Found 5 results

510(k) Data Aggregation

    K Number
    K243226

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-05-08

    (212 days)

    Product Code
    Regulation Number
    892.1550
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Ultrasound Imaging System is a software-based imaging system and accessories intended for use by qualified physicians and healthcare professionals who has the ability to conduct ultrasound scan process for evaluation by ultrasound imaging system or fluid flow analysis of the human body. The modes of operation include B mode, M mode, PWD mode, Color Doppler (CD) mode, Power Doppler mode, and the combined mode (B+M, B+CD, B+PWD). Specific clinical applications and exam types including:

    LX Series:
    Ophthalmic,Fetal,General abdominal imaging,Pediatric,Small organ (Nerve, prostate, scrotum, breast),Neonatal cephalic,Trans-rectal,Trans-vaginal,Urology,Musculoskeletal (conventional),Musculoskeletal (superficial),OB/Gyn,Cardiac (adult),Cardiac (pediatric),Peripheral vessel ,Other (Carotid),Pulmonary,Nerve

    128 Series:
    Ophthalmic,Fetal,General abdominal imaging,Pediatric,Small organ (Nerve, prostate, scrotum, breast),Neonatal cephalic,Trans-rectal,Trans-vaginal,Urology,Musculoskeletal (conventional),Musculoskeletal (superficial),OB/Gyn,Cardiac (adult),Cardiac (pediatric),Peripheral vessel ,Other (Carotid),Pulmonary,interventional guidance (free hand needle/ catheter),FAST/EFAST,Nerve

    LU700 Series:
    Ophthalmic,Fetal,General abdominal imaging,Pediatric,Small organ (Nerve, prostate, scrotum, breast),Neonatal cephalic,Trans-rectal,Trans-vaginal,Urology,Musculoskeletal (conventional),Musculoskeletal (superficial),OB/Gyn,Cardiac (adult),Cardiac (pediatric),Peripheral vessel ,Other (Carotid),Pulmonary,interventional guidance (free hand needle/ catheter),FAST/EFAST,Nerve

    The device is intended for use in environments where healthcare is provided by trained healthcare professionals, but not intended for use in emergency medical service, ambulance, or aircraft.

    Device Description

    The Ultrasound Imaging System is a portable, software controlled, handheld ultrasound system used to acquire and display hi-resolution, real-time ultrasound data through a commercial off-the-shelf (COTS) mobile device.

    I. The imaging system software runs as an app on a mobile device.

    II. The imaging system software can be download to a commercial off-the-shelf (COTS) mobile device and utilizes an icon touch-based user interface.

    III. The imaging system consists of a series of wireless transducers employing Wi-Fi-based technology to communicate with traditional tablet/smartphone devices via direct Wi-Fi. This allows the user to export ultrasound images and display them across a range portable personal device.

    IV. The imaging system houses a built-in battery, multichannel beamformer, prescan converter and Wi-Fi components

    A complete list of all the probes and indications including:

    Dual headed probe

    LX Series:

    LX128LC
    Linear transducer: Ophthalmic,General abdominal imaging,Pediatric,Small organ (thyroid, prostate, scrotum, breast),Neonatal cephalic,Musculoskeletal (conventional),Musculoskeletal (superficial),Peripheral vessel ,Other (Carotid),Pulmonary,interventional guidance (free hand needle/ catheter),Nerve

    Convex transducer: Fetal,General abdominal imaging,Pediatric,Small organ (thyroid, prostate, scrotum, breast),Urology,Musculoskeletal (conventional),OB/Gyn,Cardiac (adult),Cardiac (pediatric),Peripheral vessel ,interventional guidance (free hand needle/ catheter),FAST/EFAST,Nerve

    LX128LP
    Linear transducer: Ophthalmic,General abdominal imaging,Pediatric,Small organ (thyroid, prostate, scrotum, breast),Neonatal cephalic,Musculoskeletal (conventional),Musculoskeletal (superficial),Peripheral vessel ,Other (Carotid),Pulmonary,interventional guidance (free hand needle/ catheter),Nerve

    Phased Array transducer: ,Fetal,General abdominal imaging,Pediatric,Cardiac (adult),Cardiac (pediatric),Pulmonary,FAST/EFAST,Nerve

    LX192LC
    Linear transducer: Ophthalmic,General abdominal imaging,Pediatric,Small organ (thyroid, prostate, scrotum, breast),Neonatal cephalic,Musculoskeletal (conventional),Musculoskeletal (superf-dotorial),Peripheral vessel ,Other (Carotid),Pulmonary,interventional guidance (free hand needle/ catheter),Nerve

    Convex transducer: ,Fetal,General abdominal imaging,Pediatric,Small organ (thyroid, prostate, scrotum, breast),Urology,Musculoskeletal (conventional),OB/Gyn,Cardiac (adult),Cardiac (pediatric),Peripheral vessel ,interventional guidance (free hand needle/ catheter),FAST/EFAST,Nerve

    Single headed probe

    128 Series:

    LK128C
    Fetal,General abdominal imaging,Pediatric,Small organ (thyroid, prostate, scrotum, breast),Urology,Musculoskeletal (conventional),OB/Gyn,Cardiac (adult),Cardiac (pediatric),Peripheral vessel ,interventional guidance (free hand needle/ catheter),FAST/EFAST,Nerve

    LK128L
    Ophthalmic,General abdominal imaging,Pediatric,Small organ (thyroid, prostate, scrotum, breast),Neonatal cephalic,Musculoskeletal (conventional),Musculoskeletal (superficial),Peripheral vessel ,Other (Carotid),Pulmonary,interventional guidance (free hand needle/ catheter),Nerve

