Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K991294

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1999-04-23

    (8 days)

    Product Code
    Regulation Number
    880.5430
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The over the counter Medi-Jector Choice Needle-Free Insulin Delivery System is intended for the subcutaneous injection of U-100 insulin. The injector is labeled as suitable for administration of 2 to 50 insulin units in a 0.5 unit increment.

    Device Description

    Needle-Free Insulin Delivery System

    AI/ML Overview

    This submission is for a labeling change only, specifically to allow the Medi-Jector Choice Needle-Free Insulin Delivery System to be sold over the counter. No new studies were performed or are included in this document to demonstrate device performance against acceptance criteria.

    Therefore, I cannot provide information on acceptance criteria, reported device performance, sample sizes for test or training sets, expert qualifications, adjudication methods, MRMC studies, or standalone algorithm performance, as these items are typically associated with studies demonstrating device efficacy or safety, which are absent in this submission.

    The document explicitly states:

    • "Summary of Nonclinical Tests: None were performed in support of this submission. This request is for a labeling change only." (Section D, Page 0)

    This implies that the device's performance characteristics were previously established and deemed acceptable under its prior 510(k) clearance (K962956), and this submission does not involve re-evaluating those performance aspects.

    Ask a Question

    Ask a specific question about this device

    K Number
    K962956

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1996-10-07

    (69 days)

    Product Code
    Regulation Number
    880.5430
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Medi-Jector Choice Needle-Fee Insulin Delivery System is intended for the subcutaneous injection of U-100 insulin.

    Device Description

    The Medi-Jector Choice Needle-Free Insulin Delivery System consists of a spring powered windable power pack injector assembly, a detachable sterile reusable needle-free syringe and a sterile adapter for connecting the needle-free syringe to a drug the vial. Spring energy is provided by winding the power pack to compress the spring. The power pack also contains a dial mechanism that defines the dosage to be injected. The needle-free syringe consists of a chamber or barrel and a plunger. Both the chamber and the plunger connect to the power pack. One end of the syringe barrel or chamber reduces to a micro-orifice and comes in contact with the injection site; the other end of the syringe connects to the power pack. The adapter consists of a plastic body and seal and connects the needle-free syringe on the power pack to the drug vial. An elastomeric seal in the adapter is compressed against the syringe tip aligning at the micro-orifice to make connection; the connection with the drug vial is by means of plastic spike capable of puncturing the vial septum. Additionally, the adapter end connection with the drug vial has locking tabs to stabilize the vial and to promote the discarding of the adapter with the vial (which is the intended usage).

    AI/ML Overview

    Here's an analysis of the provided text, focusing on the acceptance criteria and the study details:

    Device: Medi-Jector Choice Needle-Free Insulin Delivery System

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific CriteriaReported Device Performance
    Dosage DeliveryDelivered dosage within ISO insulin syringe standards for both low (10 units) and high (50 units) volumes."Dosages delivered... was well within the ISO standards and was therefore deemed to be safe." (Tested over 30 injections, exceeding 100% of indicated usage.)
    Delivered dosage within ISO insulin syringe standards over 14 days, with two ejections per day, using three common insulin types (R, NPH, Ultralente)."Dosages delivered at 100% over the indicated usage were found to be safely within the ISO standards for all types of insulin."
    PenetrationPenetration of saline streams into an artificial test matrix (Whatman #3 filter paper) comparable to a cited predicate device (Medi-Jector Needle-Free Bio-Tropin™ Drug Delivery System)."The penetration of saline streams produced by the Medi-Jector Choice Needle-Free Insulin Delivery Systerm into an artificial test matrix (Whatman #3 filter paper) was comparable to that seen with cited predicate device."

    2. Sample Size and Data Provenance

    • Test Set Sample Size:
      • Dosage Delivery:
        • "thirty (30) injections" for the initial saline test (more than 100% of indicated usage for a single syringe).
        • "over the course of 14 days, with two ejections per day," for insulin types (totaling 28 injections per insulin type, or 84 injections across the three insulin types).
      • Penetration: Not explicitly stated, but implies a sufficient number of tests to establish comparability.
    • Data Provenance: Not explicitly stated, but given the nature of the device (insulin delivery) and the regulatory submission, the testing would likely have been conducted in a controlled laboratory environment. There is no mention of country of origin or whether it was retrospective or prospective, but these types of tests are typically prospective.

    3. Number of Experts and Qualifications for Ground Truth

    • Not Applicable. This device is a mechanical fluid injector. Its performance is evaluated against objective, measurable physical criteria (ISO standards for dosage delivery, physical comparison for penetration). There is no "ground truth" established by human experts in the context of diagnostic interpretation.

    4. Adjudication Method for Test Set

    • None. As the evaluation relies on objective measurements against ISO standards and physical comparability, no human adjudication was required.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No. This type of study is relevant for diagnostic devices where human readers (e.g., radiologists) interpret results, often with and without AI assistance, to assess performance improvement. The Medi-Jector Choice is a drug delivery device, not a diagnostic one, so an MRMC study is not applicable.

