510(k) Data Aggregation

    K Number
    K182749

    Validate with FDA (Live)

    Date Cleared
    2019-06-12

    (257 days)

    Product Code
    Regulation Number
    884.5160
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Lansinoh® Signature Pro™ Double Electric Breast Pump is intended to express and collect the breastmilk of a nursing woman for the purpose of feeding collected breastmilk to a baby. The pump is intended for multiple users, and single users.

    The Lansinoh® Smartpump™M Double Electric Breast Pump is intended to express and collect the breasmilk of a nursing woman for the purpose of feeding collected breastmilk to a baby. The pump is intended for multiple users, and single users.

    Device Description

    The Lansinoh Powered Breast Pumps are intended to express the breast milk of a nursing woman. The pumping can be performed on one breast or on both breasts at the same time. The pump can be powered by 6 AA batteries or by an AC adaptor that is provided with the pump. The pumping system consists of a diaphragm-type vacuum pump which is driven by a microprocessor controlled DC electric motor. The user interface consists of a front panel keypad and LCD display. The user is able to adjust cycle mode and vacuum level based on personal comfort and preference.

    The purpose of this 510(k) is to update the indications for use statement to include use by multiple users. The labeling has been revised to clarify that the collection kits are for single users only and are not to be shared with other users.

    AI/ML Overview

    This document describes the premarket notification (510(k)) for the Lansinoh® Signature Pro™ Double Electric Breast Pump (DEBP 2.1) and Lansinoh® Smartpump™ Double Electric Breast Pump (DEPB 2.2). The purpose of this submission is to update the indications for use to include multiple users, leveraging the previously cleared Lansinoh Double Electric Breast Pump (K122474) as the predicate.

    Here's an analysis of the provided information concerning acceptance criteria and the supporting study:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't explicitly present a formal "acceptance criteria" table with pass/fail thresholds in the typical sense for a diagnostic AI device. Instead, it focuses on demonstrating that the new devices' performance characteristics are similar to or do not raise new questions of safety and effectiveness compared to the predicate device, especially considering the expanded "multiple user" indication. The performance data presented are primarily focused on hardware and safety aspects of the breast pump, not an AI or diagnostic algorithm.

    However, we can infer some criteria and reported performance from the "Device Comparison Table" and "Performance Data" sections:

    Feature/TestAcceptance Criteria (Inferred from Predicate/Standard)Reported Device Performance
    Primary Indication for UseBe intended to express and collect breastmilk for feeding to a baby. (No new questions of safety/effectiveness compared to predicate, despite "multiple user" inclusion).The Lansinoh® Signature Pro™ and Smartpump™ are intended to express and collect breastmilk of a nursing woman for the purpose of feeding collected breastmilk to a baby. The pump is intended for multiple users, and single users. (Different from predicate in that it allows multiple users, but this difference does not alter the intended use or raise different questions of safety/effectiveness).
    Suction Levels (Stimulation)Within a clinically acceptable and safe range (Predicate: 55-140 mmHg).46-140 mmHg. (Similar; slightly lower suction level but does not raise different questions of safety and effectiveness).
    Cycles per Second (Stimulation)Within a clinically acceptable and safe range (Predicate: 1.55-2.4).1.61-2.33. (Similar; slightly different cycle rates but does not raise different questions of safety and effectiveness).
    Suction Levels (Expression)Within a clinically acceptable and safe range (Predicate: 80-220 mmHg).95-280 mmHg. (Different; higher pressure levels for expression than the predicate, but this difference does not raise different questions of safety and effectiveness).
    Cycles per Second (Expression)Within a clinically acceptable and safe range (Predicate: 0.61-1.52).0.58-1.69. (Similar; slightly different cycle rate but does not raise different questions of safety and effectiveness).
    Suction SettingsSame as predicate (8).8. (Same)
    Expression ModesSame as predicate (3).3. (Same)
    Power SupplySame as predicate (6 AA batteries, AC adapter).a) 6 AA alkaline batteries b) AC Adapter. (Same)
    Pumping OptionSame as predicate (Single or Double).Single or Double. (Same)
    Back Flow ProtectionPresence and effectiveness in preventing milk from entering tubing/pump. (Yes).Yes. (Same). Demonstrated at maximum pressure/cycle settings (Expression Mode 3, Level 8) and various orientations, no milk was present in tubing, confirming diaphragm prevents backflow.
    Let Down FunctionPresence. (Yes).Yes. (Same)
    Cycling/Suction Control MechanismSame as predicate (Microprocessor).Microprocessor. (Same)
    Communication with mobile app (Smartpump only)No alteration of intended use or new questions of safety/effectiveness (N/A for predicate).Bluetooth. (Different; SmartPump has Bluetooth capability, but this does not alter the intended use or raise different questions of safety and effectiveness).
    Bench Performance (Suction Curves)Meets specifications and performs within specified working ranges of pressure and cycle speed for each mode/level.Demonstrated that the device meets its specifications and performs within the specified working ranges of pressure and cycle speed for each mode/level.
    Battery LifeMeets specified battery life (approx. 90 minutes).Approximately 90 minutes at the highest level (Expression Mode 3, Level 8).
    CleaningInstructions provided for reusable, non-sterile device.Cleaning instructions are provided in the labeling.
    Shelf-lifeNot applicable due to low degradation likelihood; expected use-life of 500 hours (per IEC 60601-1:2005/(R)2012).Demonstrated to operate within specifications for up to 500 hours of continuous operation at the maximum setting.
    BiocompatibilityPatient-contacting materials meet FDA's Food Additive criteria (21 CFR 175-179) and are equivalent to predicate (K122474).No changes to patient-contacting materials since previous clearance (K122474). All milk-contacting components tested to meet FDA's Food Additive criteria. (Leverages K122474 data).
    Electrical Safety & EMCMeets applicable test limits per ANSI/AAMI ES60601-1:2005/(R)2012 and IEC 60601-1-2:2007.Demonstrated to meet applicable test limits for electrical safety and EMC per the cited standards.
    Software Function (if applicable)Functions as intended, consistent with a minor level of concern (per FDA guidance "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices," May 11, 2005).Documentation consistent with a minor level of concern demonstrated the software functions as intended.
    Cybersecurity (if applicable)Appropriate mitigations implemented based on device risk level.Evaluated through a risk-based approach; appropriate mitigations implemented based upon the level of risk of the devices.