    LK128LH
    Ophthalmic,General abdominal imaging,Pediatric,Small organ (thyroid, prostate, scrotum, breast),Neonatal cephalic,Musculoskeletal (conventional),Musculoskeletal (superficial),Peripheral vessel ,Other (Carotid),Pulmonary,interventional guidance (free hand needle/ catheter),Nerve

    LK128M
    Fetal,General abdominal imaging,Pediatric,Small organ (thyroid, prostate, scrotum, breast),Neonatal cephalic,Urology,Musculoskeletal (conventional),OB/Gyn,Cardiac (adult),Cardiac (pediatric),Peripheral vessel ,interventional guidance (free hand needle/ catheter)

    LK128PA
    Fetal,General abdominal imaging,Pediatric,Cardiac (adult),Cardiac (pediatric),Pulmonary,FAST/EFAST,Nerve

    LK128E
    Fetal,General abdominal imaging,Pediatric,Small organ (thyroid, prostate, scrotum, breast),Trans-rectal,Trans-vaginal,Urology,OB/Gyn,interventional guidance (free hand needle/ catheter)

    LU700 Series:

    LU700L
    General abdominal imaging,Pediatric,Small organ (thyroid, prostate, scrotum, breast),Neonatal cephalic,Musculoskeletal (conventional),Musculoskeletal (superficial),Peripheral vessel ,Other (Carotid),Pulmonary,interventional guidance (free hand needle/ catheter),Nerve

    LU710L
    Ophthalmic,General abdominal imaging,Pediatric,Small organ (thyroid, prostate, scrotum, breast),Neonatal cephalic,Musculoskeletal (conventional),Musculoskeletal (superficial),Peripheral vessel ,Other (Carotid),Pulmonary,interventional guidance (free hand needle/ catheter),Nerve

    LU710LH
    Ophthalmic,General abdominal imaging,Pediatric,Small organ (thyroid, prostate, scrotum, breast),Neonatal cephalic,Musculoskeletal (conventional),Musculoskeletal (superficial),Peripheral vessel ,Other (Carotid),Pulmonary,interventional guidance (free hand needle/ catheter),Nerve

    LU700C
    General abdominal imaging,Musculoskeletal (conventional),Musculoskeletal (superficial),OB/Gyn,Peripheral vessel ,interventional guidance (free hand needle/ catheter),FAST/EFAST,Nerve

    LU710C
    Fetal,General abdominal imaging,Pediatric,Small organ (thyroid, prostate, scrotum, breast),Urology,Musculoskeletal (conventional),OB/Gyn,Cardiac (adult),Cardiac (pediatric),Peripheral vessel ,interventional guidance (free hand needle/ catheter),FAST/EFAST,Nerve

    LU710M
    Fetal,General abdominal imaging,Pediatric,Small organ (thyroid, prostate, scrotum, breast),Neonatal cephalic,Urology,Musculoskeletal (conventional),OB/Gyn,Cardiac (adult),Cardiac (pediatric),Peripheral vessel ,interventional guidance (free hand needle/ catheter)

    LU710PA
    Fetal,General abdominal imaging,Pediatric,Cardiac (adult),Cardiac (pediatric),Pulmonary,FAST/EFAST,Nerve

    LU710E
    Fetal,General abdominal imaging,Pediatric,Small organ (thyroid, prostate, scrotum, breast),Trans-rectal,Trans-vaginal,Urology,OB/Gyn,interventional guidance (free hand needle/ catheter)

    AI/ML Overview

    The provided FDA 510(k) clearance letter and summary for Leltek Inc.'s Ultrasound Imaging Systems (K243226) address the addition of new dual-headed probes (LX series) to existing cleared devices. While it asserts that the new models meet safety and performance standards, the document does not provide specific acceptance criteria or detailed study results for device performance in the format requested.

    The document primarily focuses on demonstrating substantial equivalence to a predicate device (K241161) and compliance with various recognized standards (e.g., IEC 60601 series, ISO 10993 series). It states that "No additional clinical testing is required" due to the nature of the changes being additions of transducers with established technologies.

    Therefore, the requested information cannot be fully extracted directly from this document. However, I can reconstruct what parts are mentioned or implied.

    Here's a breakdown of what can be inferred or directly stated from the provided text, and what is missing:


    Request for Acceptance Criteria and Study Details

    The provided document (FDA 510(k) Clearance Letter - Leltek Inc. Ultrasound Imaging Systems, K243226) details a submission for adding new dual-headed probes (LX series) to an already cleared Ultrasound Imaging System. It asserts that the new models meet safety and performance standards equivalent to the predicate device, thereby requiring no additional clinical testing.

    The document does not provide specific acceptance criteria or detailed study results for device performance in the format of a comparative table or a comprehensive study description as requested for a new AI/software-based medical device. Instead, it relies on the principle of substantial equivalence and compliance with recognized standards for hardware modifications.

    However, it does provide information on the safety and performance evaluations conducted to support the substantial equivalence.


    Requested Information Breakdown:

    1. Table of Acceptance Criteria and Reported Device Performance

    Not explicitly provided as clear, quantifiable acceptance criteria or reported device performance metrics in the given document.
    The document states that the devices comply with various standards, which inherently means they meet the performance requirements within those standards. It mentions:

    • "full compliance with acoustic output, biocompatibility, cleaning, and disinfection effectiveness"
    • "maximum acoustic output level is under the FDA recommended limit"
    • "All the safety and performance tests of the device meet the essential requirements."

    Without specific numerical targets for metrics like accuracy, sensitivity, or specificity (which are common for AI/software devices, but not the primary focus here for a hardware add-on to an imaging system), a table cannot be constructed. The performance is assessed against the established safety and performance of the predicate device and relevant industry standards.

