510(k) Data Aggregation

    K Number
    K241806

    Validate with FDA (Live)

    Date Cleared
    2025-01-08

    (201 days)

    Product Code
    Regulation Number
    866.3981
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Applied Biosystems TaqPath COVID-19, Flu A, Flu B, RSV Select Panel is a multiplex, real-time reverse transcription polymerase chain reaction (RT-PCR) in vitro diagnostic test for the qualitative detection and differentiation of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), influenza B virus, and respiratory syncytial virus (RSV). Nucleic acids are isolated and purified from nasopharyngeal (NP) swab and anterior nasal (AN) swab specimens obtained from individuals exhibiting signs and symptoms of a respiratory tract infection. Clinical signs and symptoms of respiratory viral infection due to SARS-CoV-2, influenza, and RSV can be similar. This test is intended to aid in the differential diagnosis of SARS-CoV-2, influenza B, and RSV A/B (undifferentiated) infections in humans and is not intended to detect influenza C virus infections.

    Nucleic acids from the viral organisms identified by this test are generally detectable in NP and AN swab specimens during the acute phase of infection and identification of specific viral nucleic acids from individuals exhibiting signs and symptoms of respiratory tract infective of the presence of the identified virus and aids in diagnosis if used in conjunction with other clinical and epidemiological information, and laboratory findings.

    The results of this test should not be used as the sole basis for diagnosis, treatment, or other patient management decisions. Positive results do not rule out coinfection with other organism(s) detected by the Applied Biosystems TaqPath COVID-19, Flu A, Flu B, RSV Select Panel may not be the definite cause of disease. Negative results do not preclude SARS-CoV-2, influenza A virus, influenza B virus, or RSV infections.

    The Applied Biosystems TaqPath COVID-19, Flu A, Flu B, RSV Select Panel is intended for use by qualified and trained clinical laboratory personnel specifically instructed and trained in the techniques of real-time PCR and in vitro diagnostic procedures.

    Device Description

    The Applied Biosystems™ TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel is a multiplex, real-time reverse transcription polymerase chain reaction (RT-PCR) test. The severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), influenza A, influenza B, respiratory syncytial virus (RSV) A/B and RNase P primer and probe sets are designed to detect viral RNA in nasopharyngeal (NP) and anterior nasal (AN) swab specimens from individuals exhibiting signs and symptoms of a respiratory tract infection.

    Each TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel includes the following components:

    • TaqPath™ COVID-19, Flu A, Flu B, RSV Select Assay-Multiplex assays that contain . primer and probe sets specific to the following targets:
      • Three SARS-CoV-2 targets (Orfla, Orf1b, and N genes) .
      • . Two influenza A virus targets (PB1 & M genes)
        • I Two influenza B virus targets (M & NS genes)
        • . Three RSV targets (NP, M, and L protein genes)
        • l RNase P (internal human sample collection control)
    • TaqPath™ COVID-19, Flu A, Flu B, RSV Select Positive Control—Inactivated viral . control that contains SARS-CoV-2, influenza A, influenza B, and RSV.
    • TaqPath™ COVID-19, Flu A, Flu B, RSV Select Negative Control—MS2 packaged RNA . control that contains targets specific to RNase P genomic regions targeted by the assay.
    • TaqPath™ 1-Step Select Master Mix (No ROX)—Ready-to-use PCR mix, including . reverse transcriptase, polymerase, dNTPs, salts, and buffer.
    • . Package Insert -- Provides the instructions and the link to download the instructions for use and other assets (including the ADF)
    • An Assay Definition File (ADF) applicable to the instrument used in the workflow . (available via download).

    In addition to the SARS-CoV-2, influenza A, influenza B and RSV viral assay targets, the assay portion of the panel includes RNase P, which serves as an endogenous internal process control to monitor extraction and amplification of each clinical sample. The TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel also contains external process positive and negative controls. The positive control (PC) component included is an inactivated viral control that contains SARS-CoV-2, influenza A, influenza B, and RSV viruses. The PC monitors extraction and real-time RT-PCR by demonstrating that each of the four viruses can be detected when present and that RNase P is not detected when absent. The negative control (NC) component included is an MS2 packaged RNA control that contains targets specific to RNase P genomic regions targeted by the assay. The NC also monitors extraction and real-time RT-PCR by demonstrating RNase P can be detected when present and that the four viruses are not detected when absent. The TaqPath™ 1-Step Select Master Mix (No ROX) included as a component of the kit is a ready-to-use PCR mix which contains a deoxyribonucleotide triphosphate mix (dNTPs), enzymes, and other components to permit reverse transcription and amplification of the assay targets. The TagPath™ 1-Step Select Master Mix (No ROX) also contains ribonuclease (RNase) inhibitors as well as deoxyuridine triphosphate (dUTP) and uracil N-glycosylase (known as UNG or UDG).

    The TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel is provided in two overall kit configurations: either 200 reactions or 1.000 reactions.

    AI/ML Overview

    The provided text describes the analytical and clinical performance studies of the "Applied Biosystems TaqPath COVID-19, Flu A, Flu B, RSV Select Panel" (hereafter referred to as "the Device"). This device is an in vitro diagnostic test for the qualitative detection and differentiation of SARS-CoV-2, influenza A, influenza B, and respiratory syncytial virus (RSV).

    Below is a breakdown of the acceptance criteria and study details based on the provided information, presented in the requested format.


    Acceptance Criteria and Device Performance

    The acceptance criteria for this diagnostic device are primarily demonstrated through various analytical performance evaluations (Limit of Detection, Inclusivity/Reactivity, Precision, Interfering Substances, Cross-Reactivity, Stability, Carryover) and clinical performance studies (PPA and NPA against a comparator device). The studies aim to show that the device performs as expected and is "substantially equivalent" to a legally marketed predicate device.

    1. Table of Acceptance Criteria and Reported Device Performance

    Study Aspect / Performance MetricAcceptance Criteria (Implied/Directly Stated where available)Reported Device Performance (Summary)
    Limit of Detection (LoD)≥ 95% positive detection at the determined LoD concentrations.All viruses confirmed at ≥ 95% positive detection at the determined LoD concentrations (e.g., SARS-CoV-2 at 107 GCE/mL; Flu A H1N1 at 340 GCE/mL). WHO Standard LoD for SARS-CoV-2 confirmed at 150.36 IU/mL.
    Inclusivity (in silico)Less than 1% (SARS-CoV-2) and less than 5% (Flu A, Flu B, and RSV) of publicly available sequences have mutations that reduce melting temperature below annealing temperature for all gene targets.> 99.99% of SARS-CoV-2, 99.9% of Flu B, and 99.49% of RSV strains met the criteria. For Flu A, 95.27% of strains were considered reactive after wet testing.
    Reactivity (wet-lab)100% detection at approximately 3x LoD.Most tested strains showed 100% detection at approximately 3x LoD. Strains not at 100% initially were further tested until 100% detection was achieved at higher concentrations.
    Precision (Within-Laboratory)N/A (Statistical measures like %CV and %PPA are reported).Positivity Rate (Negative Sample): 0%.
    Overall PPA (1x LoD, 3x LoD): 100.00% for all targets.
    Observed Precision and Repeatability %CV (1x and 3x LoD): < 2%CV for all targets.
    Site-to-Site ReproducibilityQualitative PPA performance: Generally high (e.g., above 90-95% at 1x LoD, 100% at 3x LoD). Quantitative %CV for Cq variability: Generally low (e.g., usually < 5%).NPA (Negative Sample): 99.63%.
    Overall Site-to-Site Reproducibility PPA (1x LoD): 92.59% (SARS-CoV-2) to 98.89% (Flu A).
    Overall Reproducibility PPA (3x LoD): 100.00% for all targets.
    Observed Reproducibility, Within-lab precision, Repeatability %CV (1x & 3x LoD): < 5% CV for all targets.
    Interfering SubstancesAll positive samples remain positive; negative samples remain negative. No false positives (FP) or false negatives (FN).All conditions produced expected results. No interference observed from 26 substances, 2 solvents, 5 VTMs, or extraction reagent carryover, except for expected positive results from FluMist® due to live influenza components.
    Competitive Interference100% positivity for both low-concentration and high-concentration viruses in multi-target samples. No off-target false positives.All conditions satisfied acceptance criteria, showing 100% positivity for both viruses in pairs and no off-target false positives.
    Cross-Reactivity & Microbial InterferenceNo false positive results from tested microbes in negative samples. No inhibition of PCR or interference with extraction/RNase P detection in positive samples with microbes.None of the 49 tested microorganisms produced false positive results or interfered with detection. One of 39 CoV-HKU1 clinical samples produced a positive SARS-CoV-2 result, the rest were negative.
    Specimen Stability100% positivity for all targets at specified storage conditions and time points.100% positivity for NP and AN specimens at all time points tested (RT for 24h, refrigerated for 72h, ≤ -70°C for 30 days).
    Eluate Stability100% positivity for contrived positive sample eluates at specified conditions.100% of contrived positive sample eluates produced expected results for up to 24 hours at 2-8°C, 5 days at ≤ -70°C, and after one freeze/thaw cycle.
    Fresh vs Frozen EquivalencySupport the use of frozen samples in clinical studies and stated freeze/thaw cycles in IFU.All samples showed 100% concordance with expected results, supporting use of frozen samples and specific F/T cycles (up to 2 F/T for Copan/BD, 1 F/T for others).
    In-Use and Hold Time StabilityAll components meet specified stability criteria for storage conditions and freeze/thaw cycles.Assay Mix Tube and Master Mix stable for 72h at 2-8°C and up to 4 F/T cycles. PC and NC stable for 24h at 2-8°C.
    Carryover Cross-ContaminationCarryover/cross-contamination rate of 0%.0% carryover/cross-contamination rate reported among 492 replicates tested.
    RNase P Internal Control Cutoff ConfirmationOverall PPA of 98.07% (NPA 100%) for NP swabs and 92.30% (NPA 100%) for AN swabs. Improved AN swab collection technique reduces invalid rate and shows 100% RNase P detection.NP PPA: 98.07%, NPA: 100%. AN PPA: 92.30%, NPA: 100%. Subsequent study with improved AN collection showed 100% RNase P detection and 100% validity for AN samples.
    Matrix Equivalency100% concordance between NP and AN swab specimens.All samples produced 100% concordance with expected results.
    Clinical Performance (Prospective Study)Demonstrate substantial equivalence based on PPA and NPA compared to a comparator device. Target typically high PPA/NPA (e.g., >85-90%).SARS-CoV-2: NP Overall PPA 96.6%, NPA 96.47%; AN Overall PPA 95.3%, NPA 96.46%.
    Flu A: NP Overall PPA 94.9%, NPA 99.36%; AN Overall PPA 97.8%, NPA 99.06%.
    Flu B: NP Overall PPA 98.0%, NPA 99.87%; AN Overall PPA 100.0%, NPA 99.87%.
    RSV: NP Overall PPA 93.8%, NPA 99.75%; AN Overall PPA 100.0%, NPA 99.6%.
    Clinical Performance (Enrichment Study)Maintain high PPA/NPA for Flu A, Flu B, RSV (where prospective numbers might be low).Flu A: NP Overall PPA 97.9%, NPA 100.0%; AN Overall PPA 100.0%, NPA 95.5%.
    Flu B: NP Overall PPA 92.3%, NPA 98.2%; AN Overall PPA 92.3%, NPA 100.0%.
    RSV: NP Overall PPA 100.0%, NPA 94.9%; AN Overall PPA 100.0%, NPA 96.6%.

