510(k) Data Aggregation
(115 days)
The LZI Buprenorphine II Enzyme Immunoassay is intended for the qualitative and semi-quantitative determination of norbuprenorphine in human urine at the cutoff value of 5 ng/mL when calibrated against norburprenorphine. The assay is designed for prescription use with a number of automated clinical chemistry analyzers.
The semi-quantitative mode is for purposes of (1) enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as gas or liquid chromatography/mass spectrometry (GC/MS or LC/MS) or (2) permitting laboratories to establish quality control procedures
The assay provides only a preliminary analytical result. A more specific alternative chemical method (e.g., gas or liquid chromatography and mass spectrometry) must be used in order to obtain a confirmed analytical result. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary test result is positive.
The LZI Buprenorphine II Enzyme Immunoassay is a homogeneous enzyme immunoassay ready-to-use liquid reagent. The assay is based on competition between drug in the sample and drug labeled with the enzyme glucose-6-phosphate dehydrogenase (G6PDH) for a fixed amount of antibody in the reagent. Enzyme activity decreases upon binding to the antibody, and the drug concentration in the sample is measured in terms of enzyme activity. In the absence of drug in the sample, norbuprenorphine-labeled G6PDH conjugate is bound to antibody, and the enzyme activity is inhibited. On the other hand, when drug is present in the sample, antibody would bind to free drug; the unbound norbuprenorphine-labeled G6PDH then exhibits its maximal enzyme activity. Active enzyme converts nicotinamide adenine dinucleotide (NAD) to NADH, resulting in an absorbance change that can be measured spectrophotometrically at 340 nm.
The LZI Buprenorphine II Enzyme Immunoassay is a kit comprised of two reagents, R₁ and R₂, which are bottled separately but sold together within the kit.
The R₁ solution contains mouse monoclonal anti-norbuprenorphine antibody, glucose-6-phosphate (G6P), nicotinamide adenine dinucleotide (NAD), stabilizers, and sodium azide (0.09%) as a preservative. The R₂ solution contains glucose-6-phosphate dehydrogenase (G6PDH) labeled with norbuprenorphine in buffer with sodium azide (0.09%) as a preservative.
N/A
Ask a specific question about this device
(20 days)
The LZI Fentanyl III Enzyme Immunoassay is intended for the qualitative determination of fentanyl in human urine at the cutoff value of 1 ng/mL when calibrated against fentanyl. The assay is designed for prescription use with a number of automated clinical chemistry analyzers.
The assay provides only a preliminary analytical result. A more specific alternative chemical method (e.g., gas or liquid chromatography and mass spectrometry) must be used in order to obtain a confirmed analytical result. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary test result is positive.
The LZI Fentanyl III Enzyme Immunoassay is a homogeneous enzyme immunoassay with ready-to-use liquid reagents. The assay is based on competition between the drug in the sample and the drug labeled with the enzyme glucose-6-phosphate dehydrogenase (G6PDH) for a fixed amount of antibody in the reagent. The drug-labeled G6PDH conjugate is traceable to a commercially available fentanyl standard and referred to as fentanyl-labeled G6PDH conjugate. Enzyme activity decreases upon binding to the antibody, and the drug concentration in the sample is measured in terms of enzyme activity. In the absence of a drug in the sample, fentanyl-labeled G6PDH conjugate is bound to the antibody, and the enzyme activity is inhibited. On the other hand, when the free drug is present in the sample, the antibody would bind to the free drug; the unbound fentanyl-labeled G6PDH then exhibits its maximal enzyme activity. Active enzyme converts nicotinamide adenine dinucleotide (NAD) to NADH, resulting in an absorbance change that can be measured spectrophotometrically at 340 nm.
The LZI Fentanyl III Enzyme Immunoassay is a kit comprised of two reagents, R1 and R2, which are bottled separately but sold together within the kit.
The R1 solution contains mouse monoclonal anti-fentanyl antibody, glucose-6-phosphate (G6P), nicotinamide adenine dinucleotide (NAD), stabilizers, and sodium azide (0.09%) as a preservative. The R2 solution contains glucose-6-phosphate dehydrogenase (G6PDH) labeled with fentanyl in buffer with sodium azide (0.09%) as a preservative.
The provided FDA 510(k) Clearance Letter for the LZI Fentanyl III Enzyme Immunoassay details the device's acceptance criteria (in terms of performance characteristics) and the study results that demonstrate the device meets these criteria.
Here's a breakdown of the requested information:
1. Acceptance Criteria and Reported Device Performance
The acceptance criteria are implicitly defined by the satisfactory performance demonstrated in the various studies, aiming to prove substantial equivalence to the predicate device. The performance characteristics sections show the device meets the expected qualitative and quantitative accuracy, precision, and specificity.
| Acceptance Criteria Category | Specific Criteria (Implicitly Derived from Study Design/Outcomes) | Reported Device Performance |
|---|---|---|
| Intended Use Equivalence | Device is substantially equivalent to the predicate (LZI Fentanyl II Enzyme Immunoassay) in terms of intended use, method principle, device components, and clinical performance. | The submission states, "The LZI Fentanyl III Enzyme Immunoassay is substantially equivalent to the LZI Fentanyl II Enzyme Immunoassay (K201938) manufactured by LZI in terms of intended use, method principle, device components, and clinical performance." The changes are clearly identified (target analyte from norfentanyl to fentanyl, updated cutoff, and assay application parameters), and subsequent studies address the impact of these changes. |
| Qualitative Accuracy | High concordance with LC/MS confirmation for true positives and true negatives, especially at and around the cutoff concentration. | Method Comparison - Clinical Samples (1 ng/mL Cutoff):- Negative by LC/MS analysis: 35 negatives, 0 positives by EIA.- < 50% of cutoff by LC/MS: 0 negatives, 20 positives by EIA (discrepant, attributed to norfentanyl cross-reactivity).- Near Cutoff Negative (50% below to cutoff) by LC/MS: 0 negatives, 12 positives by EIA (discrepant, attributed to norfentanyl cross-reactivity).- Near Cutoff Positive (cutoff to 50% above) by LC/MS: 0 negatives, 22 positives by EIA.- High Positive (> 50% above cutoff) by LC/MS: 0 negatives, 61 positives by EIA.Overall, 35/35 true negatives, 83/83 true positives (excluding near-cutoff positives and positives above cutoff). The remaining 32 discrepant samples (20 + 12) were positive by EIA but negative by LC/MS for fentanyl, indicating cross-reactivity with norfentanyl as a contributing factor. |
| Precision | Consistent and reliable qualitative results (positive/negative) across multiple runs and days, especially around the cutoff concentration. | Precision: 1 ng/mL Cutoff (N=88 total replicates over 22 days):- 0 ng/mL, 0.25 ng/mL, 0.5 ng/mL, 0.75 ng/mL: 88/88 Negative (100% agreement).- 1 ng/mL (Cutoff): 4 Neg / 84 Pos (95.5% Positive, 4.5% Negative - demonstrating near-cutoff variability, which is expected for qualitative assays at the cutoff).- 1.25 ng/mL, 1.5 ng/mL, 1.75 ng/mL, 2 ng/mL: 88/88 Positive (100% agreement). |
| Cross-Reactivity (Specificity) | Minimal or no false positives from structurally unrelated compounds. Acceptable cross-reactivity with fentanyl metabolites and structurally related compounds as defined by the assay's intended specificity. | Fentanyl and Metabolites: Fentanyl (100%), Norfentanyl (40%).Structurally Related Compounds: Varying levels of cross-reactivity (e.g., Acetyl fentanyl 25%, Acryl fentanyl 100%, Thiofentanyl 200%). Many substances showed "ND" (Not Detected) at high concentrations (e.g., 100,000 ng/mL).Structurally Unrelated Pharmacological Compounds: No interference (all Neg/Neg/Pos results for 0 ng/mL Fentanyl, -50% Cutoff, +50% Cutoff respectively) at 100,000 ng/mL for almost all listed compounds, except Dextromethorphan, which showed interference at 20,000 ng/mL (Neg/Pos/Pos) but was acceptable at 15,000 ng/mL (Neg/Neg/Pos). This demonstrates good specificity against common drugs. |
| Interference | Test performance (positive/negative results) should not be significantly affected by common endogenous substances or preservatives found in urine, or by varying specific gravity and pH within physiological ranges. | Endogenous and Preservative Compound Interference: No interference observed for most compounds (e.g., Acetone, Ascorbic acid, Bilirubin, Glucose, Hemoglobin) at high concentrations (e.g., 100,000 mg/dL), showing Neg/Neg/Pos results for 0 ng/mL, -50% Cutoff, +50% Cutoff respectively. Boric acid showed interference at 1,000 mg/dL (Neg/Neg/Neg), indicating a potential issue if present at high concentrations. This is a known interference for certain urine assays.Specific Gravity Interference: No interference observed across the range of 1.000 to 1.030 (all Neg/Neg/Pos).pH Interference: No major interference observed between pH 3 to pH 11 (all Neg/Neg/Pos). |
2. Sample Size and Data Provenance
- Test Set Sample Size:
- Precision Study: 88 replicates for each concentration level (0 ng/mL to 2 ng/mL fentanyl).
- Method Comparison (Clinical Samples): 150 unaltered clinical samples.
- Cross-Reactivity: Duplicates for each compound and concentration tested.
- Interference (Endogenous/Preservative): Duplicates for each compound and concentration tested.
- Specific Gravity/pH Interference: Duplicates for each specific gravity/pH level and fentanyl concentration.
- Data Provenance:
- The clinical samples for the Method Comparison study were obtained by LZI and "through collaboration with various clinical laboratories across the United States and Canada." This indicates a prospective and/or retrospective collection of real-world clinical samples from diverse geographical regions. The nature (retrospective/prospective) of the collection for these specific 150 samples isn't explicitly stated but "clinical samples" usually implies samples collected from patients.
3. Number of Experts and Qualifications for Ground Truth
- The document does not specify the number of experts used to establish ground truth.
- Qualifications of Experts: Not explicitly stated. However, for LC/MS confirmation, it implicitly relies on the expertise of the laboratory personnel performing and interpreting the LC/MS (Liquid Chromatography-Mass Spectrometry) analysis, which is a gold standard for chemical identification and quantification. These would typically be trained analytical chemists or lab technicians with experience in mass spectrometry.
4. Adjudication Method for the Test Set
- The document does not describe an adjudication method for the test set in the typical sense of multiple human readers or reviewers resolving discrepancies.
- For the Method Comparison study, the "ground truth" was established independently by LC/MS. Discrepancies between the EIA result and the LC/MS result were reported and attributed to norfentanyl cross-reactivity. This is a comparison against a reference method, not a subjective adjudication by experts.
5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study
- No, an MRMC comparative effectiveness study was not done. This type of study is more common for diagnostic imaging AI algorithms where human interpretation is a primary component and AI aims to augment or replace it. This document describes an in vitro diagnostic (IVD) assay, an enzyme immunoassay, which is an automated lab test. Its performance is evaluated against a reference standard (LC/MS), not against human reader performance.
6. Standalone Performance
- Yes, standalone performance was done. The entire submission details the performance of the LZI Fentanyl III Enzyme Immunoassay as a standalone device. The qualitative accuracy, precision, cross-reactivity, and interference studies all evaluate the algorithm/device's performance independently in generating a preliminary analytical result from a urine sample. It does not involve human-in-the-loop performance; rather, it provides a result that a clinician then interprets in conjunction with other clinical data.
7. Type of Ground Truth Used
- The primary type of ground truth used for the quantitative confirmation of fentanyl concentration in the clinical samples (Method Comparison study) was LC/MS (Liquid Chromatography-Mass Spectrometry) analysis. This is considered a highly specific and sensitive reference method for drug quantification.