    2. Sample size used for the test set and the data provenance

    Not applicable in the context of this 510(k) submission for new transducers.
    The submission focuses on the safety and performance of the hardware (new transducers) and their integration with existing software. It explicitly states, "No additional clinical testing is required." Therefore, there is no mention of a clinical "test set" with patient data or its provenance. The testing would be on the physical device's characteristics (e.g., acoustic output, EMC, biocompatibility).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. As no clinical "test set" or diagnostic performance study is described, there's no mention of experts establishing ground truth for such a test.

    4. Adjudication method for the test set

    Not applicable. As no clinical "test set" or diagnostic performance study is described, there's no mention of an adjudication method.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No. This is not an AI/Software as a Medical Device (SaMD) submission focused on aiding diagnosis, but rather a submission for a new hardware component (ultrasound transducers) to an imaging system. Therefore, no MRMC study or AI assistance effect size is mentioned.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    No. This is not an AI/SaMD submission.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    Not applicable. For this type of submission, the "ground truth" for demonstrating safety and performance relates to engineering specifications, physical measurements (e.g., acoustic output, electrical safety), and material properties (e.g., biocompatibility testing results), rather than clinical diagnostic accuracy against pathology or expert consensus.

    8. The sample size for the training set

    Not applicable. This submission is for hardware. There is no mention of a "training set" in the context of an algorithm or AI.

    9. How the ground truth for the training set was established

    Not applicable. As there is no training set mentioned, this information is not available.


    Summary of Device Safety and Performance Assessment (from the Document):

    The document indicates that the substantial equivalence is established based on the following:

    • Compliance with recognized standards: A comprehensive list of IEC and ISO standards for medical electrical equipment, usability, software lifecycle, biological evaluation (biocompatibility), risk management, and quality management is provided (Pages 14-15).
    • Acoustic Output: The device is tested and determined to be in "full compliance with acoustic output" and the "maximum acoustic output level is under the FDA recommended limit" (Page 16).
    • Biocompatibility: The new models are deemed to have "biosafety equivalence" and passed "biocompatibility" tests (Page 16).
    • Cleaning and Disinfection: Demonstrated "cleaning, and disinfection effectiveness" (Page 16).
    • Software and Hardware Equivalence: The new LX Series probes "utilizes the same software, cover material, and transducer specifications as the 128 Series but replaces a similar PCBA with a dual-headed probe, maintaining the same block diagram." (Page 16)
    • No Pragmatic Detriments: The changes "have no pragmatic detriments" (Page 16).

    In essence, Leltek Inc. demonstrated that the new LX series transducers behave similarly to their predicate devices and comply with relevant safety standards, thus no further clinical performance studies were deemed necessary for this specific 510(k) submission.

    Ask a Question

    Ask a specific question about this device

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Ultrasound Imaging System (Model: 128 Series) and (Model: LU700 Series) is a software-based imaging system and accessories intended for use by qualified physicians and healthcare professionals who has the ability to conduct ultrasound scan process for evaluation by ultrasound imaging system or fluid flow analysis of the human body. The modes of operation include B mode, M mode, PWD mode, Color Doppler (CD) mode, Power Doppler mode, and the combined mode (B+M, B+CD, B+PWD) Specific clinical applications and exam types including: LK128C Fetal, General abdominal imaging, Pediatric, Small organ (thyroid, prostate, scrotum, breast), Urology, Musculoskeletal (conventional), OB/Gyn, Cardiac (adult), Cardiac (pediatric).Peripheral vessel .interventional quidance (free hand needle/ catheter).FAST/ EFAST,Nerve LK128L Ophthalmic General abdominal imaging Pediatric, Small organ (thyroid, prostate, scrotum, breast).Neonatal cephalic.Musculoskeletal (conventional).Musculoskeletal (superficial).Peripheral vessel , Other (Carotid), Pulmonary, interventional guidance (free hand needle/ catheter), Nerve LK128LH Ophthalmic General abdominal imaging Pediatric, Small organ (thyroid, prostate, scrotum, breast),Neonatal cephalic,Musculoskeletal (conventional),Musculoskeletal (superficial),Peripheral vessel , Other (Carotid), Pulmonary, interventional guidance (free hand needle/ catheter), Nerve LK128M Fetal General abdominal imaging Pediatric, Small organ (thyroid, prostate, scrotum, breast), Neonatal cephalic,Urology,Musculoskeletal (conventional),OB/Gyn,Cardiac (adult),Cardiac (pediatric),Peripheral vessel interventional guidance (free hand needle/ catheter) LK128PA Fetal, General abdominal imaging, Pediatric, Cardiac (adult), Cardiac (pediatric), Pulmonary, FAST/ EFAST,Nerve LK128E Fetal, General abdominal imaging, Pediatric, Small organ (thyroid, prostate, scrotum, breast), Trans-rectal, Trans-vaginal, Urology, OB/Gyn, interventional guidance (free hand needle/ catheter) LU700L General abdominal imaging Pediatric, Small organ (thyroid, prostate, scrotum, breast).Neonatal cephalic,Musculoskeletal (conventional),Musculoskeletal (superficial).Peripheral vessel ,Other (Carotid),Pulmonary,interventional guidance (free hand needle/ catheter),Nerve LU710L Ophthalmic, General abdominal imaging, Pediatric, Small organ (thyroid, prostate, scrotum, breast),Neonatal cephalic,Musculoskeletal (conventional),Musculoskeletal (superficial),Peripheral vessel , Other (Carotid), Pulmonary, interventional guidance (free hand needle/ catheter), Nerve LU710LH Ophthalmic, General abdominal imaging, Pediatric, Small organ (thyroid, prostate, scrotum, breast),Neonatal cephalic,Musculoskeletal (conventional),Musculoskeletal (superficial),Peripheral vessel . Other (Carotid),Pulmonary.interventional quidance (free hand needle/ catheter),Nerve LU700C General abdominal imaging, Musculoskeletal (conventional), Musculoskeletal (superficial), OB/ Gyn, Peripheral vessel .interventional guidance (free hand needlel catheter),FAST/EFAST,Nerve LU710C Fetal, General abdominal imaging, Pediatric, Small organ (thyroid, prostate, scrotum, breast).Urology.Musculoskeletal (conventional).OB/Gyn.Cardiac (adult).Cardiac (pediatric),Peripheral vessel ,interventional guidance (free hand needle/ catheter),FAST/ EFAST,Nerve LU710M Fetal General abdominal imaging, Pediatric, Small organ (thyroid, prostate, scrotum, breast), Neonatal cephalic.Urology.Musculoskeletal (conventional).OB/Gvn.Cardiac (adult).Cardiac (pediatric),Peripheral vessel ,interventional guidance (free hand needle/ catheter) LU710PA Fetal,General abdominal imaging,Pediatric,Cardiac (adult),Cardiac (pediatric),Pulmonary,FAST/ EFAST.Nerve LU710E Fetal, General abdominal imaging, Pediatric, Small organ (thyroid, prostate, scrotum, breast), Trans-rectal, Trans-vaginal, Urology, OB/Gyn, interventional guidance (free hand needle/ catheter)