    2. Sample Size Used for the Test Set and Data Provenance

    • Clinical Performance Study (Prospective Cohort):

      • Subjects: 1,909 subjects enrolled (1,840 prospectively, 69 enrichment).
      • Test Set (Evaluable Specimens): 1,620 Nasopharyngeal (NP) swabs and 1,541 Anterior Nasal (AN) swabs were included in the final data analysis for the prospective cohort. (Minor adjustments for inconclusive comparator results in Flu A, specifically 1,539 AN and 1,616 NP for Flu A analysis).
      • Enrichment Study: 69 subjects, yielding 69 NP swabs and 68 AN swabs for analysis.
      • Data Provenance:
        • Country of Origin: Not explicitly stated as a single country, but "14 different collection sites across the U.S." indicate United States data.
        • Retrospective or Prospective: Primarily prospective data, with an "enrichment phase" also involving prospective collection from subjects with confirmed positive PCR results for Flu A, Flu B, and/or RSV.
        • Specimen Handling: Specimens were tested either fresh (Category I) or frozen (Category II).
    • Analytical Studies (Test Sets):

      • LoD Confirmation: At least 24 replicates per virus strain/specimen type.
      • Precision (within-lab): 96 replicates for each panel member.
      • Reproducibility (site-to-site): 9 replicates per panel member (3 sites x 3 reagent lots x 1 replicate/lot). Total ~270 replicates per analyte level across 3 sites.
      • Interfering Substances: 3 replicates per substance/condition.
      • Competitive Interference: 3 replicates per condition.
      • Cross-Reactivity & Microbial Interference: 3 replicates per organism/condition.
      • Specimen Stability: Multiple replicates per time point/storage condition.
      • Eluate Stability: Multiple replicates per time point/storage condition.
      • Fresh vs Frozen Equivalency: Multiple replicates per F/T cycle and concentration.
      • In-Use and Hold Time Stability: Multiple replicates per time point/F/T cycle.
      • Carryover Cross-Contamination: 492 replicates (negative wells).
      • RNase P Internal Control Cutoff Confirmation: Individually collected NP and AN swabs (specific number not given for the initial confirmation, but the follow-up AN study used 138 AN samples).
      • Matrix Equivalency: Multiple replicates per matrix and concentration.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • For the clinical performance study, the ground truth appears to be established by a comparator FDA-cleared molecular test.
    • The document does not specify the number or qualifications of human experts (e.g., radiologists, pathologists) involved in establishing the ground truth for any of the studies. This is typical for molecular diagnostic assays where the "ground truth" is often defined by the results of a highly sensitive and specific reference method (e.g., another molecular test or a composite reference method incorporating multiple tests or clinical findings).

    4. Adjudication Method for the Test Set

    • For the clinical performance study, samples that produced a discordant call between the Device and the comparator test were further tested with "another FDA cleared molecular assay" (a resolver assay). This implies a reconciliation method where a third test acts as an adjudicator for discordant results between the device under evaluation and the primary comparator.
    • The specific rules for how the resolver assay's result determined the final ground truth (e.g., 2+1 majority rule, or if the resolver result was the final truth) are not explicitly detailed but are common in diagnostic test evaluation.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    • No, an MRMC comparative effectiveness study was not done. The evaluated device is a molecular diagnostic test (RT-PCR kit), not an imaging AI algorithm that typically involves human readers. The performance is assessed based on the device's ability to detect viral nucleic acids against a comparator molecular test, not on improvements in human reader performance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • Yes, the primary evaluation of this device is a standalone performance study. The "TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel" is a laboratory-based RT-PCR diagnostic kit. Its performance, as detailed in the analytical and clinical studies, is measured by its output (detection/differentiation of viral targets) directly from patient samples, independent of human interpretation or "human-in-the-loop" assistance for the test result itself. The Diomni™ Software provides automated interpretation, making it an algorithm-only result at the interpretation stage.

    7. The Type of Ground Truth Used

    • For analytical studies (e.g., LoD, Inclusivity, Precision, Interference, Cross-Reactivity, Stability), the ground truth was established by contrived samples with known concentrations of viral material or known presence/absence of interfering/cross-reactive substances. This is a common and appropriate method for analytical validation.
    • For the clinical performance study, the ground truth was established by an FDA-cleared molecular test (comparator device), supplemented by a resolver FDA-cleared molecular assay for discordant results. This constitutes a form of "composite reference standard" where the truth is determined by a highly reliable, already approved diagnostic method.

    8. The Sample Size for the Training Set

    • The document describes the validation of a diagnostic kit, not an AI model requiring a separate "training set" in the machine learning sense. The information provided heavily details the test set performance (analytical and clinical validation).
    • While the "Diomni™ Software" performs data analysis and interpretation based on "assay-specific parameters from the Assay Definition File (ADF)," the document does not specify a distinct training set size typically associated with machine learning model development. The "training" of such a system would typically involve internal development data used to establish parameters like Cq cutoffs, but this is not detailed as a separate "training set" with specific sample numbers. The development study that determined the RNase P IC cutoff of Cq = 33.0 could be considered part of the "training" or parameter establishment phase, but no sample size is given for it.

    9. How the Ground Truth for the Training Set Was Established

    • As noted above, a distinct "training set" for an AI/algorithm is not explicitly described in the context of this diagnostic kit's submission. The "ground truth" for establishing device parameters (like Cq cutoffs) would generally come from extensive analytical experiments and internal validation studies using characterized samples (e.g., known concentrations of viruses, negative controls).
    • For example, the preliminary LoD determined by probit analysis and the RNase P IC cutoff determined in a development study are examples of how internal parameters were established. These processes rely on highly controlled experiments with well-defined inputs (the "ground truth" being the known composition of the contrived samples).
    Ask a Question

    Ask a specific question about this device

    K Number
    K233453

    Validate with FDA (Live)

    Date Cleared
    2024-07-10

    (264 days)

    Product Code
    Regulation Number
    866.3981
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TaqPath™ COVID-19 Diagnostic PCR Kit is a real-time reverse transcription polymerase chain reaction (RT-PCR) test intended for the qualitative detection of nucleic acid from SARS-CoV-2 in nasopharyngeal and anterior nasal specimens from individuals with signs and symptoms of respiratory tract infection.