    2. Sample Size Used for the Test Set and the Data Provenance:

    The document describes various performance tests, but it does not specify exact sample sizes (e.g., number of units tested) for each. For example:

    • Bench Performance (Suction Curves, Battery Life, Backflow Test, Shelf-Life): "Devices were evaluated," "Devices were tested." No specific number of devices is mentioned.
    • Biocompatibility: Leveraged data from K122474, implying no new testing sample for this submission. The materials are the same.
    • Electrical Safety and EMC: "The device was tested." No specific number of devices is mentioned.
    • Software and Cybersecurity: "The software was evaluated," "Cybersecurity for the subject devices was evaluated."

    Data Provenance: The tests appear to be conducted by the manufacturer, Lansinoh Laboratories Saglik Gerecleri Tasarim San. Tic. Ltd. Sti., which is based in Turkey. The studies mentioned (Bench Performance, Battery Life, Backflow Test, Shelf-Life, Electrical Safety, EMC, Software, Cybersecurity) are retrospective relative to the submission date, as they were completed to support the 510(k). They are engineering/performance verification tests conducted in a lab environment.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts:

    This information is not applicable as the device is a breast pump, not a diagnostic device that requires expert interpretation for ground truth. The "ground truth" here is determined by engineering specifications and objective measurements (e.g., pressure, cycle rates, battery life, absence of milk in tubing).

    4. Adjudication Method for the Test Set:

    This information is not applicable as the device is a breast pump. Adjudication methods (like 2+1 or 3+1) are typically used in clinical studies for diagnostic devices where human experts disagree on interpretations (ground truth). The performance tests described are objective engineering measurements.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This information is not applicable. The device is a breast pump and does not involve AI assistance for human readers or interpretation of medical cases.

    6. If a standalone (i.e. algorithm only, without human-in-the-loop performance) was done:

    This information is not applicable. The device is a breast pump. While it has a microprocessor and software, it's not an "algorithm" in the diagnostic AI sense, and its performance is about mechanical function, not diagnostic accuracy.

    7. The Type of Ground Truth Used:

    The "ground truth" for the performance tests is based on:

    • Engineering Specifications: Defined parameters for suction levels, cycle rates, battery life, etc.
    • Objective Measurements: Using testing equipment to measure pressure, time, and physical presence of milk (for backflow).
    • Industry Standards: Compliance with standards like ANSI/AAMI ES60601-1:2005/(R)2012 for electrical safety and IEC 60601-1-2:2007 for EMC.
    • FDA Regulations: Compliance of patient-contacting materials with FDA's Food Additive criteria (21 CFR 175-179).
    • Prior Clearance Data (K122474): For biocompatibility, the ground truth established for the predicate device was leveraged.

    8. The Sample Size for the Training Set:

    This information is not applicable. This is not an AI or machine learning device requiring a training set. The software mentioned is likely embedded control software for the pump's operation, not a learning algorithm.