    Device Description

    The Ultrasound Imaging System is a portable, software controlled, handheld ultrasound system used to acquire and display hi-resolution, real-time ultrasound data through a commercial off-the-shelf (COTS) mobile device..

    • I. The imaging system software runs as an app on a mobile device.
    • II. The imaging system software can be download to a commercial off-the-the-shelf (COTS) mobile device and utilizes an icon touch-based user interface.
    • III. The imaging system consists of a series of wireless transducers employing Wi-Fi-based technology to communicate with traditional tablet/smartphone devices via direct Wi-Fi. This allows the user to export ultrasound images and display them across a range portable personal device.
    • IV. The imaging system houses a built-in battery, multichannel beamformer, prescan converter and Wi-Fi components
    AI/ML Overview

    This document is a 510(k) Premarket Notification from Leltek Inc. to the FDA for their Ultrasound Imaging System (LK128L; LK128LH; LK128C; LK128M; LK128PA; LK128E; LU700C; LU700L; LU710L; LU710LH; LU710C; LU710M; LU710PA; LU710E).

    The information provided does not contain details about acceptance criteria or a study proving the device meets those criteria, specifically in the context of AI/algorithm performance. The document focuses on demonstrating substantial equivalence to a predicate device (Leltek Ultrasound Imaging System, K222365) and a reference device (Aco Apache Ultrasound System, K231509) for an ultrasound imaging system, which is a hardware and software system for acquiring and displaying ultrasound data, not an AI or algorithm for diagnostic interpretation that would typically require the detailed performance metrics you're asking for.

    The submission is primarily focused on:

    • Adding new specific clinical applications (FAST/EFAST, Nerve)
    • Adding new transducers for the LK128 series.

    The "Performance standards" section (Section 8) lists compliance with various electrical and medical device safety and performance standards (e.g., IEC 60601-1, ISO 10993). It mentions "acoustic output" and "cleaning, and disinfection effectiveness" tests, but these are general safety and performance aspects of an ultrasound machine, not a study to evaluate an AI's diagnostic accuracy.

    Therefore, I cannot provide the requested information regarding acceptance criteria and an AI performance study from this document because it is not relevant to the scope of this 510(k) submission. The device described is a diagnostic ultrasound imaging system, not an AI/algorithm-based diagnostic aid that would involve the kind of performance study you're asking for (e.g., sensitivity, specificity, MRMC studies, ground truth establishment methodologies, and sample sizes for AI training/testing).

    Ask a Question

    Ask a specific question about this device

    K Number
    K222365

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-10-25

    (82 days)

    Product Code
    Regulation Number
    892.1550
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Leltek Ultrasound Imaging System (Model: LU700 Series) is a software-based imaging system and accessories intended for use by qualified physicians and healthcare professionals who has the ability to conduct ultrasound scan process for evaluation by ultrasound imaging system or fluid flow analysis of the human body. The modes of operation include B mode, PWD mode, Color Doppler (CD) mode, Power Doppler mode, and the combined mode (B+M, B+CD, B+PWD). Specific clinical applications and exam types including:

    LU700L
    General abdominal imaging, Pediatric, Small organ (thyroid, prostate, scrotum, breast), Neonatal cephalic, Musculoskeletal (conventional),Musculosketal (superficial),Peripheral vessel ,Other (Carotid),Pulmonary,interventional guidance (free hand needle/ catheter)

    LU710L
    Ophthalmic, General abdominal imaging, Pediatric, Small organ (thyroid, prostate, scrotum, breast), Neonatal cephalic.Musculoskeletal (conventional),Musculoskeletal (superficial),Peripheral vessel ,Other
    (Carotid),Pulmonary,interventional guidance (free hand needle/ catheter)

    LU710LH
    Ophthalmic, General abdominal imaging, Pediatric, Small organ (thyroid, prostate, scrotum, breast), Neonatal cephalic,Musculoskeletal (conventional),Musculoskeletal (superficial),Peripheral vessel ,Other
    (Carotid),Pulmonary,interventional guidance (free hand needle/ catheter)

    LU700C
    General abdominal imaging,Musculoskeletal (conventional),Musculoskeletal (superficial),OB/Gyn,Peripheral vessel ,interventional guidance (free hand needle/ catheter)

