510(k) Data Aggregation

    K Number
    K953649

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1996-06-21

    (322 days)

    Product Code
    Regulation Number
    870.2360
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    LMI Medical considers their ElectroBlue® ECG Electrodes to be safe and effective for use with electrocardiograph monitors which are used for recording the electric activity of the myocardium to detect abnormal transmission of small electrical impulses through the conductive tissue surrounding the heart, when used according to the intended procedures as related in the instructions accompanying the product. The function of the LMI Medical ElectroBlue® Electrodes is to detect and transmit these small electrical impulses to the electrocardiograph monitors where they are processed and plotted.

    Device Description

    The products are all the same in their construction using medical grade foam, polypropylene cloth, clear vinyl, medical grade adhesive, silver coated carbon loaded ABS snaps, and a conductive hydrogel with a polyurethane cover release liner.

    AI/ML Overview

    This document is a 510(k) Premarket Notification for ElectroBlue® ECG Electrodes. It describes the safety and effectiveness of the device by comparing it to an established standard and existing market products.

    Here's an analysis of the provided text based on your request:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (from ANSI/AAMI EC12-1991)Reported Device Performance
    Electrical Performance:Met all listed requirements
    - AC impedance
    - DC offset Voltage
    - Combined offset instability and internal noise
    - Defibrillation overload recovery
    - Bias current tolerance
    Biological Response:Met all listed requirements
    - Cytotoxicity
    - Sensitization
    - Irritation
    - Intracutaneous reactivity
    Performance, Packaging and Shelf life:Met all listed requirements

    2. Sample size used for the test set and the data provenance

    The document does not explicitly state the sample size used for the test set (i.e., the number of ElectroBlue® ECG Electrodes tested). It also does not provide details on the data provenance (e.g., country of origin, retrospective or prospective nature of the testing). It only states that the "results are on file at LMI Medical."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable and not provided. The "ground truth" for this device's performance is established by meeting the objective, standardized physical and biological requirements outlined in the ANSI/AAMI EC12-1991 standard, not by expert consensus on interpretations of data.

    4. Adjudication method for the test set

    Not applicable. As mentioned above, the performance is measured against objective standards, not subject to expert adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This type of study is typically used for diagnostic devices where human interpretation is involved. For ECG electrodes, the effectiveness is determined by their adherence to electrical and biological performance standards.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This question is not applicable to an ECG electrode. An ECG electrode is a passive device; it does not have an "algorithm" or standalone "performance" in the sense of AI or software. Its performance is measured by its material properties and ability to conduct electrical signals and be biologically safe.

    7. The type of ground truth used

    The "ground truth" for demonstrating the device's safety and effectiveness is compliance with the established technical and biological performance requirements of the ANSI/AAMI standard EC12-1991. This standard specifies objective, measurable criteria for various aspects like electrical performance, biological response, and packaging/shelf life.

    8. The sample size for the training set

    This concept is not applicable to the evaluation of this device. A "training set" is relevant for machine learning algorithms. The ElectroBlue® ECG Electrodes are physical medical devices, and their performance is proven through direct testing against a standard.

    9. How the ground truth for the training set was established

    Not applicable. As explained above, there is no "training set" in the context of this device's evaluation.