Found 2 results

510(k) Data Aggregation

    K Number
    K132921

    Validate with FDA (Live)

    Device Name
    GALAXY PLUS
    Date Cleared
    2014-01-17

    (122 days)

    Product Code
    Regulation Number
    892.1680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Galaxy Plus Digital Radiography System is indicated for use in generating radiographic images of human anatomy. This device is not intended for mammographic applications.

    Device Description

    The digital diagnostic x-ray system consists of a generator, a tube, an x-ray console, a beam limiting device, an image receptor, a u-arm stand and software. The digital panels supplied are intended to replace radiographic film/screen systems in all general purpose diagnostic procedures. This device is not intended for mammographic applications. Product features: This system is designed to maintain the alignment between a tube and the image receptor, regardless of the angle on the image receptor or image tilt positions. This typical configuration permits a qualified/trained doctor or technologist to take a range of head-to-toe diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts on patients. Its flexibility makes the system ideal for all patient examinations including the standing, sitting and lying patient positions.

    AI/ML Overview

    The provided 510(k) summary for K132921 describes the Galaxy Plus Digital Radiography System.

    Here's an analysis based on the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The 510(k) summary does not explicitly state quantitative "acceptance criteria" in terms of performance metrics (e.g., sensitivity, specificity, resolution targets) for the device. Instead, the primary acceptance criterion appears to be demonstrated substantial equivalence to the predicate device (DRGEM Corporation's DIAMOND-5A, 6A, 8A, K102408) through non-clinical and clinical testing, and the ability to produce images of "good diagnostic quality."

    Given this, the table would look like this:

    Acceptance Criteria CategorySpecific Criteria (Implicit/Explicit)Reported Device Performance
    Electrical SafetyCompliance with EN/IEC 60601-1: 2006 (for Stand & System)Test reports submitted; unit successfully underwent electrical safety testing.
    EMCCompliance with IEC 60601-1-2:2007, EN/IEC 60601-1-2:2007, CISPR 11:2009/A1:2010, IEC 61000-3-2:2005/A1:2008/A2:2009, IEC 61000-3-3:2008Test reports submitted; unit successfully underwent electromagnetic compatibility testing.
    Radiation ProtectionCompliance with EN/IEC 60601-1-3; EN/IEC 60601-2-28; EN/IEC 60601-2-54Test reports submitted; compliance demonstrated.
    Software ValidationSoftware validation conducted.Software validation performed; updated software (previously cleared in K093816) employed.
    Risk AnalysisRisk analysis conducted.Risk analysis performed.
    BiocompatibilityPerformed in accordance with applicable ISO standards.Biocompatibility testing performed.
    Image Quality (Clinical)Images for all four panels to be of "good diagnostic quality" as evaluated by board-certified radiologists, and obtained in accordance with FDA Guidance Document on Solid State Imaging Devices. The device is "as safe, as effective, and performs as well as or better than" the predicate device.Clinical images for all four panels obtained, evaluated by board-certified radiologists, and found to be of "good diagnostic quality." This implicitly means the device performs at least as well as the predicate in producing diagnostically acceptable images for its intended use (generating radiographic images of human anatomy, excluding mammography).
    Non-clinical PerformanceDemonstrated by respective panel manufacturers.Non-clinical performance tests provided by the panel manufacturers.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document does not specify a numerical sample size for the clinical test set (i.e., the number of patients or images). It vaguely states "Clinical images for all four panels were obtained."
    • Data Provenance: Not explicitly stated (e.g., country of origin). It states the images were obtained "in accordance with the FDA Guidance Document on Solid State Imaging Devices," implying they met U.S. regulatory standards for data collection, but no further details on location or retrospective/prospective nature are provided. It is likely retrospective or a limited prospective case series given the summary nature.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • Number of Experts: Not specified. The document states "They were evaluated by board certified radiologist." This could imply one or more radiologists.
    • Qualifications: "board certified radiologist." The experience level in years is not provided.

    4. Adjudication Method for the Test Set

    The document does not describe an adjudication method (e.g., 2+1, 3+1, none). It simply states the images "were evaluated by board certified radiologist." This suggests either independent evaluations or a consensus approach, but the specific process is not detailed.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No. The document does not mention an MRMC study. The comparison is primarily against the predicate device's technological characteristics and the general standard of "good diagnostic quality" for radiographic images, rather than a comparative effectiveness study with human readers.

    6. Standalone Performance

    Yes, a standalone evaluation was performed. The clinical images generated by the algorithm/system only (i.e., the Galaxy Plus Digital Radiography System itself) were evaluated by radiologists for their diagnostic quality. This evaluation was independent of human interpretation improvements; it assessed the quality of the raw output from the device.

    7. Type of Ground Truth Used

    The ground truth used for the clinical evaluation was expert consensus/evaluation by "board certified radiologist[s]" who determined the images were of "good diagnostic quality." There is no mention of pathology, outcomes data, or other objective measures of truth. For a diagnostic imaging device like an X-ray system, "good diagnostic quality" determined by an expert is a common form of ground truth for substantial equivalence.