    LU710C
    Fetal.General abdominal imaging.Small organ (thyroid, prostate, scrotum, breast).Urology.Musculoskeletal (conventional),OB/Gyn,Cardiac (pediatric),Peripheral vessel ,interventional guidance (free hand needle/ catheter)

    LU710M
    Fetal,General abdominal imaging,Pediatric,Small organ (thyroid, prostate, scrotum, breast),Neonatal cephalic. Urology. Musculoskeletal (conventional), OB/Gyn. Cardiac (pediatric), Peripheral vessel ,interventional guidance (free hand needle/ catheter)

    LU710PA
    Fetal, General abdominal imaging, Pediatric, Cardiac (adult), Cardiac (pediatric), Pulmonary

    LU710E
    Fetal, General abdominal imaging, Pediatric, Small organ (thyroid, prostate, scrotum, breast), Transvaginal,Urology,OB/Gyn,interventional guidance (free hand needle/ catheter)

    The clinical environments where the system can be used include physician offices, clinics, and clinical point-of-care for diagnosis of patients.

    Device Description

    The Leltek Ultrasound Imaging System (Model: LU700 Series) is a portable, software controlled, handheld ultrasound system used to acquire and display hi-resolution, real-time ultrasound data through a commercial off-the-shelf (COTS) mobile device.

    • I. The imaging system software runs as an app on a mobile device.
    • II. The imaging system software can be download to a commercial off-the-shelf (COTS) mobile device and utilizes an icon touch-based user interface.
    • III. The imaging system consists of a series of wireless transducers employing Wi-Fi-based technology to communicate with traditional tablet/smartphone devices via direct Wi-Fi. This allows the user to export ultrasound images and display them across a range portable personal device.
    • IV. The imaging system houses a built-in battery, multichannel beamformer, prescan converter and Wi-Fi components
    AI/ML Overview

    The provided text describes the Leltek Ultrasound Imaging System (Model: LU700 Series) seeking 510(k) clearance from the FDA. Based on the document, here's a breakdown of the acceptance criteria and study information:

    1. A table of acceptance criteria and the reported device performance

    The document does not provide a specific table of quantitative acceptance criteria and reported device performance in terms of diagnostic accuracy metrics (e.g., sensitivity, specificity, accuracy). Instead, the acceptance criteria are generally based on compliance with international safety and performance standards for medical electrical equipment and the device demonstrating "substantially equivalent" safety and effectiveness to predicate devices.

    The reported device performance is primarily described as:

    Acceptance Criteria CategoryReported Device Performance
    Safety and Performance Standards ComplianceThe device has been designed, manufactured, tested, and certified to comply with the following internationally recognized standards:
    - AAMI/ANSI ES60601-1 (2009 & 2012): Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
    - IEC 60601-1-2 (2014): Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
    - IEC 60601-1-6 (2013): Medical electrical equipment Part 1-6 General requirements for basic safety and essential performance Collateral standard Usability
    - IEC 60601-2-37/AMD1 (2008 & 2015): Medical electrical equipment - Part 2-37: Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment
    - IEC 62133 (2012): Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications
    - IEC 62304 (2014): Medical device software - Software life-cycle processes
    - IEC 62366-1 (2015): Medical devices -- Part 1: Application of usability engineering to medical devices
    - ISO 10993-1 (2009 & 2018): Biological evaluation of medical devices Part 1: Evaluation and testing within a risk management process
    - ISO 10993-5 (2009): Biological evaluation of medical devices Part 5: Tests for in vitro cytotoxicity
    - ISO 10993-10 (2010): Biological evaluation of medical devices Part 10: Tests for irritation and skin sensitization
    - ISO 13485 (2016): Medical devices - Quality management systems - Requirements for regulatory purposes
    - ISO 14971 (2019): Medical devices - Application of risk management to medical devices
    - ISO 15223-1 (2016): Medical devices -- Symbols to be used with medical device labels, labelling and information to be supplied -- Part 1: General requirements
    Device Performance Specifics (as mentioned for new models/changes)- Acoustic output: The new models (LU710L, LU710LH) were tested and determined to be in full compliance. The maximum acoustic output level is under the FDA recommended limit, and the power level is displayed all the time.
    - Biocompatibility: The new models were tested and determined to be in full compliance.
    - Cleaning and disinfection effectiveness: The new models were tested and determined to be in full compliance.
    - Wireless, thermal, electrical, electromagnetic and mechanical safety: The system has been evaluated and found to conform with applicable medical device safety standards.
    Substantial EquivalenceThe device is deemed "substantially equivalent" to predicate devices, meaning it has "no pragmatic detriments" compared to the cleared device (K210432) and other reference devices (K202035). All safety and performance tests meet essential requirements.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    The document explicitly states: "The Leltek Ultrasound Imaging System did not require clinical testing to establish substantial equivalence." This indicates that no test set involving patient data was used for a clinical performance study. The evaluation was based on compliance with performance standards and comparison with predicate devices.

    Therefore, information regarding sample size, country of origin, and retrospective/prospective nature of a test set is not applicable as no clinical test set was used.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable, as no clinical test set was used and therefore no ground truth needed to be established by experts for a performance evaluation study.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable, as no clinical test set was used.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. The document states "The Leltek Ultrasound Imaging System did not require clinical testing to establish substantial equivalence." This implies no clinical performance study, including an MRMC study with AI assistance, was conducted or required for this 510(k) submission.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The Leltek Ultrasound Imaging System is an ultrasound imaging system, not an AI algorithm. It provides imaging data for qualified physicians and healthcare professionals to conduct ultrasound scans and fluid flow analysis. Therefore, a "standalone algorithm only" performance study is not applicable to this device. Its performance is evaluated on its ability to produce safe and effective ultrasound images, not on an AI's diagnostic accuracy.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    Not applicable in the context of a clinical performance study for an AI algorithm. The device itself is an imaging tool. Its "ground truth" for regulatory purposes revolves around its technical specifications, safety, and ability to generate images comparable to legally marketed predicate devices, as demonstrated through engineering tests and adherence to standards.

