510(k) Data Aggregation

    K Number
    K193010

    Validate with FDA (Live)

    Date Cleared
    2020-11-18

    (386 days)

    Product Code
    Regulation Number
    880.5440
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Liebel-Flarsheim™ IBP Transfer Set is indicated for the transfer of Optiray® (loversol) Imaging Bulk Package contrast media to empty, single-use only, sterile syringes on syringe based contrast injection systems indicated for the controlled, automatic intra-vascular administration of radiopaque contrast agents for their indicated procedures. The IBP Transfer Set is to be discarded after the contrast media container has been emptied or after 12 hours have elapsed since the container has been penetrated, whichever occurs first.

    Contra-indications for the use of this device are determined by the prescribing physician at the time of use based upon the contrast media package inserts.

    Device Description

    The Liebel-Flarsheim™ IBP Transfer Set is a sterile tubing set connection between Optiray® (Ioversol) contrast media imaging bulk packages (hereinafter Optiray IBPs) and empty sterile singleuse syringes on syringe-based automatic contrast injection systems. The Liebel- Flarsheim™ IBP Transfer Set consists of a small bore (1/8" ID) clear polyvinyl chloride (PVC) tube with standard medical component end terminations. Bonded to the proximal end of the PVC tube is a vented IV spike that is used to pierce a rubber stopper belonging to an imaging bulk contrast media container. Distally, a self-sealing Luer type connector (Smiths Medical Nu- Site®) is bonded to the opposite end of the PVC tube. The Liebel-Flarsheim™ IBP Transfer Set, shown in the image below, has an overall length of 19-inches, a protective cap on the IV Spike, and both untethered and tethered protective caps on the Luer end to protect the Nu-Site® Luer Activated fitting between syringe fills. The transfer sets are individually packaged in blister packs with a perimeter heat sealed Tyvek lid. Sealed blister packs are packaged in an outer box containing 50 units (50 units/box) which are terminally sterilized with ethylene oxide.

    The Liebel-Flarsheim™ IBP Transfer Set is intended for the transfer of fluids from bulk containers to empty sterile syringes on syringe-based contrast injection systems.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Liebel-Flarsheim™ IBP Transfer Set, based on the provided text:

    Important Note: The provided document is a 510(k) summary for a medical device (a transfer set) and does not describe an AI/ML-based device. Therefore, many of the requested fields related to AI/ML studies (like test set size for AI, number of experts for ground truth, MRMC studies, training set size for AI, etc.) are not applicable to this device. The study described is a traditional medical device verification and validation (V&V) study, comparing a new device to a predicate device.


    Acceptance Criteria and Device Performance (Verification and Validation)

    The document describes various tests performed to demonstrate substantial equivalence to a predicate device, focusing on safety and effectiveness. The acceptance criteria are generally "compliance with the standards" or "performs as intended."

    Acceptance Criteria CategoryAcceptance Criteria (Implied)Reported Device Performance/Verification
    SterilizationSterility Assurance Level (SAL) of 10⁻⁶; EtO residuals comply with ISO 10993-7:2008 limits.Validated to SAL of 10⁻⁶ in accordance with ISO 11135:2014 and ISO 10993-7:2008. Validation results indicate compliance.
    Shelf LifeMaintains integrity of primary package and functional user requirements for 3 years.Verification and validation testing in accordance with ISO 11607 (framework described, not full conformity) demonstrated maintenance of primary package integrity. Functional user requirements were also verified over the proposed three-year shelf life.
    BiocompatibilityVerified in accordance with ISO 10993-1:2018 for cytotoxicity, hemocompatibility, irritation, sensitization, acute systemic toxicity, material mediated pyrogenicity; complies with ISO 10993-18:2005 (chemical characterization), USP <788> (particulate matter), and USP <85> (bacterial endotoxin).Verification results indicate that the materials comply with the standards listed: ISO 10993-1:2018 (cytotoxicity, hemocompatibility, irritation, sensitization, acute systemic toxicity, material mediated pyrogenicity), ISO 10993-18:2005, USP <788>, and USP <85>.
    Bench Performance (General/Functional)Performance and functional user requirements verified according to product specifications and relevant ISO standards.Tested for performance and verified in accordance with ISO 8536-4 (infusion sets), ISO 594-2 (Luer taper fittings), and ISO 22413:2010 (transfer sets). Additional testing included Microbial Ingress Testing, Chemical Compatibility (with Optiray®), Extractables and simulation testing for leachable compounds and particulates, and Verification of functional performance according to product specifications. The device "performs as intended."
    Microbial IngressNo microbial ingress (implied good performance).Performed, implied successful.
    Chemical CompatibilityCompatible with Optiray® (Ioversol) Imaging Bulk Package contrast media.Approved Release Specifications for Optiray® (Ioversol) Imaging Bulk Package.
    Extractables/LeachablesAcceptable levels of extractable and leachable compounds and particulates (implied).Extractables and simulation testing for leachable compounds and particulates was performed.

