510(k) Data Aggregation

    K Number
    K092550

    Validate with FDA (Live)

    Device Name
    TIGRT IVS
    Manufacturer
    Date Cleared
    2009-12-15

    (118 days)

    Product Code
    Regulation Number
    892.5050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    TiGRT IVS is intended to be used in conjunction with Linac and Digital Panel for analyzing the current patient position and calculating the patient positioning shift correction factor, as well as the treatment portal verification and record.

    Device Description

    TiGRT IVS is intended to be used in conjunction with Linac and Digital Panel for analyzing the current patient position and calculating the patient positioning shift correction factor, as well as the treatment portal verification and record.

    AI/ML Overview

    The provided text does not contain any information about acceptance criteria or a study proving the device meets acceptance criteria.

    The document is a 510(k) premarket notification for an image viewing system called TiGRT IVS. It primarily focuses on demonstrating substantial equivalence to a predicate device (ExacTrac K072046) based on intended use, design, and performance characteristics.

    Here's what the document does include:

    • Device Name: TiGRT IVS
    • Intended Use: To be used in conjunction with Linac and Digital Panel for analyzing the current patient position, calculating patient positioning shift correction factor, and for treatment portal verification and record.
    • Predicate Device: ExacTrac (K072046)
    • Classification: Class II device (21 CFR 892.5050)
    • Comparison with Predicate Device: A table comparing characteristics like operating system, networking, intended use, application, and DICOM compatibility. All characteristics are either identical or considered substantially equivalent.

    Therefore, I cannot provide the requested information in the table format because the source text does not contain it. This type of document (a 510(k) summary) focuses on substantial equivalence rather than explicit performance data against acceptance criteria.

    Ask a Question

    Ask a specific question about this device

    K Number
    K090893

    Validate with FDA (Live)

    Device Name
    TIGRT TPS
    Manufacturer
    Date Cleared
    2009-07-15

    (106 days)

    Product Code
    Regulation Number
    892.5050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    TiGRT TPS is a Radiation Therapy Treatment Planning System for radiation dose planning of patients undergoing external beam treatment in the oncology clinic. TiGRT TPS is used to plan radiation treatments with linear accelerators and other similar teletherapy devices with x-ray and/or electron energies from 1 to 25MV, as well as Cobalt-60.

    Device Description

    TiGRT TPS is a Radiation Therapy Treatment Planning System for radiation dose planning of patients undergoing external beam treatment in the oncology clinic. TiGRT TPS is used to plan radiation treatments with linear accelerators and other similar teletherapy devices with x-ray and/or electron energies from 1 to 25MV, as well as Cobalt-60.

    AI/ML Overview

    The provided text describes a 510(k) submission for a Radiation Treatment Planning System called TiGRT TPS. This submission focuses on demonstrating substantial equivalence to a predicate device (WiMRT, K041971) rather than presenting a detailed performance study with specific acceptance criteria and outcome metrics for standalone or human-in-the-loop performance.

    Here's an analysis of the requested information based on the provided document:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state quantitative acceptance criteria for device performance (e.g., accuracy of dose calculation, speed of optimization) or present specific reported device performance metrics against such criteria. Instead, the acceptance is based on demonstrating substantial equivalence to a predicate device. The "performance documentation" listed is about the process of development and risk management, not a direct measurement of the device's functional output against numerical benchmarks.

    Acceptance Criteria (Implied by Substantial Equivalence)Reported Device Performance
    Intended Use is the same as the predicate.TiGRT TPS has the same intended use as WiMRT.
    Technological Characteristics are similar to the predicate.TiGRT TPS has similar design, function, application, operating system, networking, application use (Conformal, IMRT, SRS/SRT), dose calculation algorithm (Super-position Convolution), and IMRT optimization algorithm (Genetic Algorithm) as WiMRT. It uses DICOM 3/RT.
    No new issues of safety or effectiveness are introduced."No new issues of safety or effectiveness are introduced by using this device." Also, "No new issues of biocompatibility are raised."
    Compliance with design and development controls.Provided documentation including: Level of Concern of TPS (Major), TiGRT TPS Description, Risk Management, Hazard Analysis Report, Product Requirement Specification, Architecture Design Chart, Software Design Description, PRS/SDD Traceability Matrix, PRS/STT Traceability Matrix, Development Environment Description, Verification and Validation Documents, Revision Level History, and Unresolved Anomalies.

