring FEF25-75%Meets ATS recommendationsMeets ATS recommendations
    Measuring PEFMeets ATS recommendationsMeets ATS recommendations
    Resistance to flowMeets ATS recommendationsMeets ATS recommendations
    Testing under BTPS conditionsMeets ATS recommendationsMeets ATS recommendations

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not explicitly state the sample size for the test set in terms of number of patients or spirometry maneuvers. Instead, it refers to "American Thoracic Society (ATS) 24 standard waveforms test". This suggests the use of a standardized set of simulated waveforms rather than patient data. The provenance of these standard waveforms is the American Thoracic Society. It is a retrospective evaluation against established synthetic data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    The ground truth for this study was established by the "American Thoracic Society (ATS) 24 standard waveforms". These are not "expert-generated ground truth" in the typical sense of a clinical interpretation, but rather predefined, standardized synthetic waveforms with known, precise values for spirometry parameters. Therefore, no human experts were involved in establishing the "ground truth" for the test set in this context.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. The test involved comparing the device's measurements against the known values of the ATS 24 standard waveforms. This is a direct measurement comparison, not an expert-based adjudication process.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a performance study for a diagnostic spirometer, which directly measures physiological parameters. It is not an AI-based interpretation system, and therefore, no MRMC study or evaluation of human reader improvement with AI assistance was performed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, a standalone performance evaluation was conducted. The Medikro® SpiroStar units (USB and DX) were tested directly against the ATS 24 standard waveforms to determine if their measurements met the specified recommendations. This is an algorithm-only (device-only) performance assessment.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The type of ground truth used was standardized synthetic waveforms with known, precise values, specifically the American Thoracic Society (ATS) 24 standard waveforms.

    8. The sample size for the training set

    The document does not mention a training set. This type of device (diagnostic spirometer) typically relies on engineering and physiological principles for its design and calibration, rather than machine learning models that require training data. The testing described is a validation of the device's accuracy against established standards.

    9. How the ground truth for the training set was established

    Not applicable, as no training set or machine learning components are described in the context of this device's performance evaluation.

    Ask a Question

    Ask a specific question about this device

    K Number
    K133428

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2014-05-21

    (194 days)

    Product Code
    Regulation Number
    868.1840
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Medikro® Spirometer is a device that measures lung air volume and airflow rate for pulmonary disease diagnosis and screening. These measurements provide information about a patient's pulmonary function which may be compared with normal values or the patient's previous values.

    The device is designed to test pulmonary function and obtain spirometric indices for

    • adult and pediatric patients 12 years and older,
    • hospital and clinic use only.
    Device Description

    Medikro® spirometer runs on a personal computer with Microsoft Windows operating systems.

    Medikro@ spirometer unit is connected to PC via USB port (USB model). SpiroSafe disposable flow transducer is connected to the spirometer unit via pressure tube. Nose clip is used to prevent air flow from nose during measurements. Optional calibration syringe is used for recommended daily volume calibrations.

    Medikro® spirometers are used to measure lung air volume and airflow rate. Medikro® Spirometry Software is used to perform the measurement and calculate the measurement volume based on chosen reference value. The reference value is based on patient's gender, race and age. User can analyze the results in different presentations and create a final report based on the results and patient information.

    The spirometers take all the power that it needs from the USB port, so no other external or internal power supply is needed.

