Found 2 results

510(k) Data Aggregation

    K Number
    K201600

    Validate with FDA (Live)

    Date Cleared
    2021-12-10

    (546 days)

    Product Code
    Regulation Number
    880.2910
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The infrared thermometer, models KFT-22M, KFT-22, KFT-24, KFT-25, KFT-26, KFT-27, KFT-28 is designed to measuring human body temperature through ear or forehead in the healthcare environments or home use and is non-sterile. The thermometer is a reusable device intended for people of all ages except neonates/newborns.

    Device Description

    The KFT series infrared thermometer measures the body temperature through receiving the infrared energy radiation from the surface of human body. It is capacified with forehead mode and ear temperature mode, directly shown on the LCD screen.

    • The device is widely used for home healthcare and medical environment.
    • . The product is mainly comprised of infrared temperature sensors, signal receiving processor, buttons, buzzer, LCD display, battery, etc.
    • . It focuses the infrared from the human's forehead or ear by the Fresnel lens.
    AI/ML Overview

    The provided document is a 510(k) summary for an infrared thermometer, focusing on its substantial equivalence to a predicate device. It contains information about performance data, including clinical and non-clinical testing. However, it does not describe an AI/ML-enabled device or a study involving human readers or expert consensus for ground truth establishment. Therefore, I cannot provide details on acceptance criteria and a study proving device performance in the context of an AI/ML device per your request.

    The document discusses the following:

    • Device: Kangfu Medical Equipment Factory Infrared thermometer, models KFT-22, KFT-23, KFT-24, KFT-25, KFT-26, KFT-27, KFT-28.
    • Predicate Device: Shenzhen Brav Electronic Technologies Co., Ltd Infrared Thermometer, Model EFT-165 (K171214).
    • Indication for Use: Measuring human body temperature through ear or forehead in healthcare environments or home use, intended for people of all ages except neonates/newborns.
    • Performance Data: Non-clinical (biocompatibility, electrical safety/EMC, bench testing, software V&V) and clinical data.

    Regarding the clinical study for this specific device (an infrared thermometer):

    The clinical study mentioned is for the accuracy validation of the thermometer itself, not for an AI/ML component or human reader performance.

    • Acceptance Criteria for Infrared Thermometer Accuracy (Implied from "Accuracy for body temperature measurement" table):

      • 32.0°C~34.9°C: ±0.3°C /0.5°F
      • 35.0°C~42.0°C: ±0.2°C /0.4°F
      • 42.1°C~42.9°C: ±0.3°C / 0.5°F
    • Reported Device Performance (Implicit from "Comparison to predicate device" and "Performance Data" sections): The document states that the "system complies with the ... ISO 80601-2-56: 2009 ... standards for performance effectiveness," and the "subject devices were found to have a performance that is similar to the predicate device," which shares the same accuracy specifications. This implies the subject device meets the listed accuracy criteria.

    • Sample Size for the test set (Clinical Data): 116 patients. The data seems to be prospective as it's a "clinical accuracy validation." Country of origin is not explicitly stated but implied to be China (Kangfu Medical Equipment Factory is in China).

    • Number of experts used to establish the ground truth for the test set and qualifications: Not applicable/not specified. For thermometer accuracy, ground truth is typically established using a highly accurate reference thermometer, not expert human readers.

    • Adjudication method for the test set: Not applicable/not specified, as it's not an image reading study.

    • Multi-reader multi-case (MRMC) comparative effectiveness study: Not applicable. This study focuses on the inherent accuracy of a temperature measurement device, not on AI assistance to human readers.

    • Standalone (algorithm only) performance: Performance data is for the device itself, which includes its internal algorithm for temperature measurement. This can be considered the standalone performance of the thermometer.

    • Type of ground truth used: For clinical accuracy validation of a thermometer, the ground truth is typically established by simultaneous measurement using a highly accurate, calibrated reference thermometer (e.g., rectal thermometer for core body temperature, or a blackbody calibrator for infrared devices), sometimes referred to as a "master thermometer" or "reference method." This is an objective measurement, not expert consensus, pathology, or outcomes data in this context.

    • Sample size for the training set: Not applicable/not specified. Infrared thermometers use established physical principles and calibration, not machine learning models that require training data in the AI sense.

    • How the ground truth for the training set was established: Not applicable, as it's not an AI/ML device requiring a training set with established ground truth labels in the typical sense.

