510(k) Data Aggregation

    K Number
    K213772

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-01-19

    (413 days)

    Product Code
    Regulation Number
    876.5990
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Lite-Med LM-9300 Plus Lithotripter is indicated for fragmentation of kidney stones and renal pelvic stones and for upper ureteral stones by extracorporeal shock wave lithotripsy (ESWL).

    Device Description

    The Lite-Med LM-9300 Plus is an Electromagnetic Extracorporeal Shock Wave Lithotripter that effectively treats urinary calculi. It is routinely used for the fragmentation of kidney and ureteral stones and offers a good combination of clinical performance, flexibility and affordability. The standard LM-9300 Plus system consists of a shockwave generator, an operator interface (industrial PC with dual monitors), a water circulation subsystem and a patient handling subsystem. For the Extracorporeal Shock Wave Lithotripsy (ESWL) operation to be fully functional, two more optional subsystems are necessary. The first is a C-arm X-ray fluoroscopy device and the second is an ultrasound imaging unit. Normally one of the imaging devices is sufficient. For most advanced ESWL designs such as LM-9300 Plus both X-ray and ultrasound are used for patient positioning and monitoring purposes.

    Shock waves are generated on the basis of a principle similar to that used in loudspeakers. An electrical impulse is sent through an inductance coil, generating a magnetic field which repulses a metallic membrane. The acoustic impulse created by this repulsion is focused by an acoustic lens to form a shock wave. A water circulation subsystem is used to provide transmission of shockwaves and cooling of the generator.

    AI/ML Overview

    The provided text describes a medical device, the Lite-Med LM-9300 Plus Lithotripter, and its clearance by the FDA based on substantial equivalence to a predicate device. However, it does not contain information about acceptance criteria, device performance metrics, or any study involving AI or human readers.

    The document specifically states: "Clinical performance is not required to demonstrate substantial equivalence to the predicate devices." This indicates that no clinical performance study, multi-reader multi-case study, or standalone algorithm performance assessment was conducted or submitted for this device in the context of its 510(k) clearance. The clearance is based on the device's technical characteristics being similar to a previously cleared predicate device and compliance with relevant safety standards.

    Therefore, I cannot provide the requested information. The document focuses on regulatory compliance and technical equivalence, not clinical performance data or AI model evaluation.

    Ask a Question

    Ask a specific question about this device

    K Number
    K142561

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2015-03-20

    (190 days)

    Product Code
    Regulation Number
    876.5990
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Lite-Med LM-9300 ELMA Lithotripter is indicated for fragmentation of kidney stones such as renal calyx stones and renal pelvic stones and for upper, middle and lower ureteral stones by extracorporeal shock wave lithotripsy (ESWL).

    Device Description

    The Lite-Med LM-9000 ELMA is an Electromagnetic Extracorporeal Shock Wave Lithotripter that effectively treats urinary calculi. It is routinely used for the fragmentation of kidney and ureteral stones and offers a good combination of clinical performance, flexibility and affordability. The standard LM-9300 ELMA system consists of a shockwave generator, an operator interface (industrial computer with dual monitors), a water circulation subsystem and a patient handling subsystem. For the Extracorporeal Shock Wave Lithotripsy (ESWL) operation to be fully functional, one or two more subsystems are necessary. The first is a C-arm X-ray fluoroscopy device and the second is an ultrasound imaging unit. Normally one of the imaging devices is sufficient. For most advanced ESWL designs such as LM-9300 ELMA both X-ray and ultrasound are used for patient positioning and monitoring purposes.

    Shock waves are generated on the basis of a principle similar to that used in loudspeakers. An electrical impulse is sent through an inductance coil, generating a magnetic field which repulses a metallic membrane. The acoustic impulse created by this repulsion is focused by an acoustic lens to form a shock wave. A water circulation subsystem is used to provide transmission of shockwaves and cooling of the generator.

    AI/ML Overview

    Here's an analysis of the provided text to extract the acceptance criteria and study details for the LM-9300 ELMA Lithotripter:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" in a typical quantitative pass/fail format for a diagnostic device. Instead, it describes performance based on the clinical outcome of stone fragmentation. The key performance indicator is the "stone-free ratio."

    Acceptance Criteria Category (Implied)Reported Device Performance (LM-9300 ELMA)
    Clinical Effectiveness (Stone-Free)Averaged stone-free ratio at one month after treatment: 80%
    Effectiveness (Residual Fragments)Effectiveness (fragments < 5mm x 4mm): better than 90%

    It's important to note that these "acceptance criteria" are derived from the observed clinical performance and are presented as evidence of the device's efficacy, rather than pre-defined thresholds the device had to meet to be approved. The document primarily focuses on demonstrating substantial equivalence to predicate devices and adherence to safety standards.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: 44 patients with 44 stones.
    • Data Provenance: The study was a prospective clinical investigation performed at 2 sites. The country of origin is not explicitly stated, but the manufacturer is Lite-Med Inc. from Taipei City, Taiwan.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This information is not provided in the document. For a lithotripter, ground truth would typically refer to the confirmed presence and size of stones before treatment, and the assessment of stone fragmentation and stone-free status after treatment. This assessment would usually be done by medical professionals (e.g., urologists, radiologists) using imaging techniques (X-ray, ultrasound). The document states "follow-up to support this application," implying expert assessment, but does not detail the number or qualifications.

    4. Adjudication Method for the Test Set

    This information is not provided in the document.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No, a multi-reader multi-case comparative effectiveness study was not conducted.
    • The study described is a clinical investigation focused on the LM-9300 ELMA's performance, not a comparison of human readers with and without AI assistance. The LM-9300 ELMA is a physical medical device (lithotripter) that treats stones, not a diagnostic AI algorithm.