    6. Standalone (Algorithm Only) Performance Study

    • Yes, in essence. The entire evaluation described is a "standalone" performance study of the device itself. There is no human-in-the-loop component being evaluated. The device's mechanical performance is directly measured without human intervention in the delivery process being part of the performance metric.

    7. Type of Ground Truth Used

    • Objective Standards and Physical Comparability:
      • Dosage Delivery: ISO insulin syringe standards. These are internationally recognized, objective benchmarks for accuracy.
      • Penetration: Direct comparison to physical streams produced by a predicate device (Medi-Jector Needle-Free Bio-Tropin™ Drug Delivery System) into an artificial test matrix (Whatman #3 filter paper). This is a physical, measurable comparison rather than a human-interpreted "ground truth."

    8. Sample Size for the Training Set

    • Not Applicable. As a mechanical device, there is no "training set" in the machine learning sense. The device's design is based on engineering principles and materials science, not on learning from a dataset.

    9. How the Ground Truth for the Training Set Was Established

    • Not Applicable. There is no training set for a mechanical device like this. Its performance is inherent in its design and manufacturing.
    Ask a Question

    Ask a specific question about this device

    K Number
    K960285

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1996-04-17

    (89 days)

    Product Code
    Regulation Number
    880.5430
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is intended only for the long-term treatment of children who have growth failure due to an inadequate secretion of normal endogenous growth hormone. It is intended for the subcutaneous administration of the reconstituted drug by highpressure injection either by health care practitioners or directly by patients.

    Device Description

    The Medi-Jector Needle-Free Human Growth Hormone Injection System is a needle-free, springpowered human growth hormone (hGH) injection system. It is comprised of a "power pack" with a detachable nozzle. The power pack contains the spring and has a plastic housing, injection release button, safety latch, and dosage display window; it is wound to draw up the desired dose of reconstituted hGH (up to 0.5 ml of hGH in bacteriostatic saline). The reusable stainless steel nozzle is supplied non-sterile and requires cleaning and boiling sterilization by the user prior to first use and every two weeks thereafter. The device is not supplied pre-loaded with hGH; rather, it utilizes a supplied sterile "bottle adapter" accessory, with an integral spike, to aseptically penetrate the rubber seal of a standard vial of reconstituted hGH, accept the Medi-Jector's nozzle end, and thus draw a calibrated dose of reconstituted hGH into the chamber of the opaque nozzle assembly by way of winding motion of the power pack.

    AI/ML Overview

    Here's an analysis of the provided 510(k) summary, extracting the requested information about acceptance criteria and the supporting studies:

    Acceptance Criteria and Device Performance for Medi-Jector Needle-Free Bio-Tropin™ Drug Delivery System

    1. Table of Acceptance Criteria and Reported Device Performance

    The 510(k) summary for the Medi-Jector Needle-Free Bio-Tropin™ Drug Delivery System does not explicitly present acceptance criteria in a formal table with specific thresholds. Instead, it describes performance characteristics where the device was assessed against predicate devices or general principles of safety and effectiveness. The "acceptance criteria" are implied by the conclusions drawn about "no differences," "similar," or "bioequivalent."