    8. The sample size for the training set

    Not applicable. The Leltek Ultrasound Imaging System is an imaging device, not an AI algorithm that requires a training set of data.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K210432

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-11-03

    (264 days)

    Product Code
    Regulation Number
    892.1550
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Leltek Ultrasound Imaging System (Model: LU700 Series) is a software-based imaging system and accessories intended for use by qualified physicians and healthcare professionals who has the ability to conduct ultrasound scan process for evaluation by ultrasound imaging system or fluid flow analysis of the human body.

    The clinical environments where the system can be used include physician offices, clinics, hospitals, and clinical point-of-care for diagnosis of patients

    The modes of operation include B mode, PWD mode, Color Doppler (CD) mode, Power Doppler mode, and the combined mode (B+M, B+CD, B+PWD). Specific clinical applications and exam types including: LU700C General abdominal imaging, musculoskeletal (conventional), musculoskeletal (superficial), peripheral vessel and OB/Gyn. LU700L General abdominal imagine, small organ (breast, thyroid), musculoskeletal (superficial) and peripheral vessel. LU710C Fetal, abdominal, pediatric, small organ (thyroid, prostate, scrotum, breast), musculoskeletal (conventional), urology, gynecology, cardiac adult, cardiac pediatric and peripheral vessel. LU710M Fetal, abdominal, pediatric, small organ (thyroid, prostate, scrotum, breast), musculoskeletal (conventional), urology, gynecology, cardiac adult, cardiac pediatric and peripheral vessel. LU710PA Fetal, abdominal, pediatric, cardiac adult, cardiac pediatric. LU710E Fetal, abdominal, small organ (thyroid, prostate, scrotum, breast), trans-vaginal, urology, gynecology.

    Device Description

    The Leltek Ultrasound Imaging System (Model: LU700 Series) is a portable, software controlled, handheld ultrasound system used to acquire and display hi-resolution, real-time ultrasound data through a commercial off-the-shelf (COTS) mobile device.

    • l. The imaging system software runs as an app on a mobile device.
    • II. The imaging system software can be download to a commercial off-the-shelf (COTS) mobile device and utilizes an icon touch-based user interface.
    • III. The imaging system consists of a series of wireless transducers employing Wi-Fi-based technology to communicate with traditional tablet/smartphone devices via direct Wi-Fi. This allows the user to export ultrasound images and display them across a range portable personal device.
    • IV. The imaging system houses a built-in battery, multichannel beamformer, prescan converter and Wi-Fi components

    The device is intended for use in environments where healthcare is provided by trained healthcare professionals, but not intended for use in emergency medical service, ambulance, or aircraft.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study proving device performance:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document (K210432) is a 510(k) summary for premarket notification of a medical device (Leltek Ultrasound Imaging System). It focuses on establishing substantial equivalence to predicate devices rather than defining and meeting specific analytical or clinical performance acceptance criteria in the traditional sense of a clinical trial.

    Therefore, the "acceptance criteria" here are compliance with recognized safety and performance standards and demonstrating similar performance characteristics to the predicate devices. The "reported device performance" is the system's demonstrated compliance with these standards and its functional equivalence.

    Acceptance Criteria (Implied by Substantial Equivalence and Standards Compliance)Reported Device Performance
    Similar Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body.Same as predicate and reference devices. The LU700 Series is a software-based imaging system and accessories intended for use by qualified physicians and healthcare professionals for evaluation by ultrasound imaging system or fluid flow analysis of the human body. (K210432, page 6, 9)
    Functional Equivalence in Operating Modes: B Mode, M mode, PWD, Color Doppler, Power Doppler, and combined modes.Same as predicate and reference devices. (K210432, page 6)
    Wireless Communication: Communication via Wi-Fi.Same as predicate. The device communicates wirelessly via Wi-Fi. (K210432, page 6)
    Portability: Portable ultrasound system.Same as predicate and reference devices. The Leltek Ultrasound Imaging System (Model: LU700 Series) is a portable, software-controlled, handheld ultrasound system. (K210432, page 5, 6)
    Power Source: Rechargeable battery (Li-ion).Same as predicate and reference devices. (K210432, page 6)
    Display: iOS or Android mobile device.Same as predicate and reference devices. The imaging system software runs as an app on a mobile device and utilizes an icon touch-based user interface, displaying images across a range of portable personal devices. (K210432, page 5, 6)
    Compliance with Safety Standards: AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-2-37, IEC 62133, IEC 62304, IEC 62366-1, ISO 10993 (1, 5, 10), ISO 13485, ISO 14971, ISO 15223-1.The device has been designed, manufactured, tested, and certified to comply with all listed international standards. Acoustic output, biocompatibility, cleaning and disinfection effectiveness, wireless, thermal, electromagnetic, and mechanical safety have been evaluated and found to conform. The maximum acoustic output level is under the FDA recommended limit. (K210432, page 9, 10)
    Biosafety Equivalence: (Implied biological compatibility as per ISO 10993 series)Biosafety equivalence is asserted, and the device conforms to applicable electromedical device safety standards. Biocompatibility testing has been conducted for the new transducers. (K210432, page 8, 10)
    No Pragmatic Detriments: Differences from predicate devices (e.g., expanded Indications for Use, additional transducer types) do not negatively impact safety or effectiveness.The differences specified have no pragmatic detriments. All safety and performance tests of the essential requirements are met. The new models (LU710C, LU710PA, LU710E) were tested and determined to be in full compliance with acoustic output, biocompatibility, cleaning, and disinfection effectiveness, and have no pragmatic detriments. (K210432, page 8, 10)

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not provide details on specific sample sizes for a "test set" in the context of clinical performance or image-based evaluation.