    8. Sample Size for the Training Set

    The document does not mention a "training set" or provide any sample size for training. This device is a digital radiography system, not an AI/ML algorithm that typically requires a large training set in the way a diagnostic AI would. The software components are an update of previously cleared software (K093816), implying its core functionality was already established. The emphasis is on the image acquisition hardware and its ability to produce images.

    9. How the Ground Truth for the Training Set Was Established

    As no training set is mentioned in the context of this 510(k) summary (because it's not a de novo AI/ML solution), the method for establishing ground truth for a non-existent training set is not applicable here. The software is an update of a previously cleared version, suggesting its underlying principles and performance were evaluated during its initial clearance (K093816).

    Ask a Question

    Ask a specific question about this device

    K Number
    K093816

    Validate with FDA (Live)

    Date Cleared
    2010-10-05

    (298 days)

    Product Code
    Regulation Number
    892.1680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Galaxy 900 is integrated into the user's stationary radiography system. This typical configuration permits a qualified/trained doctor or technologist to take a range of head-to-toe diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts on both adult and pediatric patients. Applications can be performed with patient sitting, standing or lying in the prone or supine positions in an integrated system. The Galaxy 900 is not intended for mammography.

    The Galaxy 1600 Digital CCD X-Ray Detector is indicated for digital imaging solution designed for general radiographic system for human anatomy. It is intended to replace film or screen based radiographic systems in all general purpose diagnostic procedures. The Galaxy 1600 is not intended for mammography.

    Device Description

    The Digital Radiography System, Galaxy is an optical based digital x-ray imager. This device should be integrated with an operating PC. The visible light is deflected by a mirror to a high-resolution CCD camera that produces a digital image. It can do to utilize as digitalizing x-ray images and transfer for radiography diagnostic. The BLADE software acquires X-ray image and viewing the image. Get image from detector, process it to ease the diagnostic, save it in database and manage it.

    AI/ML Overview

    The provided text is a 510(k) summary for the Medien International Co., Ltd. Digital Radiography System, Galaxy 900 and Galaxy 1600. It details the device's classification, features, and intended use, along with comparisons to predicate devices and declarations of safety and effectiveness. However, it does not contain the specific information required to answer all the questions about acceptance criteria, detailed study design, sample sizes, expert qualifications, or ground truth establishment.

    Here's a breakdown of what can be extracted and what information is missing based only on the provided text:


    1. Table of Acceptance Criteria and Reported Device Performance

    The 510(k) summary does not explicitly list specific acceptance criteria in a quantitative table or directly report performance metrics against such criteria. It states that:

    • "Electrical, mechanical, environmental safety and performance testing according to standard IEC 60601-1 was performed."
    • "EMC testing was conducted in accordance with standard IEC 60601-1-2(2001)."
    • "Biocompatibility testing was conducted in accordance with Standard ISO 10993-1."
    • "Non-clinical & Clinical considerations according to FDA Guidance 'Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices' was performed."
    • "All test results were satisfactory."

    The "Product features" section for each device (Galaxy 900 and Galaxy 1600) lists technical specifications, which implicitly serve as performance requirements. Assuming "satisfactory" means these specifications were met, the table can be constructed using these features:

    Acceptance Criteria / SpecificationGalaxy 900 Reported PerformanceGalaxy 1600 Reported Performance
    Imaging Area17" x 17"17" x 17"
    Digitization16-bit16-bit
    Operating SystemMS Windows XPMS Windows XP
    DICOM ComplianceDICOM3.0DICOM3.0
    Image Format (Pixels)3,056 x 3,0564,096 x 4,096
    High Resolution Image (lp/mm)3.5 lp/mm4.6 lp/mm
    Safety StandardsIEC 60601-1 (Satisfactory)IEC 60601-1 (Satisfactory)
    EMC StandardsIEC 60601-1-2 (2001) (Satisfactory)IEC 60601-1-2 (2001) (Satisfactory)
    Biocompatibility StandardsISO 10993-1 (Satisfactory)ISO 10993-1 (Satisfactory)

    2. Sample size used for the test set and the data provenance

    The document does not specify the sample size for any test set or the data provenance (country of origin, retrospective/prospective). It generally refers to "Non-clinical & Clinical considerations" as being performed.


    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The document does not provide information on the number or qualifications of experts used to establish ground truth for any test set.


    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    The document does not mention any adjudication method for test sets.


    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    The document makes no mention of an MRMC study or any AI assistance for human readers. This device is a Digital Radiography System, a hardware component for acquiring images, not an AI-based diagnostic tool.


    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The document does not describe any standalone algorithm performance testing. The device itself is a digital radiography system, not an algorithm.


    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The document does not specify the type of ground truth used or how it was established for any clinical or non-clinical considerations.


    8. The sample size for the training set

    The document does not mention a training set, as the device is a hardware imaging system, not a machine learning algorithm that typically requires a training set.


    9. How the ground truth for the training set was established

    Not applicable, as no training set is mentioned for this hardware device.