    K Number
    K201983

    Validate with FDA (Live)

    Date Cleared
    2020-10-09

    (84 days)

    Product Code
    Regulation Number
    874.3340
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BONEBRIDGE bone conduction hearing implant system is intended for the following patients and indications: · Patients 12 years of age or older. · Patients who have a conductive or mixed hearing loss and still can benefit from sound amplification. The pure tone average (PTA) bone conduction (BC) threshold (measured at 0.5, 1, 2 and 3 kHz) should be better than or equal to 45 dB HL. · Bilateral fitting of the BONEBRIDGE is intents having a symmetrically conductive or mixed hearing loss. The difference between the left and right sides' BC thresholds should be less than 10 dB on average measured at 0.5, 1, 2 and 3 kHz, or less than 15 dB at individual frequencies. · Patients who have profound sensorineural hearing loss in one ear and normal hearing in the opposite ear (i.e., singlesided deafness or "SSD"). The pure tone average air conduction hearing ear should be better than or equal to 20 dB HL (measured at 0.5, 1, 2 and 3 kHz). · The BONEBRIDGE for SSD is also indicated for any patient who is indicated for an air-conduction contralateral routing of signals (AC CROS) hearing aid, but who for some reason cannot or will not use an AC CROS. · Prior to receiving the device, it is recommended that an individual has experience with appropriately fit air conduction or bone conduction hearing aids.

    Device Description

    The BCI Lifts are optional accessories that can be implanted together with the BONEBRIDGE Bone Conduction Implant (model BCI 601 or BCI 602). They are attached to the fixation wings of the BCI and serve as a spacer to increase the distance between the fixation wings and the skull bone. The devices can be used if the necessary drill depth for the BCI implant cannot be achieved for anatomical reasons. They are attached to the fixation wings of the BCI implant to reduce the necessary drill depth for the Bone Conduction - Floating Mass Transducer (BC-FMT) in the skull bone.

    AI/ML Overview

    This document is a 510(k) summary for the BCI Lifts, which are accessories for the BONEBRIDGE bone conduction hearing implant system. The purpose of this submission is to request clearance to update the materials of the BCI Lifts from PEKK to PEEK.

    Here's an analysis of the provided text in relation to the requested information:

    1. Table of acceptance criteria and the reported device performance:

    The document focuses on demonstrating bioequivalence and functional equivalence of the new material (PEEK) for the BCI Lifts compared to the existing material (PEKK). It does not present specific quantitative acceptance criteria or performance metrics in a tabular format as one might expect for an AI/ML device. Instead, it describes general categories of evaluation.

    Acceptance Criteria CategoryReported Device Performance
    Biocompatibility"The BCI Lifts made of PEEK have been evaluated and shown to be biocompatible." (Per ISO 10993-1)
    Sterilization"The BCI Lifts made of PEEK have been evaluated as per AAMI TIR28:2016 and it was concluded that the new Lifts can be adopted into the existing sterilization of the BCI Lifts (PEKK) without further study."
    Mechanical Properties (Material)"An assessment and comparison of the mechanical properties of the two materials was performed." (Implies equivalence was established, but no specifics given).
    Performance (Device-Level)"Performance testing was performed on BCI Lifts made of PEEK according to existing test methods."
    Torque/Force Stability (with BCI implants)"Testing of the new BCI Lifts with the BCI implants was performed according to existing test methods that covered: Torque/Force Stability." (Implies satisfactory performance, but no specific values).
    Dynamic Impact at 2.5J (with BCI implants)"Testing of the new BCI Lifts with the BCI implants was performed according to existing test methods that covered: Dynamic Impact at 2.5J." (Implies satisfactory performance, but no specific values).

    The overall conclusion is that the new material is "at least as safe and effective" as the predicate.

    2. Sample size used for the test set and the data provenance:

    The document describes "bench testing" and evaluations of materials and devices. It does not provide specific sample sizes (e.g., number of units tested) for these engineering tests. There's no indication of clinical test sets or data provenance (country of origin, retrospective/prospective) because this submission is about a material change, not a new clinical indication or AI/ML algorithm.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This is not applicable to this 510(k) submission. This submission does not involve an AI/ML device that requires human expert review for establishing ground truth on image data or similar diagnostic tasks. The evaluation is focused on material science and mechanical engineering properties.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    Not applicable. As there's no expert review of data for ground truth establishment relevant to AI/ML, there's no adjudication method described.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This 510(k) is not for an AI/ML device that assists human readers. It's for a change in material for an implant accessory.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Not applicable. This is not an AI/ML algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    The "ground truth" in this context is established through industry standards and established engineering test methods for biocompatibility (ISO 10993-1), sterilization (AAMI TIR28:2016), and mechanical performance (e.g., torque/force stability, dynamic impact). This is not a "ground truth" in the clinical AI sense.

    8. The sample size for the training set:

    Not applicable. There is no training set mentioned or implied as this is not an AI/ML device.

    9. How the ground truth for the training set was established:

    Not applicable. There is no training set or clinical ground truth in the context of an AI/ML device. The "ground truth" for the engineering tests mentioned is inherent in the established test methods and material science principles.

    Ask a Question

    Ask a specific question about this device

    K Number
    K200504

    Validate with FDA (Live)

    Date Cleared
    2020-07-29

    (149 days)

    Product Code
    Regulation Number
    874.3340
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Bonebridge is intended to improve hearing for patients with conductive or mixed hearing losses, bilateral fitting, and single-sided deafness.