    Instead, the submission relies on:

    • Compliance with recognized performance and safety standards: This implicitly means testing against defined criteria within those standards.
    • Bench testing and engineering verification: "Acoustic output, biocompatibility, cleaning, and disinfection effectiveness as well as wireless, thermal, electromagnetic and mechanical safety" were evaluated by tests, but specific sample sizes for these tests are not mentioned.
    • Substantial equivalence argument: The core of the 510(k) relies on demonstrating that the new device performs as safely and effectively as a legally marketed predicate device (K191235) and a reference device (K192107).

    The data provenance is related to the testing performed to ensure compliance with the listed standards. This would typically be internal testing by the manufacturer (LELTEK, Inc. in Taiwan) as part of their design, manufacturing, and quality control processes. The document does not indicate the use of external or clinical data for the performance evaluation in the way a clinical trial would. It states, "The Leltek Ultrasound Imaging System did not require clinical testing to establish substantial equivalence." (K210432, page 10).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    Since the document explicitly states that clinical testing was not required, there was no "test set" in the clinical sense that would require expert-established ground truth for diagnostic accuracy. The evaluation focused on engineering, safety, and performance characteristics against established standards.

    4. Adjudication Method for the Test Set

    Not applicable, as a clinical "test set" requiring adjudication for diagnostic accuracy as described in the prompt was not part of this 510(k) submission.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. The document states, "The Leltek Ultrasound Imaging System did not require clinical testing to establish substantial equivalence." (K210432, page 10). The submission is based on demonstrating technical and safety equivalence to predicate devices, not on a clinical comparison of diagnostic efficacy with human readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This question is not directly applicable in the context of this device. The Leltek Ultrasound Imaging System is an ultrasound imaging device, not an AI algorithm intended for automated diagnosis or interpretation. The "software" aspect refers to the app running on a mobile device and controlling the ultrasound hardware, not a standalone diagnostic algorithm. Therefore, "standalone" performance for an algorithm is not a relevant concept for this type of device based on the provided information.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    Given that clinical testing for diagnostic performance was not required, there was no "ground truth" of the type typically used for evaluating diagnostic accuracy (e.g., pathology, surgical findings, long-term outcomes, expert consensus on images). The ground truth for this submission was based on:

    • Engineering specifications and measurements: To ensure parameters like acoustic output, image quality (implied by functional equivalence), and electrical safety met defined standards.
    • Biocompatibility testing results: Against ISO 10993 standards.
    • Compliance with cleaning/disinfection protocols.

    8. The Sample Size for the Training Set

    Not applicable. This device is an ultrasound imaging system, and the submission does not describe it as an AI/ML-based diagnostic algorithm that would require a "training set" of data in the sense used for machine learning. The "software" is for image acquisition, processing, and display, not for learning to diagnose from a dataset.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no mention of a "training set" for an AI/ML algorithm in this 510(k) submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K191235

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-02-12

    (280 days)

    Product Code
    Regulation Number
    892.1550
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Leltek Ultrasound Imaging System (Model: LU700) is a software-based imaging system and accessories intended for use by qualified physicians and healthcare professionals who has the ability to conduct ultrasound scan process for evaluation by ultrasound imaging system or fluid flow analysis of the human body. The device is intended for use in environments where healthcare is provided by trained healthcare professionals, but not intended for use in emergency medical service, ambulance, or aircraft.

    Specific clinical applications and exam types including:

    For LU700C:

    General abdominal imaging, musculoskeletal (conventional), musculoskeletal (superficial), peripheral vessel and OB/ Gyn. The modes of operation include B mode. M mode. PWD mode, Color Doppler (CD) mode. Power Doppler mode. and the combined mode(B+M, B+CD, B+PWD).

    For LU700L:

    General abdominal imaging, small organ (breast, thyroid), musculoskeletal (conventional), musculoskeletal (superficial) and peripheral vessel. The modes of operation includes B mode, M mode, Color Doppler (CD) mode, Power Doppler mode, and the combined mode(B+M, B+CD, B+PWD).

    Device Description

    The Leltek Ultrasound Imaging System is a portable, software controlled, handheld ultrasound system used to acquire and display hi-resolution, real-time ultrasound data through a commercial off-the-shelf (COTS) Android mobile device.

    l. The imaging system software runs as an app on a mobile device.

    II. The imaging system software can be download to a commercial off-the-shelf (COTS) Android mobile device and utilizes an icon touch-based user interface.

    lll. The imaging system consists of a series of wireless transducers employing Wi-Fi-based technology to communicate with traditional tablet/smartphone devices via direct Wi-Fi. This allows the user to export ultrasound images and display them across a range portable personal device.

    IV. The imaging system houses a built-in battery, multichannel beamformer, prescan converter and Wi-Fi components

    AI/ML Overview

    This document describes the regulatory clearance (K191235) for the Leltek Ultrasound Imaging System (Model: LU700) by the US FDA. The information provided is a summary of the device, its intended use, and a comparison to predicate devices, focusing on demonstrating substantial equivalence rather than explicit acceptance criteria and a detailed study report. As such, the document does not contain a typical detailed clinical study report with acceptance criteria, sample sizes for test and training sets, expert qualifications, or adjudication methods for ground truth, nor MRMC study results.

    However, based on the provided text, we can infer and construct some of the requested information, particularly regarding the performance standards (which act as implicit acceptance criteria for safety and effectiveness) and the basis for proving compliance.