    The TaqPath™ COVID-19 Diagnostic PCR Kit is intended for use as an aid in the diagnosis of COVID-19 if used in conjunction with other clinical observations, epidemiological information and laboratory findings. The SARS-CoV-2 RNA is generally detectable in upper respiratory (anterior nasal and nasopharyngeal swabs) specimens during the acute phase of infection. Positive results are indicative of the presence of SARS-CoV-2 RNA. Positive results do not rule out bacterial infection or co-infection with other pathogens. The agent detected may not be the definite cause of disease.

    Negative results do not preclude SARS-CoV-2 infection and should not be used as the sole basis for patient management decisions.

    The TaqPath™ COVID-19 Diagnostic PCR Kit is intended for use by qualified clinical laboratory personnel specifically instructed and trained in the techniques of real-time PCR and in vitro diagnostic procedures.

    Device Description

    The Applied BioSystems™ TaqPath™ COVID-19 Diagnostic PCR Kit (TaqPath™ COVID-19 Diagnostic PCR Kit) includes the assays and controls for a multiplex real-time RT-PCR test for the qualitative detection of RNA from SARS-CoV-2 in nasopharyngeal and anterior nasal specimens from individuals with signs and symptoms of respiratory tract infection.

    Each kit includes the following components:

    • Multiplexed assays that contain three primer/probe sets specific to different SARS-CoV-2 genomic regions and primers/probes for bacteriophage MS2
    • MS2 Phage Control as an internal process control for nucleic acid extraction
    • TaqPath™ COVID-19 Diagnostic PCR Control as a positive RNA control that contains targets specific to the SARS-CoV-2 genomic regions targeted by the assays.

    The workflow begins with nucleic acid extraction from upper respiratory specimens (nasopharyngeal and anterior nasal swabs) that arrive in the testing site in transport media. Nucleic acids are isolated and purified from the specimens using the MagMAX™ Viral/Pathogen II Nucleic Acid Isolation Kit. Nucleic acid isolation is performed via an automated process using the KingFisher™ Flex Purification System. The nucleic acid is reverse transcribed into cDNA and amplified using the TaqPath™ COVID-19 Diagnostic PCR Kit and one of the following real-time PCR instruments:

    • Applied Biosystems™ 7500 Fast Dx Real-Time PCR Instrument
    • Applied Biosystems™ QuantStudio™ 5 Dx Real-Time PCR Instrument

    In the process, the probes anneal to three (3) specific SARS-CoV-2 target sequences located between three (3) unique forward and reverse primers for the following genes:

    • ORF1ab
    • N gene
    • S gene

    During the extension phase of the PCR cycle, the 5' nuclease activity of Taq polymerase degrades the probe, causing the reporter dye to separate from the quencher dye, generating a fluorescent signal. With each PCR cycle, additional reporter dye molecules are cleaved from their respective probes, increasing the fluorescent intensity, which is measured at each cycle by the real-time PCR instrument.

    Following RT-PCR, the data from the instrument's data collection software are imported into COVID-19 Interpretive Software IVD Edition for analysis and interpretation. After data import, the software analyzes the run data, performs quality check (QC) analysis, and calculates the interpretive results for each sample and control. The imported data and interpretive results for each run are saved as a batch in the software. Results can be exported as CSV files and reports can be generated in PDF format.

    Validation of the results is performed automatically by the COVID-19 Interpretive Software based on performance of the positive and negative controls. The following results are automatically generated using the calling rules, plate validity and the CT cutoff values for assay targets:

    ORF1abN geneS geneMS2StatusResultAction
    NEGNEGNEGNEGINVALIDNARETESTRepeat test by re-extracting theoriginal sample and repeating the RT-PCR. If therepeat result remains invalid, considercollecting a new specimen.
    NEGNEGNEGPOSVALIDSARS-CoV-2Not DetectedREPORTReport the results to the healthcareprovider.
    Only one SARS-CoV-2 target= POSPOS orNEGVALIDSARS-CoV-2InconclusiveRETEST/REPORT1. Repeat the test by re-extracting theoriginal sample and repeating theRT-PCR.IMPORTANT! Samples with aresult of SARS-CoV-2 Inconclusiveshall be retested one time.2. After retesting one time, report theresults to the healthcare provider.3. If the repeat result remainsinconclusive, the healthcareprovider should conduct additionalconfirmation testing with a newspecimen, if clinically indicated.
    Two or more SARS-CoV-2 targets= POSPOS orNEGVALIDPositive SARS-CoV-2REPORTReport the results to the healthcareprovider.

    A minimum of one negative control and one positive control must be present for each run. Additional negative control wells shall be run for each extraction that is represented on a realtime RT-PCR plate. All control wells must pass for the real-time RT-PCR plate to be considered valid. Recommended actions of retest or report are also provided depending on the results generated.

    AI/ML Overview

    The provided FDA 510(k) summary for the "Applied BioSystems™ TaqPath™ COVID-19 Diagnostic PCR Kit" details various studies to establish its performance. Here's a breakdown of the acceptance criteria and study proving device performance, as per your request:

    Acceptance Criteria and Reported Device Performance

    The acceptance criteria for each study are implicitly demonstrated by the reported performance meeting the standards required for the 510(k) clearance, primarily aiming for high Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA) compared to a composite comparator method, along with demonstrating robustness in analytical performance.

    Here's a table summarizing key performance characteristics and the reported results:

    Study CategoryPerformance Metric/CriterionAcceptance Criteria (Implicitly Met)Reported Device Performance (TaqPath™ COVID-19 Diagnostic PCR Kit)
    Limit of Detection (LoD)Lowest concentration resulting in ≥ 95% positivity.≥ 95% positivity at claimed LoDConfirmed LoD: 50 GCE/mL (SARS-CoV-2 USA-WA1/2020) for both QuantStudio™ 5 Dx and 7500 Fast Dx (100% positivity at 50 GCE/mL). WHO Standard LoD: 150 IU/mL for 7500 Fast Dx, 50 IU/mL for QuantStudio™ 5 Dx.
    Reproducibility/Precision100% PPA for positive samples (1.5x LoD, 5x LoD) and 100% NPA for negative samples (0x LoD) across sites, operators, and lots.100% PPA/NPA at specified LoD levels; acceptable Ct SD/CV values.100% PPA for 5x LoD and 1.5x LoD samples, 100% NPA for 0x LoD samples across all three sites for both PCR instruments. Low Ct SD and %CV values demonstrating high precision.
    In-Use StabilityStable performance under claimed storage/handling conditions.Maintain performance characteristics.TaqPath™ COVID-19 Diagnostic PCR Assay Multiplex: Stable up to 10 Freeze-thaw cycles. TaqPath™ COVID-19 Diagnostic PCR Control (Stock & Working): Stable up to 48 hours refrigerated (2°C to 8°C).
    Transport SimulationMaintenance of packaging integrity and reagent performance.No adverse impact on performance.Demonstrated that packaging, temperature, and reagent performance were maintained.
    Interfering SubstancesNo false positive/negative interference from common substances.100% PPA and 100% NPA for tested substances (some exceptions where interference started at very high concentrations, e.g., Triamcinolone, Oxymetazoline).Generally 100% PPA (3/3) and 100% NPA (3/3) for most substances. Interference observed for Oxymetazoline (Afrin - No drip, extra moisturizing) and Triamcinolone at >5% (v/v), but not at 5% (v/v) (100% PPA/NPA).
    Cross-Reactivity & Microbial Interference (Wet Lab)No cross-reactivity with specified microorganisms at high concentrations.No false positive signals for other organisms; no significant interference with SARS-CoV-2 detection.No cross-reactivity or microbial interference observed for all 46 tested organisms.
    In silico Cross-ReactivityNo significant homology between assay components and other respiratory pathogens.No matches to two or more individual components.101 sequences aligned with exactly one component (primer or probe) with ≥ 80% homology, but none matched two or more, thus predicted no cross-reactivity.
    Carry-Over Cross-ContaminationLow rate of false positives due to carry-over contamination.Acceptably low contamination rate.0.85% (4/470 false positives).
    Analytical Reactivity (Inclusivity)Detection of various SARS-CoV-2 variants.100% detection (PPA) for tested variants at 3x LoD.100% analytical inclusivity for all 15 tested SARS-CoV-2 strains (including Alpha, Beta, Gamma, Delta, Omicron BA.1.1.529, etc.).
    In silico Reactivity-InclusivityDetection of all known SARS-CoV-2 strains/isolates from databases.>99% reactivity based on 100% homology or Tm > annealing temperature for at least two of three gene targets.Analysis indicates >99% of sequences are reactive; most primer/probe mismatches are unlikely to affect function. Highly inclusive for SARS-CoV-2.
    Specimen Storage StabilityStable performance over specified storage conditions and timeframes.Maintain 100% PPA and 100% NPA.100% PPA and 100% NPA for all timepoints: up to 4 hours ambient, 72 hours refrigerated, 30 days frozen (-30°C to -10°C), 30 days frozen (≤-70°C).
    Fresh vs. Frozen EquivalencyEquivalent performance for fresh and freeze-thawed samples.100% PPA at 1.5x LoD and 5x LoD; 100% NPA at 0x LoD.100% PPA at 1.5x LoD/5x LoD and 100% NPA at 0x LoD for various freeze-thaw cycles.
    Clinical Validation (NP Specimens)High PPA and NPA against composite comparator for NP samples.High PPA and NPA, with 95% CI.QuantStudio™ 5 Dx (NP): PPA 98.9% (95% CI: 93.9%-99.8%), NPA 98.7% (95% CI: 97.7%-99.2%). 7500 Fast Dx (NP): PPA 98.9% (95% CI: 93.9%-99.8%), NPA 98.4% (95% CI: 97.4%-99.1%).
    Clinical Validation (AN Specimens)High PPA and NPA against composite comparator for AN samples.High PPA and NPA, with 95% CI.QuantStudio™ 5 Dx (AN): PPA 98.8% (95% CI: 93.6%-99.8%), NPA 98.0% (95% CI: 97.0%-98.7%). 7500 Fast Dx (AN): PPA 98.8% (95% CI: 93.6%-99.8%), NPA 97.8% (95% CI: 96.7%-98.6%).

