Search Filters

Search Results

Found 5 results

510(k) Data Aggregation

    K Number
    K063855

    Validate with FDA (Live)

    Date Cleared
    2007-02-01

    (35 days)

    Product Code
    Regulation Number
    862.1660
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For in vitro diagnostic use (i.e. for external use only) by healthcare professionals and in the home by people with diabetes mellitus to assess the performance of the Liberty Blood Glucose Monitoring System.

    Device Description

    The Liberty Control Solutions consist of a viscosity-adjusted, aqueous liquid control solution containing known quantities of glucose. They are packaged in plastic dropper tipped bottles for easy application of the control solutions to the test strips and a red coloration to aid the user to visually confirm application of the control. The product is nonhazardous and contains no human or animal derived materials.

    AI/ML Overview

    Here's an analysis of the provided information regarding the Liberty Control Solutions:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided 510(k) summary does not explicitly state specific acceptance criteria (e.g., numerical thresholds for stability or precision). It lists the types of performance studies conducted but does not detail the pass/fail criteria used. Similarly, it does not provide detailed reported device performance in terms of specific values (e.g., actual stability ranges, precision CV%).

    However, based on the general information, we can infer that the acceptance criteria would be linked to demonstrating substantial equivalence to the predicate devices. This implies that the performance (stability, precision) of the Liberty Control Solutions should be comparable to or better than that of the predicate devices.

    Acceptance Criterion (Inferred)Reported Device Performance (Summary)
    Stability (Shelf-life) is acceptable.Tests were performed to verify stability.
    Open Vial Stability is acceptable.Tests were performed to verify open vial stability.
    Microbial Stress Stability is acceptable.Tests were performed to verify microbial stress stability.
    Test precision is acceptable.Tests were performed to verify test precision.
    Performance characteristics (formulation, intended use) support substantial equivalence to predicate devices.Comparison of the performance characteristics, formulation, and intended use support the claim of substantial equivalence.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the sample size used for the test set for any of the performance studies (stability, open vial stability, microbial stress stability, test precision).

    The document does not provide information on the data provenance (e.g., country of origin, retrospective or prospective nature of data collection).

    3. Number of Experts Used to Establish Ground Truth and Qualifications of Experts

    This information is not applicable to this device. The Liberty Control Solutions are quality control materials, and their performance is assessed against established analytical methods and specifications, not against expert interpretation of biological samples or images. Therefore, there is no "ground truth" in the sense of expert consensus on patient diagnoses or pathology for this type of device.

    4. Adjudication Method for the Test Set

    This information is not applicable for the same reasons as #3. Adjudication methods are typically used in studies involving human interpretation or subjective assessments, which are not relevant to the evaluation of a quality control solution.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This type of study is relevant for diagnostic devices that involve human interpretation (e.g., radiologists reading images) to assess the impact of AI assistance on reader performance. The Liberty Control Solutions are quality control materials for blood glucose monitoring systems and do not involve human readers in this context.

    6. If a Standalone (i.e., Algorithm Only Without Human-in-the-Loop Performance) Was Done

    No, a standalone (algorithm only) performance study was not done in the context of an AI algorithm. The device is a chemical control solution, not an AI algorithm. Its performance is inherent to its chemical composition and manufacturing, assessed through laboratory tests.

    7. The Type of Ground Truth Used

    For quality control materials like the Liberty Control Solutions, the "ground truth" for performance studies would be:

    • Reference Method/Standard Measurements: For glucose concentration, the accurate measurement of glucose using a highly precise and validated reference method (e.g., hexokinase-based methods) would serve as the ground truth.
    • Established Specifications: Manufacturers define specifications for stability (e.g., acceptable percentage change over time) and precision (e.g., maximum coefficient of variation). The "ground truth" for these tests is whether the measured performance falls within these predefined acceptable ranges.

    The document does not explicitly detail the specific reference methods or internal specifications used to establish ground truth for each performance study.

    8. The Sample Size for the Training Set

    This information is not applicable. The Liberty Control Solutions are chemical control solutions, not machine learning algorithms or AI models that require a "training set."

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable for the same reasons as #8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K060706

    Validate with FDA (Live)

    Date Cleared
    2006-04-28

    (43 days)

    Product Code
    Regulation Number
    862.1660
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For in vitro diagnostic use (i.e. for external use only) by healthcare professionals and in the home by people with diabetes mellitus to assess the performance of the Roche Accu-chek Active, the Bayer Ascensia Microfill, and the LifeScan OneTouch Ultra and FastTake Blood Glucose Monitors.