    INDICATIONS:
    The Bonebridge bone conduction hearing implant system is intended for the following patients and indications:

    • Patients 12 years of age or older.
    • Patients who have a conductive or mixed hearing loss and still can benefit from sound amplification. The pure tone average (PTA) bone conduction (BC) threshold (measured at 0.5, 1, 2 and 3kHz) should be better than or equal to 45 dB HL.
    • Bilateral fitting of the Bonebridge is intended for patients having a symmetrically conductive or mixed hearing loss. The difference between the left and right sides' BC thresholds should be less than 10 dB on average measured at 0.5, 1, 2 and 3kHz, or less than 15 dB at individual frequencies.
    • Patients who have profound sensorineural hearing loss in one ear and normal hearing in the opposite ear (i.e., single-sided deafness or "SSD"). The pure tone average air conduction hearing thresholds of the hearing ear should be better than or equal to 20 dB HL (measured at 0.5, 1, 2 and 3kHz).
    • The Bonebridge for SSD is also indicated for any patient who is indicated for an air-conduction contralateral routing of signals (AC CROS) hearing aid, but who for some reason cannot or will not use an AC CROS.
    • Prior to receiving the device, it is recommended that an individual has experience with appropriately fit air conduction or bone conduction hearing aids.
    Device Description

    The MED-EL BONEBRIDGE System augments hearing by providing acoustic input to the inner ear via bone conduction. The MED-EL BONEBRIDGE System consists of two major components: The implant, called Bone Conduction Implant (BCI) and the external audio processor, e.g. the SAMBA 2 BB. The SYMFIT 8.0 software enables the fitting and configuration of the SAMBA 2 BB audio processor and is indicated for professional use only, i.e. used by hearing healthcare professionals during a fitting session of the external audio processor SAMBA 2 BB. Additionally, the SAMBA 2 GO and SAMBA 2 Remote are also subject to this 510(k). Both devices offer a convenient option to adjust simple, pre-defined settings of the compatible SAMBA 2 BB audio processor. Additionally, the SAMBA 2 GO can also be used as an assistive streaming device. The MED-EL BONEBRIDGE System is an active implantable bone conduction hearing system. The MED-EL BONEBRIDGE System is a prescription device consisting of an implanted transducer, implanted electronics components, and an audio processor. The active implantable bone conduction hearing system is intended to compensate for conductive or mixed hearing losses by conveying amplified acoustic signals to the cochlea via mechanical vibrations on the skull bone.

    AI/ML Overview

    The provided text is a 510(k) summary for the MED-EL Elektromedizinische Geraete GmbH's BONEBRIDGE System components (SAMBA 2 BB, SYMFIT 8.0, SAMBA 2 GO, SAMBA 2 Remote). This document focuses on demonstrating substantial equivalence to a predicate device rather than providing a detailed study proving acceptance criteria for a novel device or AI.

    Therefore, much of the requested information, such as sample sizes for test sets, data provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone algorithm performance, and training set details, is not applicable or not present in this type of regulatory submission.

    The document primarily describes comparative bench testing against a predicate device.

    Here's the information that can be extracted from the provided text based on your request:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state quantitative acceptance criteria or specific numerical performance metrics in the format of a table you've requested. Instead, it relies on demonstrating that the new devices perform "as intended" and are "at least as safe and effective" as the predicate devices, based on various bench tests.

    Acceptance Criteria (Inferred from comparison to predicate)Reported Device Performance
    Same basic design concept and technological characteristics as predicate.SAMBA 2 BB maintains basic design, electronics, housing material, and similar stimulation strategies to predicate. Differences (battery door, dimensions, weight, magnet strength, microphone protection, microphone type, IP rating) do not affect safety or effectiveness.
    Same energy source as predicate.SAMBA 2 BB uses the same energy source as predicate.
    Biocompatible materials in skin contact.Main materials in skin contact for SAMBA 2 BB evaluated per 10993-1 and shown to be biocompatible and safe. Material identical to predicate.
    Signal processing maintains basic functionality and overall performance compared to predicate.SAMBA 2 BB signal processing features same basic functionality as predicate, with new optional noise reduction, compression settings, and improved acoustic classification. Bench testing confirms differences do not affect safety or effectiveness, and overall performance remains identical.
    Connectivity options maintain functionality compared to predicate.SAMBA 2 BB offers identical connectivity options as predicate, plus an acoustic interface for MED-EL Wireless Accessories. Bench testing confirms differences do not affect safety or effectiveness.
    SYMFIT 8.0 software maintains basic working principle, operational environment, moderate level of concern, and purpose as predicate.SYMFIT 8.0 works to program SAMBA 2 BB, maintains basic working principle and operational environment. Software architecture improved, workflow identical but with adapted UI. Bench and usability testing confirm differences do not affect safety or effectiveness.
    SYMFIT 8.0 fitting features maintain basic functionality and overall performance compared to predicate.SYMFIT 8.0 fitting features provide same basic functionality as predicate, with new optional fine-tuning for compression settings. Bench testing confirms differences do not affect safety or effectiveness, and overall performance remains identical.
    SAMBA 2 GO design concept and SAMBA 2 Remote app-specific design do not impact safety or effectiveness.SAMBA 2 GO has slightly updated design (neckloop); SAMBA 2 Remote is a mobile app. Bench testing confirms differences do not affect safety or effectiveness.
    SAMBA 2 GO and SAMBA 2 Remote remote-control features maintain functionality compared to predicate.SAMBA 2 GO and SAMBA 2 Remote provide identical remote-control features, with SAMBA 2 Remote offering an additional "Sound Balancing" feature for comfort. Bench testing confirms differences do not affect safety or effectiveness.
    SAMBA 2 GO and SAMBA 2 Remote connectivity maintains functionality compared to predicate.SAMBA 2 GO uses identical wireless connection, and enables streaming (non-medical purpose). SAMBA 2 Remote uses acoustic interface. Bench testing confirms differences do not affect safety or effectiveness.
    Device components meet established specifications for safe and consistent performance.Non-clinical bench tests (electrical, performance, compatibility, reliability, shipping, environmental, electrical safety, EMC, software testing) demonstrated that all components meet established specifications ensuring consistent and safe performance for intended use.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document mentions "bench testing" and "non-clinical performance testing." This usually implies testing on physical devices or software in a laboratory setting, not patient data. Therefore, concepts like sample size of a patient test set, data provenance (country, retrospective/prospective), etc., are not explicitly provided or relevant in the context of this 510(k) summary, which focuses on device modifications and substantial equivalence via engineering and verification testing.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. Ground truth typically refers to clinical data interpreted by experts. This document describes bench testing of hardware and software components.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. Adjudication methods are for human interpretation of clinical data sets.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI-assisted diagnostic device, and no MRMC study is mentioned. The device is a hearing aid system.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This refers to software performance, specifically for SYMFIT 8.0, SAMBA 2 GO, and SAMBA 2 Remote. The document states "Software testing according to software's defined Level of Concern" was conducted. This implies standalone testing of the software components. However, specific performance metrics for the software in a "standalone" fashion are not provided, beyond confirming it meets specifications and does not affect safety or effectiveness compared to the predicate.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    For the "bench testing," the "ground truth" would be established engineering specifications, industry standards, and the performance of the predicate device. This is not expert consensus, pathology, or outcomes data.