    Study Details to Prove Device Meets Acceptance Criteria

    The studies primarily focus on analytical and clinical performance of the PCR kit itself, rather than an AI-driven device with human-in-the-loop interaction. Therefore, aspects related to MRMC studies or expert ground truth for image interpretation are not applicable here.

    1. Sample sizes used for the test set and the data provenance:

      • Limit of Detection (LoD):
        • Preliminary LoD: Serial dilutions (number of replicates not specified beyond "three (3) replicates per concentration level" in the WHO Standard test).
        • Confirmatory LoD: 20 replicates at three concentration levels (3x, 1x, 0.33x LoD) for both SARS-CoV-2 USA-WA1/2020 and WHO Standard.
        • Provenance: Contrived samples using inactivated SARS-CoV-2 virus spiked into pooled negative NP specimen matrix. No geographical provenance for these samples is specified, but likely from US. Retrospective in nature as samples are manipulated.
      • Reproducibility and Within-Laboratory Precision: 270 replicates per sample at each test level (0x LoD, 1.5x LoD, 5x LoD) on each PCR instrument model.
        • Provenance: Contrived samples (inactivated SARS-CoV-2 virus in pooled negative NP specimen matrix). Testing performed at three sites (two external, one internal), indicating multi-center analytical study. No geographical provenance for specimens specified. Retrospective.
      • Interfering Substances: 3 replicates per substance level (positive and negative). Some substances tested with 6 replicates at certain concentrations (e.g., Oxymetazoline, Triamcinolone).
        • Provenance: Contrived samples (inactivated SARS-CoV-2 virus at 3x LoD or negative matrix) spiked with interferents. Negative pooled NP samples were used as matrix.
      • Cross-Reactivity and Microbial Interference: 3 replicates of each of 46 microorganisms.
        • Provenance: Microorganisms spiked into pooled negative NP specimen matrix (for cross-reactivity) or contrived positive NP specimen matrix (for microbial interference).
      • Analytical Reactivity (Inclusivity): 3 replicates for each of 15 SARS-CoV-2 variants.
        • Provenance: Contrived samples (inactivated SARS-CoV-2 virus spiked into negative pooled NP specimen).
      • Carry-Over Cross-Contamination: 470 negative samples tested (47 replicates across 10 extraction runs x 2 instruments).
        • Provenance: Contrived samples with high viral titers and negative NP samples.
      • Specimen Storage Stability: Not explicitly stated, but implies multiple replicates at various time points for both positive and negative contrived samples.
        • Provenance: Contrived positive and negative NP samples in VTM.
      • Fresh vs Frozen Equivalency Study: 10 replicates for 0x and 5x LoD; 40 replicates for 1.5x LoD.
        • Provenance: Contrived NP samples in VTM.
      • Clinical Validation (Primary Test Set):
        • Total Subjects: 1076 subjects enrolled across 11 geographically diverse sites in the U.S. (April 2023 - August 2023).
        • Evaluable Samples: 1055 nasopharyngeal swabs (NP) and 1052 anterior nasal swabs (AN). Exclusions resulted in 1053 evaluable NP specimens and 1049 (7500 Fast Dx) / 1052 (QS5 Dx) evaluable AN specimens.
        • Provenance: Prospectively collected from individuals with signs and symptoms of respiratory tract infection in the U.S. "All-comers fashion". Both fresh (tested within 72 hours) and frozen (stored ≤-70°C, tested within 30 days) Category I and Category II specimens. This is prospective data from the US.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • For this PCR diagnostic kit, the "ground truth" for clinical validation is established by a composite comparator method using multiple highly sensitive SARS-CoV-2 molecular assays (other FDA-cleared or authorized PCR tests). Qualified clinical laboratory personnel presumably performed these comparator tests. There is no mention of human expert consensus/adjudication typically seen in AI imaging studies.
    3. Adjudication method for the test set:

      • Clinical Validation: A composite comparator approach was used. Samples were tested by two comparator assays (Test A and Test B). If there was discordance between the first two assays, a third comparator assay (Test C) was used to resolve the discrepancy. The "two-out-of-three" rule determined the composite comparator result, which served as the ground truth.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No. This is a diagnostic PCR kit, not an AI-based imaging or interpretive device that requires human readers. Therefore, an MRMC study is not applicable. The device provides a direct biological result (presence/absence of viral RNA).
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Yes, in essence. The analytical performance studies (LoD, reproducibility, cross-reactivity, inclusivity, etc.) evaluate the kit's performance independently of human interpretation nuances beyond standard laboratory procedures and instrument operation. The clinical validation also assesses its performance against a composite ground truth, effectively as a "standalone" diagnostic test. The "COVID-19 Interpretive Software IVD Edition" analyzes data and generates results automatically, which is an algorithmic standalone component.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

      • A composite comparator method using other molecular diagnostic (RT-PCR) assays. This is considered the clinical reference standard for nucleic acid detection of SARS-CoV-2.
    7. The sample size for the training set:

      • For a molecular diagnostic kit like this, a traditional "training set" in the machine learning sense isn't explicitly defined as it would be for an AI algorithm. The development and optimization of the primers, probes, and reaction conditions (which is analogous to "training" in developing molecular assays) are based on extensive genomic data of SARS-CoV-2 and related viruses. The document describes validation studies, not a separate training phase. The "control" elements (positive/negative controls, MS2 Phage internal control) are part of the kit's design to ensure proper function and validate each run.
    8. How the ground truth for the training set was established:

      • Not applicable as there is no explicitly defined "training set" in the context of an AI algorithm or a separately identified ground truth for such a set. The "training" for such a device is likely the iterative design and optimization of the molecular components to achieve specificity and sensitivity based on known viral sequences and biological principles.
    Ask a Question

    Ask a specific question about this device

    K Number
    K191030

    Validate with FDA (Live)

    Date Cleared
    2020-02-21

    (309 days)

    Product Code
    Regulation Number
    862.2570
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Applied Biosystems™ 3500 Dx Genetic Analyzer and the Applied Biosystems™ 3500xL Dx Genetic Analyzer are in vitro diagnostic devices intended for detection of fluorescently-labeled human genomic deoxyribonucleic acid (DNA) nucleotides by capillary electrophoresis.

    The Applied Biosystems™ 3500 Dx Genetic Analyzer and the Applied Biosystems™ 3500xL Dx Genetic Analyzer are indicated for sequencing and fragment analysis using FDA- cleared or approved assays.

    Device Description

    The Applied Biosystems™ 3500 Dx Genetic Analyzer and the Applied Biosystems™ 3500xL Dx Genetic Analyzer are fluorescence-based DNA analysis instruments that use capillary electrophoresis technology with 8 and 24 capillaries, respectively.

    The 8-capillary system and the 24-capillary system include the following components:

    • 8-capillary or 24-capillary array and POP™ polymer .
    • . Consumables for system qualification
    • Computer workstation and monitor
    • Integrated software for instrument control, data collection, quality control, . basecalling and sizecalling of samples

    The following consumables (branded with the Applied Biosystems name) are required to operate the Applied Biosystems™ 3500 Dx Genetic Analyzer and the Applied Biosystems™ 3500xL Dx Genetic Analyzer.