    Device Description

    The Liberty Normal Glucose Control consists of a viscosity-adjusted, aqueous liquid control solution containing a known quantity of glucose. The product is packaged in plastic dropper tipped bottles for easy application of the control solutions to the test strips and a red coloration to aid the user to visually confirm application of the control. The product is non-hazardous and contains no human or animal derived materials.

    AI/ML Overview

    The provided document describes a Quality Control Material for blood glucose monitors, not an AI/ML-powered diagnostic device. Therefore, many of the requested categories for AI/ML device studies (such as sample size for test sets, data provenance, number of experts, adjudication methods, MRMC studies, standalone performance, training set data, etc.) are not applicable.

    The document focuses on demonstrating substantial equivalence to existing predicate devices based on comparable characteristics and performance studies relevant to control solutions.

    Here's a breakdown of the available information relevant to acceptance criteria and performance, alongside an explanation of why certain AI/ML-specific questions cannot be answered:

    1. Table of Acceptance Criteria and Reported Device Performance

    Characteristic/AspectAcceptance Criteria (Implied by Predicate Equivalence)Reported Device Performance (Liberty Normal Glucose Control)
    Device TypeSingle Analyte Control Solution (Glucose)Single Analyte Control Solution (Glucose)
    Number of Levels1 (matching most predicates)1
    AnalytesGlucoseGlucose
    ContainerPlastic bottle with dropper-tipPlastic bottle with dropper-tip
    Fill VolumeComparable to predicates (e.g., 2.5 mL - 4 mL range)3.6 mL
    ColorRed (matching most predicates)Red
    MatrixBuffered aqueous solution of D-Glucose, viscosity modifier, preservatives, and other non-reactive ingredientsBuffered aqueous solution of D-Glucose, viscosity modifiers, preservatives, and other non-reactive ingredients
    Intended UseTo assess performance of specific blood glucose monitorsTo assess performance of specific blood glucose monitors (Bayer Ascensia Microfill, Roche Accu-chek Active, LifeScan OneTouch Ultra, and FastTake)
    Target PopulationProfessional and home useProfessional and home use
    Performance StudiesStability (Accelerated and Real-time), Open Vial, Microbial Stress Stability, Test precision data comparable to predicates"Tests were performed to verify specific performance characteristics: Stability (Accelerated and Real-time), Open Vial, Microbial Stress Stability, Test precision." (Details not provided in this summary, but concluded to support substantial equivalence).

    Explanation of Applicability/Non-Applicability to AI/ML Questions:

    • 2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

      • Not applicable for a control solution. This device is a consumable quality control material, not a diagnostic algorithm. Performance is assessed through chemical stability and precision testing, not diagnostic accuracy on a patient dataset. The "test set" here would refer to vials of the control solution and measurements taken by various blood glucose meters.
    • 3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

      • Not applicable. Ground truth for a glucose control solution is established by precise chemical formulation and analytical testing, not by expert interpretation of medical images or patient data.
    • 4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not applicable. Adjudication methods are relevant for ambiguous diagnostic interpretations, not for the chemical properties and performance of a control solution.
    • 5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not applicable. This pertains to AI-assisted diagnostic reads, which is not what this device is.
    • 6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Not applicable. This device is a control solution, not an algorithm.
    • 7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

      • Ground truth: For a glucose control solution, the ground truth is the known, precisely manufactured concentration of glucose within the solution, verified through analytical chemistry methods.
    • 8. The sample size for the training set:

      • Not applicable. This device is not an AI/ML algorithm that requires a training set. Its formulation is based on chemical engineering and existing predicate control solutions.
    • 9. How the ground truth for the training set was established:

      • Not applicable. As above, no training set for this type of device.

    Conclusion:

    The Liberty Normal Glucose Control is a chemical control solution designed to verify the performance of blood glucose monitors. Its acceptance criteria and performance are established through demonstrating substantial equivalence to legally marketed predicate devices by comparing their physical characteristics, chemical matrix, intended use, and through standard performance studies for such products (stability, open vial, microbial stress, and test precision). The 510(k) summary indicates that these comparisons and studies support the claim of substantial equivalence. The document is not for an AI/ML diagnostic device, so questions related to AI/ML specific study designs are not relevant.