    AI/ML Overview

    The Medikro® Spirometer (M913 Medikro® Nano, M914 Medikro® Primo, M915 Medikro® Pro) is designed to measure lung air volume and airflow rate for pulmonary disease diagnosis and screening. The devices were tested to ensure they meet the essential performance criteria for spirometers as defined by the American Thoracic Society (ATS) and ISO 26782 standards.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (ATS/ERS & ISO 26782 Standards)Reported Device Performance (Medikro® Spirometers)
    ATS/ERS Standards (Eur Respir J, 2005, Vol 26, pp. 948-968):All units meet the recommendations for:
    Measuring FVCMet
    Measuring FEV1Met
    Measuring FEF25-75%Met
    Measuring PEFMet
    Measuring MVVMet
    Resistance to flowMet
    ISO 26782:2009 Standards:All units meet the ISO 26782 requirements for:
    Measuring FVC accuracy and repeatabilityMet
    Measuring FEV1 accuracy and repeatabilityMet
    Measuring FEV6 accuracy and repeatabilityMet
    LinearityMet
    ImpedanceMet
    Safety and EMC Standards:
    AAMI / ANSI ES 60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012 (Consolidated Text) Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD)Complied
    IEC 60601-1-2 Edition 3:2007-03, Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and testsComplied

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample size in terms of number of human subjects or unique cases for the performance testing. Instead, the testing involved exposing all three device models (M913 Medikro® Nano, M914 Medikro® Primo, and M915 Medikro® Pro) to standardized conditions using "24 standard waveforms" as specified by the American Thoracic Society (ATS) and further tests as per ISO 26782.
    The data provenance is not explicitly stated as "country of origin" for the ATS/ERS and ISO 26782 standard waveforms. These are international standards, implying a general applicability rather than specific regional data. The testing itself was performed at CSA Group CB testing laboratory and Savonia Polytechnic EMC laboratory, both located in Finland, which is the manufacturer's country of origin. The study appears to be prospective in nature, as it involved physical testing of the manufactured devices against predefined international standards.

    3. Number, Qualifications, and Adjudication Method of Experts

    This study did not involve human experts to establish ground truth for clinical performance in the traditional sense (e.g., radiologists interpreting images). Instead, the "ground truth" or reference for the device's performance was established by adherence to universally accepted engineering and medical device standards for spirometry.

    • Experts: Not applicable for establishing ground truth of the test set in this context. The "experts" in this scenario would be the bodies that define the ATS/ERS and ISO standards, which are internationally recognized professional organizations in respiratory medicine and standardization.
    • Qualifications: Not applicable.
    • Adjudication method: Not applicable.

    4. Adjudication Method (for the test set)

    Not applicable. The performance was assessed against objective, predefined values and ranges specified by the ATS/ERS and ISO 26782 standards. There was no expert adjudication process in this technical validation.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not conducted as this device is a diagnostic spirometer, not an AI-assisted diagnostic tool for interpretation of human expert readings. The study focuses on the technical performance of the device in accurately measuring lung parameters, not on how human readers' interpretations improve with or without AI assistance.

    6. Standalone Performance Study

    Yes, a standalone performance study was conducted. The performance of the Medikro® spirometers was evaluated in isolation (algorithm only, as they calculate measurements based on raw data) against established international standards (ATS/ERS and ISO 26782) without human intervention in the measurement process. The device's ability to accurately measure FVC, FEV1, FEF25-75%, PEF, MVV, and FEV6, as well as its linearity and impedance characteristics, were tested directly against these benchmarks.

    7. Type of Ground Truth Used

    The ground truth used was based on internationally recognized performance standards and simulated waveforms. Specifically:

    • Standardized Waveforms: The "24 standard waveforms" from the American Thoracic Society (ATS) served as the reference for dynamic performance.
    • Engineering Specifications: The ISO 26782 standard provides precise numerical acceptance criteria for accuracy, repeatability, linearity, and impedance.
    • Safety and EMC Standards: AAMI / ANSI ES 60601-1 and IEC 60601-1-2 define the ground truth for electrical safety and electromagnetic compatibility.

    8. Sample Size for the Training Set

    The document does not mention a "training set" in the context of machine learning or AI algorithms. The device performs calculations based on established physiological formulas and measurement principles, not through a learned model from a training dataset. Therefore, this question is not applicable to the described device.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no mention of a training set for machine learning. The device's operation relies on known physics and physiological models, not on a data-trained algorithm.