    • . 50cm Capillary Array: enable the labeled DNA fragments to migrate from the cathode toward the anode for detection
    • POP-6TM Polymer: used as a separation matrix to separate DNA fragments by size ● during electrophoresis for sequencing
    • POP-7TM Polymer: used as a separation matrix for separating DNA fragments by size during electrophoresis for fragment analysis
    • Hi-Di™ Formamide: sample re-suspension solution used for electrokinetic ● injection and denaturing the DNA
    • Sequencing Standard v1.1: used for spectral calibration of the instrument and instrument performance check
    • Cathode Buffer Container: pre-filled with running buffer which maintains a source . of ions and the correct pH for electrophoresis
    • Anode Buffer Container: pre-filled with running buffer which maintains a source ● of ions and the correct pH for electrophoresis
    • Conditioning Reagent: pre-filled pouch used for priming the polymer pump, ● washing the pump between polymer type changes, and during instrument shutdown
    • DS-30 Matrix Standard Dx: used for spectral calibration ●
    • DS-33 Matrix Standard - Dx: used for spectral calibration
    • DS-33 GeneScan™ Install Kit Dx: used for instrument operational qualification ●
    • GeneScan™ 600 LIZ® Size Standard v2.0 Dx: used as a ladder for sizing DNA ● fragments
    • Other accessories (e.g. sample plate holders, plate retainers, septa) .
    AI/ML Overview

    The provided text describes the Applied Biosystems™ 3500 Dx Genetic Analyzer and Applied Biosystems™ 3500xL Dx Genetic Analyzer, which are in vitro diagnostic devices. It details their intended use, comparison to a predicate device, and supporting performance studies.

    Here's an analysis of the acceptance criteria and study information provided:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document states that for both non-clinical and clinical performance studies, "all pre-established acceptance criteria were met." However, it does not explicitly list the quantitative acceptance criteria themselves.

    Performance MetricAcceptance CriteriaReported Device Performance
    Non-ClinicalNot explicitly stated (e.g., specific thresholds for accuracy, precision, reproducibility)"The pre-established acceptance criteria were met."
    ClinicalNot explicitly stated (e.g., specific thresholds for clinical accuracy, sensitivity, specificity)"All pre-established performance criteria were met."

    2. Sample Size Used for the Test Set and Data Provenance

    • Non-Clinical Study: "a reproducibility study using a representative fragment analysis assay. The reproducibility study was performed across multiple sites, using different instruments, multiple operators, and across several days."
      • Sample Size: Not specified.
      • Data Provenance: Implied to be prospective, collected across multiple sites for the study. Country of origin not specified, but the device is manufactured by Life Technologies Holdings Pte Ltd in Singapore and the submission is to the US FDA.
    • Clinical Study: "Clinical performance studies were conducted across 3 US clinical laboratory sites, with multiple instruments using a representative fragment analysis assay and an appropriate method comparison assay."
      • Sample Size: Not specified.
      • Data Provenance: Prospective, conducted across 3 US clinical laboratory sites.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not mention the use of experts to establish ground truth. The studies appear to be focused on the analytical performance of the instrument using "a representative fragment analysis assay and an appropriate method comparison assay," rather than clinical diagnosis by human experts.

    4. Adjudication Method for the Test Set

    Not applicable, as ground truth was not established by human experts requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    This section is not applicable. The device described is a genetic analyzer (an instrument for detecting fluorescently-labeled DNA nucleotides). It is not an AI-based diagnostic tool designed to assist human readers or clinicians in interpreting images or other data that typically involve MRMC studies.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) Was Done

    The device itself is a standalone instrument (analyzer) with integrated software for control, data collection, quality control, basecalling, and sizecalling. The performance studies evaluate the instrument's (algorithm's) ability to perform these functions. Therefore, the non-clinical and clinical performance data effectively describe the standalone performance of the instrument.

    7. The Type of Ground Truth Used

    The ground truth for the non-clinical and clinical studies appears to be based on the analytical results obtained from "a representative fragment analysis assay and an appropriate method comparison assay." This implies that the ground truth is established by the expected results of these assays or comparison to a reference method, rather than pathology, expert consensus, or outcomes data.

    8. The Sample Size for the Training Set

    The document does not provide information about a separate "training set" in the context of machine learning or AI. This device is an instrument, and its software (Data Collection Software 3 IVD v3.2) controls its operations and processes data. While software development involves testing, the concept of a "training set" as used in machine learning for algorithm development is not addressed here.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no mention of a traditional machine learning "training set" with established ground truth outlined in the document. The device's functionality is based on established principles of capillary electrophoresis and DNA analysis.

    Ask a Question

    Ask a specific question about this device

    K Number
    K170299

    Validate with FDA (Live)

    Date Cleared
    2017-06-22

    (142 days)

    Product Code
    Regulation Number
    862.2265
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The lon PGM™ Dx Instrument System is intended for targeted sequencing of human genomic DNA (gDNA) from peripheral whole-blood samples and DNA and RNA extracted from formalin-fixed, paraffin-embedded (FFPE) samples. The lon PGM™ Dx Instrument System is not intended for whole genome or de novo sequencing.

    Device Description

    The Ion PGM™ Dx System is used for detection of human variant sequences from DNA from whole blood samples or RNA and DNA from FFPE tissue samples. Detectable variants include substitutions, insertions, and deletions.

    The Ion PGMTM Dx System consists of the following:

    • Ion OneTouch™ Dx Instrument
    • Ion OneTouch™ ES Dx Instrument
    • Ion OneTouch™ Rack Kit
    • Ion PGM™ Dx Chip Minifuge
    • Ion PGM™ Dx Sequencer
    • Ion PGMTM Wireless Scanner
    • DynaMag™ Dx Kit—Tube & Plate
    • Ion Torrent™ Server
    • Torrent Suite™ Dx Software

    The Ion PGM™ Dx System is used in conjunction with the following kits:

    • Ion PGM™ Dx Library Kit
    • Ion OneTouch™ Dx Template Kit
    • Ion PGM™ Dx Sequencing Kit
    • Ion 318™ Dx Chip Kit

    The system should be used only by professionals trained in laboratory techniques and procedures and in the use of the system.

    AI/ML Overview

    The provided text describes the acceptance criteria and the studies performed for the Ion PGM™ Dx System. Here's a structured breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are not explicitly stated in a single section as "acceptance criteria," but rather derived from the "Special conditions statement for performance" for both gDNA from whole blood and DNA/RNA from FFPE samples. The "Reported Device Performance" is taken from the "Non-Clinical Performance Data" section.