    Study Details (Traditional Medical Device V&V)

    1. Sample size used for the test set and the data provenance:

      • The document does not specify a "test set" in the context of an AI/ML model where data provenance like country of origin or retrospective/prospective is relevant.
      • Instead, for traditional device testing, sample sizes would be determined by statistical methods for each specific test (e.g., sterilization validation, biocompatibility, functional performance). These specific sample sizes are not explicitly stated in this summary for each test, but standard good manufacturing practices and regulatory requirements would apply.
      • The data provenance is implicitly from manufacturing and laboratory testing conducted by Liebel-Flarsheim Company LLC (part of Guerbet LLC).
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • N/A. This is not an AI/ML device, so there is no "ground truth" in the sense of expert annotation for an algorithm. The "ground truth" for this type of device is established through adherence to recognized international standards (ISO, USP) and verified performance against predefined specifications, often evaluated by qualified laboratory personnel and engineers.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • N/A. Adjudication methods are specific to expert consensus in AI/ML model evaluation. For this device, compliance is assessed against objective technical specifications and performance limits defined by the relevant standards and internal company requirements.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • N/A. This is not an AI/ML device. MRMC studies are for evaluating clinical performance of diagnostic algorithms with human readers.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • N/A. This is not an AI/ML device.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • N/A (in the AI/ML context). For this device, "ground truth" is based on:
        • Objective physical measurements and chemical analyses: e.g., sterility levels, EtO residuals, particulate matter, leachable compounds.
        • Functional performance tests: verifying fluid transfer, connection integrity (Luer taper), and absence of microbial ingress.
        • Biocompatibility evaluations: ensuring the materials are safe for patient contact according to established biological testing standards.
        • Adherence to recognized industry standards: ISO, USP.
    7. The sample size for the training set:

      • N/A. This is not an AI/ML device. There is no "training set."
    8. How the ground truth for the training set was established:

      • N/A. This is not an AI/ML device. There is no "training set" or "ground truth" for AI training. The safety and effectiveness of this device are established through direct physical, chemical, and biological testing, manufacturing controls, and adherence to design specifications and regulatory standards.
    Ask a Question

    Ask a specific question about this device

    K Number
    K152361

    Validate with FDA (Live)

    Date Cleared
    2016-01-14

    (147 days)

    Product Code
    Regulation Number
    870.1650
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The OptiOne Single-Head Contrast Delivery System is a contrast delivery system and is designed to inject radiopaque contrast media into a patient's vascular system to obtain diagnostic images with computed tomography (i.e. "CT") equipment.

    Device Description

    The OptiOne Single-Head Contrast Delivery System (OptiOne) is intended to inject radiopaque contrast media into a patient's vascular system to obtain diagnostic images when used in conjunction with computed tomography equipment. Each injection is accomplished with a motor-driven syringe mechanism with microprocessor control of the flow rate, volume, pressure and timing. The OptiOne consists of the following key components: a Powerhead, a Console, and a Power Supply. The Power Supply delivers power to the Powerhead and Console. The Console provides remote control and programmability of the injection through a touch screen display system. The Powerhead contains the components to perform the actual injection such as motor control circuits, ball screw and syringe push ram.

    AI/ML Overview

    The provided document is a 510(k) summary for the Liebel-Flarsheim Company LLC's OptiOne Single-Head Contrast Delivery System. It details the device's characteristics and its comparison to a predicate device, focusing on demonstrating substantial equivalence.

    However, the document does not contain the specific acceptance criteria, reported device performance data, detailed descriptions of studies (sample sizes, data provenance, expert ground truth, adjudication methods, MRMC studies, standalone performance), or information on the ground truth establishment for either test or training sets as requested in your prompt.