8. Sample Size for the Training Set
- The document does not explicitly state the sample size for the training set. The described studies are verification and validation activities for a modified assay (Special 510(k) for an existing predicate device, LZI Fentanyl II). For IVD assays, "training" often refers to the internal development and optimization of the assay performed by the manufacturer, which precedes the formal V&V studies presented for regulatory submission. The reported studies are the test set performance.
9. How the Ground Truth for the Training Set Was Established
- Since the training set size is not explicitly stated, the method for establishing ground truth for it is also not explicitly described. However, it is highly probable that internal development and optimization of the assay would have utilized similar rigorous analytical methods, likely LC/MS or other established analytical techniques, to calibrate and refine the assay's performance against known concentrations of fentanyl and its metabolites. This would be part of the manufacturer's design control and development process.
Ask a specific question about this device
(58 days)
The LZI Oxycodone III Enzyme Immunoassay is intended for the quantitative determination of oxycodone in human urine at the cutoff values of 100 ng/mL when calibrated against oxycodone. The assay is designed for prescription use with a number of automated clinical chemistry analyzers.
The semi-quantitative mode is for purposes of (1) enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as GC/MS or (2) permitting laboratories to establish quality control procedures.
The assay provides only a preliminary analytical result. A more specific alternative chemical method (e.g., gas or liquid chromatograpy and mass spectrometry) must be used in order to obtain a confirmed analytical result. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary test result is positive.
The LZI Oxycodone III Enzyme Immunoassay is a homogeneous enzyme immunoassay with ready-to-use liquid reagents. The assay is based on competition between drug in the sample and drug labeled with the enzyme glucose-6-phosphate dehydrogenase (G6PDH) for a fixed amount of antibody in the reagent. The drug-labeled G6PDH conjugate is traceable to a commercially available oxycodone standard and referred to as oxycodone-labeled G6PDH conjugate. Enzyme activity decreases upon binding to the antibody, and the drug concentration in the sample is measured in terms of enzyme activity. In the absence of drug in the sample, oxycodone-labeled G6PDH conjugate is bound to antibody, and the enzyme activity is inhibited. On the other hand, when free drug is present in the sample, antibody would bind to free drug; the unbound oxycodone-labeled G6PDH then exhibits its maximal enzyme activity. Active enzyme converts nicotinamide adenine dinucleotide (NAD) to NADH, resulting in an absorbance change that can be measured spectrophotometrically at 340 nm.
The LZI Oxycodone III Enzyme Immunoassay is a kit comprised of two reagents, an R1 and R2, which are bottled separately but sold together within the kit. The LZI Oxycodone III Enzyme Immunoassay is traceable to a commercially available oxycodone standard.
The Ri solution contains mouse monoclonal anti-oxycodone antibody, glucose-6-phosphate (G6P) nicotinamide adenine dinucleotide (NAD), stabilizers, and sodium azide (0.09 %) as a preservative. The R2 solution contains glucose-6-phosphate dehydrogenase (G6PDH) labeled with oxycodone in buffer with sodium azide (0.09 %) as a preservative.
The document provided describes the LZI Oxycodone III Enzyme Immunoassay, a device for detecting oxycodone in human urine. This is a chemical assay, not an AI/ML-based device. Therefore, many of the requested criteria related to AI/ML or imaging studies (like number of experts, adjudication methods, MRMC studies, training/test set sample sizes, and ground truth establishment for AI/ML models) are not applicable.
However, I can extract and present the acceptance criteria for a chemical assay and the study that proves the device meets those criteria based on the provided text.
Device Name: LZI Oxycodone III Enzyme Immunoassay
Device Type: Homogeneous enzyme immunoassay (chemical assay)
Intended Use: Qualitative and semi-quantitative determination of oxycodone in human urine at cutoff values of 100 ng/mL and 300 ng/mL. The assay provides only a preliminary analytical result, requiring a more specific alternative chemical method (e.g., GC/MS or LC/MS) for confirmation.
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria for a chemical assay are typically defined by performance characteristics such as precision, analytical recovery, accuracy (method comparison with a gold standard), cross-reactivity, and interference studies.
Cutoff 1: 100 ng/mL
| Acceptance Criteria Category | Specific Acceptance Criteria (Inferred from study design) | Reported Device Performance and Study Results |
|---|---|---|
| Precision | Consistent results (negative/positive) for samples at various concentrations around the cutoff. For 100 ng/mL cutoff, typically 100% agreement for concentrations ± 25% and 100% agreement for concentrations further from cutoff, with some expected variability at the cutoff itself. | Within Run (N=22 determinations):- 0-75 ng/mL: 100% Negative- 125-200 ng/mL: 100% Positive- 100 ng/mL: 5 Neg / 17 Pos (Semi-quantitative), 9 Neg / 13 Pos (Qualitative)Total Precision (N=88 determinations):- 0-75 ng/mL: 100% Negative- 125-200 ng/mL: 100% Positive- 100 ng/mL: 26 Neg / 62 Pos (Semi-quantitative), 33 Neg / 55 Pos (Qualitative) |
| Analytical Recovery | Determined concentration should be within a specified percentage of the target concentration across the assay range. | Performed across a range of 0-300 ng/mL by serial dilution of a 300 ng/mL spiked sample. Average % Recovery ranged from 90.3% to 110.0% for concentrations 30-300 ng/mL. The 0 ng/mL sample showed 0.7 ng/mL determined average. |
| Method Comparison (Accuracy) | High agreement with a confirmatory method (LC/MS) across different concentration ranges (negative, near cutoff negative/positive, high positive). | Clinical Samples (N=82), compared to LC/MS:Semi-Quantitative Results:- Positive agreement: 90.2% (2 discordants in "Near Cutoff Negative" range, 4 discordants in "Near Cutoff Positive" range)- Negative agreement: 95.1% (4 discordants in "Near Cutoff Positive" range)Qualitative Accuracy Study (same data):- Positive agreement: 90.2%- Negative agreement: 95.1%Discordant samples often contained oxymorphone, indicating potential cross-reactivity or combined effect of oxycodone and its metabolite. |
| Cross-reactivity | Low or no cross-reactivity with structurally unrelated compounds and expected cross-reactivity with major metabolites and structurally related compounds. | Oxycodone & Major Metabolites:- Oxycodone, Oxymorphone: 100.00%- Noroxycodone: 0.40%- Noroxymorphone: 0.17%Structurally Related Compounds: Hydrocodone (0.40%), Hydromorphone (0.40%), Oxymorphone-3β-D-Glucuronide (43.48%). Most others (e.g., Codeine, Morphine, Buprenorphine, Naloxone) showed ND (Not Detected) at 100,000 ng/mL. The study confirmed expected positive and negative results at ±25% of cutoff for various other pharmacological compounds. |
| Interference (Endogenous & Preservative) | No significant interference observed with common endogenous substances or preservatives at relevant concentrations. | No significant interference observed for most compounds (e.g., Acetone, Ascorbic Acid, Bilirubin, Creatinine, Ethanol, Glucose, Hemoglobin, Urea, etc.) at tested concentrations (up to 6000 mg/dL).Boric Acid: Showed interference at 1000 mg/dL at ±25% of cutoff (Negative result for both 75 ng/mL and 125 ng/mL oxycodone samples), but not at ±50% of cutoff. This indicates a potential suppression of the signal by high concentrations of boric acid. |
| Specific Gravity Interference | No deviations from expected results over a range of specific gravity values. | No deviations from expected positive or negative results across specific gravity 1.000 to 1.030 when spiked at ±25% of cutoff (75 ng/mL and 125 ng/mL oxycodone). |
| pH Interference | No deviations from expected results over a physiological pH range. | No deviations from expected results across pH levels from 3 to 11 when spiked at ±25% of cutoff (75 ng/mL and 125 ng/mL oxycodone). |
Cutoff 2: 300 ng/mL
| Acceptance Criteria Category | Specific Acceptance Criteria (Inferred from study design) | Reported Device Performance and Study Results |
|---|---|---|
| Precision | Consistent results (negative/positive) for samples at various concentrations around the cutoff. For 300 ng/mL cutoff, similar criteria as above. | Within Run (N=22 determinations):- 0-225 ng/mL: 100% Negative- 375-600 ng/mL: 100% Positive- 300 ng/mL: 6 Neg / 16 Pos (Semi-quantitative), 10 Neg / 12 Pos (Qualitative)Total Precision (N=88 determinations):- 0-225 ng/mL: 100% Negative- 375-600 ng/mL: 100% Positive- 300 ng/mL: 27 Neg / 61 Pos (Semi-quantitative), 40 Neg / 48 Pos (Qualitative) |
| Analytical Recovery | Determined concentration should be within a specified percentage of the target concentration across the assay range. | Performed across a range of 0-800 ng/mL by serial dilution of an 800 ng/mL spiked sample. Average % Recovery ranged from 104.5% to 111.5% for concentrations 80-800 ng/mL. The 0 ng/mL sample showed 0.3 ng/mL determined average. |
| Method Comparison (Accuracy) | High agreement with a confirmatory method (LC/MS) across different concentration ranges. | Clinical Samples (N=90), compared to LC/MS:Semi-Quantitative Results:- Positive agreement: 97.8% (2 discordants in "Near Cutoff Negative" range, 1 discordant in "Near Cutoff Positive" range)- Negative agreement: 95.6% (1 discordant in "Near Cutoff Positive" range)Qualitative Accuracy Study (same data):- Positive agreement: 95.6%- Negative agreement: 95.6%Discordant samples often involved significant oxymorphone concentrations relative to oxycodone. |
| Cross-reactivity | Low or no cross-reactivity with structurally unrelated compounds and expected cross-reactivity with major metabolites and structurally related compounds. | Oxycodone & Major Metabolites:- Oxycodone, Oxymorphone: 100.00%- Noroxycodone: 0.40%- Noroxymorphone: ND (Not Detected) at 100,000 ng/mL.Structurally Related Compounds: Hydrocodone (0.40%), Hydromorphone (0.40%), Oxymorphone-3β-D-Glucuronide (42.86%). Most others (e.g., Codeine, Morphine, Buprenorphine, Naloxone) showed ND at 100,000 ng/mL. The study confirmed expected positive and negative results at ±25% of cutoff for various other pharmacological compounds. |
| Interference (Endogenous & Preservative) | No significant interference observed with common endogenous substances or preservatives at relevant concentrations. | No significant interference observed for most compounds (e.g., Acetone, Ascorbic Acid, Bilirubin, Creatinine, Ethanol, Glucose, Hemoglobin, Urea, etc.) at tested concentrations (up to 6000 mg/dL).Boric Acid: Showed interference at 1000 mg/dL at ±25% of cutoff (Negative result for both 225 ng/mL and 375 ng/mL oxycodone samples), but for 300 ng/mL cutoff, it showed Negative at 150 ng/mL and Positive at 450 ng/mL, indicating it might behave differently at higher concentrations or different % of cutoff. |
| Specific Gravity Interference | No deviations from expected results over a range of specific gravity values. | No deviations from expected positive or negative results across specific gravity 1.000 to 1.030 when spiked at ±25% of cutoff (225 ng/mL and 375 ng/mL oxycodone). |
| pH Interference | No deviations from expected results over a physiological pH range. | No deviations from expected results across pH levels from 3 to 11 when spiked at ±25% of cutoff (225 ng/mL and 375 ng/mL oxycodone). |
Study Details:
As this is a chemical immunoassay and not an AI/ML device, certain questions are not relevant. However, I will answer the applicable points:
2. Sample sizes used for the test set and data provenance:
- Test set (Clinical Samples for Method Comparison):
- 100 ng/mL Cutoff: 82 unaltered clinical samples
- 300 ng/mL Cutoff: 90 unaltered clinical samples
- Precision Studies: Spiked pooled negative urine samples, tested in replicates, resulting in 88 total determinations per concentration level across 22 days.
- Analytical Recovery Studies: Serially diluted spiked pooled negative urine, each sample run in 10 replicates.