    2. Sample size used for the test set and the data provenance

    The document does not describe a traditional "test set" in the context of clinical or performance validation data (e.g., patient cases, imaging data) for evaluating the device's diagnostic or treatment accuracy. The submission focuses on comparing the new device's specifications and design to a predicate, and the provided "Performance Documentation" relates to software development and risk management. Therefore, no specific sample size or data provenance for a performance test set is mentioned.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Since no specific test set involving medical data (e.g., images for diagnosis, treatment plans for review) is described, this information is not applicable and not provided in the document. The "ground truth" for this submission revolves around the technical specifications and safety profile being equivalent to an already approved device.

    4. Adjudication method for the test set

    As there is no described test set requiring expert review or ground truth establishment, no adjudication method is mentioned.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance

    The TiGRT TPS is a Radiation Treatment Planning System, not a diagnostic AI tool that assists human readers. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance is not described and not relevant to this type of device. The document does not mention any AI capabilities that would directly assist human interpretation in a diagnostic sense.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The document does not provide details of a standalone performance study in terms of metrics like dose calculation accuracy against physical measurements or clinical outcomes for a specific patient cohort. The "performance documentation" relates to the software development process and risk analysis rather than direct performance measurement of the algorithm itself against ground truth. The acceptance is based on design and functional similarity to the predicate, implying that if the design is substantially equivalent, the performance is also considered equivalent.

    7. The type of ground truth used

    The concept of "ground truth" in this 510(k) submission is primarily established by:

    • Predicate Device Specifications: The technical characteristics and intended use of the legally marketed predicate device (WiMRT) serve as the "ground truth" or benchmark against which TiGRT TPS is compared for substantial equivalence.
    • Industry Standards and Best Practices: Implied "ground truth" for software development and risk management practices (e.g., "Hazard Analysis Report," "Verification and Validation Documents," "Software Design Description") ensure the device is developed safely and effectively, analogous to how the predicate was developed.

    8. The sample size for the training set

    The document does not describe any machine learning or AI components that would require a "training set" of data. The device relies on established dose calculation algorithms (Super-position Convolution) and optimization algorithms (Genetic Algorithm) which typically do not involve statistical machine learning training on a large dataset in the same way modern AI systems do. Therefore, this information is not applicable and not provided.

    9. How the ground truth for the training set was established

    As there is no described training set for a machine learning model, this information is not applicable and not provided.

    Ask a Question

    Ask a specific question about this device

    K Number
    K090802

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2009-07-07

    (105 days)

    Product Code
    Regulation Number
    892.5050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    TIGRT MLC is a dynamic multileaf collimator designed to be mounted on the linear accelerator. It is intended to shape the specific fields, either in static or dynamic mode, to assist the radiation oncologist in the delivery of radiation to well-defined target volumes while sparing surrounding normal tissue and critical organs from excess radiation.

    Device Description

    TIGRT MLC is a dynamic multileaf collimator designed to be mounted on the linear accelerator. It is intended to shape the specific fields, either in static or dynamic mode, to assist the radiation oncologist in the delivery of radiation to well-defined target volumes while sparing surrounding normal tissue and critical organs from excess radiation.

    AI/ML Overview

    The provided document, KO90802 from July 7, 2009, is a 510(k) summary for the TiGRT MLC (Dynamic Multileaf Collimator). This document establishes substantial equivalence to a predicate device and does not contain information about acceptance criteria or a study that specifically proves the device meets such criteria in terms of analytical or clinical performance metrics.

    Instead, this submission focuses on demonstrating substantial equivalence to the DMLC (K060187) based on similarity in design, construction, intended use, and performance characteristics. The comparison table (Table 1) highlights minor differences in operating system, number of COM ports, and selectable leaf pairs, which are not presented as acceptance criteria or performance metrics derived from a study.

    Therefore, I cannot fulfill your request for:

    • A table of acceptance criteria and the reported device performance: This information is not present in the provided document. The table in the document is for comparing characteristics with a predicate device, not for reporting performance against acceptance criteria.
    • Sample sized used for the test set and the data provenance: No test set or data provenance is mentioned as part of a performance study.
    • Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Ground truth establishment for a test set is not discussed.
    • Adjudication method for the test set: No test set adjudication is mentioned.
    • If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: No MRMC study is described. This device is a mechanical component for radiation delivery, not an AI-assisted diagnostic tool for human readers.
    • If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: No standalone algorithm performance is discussed.
    • The type of ground truth used: No ground truth is mentioned in the context of a performance study.
    • The sample size for the training set: No training set is mentioned.
    • How the ground truth for the training set was established: No training set ground truth establishment is mentioned.

    The document primarily states:

    • "This device is similar in design and construction, and has the same intended use and performance characteristics to the predicate device."
    • "No new issues of safety or effectiveness are introduced by using this device."

    This type of submission relies on demonstrating that the new device functions similarly to an already approved device, rather than providing novel clinical or analytical performance data against specific, quantifiable acceptance criteria.