    Ask a Question

    Ask a specific question about this device

    K Number
    K060481

    Validate with FDA (Live)

    Date Cleared
    2006-03-09

    (14 days)

    Product Code
    Regulation Number
    862.1660
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For in vitro diagnostic use (i.e. for external use only) by healthcare professionals and in the home by people with diabetes mellitus to assess the performance of the TheraSense FreeStyle Blood Glucose System.

    Device Description

    The Liberty Glucose Control consists of a viscosityadjusted, aqueous liquid control solution containing a known quantity of glucose. The product is packaged in plastic dropper tipped bottles for easy application of the control solutions to the test strips and a red coloration to aid the user to visually confirm application of the control. The product is nonhazardous and contains no human or animal derived materials.

    AI/ML Overview

    The provided 510(k) summary for the Liberty Glucose Control device does not contain specific acceptance criteria with numerical thresholds, nor does it detail a comprehensive study with reported device performance against such criteria. The document states that "Tests were performed to verify specific performance characteristics" including "1. Stability (Accelerated and Real-time), 2. Open Vial, 3. Microbial Stress Stability, 4. Test precision." However, it does not provide the results of these tests or the predefined acceptance limits.

    Instead, the submission focuses on demonstrating substantial equivalence to predicate devices based on a comparison of characteristics and general performance categories.

    Therefore, many of the requested details cannot be extracted directly from this document.

    Here's an attempt to answer based on the available information:

    1. A table of acceptance criteria and the reported device performance

    This information is not provided in the document. The document states that "Tests were performed to verify specific performance characteristics" but does not enumerate the acceptance criteria for these tests nor report the specific performance values obtained from these tests.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided in the document. The submission mentions "Tests were performed," but no details about sample sizes, study design (retrospective/prospective), or data provenance are given.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This is not applicable and therefore not provided. The device is a "Single Analyte Control Solution," meaning it's a quality control material with a known quantity of glucose. The "ground truth" for this type of device would be the established concentration of glucose in the control solution itself, determined through analytical chemistry methods, not expert consensus on diagnostic images or patient outcomes.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This is not applicable and therefore not provided. Adjudication methods are typically used in studies involving human interpretation or subjective assessments, which is not the case for a quality control solution.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable and therefore not provided. The device is a glucose control solution, not an AI-powered diagnostic tool requiring human reader studies.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable and therefore not provided. The device is a physical control solution, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth for a glucose control solution is the established, precisely measured concentration of glucose within the control solution itself, determined through laboratory analytical methods.

    8. The sample size for the training set

    This is not applicable and therefore not provided. The device is a quality control material and does not involve AI or machine learning, thus there is no "training set."

    9. How the ground truth for the training set was established

    This is not applicable and therefore not provided. As there is no training set for this type of device, the concept of establishing ground truth for it does not apply.

    Summary based on available information:

    The 510(k) summary for the Liberty Glucose Control focuses on demonstrating substantial equivalence to predicate devices (TheraSense FreeStyle Control K031260 and Liberty Glucose Control K052980) based on shared characteristics and intended use. While it mentions performance studies (Stability, Open Vial, Microbial Stress Stability, Test Precision), it does not provide explicit acceptance criteria, detailed study designs, or the quantitative results of these studies.

    The "study that proves the device meets the acceptance criteria" is broadly stated as "Performance Studies: Tests were performed to verify specific performance characteristics." However, no specific data or direct comparisons against pre-defined acceptance criteria are presented in this summary document. The conclusion states that "Comparison of the performance characteristics, formulation and intended use support the claim of substantial equivalence," implying that general performance testing was conducted and deemed acceptable, but the specifics are not disclosed in this public summary.

    Ask a Question

    Ask a specific question about this device

    K Number
    K060426

    Validate with FDA (Live)

    Date Cleared
    2006-03-09

    (16 days)

    Product Code
    Regulation Number
    862.1660
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For in vitro diagnostic use (i.e. for external use only) by healthcare professionals and in the home by people with diabetes mellitus to assess the performance of Bayer Ascensia DEX 2/DEX and BREEZE Blood Glucose Monitors.