    9. How the Ground Truth for the Training Set Was Established:

    This information is not applicable, as there is no training set mentioned or implied for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K122474

    Validate with FDA (Live)

    Date Cleared
    2012-12-28

    (136 days)

    Product Code
    Regulation Number
    884.5160
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Powered Breast Pump is intended to express and collect the breast milk of a nursing woman for the purpose of feeding the collected milk to a baby. The Powered Breast Pump is intended for a single user.

    Device Description

    The Powered Breast Pump provides the option of pumping on one breast or on both breasts at the same time. The Powered Breast Pump can be powered by 6 AA batteries or an AC adaptor provided with the pump. The pumping system consists of a diaphragm-type vacuum pump which is driven by a microprocessor. The user interface consists of a front panel keypad and LCD display. The user is able to adjust cycle-mode and suction level. The Powered Breast Pump is capable of providing vacuum levels from 75 to 220 mmHg with 3 different cycle modes. Mothers are able to choose the most comfortable / preferred vacuum level and cyclemode.

    AI/ML Overview

    This is a device for a powered breast pump. The provided content indicates that this is a Special 510(k) submission for modifications to a previously cleared breast pump, primarily to cycle speeds and suction settings to enhance user comfort.

    Here's an analysis based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The submission does not explicitly state "acceptance criteria" for the performance tests in a quantitative manner. Instead, it states that the tests were conducted to ensure "satisfactory performance" or to "illustrate substantial equivalence." The performance data is primarily presented as a comparison to the predicate device.

    Acceptance Criterion (Implicit)Reported Device Performance
    Suction curves demonstrate substantial equivalence to predicateSuction curves provided (details not explicitly given)
    Satisfactory backflow performanceBackflow test conducted (details not explicitly given)
    Validated cleaning instructionsCleaning validation study conducted (details not explicitly given)

    Device Comparison Details (from the table provided):

    FeatureNew Device (Powered Breast Pump)Predicate Device (Affinity Double Electric Breast Pump)
    Suction Levels (stimulation)55 - 140 mmHg50 - 150 mmHg
    Cycles per Second (stimulation)1.55 - 2.41.85 (fixed)
    Suction Levels (expression)80 - 220 mmHg50 - 250 mmHg
    Cycles per Second (expression)0.61 - 1.520.51 - 1.0
    Suction Settings88
    Power Supply6 AA batteries, AC Adapter6 AA batteries, AC Adapter
    Pumping OptionSingle or DoubleSingle or Double
    Back Flow ProtectionYesYes
    Let Down FunctionYesYes
    Cycling/Suction Control MechanismMicroprocessorMicroprocessor

    2. Sample Size Used for the Test Set and Data Provenance:

    The document mentions "Performance Testing" including "Suction Curves," "Back Flow Test," and "Cleaning Validation Study." However, it does not specify the sample sizes used for these tests. There is no information regarding the country of origin of the data or whether the studies were retrospective or prospective.

    3. Number of Experts Used to Establish Ground Truth and Qualifications:

    This information is not provided in the document. The performance tests described (suction curves, backflow, cleaning validation) are technical tests of the device's functionality rather than assessments requiring expert medical interpretation or "ground truth" derived from patient data.

    4. Adjudication Method for the Test Set:

    This information is not applicable or provided. The tests mentioned are objective engineering and validation tests, not clinical studies requiring adjudication of results based on expert consensus.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    An MRMC study was not conducted as this is a breast pump, not a diagnostic imaging device typically evaluated with such studies. The document focuses on demonstrating substantial equivalence to a predicate device through technical performance characteristics and intended use.

    6. Standalone (Algorithm Only) Performance Study:

    A standalone performance study, as typically understood for an algorithm or AI without human interaction, was not done. The "device" is a physical breast pump, and its performance is evaluated through its functional capabilities (suction, cycles, backflow protection).

    7. Type of Ground Truth Used:

    The concept of "ground truth" as typically applied to diagnostic algorithms (e.g., pathology, outcomes data) is not relevant for this breast pump submission. The performance assessment is based on physical and functional measurements and validation against pre-defined engineering or safety standards (e.g., suction curves, backflow prevention, cleaning effectiveness). The "ground truth" for these types of devices is their ability to perform their intended mechanical function within acceptable parameters.

    8. Sample Size for the Training Set:

    This information is not provided and is likely not applicable in the context of this device. A breast pump does not typically involve machine learning algorithms that require a "training set" of data in the sense that an AI diagnostic tool would.

    9. How the Ground Truth for the Training Set Was Established:

    This information is not provided and is not applicable for the reasons stated in point 8.