- Cross-reactivity and Interference Studies: Spiked pooled negative human urine, tested in duplicates.
- Data Provenance: Not explicitly stated, but clinical samples typically imply real-world specimens. The location (country of origin) and retrospective/prospective nature are not specified in this summary.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable as the ground truth for this chemical assay was established by a well-defined analytical standard: Liquid Chromatography–Mass Spectrometry (LC/MS), which is considered a gold standard for drug quantitative analysis in urine. This is a laboratory-based, objective ground truth, not reliant on human expert interpretation of images or other subjective data.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- Not applicable. Ground truth for chemical assays is based on quantitative analytical methods (LC/MS), not on expert consensus or adjudication.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable. This is a standalone chemical immunoassay, not an AI-assisted diagnostic tool that involves human readers.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Yes, the performance characteristics described (Precision, Analytical Recovery, Method Comparison, Cross-reactivity, Interference, Specific Gravity, pH Interference) represent the standalone performance of the LZI Oxycodone III Enzyme Immunoassay. The device generates a quantitative or semi-quantitative result directly from the sample. Humans are involved in operating the automated clinical chemistry analyzer and interpreting the results, but a "human-in-the-loop" performance study as typically understood for AI models (e.g., radiologists interpreting images with or without AI guidance) is not relevant here.
7. The type of ground truth used:
- Confirmed Analytical Result: Gas Chromatography/Mass Spectrometry (GC/MS) or Liquid Chromatography/Mass Spectrometry (LC/MS). For the method comparison studies, LC/MS was explicitly stated as the confirmation method for oxycodone and oxymorphone concentrations.
8. The sample size for the training set:
- Not applicable. This is a laboratory-developed and validated chemical assay, not an AI/ML model that requires a "training set" in the machine learning sense. The assay's parameters (e.g., reagent concentrations, antibody specificity) are developed and optimized through traditional biochemical and analytical chemistry methods, rather than by training on a dataset.
9. How the ground truth for the training set was established:
- Not applicable for the same reason as #8. The "ground truth" for developing such a chemical assay lies in analytical chemistry principles and the use of well-characterized reference standards and controls to determine its performance characteristics.
Ask a specific question about this device
(25 days)
The LZI Fentanyl II Enzyme Immunoassay is intended for the qualitative determination of norfentanyl in human urine at the cutoff value of 5 ng/mL. The assay is designed for prescription use with a number of automated clinical chemistry analyzers.
The assay provides only a preliminary analytical result. A more specific alternative chemical method (e.g., gas or liquid chromatography and mass spectrometry) must be used in order to obtain a confirmed analytical result. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary test result is positive.
The LZI Fentanyl II Enzyme Immunoassay is a homogeneous enzyme immunoassay with readyto-use liguid reagents. The assay is based on competition between drug in the sample and drug labeled with the enzyme glucose-6-phosphate dehydrogenase (G6PDH) for a fixed amount of antibody in the reagent. The drug-labeled G6PDH conjugate is traceable to a commercially available fentanyl standard and referred to as fentanyl-labeled G6PDH conjugate. Enzyme activity decreases upon binding to the antibody, and the drug concentration in the sample is measured in terms of enzyme activity. In the absence of drug in the sample, fentanyl-labeled G6PDH conjugate is bound to antibody, and the enzyme activity is inhibited. On the other hand, when free drug is present in the sample, antibody would bind to free drug; the unbound fentanyllabeled G6PDH then exhibits its maximal enzyme activity. Active enzyme converts nicotinamide adenine dinucleotide (NAD) to NADH, resulting in an absorbance change that can be measured spectrophotometrically at 340 nm.
The LZI Fentanyl II Enzyme Immunoassay is a kit comprised of two reagents, an R1 and R2 which are bottled separately but sold together within the kit. The LZI Fentanyl II Enzyme Immunoassay is traceable to a commercially available fentanyl standard.
The R1 solution contains mouse monoclonal anti-fentanyl antibody, glucose-6-phosphate (G6P) nicotinamide adenine dinucleotide (NAD), stabilizers, and sodium azide (0.09 %) as a preservative. The R2 solution contains glucose-6-phosphate dehydrogenase (G6PDH) labeled with fentanyl in buffer with sodium azide (0.09 %) as a preservative.
The LZI Fentanyl II Enzyme Immunoassay is intended for the qualitative determination of norfentanyl in human urine at a cutoff value of 5 ng/mL. The device provides a preliminary analytical result, and a more specific alternative chemical method (e.g., gas or liquid chromatography and mass spectrometry) must be used for a confirmed analytical result.
Here's an analysis of the acceptance criteria and the study that proves the device meets them:
1. A table of acceptance criteria and the reported device performance:
| Acceptance Criteria Category | Specific Criteria | Reported Device Performance |
|---|---|---|
| Precision | Ability to consistently produce correct qualitative results (Negative/Positive) for samples with concentrations at various percentages around the 5 ng/mL cutoff (0%, 25%, 50%, 75%, 100%, 125%, 150%, 175%, 200%). | Within Run (N=22): All 0%, 25%, 50%, 75% negative samples were Negative. All 125%, 150%, 175%, 200% positive samples were Positive. 100% cutoff samples were 13 Neg/9 Pos (indicating some variability around the cutoff, which is expected for qualitative assays at the cutoff). Total Precision (N=88): All 0%, 25%, 50%, 75% negative samples were Negative. All 125%, 150%, 175%, 200% positive samples were Positive. 100% cutoff samples were 59 Neg/29 Pos. |
| Method Comparison (Qualitative Accuracy) | Comparison of EIA results with LC/MS confirmation for clinical samples, especially focusing on false positives and false negatives within different concentration ranges relative to the cutoff. Acceptable levels of agreement and defined explanations for discrepancies. | Overall Accuracy: 20/20 negative samples (by LC/MS) were EIA Negative. 19/19 samples <50% of cutoff (by LC/MS) were EIA Negative. 2/10 samples Near Cutoff Negative (by LC/MS) were EIA Negative, while 8 were EIA Positive. 10/10 Near Cutoff Positive samples (by LC/MS) were EIA Positive. 40/40 High Positive samples (by LC/MS) were EIA Positive. Discrepant Samples: 9 samples with LC/MS norfentanyl concentrations between 1.5 ng/mL and 4.2 ng/mL (below the 5 ng/mL cutoff) were reported as EIA Positive. These were attributed to "levels of fentanyl that contributed to the false positive result" (cross-reactivity from fentanyl itself). |
| Cross-reactivity | Identification of compounds (structurally related, unrelated pharmacological, endogenous, preservatives) that might interfere with the assay, and quantification of their cross-reactivity. Acceptable cross-reactivity levels (e.g., "Not Detected" or specific percentage) for various substances. | Fentanyl and Metabolites: Fentanyl showed 131.58% cross-reactivity at 3.8 ng/mL. Norfentanyl showed 100.00% cross-reactivity at 5 ng/mL (as expected). Many other fentanyl analogs and metabolites showed significant cross-reactivity (e.g., Cyclopropyl Fentanyl 156.25% at 3.2 ng/mL). Structurally Unrelated Compounds: Most tested at 100,000 ng/mL showed "ND" (Not Detected). Dextromethorphan interfered at 40,000 ng/mL, showing 0.01% cross-reactivity and causing a positive result at the -25% Norfentanyl Cutoff (3.75 ng/mL). Endogenous & Preservative Compounds: Most showed no interference. Boric Acid showed interference at 1000 mg/dL, causing a negative result even at the +50% Norfentanyl Cutoff (7.5 ng/mL). |
| Specific Gravity Interference | No qualitative result interference across a range of specific gravity values (1.000 to 1.027) for negative, -25% cutoff, and +25% cutoff samples. | No interference was observed across the tested specific gravity range (1.000 to 1.027) for negative, -25% cutoff, and +25% cutoff samples. |
| pH Interference | No qualitative result interference across a range of pH values (pH 3 to pH 11) for negative, -25% cutoff, and +25% cutoff samples. | No major interference was observed between pH 3 to pH 11 for negative, -25% cutoff, and +25% cutoff samples. |
Important Note: The document focuses on demonstrating substantial equivalence to the predicate device and does not explicitly state numerical "acceptance criteria" for precision or accuracy, but rather reports the performance of the device against established test designs (e.g., NCCLS-EP5 for precision) and confirms qualitative results against a gold standard (LC/MS). The tables demonstrate the device's performance characteristics.
2. Sample size used for the test set and the data provenance:
- Precision Test Set (Qualitative): 88 replicates per concentration level (for 9 concentration levels from 0 to 200% of cutoff), making a total of 792 individual tests. These were prepared by spiking known norfentanyl standards into a pool of negative human urine.
- Method Comparison - Clinical Samples (Qualitative Accuracy): 100 unaltered clinical samples.
- Data Provenance: Samples were collected by Lin-Zhi International, Inc. and from various clinical labs: University of California, San Francisco (UCSF) (San Francisco, CA), Northwest Physicians Labs (NWPL) (Bellevue, Washington), Soloniuk Pain Clinic (Redding, CA), and Calgary Labs (Calgary, Canada). The study represents a retrospective analysis of these collected samples.
- Cross-reactivity, Endogenous, Preservative, Specific Gravity, and pH Interference Tests: Samples were prepared by spiking various compounds into pooled negative human urine or by adjusting the urine's properties. These were not external clinical samples in the same way as the method comparison set.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):
The ground truth for clinical samples (method comparison) was established using LC/MS (Liquid Chromatography/Mass Spectrometry) analysis. LC/MS is a highly specific and sensitive analytical technique, often considered the gold standard for drug quantification in toxicology. This method does not typically involve human experts establishing a subjective "ground truth" but rather relies on the analytical results of the instrument. The document does not specify the qualifications of the personnel performing the LC/MS analysis.
For the other test sets (precision, cross-reactivity, interference), the ground truth was established by the known concentrations of spiked compounds or the known properties of the prepared urine samples.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
There was no adjudication method involving multiple human readers described for establishing the ground truth or interpreting the results.
- For the precision study, the qualitative results (Negative/Positive) were derived directly from the automated clinical analyzer's response compared to the cutoff calibrator.
- For the method comparison (clinical samples), the ground truth was solely determined by the LC/MS reference method.
- For interference studies, the outcomes were based on the instrument's qualitative determination relative to the cutoff.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
No MRMC comparative effectiveness study was done. This device is an in vitro diagnostic (IVD) immunoassay performed on an automated clinical chemistry analyzer, not an AI-assisted diagnostic imaging or interpretative tool that would involve human readers.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
Yes, the performance studies described are essentially standalone (algorithm/device-only performance). The LZI Fentanyl II Enzyme Immunoassay operates on an automated clinical analyzer (Beckman Coulter AU480), and its performance metrics (precision, qualitative accuracy against LC/MS, cross-reactivity, interference) are determined directly from the instrument's output based on its chemical reactions and detection system, without human interpretation in the diagnostic process.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
The primary ground truth used was analytical reference methods, specifically:
- LC/MS (Liquid Chromatography/Mass Spectrometry) for confirming the concentration of norfentanyl in clinical urine samples. This is considered an objective gold standard for drug testing.
- Known concentrations/properties for prepared spiked samples (for precision, cross-reactivity, and interference studies).
8. The sample size for the training set:
This document describes a premarket notification for an immunoassay kit, which is a chemical and enzymatic assay, not a machine learning or AI algorithm. Therefore, there is no "training set" in the context of data used to train an AI model. The device's performance is based on its reagents' chemical properties and reaction kinetics.
9. How the ground truth for the training set was established:
As there is no training set for an AI model, this question is not applicable. The device's performance is intrinsically linked to the validated formulation of its reagents and the instrument's analytical capabilities, which are established through general scientific and manufacturing practices, not data-driven AI training.