    8. The sample size for the training set

    Not applicable. This device is not an AI/Machine Learning device that requires a training set in the typical sense.

    9. How the ground truth for the training set was established

    Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K191457

    Validate with FDA (Live)

    Date Cleared
    2019-09-18

    (110 days)

    Product Code
    Regulation Number
    874.3340
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    INTENDED USE:

    The Bonebridge is intended to improve hearing for patients with conductive or mixed hearing losses, bilateral fitting, and single-sided deafness.

    INDICATIONS:

    The Bonebridge bone conduction hearing implant system is intended for the following patients and indications:

    • Patients 12 years of age or older.

    • Patients who have a conductive or mixed hearing loss and still can benefit from sound amplification. The pure tone average (PTA) bone conduction (BC) threshold (measured at 0.5, 1, 2 and 3 kHz) should be better than or equal to 45 dB HL.

    • Bilateral fitting of the Bonebridge is intended for patients having a symmetrically conductive or mixed hearing loss. The difference between the left and right sides' BC thresholds should be less than 10 dB on average measured at 0.5, 1, 2 and 3 kHz, or less than 15 dB at individual frequencies.

    • Patients who have profound sensorineural hearing loss in one ear and normal hearing in the opposite ear (i.e., single-sided deafness or "SSD"). The pure tone average air conduction hearing thresholds of the hearing ear should be better than or equal to 20 dB HL (measured at 0.5, 1, 2 and 3 kHz).

    -The Bonebridge for SSD is also indicated for any patient who is indicated for an air-conduction contralateral routing of signals (AC CROS) hearing aid, but who for some reason cannot or will not use an AC CROS.

    • Prior to receiving the device, it is recommended that an individual has experience with appropriately fit air conduction or bone conduction hearing aids.
    Device Description

    Not Found

    AI/ML Overview

    The provided text is a 510(k) clearance letter from the FDA for a medical device called the "BONEBRIDGE System." This document primarily focuses on the regulatory clearance process, outlining the device's intended use, indications, and contraindications. It does not contain any information about acceptance criteria, study methodologies (like sample sizes, expert qualifications, adjudication methods), or performance data that would prove the device meets specific acceptance criteria.

    The purpose of a 510(k) submission is to demonstrate substantial equivalence to a legally marketed predicate device, not necessarily to present detailed performance study results against predefined acceptance criteria in the format requested. While performance data may have been submitted as part of the 510(k) application, that information is not present in this public clearance letter.

    Therefore, I cannot fulfill your request for the detailed information regarding acceptance criteria and the study proving the device meets them based solely on the provided text. The document is a regulatory communication, not a clinical study report.

    Ask a Question

    Ask a specific question about this device

    K Number
    K183373

    Validate with FDA (Live)

    Device Name
    Bonebridge
    Date Cleared
    2019-03-07

    (91 days)

    Product Code
    Regulation Number
    874.3340
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BONEBRIDGE bone conduction hearing implant system is intended for the following patients and indications:

    • · Patients 12 years of age or older.
      · Patients who have a conductive or mixed hearing loss and still can benefit from sound amplification. The pure tone average (PTA) bone conduction (BC) threshold (measured at 0.5,1, 2, and 3 kHz) should be better than or equal to 45 dB HL.

    · Bilateral fitting of the BONEBRIDGE is intended for patients having a symmetrically conductive or mixed hearing loss. The difference between the left and right sides' BC thresholds should be less than 10 dB on average measured at 0.5. 1, 2. and 3 kHz, or less than15 dB at individual frequencies.