    6. Standalone (Algorithm Only) Performance Study

    • No, a standalone performance study was not conducted in the context of an AI algorithm.
    • The LM-9300 ELMA is a physical lithotripter. The performance study refers to its efficacy in fragmenting stones in human patients.

    7. Type of Ground Truth Used

    The ground truth implicitly used for the clinical study was based on clinical outcomes/assessments of stone presence, size, fragmentation, and "stone-free" status at 1, 2, and 4 weeks follow-up. This likely involved medical imaging (X-ray, ultrasound) and clinical judgment by treating physicians, although specific methods are not detailed.

    8. Sample Size for the Training Set

    The document describes a clinical investigation (the "test set" or evaluation dataset), but it does not mention a separate "training set" for an algorithm. This is consistent with the device being a physical lithotripter rather than a machine learning or AI-driven diagnostic tool.

    9. How the Ground Truth for the Training Set Was Established

    Since no training set for an algorithm is mentioned, this information is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K103217

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2011-09-28

    (331 days)

    Product Code
    Regulation Number
    876.5990
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Lite-Med LM-9200ELMA Lithotripter is indicated for fragmentation of kidney stones such as renal calyx stones and renal pelvic stones and for upper, middle, and lower ureteral stones.

    Device Description

    The Lite-Med LM-9200 ELMA is an Electromagnetic Extracorporeal Shock Wave Lithotripter that effectively treats urinary calculi. It is routinely used for the fragmentation of kidney and ureteral stones and offers a good combination of clinical performance, flexibility and affordability. The standard LM-9200 ELMA device consists of a Shockwave Generator, an operator interface/touch panel, and a water circulation subsystem. Shock waves are generated on the basis of a principle similar to that used in loudspeakers. An electrical impulse is sent through an inductance coil, generating a magnetic ficld which repulses a metallic membrane. The acoustic impulse created by this repulsion is focused by an acoustic lens to form a shock wave. A water circulation subsystem is used to provide transmission of shockwaves and cooling of the generator.

    For the ESWL operation to be fully functional, two or three optional subsystems are needed. The first is a special treatment table. The second and third are a C-arm X-ray fluoroscope and an ultrasound imaging unit. The treatment table is a motorized floating table which can be moved electrically in all three axes.

    AI/ML Overview

    The provided document describes the Lite-Med LM-9200 ELMA Lithotripter, an Extracorporeal Shock Wave Lithotripter (ESWL) for fragmenting kidney and ureteral stones.

    1. Table of Acceptance Criteria and Reported Device Performance

    The direct acceptance criteria for clinical performance are not explicitly stated as numerical targets in the provided text. Instead, the study reports an "overall success rate" against which its performance is implicitly measured, and also highlights safety and compliance with standards.

    Acceptance Criteria (Implied)Reported Device Performance
    Overall clinical success in stone fragmentation85% success rate
    Safety and high evaluation for device functionPatients treated are safe and have high evaluations for device function. No device malfunction incidence reported.
    Compliance with relevant medical device safety standardsIEC 60601-2-36, IEC 61846, IEC 60601-1-1, IEC 60601-1-2, IEC 60601-1-4, ISO 13485, ISO 14971
    Adequacy of user's manualUser's manual is adequate for operation.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: 40 patients (30 male, 10 female)
    • Data Provenance: Not explicitly stated, but the company is based in Taipei City, Taiwan. The clinical investigations were performed at 2 sites, implying prospective data collection for the purpose of this application.

    3. Number of Experts Used to Establish Ground Truth and Their Qualifications

    The document refers to "experiences of physicians" but does not specify the number of experts used to establish a formal "ground truth" for the test set, nor their specific qualifications (e.g., years of experience). The assessment of device function and safety, as well as the success rate, would presumably be based on clinical evaluations by the treating physicians at the two sites.

    4. Adjudication Method for the Test Set

    The document does not describe any specific adjudication method (e.g., 2+1, 3+1) for establishing the success rate or safety outcomes. Clinical outcomes were reported, implying a direct assessment by the treating physicians without an explicit external adjudication panel.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No Multi-Reader Multi-Case (MRMC) comparative effectiveness study was mentioned. The study focused on the performance of the LM-9200 ELMA Lithotripter alone.

    6. Standalone Performance Study

    Yes, a standalone study was performed. The clinical investigation directly assessed the performance of the LM-9200 ELMA Lithotripter without human-in-the-loop assistance for the primary function of stone fragmentation. The clinical success rate of 85% for stone fragmentation and the safety profile were assessed based on the device's use.

    7. Type of Ground Truth Used

    The ground truth appears to be based on clinical outcomes as assessed by treating physicians. The "overall success rate" would be defined by the clinical fragmentation of kidney and ureteral stones, likely confirmed through imaging or other medical follow-ups. Safety was also assessed based on clinical observations and patient evaluations.

    8. Sample Size for the Training Set

    The document does not mention a "training set" in the context of an algorithm or AI. The device is an Extracorporeal Shock Wave Lithotripter, a physical medical device, not an AI/software as a medical device (SaMD). Therefore, the concept of a training set as typically applied to machine learning models is not relevant here.

    9. How the Ground Truth for the Training Set Was Established

    As noted above, the device is a physical lithotripter, not an AI/ML algorithm. Thus, there is no "training set" or ground truth establishment method for a training set in this context. The physical and technological characteristics of the device were established through measurements based on standards like IEC 61846.