    The document states: "Extensive verification and validation activities were performed to ensure the OptiOne Single-Head Contrast Delivery System design outputs met all design inputs and the customer requirements were successfully met. The testing confirms the OptiOne Single-Head Contrast Delivery System meets the required specifications." And, "The OptiOne Single-Head Contrast Delivery System was developed under the Liebel-Flarsheim Company LLC Design Control processes. The verification and validation activities ensure the OptiOne meets all customer requirements and product specifications."

    This indicates that internal testing was conducted to verify that the device met its specifications, but the specifics of those tests, their results against acceptance criteria, or the methodology of the studies are not included in this publicly available summary.

    Therefore, I cannot provide the requested information from the given text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K121838

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-07-11

    (19 days)

    Product Code
    Regulation Number
    892.1680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Liebel-Flarsheim Direct Digital Imaging System is a solid state detector X-ray system that facilitates digital radiologic and/or fluoroscopic procedures requiring a beam of diagnostic quality radiation and a flat imaging table primarily for urological applications such as functional x-ray diagnostics, endourology and minimal invasive urology/surgery. The system may be used for urological, gastroenterological, and gynecological treatment, planning and diagnostic procedures. It is intended to replace fluoroscopic images formerly obtained through image intensifier technology. Not intended for mammography applications.

    Device Description

    The Liebel-Flarsheim Direct Digital Imaging System is a radiographic and fluoroscopy examination' table with the X-ray tube over the table and the detector underneath the patient table. The table top can be moved longitudinally and laterally as well as vertically. The imaging system and table top can be rotated in the Trendelenburg and reverse-Trendelenburg positions. The system is a modified version of the Hydra Vision Urological X-Ray System (marketed as Hydra Vision Plus DR System). The modification features a multipurpose (radiography and fluoroscopy) solid state digital detector in place of the Image Intensifier and CCD camera for image acquisition. The table design remains unchanged.

    AI/ML Overview

    Here's a summary of the acceptance criteria and study information for the Liebel-Flarsheim Direct Digital Imaging System based on the provided 510(k) summary:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria/TestReported Device Performance
    Safety TestingElectrical SafetySatisfactory (conformance to IEC 60601-1, IEC 60601-1-1)
    Mechanical SafetySatisfactory (conformance to IEC 60601-1, IEC 60601-1-3, IEC 60601-2-32)
    Electromagnetic Compatibility (EMC)Satisfactory (conformance to IEC 60601-1-2 3rd edition 2007 - 03)
    Performance TestingContrast ResolutionMeets required specifications
    Maximum Entrance ExposureMeets required specifications
    Maximum Viewable ImageMeets required specifications
    Motion Imaging PerformanceMeets required specifications
    Substantial EquivalenceComparison to Predicate Device (Hydra Vision Urological X-Ray System)Confirmed. Device does not raise new significant issues of safety, effectiveness, or performance.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify a "test set" in terms of patient data or image datasets. The performance evaluation appears to be based on functional testing of the device's technical capabilities rather than a clinical study with a patient cohort. Therefore, information on data provenance (country of origin, retrospective/prospective) and sample size for a test set of patient data is not applicable/provided.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    As the evaluation was functional testing of the device's technical performance and not a clinical study assessing image interpretation against a ground truth, this information is not applicable/provided.

    4. Adjudication Method for the Test Set

    Not applicable, as no clinical test set requiring adjudication is described.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, a multi-reader multi-case (MRMC) comparative effectiveness study is not mentioned or described. The comparison is made at a technological level between the new device and its predicate, not between human readers with and without AI assistance.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done

    The device is an imaging system (hardware and associated software for image acquisition and processing), not an AI algorithm for image analysis. Therefore, a standalone algorithm performance study is not applicable. The "performance data" described refers to the functional performance of the imaging system itself.

    7. The Type of Ground Truth Used

    For the functional performance criteria (contrast resolution, exposure, viewable image, motion imaging), the "ground truth" would be the expected technical specifications or standards that the device is designed to meet. For instance, contrast resolution would be measured against established technical benchmarks or the performance of the predicate device.

    8. The Sample Size for the Training Set

    Not applicable. The device is a hardware imaging system. There is no mention of an AI model or "training set" in the context of this 510(k) submission.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no mention of a training set.