Ask a specific question about this device
(29 days)
The LZI Tramadol Enzyme Immunoassay is intended for the qualitative and semi-quantitative determination of tramadol in human urine at the cutoff value of 100 ng/mL when calibrated against tramadol. The assay is designed for prescription use with a number of automated clinical chemistry analyzers.
The semi-quantitative mode is for purposes of (1) enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as gas or liquid chromatography/mass spectrometry (GC/MS or LC/MS) or (2) permitting laboratories to establish quality control procedures.
The assay provides only a preliminary analytical result. A more specific alternative chemical method (e.g., gas or liquid chromatography and mass spectrometry) must be used in order to obtain a confirmed analytical result. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary test result is positive.
The LZI Tramadol Enzyme Immunoassay is a homogeneous enzyme immunoassay with readyto-use liquid reagents. The assay is based on competition between drug in the sample and drug labeled with the enzyme glucose-6-phosphate dehydrogenase (G6PDH) for a fixed amount of antibody in the reagent. The drug-labeled G6PDH conjugate is traceable to a commercially available tramadol standard and referred to as tramadol-labeled G6PDH conjugate. Enzyme activity decreases upon binding to the antibody, and the drug concentration in the sample is measured in terms of enzyme activity. In the absence of drug in the sample, tramadol-labeled G6PDH conjugate is bound to antibody, and the enzyme activity is inhibited. On the other hand, when free drug is present in the sample, antibody would bind to free drug; the unbound tramadol-labeled G6PDH then exhibits its maximal enzyme activity. Active enzyme converts nicotinamide adenine dinucleotide (NAD) to NADH, resulting in an absorbance change that can be measured spectrophotometrically at 340 nm.
The LZI Tramadol Enzyme Immunoassay is a kit comprised of two reagents, an R1 and R2 which are bottled separately but sold together within the kit. The LZI Tramadol Enzyme Immunoassay is traceable to a commercially available tramadol standard.
The Ri solution contains mouse monoclonal anti-tramadol antibody, glucose-6-phosphate (G6P) nicotinamide adenine dinucleotide (NAD), stabilizers, and sodium azide (0.09 %) as a preservative. The R2 solution contains glucose-6-phosphate dehydrogenase (G6PDH) labeled with tramadol in buffer with sodium azide (0.09 %) as a preservative.
The provided text describes the performance characteristics of the LZI Tramadol Enzyme Immunoassay. Here's a breakdown of the requested information:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly list "acceptance criteria" for precision, recovery, or method comparison in a consolidated table with specific numerical targets like "sensitivity > X%" or "specificity > Y%". Instead, it presents the raw performance data and concludes that the results are acceptable or show no significant interference. However, using the precision study's implicit goal (correct qualitative classification for samples ± 25% of cutoff) and the method comparison's agreement, and the recovery study's target, we can construct the following:
| Acceptance Criterion (Implicit) | Reported Device Performance | Study |
|---|---|---|
| Precision: Samples at 0%, 25%, 50%, 75% of cutoff should be Negative. Samples at 125%, 150%, 175%, 200% of cutoff should be Positive. (For 100% cutoff, some variance is expected due to inherent assay variability near the cutoff). | 100% Qualitative Agreement for 0, 25, 50, 75, 125, 150, 175, 200 ng/mL samples over 22 days/88 replicates. Samples at 100 ng/mL appropriately showed mixed results. | Precision (Qualitative) |
| Analytical Recovery: Determined average percent recovery for spiked samples to be between 85% and 115% of target concentration. | Recovery ranged between 89% - 116%. | Analytical Recovery |
| Method Comparison (Qualitative Accuracy): High percentage agreement with LC/MS for clinical samples. (Implicitly close to 100% after addressing near-cutoff discrepancies). | Overall % Agreement: 97.7% (Semi-Quantitative) and 95.3% (Qualitative) | Method Comparison - Clinical Samples |
| Cross-reactivity: Structurally unrelated compounds should not cause false positives or interfere with results at ±25% of cutoff concentrations when present at high concentrations (e.g., 100,000 ng/mL). | Most tested compounds showed no cross-reactivity and did not interfere with negative or positive trampadol samples. | Cross-reactivity & Structurally Unrelated Compounds |
| Endogenous/Preservative Interference: Common endogenous and preservative compounds should not cause false positives or interfere with results at ±25% of cutoff concentrations when present at high concentrations (e.g., 1000 mg/dL). | Boric Acid interfered at 1% w/v but no other significant interference observed. | Endogenous and Preservative Compound Interference |
| Specific Gravity Interference: No interference observed across a range of specific gravity (1.000 to 1.030) for samples at 0 ng/mL, 75 ng/mL, and 125 ng/mL tramadol. | No interference observed. | Specific Gravity Interference |
| pH Interference: No major interference observed between pH 3 to pH 11 for samples at 0 ng/mL, 75 ng/mL, and 125 ng/mL tramadol. | No major interference observed. | pH Interference |
2. Sample Size Used for the Test Set and Data Provenance
- Precision Test Set: Not clinical samples. Tramadol samples were prepared by spiking a tramadol standard into a pool of negative human urine.
- For qualitative and semi-quantitative modes: 8 concentrations (0, 25, 50, 75, 100, 125, 150, 175, 200 ng/mL). Each concentration was tested in duplicate, two runs a day for 22 days, totaling 88 replicates per concentration.
- Analytical Recovery: Not clinical samples. A drug-free urine pool spiked with tramadol at 400 ng/mL was serially diluted. Each dilution was run in 10 replicates.
- Method Comparison - Clinical Samples: Eighty-six (86) unaltered clinical samples.
- Provenance: Samples were collected by Lin-Zhi International, Inc. (LZI) and the University of California at San Francisco (UCSF, San Francisco). This indicates a retrospective collection of clinical samples from a US-based origin.
- Cross-reactivity, Endogenous/Preservative Interference, Specific Gravity, pH Interference: Not clinical samples. Prepared by spiking substances into pooled negative human urine. All samples were tested in duplicates (except for Analytical Recovery, which was 10 replicates).
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
N/A. This device is an in-vitro diagnostic assay for tramadol in urine. The ground truth for the clinical samples was established using a "more specific alternative chemical method," specifically LC/MS (Liquid Chromatography/Mass Spectrometry), which is a gold standard analytical technique for drug quantification, not human expert interpretation. For the spiked samples (precision, recovery, interference studies), the ground truth was the known concentration of the spiked substance, not expert consensus.
4. Adjudication Method for the Test Set
N/A. The ground truth for clinical samples was established by LC/MS analysis, which is an objective chemical measurement. There was no human expert adjudication of results, as there would be in an image analysis or diagnostic interpretation study.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done
No, an MRMC comparative effectiveness study was not done. This device is an automated enzyme immunoassay, not an AI-assisted diagnostic tool that involves human reader interpretation. Therefore, there is no "human readers improve with AI vs without AI assistance" effect size to report. The study focuses on the analytical performance of the immunoassay itself in comparison to a confirmatory method (LC/MS).
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done
Yes, the studies presented are standalone (algorithm only) performance evaluations of the LZI Tramadol Enzyme Immunoassay. The assay is automated on a clinical chemistry analyzer, and its results are compared directly to known concentrations or LC/MS results without human interpretative input as part of its primary performance.
7. The Type of Ground Truth Used
- For Clinical Samples (Method Comparison): LC/MS (Liquid Chromatography/Mass Spectrometry) was used as the confirmatory gold standard for tramadol concentrations.
- For Spiked Samples (Precision, Analytical Recovery, Cross-reactivity, Interference Studies): The ground truth was the known concentration of tramadol or interfering substance spiked into negative human urine.
8. The Sample Size for the Training Set
N/A. This is an immunoassay, not a machine learning model. It does not have a "training set" in the context of AI/ML. The assay's performance is based on its chemical and enzymatic reactions, calibrated using known standards.
9. How the Ground Truth for the Training Set was Established
N/A. As mentioned above, there is no "training set" for this immunoassay device. The assay uses calibrators (0, 50, 100, 225, and 400 ng/mL tramadol) to establish its calibration curve, which determines the relationship between enzyme activity and tramadol concentration. These calibrators represent known, established concentrations.
Ask a specific question about this device
(90 days)
The LZI Cotinine II Enzyme Immunoassay is intended for the quantitative determination of cotinine in human urine at the cutoff value of 200 ng/mL when calibrated against cotinine. The assay is intended as an aid in the detection of cotinine after use or exposure to tobacco products. The assay is designed for prescription use with a number of automated clinical chemistry analyzers.
The semi-quantitative mode is for purposes of (1) enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as gas or liquid chromatography/mass spectrometry (GC/MS or LC/MS) or (2) permitting laboratories to establish quality control procedures.
The assay provides only a preliminary analytical result. A more specific alternative chemical method (e.g., gas or liquid chromatography and mass spectrometry) must be used in order to obtain a confirmed analytical result. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary test result is positive.
The LZI Cotinine II Enzyme Immunoassay is a homogeneous enzyme immunoassay with readyto-use liguid reagents. The assay is based on competition between drug in the sample and drug labeled with the enzyme glucose-6-phosphate dehydrogenase (G6PDH) for a fixed amount of antibody in the reagent. Enzyme activity decreases upon binding to the antibody, and the drug concentration in the sample is measured in terms of enzyme activity. In the absence of drug in the sample, cotinine-labeled G6PDH conjugate is bound to antibody, and the enzyme activity is inhibited. On the other hand, when free drug is present in the sample, antibody would bind to free drug; the unbound cotinine-labeled G6PDH then exhibits its maximal enzyme activity. Active enzyme converts nicotinamide adenine dinucleotide (NAD) to NADH, resulting in an absorbance change that can be measured spectrophotometrically at 340 nm.
The LZI Cotinine II Enzyme Immunoassay is a kit comprised of two reagents, an R1 and R2 which are bottled separately but sold together within the kit.
The R1 solution contains mouse monoclonal anti-cotinine antibody, glucose-6-phosphate (G6P) nicotinamide adenine dinucleotide (NAD), stabilizers, and sodium azide (0.09 %) as a preservative. The R2 solution contains glucose-6-phosphate dehydrogenase (G6PDH) labeled with cotinine in buffer with sodium azide (0.09 %) as a preservative.
The provided text describes the LZI Cotinine II Enzyme Immunoassay, its intended use, and performance characteristics, but it does not describe a study that directly establishes acceptance criteria and directly proves the device meets specific acceptance criteria in a structured manner you requested.
The document is a 510(k) Summary of Safety and Effectiveness, which typically outlines the device's characteristics and demonstrates substantial equivalence to a predicate device. It includes performance data such as precision, linearity, and method comparison (accuracy) with clinical samples, which implicitly serve to show the device is safe and effective for its stated intended use. However, explicit acceptance criteria values that the device must meet are not directly listed in a table format with corresponding proof of meeting them.
I will extract the closest information available to address your request, making inferences where necessary based on the context of common performance expectations for such devices.
Here's a breakdown of the information as requested, largely derived from the "Method Comparison - Clinical Samples" and "Precision" sections.
Acceptance Criteria and Device Performance for LZI Cotinine II Enzyme Immunoassay
Note: The document does not explicitly state "acceptance criteria" in a definitive table format with corresponding proof. Instead, it presents performance study results that are implicitly considered acceptable for demonstrating substantial equivalence. The "Reported Device Performance" below is directly from the study results. The "Acceptance Criteria" are inferred based on the observed performance and typical expectations for diagnostic assays.