    Here's an analysis of the provided information relative to your request:


    Acceptance Criteria and Device Performance

    The document does not present explicit "acceptance criteria" in the format of defined metrics and thresholds (e.g., "sensitivity must be > X%, specificity > Y%"). Instead, the acceptance of the device is based on its compliance with a robust set of internationally recognized performance and safety standards, and its substantial equivalence to legally marketed predicate devices. The "reported device performance" is primarily the assertion that the device meets these standards and, therefore, is safe and effective for its intended use.

    Inferred Acceptance Criteria (Based on Compliance Standards) and Reported Performance:

    CategoryInferred Acceptance Criteria (Standard Compliance)Reported Device Performance
    General Safety & PerformanceCompliance with AAMI/ANSI/ES60601-1 (General requirements for basic safety and essential performance). This implies meeting requirements for electrical safety, mechanical safety, thermal safety, protection against hazardous output, etc."The Leltek Ultrasound Imaging System has been evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness as well as wireless, thermal, electrical, electromagnetic and mechanical safety and has been found to conform with applicable medical device safety standards." "All the safety and performance tests meet the essential requirement to its intended use."
    Electromagnetic Compatibility (EMC)Compliance with IEC 60601-1-2 (Electromagnetic disturbances - Requirements and tests). Ensures the device doesn't interfere with other devices and is not unduly affected by external electromagnetic fields."found to conform with applicable medical device safety standards." (Implied by overall safety compliance)
    UsabilityCompliance with IEC 60601-1-6 and IEC 62366-1 (Application of usability engineering to medical devices). Ensures the device is safe and effective to use by its intended users."found to conform with applicable medical device safety standards." (Implied by overall safety compliance)
    Ultrasonic PerformanceCompliance with IEC 60601-2-37 (Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment). This would cover aspects like acoustic output limits, image quality (though specific quantitative metrics are not provided in this document), and other specific ultrasound safety aspects. It also mentions AIUM/NEMA UD 2- 2004 R2009 and AIUM/NEMA UD 3- 2004 R2009 which are standards for diagnostic ultrasound measurement and reporting."Maximum acoustic output level is under by the FDA recommended limit and power level is displayed all the time." "All the safety and performance tests meet the essential requirement to its intended use."
    BiocompatibilityCompliance with ISO 10993-1, -5, -10 (Biological evaluation of medical devices for cytotoxicity, irritation, and skin sensitization)."biocompatibility... has been found to conform with applicable medical device safety standards."
    Software Life CycleCompliance with IEC 62304 (Medical device software - Software life-cycle processes). Ensures robust software development, verification, and validation.Implied by "Verification and validation testing has been conducted... and ascertain that it is safe for use by physicians." and "found to conform to applicable electromedical devices safety standards."
    Quality Management SystemCompliance with ISO 13485 (Medical devices - Quality management systems - Requirements for regulatory purposes).Not directly stated as a performance result, but a prerequisite for manufacturing and maintaining safety/effectiveness. This is usually audited.
    Risk ManagementCompliance with ISO 14971 (Medical devices - Application of risk management to medical devices).Not directly stated as a performance result, but a process followed.
    Substantial EquivalenceDevice performance (including safety and effectiveness) must be comparable to the predicate device to establish substantial equivalence."The differences between the proposed device, Leltek Ultrasound Imaging system and the predicate devices do not raise new questions of safety or effectiveness." "can be considered substantially equivalent to the listed predicate devices."

    Study Details (Based on available information)

    This document is a 510(k) summary, which generally focuses on demonstrating substantial equivalence through non-clinical performance testing and comparison to predicates, rather than extensive clinical trials for a novel device.

    1. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):

      • Test Set Sample Size: Not specified. The document states that "The Leltek Ultrasound Imaging System did not require clinical testing to establish substantial equivalence." This implies that the 'test set' primarily encompasses data from bench testing, engineering validation, and comparison to technical specifications and predicate device performance, rather than a clinical dataset.
      • Data Provenance: Not specified, but given the manufacturing location (Taiwan R.O.C.), it's plausible that much of the engineering and bench test data would originate from there. It's non-clinical data, likely from internal verification and validation activities.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

      • Not Applicable / Not Specified: Since clinical testing was not required for substantial equivalence, clinical ground truth established by experts (like radiologists) on patient data was not a primary component of this submission. The "ground truth" for the non-clinical tests would be defined by the technical specifications, reference standards, and established engineering principles.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • Not Applicable: No clinical test set requiring expert adjudication is described in the document.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No: The document explicitly states, "The Leltek Ultrasound Imaging System did not require clinical testing to establish substantial equivalence." This is not an AI-assisted device, it's a diagnostic ultrasound imaging system; therefore, an MRMC study comparing human readers with and without AI assistance would not be relevant or expected for this type of device and submission.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Not Applicable: This is a diagnostic ultrasound imaging system, not an AI algorithm. Its performance is based on its ability to acquire and display images/Doppler data, not standalone diagnostic interpretation.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • Technical Specifications and Standard Compliance: The "ground truth" for this device's performance demonstration is its adherence to the technical specifications outlined in its design and its compliance with the various international safety and performance standards listed (e.g., IEC 60601 series, ISO 10993 series). This is evidenced through a combination of engineering tests (e.g., acoustic power measurements against limits, electrical safety tests, EMC tests), biocompatibility tests, and software verification and validation against requirements and design specifications.
    7. The sample size for the training set:

      • Not Applicable / Not Specified: This is not a machine learning/AI device requiring a "training set." The device is a hardware and software system for image acquisition and display.
    8. How the ground truth for the training set was established:

      • Not Applicable: As above, there is no "training set" for this type of device.