    Device Description

    The Liberty Glucose Normal Control Solution consists of a viscosity-adjusted, aqueous liquid control solution containing a known quantity of glucose. The product is packaged in plastic dropper tipped bottles for easy application of the control solutions to the test strips and a red coloration to aid the user to visually confirm application of the control. The product is nonhazardous and contains no human or animal derived materials.

    AI/ML Overview

    The provided text describes the 510(k) Premarket Notification for the "Liberty Glucose Normal Control Solution."

    Here's an analysis of the requested information based on the provided text:

    1. Table of Acceptance Criteria and the Reported Device Performance

    The document does not explicitly state numerical acceptance criteria with corresponding reported device performance values in a clear table format for the Liberty Glucose Normal Control Solution itself. Instead, it refers to performance characteristics that were verified.

    The performance studies mentioned were conducted to verify:

    • Stability (Accelerated and Real-time)
    • Open Vial
    • Microbial Stress Stability
    • Test precision

    The conclusion stated: "Comparison of the performance characteristics, formulation and intended use support the claim of substantial equivalence," implying that the device met the internal criteria for these characteristics,
    but the specific numerical targets or results are not detailed.

    2. Sample Size Used for the Test Set and the Data Provenance

    The document does not specify sample sizes for any of the tests performed (Stability, Open Vial, Microbial Stress Stability, Test Precision).
    The data provenance (country of origin, retrospective/prospective) is also not specified.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This information is not provided in the document. As this is a control solution for glucose meters, the "ground truth" would likely be based on established analytical methods for glucose concentration, not expert interpretation.

    4. Adjudication Method for the Test Set

    This information is not provided. Given the nature of a control solution, adjudication methods common in diagnostic imaging or clinical trials involving subjective interpretation would not typically apply.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. This type of study is typically relevant for interpretative diagnostic devices (e.g., imaging AI) and explicitly measures human reader performance with and without AI assistance. This device is a control solution for glucose meters.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    While the device itself is a standalone control solution (used to verify a glucose meter's performance), the document does not describe a "standalone algorithm-only performance" study in the context of an AI/algorithm. The performance studies listed (Stability, Open Vial, Microbial Stress Stability, Test Precision) refer to the chemical and physical properties of the control solution.

    7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)

    For a glucose control solution, the 'ground truth' for its performance characteristics (stability, precision) would be established by:

    • Analytical Chemistry Methods: The known quantity of glucose in the solution would be precisely measured using a highly accurate reference method.
    • Stability Testing Protocols: Standard protocols for assessing the degradation or change in concentration over time under various conditions.
    • Precision Testing Protocols: Repeated measurements to determine the reproducibility of the glucose concentration.

    The document does not explicitly state the specific methods used to establish these ground truths, but it implies standard laboratory practices for quality control materials.

    8. The Sample Size for the Training Set

    This information is not applicable or provided. The device is a physical control solution, not an algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable or provided, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K052980

    Validate with FDA (Live)

    Date Cleared
    2005-11-30

    (37 days)

    Product Code
    Regulation Number
    862.1660
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For in vitro diagnostic use (i.e. for external use only) by healthcare professionals and in the home by people with diabetes mellitus to assess the performance of the Abbott Medisense Optium, Precision, Precision 3, and Precision Xtra Blood Glucose Monitors.

    Device Description

    The Liberty Glucose Control MID consists of a viscosity-adjusted, aqueous liquid control solution containing a known quantity of qlucose. The product is packaged in plastic dropper tipped bottles for easy application of the control solutions to the test strips and a red coloration to aid the user to visually confirm application of the control. The product is non-hazardous and contains no human or animal derived materials.

    AI/ML Overview

    The provided text describes a 510(k) submission for the "Liberty Glucose Control MID" and its substantial equivalence to a predicate device. However, it does not contain specific acceptance criteria, detailed study designs, or performance results in the format requested.

    The document states that "Tests were performed to verify specific performance characteristics: 1. Stability (Accelerated and Real-time), 2. Open Vial, 3. Microbial Stress Stability, 4. Test precision." It then concludes that "Comparison of the performance characteristics, formulation and intended use support the claim of substantial equivalence," implying these tests were sufficient to meet regulatory requirements.

    Therefore, I cannot fulfill your request for a table of acceptance criteria and reported device performance, or details about sample sizes, ground truth establishment, or expert involvement, as this information is not present in the provided text.

    The document focuses on the regulatory submission rather than a detailed scientific study report with specific performance metrics against defined acceptance criteria.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1