1. Table of Acceptance Criteria and Reported Device Performance
| Performance Metric | Inferred Acceptance Criterion (Approximate) | Reported Device Performance (from Clinical Samples) | Study Comments |
|---|---|---|---|
| Accuracy (Semi-Quantitative) | High overall agreement with LC/MS, with minimal discrepancies, especially away from the cutoff concentration. | Overall Agreement: ~96.4% for positive, ~98.7% for negative. Discrepant Samples: - 1 sample (128.6 ng/mL LC/MS) was positive by EIA but was between 50% below cutoff and cutoff concentration (100-199.9 ng/mL). The report states it was "discrepant" but the EIA result (positive) aligns with the stated LC/MS value being within the range where a positive result starts to become more likely if it's "near cutoff positive". However, interpreting that specific discrepancy is complex without a clear cut-off definition of "positive" for LC/MS in the table. - 1 sample (204.2 ng/mL LC/MS) was negative by EIA but was between cutoff and 50% above cutoff concentration (200-299.9 ng/mL). The report states it was "discrepant". | The accuracy study uses LC/MS as the gold standard. The reported percentages reflect agreement for positive and negative results with respect to the 200 ng/mL cut-off. The discrepancies highlight inherent variability near the cutoff. |
| Accuracy (Qualitative) | High overall agreement with LC/MS, similar to semi-quantitative. | Overall Agreement: ~96.4% for positive, ~98.7% for negative. Discrepant Samples: - 1 sample (128.6 ng/mL LC/MS) was positive by EIA (146.6 mAU) with a qualitative cutoff rate of 126.4 mAU. This sample was LC/MS positive but between 100-199.9 ng/mL, the EIA result was positive. - 1 sample (204.2 ng/mL LC/MS) was negative by EIA (93.8 mAU) with a qualitative cutoff rate of 126.4 mAU. This sample was LC/MS positive (204.2 ng/mL) but the EIA result was negative. | Similar to semi-quantitative accuracy, with concordance defined by the qualitative response based on mAU compared to a cutoff rate. |
| Precision (Qualitative) | At 200 ng/mL cutoff, some variability is expected. For concentrations significantly above or below the cutoff, 100% agreement with expected result (positive/negative) is expected. | At 200 ng/mL: - Within Run (N=22): 10 Neg/12 Pos - Total Precision (N=88): 51 Neg/37 Pos At 250-400 ng/mL: 100% Positive At 0-150 ng/mL: 100% Negative | This indicates that at the exact cutoff concentration (200 ng/mL), there is expected variability in classification. This is typical for assays and is why results near the cutoff require careful interpretation and often confirmatory testing. |
| Precision (Semi-Quantitative) | Same as qualitative. At 200 ng/mL cutoff, some variability is expected. For concentrations significantly above or below the cutoff, 100% agreement with expected result (positive/negative) is expected. | At 200 ng/mL: - Within Run (N=22): 14 Neg/8 Pos - Total Precision (N=88): 55 Neg/33 Pos At 250-400 ng/mL: 100% Positive At 0-150 ng/mL: 100% Negative | The semi-quantitative precision results also show variability at the 200 ng/mL cutoff, with 100% agreement for samples sufficiently above or below the cutoff. This confirms the device's ability to consistently classify samples away from the decision point. |
| Linearity | Percent recovery between 85% and 115% of expected values across the assay's linear range. | Range 100 ng/mL to 1000 ng/mL: Recovery ranging from 97.9% to 110.8%. At 20 ng/mL: 167.5% recovery. At 0 ng/mL: N/A (reported 19.3 ng/mL determined). | The device demonstrated good linearity within the critical range for its intended use (100-1000 ng/mL). The higher recovery at 20 ng/mL suggests reduced accuracy at very low concentrations, which is generally not a concern given the 200 ng/mL cutoff. |
| Cross-reactivity | Minimal or no significant cross-reactivity with common substances, particularly structurally unrelated compounds, at physiologically relevant concentrations or concentrations higher than expected. | Cotinine-related compounds: Some cross-reactivity with compounds like (-) Norcotinine (20.00%) and (R, S)-Norcotinine (23.53%). Structurally Unrelated Pharmacological Compounds: "ND" (Not Detected) at high concentrations (e.g., 100,000 ng/mL) for most tested drugs. Negative results for 0 ng/mL cotinine spiked with these compounds, and correct positive/negative results for cotinine controls. | The device largely demonstrates specificity against a wide panel of common drugs and related substances, indicating minimal interference for its intended purpose. Some expected cross-reactivity with specific cotinine metabolites is noted. |
| Endogenous Interference | No significant interference from common endogenous substances at relevant concentrations. | No significant undesired cross-reactants or endogenous substance interference was observed except for Boric Acid (1000 ng/mL) and Citric Acid (pH 3) (800 ng/mL) which showed interference at ±25% of cutoff and also at ±50% of cutoff. | Most endogenous substances showed no interference. Boric acid and citric acid showed interference, which is important for laboratories to be aware of when testing samples that might contain these substances. |
| Specific Gravity Interference | No interference observed across the tested range. | No interference was observed in samples ranging from 1.005 to 1.028 specific gravity when spiked with cotinine at control concentrations. | This indicates the device is robust to variations in urine specific gravity, which is a common factor in urine drug testing. |
| pH Interference | No major interference observed across the tested pH range. | No major interference between pH 3 to pH 11. | This demonstrates the device's robustness across a wide physiological pH range for urine, an important aspect for clinical utility. |
2. Sample Size Used for the Test Set and Data Provenance
-
Accuracy (Method Comparison - Clinical Samples):
- Sample Size: 104 unaltered clinical samples.
- Data Provenance: Samples were collected by Lin-Zhi International, Inc. (LZI). The country of origin is not specified, but LZI is based in Santa Clara, CA, USA. The samples are indicated to be "clinical samples," implying they were prospectively collected from human subjects for diagnostic purposes. They are referred to as "unaltered," suggesting they were tested as received without intentional modification. The study is retrospective in the sense that these samples were then tested against a new assay.
-
Precision:
- Sample Size: For each cotinine concentration, 22 determinations were made within a run, and a total of 88 determinations were made across multiple runs (2 runs/day for 22 days).
- Data Provenance: Samples were prepared by spiking a cotinine standard into a "pool of negative human urine." This indicates a controlled laboratory setting (likely in-house at LZI) rather than clinical samples, and thus not tied to a specific country of origin from patients.
-
Linearity, Cross-reactivity, Endogenous Interference, Specific Gravity, pH Interference: These studies used spiked samples prepared in a laboratory setting (e.g., "drug free-urine pool," "pool of negative human urine"). The sample sizes for each condition are detailed in the tables (e.g., "10 replicates" for linearity, "replicates" for others).
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
- For the Accuracy (Method Comparison - Clinical Samples): The ground truth was established by LC/MS (Liquid Chromatography/Mass Spectrometry). This is an analytical laboratory method, not typically performed by "experts" in the sense of clinicians or radiologists. The result from the LC/MS machine is the ground truth. Therefore, the concept of "number of experts" and their "qualifications" is not applicable here as the ground truth is an objective chemical measurement.
4. Adjudication Method for the Test Set
- For the accuracy study, the "ground truth" was established by LC/MS, which is a definitive analytical method. Therefore, no human adjudication method (like 2+1, 3+1 consensus) was used or required for determining the confirmed cotinine concentration. The comparison was directly between the LZI Cotinine II Enzyme Immunoassay results and the LC/MS results.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
- No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is typical for imaging devices where human readers interpret images, sometimes with and without AI assistance, to assess AI's impact on human performance. The LZI Cotinine II Enzyme Immunoassay is an in-vitro diagnostic (IVD) device (a laboratory test), not an imaging device that would involve human readers for interpretation in this manner.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done
- Yes, the performance studies described are essentially standalone evaluations of the LZI Cotinine II Enzyme Immunoassay. The accuracy, precision, linearity, and interference studies assess the analytical performance of the device itself (the "algorithm" in a broad sense, referring to the assay's methodology and reagents) against established reference methods or known concentrations, without requiring human intervention for interpretation beyond operating the automated clinical analyzer and analyzing the data. The results (e.g., ng/mL concentrations or positive/negative classifications) are generated directly by the assay on the automated analyzer.
7. The Type of Ground Truth Used
- Primary Ground Truth for Clinical Accuracy: LC/MS (Liquid Chromatography/Mass Spectrometry) for cotinine concentrations. This is a highly specific and sensitive analytical technique, often considered the gold standard for drug confirmation testing.
- Ground Truth for Other Studies (Precision, Linearity, Cross-reactivity, etc.): Known concentrations of cotinine or other substances (spiked samples) in negative human urine pools.
8. The Sample Size for the Training Set
- The document describes performance studies for device validation. It does not provide information about a "training set" in the context of an AI/machine learning model. This device is an enzyme immunoassay, a chemical assay, not an AI-based diagnostic algorithm that typically undergoes a separate training phase with a large dataset. The "training" for such an assay would be its initial development and optimization, not a dataset-driven training process in the AI sense.
9. How the Ground Truth for the Training Set Was Established
- As explained in point 8, the concept of a "training set" for an AI/machine learning model is not applicable to this enzyme immunoassay device. Therefore, how its ground truth was established is not relevant here. The development of the assay's chemical reagents and methodology would have involved internal validation and optimization, but not a "training set" in the context of AI.
Ask a specific question about this device
(29 days)
The LZI Methadone II Enzyme Immunoassay is an in vitro diagnostic test intended for the qualitative and semi-quantitative determination of methadone in human urine. The cutoff for both the qualitative and semi-quantitative modes of the assay is 300 ng/mL for methadone. The assay is designed for prescription use on automated clinical chemistry analyzers.
The semi-quantitative mode is for purposes of (1) enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as gas or liquid chromatography/mass spectrometry (GCMS or LC/MS) or (2) permitting laboratories to establish quality control procedures.
The assay provides only a preliminary analytical result. A more specific alternative analytical chemistry method must be used in order to obtain a confirmed analytical result. Gas or liquid chromatography/mass spectrometry (GC/MS or LC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary test result is positive.
The LZI Methadone II Enzyme Immunoassay is a homogeneous enzyme immunoassay with ready-to-use liquid reagents. The assay is based on competition between methadone in the sample and methadone labeled with the enzyme glucose-6-phosphate dehydrogenase (G6PDH) for a fixed amount of antibody in the reagent. Enzyme activity decreases upon binding to the antibody, and the methadone concentration in the sample is measured in terms of enzyme activity. In the absence of methadone in the sample, methadone-labeled G6PDH conjugate is bound to antibody, and the enzyme activity is inhibited. On the other hand, when free methadone is present in the sample, antibody would bind to free methadone; the unbound methadone-labeled G6PDH then exhibits its maximal enzyme activity. Active enzyme converts nicotinamide adenine dinucleotide (NAD) to NADH, resulting in an absorbance change that can be measured spectrophotometrically at 340 nm.
The LZI Methadone II Enzyme Immunoassay is a kit comprised of two reagents, an R1 and R2 which are bottled separately but sold together within the kit.
The Ri solution contains mouse monoclonal anti-methadone antibody, glucose-6-phosphate (G6P) nicotinamide adenine dinucleotide (NAD), stabilizers, and sodium azide (0.09 %) as a preservative. The R2 solution contains glucose-6-phosphate dehydrogenase (G6PDH) labeled with methadone in buffer with sodium azide (0.09 %) as a preservative.
The LZI Methadone II Enzyme Immunoassay is an in-vitro diagnostic test for qualitative and semi-quantitative determination of methadone in human urine, with a cutoff of 300 ng/mL. The device's performance was evaluated through various studies demonstrating its precision, linearity, accuracy against LC/MS, cross-reactivity with other substances, and interference from endogenous compounds and pH levels.