    Feature / MetricAcceptance Criteria (from "Special Conditions")Reported Device Performance (from "Non-Clinical Performance Data")
    gDNA from Whole Blood (SVA Panel)
    Sequencing output> 0.7 gigabasesNot explicitly reported for gDNA from whole blood in the performance data, but indicated as "validated to deliver" in the special conditions.
    Reads> 4 millionNot explicitly reported for gDNA from whole blood in the performance data, but indicated as "validated to deliver" in the special conditions.
    Read lengthup to 200 base pairsNot explicitly reported for gDNA from whole blood in the performance data, but indicated as "validated to deliver" in the special conditions.
    Mean Raw Read Accuracy99.0% when compared to hg19Not explicitly reported for gDNA from whole blood in the performance data, but indicated as "validated to deliver" in the special conditions.
    SNV Detection Reproducibility100% reproducibility for 440 unique SNV positionsNot explicitly reported in the Non-Clinical Performance Data for gDNA from whole blood. This is stated as a system capability in the "Special conditions statement for performance derived from gDNA from whole blood."
    Indel Detection Reproducibility100% reproducibility for various insertion/deletion lengths (1-4 bp insertions, 1-14 bp deletions)Not explicitly reported in the Non-Clinical Performance Data for gDNA from whole blood. This is stated as a system capability in the "Special conditions statement for performance derived from gDNA from whole blood."
    HPT limitationVariants in homoploymer tracts exceeding 8 bases called as no callsNot applicable; this is a known limitation, not a performance metric to be achieved.
    Min coverage for germline DNA>30XNot explicitly reported in performance data. (This is a recommended use parameter)
    FFPE Samples (Representative Assay)
    Sequencing output> 0.7 gigabasesNot explicitly reported for a representative assay in the performance data, but indicated as "validated to deliver" in the special conditions.
    Reads> 3 millionNot explicitly reported for a representative assay in the performance data, but indicated as "validated to deliver" in the special conditions.
    Read lengthup to 141 base pairsNot explicitly reported for a representative assay in the performance data, but indicated as "validated to deliver" in the special conditions.
    PPA (excluding no calls)Not explicitly stated as a minimum.Variant: 98.5% (195/198)Bin: 97.2% (176/181)Sample: 96.9% (158/163)
    NPA (excluding no calls)Not explicitly stated as a minimum.Variant: 100.0% (118,155/118,159)Bin: 99.8% (942/944)Sample: 98.4% (124/126)
    OPA (excluding no calls)Not explicitly stated as a minimum.Variant: 100.0% (118,350/118,357)Bin: 99.4% (1,118/1,125)Sample: 97.6% (282/289)
    Repeatability (DNA variants, excl. no calls)≥97.5% (95% CI lower limit)≥98.8% (95% CI lower limit of ≥97.5%)
    Repeatability (RNA variants, excl. no calls)Not explicitly stated as a minimum for individual RNA variant locations, but overall ≥87.5% for positive variant location.94.4% for each RNA clinical variant location. (For a specific ROS1 RNA variant for Sample C, it was 87.5% with 95% CI lower bound of 61.7%).
    Call Rate (DNA pos. variants, excl. no calls)Not explicitly stated as a minimum.Mean: 96.60%, Median: 97.10%
    Call Rate (RNA pos. variants, excl. no calls)Not explicitly stated as a minimum.Mean: 94.80%, Median: 95.50%
    Call Rate (WT DNA, neg. calls, excl. no calls)Not explicitly stated as a minimum.Mean: 96.10%, Median: 95.00%
    Call Rate (WT RNA, neg. calls, excl. no calls)Not explicitly stated as a minimum.Mean: 99.30%, Median: 99.30%
    Tissue Input (% meeting conc.)98.3% (59/60) had DNA ≥0.83 ng/uL and RNA >1.43 ng/uL. (This is a study finding, implicitly demonstrating the ability to meet the given concentration requirements for the assay under specific tissue input conditions.)98.3% (59/60) met the concentration requirements (DNA ≥0.83 ng/uL, RNA >1.43 ng/uL).
    DNA/RNA Input (Positive Call Rate)100% positive variant call rate within 5-15 ng input range for a representative assay.100% positive variant call rate within the input range tested (5-15 ng), supporting the 10 ng specified input. For clinical samples, one CD74-ROS1 fusion variant showed 100% positive calls at all input combinations, while the other showed rates as low as 50% at specific high input combinations (attributed to likely operator error).
    DNA/RNA Input (Negative Call Rate)Not explicitly stated as a minimum.>95% for all except 4 sample/input combinations; cases with <95% were due to no calls. For clinical samples, DNA variants showed >95% negative call rates. The second CD74-ROS1 clinical sample showed 100% negative call rates for all test conditions where it was expected to be wild type.
    Interfering SubstancesPositive/Negative/Overall concordance with control was 100% (excluding no calls) for most interferents. For hemoglobin, positive concordance was 100%, negative 99.99%, overall 99.99%. (These are the observed results which met the study's goal of demonstrating non-interference).For most interferents (Paraffin, Xylene, Ethanol, Protease, Wash buffer): 100% positive, negative, and overall concordance with control (excluding no calls). For Hemoglobin: 100% positive concordance, 99.99% negative concordance, 99.99% overall concordance (excluding no calls).
    Cross Contamination RateNot explicitly stated as a specific rate, but the study aims to evaluate cross-contamination.False-positive rate at DNA variant locations: 0% (0/100). False-positive rate at RNA variant locations: 1.25% (1/80), attributed to likely sample cross-contamination.
    Minimal coverage needed for FFPE calling≥347X for SNV, MNV, deletion; ≥41X for fusion.Not explicitly reported in performance data. (This is a recommended use parameter)

    2. Sample size used for the test set and the data provenance

    • Accuracy Study (FFPE):
      • Sample Size: 290 FFPE tumor samples.
      • Data Provenance: The document implies these are clinical samples ("human specimens," "FFPE tumor samples") but does not specify country of origin. It is a retrospective study since variants were compared against "validated reference detection methods."
    • Sample Reproducibility Study (FFPE):
      • Sample Size: 2 WT (Wild Type) samples and 10 variant-positive samples. Each sample tested 8 times at each of 4 sites, for a total of 32 replicates per sample.
      • Data Provenance: Not explicitly stated (e.g., country of origin, retrospective/prospective). These appear to be characterized samples used in a prospective, controlled study.
    • Assay Reproducibility Study (FFPE):
      • Sample Size: 18 DNA samples (6 plasmid/clinical DNA blends, 12 clinical DNA samples) and 9 RNA samples (1 WT, 8 with RNA variants). Each pre-extracted sample run in duplicate with 2 different reagent lots (3 sites) or 3 reagent lots (1 site), resulting in 72 test determinations per DNA sample and 144 per RNA sample, total of at least 1,296 sequencing reactions.
      • Data Provenance: Mixtures of plasmid and clinical DNA/RNA. Not explicitly stated (e.g., country of origin, retrospective/prospective). This is a prospective, controlled study.
    • Tissue Input Study (FFPE):
      • Sample Size: 60 slide-mounted FFPE samples (30 resection with >20% tumor, 15 resection with <20% to ≥10% tumor (macrodissected), 15 CNB samples).
      • Data Provenance: Clinical FFPE samples. Not explicitly stated (e.g., country of origin, retrospective/prospective). Presumed retrospective collection for the purpose of testing input requirements.
    • DNA and RNA Input Study (FFPE):
      • Sample Size: 8 cell-line samples (as FFPE sections) for blending. 540 individual DNA and RNA libraries tested. Additionally, 4 clinical samples (2 DNA variants, 2 CD74-ROS1 fusion variants).
      • Data Provenance: Cell-line samples prepared as FFPE sections, and clinical FFPE samples. Not explicitly stated (e.g., country of origin, retrospective/prospective). This is a prospective, controlled study.
    • Interfering Substances Study (FFPE):
      • Sample Size: 8 FFPE samples (1 WT, 7 mutants) with 6 replicates each.
      • Data Provenance: Not explicitly stated (e.g., country of origin, retrospective/prospective). This is a prospective, controlled study using selected samples.
    • Cross Contamination Study (FFPE):
      • Sample Size: 8 FFPE cell line samples. 100 DNA and 80 RNA data points analyzed.
      • Data Provenance: FFPE cell line samples. Not explicitly stated (e.g., country of origin, retrospective/prospective). This is a prospective, controlled study.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The document does not specify the number or qualifications of experts used to establish ground truth for any of the studies.

    4. Adjudication method for the test set

    The document does not describe an adjudication method for the test set. For the accuracy study, variants were compared against "validated reference detection methods" (NGS assay, ROS1 FISH test), which implies pre-established ground truth, not an adjudication process during the study.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    There is no indication that a multi-reader multi-case (MRMC) comparative effectiveness study was done, nor is there any mention of human readers or AI assistance. This device is a high-throughput genomic sequencer, not an imaging device typically involving human readers for interpretation in the way an MRMC study would apply.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The performance studies described (Accuracy, Reproducibility, etc.) are evaluating the Ion PGM™ Dx System as a whole, which includes the instrument and its associated software (Torrent Suite™ Dx Software). The "Performance Data" sections appear to describe the standalone performance of the entire system (instrument + software) without human intervention in the interpretation of the calls as the primary performance metric. The study results (PPA, NPA, call rates) reflect the algorithm's performance in detecting variants.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Accuracy Study (FFPE): Ground truth was established using "validated reference detection methods," specifically:
      • A validated NGS assay for SNV and deletion hotspot variants.
      • A ROS1 FISH reference test for ROS1 fusions.
    • Reproducibility Studies (Sample, Assay, DNA/RNA Input, Interfering Substances, Cross Contamination): Ground truth was based on:
      • Characterized samples (WT samples, variant-positive samples, plasmid/clinical DNA blends, FFPE cell line samples). This implies a pre-existing understanding of the variants present in these samples, likely established through prior sequencing or genetic analysis, serving as the reference standard.

    8. The sample size for the training set

    The document does not provide a specific sample size for a "training set." The studies described are performance validation studies, not studies related to the development or training of an AI/ML algorithm. The Ion PGM™ Dx System measures hydrogen ions during sequencing; its algorithms for base calling and variant detection are typically deterministic or rule-based, or if they involve machine learning, the training details are not disclosed in this document.

    9. How the ground truth for the training set was established

    As no "training set" is described for an AI/ML context, this question is not directly applicable. If the underlying algorithms for variant calling involved machine learning, the method for establishing ground truth for their training is not provided in this document. The document describes validation against characterized samples and reference methods.

    Ask a Question

    Ask a specific question about this device

    K Number
    K123955

    Validate with FDA (Live)

    Date Cleared
    2013-03-08

    (77 days)

    Product Code
    Regulation Number
    862.2570
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The QuantStudio™ Dx Real-Time PCR Instrument with QuantStudio™ Dx Software is intended to perform fluorescence-based PCR to provide detection of FDA cleared and approved nucleic acid sequences in human-derived specimens. The QuantStudio™ Dx Real-Time PCR Instrument with QuantStudio™ Dx Software is intended for in vitro diagnostic use by trained laboratory technologists in combination with nucleic acid reagent kits/tests manufactured and labeled for diagnostic purposes on this instrument.