    Area of AssessmentImplied Acceptance Criteria (Goal)Reported Device Performance
    SterilityEnsure adequate sterility of medication-contacting components (bottle adapter).Medi-Jector bottle adapter validated to a sterility assurance level of 10^-6.
    BiocompatibilityNo adverse tissue reactions or cytotoxicity from drug-contacting materials.No differences in macroscopic or microscopic porcine dermal alterations over 72 hours after Medi-Jector or standard needle injection. Drug-contacting materials verified.
    Biochemical Integrity (hGH)No detrimental effect on the chemical integrity (e.g., dimerization, oxidation, deamidation) of hGH due to injection/contact.No evidence that Medi-Jector injection changes the chemical integrity of injected hGH molecule at various power settings; no molecular differences between jet-injected and standard syringe/needle-injected hGH.
    Leaching (Heavy Metals)Negligible leaching of hazardous substances (e.g., cadmium, lead) from adapter materials into hGH solution.Spectrophotometric analysis showed concentrations of cadmium and lead generally within the range of blank reference controls.
    hGH Stability in AdapterAdapter components should not adversely affect the stability or purity of hGH over time.Chromatographic analysis of adapter/hGH solution showed no peaks different from control hGH solution. Analysis of adapter parts incubated in hGH demonstrated no progressive increase in polymeric, deamidated, oxidized, or unknown forms of hGH, or a difference from controls, at 5°C. At 37°C, while forms increased for both, sample levels remained equivalent to controls at all time points.
    Nozzle Seal IntegrityNo leakage or degradation of nozzle friction seals after repeated exposure and sterilization.Repeated fillings with dyes showed no moistening of seals. No change in dimensional conformance, tear strength, or surface quality after exposure to hGH for 5 weeks or 5 cycles of boiling.
    Ink Printing Rub ResistancePrinting on power pack should withstand expected cleaning procedures.Withstood 25 double rub cycles with a solvent-challenged rubbing cloth (2 lbs force, recommended cleaning solutions).
    Pressure Curve IdentityPressure curve characteristics should be identical to predicate devices.Pressure curve experiments confirmed identity with proposed predicate devices.
    Penetration Depth/DispersionSimilar penetration and dispersion in artificial matrix compared to predicate devices.Penetration studies confirmed similar injectate penetration between Medi-Jector and Biojector 2000.
    Delivery AccuracyAccurate delivery of target dose within an acceptable margin of error.Mean Medi-Jector error of 1.4% at a target dose of 0.125 ml and 0.37% at a target dose of 0.5 ml.
    Reuse DurabilityExtensive use should not significantly impact volume delivery.Supported the notion that extensive use does not significantly impact volume delivery.
    Contamination PreventionBottle adapter should not contaminate hGH vial; nozzle should maintain hygienic status between sterilizations.Absence of bacterial contamination (Pseudomonas putida) in vials after adaptor spike penetration, validating integrity of closure. Once-daily use of disinfected Medi-Jector ensured hygienic status of hGH-contacting nozzle components for at least 28 days.
    Bioequivalence (Human)Systemic exposure to hGH (amount absorbed) should be similar to conventional syringe administration.Systemic exposure to hGH (in terms of amount absorbed into circulation) was similar (bioequivalent) for Medi-Jector and conventional syringe administration. Higher and more rapid onset of peak concentration with Medi-Jector due to maximal spreading (favoring absorption).
    Bioequipotency (IGF-I)Increase in IGF-I serum concentration should be similar to conventional syringe administration.Marked increase in IGF-I observed after hGH administration by both treatments; no statistically significant difference in IGF-I between treatments at any time point measured during the 24-hour post-injection period (bioequipotency).
    Safety in UseNo serious adverse reactions or significant complications compared to conventional methods.Two mild skin lesions after incomplete Medi-Jector injection, one mild hematoma 24 hours after Medi-Jector use; no serious adverse reactions. Incomplete injections were attributed to fixed penetration setting. Another clinical study cited comparable height increase and growth rate with Medi-Jector vs. syringe.

    2. Sample Size Used for the Test Set and Data Provenance

    • Clinical Study (Bioequivalence/Bioequipotency):
      • Sample Size: 12 healthy young male volunteers.
      • Data Provenance: Prospective (clinical study conducted for this submission), likely from the country of the manufacturer (USA) unless specified otherwise.
    • Clinical Study (Growth Hormone Therapy in Children):
      • Sample Size: Not explicitly stated, refers to "hGH-deficient children."
      • Data Provenance: Retrospective (cited published report: Igarashi et al., 1993), likely Japanese given author names and publication title (Clin. Rep.).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    • Clinical Studies: No external "experts" were used to establish ground truth in the sense of adjudication for the clinical test sets. The ground truth was based on objective physiological measurements (hGH, IGF-I, free fatty acids) or clinical outcomes (height increase, growth rate) from the subjects themselves, compared against conventional syringe injection. Medical professionals (nurses, physicians) would have administered the injections and collected the data, but their "expertise" was in performing the study, not in establishing a separate ground truth.

    4. Adjudication Method for the Test Set

    • Clinical Studies: Not applicable. The studies involved direct measurement of physiological responses (bioavailability, bioactivity markers) or direct clinical outcomes, not subjective interpretations requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This device is a drug delivery system, not an AI-powered diagnostic or imaging interpretation tool. Therefore, MRMC studies and "human readers improved with AI" metrics are not relevant. The comparison was between the device and conventional syringe injection.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This is a mechanical drug delivery device. Its performance is inherent to its design and function, not an algorithm's output. The "standalone" performance is simply the device's ability to inject safely and effectively, as demonstrated by the various in vitro and clinical tests.

    7. The Type of Ground Truth Used

    • Sterility: Sterility Assurance Level (SAL) of 10^-6 for the bottle adapter.
    • Biocompatibility: Microscopic and macroscopic evaluation of porcine dermal alterations.
    • Biochemical Integrity: Chemical assays (dimer, oxidized, deamidated forms of hGH) and spectrophotometric/chromatographic analysis.
    • Delivery Accuracy: Measured volume of liquid (bidistilled water) delivered vs. target dose.
    • Contamination Prevention: Absence of bacterial contamination (Pseudomonas putida) after exposure.
    • Bioequivalence/Bioequipotency (Clinical): Measured serum concentrations of hGH, IGF-I, and free fatty acids in human volunteers.
    • Clinical Efficacy (Children): Measured height increase and growth rate.

    8. The Sample Size for the Training Set

    • Not applicable. This device does not use machine learning or AI, so there is no concept of a "training set" in the context of an algorithm. All samples and tests described are for evaluation of the physical device's performance.

    9. How the Ground Truth for the Training Set was Established

    • Not applicable. As there is no training set for an algorithm, this question is not relevant.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1