Here is a summary of the acceptance criteria and reported device performance from the provided document:
1. Table of Acceptance Criteria and Reported Device Performance
| Device Feature | Acceptance Criteria (Implied / Expected) | Reported Device Performance |
|---|---|---|
| Qualitative Precision | Consistent positive/negative results for samples 25% away from cutoff (e.g., all negative for ≤225 ng/mL, all positive for ≥375 ng/mL). | Within Run (N=22):- 0 ng/mL, 75 ng/mL, 150 ng/mL, 225 ng/mL: 22 Negative- 300 ng/mL: 17 Neg / 5 Pos- 375 ng/mL, 450 ng/mL, 525 ng/mL, 600 ng/mL: 22 PositiveTotal Precision (N=88):- 0 ng/mL, 75 ng/mL, 150 ng/mL, 225 ng/mL: 88 Negative- 300 ng/mL: 59 Neg / 29 Pos- 375 ng/mL, 450 ng/mL, 525 ng/mL, 600 ng/mL: 88 Positive |
| Semi-Quantitative Precision | Consistent positive/negative results for samples 25% away from cutoff. | Within Run (N=22):- 0 ng/mL, 75 ng/mL, 150 ng/mL, 225 ng/mL: 22 Negative- 300 ng/mL: 18 Neg / 4 Pos- 375 ng/mL, 450 ng/mL, 525 ng/mL, 600 ng/mL: 22 PositiveTotal Precision (N=88):- 0 ng/mL, 75 ng/mL, 150 ng/mL, 225 ng/mL: 88 Negative- 300 ng/mL: 66 Neg / 22 Pos- 375 ng/mL, 450 ng/mL, 525 ng/mL, 600 ng/mL: 88 Positive |
| Linearity | Recovery values between 85% - 115% of expected values. | Samples from 100 ng/mL to 1000 ng/mL showed recovery ranging from 93.7% to 104.9%. |
| Semi-Quantitative Accuracy (vs. LC/MS) | High percentage agreement with LC/MS results. | % Agreement: 97.8% (Positive), 97.9% (Negative) |
| Qualitative Accuracy (vs. LC/MS) | High percentage agreement with LC/MS results. | % Agreement: 97.8% (Positive), 97.9% (Negative) |
| Cross-reactivity (Structurally Related) | Low cross-reactivity for other methadone-related compounds, and no interference with unrelated compounds. | Methadone: 100.00%EDDP, EMDP: <0.30%Nor-LAAM HCl: 0.40%LAAM HCl: 1.20%Alpha-Methadol: 1.13%Isomethadone HCl: 0.86%Structurally Unrelated Pharmacological Compounds: No significant interference observed for 50+ compounds at high concentrations (100,000 ng/mL, except Duloxetine at 50,000 ng/mL). |
| Endogenous Compound Interference | No significant interference with common endogenous substances at specified concentrations. | Boric Acid: Interference observed at 1% w/v (1000 ng/mL) causing negative results for both 225 ng/mL and 375 ng/mL methadone. No other significant interference observed for the other 12 compounds tested. |
| Specific Gravity Interference | No interference across a range of specific gravity. | No interference observed for samples ranging from 1.003 to 1.028 with and without methadone spikes. |
| pH Interference | No interference across a range of pH levels. | No major interference between pH 3 to pH 11. |
2. Sample Size Used for the Test Set and Data Provenance
- Precision Studies (Qualitative & Semi-quantitative):
- Within Run: 22 data points (2 replicates, 1 run/day for 22 days) for each concentration level (0 ng/mL, 75 ng/mL, 150 ng/mL, 225 ng/mL, 300 ng/mL, 375 ng/mL, 450 ng/mL, 525 ng/mL, 600 ng/mL).
- Total Precision: 88 data points (2 replicates, 2 runs/day for 22 days) for each concentration level.
- Provenance: Samples were prepared by spiking a methadone standard into a pool of negative human urine. Therefore, this data is from controlled laboratory conditions, not directly from patients.
- Method Comparison - Clinical Samples (Accuracy Study):
- Sample Size: A total of ninety-four (94) unaltered clinical samples.
- Provenance: Samples were collected by Lin-Zhi International, Inc. (LZI) and from the University of California, San Francisco (UCSF). This indicates human clinical data.
- Linearity: Serially diluted methadone spiked urine, each run in 10 replicates for a range of 0 to 1000 ng/mL.
- Cross-reactivity and Endogenous Compound Interference: Various concentrations of each substance spiked into pooled negative human urine. (Not specified if replicates were performed for all, but for structurally unrelated compounds, it states "All samples were tested in replicates.")
- Specific Gravity and pH Interference: Urine samples (negative and spiked with methadone) adjusted to different specific gravity and pH levels respectively.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
The document does not explicitly state the number or qualifications of experts used to establish ground truth.
For the accuracy studies with clinical samples, the ground truth was established by LC/MS (Liquid Chromatography/Mass Spectrometry), which is an independent and highly sensitive analytical chemistry method. This method itself acts as the "gold standard" or "ground truth" for drug concentration determination, rather than relying on human expert interpretation of the test results.
4. Adjudication Method for the Test Set
Not applicable. The ground truth for concentrations (precision, linearity, cross-reactivity, interference studies) was determined analytically. For clinical samples, the ground truth for methadone concentration was established by LC/MS, acting as an objective reference standard. No human adjudication process involving multiple readers is described for the test results.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done
No, an MRMC comparative effectiveness study was not done. This device is an in-vitro diagnostic assay for chemical analysis, not an imaging device or AI diagnostic tool that typically involves human reader interpretation. Therefore, the concept of human readers improving with or without AI assistance is not relevant here.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
Yes, the studies presented are all standalone performance evaluations of the LZI Methadone II Enzyme Immunoassay device itself, without human interpretation in the results. The device outputs a quantitative or qualitative result, which is then compared against established ground truths (e.g., LC/MS, spiked concentrations).
7. The Type of Ground Truth Used
- For Precision, Linearity, Cross-reactivity, and Interference Studies: The ground truth was established by preparing urine samples with known, spiked concentrations of methadone or other compounds.
- For Method Comparison (Accuracy) with Clinical Samples: The ground truth for methadone and methadone metabolite concentrations was established using LC/MS (Liquid Chromatography/Mass Spectrometry).
8. The Sample Size for the Training Set
The document describes performance studies, which represent a test set. It does not mention a specific "training set" or "validation set" in the context of an algorithm or machine learning. For an immunoassay, the "training" typically refers to the assay development and optimization process, not a separate data set in the same way as an AI model.
9. How the Ground Truth for the Training Set Was Established
As noted above, the concept of a "training set" for an algorithm with specific ground truth establishment is not directly applicable to this immunoassay device in the provided document. The development and optimization of the immunoassay (analogous to "training" in AI) would involve establishing reactivity to methadone and minimal cross-reactivity via laboratory experiments and chemical principles, rather than a distinct training data set with ground truth in the AI sense.
Ask a specific question about this device
(216 days)
The LZI Fentanyl Enzyme Immunoassay is intended for the qualitative determination of norfentanyl in human urine at the cutoff value of 5 ng/mL when calibrated against norfentanyl. The assay is designed for prescription use with a number of automated clinical chemistry analyzers.
The assay provides only a preliminary analytical result. A more specific alternative chemical method (e.g., gas or liquid chromatography and mass spectrometry) must be used in order to obtain a confirmed analytical result.
Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary test result is positive.
The LZI Fentanyl Enzyme Immunoassay is a homogeneous enzyme immunoassay with ready-to-use liquid reagents. The assay is based on competition between drug in the sample and drug labeled with the enzyme glucose-6-phosphate dehydrogenase (G6PDH) for a fixed amount of antibody in the reagent. Enzyme activity decreases upon binding to the antibody, and the drug concentration in the sample is measured in terms of enzyme activity. In the absence of drug in the sample, norfentanyl-labeled G6PDH conjugate is bound to antibody, and the enzyme activity is inhibited. On the other hand, when free drug is present in the sample, antibody would bind to free drug; the unbound norfentanyl-labeled G6PDH then exhibits its maximal enzyme activity. Active enzyme converts nicotinamide adenine dinucleotide (NAD) to NADH, resulting in an absorbance change that can be measured spectrophotometrically at 340 nm.
The LZI Fentanyl Enzyme Immunoassay is a kit comprised of two reagents, an R1 and R2 which are bottled separately but sold together within the kit.
The R1 solution contains mouse monoclonal anti-fentanyl antibody, glucose-6-phosphate (G6P) nicotinamide adenine dinucleotide (NAD), stabilizers, and sodium azide (0.09 %) as a preservative. The R2 solution contains glucose-6-phosphate dehydrogenase (G6PDH) labeled with fentanyl in buffer with sodium azide (0.09 %) as a preservative.
The LZI Fentanyl Enzyme Immunoassay calibrators and controls are designated for use at the 5 ng/mL cutoff contain 0, 2.5, 3.75, 5, 6.25, 10, and 20 ng/mL of norfentanyl in human urine with sodium azide (0.09 %) as a preservative. These five calibrators and two controls are sold as individual bottles.
Here's an analysis of the acceptance criteria and study findings for the LZI Fentanyl Enzyme Immunoassay, based on the provided document:
This document is a 510(k) summary for a diagnostic device, which typically focuses on demonstrating substantial equivalence to a predicate device rather than comprehensive clinical effectiveness. Therefore, some of the requested information (like multi-reader multi-case studies, effect size of AI, or detailed training set information for an AI algorithm) is not applicable or present in this type of submission. The device described is a traditional enzyme immunoassay, not an AI-based system.
1. Acceptance Criteria and Reported Device Performance
The document describes performance characteristics typically assessed for an immunoassay. The primary acceptance criteria are implied by the precision and method comparison studies. For qualitative drug tests, the key is the accuracy of positive and negative determinations around the cutoff.
| Acceptance Criteria Category | Specific Metric (Implied) | Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|---|---|
| Precision | % CV near cutoff | Low % CV (e.g., < 20%) | 3.1% at 5 ng/mL cutoff (Within Run), 4.0% (Total Precision) |
| Qualitative Agreement | Agreement for positives | 100% (or very high) | 100.0 % agreement with positive samples |
| Agreement for negatives | High (e.g., > 90%) | 86.5 % agreement with negative samples | |
| Interference | No significant interference | No significant interference | Interference observed with Boric Acid (1% w/v) and dextromethorphan; no other significant undesired cross-reactants or endogenous substance interference observed. |
Note on Acceptance Criteria: For 510(k) submissions, the acceptance criteria are often benchmarked against the predicate device's performance or established standards for that device type. Explicit numerical acceptance criteria are not always stated directly in the summary document but are implicit in the study design and the finding of "substantial equivalence."
2. Sample Size and Data Provenance
- Test Set Sample Size:
- Precision (Qualitative): 88 determinations per concentration (N=88) spread across different runs/days.
- Method Comparison - Clinical Samples: 101 clinical unaltered samples.
- Data Provenance: The document does not explicitly state the country of origin or whether the data was retrospective or prospective. Given the nature of a 510(k) submission for an in-vitro diagnostic, these studies are typically conducted by the manufacturer in a controlled environment, likely using a mix of spiked and anonymized clinical samples. The "unaltered clinical samples" suggests prospective or retrospective collection from a clinical setting, but further details are not provided.
3. Number of Experts used to establish Ground Truth & Qualifications
- Number of Experts: Not applicable/not specified. For an immunoassay measuring a chemical compound, the "ground truth" is typically established by an independent, highly accurate analytical method, not human expert interpretation.
- Qualifications of Experts: Not applicable.
4. Adjudication Method
- Adjudication Method: Not applicable. The ground truth (confirmatory method) for chemical analysis in this context does not involve human adjudication.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- Was it done?: No. This type of study is relevant for imaging or interpretation-based AI systems where human readers are involved. This device is an automated immunoassay.
6. Standalone (Algorithm Only) Performance Study
- Was it done?: Yes, in a sense. The performance characteristics (precision, method comparison, interference) describe the standalone performance of the LZI Fentanyl Enzyme Immunoassay on an automated clinical chemistry analyzer (AU680 Analyzer), without human interpretation being part of the primary measurement process. The results are quantitative (ΔOD, mAU) or qualitative (positive/negative) based on predefined cutoff values.
7. Type of Ground Truth Used
- Ground Truth: For the method comparison study with clinical samples, the ground truth was established by a "more specific alternative chemical method," specifically gas or liquid chromatography/mass spectrometry (GC/MS or LC/MS). This is considered the gold standard for confirming the presence and concentration of chemical compounds in forensic and clinical toxicology.
8. Sample Size for the Training Set
- Training Set Sample Size: Not applicable. This is a traditional immunoassay, not an AI/machine learning algorithm that requires a "training set" in that sense. The device's performance is based on its chemical and enzymatic reactions, which are inherent to its design and formulation, not learned from data. Development might involve optimizing reagents, but not "training" like an AI model.
9. How the Ground Truth for the Training Set was Established
- Ground Truth for Training Set: Not applicable, as there is no "training set" in the AI sense. The development of such an immunoassay involves analytical chemistry principles and empirical testing to optimize sensitivity, specificity, and other performance parameters.
Ask a specific question about this device
(423 days)
The LZI Carisoprodol Metabolite (Meprobamate) Enzyme Immunoassay is intended for the qualitative and semi-quantitative determination of carisoprodol metabolite (meprobamate) in human urine at a cutoff value of 100 ng/mL when calibrated against meprobamate. The assay is designed for prescription use with a number of automated clinical chemistry analyzers.
The semi-quantitative mode is for purposes of enabling laboratories to determine an appropriate dilution of the specimen for verification by a confirmatory method such as GC/MS, LC/MS or permitting laboratories to establish quality control procedures.
The assay provides only a preliminary analytical result. A more specific alternative chemical confirmatory method (e.g., gas or liquid chromatography and mass spectrometry) must be used to obtain a confirmed analytical result. Clinical consideration and professional judgment must be exercised with any drug of abuse test, particularly when the preliminary test result is positive.
The LZI Carisoprodol Metabolite (Meprobamate) Enzyme Immunoassay is a homogeneous enzyme immunoassay with ready-to-use liquid reagents. The assay is a kit comprised of two reagents (antibody/substrate reagent R1 and enzyme-drug conjugate reagent R2), which are bottled separately but sold together within the kit. The Ri solution contains mouse monoclonal anti-meprobamate antibody, glucose-6- phosphate (G6P), nicotinamide adenine dinucleotide (NAD), stabilizers, and sodium azide (0.09 %) as a preservative. The R2 solution contains meprobamate-labeled glucose-6-phosphate dehydrogenase (G6PDH) in buffer with sodium azide (0.09 %) as a preservative.
Here's a breakdown of the acceptance criteria and the study details for the LZI Carisoprodol Metabolite (Meprobamate) Enzyme Immunoassay, based on the provided document:
Acceptance Criteria and Reported Device Performance
| Acceptance Criteria Category | Specific Acceptance Criteria (Inferred from study results and regulatory requirements) | Reported Device Performance |
|---|---|---|
| Accuracy (Qualitative Mode) | High agreement with confirmatory method (LC/MS or GC/MS) for positive and negative samples, particularly around the cutoff. | Qualitative Analysis Agreement: - Positive Agreement: 98.5% - Negative Agreement: 96.6% *2 false positives in "Near Cutoff Negative" (92 and 98 ng/mL, both identified as Negative by GC/MS/LC/MS) *1 false negative in "Near Cutoff Positive" (103 ng/mL, identified as Positive by GC/MS/LC/MS) |
| Accuracy (Semi-Quantitative Mode) | High agreement with confirmatory method (LC/MS or GC/MS) for positive and negative samples, particularly around the cutoff. | Semi-Quantitative Analysis Agreement: - Positive Agreement: 98.5% - Negative Agreement: 98.3% *1 false positive in "Near Cutoff Negative" (98 ng/mL, identified as Negative by GC/MS/LC/MS) *1 false negative in "Near Cutoff Positive" (103 ng/mL, identified as Positive by GC/MS/LC/MS) |
| Reproducibility/Precision (Qualitative Mode) | Consistent results at specific concentrations, especially near the cutoff. | At cutoff (100 ng/mL, GC/MS confirmed 94.9 ng/mL): 40 Negative / 48 Positive out of 88 determinations. At -25% Cutoff (75 ng/mL, GC/MS confirmed 76.8 ng/mL): 88 Negative / 0 Positive. At +25% Cutoff (125 ng/mL, GC/MS confirmed 122.3 ng/mL): 0 Negative / 88 Positive. |
| Reproducibility/Precision (Semi-Qualitative Mode) | Consistent results at specific concentrations, especially near the cutoff. | At cutoff (100 ng/mL, GC/MS confirmed 94.9 ng/mL): 60 Negative / 28 Positive out of 88 determinations. At -25% Cutoff (75 ng/mL, GC/MS confirmed 76.8 ng/mL): 88 Negative / 0 Positive. At +25% Cutoff (125 ng/mL, GC/MS confirmed 122.3 ng/mL): 0 Negative / 88 Positive. |
| Linearity/Reportable Range | Acceptable recovery across a range of concentrations. | Mean Recovery: 66.6% to 109.2% for expected concentrations from 10 ng/mL to 400 ng/mL. (Note: 10 ng/mL showed lower recovery, but higher concentrations were generally good). |
| Analytical Specificity (Endogenous Compounds) | No interference from common endogenous compounds at specified concentrations. | No positive or negative interference observed for the listed endogenous compounds when meprobamate was at -25% or +25% of the cutoff. |
| Analytical Specificity (Urine Preservatives) | No significant interference from common urine preservatives. | Sodium azide and sodium fluoride: No interference. Boric acid: Caused false negative results at +25% cutoff (125 ng/mL) and up to +125% cutoff (225 ng/mL). Labeling mitigation required. |
| Analytical Specificity (pH) | No interference across a physiological pH range. | No positive or negative interference observed at urine pH values ranging from 3 to 11. |
| Analytical Specificity (Specific Gravity) | No interference across a physiological specific gravity range. | No positive or negative interference observed at urine specific gravities ranging from 1.002 to 1.029. |
| Cross-Reactivity (Structurally Related Compounds) | Acceptable cross-reactivity profile. | Carisoprodol (90.9%), Felbamate (25.0%), Meprobamate-N-Glucuronide (0.5%), Hydroxymeprobamate (0.2%). Other listed compounds showed <0.1% cross-reactivity. |
| Interference (Structurally Unrelated Compounds) | No interference from common structurally unrelated drugs. | No positive or negative interference observed from a wide range of common drugs/substances (e.g., Acetaminophen, Codeine, Ibuprofen) at high concentrations (e.g., 100,000 ng/mL) when meprobamate was at -25% or +25% of the cutoff. |
| Clinical Validity of Cutoff | Meprobamate levels measurable for a relevant post-dose period. | Pharmacokinetic studies of carisoprodol and meprobamate in plasma, and meprobamate elimination in urine, indicated levels above 100 ng/mL are present for "up to a few days" post-single dose. |
Study Details
2. Sample size used for the test set and the data provenance
-
Sample Size for Test Set:
- Precision Study: 88 determinations per concentration level for
each of 8 concentrations (a total of 704 determinations for qualitative and 704 for semi-quantitative). - Method Comparison Study: 127 clinical urine samples.
- Linearity Study: Each sample run in 10 replicates across 11 concentrations.
- Analytical Specificity (Endogenous, Preservatives, pH, Specific Gravity, Unrelated Compounds): Samples tested in duplicate.
- Cross-reactivity (Structurally Related Compounds): Not explicitly stated, but typically involves multiple runs.
- Precision Study: 88 determinations per concentration level for
-
Data Provenance: Retrospective clinical urine samples "from subjects that were (b) (4) taking prescribed or nonprescribed Carisoprodol." The origin country is not explicitly stated, but given this is an FDA submission, it's likely primarily US-based or from regions with comparable medical practices.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- The ground truth for the test set was established by Gas Chromatography/Mass Spectrometry (GC/MS) or Liquid Chromatography/Mass Spectrometry (LC/MS). These are established analytical methods for drug confirmation. There is no mention of "experts" in the sense of medical professionals (e.g., radiologists) establishing ground truth, as this is an in vitro diagnostic device. The "experts" in this context would be the technicians/scientists operating and interpreting the GC/MS/LC/MS results, who are implied to be qualified in their field based on the use of these validated methods.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- There was no human adjudication process described. The reference method (GC/MS or LC/MS) provided the definitive "ground truth" for each sample.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No, this was not a multi-reader multi-case (MRMC) comparative effectiveness study. This device is an automated enzyme immunoassay for chemical analysis, not an imaging-based AI system that would typically involve human readers. Therefore, the concept of "human readers improve with AI vs without AI assistance" does not apply here.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Yes, this was a standalone (algorithm only) performance study. The LZI Carisoprodol Metabolite (Meprobamate) Enzyme Immunoassay is an automated system designed to provide a preliminary analytical result without human interpretation of the assay itself. Human involvement comes in confirming positive results with alternative chemical methods and clinical judgment, as stated in the intended use.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- The ground truth used was confirmatory analytical methods, specifically Gas Chromatography/Mass Spectrometry (GC/MS) or Liquid Chromatography/Mass Spectrometry (LC/MS). These methods are considered the gold standard for drug quantification and identification in urine.
8. The sample size for the training set
- The document does not explicitly describe a separate "training set" for the device. As an enzyme immunoassay, the "training" (calibration) involves using known calibrator concentrations (0, 50, 100, 200, and 400 ng/mL meprobamate, as mentioned in the linearity section) to establish the standard curve against which unknown samples are measured. There isn't an "algorithm" in the conventional AI sense that undergoes training with a large dataset. The performance characteristics described are related to the analytical validation of the assay itself.
9. How the ground truth for the training set was established
- As there isn't a traditional "training set" in the AI sense, the concept of ground truth for it doesn't directly apply. The "ground truth" for the calibrators used in the assay system would be their known, precisely manufactured concentrations of meprobamate, traceable to a commercially available meprobamate source material with 99% analytical purity (as mentioned in section L.1.c).
Ask a specific question about this device
(136 days)
The LZI Methadone Metabolite (EDDP) Enzyme Immunoassay is an in vitro diagnostic test intended for the qualitative and semi-quantitative determination of Methadone Metabolite in human urine. The cutoff for both the qualitative and semi-quantitative modes of the assay are 100 ng/mL and 300 ng/mL for methadone metabolite. The assay is designed for prescription use on automated clinical chemistry analyzers.
The semi-quantitative mode is for purposes of (1) enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as gas or liquid chromatography/mass spectrometry (GCMS or (2) permitting laboratories to establish quality control procedures.
The assay provides only a preliminary analytical result. A more specific alternative analytical chemistry method must be used in order to obtain a confirmed analytical result. Gas or liguid chromatography/mass spectrometry (GC/MS or LC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary test result is positive.
The LZI Methadone Metabolite (EDDP) (100 and 300) Calibrators are for use as calibrators in the qualitative and semi-quantitative calibration of the LZI Methadone Metabolite (EDDP) Enzyme Immunoassay at the cutoff values of 100 ng/mL and 300 ng/mL.
The LZI Methadone Metabolite (EDDP) Enzyme Immunoassay is a homogeneous enzyme immunoassay with ready-to-use liquid reagents. The assay is based on competition between EDDP in the sample and EDDP labeled with the enzyme glucose-6-phosphate dehydrogenase (G6PDH) for a fixed amount of antibody in the reagent. Enzyme activity decreases upon binding to the antibody, and the EDDP concentration in the sample is measured in terms of enzyme activity. In the absence of EDDP in the sample, EDDP-labeled G6PDH conjugate is bound to antibody, and the enzyme activity is inhibited. On the other hand, when free EDDP is present in the sample, antibody would bind to free EDDP; the unbound EDDP-labeled G6PDH then exhibits its maximal enzyme activity. Active enzyme converts nicotinamide adenine dinucleotide (NAD) to NADH, resulting in an absorbance change that can be measured spectrophotometrically at 340 nm.