    Device Description

    The QuantStudio™ Dx Real-Time PCR Instrument is a bench top Real-Time PCR instrument that uses fluorescent-based polymerase chain reaction (PCR) reagents to provide qualitative or quantitative detection of target nucleic acid sequences (targets) using real-time analysis.
    The QuantStudio™ Dx Real-Time PCR Instrument system includes the following components:

    • QuantStudio™ Dx Real-Time PCR instrument with embedded graphical user . interface (eGUI) Touchscreen
    • Thermal Block, also referred to as the sample block, with associated Heated Cover . and Plate Adaptor
    • Calibration and verification materials for instrument qualification .
    • . Computer workstation with a monitor, keyboard and mouse
    • QuantStudio™ Dx instrument software .
    AI/ML Overview

    Here's an analysis of the provided text to extract information about the acceptance criteria and the study demonstrating the device meets them:

    K123955: QuantStudio™ Dx Real-Time PCR Instrument

    This submission concerns the QuantStudio™ Dx Real-Time PCR Instrument, which is a benchtop real-time PCR instrument intended for in vitro diagnostic use to detect FDA-cleared and approved nucleic acid sequences in human-derived specimens. The performance data presented focuses on its use with the Quidel® Molecular Real-Time PCR Direct C. difficile Tox A/B Assay.


    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implicitly derived from the "Precision/Reproducibility" and "Detection Limit" sections, and then more explicitly in the "Comparison Studies" section through performance metrics like sensitivity and specificity against a reference method (Tissue Culture Cytotoxicity Assay and Enhanced Toxigenic Culture).

    Implicit Acceptance Criteria (Analytical Performance):

    Performance MetricAcceptance Criteria (Implicit)Reported Device Performance (QuantStudio™ Dx with Quidel C. difficile Assay)
    Precision (Within-Lab Repeatability)Consistent detection rates across different LoD levels and operators; low %CV for Ct values. (Generally, 100% detection for 2X LoD and above, and high detection for 0.3X LoD)5X LoD: 100% Detection, Avg Ct: 16.51, STDEV: 0.42, %CV: 2.6% 2X LoD: 100% Detection, Avg Ct: 17.70, STDEV: 0.76, %CV: 4.3% 0.3X LoD: 88% Detection, Avg Ct: 21.13, STDEV: 1.37, %CV: 6.5% Negative: 0% Detection
    Reproducibility (Multi-Site)Consistent detection rates and low %CV for Ct values across multiple sites, operators, and days. (Generally, 100% detection for 2X LoD and above, and high detection for 0.3X LoD)High Negative 0.3x LoD: 38/90 positive across 3 sites (individual site results: 8/30, 15/30, 15/30); AVE Ct %CV range: 1.3-5.0 Low Positive 2x LoD: 90/90 positive across 3 sites; AVE Ct %CV range: 0.8-5.9 Med Positive 5x LoD: 90/90 positive across 3 sites; AVE Ct %CV range: 0.4-4.2 Negative Specimen: 0/90 positive Negative Control: 0/90 positive Positive Control: 90/90 positive; AVE Ct %CV range: 0.1-0.6
    Detection Limit (LoD)Defined as the lowest concentration at which 95% of all replicates tested positive.Final assay LoD: 4.2E-01 CFU/assay (based on ATCC BAA-1870 and ATCC BAA-1872 strains, where 95% positivity was observed)

    Explicit Acceptance Criteria (Clinical Performance against Reference Method):

    Performance MetricAcceptance Criteria (Implicit - based on comparison study to predicate/reference)Reported Device Performance (QuantStudio™ Dx with Quidel C. difficile Assay)
    Sensitivity (vs. Tissue Culture Cytotoxicity Assay)High sensitivity (specific numerical target not provided, but typically >90% for diagnostic assays)93.3% (95% CI: 86.9% - 96.7%)
    Specificity (vs. Tissue Culture Cytotoxicity Assay)High specificity (specific numerical target not provided, but typically >90% for diagnostic assays)93.4% (95% CI: 91.3% - 95.0%)
    Sensitivity (vs. Enhanced Toxigenic Culture)High sensitivity87.3% (95% CI: 81.1% - 91.6%)
    Specificity (vs. Enhanced Toxigenic Culture)High specificity98.7% (95% CI: 97.5% - 99.4%)

    2. Sample Size and Data Provenance for the Test Set

    • Analytical Test Set (Precision/Reproducibility):

      • Precision: A "blinded four-member panel consisting of C. difficile positive and negative sample" tested over 12 days by 2 operators, twice a day, using a single assay lot. The total number of tests for each panel member would be 48 (2 operators * 2 times/day * 12 days). For example, 0.3x LoD shows 88% detection, implying 42 positive results out of 48 total tests.
      • Reproducibility: A "blinded and randomized study panel" tested at three (3) test sites. Each site tested the panel for five (5) days in triplicate on each instrument, by two operators. Each operator ran the panel once a day. This means for each panel member, at each site, there were 30 tests (2 operators * 5 days * 3 replicates). Total tests for each panel member across all three sites would be 90 (30 tests/site * 3 sites).
      • Data Provenance: Not explicitly stated for analytical studies, but given the manufacturer's location and the mention of Quidel, it's likely primarily US-based or an international corporate R&D setting. The study is prospective in nature, designed specifically for this validation.
    • Clinical Test Set (Comparison Studies):

      • Sample Size: 792 samples initially collected. 788 specimens were used for the Tissue Culture Cytotoxicity Assay comparison, and 791 specimens were used for the Enhanced Toxigenic Culture comparison (due to removed invalid/indeterminate results).
      • Data Provenance: Prospective study conducted from August to November 2012. Samples were "collected from patients suspected of having Clostridium difficile-associated disease (CDAD) at four (4) distinct geographical sites across the United States."

    3. Number of Experts and Qualifications for Ground Truth

    • Analytical Studies: Ground truth for analytical studies (precision, reproducibility, LoD) is typically based on known concentrations of the target analyte (like C. difficile strains in CFU/mL) prepared in a laboratory setting. No external "experts" are typically involved in establishing this ground truth beyond the scientific personnel designing and executing the standard preparations and dilutions.

    • Clinical Studies:

      • Tissue Culture Cytotoxicity Assay: This is a laboratory-based reference method for detecting C. difficile toxin. The "experts" involved would be the laboratory personnel performing and interpreting this gold standard assay. Their qualifications are not specified but would typically involve trained medical technologists or microbiologists.
      • Enhanced Toxigenic Culture: This is another laboratory-based reference method. Similar to the cytotoxicity assay, the "experts" would be trained laboratory personnel. Their qualifications are not specified.
      • The document does not mention a panel of experts for consensus reading of raw data or images. The ground truth relies on established laboratory methodologies.

    4. Adjudication Method for the Test Set

    The document does not describe an adjudication method involving multiple human readers for the test set.

    • For the analytical studies, results are based on direct output from the instrument (detection yes/no, Ct values) against predetermined concentrations, so no adjudication is required.
    • For the clinical comparison studies, the "ground truth" (reference standard) is provided by the Tissue Culture Cytotoxicity Assay and Enhanced Toxigenic Culture. Discordant results between the Quidel Molecular Direct C. difficile Assay on the QuantStudio™ Dx and these reference methods were further investigated by testing with an "FDA-cleared molecular device." This serves as a form of secondary adjudication or reconciliation for discordant results but does not involve human readers adjudicating an algorithm's output. For example:
      • For the Cytotoxicity comparison, 45 Quidel Positive/Tissue Culture Negative were re-tested with an FDA-cleared molecular device: 35 positive, 9 negative.
      • 7 Quidel Negative/Tissue Culture Positive were re-tested with an FDA-cleared molecular device: 2 positive, 5 negative.
      • Similar reconciliation was done for the Enhanced Toxigenic Culture comparison.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This study focuses on the standalone performance of the instrument with a specific diagnostic assay, and its concordance with established reference laboratory methods. There is no mention of human readers evaluating cases with and without AI assistance.


    6. Standalone (Algorithm Only) Performance

    Yes, the studies presented demonstrate standalone performance of the QuantStudio™ Dx Real-Time PCR Instrument (in conjunction with the Quidel Molecular Direct C. difficile Assay). The data provided (detection rates, Ct values, sensitivity, specificity) represent the performance of the instrument/assay system without direct human-in-the-loop intervention for result interpretation beyond running the assay and reviewing its output.


    7. Type of Ground Truth Used

    • Analytical Studies (Precision, Reproducibility, LoD): Known concentrations of specific C. difficile strains (e.g., ATCC BAA-1870 and ATCC BAA-1872) diluted in a negative fecal matrix. This is a form of analytical (spiked) ground truth.
    • Clinical Studies (Comparison Studies):
      • Tissue Culture Cytotoxicity Assay: An established laboratory method for detecting C. difficile toxin.
      • Enhanced Toxigenic Culture: An established laboratory method for detecting toxigenic C. difficile.
      • For discordant results, an "FDA-cleared molecular device" was used for reconciliation.
        These are all forms of reference standard (laboratory-based) ground truth.