    · Patients who have profound sensorineural hearing loss in one ear and normal hearing in the opposite ear (i.e., singlesided deafness or "SSD"). The pure tone average air conduction hearing thresholds of the hearing ear should be better than or equal to 20 dB HL (measured at0.5, 1, 2, and 3 kHz).

    · The BONEBRIDGE for SSD is also indicated for any patient who is indicated for an air-conduction contralateral routing of signals (AC CROS) hearing aid, but who for some reason cannot or will not use an AC CROS.

    · Prior to receiving the device, it is recommended that an individual have experience with appropriately fit air conduction or bone conduction hearing aids.

    Device Description

    BONEBRIDGE augments hearing by providing acoustic input to the inner ear via bone conduction. BONEBRIDGE consists of the following components: The Bone Conduction Implant (BCI) and the externally worn audio processor.

    The BCI 601 implant is provided in an implant kit together with two templates to determine optimal implant placement, one drill bit, two regular cortical titanium screws (gold surface) and one emergency screw (blue surface).

    All devices are shipped in one sterile tray and are for single-use only.

    The Bone Conduction Implant (BCI) is the implantable part of BONEBRIDGE and can only be used together with compatible MED-EL external components. The device is an osseointegrated bone conduction implant system, intended to provide a level of useful sound perception for individuals with conductive and mixed hearing loss, as well as those suffering from single-sided deafness. The BCI is surgically implanted into the mastoid bone.

    The BCI 601 has two main sections, the coil section and the transducer section.

    The BCl 601 consists of a magnet surrounded by the receiver coil, the electronics (demodulator), the transition and the Bone Conduction - Floating Mass Transducer (BC-FMT). The BCI 601 is activated by placing the external audio processor over the magnet of the BCI 601. The signal and the energy to drive the BC-FMT are transferred via an inductive link to the internal coil, processed by the demodulator and then relayed to the BC-FMT. The BC-FMT transduces the signal into mechanical vibrations, which are conducted to the mastoid bone via the cortical titanium screws. These vibrations stimulate the auditory system through the bone conduction pathway to allow the patient to hear.

    AI/ML Overview

    The provided text is a 510(k) summary for the MED-EL BONEBRIDGE device. This document primarily focuses on obtaining clearance for labeling updates regarding Electromagnetic Compatibility (EMC) and Magnetic Resonance Imaging (MRI) compatibility, rather than presenting a performance study for the device's primary function as a hearing system, or an AI-based system.

    Therefore, much of the requested information about acceptance criteria, efficacy studies, sample sizes, expert ground truth, and AI specifics are not available in this document. The document states that "No changes have been made to the devices themselves. Only the product labelling is affected." and the updates "do not affect the device intended use, fundamental operating principles or functional characteristics." This implies that the performance related to hearing augmentation was presumed to be established by previous clearances (DEN170009).

    Here's what can be extracted from the document:

    1. A table of acceptance criteria and the reported device performance

    The document does not provide a table of acceptance criteria for the hearing performance of the device, nor does it report the device's clinical performance in that regard. It focuses on the safety aspects related to EMC and MRI.

    Acceptance Criteria (related to labelling update)Reported Device Performance (related to labelling update)
    Electromagnetic Compatibility (EMC): Device meets established specifications for EMC."Electromagnetic compatibility testing" was conducted. "The collective results of the non-clinical testing demonstrate that BONEBRIDGE meets the established specifications to ensure consistent and safe performance for its intended use." (Specific numerical criteria or detailed results are not provided).
    Magnetic Resonance Imaging (MRI) Compatibility: BCI 601 implant is MR conditional at 1.5T."MRI compatibility testing" was conducted. The labelling update is "to label the BCI 601 implant as MR conditional at 1.5T." (Specific numerical criteria or detailed results are not provided).

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Test Set Sample Size: Not applicable. The document describes "bench testing" for EMC and MRI compatibility, not a clinical test set of patients or data in the typical sense for evaluating clinical performance. These tests are conducted on the device itself.
    • Data Provenance: Not applicable for clinical data. The tests are non-clinical bench tests. The document does not specify the location where these bench tests were conducted.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. Ground truth for EMC and MRI compatibility is established through standardized testing protocols (e.g., ISO, IEC standards) and engineering evaluations, not typically by clinical experts in the same way as, for example, image interpretation.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This concept is relevant for studies involving human interpretation or clinical endpoints, which this document does not cover for the labeling update.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is not an AI-assisted diagnostic tool, and no MRMC study is mentioned or relevant to the reported labeling update.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is not an algorithm, but an implantable hearing system.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the EMC and MRI compatibility bench tests, the "ground truth" would be objective measurements against established engineering and safety standards (e.g., electromagnetic interference limits, heating, artifact, or force limits within an MRI environment). The document does not explicitly detail these standards or the specific measurements. For the hearing performance (which is not evaluated in this document), the ground truth would typically be audiometric data and patient-reported outcomes.

