Found 4 results

510(k) Data Aggregation

    K Number
    K172271

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-05-04

    (280 days)

    Product Code
    Regulation Number
    882.1400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Trackit T4 EEG Amplifier is intended to be used as a front-end amplifier to acquire, store and transmit electrophysiological signals (wireless or cabled).

    Device Description

    The Trackit T4 EEG Amplifier is a multi-channel electroencephalograph designed for use in routine EEG and lab monitoring applications and due to its small size, can be used in ambulatory applications. In this situation, the EEG electrodes are fitted to the patient by a trained clinician prior to the patient being sent home. No subsequent intervention is required by the patient. Upon completion of the recording, the data which is stored on a memory card is reviewed by a clinician using review and analysis software on a PC. It is a compact USB amplifier which provides 32 channels with internal expansion option) with built-in calibration and electrode impedance measurement. Also provided is a Nonin pulse oximeter interface, a Patient Event input and an Aux DC input. Optional wireless communication is available (Bluetooth and WLAN WiFi). There are two variants of the Trackit T4 EEG Amplifier: - Trackit T4-32 providing 24 referential + 8 poly channels. . - . Trackit T4-68 providing 64 referential + 4 poly channels (using internal expansion board). Plug-on Patient Connection Units (PCUs) provide 32 channel touchproof inputs (model T4-PCU 24+8) or 68 channels (model T4-PCU 64+4). The Amplifier is intended to be connected to a USB port on a PC which is powered from a medically approved power supply. In addition it can be battery powered in ambulatory applications. This equipment is intended only as an adjunct device in patient assessment; it must be used in conjunction with other methods of patient diagnosis. The equipment does not sustain or support life.

    AI/ML Overview

    This document describes the Lifelines Trackit T4 EEG Amplifier, which is substantially equivalent to the predicate device, Lifelines R-40 EEG Amplifier (K151600). The information provided is primarily focused on pre-clinical performance testing and safety, not on a study proving a device meets specific acceptance criteria based on diagnostic accuracy or clinical outcomes.

    Here's an analysis of the provided text in the context of your request:

    1. Table of acceptance criteria and the reported device performance:

    The document primarily shows that the new device (T4) performs similarly to its predicate (R-40) in various technical tests. The "acceptance criteria" are implied by the "Test specification" and the "Conclusion" indicating "Same test as predicate" or noting the specific values the predicate device achieved.

    TestTest MethodT4 subject device (Reported Performance)R-40 predicate deviceAcceptance Criteria (Implied)
    Insulation resistanceMeasure resistance between TF1 'Aux. B' socket and USB plug metalwork.> 200 MΩ (3.1)> 200 MΩ (3.1)> 200 MΩ
    Dielectric strengthIncrease voltage to 1500Vrms over 10s and hold for 1 min.No flash-over/breakdown (3.2)No flash-over/breakdown (3.2)No signs of flash-over or breakdown
    Mains on Applied PartsMeasure leakage current.≤ 5000uA AC (3.6)≤ 5000uA AC (3.6)≤ 5000uA AC
    Patient leakageMeasure leakage current.≤ 100uA AC and ≤ 10uA DC (3.8)≤ 100uA AC and ≤ 10uA DC (3.8)≤ 100uA AC and ≤ 10uA DC
    CalibrationCheck for 8mVpp ±0.4 square wave signals on all channels at 1Hz frequency.8mVpp ±0.4 square wave (4.4)8mVpp ±0.4 square wave (4.4)8mVpp ±0.4 square wave signals on all channels at 1Hz frequency
    Short circuit noiseCheck short circuit noise for all channels.< 1.5uVpp (4.8)< 1.5uVpp (4.7)< 1.5uVpp
    HF responseCheck HF response to 80 Hz.(4.10)(4.9)Response to 80 Hz maintained (implied "Same test as predicate")
    DC responseCheck for a +300mV step and a -300mV step.(4.11)(4.10)+300mV step and -300mV step detected (implied "Same test as predicate")
    Impedance checkCheck all channels indicate an impedance of 3000Ω ±500.3000Ω ±500 (4.16)3000Ω ±500 (4.15)3000Ω ±500

    2. Sample size used for the test set and the data provenance:

    The document describes "pre-clinical performance testing" using the device itself and comparing it to the predicate device. It does not refer to a "test set" of patient data or clinical samples. The tests were performed on the device hardware.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    Not applicable. This was hardware performance testing, not a clinical study involving experts establishing ground truth for diagnostic accuracy.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    Not applicable. As above, this was not a clinical study requiring adjudication of expert opinions.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This device is an EEG amplifier, not an AI-powered diagnostic tool, and no MRMC study was conducted or mentioned. The document explicitly states: "Clinical testing was not performed with this device."

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Not applicable. This device is an EEG amplifier; it does not operate as a standalone algorithm in the way a diagnostic AI would. The performance described is of the hardware itself.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    The "ground truth" for the performance tests was defined by engineering specifications and direct measurements against those specifications, comparing the subject device's performance to that of the predicate device, or standard electrical and safety parameters.

    8. The sample size for the training set:

    Not applicable. This is not a machine learning or AI device that requires a training set.