    In summary, the provided text describes a traditional medical device (infrared thermometer), not an AI/ML-enabled device. Therefore, many of the requested details pertaining to AI/ML device studies (like MRMC studies, expert ground truth, training sets) are not found in this document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K182652

    Validate with FDA (Live)

    Date Cleared
    2018-11-20

    (57 days)

    Product Code
    Regulation Number
    880.2910
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    COCET Digital Thermometer measures body temperature. This digital thermometer is used in the armpit (axillary) to take temperature measuring results. The device is for adult use.

    Device Description

    COCET Digital Thermometer (Models: KFT-01/KFT-02/KFT-03/KFT-04/ KFT-05/KFT-06/ KFT-07/KFT-08/ KFT-09/KFT-10) is a hand-held electronic thermometer, it comprised of an electronic thermo resistor (thermistor) sensor connected to printed circuit board(PCB) with liquid crystal display(LCD) user readout. The outer material of thermometer is made of ABS and the measuring tip is stainless steel. The thermometer using a temperature sensor to output an electrical signal, and then converting the current signal into a liquid crystal digital display temperature. The thermometers are powered by LR41 battery. The battery can be replaced.

    AI/ML Overview

    The provided text describes the 510(k) summary for the COCET Digital Thermometer. It outlines a comparison to a predicate device and details non-clinical tests performed. However, it does not contain the specific information required to answer all parts of your request regarding acceptance criteria and a study proving the device meets those criteria, particularly in the context of an AI/human-in-the-loop study.

    The document is for a digital thermometer, which is a fairly straightforward medical device. The "study" described is primarily bench testing to demonstrate performance according to recognized standards, rather than a clinical trial involving a large number of human subjects and complex ground truth establishment as would be seen for an AI diagnostic device.

    Here's what can be extracted and what is missing based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (from text, implied by standards adherence)Reported Device Performance (from text)
    Accuracy: +/- 0.1 °C (±0.2°F) from 35.5 °C to 42 °C (95.9°F-107.6°F)+/- 0.1 °C (±0.2°F) from 35.5 °C to 42 °C (95.9°F-107.6°F)
    Accuracy: +/- 0.2 °C (± 0.4 °F) in the rest measuring range+/- 0.2 °C (± 0.4 °F) in the rest measuring range
    Biocompatibility: Compliance with ISO 10993-1, ISO 10993-5, ISO 10993-10Device passed cytotoxicity, skin irritation, and sensitization testing.
    Safety and EMC: Compliance with IEC 60601-1 and IEC 60601-1-2Performed in accordance with IEC 60601-1:2005 and IEC 60601-1-2:2014.
    Clinical Electronic Thermometer Standards: Compliance with specific standardsComplies with ISO 80601-2-56:2017.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    • Sample Size: Not explicitly stated for specific tests. The tests performed are primarily bench testing (e.g., testing multiple physical units of the thermometer for accuracy, biocompatibility, and electrical safety). It's typically a small number of physical samples (e.g., n=3 or n=5) from a production lot for regulatory testing, not a "data set" in the AI sense.
    • Data Provenance: Not applicable in the sense of patient data. The tests are laboratory bench tests.
    • Retrospective or Prospective: Not applicable as it's not a clinical study on patients. These were tests conducted on the physical device itself.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not Applicable. For a digital thermometer, "ground truth" for temperature is established using calibrated reference thermometers and controlled environments (e.g., water baths). There are no human experts "adjudicating" temperature readings in the way a diagnostic image might be.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    • Not Applicable. See point 3.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No. This device is a standalone digital thermometer, not an AI-assisted diagnostic tool. Therefore, an MRMC study is not relevant or performed for this type of device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Yes, in essence. The performance data presented (accuracy, biocompatibility, safety) are inherent properties of the device itself, functioning independently of human interpretation in a diagnostic context. However, it's not an "algorithm" in the typical software sense, but the physical device's performance.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • The ground truth for temperature accuracy is established against calibrated reference standards, typically highly accurate laboratory thermometers traceable to national/international standards.
    • For biocompatibility, the ground truth is the absence of adverse biological reactions in established in-vitro and in-vivo tests according to ISO standards.
    • For safety and EMC, the ground truth is compliance with the specified electrical and electromagnetic compatibility limits/requirements defined in the IEC standards.

    8. The sample size for the training set

    • Not Applicable. This is a physical digital thermometer, not an AI model that requires a "training set."

    9. How the ground truth for the training set was established

    • Not Applicable. See point 8.

    In summary: The provided document is for a general medical device (digital thermometer) and describes its compliance with established performance, safety, and biocompatibility standards through bench testing. It does not describe an AI/ML-driven device or a clinical study that would involve expert readers, complex ground truth establishment, or MRMC studies. The "study" proving acceptance criteria is a series of non-clinical, laboratory-based tests comparing the device's performance to recognized international standards and its predicate device.