The LZI Methadone Metabolite (EDDP) Enzyme Immunoassay is a kit comprised of two reagents, an R1 and R2 which are bottled separately but sold together within the kit.
The Ri solution contains mouse monoclonal anti-methadone metabolite antibody, glucose-6phosphate (G6P) nicotinamide adenine dinucleotide (NAD), stabilizers, and sodium azide (0.09 %) as a preservative. The R2 solution contains glucose-6-phosphate dehydrogenase (G6PDH) labeled with methadone metabolite in buffer with sodium azide (0.09 %) as a preservative.
The LZI Methadone Metabolite (EDDP) Enzyme Immunoassay calibrators and controls designated for use at the 100 ng/mL cutoff contain 0, 50, 75, 100, 125, 250, 500 ng/mL of methadone metabolite (EDDP) in human urine with sodium azide (0.09 %) as a preservative. These five calibrators and two controls are sold as individual bottles.
The LZI Methadone Metabolite (EDDP) Enzyme Immunoassay calibrators and controls designated for use at the 300 ng/mL cutoff contain 0, 150, 225, 300, 375, 600, and 1000 ng/mL of methadone metabolite (EDDP) in human urine with sodium azide (0.09 %) as a preservative. These five calibrators and two controls are sold as individual bottles.
The provided document is a 510(k) premarket notification for the LZI Methadone Metabolite (EDDP) Enzyme Immunoassay and Calibrators. It focuses on demonstrating substantial equivalence to a predicate device, rather than establishing acceptance criteria and proving conformance to them in the same way a de novo or PMA submission might.
Therefore, the acceptance criteria are largely implied by the comparison to the predicate device and the analytical performance data presented. The study aims to show that the new device performs comparably to the predicate and provides accurate results for methadone metabolite detection.
Here's an attempt to extract the requested information based on the provided text, with notable limitations due to the nature of the document:
1. Table of Acceptance Criteria and Reported Device Performance
| Performance Characteristic | Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|---|
| Precision (100 ng/mL Cutoff) | Consistency in qualitative results (Negative/Positive) at various concentrations, particularly near the cutoff, demonstrating minimal variation within and between runs. For concentrations <75 ng/mL, results should be consistently negative. For concentrations >125 ng/mL, results should be consistently positive. At the 100 ng/mL cutoff, a mix of positive and negative results is expected due to inherent variability, but overall agreement with expected ranges should be demonstrated. | Semi-Quantitative Results:- 0, 25, 50, 75 ng/mL: 100% Negative (Within Run N=22, Total N=88)- 100 ng/mL (Cutoff): Within Run: 11 Neg/11 Pos (50%); Total: 40 Pos/48 Neg (45.5% Pos)- 125, 150, 175, 200 ng/mL: 100% Positive (Within Run N=22, Total N=88)Qualitative Results:- 0, 25, 50, 75 ng/mL: 100% Negative (Within Run N=22, Total N=88)- 100 ng/mL (Cutoff): Within Run: 13 Neg/9 Pos (40.9% Pos); Total: 34 Pos/54 Neg (38.6% Pos)- 125, 150, 175, 200 ng/mL: 100% Positive (Within Run N=22, Total N=88) |
| Precision (300 ng/mL Cutoff) | Consistency in qualitative results (Negative/Positive) at various concentrations, particularly near the cutoff, demonstrating minimal variation within and between runs. For concentrations <225 ng/mL, results should be consistently negative. For concentrations >375 ng/mL, results should be consistently positive. At the 300 ng/mL cutoff, a mix of positive and negative results is expected due to inherent variability, but overall agreement with expected ranges should be demonstrated. | Semi-Quantitative Results:- 0, 75, 150, 225 ng/mL: 100% Negative (Within Run N=22, Total N=88)- 300 ng/mL (Cutoff): Within Run: 6 Neg/16 Pos (72.7% Pos); Total: 52 Pos/36 Neg (59.1% Pos)- 375, 450, 525, 600 ng/mL: 100% Positive (Within Run N=22, Total N=88)Qualitative Results:- 0, 75, 150, 225 ng/mL: 100% Negative (Within Run N=22, Total N=88)- 300 ng/mL (Cutoff): Within Run: 7 Neg/15 Pos (68.2% Pos); Total: 55 Pos/33 Neg (62.5% Pos)- 375, 450, 525, 600 ng/mL: 100% Positive (Within Run N=22, Total N=88) |
| Method Comparison - Clinical Samples (100 ng/mL Cutoff) | High concordance with LC/MS results, especially for samples clearly positive or negative relative to the cutoff. Discrepancies should be understood and ideally minimal, particularly for samples significantly above/below the cutoff. The device should demonstrate appropriate sensitivity and specificity compared to a confirmatory method. | Qualitative/Semi-Quantitative Accuracy Study (N=87):- Agreement for 23 Negative, 11 <50% of cutoff, 9 Near Cutoff Neg, 40 High Pos.- 2 discordant samples: Sample #45 (LC/MS 103.1 ng/mL, EIA Negative); Sample #46 (LC/MS 126.0 ng/mL, EIA Negative). These were "Near Cutoff Pos" by LC/MS but negative by EIA. This indicates the EIA may be less sensitive for results very close to the cutoff, which is acceptable for a screening assay requiring confirmation. |
| Method Comparison - Clinical Samples (300 ng/mL Cutoff) | High concordance with LC/MS results, especially for samples clearly positive or negative relative to the cutoff. Discrepancies should be understood and ideally minimal, particularly for samples significantly above/below the cutoff. The device should demonstrate appropriate sensitivity and specificity compared to a confirmatory method. | Qualitative/Semi-Quantitative Accuracy Study (N=84):- Agreement for 21 Negative, 15 <50% of cutoff, 6 Near Cutoff Neg, 38 High Pos.- 4 discordant samples: These were "Near Cutoff Pos" by LC/MS but positive by EIA (no specific concentrations provided, but implies the EIA correctly identified them as positive, while LC/MS categorized them differently near the cutoff based on the table structure). The table indicates 4 samples were "Near Cutoff Pos" by LC/MS and "Positive" by EIA. No 'Negative' results for these 'Near Cutoff Pos' samples are shown as in the 100ng/mL cutoff. |
| Cross-reactivity | Minimal or no cross-reactivity with structurally related compounds that are not the target analyte, ensuring specificity of the assay. | 100 ng/mL Cutoff:- EDDP: 100% Cross Reactivity- EMDP, Methadone, LAAM HCl, (±)-α-Methadol, (-)-Isomethadone HCl, (-)-α-Noracetylmethadol (Nor-LAAM) HCl: <0.1% or <0.2% Cross Reactivity (i.e., consistently Negative at high spiked concentrations).300 ng/mL Cutoff:- EDDP: 100% Cross Reactivity- EMDP, Methadone, LAAM HCl, (±)-α-Methadol, (-)-Isomethadone HCl, (-)-α-Noracetylmethadol (Nor-LAAM) HCl: <0.1% Cross Reactivity (i.e., consistently Negative at high spiked concentrations). |
| Endogenous Compound Interference & Specific Gravity | No significant undesired interference from common endogenous compounds or variations in specific gravity. | "No significant undesired cross-reactants or endogenous substance interference was observed." |
| Intended Use | The device should be reliably used for qualitative and semi-quantitative determination of Methadone Metabolite in human urine at specified cutoffs (100 ng/mL and 300 ng/mL) on automated clinical chemistry analyzers for prescription use, providing preliminary analytical results requiring confirmatory methods. | The performance data supports the claims for qualitative and semi-quantitative determination at both 100 ng/mL and 300 ng/mL cutoffs, indicating it functions as intended for preliminary screening. |
2. Sample Size Used for the Test Set and Data Provenance
-
Precision Studies:
- For each concentration level tested (e.g., 0, 25, 50, 75, 100, 125, 150, 175, 200 ng/mL for 100 ng/mL cutoff, and similarly for 300 ng/mL cutoff), there were N=22 samples for "Within Run" analysis and N=88 samples for "Total Precision" analysis. This suggests 4 runs (88/22).
- Total number of unique samples for precision is not explicitly stated as distinct samples across all concentrations, but it's at least 9 concentrations x 22 samples for within-run and 9 concentrations x 88 for total precision for each cutoff. These are likely replicates of specific concentrations rather than 88 unique patient samples.
- Data Provenance: Not specified (e.g., country of origin). Likely laboratory-prepared spiked samples for precision. Retrospective or Prospective is not stated but typically spiked samples are prepared for the study.
-
Method Comparison - Clinical Samples:
- 100 ng/mL Cutoff: N=87 clinical unaltered samples.
- 300 ng/mL Cutoff: N=84 clinical unaltered samples.
- Data Provenance: "clinical unaltered samples." Not specified (e.g., country of origin, retrospective or prospective), but implies real patient samples.
-
Cross-reactivity: The sample sizes are implied by the "Spiked []" concentrations for each cross-reactant. Each cross-reactant was tested at a specific concentration. This is generally a set of controlled laboratory-prepared samples.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
- Ground Truth Method: The primary ground truth for the clinical sample method comparison and for validating the calibrators/controls is Gas Chromatography/Mass Spectrometry (GC/MS) or Liquid Chromatography/Mass Spectrometry (LC/MS).
- Number/Qualifications of Experts: The document does not specify the number of experts or their qualifications involved in establishing the GC/MS or LC/MS ground truth. This is a common characteristic of device submissions where a recognized "gold standard" analytical method (like GC/MS or LC/MS for drug testing) is used, and the expertise is assumed to be inherent in the execution of that method by a qualified laboratory.
4. Adjudication Method for the Test Set
- The document does not describe an adjudication method involving multiple human readers or a formal adjudication process beyond using GC/MS/LC/MS as the confirmatory method. The presented data primarily compares the EIA results directly to LC/MS results. For the two discordant samples at the 100 ng/mL cutoff, their LC/MS concentrations are noted, but no further "adjudication" is detailed in the text beyond acknowledging the discrepancy.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done as described in the document. This type of study typically assesses human reader performance with and without AI assistance for tasks like image interpretation. This submission is for an in vitro diagnostic (IVD) immunoassay, not an AI-powered diagnostic imaging device.
6. Standalone Performance Study (Algorithm Only)
- Yes, this entire submission is essentially a standalone performance study of the "algorithm" (the LZI Methadone Metabolite (EDDP) Enzyme Immunoassay). The performance characteristics (precision, method comparison, cross-reactivity) demonstrate the device's analytical performance without human intervention in the interpretation of the raw assay signal, as it's designed for automated clinical chemistry analyzers. The "EIA Result" is the output of the device itself.
7. Type of Ground Truth Used
- The primary ground truth used is confirmatory analytical chemistry methods, specifically Gas Chromatography/Mass Spectrometry (GC/MS) or Liquid Chromatography/Mass Spectrometry (LC/MS). These are considered highly accurate and specific "gold standards" for drug quantification.
8. Sample Size for the Training Set
- The document does not explicitly describe a "training set" in the context of an AI/machine learning model. This is an IVD device, where performance is established through analytical validation studies (precision, accuracy, interference, cross-reactivity) using predefined reagents and methods. The device's "algorithm" is the enzymatic reaction and spectrophotometric measurement, which is inherently deterministic, not a learned model in the AI sense.
- The calibrators are used for the calibration of the assay (0, 50, 100, 250, 500 ng/mL for 100 ng/mL cutoff; 0, 150, 300, 600, 1000 ng/mL for 300 ng/mL cutoff), which is analogous to "setting up" the device for accurate measurement.
9. How the Ground Truth for the Training Set Was Established
- As there is no "training set" in the AI/ML context, this question is not directly applicable. For the calibrators, their concentrations are precisely defined (e.g., 100 ng/mL or 300 ng/mL of methadone metabolite (EDDP)). These are manufactured with known concentrations in a human urine matrix. The "ground truth" for calibrators is therefore their known, manufactured concentration.
Ask a specific question about this device