    9. How the ground truth for the training set was established:

    Not applicable. This is not a machine learning or AI device that requires a training set with established ground truth.

    Summary of the Study:

    The study described is a series of pre-clinical bench tests (hardware performance tests) conducted to demonstrate that the Lifelines Trackit T4 EEG Amplifier achieved similar technical performance and safety characteristics to its predicate device, the Lifelines R-40 EEG Amplifier. The "acceptance criteria" were based on the predicate device's established performance and relevant international safety and performance standards (e.g., IEC 60601-1, IEC 60601-2-26, IEC 60601-1-2, etc.).

    The document concludes that the T4 EEG Amplifier "meets the functional claims and intended use as described in the product labeling" and that its "safety and effectiveness are substantially equivalent to the predicate device." This determination of substantial equivalence relies on technical data rather than clinical performance data with patient populations.

    Ask a Question

    Ask a specific question about this device

    K Number
    K151600

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2015-10-23

    (133 days)

    Product Code
    Regulation Number
    882.1835
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The R-40 EEG Amplifier is intended to be used as a front-end amplifier to acquire, store and transmit electrophysiological signals (wireless or cabled).

    Device Description

    The R-40 EEG Amplifier is a multi-channel electroencephalograph designed for use in routine EEG and lab monitoring applications.

    It is a compact USB 40-channel amplifier which incorporates 32 referential channels and 8 polygraphic channels with built-in calibration and electrode impedance measurement. A Nonin pulse oximeter interface is provided, a Patient Event input, 2 Aux DC inputs and an Electrocap connector. Optional wireless communication is available (Bluetooth and WLAN WiFi).

    The Amplifier is intended to be connected to a USB port on a PC which is powered from a medically approved power supply.

    This equipment is intended only as an adjunct device in patient assessment; it must be used in conjunction with other methods of patient diagnosis. The equipment does not sustain or support life.

    AI/ML Overview

    The provided text is a 510(k) summary for the Lifelines R-40 EEG Amplifier. This document focuses on demonstrating substantial equivalence to a predicate device rather than presenting a detailed study with acceptance criteria and performance metrics in the way a clinical trial or algorithm performance study would.

    Therefore, the information required to populate the table and answer the study-related questions comprehensively is not available in the provided text.

    Here's why and what can be inferred:

    1. A table of acceptance criteria and the reported device performance

    • Not Available. The document states "The Lifeline R-40 EEG Amplifier meets the functional claims and intended use as described in the product labeling." and "The safety and effectiveness are substantially equivalent to the predicate device, Nicolet Wireless EEG Amplifier." However, it does not explicitly list quantitative acceptance criteria (e.g., sensitivity, specificity, accuracy, signal-to-noise ratio within a certain range) and corresponding device performance values against those criteria. The approval is based on demonstrating equivalence rather than meeting specific pre-defined performance thresholds for a novel diagnostic claim.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not Applicable/Not Available. The document explicitly states: "Clinical testing was not performed with this device." This implies there was no "test set" in the context of a clinical performance study. The evaluation was based on engineering and safety testing, and comparison of technical characteristics to the predicate device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not Applicable/Not Available. As no clinical testing was performed, there was no test set requiring ground truth established by experts.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable/Not Available. No test set or clinical data requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No. This device is an EEG amplifier, not an AI-powered diagnostic algorithm designed to assist human readers. Therefore, an MRMC study related to AI assistance is not relevant or applicable here.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable. This is hardware (an EEG amplifier), not an algorithm that would be evaluated for standalone performance.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    • Not Applicable. No clinical ground truth was established as no clinical testing was performed. The "ground truth" for this device's evaluation would be its adherence to engineering standards (IEC 60601-1, IEC 60601-1-2, IEC 60601-2-26) and its functional specifications, which are assessed via engineering tests, not clinical ground truth.

    8. The sample size for the training set

    • Not Applicable. This is hardware, not a machine learning model, so there is no "training set."

    9. How the ground truth for the training set was established

    • Not Applicable. (See point 8).

    Summary of what is known from the document:

    • Device Name: R-40 EEG Amplifier
    • Intended Use: To acquire, store and transmit electrophysiological signals (wireless or cabled).
    • Performance Evaluation Method: Primarily engineering safety testing (IEC 60601 series standards) and comparison of technical characteristics to a legally marketed predicate device (Nicolet Wireless EEG Amplifier, K103140).
    • Clinical Testing: Explicitly stated as "not performed."
    • Conclusion: The device meets functional claims and is substantially equivalent to the predicate device in terms of safety and effectiveness.

    This type of 510(k) submission relies on demonstrating that a new device is "substantially equivalent" to an already legally marketed device, often through technical comparisons and adherence to recognized safety standards, rather than proving novel clinical performance against strict acceptance criteria in a clinical trial.

    Ask a Question

    Ask a specific question about this device

    K Number
    K101691

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2010-08-10

    (55 days)

    Product Code
    Regulation Number
    882.1890
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Lifelines Photic Stimulator is indicated for photic activation of the EEG during an EEG study and in the generation of visual evoked potentials.