    8. Sample Size for the Training Set

    The document does not provide information on a training set sample size. The submission describes the performance validation of the instrument when used with a specific diagnostic assay, rather than the development and training of a machine learning algorithm. PCR instruments perform predefined biochemical reactions and detect fluorescence signals based on set parameters, they do not typically undergo "training" in the machine learning sense.


    9. How the Ground Truth for the Training Set was Established

    Since there is no mention of a training set for a machine learning algorithm, there is no information provided on how ground truth was established for a training set. The instrument's operation is based on established PCR principles and specified assay parameters.

    Ask a Question

    Ask a specific question about this device

    K Number
    K103302

    Validate with FDA (Live)

    Device Name
    STEMPRO MSC SFM
    Date Cleared
    2011-02-18

    (101 days)

    Product Code
    Regulation Number
    876.5885
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    StemPro® MSC SFM Medium is a liquid tissue culture medium products intended for human ex vivo tissue and cell culture processing applications.

    Device Description

    StemPro® MSC SFM is a serum-free medium (SFM) specially formulated for the growth and expansion of human mesenchymal stem cells (MSCs). StemPro® MSC SFM enables human MSC growth and increased consistency compared to classical serum-supplemented medium. In addition, human MSCs can be expanded for multiple passages while maintaining their multipotential phenotype (i.e. ability to differentiate into osteogenic, chondrogenic, adipogenic lineages). StemPro® MSC SFM contains two components: StemPro® MSC SFM Basal Medium and StemPro® MSC SFM Supplement.

    AI/ML Overview

    This is a 510(k) premarket notification for a tissue culture medium, not an AI/ML powered device. Therefore, many of the requested categories related to AI/ML device studies are not applicable.

    Here's the relevant information based on the provided text:

    Device Name: StemPro® MSC SFM Medium

    Predicate Device: Knockout™ SR Medium (K100616)

    Intended Use: StemPro® MSC SFM Medium is a liquid tissue culture media product intended for human ex vivo tissue and cell culture processing applications. This device is a chemically defined tissue culture media used to support the growth or maintenance of human tissue or cells in culture.


    Acceptance Criteria and Reported Device Performance

    The device demonstrates substantial equivalence to its predicate by meeting performance standards outlined in a "Class II Special Controls Guidance Document: Tissue Culture Media for Human Ex Vivo Tissue and Cell Culture Processing Applications".

    Acceptance Criteria (Special Control Objective)Reported Device Performance (Life Technologies Corporation's Equivalent Tests)
    Demonstrate lack of potential toxicity of materials in the media to cells or tissue and demonstrate support of tissue and cell growthStemPro® MSC SFM Performance Assay
    Demonstrate lack of endotoxin or pyrogen contaminationLimulus Ameobocyte (LAL) test (25 USP Monograph <85>)
    Validation of Aseptic Processing and Sterility Assurance Level (SAL)Determination of SAL to be ≥ 10-3 and compliance with GMP requirements
    Demonstrate Chemical purityIncoming Raw Material testing using USP, ACS, FCC, GIBCO, or Cell Culture requirements
    Stability/Shelf-Life (pH, non-cytotoxicity, support of cell growth)pH continues to meet specifications; media is not cytotoxic and supports mammalian cell growth over 12 months. Container/closure system protects from microbial contamination.

    Study Details (Not applicable for AI/ML device)

    This is a medical device submission for a tissue culture medium, not an AI/ML enabled device. Therefore, the following AI/ML-specific questions are not applicable:

    • Sample size used for the test set and the data provenance: Not applicable. Performance testing involved laboratory assays rather than clinical data sets.
    • Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for performance studies of tissue culture media relies on established laboratory testing methodologies and specifications.
    • Adjudication method for the test set: Not applicable.
    • If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable.
    • If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
    • The type of ground truth used: For performance tests like cytotoxicity, LAL, and chemical purity, the ground truth is established by the specified assay results conforming to pre-defined scientific and regulatory standards (e.g., USP monographs, internal specifications). For cell growth, it's the observed ex vivo cell proliferation and maintenance of phenotype.
    • The sample size for the training set: Not applicable.
    • How the ground truth for the training set was established: Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K100616

    Validate with FDA (Live)

    Date Cleared
    2010-05-20

    (77 days)

    Product Code
    Regulation Number
    876.5885
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Knockout™ SR Medium and Knockout™ SR Xenofree Medium are liquid tissue culture medium products intended for human ex vivo tissue and cell culture processing applications.

    Device Description

    Knockout™ SR is a serum-free medium with a defined formulation that provides consistent growth conditions for human and mouse embryonic stem cell (ESC) and patient-specific induced pluripotent cell lines (iPSCs).
    Knockout™ SR Xenofree is a serum-free and animal origin-free defined formulation that provides consistent growth conditions for human and mouse embryonic stem cell (ESC) and patient-specific induced pluripotent cell lines (iPSCs). All animal derived components have been replaced with human derived or synthetic components to yield a xenogeneic-free formulation.

    AI/ML Overview

    The provided document describes the acceptance criteria and the study that demonstrates the performance of Knockout™ SR Medium and Knockout™ SR Xenofree Medium, which are tissue culture media. This is not a medical device in the typical sense of a diagnostic or therapeutic AI-powered tool. Therefore, many of the requested categories for AI-based device performance studies (like MRMC studies, number of experts for ground truth, sample size of test/training sets, adjudication methods, and standalone performance metrics of an algorithm) are not applicable to this type of product.

    However, I can extract the relevant performance criteria and how the manufacturer demonstrates compliance.

    1. Table of Acceptance Criteria and Reported Device Performance

    Special Control Objective (Acceptance Criteria)Life Technologies Corporation Knockout™ SR Medium Performance
    Demonstrate lack of potential toxicity of materials in the media to cells or tissue and demonstrate support of tissue and cell growthES Cell Morphology: Demonstrated through specific morphology observed. Relative Plate Efficiency: Demonstrated as comparable to predicate. Relative Type 1 Colonies: Demonstrated as comparable to predicate. Also, noted that ES and iPSC grown in Knockout™ SR supplemented media are "substantially less differentiated than those grown in fetal bovine serum (FBS) supplemented media." For stability testing, it was demonstrated that the media was not cytotoxic and supported the growth of mammalian cells throughout its shelf life.
    Demonstrate lack of endotoxin or pyrogen contaminationLimulus Ameobocyte (LAL) test (25 USP Monograph <85>) was performed and met acceptance criteria.
    Validation of Aseptic Processing and Sterility Assurance Level (SAL)Determination of SAL to be ≥ 10⁻³ compliance with GMP requirements regarding aseptic processing.
    Demonstrates Chemical purityIncoming Raw Material testing using USP, ACS, FCC, GIBCO, or Cell Culture requirements.
    Stability/Shelf LifeEstablished a shelf life of 14-months when stored between 2°C-8°C. This was demonstrated by assessing: 1. pH continued to meet specifications. 2. Media was not cytotoxic and supported the growth of mammalian cells. 3. Container/closure integrity testing confirmed protection from microbial contamination.

    2. Sample Size Used for the Test Set and Data Provenance

    This is not applicable as the studies described are laboratory-based assays and stability tests for a cell culture medium, not clinical studies with patient data. The "test set" would refer to internal lab samples and batches of the media.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. The "ground truth" for chemical purity, sterility, and cell growth support is established through standardized laboratory assays and protocols, not by expert human review in the context of radiology or pathology, for example.

    4. Adjudication Method for the Test Set

    Not applicable. Laboratory test results are typically objective measurements against predefined specifications, not subjective assessments requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    Not applicable. This device is not an AI-powered diagnostic or therapeutic tool for human readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not applicable. This device is a cell culture medium, not an algorithm.

    7. The Type of Ground Truth Used

    The ground truth is based on:

    • Established laboratory assay specifications: For LAL test (<85> USP Monograph), SAL determination (≥ 10⁻³), and incoming raw material testing (USP, ACS, FCC, GIBCO, or Cell Culture requirements).
    • Scientific and biological principles: For ES cell morphology, relative plate efficiency, and relative type 1 colonies, indicating proper cell growth and differentiation status.
    • Physical and chemical specifications: For pH and container integrity over time.

    8. The Sample Size for the Training Set

    Not applicable. This is not a machine learning or AI model that requires a training set. The development of the media involved formulation and optimization, but not in the sense of an algorithm training on data.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set in the context of an AI device. The "ground truth" for the medium's development would have been empirically determined optimal conditions for cell culture based on literature, prior research, and internal experiments to achieve desired cell growth and maintenance characteristics. The company mentions a long history and substantial literature (300+ references) for Knockout™ SR, suggesting a robust scientific foundation for its formulation.