    8. The sample size for the training set

    Not applicable. This device is not an AI algorithm that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    DEN170009

    Validate with FDA (Live)

    Device Name
    Bonebridge
    Date Cleared
    2018-07-20

    (522 days)

    Product Code
    Regulation Number
    874.3340
    Type
    Direct
    Age Range
    All
    Reference & Predicate Devices
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BONEBRIDGE bone conduction hearing implant system is intended for the following patients and indications:

    • Patients 12 years of age or older.
    • Patients who have a conductive or mixed hearing loss and still can benefit from sound amplification. The pure tone average (PTA) bone conduction (BC) threshold (measured at 0.5, 1, 2, and 3 kHz) should be better than or equal to 45 dB HL.
    • . Bilateral fitting of the BONEBRIDGE is intended for patients having a symmetrically conductive or mixed hearing loss. The difference between the left and right sides' BC thresholds should be less than 10 dB on average measured at 0.5, 1, 2, and 3 kHz, or less than 15 dB at individual frequencies.
    • Patients who have profound sensorineural hearing loss in one ear and normal hearing . in the opposite ear (i.e., single-sided deafness or "SSD"). The pure tone average air conduction hearing thresholds of the hearing ear should be better than or equal to 20 dB HL (measured at 0.5, 1, 2, and 3 kHz).
    • . The BONEBRIDGE for SSD is also indicated for any patient who is indicated for an air-conduction contralateral routing of signals (AC CROS) hearing aid, but who for some reason cannot or will not use an AC CROS.
    • . Prior to receiving the device, it is recommended that an individual have experience with appropriately fit air conduction or bone conduction hearing aids.
    Device Description

    The BONEBRIDGE consists of a Bone Conduction Implant that is surgically implanted in the mastoid bone and an external Audio Processor that is held in place on the patient's scalp by magnetic attraction between the implant and the Audio Processor (Figure 1).

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study data provided for the BONEBRIDGE™ System:

    The BONEBRIDGE™ System did not undergo a study to prove meeting "acceptance criteria" in the traditional sense of a classification algorithm with performance metrics like sensitivity, specificity, or AUC. Instead, it underwent a series of non-clinical (bench) and clinical studies to demonstrate its safety and effectiveness for its intended use as an active implantable bone conduction hearing system.

    The acceptance criteria are therefore defined by the successful demonstration of:

    • Biocompatibility of all components.
    • Sterilization and Shelf Life validation.
    • Electrical Safety and Electromagnetic Compatibility (EMC), including wireless compatibility.
    • Software (SYMFIT) verification and validation at a "Minor" Level of Concern.
    • Performance Testing (Bench) for the Audio Processor (SAMBA BB) and the Bone Conduction Implant (BCI 601) and surgical tools.
    • Clinical Effectiveness for conductive/mixed hearing loss and single-sided deafness.
    • Clinical Safety in terms of adverse events and maintenance of bone conduction thresholds.

    1. Table of Acceptance Criteria and Reported Device Performance

    Given the nature of the device, the "acceptance criteria" and "reported device performance" are largely based on meeting pre-defined thresholds and "passing" various tests to demonstrate safety and functional specifications.

    CategoryAcceptance Criteria (Design Requirement/Objective)Reported Device Performance (Result)
    BiocompatibilityAll patient-contacting components to be biocompatible per ISO 10993 standards.All components (SAMBA BB, BCI 601, BCI Lifts, Surgical Tools) passed all specified biocompatibility tests (Cytotoxicity, Sensitization, Irritation, Systemic Toxicity, Implantation, Genotoxicity, Subchronic Toxicity, Carcinogenicity, Pyrogenicity, Exhaustive Extraction, Particulate Matter).
    Sterilization & Shelf LifeSterilization validation per AAMI/ANSI/ISO 11135:2014; Shelf life validation for sterility and integrity per ISO 11607-1:2006.BCI 601, surgical tools, and BCI Lifts/Sizer Kit sterilization validated (EO sterilization, overkill approach). Packaging of sterile components validated for shelf life, sterility, and integrity. SAMBA BB is non-sterile.
    Electrical Safety & EMCMeet IEC 60601-1-2 EMC standards, ANSI C63.19 wireless compatibility, electrical safety per IEC 60601-1, leakage current/voltage difference.All acceptance criteria were met for EMC (with SAMBA BB), wireless compatibility, and electrical safety (BCI 601 and SAMBA BB). Minor reported issue of "parasitic demodulation" mitigated by labeling instructions.
    MR CompatibilityN/A (Device is identified as MR Unsafe, which is an accepted characteristic rather than a performance criterion)MR Unsafe.
    SoftwareSYMFIT software to be verified, validated, and hazard analyzed at "Minor" Level of Concern.Software description, hazard analysis, traceability, verification/validation, and revision history were addressed. SYMFIT is standalone and does not affect safety; failure only impacts programming.
    Bench Performance (SAMBA BB)Force Output: > Max Power Output (MPO) at specific frequencies (e.g., 90 dB μN at 500 Hz). Transmission Scheme: Difference between force output at 7mm and 4mm coupling distance ≤ 4 dB. Battery Lifetime: > 16 hours. Noise Floor: < 10 dB HL.All force output, transmission scheme, battery lifetime (>16 hours), and noise floor (<10 dB HL) criteria were passed. Additional tests (audiological performance, dual mic, physical characteristics, adapter compatibility, vibration, shock, drop, temperature, substance resistance, solar radiation, moisture ingress, atmospheric pressure, light, shipping, accessory & wireless compatibility, EMC/ESD, ERM, biocompatibility, usability) all passed.
    Bench Performance (BCI 601 & Surgical Tools)BCI Implant Performance: Signal quality (transducer distortion), output level limitation. Reliability: <5% failure rate after 10 years continuous use. Screw Torque/Force Stability: Max torque/force withstand capability. Bending: Functional after bending. Physical Characteristics: Dimensions, symmetry, coating. Simulated Surgical Implantation: Adequacy for safe implantation/fixation. External Pressure/Vibrations/Accelerations: Resistance to vibration, shock, dynamic impact, atmospheric pressure, pressurizing (swimming/diving), chronic implant movement, skull fracture (not adversely affected), 2.5J impact. Medical Treatment Compatibility: Compatibility with diagnostic ultrasound, surgical diathermy, X-ray/CT, radiation therapy/PET. Temperature Resistance: Operational (18-43°C) and storage (-20°C to 60°C).All performance objectives passed. Reliability testing (fatigue fracture, wire guard, adhesive joint, hermeticity, screw fracture, weld seams) showed no failures. Screw torque/force stability, bending, physical characteristics, and simulated implantation passed. All external pressure/vibration/acceleration tests passed, except for one BCI unit which showed a fine leak following a 3.5J impact (acceptance criterion was "no skull fractures were present", which was met). All medical treatment compatibility tests (ultrasound, diathermy, X-ray/CT, radiation/PET) passed. All temperature resistance tests passed.
    Bench Performance (BCI Lifts & Sizer Kit)Performance (Lifts): Performance, distortion, frequency response, linearity, torque/force stability. Mechanical Safety (Lifts): Withstand daily use forces. Usability (Lifts & Sizer Kit): Appropriate for intended use by surgeons during implantation.All BCI Lifts performance (including torque/force stability) and mechanical safety tests passed. Usability testing confirmed appropriate designs for intended use (surgeons implanted BCI 601 using Lifts and Sizer Kit).
    Clinical Effectiveness (CHL/MHL)Primary Endpoint: ≥15% improvement in word/sentence recognition (Freiburger Monosyllable Word recognition) from preoperative unaided to 12-month postoperative aided. Secondary Endpoints: Functional gain (≥28.9 dB PTA improvement 500-4000 Hz, unaided vs. aided); Speech Reception Threshold (SRT; OLSA) improvement.Primary Endpoint Met: Average 63.3% improvement in Freiburger Monosyllable Word recognition (p < 0.001). Secondary Endpoints Met: Mean 28.9 dB functional gain (p < 0.001); Mean 24.0 dB SRT improvement (OLSA) (p < 0.001). HDDS: 79.6% adults, 87.9% children showed increased satisfaction.
    Clinical Effectiveness (SSD)Primary Endpoint: N/A (The text for SSD explicitly states "no significant differences between aided vs. unaided speech recognition performance in scenarios 1 and 2" but then significant benefit was observed in scenario 3.) The primary effectiveness endpoint was generally stated as "met by all three studies". Specific SSD criterion: Significant benefit in speech presented to implanted ear with noise to contralateral ear (Scenario 3).Primary Endpoint Met (Overall): The text states "Overall, the primary effectiveness endpoint was met by all three studies". SSD Specific: Significant benefit of 1.56 dB SNR difference at 12 months for Scenario 3 (speech to implanted ear, noise to contralateral normal hearing ear). No detrimental effect in other scenarios. SSQ-B: 25.7% improvement (above 10% acceptance threshold).
    Clinical Safety (CHL/MHL & SSD)Primary Endpoint: Tabulate Adverse Events (AEs) and Serious Adverse Events (SAEs) through 12-month follow-up. Secondary Endpoint: ≤ mean 10 dB change in bone conduction thresholds (PTA 500-4000 Hz) from pre-op to post-op unaided.Primary Endpoint: 31 AEs, 1 SAE (unrelated to device/procedure) reported in BB001/BB002. 18 device/procedure-related AEs (including 2 SADEs; one excluded due to inclusion criteria violation). 2 procedure-related AEs in BB003. Overall AE rate for included subjects: 7.57% (5/66). SADE rate: 3.03% (2/66). Secondary Endpoint Met: No more than mean 10 dB change in BC thresholds (PTA). Mean pre-op BC: 16.03 dB HL vs. 12-month post-op BC: 17.03 dB HL.

    2. Sample Sizes and Data Provenance

    The document describes three clinical studies: BB001, BB002, and BB003.

    • BB001 (Prospective, single-arm, open-label, pre-market):
      • Sample Size: 24 subjects (12 adult, 12 pediatric) for 3 months post-implantation, with an unspecified number followed longer in BB002.
      • Data Provenance: Not specified, but likely international given the company's origin (Austria). Prospective.
    • BB002 (Post-market follow-up to BB001):
      • Sample Size: 53 subjects (45 adult, 8 pediatric). This appears to include 10 adults and 8 children from BB001, plus an additional 35 adults.
      • Data Provenance: Not specified, likely international. Prospective.
    • BB003 (Prospective, single-arm, open-label, post-market):
      • Sample Size: 13 adult patients with Single Sided Deafness (SSD).
      • Data Provenance: Not specified, likely international. Prospective.

    Test Set for Clinical Studies: The individuals participating in BB001, BB002, and BB003 serve as the test set for clinical performance. The total number of unique individuals across these studies for the most critical outcome assessments (e.g., 12-month data) is approximately 65 (BB001/BB002 + BB003, with some overlap from BB001 into BB002).

    Training Set: Not applicable in the context of this device. This is a medical device, not an AI/ML algorithm that requires a "training set" in the computational sense. The "training" for the device here refers to audiological fitting by healthcare professionals using software, not machine learning model training. The SYMFIT software mentioned explicitly states it "does not contain software" that controls the device's safety and is a "Minor Level of Concern" for failure.


    3. Number of Experts and Qualifications for Ground Truth of Test Set

    • Ground Truth: For this type of medical device, the "ground truth" for clinical effectiveness is primarily objective audiometric measurements (e.g., word recognition scores, functional gain, speech reception thresholds) and subjective patient-reported outcomes (satisfaction, improved hearing).
    • Experts: These measurements are performed by trained audiologists and healthcare professionals. The document does not specify the exact number of experts involved in data collection across all sites or their specific qualifications (e.g., "audiologist with X years of experience"). However, audiometric testing and clinical assessments are standard procedures performed by qualified professionals in the field of audiology and otolaryngology.

    4. Adjudication Method for the Test Set

    The document does not describe a formal "adjudication method" in the context of an expert panel reviewing test results for correctness. Clinical trial data, especially objective audiometric measurements, is typically collected according to a strict protocol on validated equipment and analyzed statistically. Adverse events are reported and classified by the investigators and potentially reviewed by a central safety committee or regulatory body, but a specific "adjudication method" for the test results in the way it might be used for diagnostic image interpretation is not mentioned or typically applicable for this type of device.


    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done.
    MRMC studies are typically used for diagnostic devices, especially those involving image interpretation where multiple human readers assess cases. This device is a hearing prosthesis, and its effectiveness is measured by direct physiological and perceptual improvements in hearing, rather than by interpreting data generated by a separate diagnostic process.

    The "comparison" in the clinical studies is between the patient's preoperative unaided condition and their postoperative aided condition with the BONEBRIDGE system. There is no "human reader" involved in assessing the device's output in the way an MRMC study would imply.


    6. Standalone (Algorithm Only) Performance

    Yes, "standalone" performance testing was extensively done, but it's not "algorithm only" in the sense of AI. This device consists of hardware and associated audiological software for programming.

    • The bench testing for the SAMBA BB Audio Processor and the BCI 601 implant represents the standalone performance of the physical components. These tests evaluate the device's physical, electrical, and audiological specifications in isolation or in a controlled system setup (e.g., force output, battery life, noise floor, reliability, mechanical stability, compatibility tests).
    • The SYMFIT software itself is described as standalone computer software that does not affect the safety of the BONEBRIDGE system, and a failure could only result in inability to program the audio processor, indicating its "standalone" role in programming rather than real-time device operation.

    So, while not an "algorithm-only" performance in the AI sense, the physical device's performance was rigorously evaluated independently of human interaction during its core operation, which is then integrated via professional fitting.


    7. Type of Ground Truth Used

    For the clinical studies:

    • Objective Measures: The primary ground truth for effectiveness was derived from objective audiometric data. This includes:
      • Freiburger Monosyllable Word recognition scores.
      • Pure Tone Average (PTA) thresholds (functional gain).
      • Speech Reception Threshold (SRT) using the Oldenburger Satztest (OLSA).
      • Signal-to-Noise Ratio (SNR) for SSD patients.
    • Subjective Measures (Patient-Reported Outcomes): Additional ground truth came from patient-reported outcomes, specifically satisfaction and perceived benefits measured through:
      • Hearing Device Satisfaction Scale (HDSS) questionnaires.
      • Speech Spatial Qualities-Benefit Questionnaire (SSQ-B).
    • Safety Measures: For safety, the ground truth was derived from clinical observation and reporting of Adverse Events (AEs) and Serious Adverse Events (SAEs), as well as objective measures like changes in bone conduction thresholds at audiometric frequencies.

    8. Sample Size for the Training Set

    As mentioned above, this is a medical device, not an AI/ML algorithm that requires a "training set." Therefore, this question is not applicable in the context of the provided information.


    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no "training set" for an AI/ML algorithm. The "training" for the device (audiological fitting) is done by qualified healthcare professionals based on established audiometric principles and the patient's individual hearing profile.

    Ask a Question

    Ask a specific question about this device

    K Number
    K172460

    Validate with FDA (Live)

    Device Name
    ADHEAR System
    Date Cleared
    2018-04-27

    (256 days)

    Product Code
    Regulation Number
    874.3302
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ADHEAR system is intended to treat patients of all ages with conductive hearing loss or single-sided deafness via bone conduction. The ADHEAR system is a non-invasive bone conduction hearing device which is retained on the patient's head with an elastic headband or an adhesive adapter that is placed behind the auricle.

    Indications:

    · Unilateral or bilateral conductive hearing loss, either chronic or temporary.

    The pure tone average bone-conduction hearing threshold (measured at 0.5. 1. 2, and 3 kHz) should be better than or equal to 25 dB HL.

    · Single-sided deafness (i.e. unilateral profound sensorineural deafness) with normal hearing on the contralateral side. Normal hearing is defined as a pure tone average air-conduction hearing threshold (measured at 0.5, 1, 2, and 3 kHz) of better than or equal to 20 dB HL

    Device Description

    The ADHEAR system includes a bone conduction audio processor that can be retained on the head with an adhesive adapter or by the headband situated over the mastoid behind the auricle. The ADHEAR system is intended to be used during waking hours. While the adhesive adapter is attached to the skin for 3 to 7 days and then replaced, the audio processor is removed at night. Both the audio processor and the adhesive adapter are necessary components for the system to work. Together the components work as one system to deliver vibrations to the mastoid bone in order to conductively transmit sound to the inner ear for patients with conductive hearing loss or single sided deafness via bone conduction.

    The ADHEAR system consists of an adhesive adapter sitting behind the auricle and an audio processor mounted on the attachment. The audio processor contains microphones and signal processing technology as well as a battery. It detects, processes, amplifies and finally transmits sound to the adhesive adapter which transmits vibrations to the mastoid which conducts sounds to the inner ear. The audio processor is equipped with a push button that allows the user to switch between 4 pre-defined settings. A colour sleeve is provided for pediatric use. It features a tamper-proof battery door and dampens potential drops of the processor unit.

    The ADHEAR system contains the following components:

    • . ADHEAR Audio Processor (model 701)
    • Package of ADHEAR Adhesive Adapters
    • . ADHEAR Positioning Tool
    • ADHEAR Headband
    • . Retention clip
    • ADHEAR Sleeves
    AI/ML Overview

    This document is a 510(k) Premarket Notification for the ADHEAR System, a bone conduction hearing device. It primarily focuses on demonstrating substantial equivalence to predicate devices through non-clinical testing.

    Based on the provided text, a conventional "study" demonstrating that a device meets acceptance criteria in the context of AI/ML performance (e.g., accuracy, sensitivity, specificity) is not present. The device in question is a physical hearing aid system, not an AI/ML algorithm for image interpretation or diagnosis.

    Therefore, many of the requested items related to AI/ML performance studies (like sample size for test sets, number of experts for ground truth, MRMC studies, standalone algorithm performance, training set details) are not applicable to this specific submission.

    The document states:
    "No clinical testing was needed to support the safety, performance, and substantial equivalence of the ADHEAR System to the predicate devices. All devices are bone conduction hearing aids and all aspects to ensure consistent and safe performance can be shown through non-clinical bench testing." (Page 6)

    This confirms that the acceptance criteria and proof of meeting them are based on engineering and safety standards, not clinical performance metrics typically associated with AI/ML systems.

    However, I can extract information related to the types of testing and the conclusions regarding the device's performance against its design specifications and relevant standards, which serve as "acceptance criteria" for a hardware medical device.

    Here's a breakdown based on the information available:

    1. A table of acceptance criteria and the reported device performance

    The document describes types of testing performed and concludes that the device complies or meets these criteria. It does not provide specific numerical targets or results in a table format for each criterion. Instead, it offers generalized statements of successful testing.

    Acceptance Criteria Category (Type of Test)Reported Device Performance
    Biocompatibility Testing"All identified biological risks have been adequately addressed... provides objective evidence to support the conclusion that the ADHEAR System and its accessories can be considered biocompatible for its intended use." (Page 5)
    Electrical Safety and Electromagnetic Compatibility (EMC/EMI)"The ADHEAR system complies with the applicable sections of the following standards: IEC 60601-1:2005 +A1:2012, IEC 60601-1-11:2015, IEC 60601-1-6:2010 +A1:2013, IEC 62366:2007 +A1:2014, IEC 60601-1-2:2014, and ANSI C63.19:2011." (Page 5)
    Mechanical and Other Testing (Bench Testing)"Bench testing was successfully performed on the ADHEAR system in respect to the defined design specification requirements such as: Frequency range, OFL90, Maximum gain, Total harmonic distortion (THD), Equivalent input noise (EIN), Operating voltage, Current consumption, DAI connection, Signal processing features, Equivalent output noise (EON), Coupling distortion, Basic safety (general, EMC, usability, home healthcare environment), Removal force, Processor unit outer dimensions, Processor unit weight, Processor unit colors, Adhesive adapter colors, Batter door (battery size), Vibrator suspension, Tamper proof battery door, safety line, Push button operating force, Device marking, Drop test, Light exposure, Substance resistance, Power supply stability, Corrosion resistance, Change Battery, Switch device on/off, Adjust volume, change program, Change adhesive adapter, Place/Remove processor unit onto/from adhesive adapter, Push button longevity, DAI connector cycles, Battery door opening/closing cycles, and Protection during shipping." "The collective results of the non-clinical testing demonstrate that the ADHEAR System meets the established specifications to ensure consistent and safe performance for its intended use." (Page 6)
    Software Verification and Validation"Software verification and validation testing were conducted and documentation was provided... The software for this device was considered as a 'minor' level of concern, since failures or latent flaws in the (optional) software are unlikely to result in any harm to the patient or operator." (Page 6)

    Regarding the other requested information:

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not applicable for AI/ML performance studies. The testing described is hardware and software validation/verification, and biocompatibility, which typically uses defined engineering samples or batches, not "test sets" in the AI/ML sense. No patient data or clinical trial data is mentioned.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable. Ground truth in the AI/ML sense (e.g., expert annotations of medical images) is not relevant here as no such clinical image interpretation or diagnostic study was performed. The "ground truth" for the non-clinical tests would be the established engineering specifications and international standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. No expert adjudication of clinical data was performed.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This device is a hearing aid, not an AI-assisted diagnostic tool for human readers. No MRMC study was performed or required.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. There is no standalone AI algorithm whose performance is being evaluated independently of human interaction. The device's function is to transmit sound, and its standalone performance is evaluated through the non-clinical bench tests (e.g., frequency response, gain).

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    • For the non-clinical tests, the "ground truth" implicitly are the established engineering design specifications, safety standards (e.g., IEC 60601-1), and biocompatibility standards (ISO 10993-1). Tests were conducted to verify the device performed according to these pre-defined requirements.

    8. The sample size for the training set

    • Not applicable. This device is hardware with embedded software, not a machine learning model trained on a dataset.

    9. How the ground truth for the training set was established

    • Not applicable. No training set for an AI/ML model.