    Device Description

    The Photic Stimulator is a device for generating short-duration flashes of white light by means of a solid-state LED (light emitting diode). The flashes are controlled from a host PC and typically occur over a repetition rate of between 1 and 60Hz.

    AI/ML Overview

    This document is a 510(k) summary for the Lifelines Photic Stimulator, which is a medical device for generating flashes of white light used in EEG studies and for generating visual evoked potentials.

    The provided text does not contain a study that proves the device meets specific acceptance criteria in terms of performance metrics like sensitivity, specificity, accuracy, or other clinical outcomes. Instead, it focuses on demonstrating substantial equivalence to a predicate device (EB Neuro, S.P.A. Model BE Plus / AURA-LTM64 Amplifier) based on intended use, technological characteristics, and design.

    Therefore, many of the requested sections about acceptance criteria, study details, and performance metrics cannot be directly extracted from this document as they are not present. This type of submission relies on the established safety and effectiveness of a predicate device rather than presenting new clinical performance data.

    Here's an analysis based on the information available:


    1. A table of acceptance criteria and the reported device performance

    The document does not specify "acceptance criteria" in terms of performance metrics like sensitivity, specificity, or accuracy for a clinical outcome. Instead, the acceptance criteria for the 510(k) submission are based on demonstrating substantial equivalence to a predicate device. This involves matching or demonstrating comparable characteristics in:

    Acceptance Criteria CategoryLifelines Photic Stimulator (Reported Device Performance)Predicate Device (EB Neuro, S.P.A. Model BE Plus / AURA-LTM64)
    Intended UsePhotic activation of EEG during EEG study and generation of visual evoked potentials.Acquisition of EEG, polygraphy, and polysomnography signals and transmission to a PC during neurophysiology examinations. (Note: Predicate's stated use is broader but encompasses photic stimulation functionality)
    Technological Characteristics
    Light sourceSingle hi-intensity LED and associated optics96 LEDs
    Luminous Flux700lm typical, 900 lm max; 13,000 lux at 1 foot144.0 lumens (minimum), 182.4 lumens (typical), 230.4 lumens (maximum)
    Duration1-10ms duration fixed or auto duty cycle, 100ms max5 ms
    Flash Rate1-60Hz or single (manual) flashUp to 60 per second, externally controlled
    InterfacesTTL, USB, Serial5-pin TRIAD male connector; Power supply, controlling signal (trigger); TTL level
    Outputs100mV isolated pulseTTL level
    Power5 - 15V DC12 VDC (±5%)
    CompatibilityFor use with Trackit system, NicoletOne system, or any other compatible system providing interface to USB, TTL, or serial control signalsFor use with Grass Comet or AURA Base Station
    Physical Size (Stimulator)5" diameter x 5" long; Weight 12 ozs.Flash lamp: 8.25" W x 2.5" H x 2" D (Weight 7.1 ozs.); Adapter Box: 3.25" W x 1.75" H x 0.75" D (Weight 1.8 ozs.)
    DesignArm mounted Photic stimulator (generates flashes of white light by means light emitting diode (LED)).Arm mounted photic stimulator (generates flashes of white light by means light emitting diode (LED)).
    Safety TestingEN/IEC 60601-1, EN/IEC 60601-1-2, UL60601-1, CAN/CSA-C22.2 No.601-1-M90, IEC 60825-1:1993+ A1: 1997 + A2:2001 (implied via predicate comparison)EN 60601-1, EN 60601-1-2, ISO 15004-2 (Light Hazard Protection)

    Study Description:
    The "study" presented here is a comparison to a legally marketed predicate device (EB Neuro, S.P.A. Model BE Plus / AURA-LTM64 Amplifier, K053606) to demonstrate substantial equivalence, rather than a clinical trial or performance study on humans or specialized test sets. The report asserts that the Lifelines Photic Stimulator is similar to the predicate device in intended use, technological characteristics, and design. The FDA's letter confirms substantial equivalence based on this comparison.


    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    Not applicable. This is not a study that involved a "test set" of patient data in the conventional sense. The "test" here was a comparison of device specifications and safety standards against a predicate device.


    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. Ground truth for clinical performance was not established as there was no patient-based clinical study. The "ground truth" for the submission was the regulatory safety and performance standards met by the predicate device and the new device's adherence to relevant safety standards.


    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. There was no clinical test set requiring adjudication.


    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a photic stimulator, not an AI-assisted diagnostic tool. No MRMC study was performed.


    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This device is hardware and does not involve an algorithm for standalone performance evaluation in the context of AI.


    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" in this context is the established safety and effectiveness profile of the predicate device, as well as adherence to relevant international and national safety standards (e.g., EN/IEC 60601-1, IEC 60825-1) for medical electrical equipment and light hazard protection.


    8. The sample size for the training set

    Not applicable. This device submission does not involve machine learning or a "training set."


    9. How the ground truth for the training set was established

    Not applicable. This device submission does not involve machine learning or a "training set."


    Ask a Question

    Ask a specific question about this device

    K Number
    K010460

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2001-05-14

    (87 days)

    Product Code
    Regulation Number
    882.1400
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview