Search Filters

Search Results

Found 22 results

510(k) Data Aggregation

    K Number
    K240506

    Validate with FDA (Live)

    Date Cleared
    2024-10-22

    (244 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The KARL STORZ Holders are manually operated surgical devices intended to hold endoscopes (such as arthroscopes, etc) and accessories during general and neurologic diagnostic and therapeutic procedures.

    Device Description

    The KARL STORZ Holder is a manually operated surgical instrument intended to hold scopes (such as endoscopes, arthroscopes, etc) in a desired position during diagnostic, therapeutic, and surgical procedures (including neurologic procedures). The holder is manufactured from Stainless Steel and functions by the user manually tightening or loosening a knob, which opens and closes around a surgical scope. The device is intended to attach to the several KARL STORZ holding systems, which are a Class 1, table-mounted, mechanical and electromechanical holding arms that have many general surgical functions in addition to scope holding; it can also be used for instrument holding, retraction, and positioning.

    AI/ML Overview

    The provided text is a 510(k) summary for the KARL STORZ Holders, which are manually operated surgical devices intended to hold endoscopes and accessories. This document primarily focuses on demonstrating substantial equivalence to a predicate device rather than providing extensive details about a clinical study with acceptance criteria for an AI/CAD-like device.

    Therefore, the requested information regarding acceptance criteria, study design, sample sizes, expert involvement, ground truth establishment, and MRMC studies for an AI/CAD device is not available in the provided text. The document describes a traditional medical device (a mechanical holder) and its non-clinical performance testing.

    However, based on the information that is available in the document, here's a summary of the non-clinical performance data and the demonstration of substantial equivalence:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document does not detail specific quantitative acceptance criteria or reported values in a table format for a clinical outcome beyond demonstrating functional equivalence. The performance data is described qualitatively as meeting the needs for substantial equivalence.

    Performance CategoryAcceptance Criteria (Implied for Substantial Equivalence)Reported Device Performance
    Holding PerformanceSufficient to securely hold intended devices (scopes, accessories) during procedures and maintain position.Testing was performed (Non-Clinical Performance Data: "Holding Performance testing"). Implied to be acceptable for intended use. Max loading weight of 3kg reported.
    Direction of View/Image StabilityMaintain stable direction of view/image during use.Testing was performed (Non-Clinical Performance Data: "Direction of view/Image stability testing"). Implied to be acceptable for intended use, as it contributes to substantial equivalence.
    Material of ConstructionBiocompatible and durable for surgical use.Stainless Steel, same as predicate device.
    Mode of OperationManual, user-friendly.Manual, same as predicate device.
    Cleaning & ReprocessingSterilizable and cleanable without degradation.Manual cleaning and Steam Sterilization, same as predicate device.

    2. Sample Size Used for the Test Set and Data Provenance:

    • The document describes non-clinical performance testing rather than a clinical test set involving human subjects or AI algorithm evaluation.
    • The exact sample size for the "Holding Performance testing" and "Direction of view/Image stability testing" is not specified.
    • Data Provenance: The testing was performed on the "subject device" (KARL STORZ Holders). The country of origin of the data is not explicitly stated beyond the submitter being KARL STORZ SE & Co. KG in Tuttlingen, Germany. The data would be considered prospective as it's testing of the new device for regulatory submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:

    • This information is not applicable as the document describes non-clinical performance testing of a mechanical device, not the evaluation of an AI algorithm requiring expert ground truth for image interpretation.

    4. Adjudication Method for the Test Set:

    • This is not applicable for the same reason as point 3.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

    • No, an MRMC comparative effectiveness study was not done. This type of study is relevant for evaluating the impact of AI on human reader performance, which is not the subject of this 510(k) submission for a mechanical endoscope holder.

    6. If a Standalone (algorithm only without human-in-the-loop performance) was done:

    • No, a standalone algorithm performance study was not done. This device is a mechanical surgical instrument, not an AI algorithm.

    7. The Type of Ground Truth Used:

    • The concept of "ground truth" as typically used for AI/CAD devices (e.g., expert consensus, pathology, outcome data) is not applicable. For this mechanical device, performance is evaluated against engineering specifications and functional requirements (e.g., ability to hold weight, maintain stability).

    8. The Sample Size for the Training Set:

    • This is not applicable. The device is a mechanical instrument, not an AI algorithm that requires a training set.

    9. How the Ground Truth for the Training Set was Established:

    • This is not applicable for the same reason as point 8.

    In summary, the provided FDA 510(k) summary (K240506) pertains to a mechanical endoscope holder, not an AI/CAD-enabled medical device. Therefore, the detailed requirements for acceptance criteria and study designs related to AI performance are not present in this document. The purpose of this 510(k) is to demonstrate substantial equivalence to a predicate mechanical device through non-clinical performance testing.

    Ask a Question

    Ask a specific question about this device

    K Number
    K223885

    Validate with FDA (Live)

    Date Cleared
    2023-09-14

    (261 days)

    Product Code
    Regulation Number
    884.1690
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The HOPKINS Telescopes when used with sheaths are intended to be used to permit viewing of the cervical and uterine cavity for the purpose of performing diagnostic and surgical procedures.

    Device Description

    The HOPKINS Telescopes are rigid telescopes that utilize the rod lens technology. At the distal end of the telescope's shaft is the lens and the other end of the shaft is attached to the eyepiece. Throughout the central lumen of the HOPKINS Telescopes, optical glass rods are used to transmit and magnify the image received from the lens. The HOPKINS Telescopes are available with 0°, 12° and 30° direction of view, 2mm, 2.9mm, 4mm diameter and 26cm, 30cm, 36cm working lengths.

    AI/ML Overview

    The provided text is a 510(k) Summary for the HOPKINS Telescopes, a medical device. This document focuses on demonstrating substantial equivalence to a predicate device rather than presenting a study of its own performance against pre-defined acceptance criteria for a novel AI/software component.

    Therefore, the provided document does not contain the information required to answer your request regarding acceptance criteria and a study proving the device meets those criteria, specifically concerning AI/software performance.

    The document states:

    • "Clinical studies were not required to demonstrate substantial equivalence to the predicate device." (Page 9)
    • "The conclusions drawn from the nonclinical test demonstrate that the subject device is as safe and effective as the predicate device to support a substantial equivalence determination." (Page 9)

    This means the submission relies on bench testing and comparison of technological characteristics to a legally marketed predicate device, not on a clinical performance study with defined acceptance criteria and human readers (or AI algorithms) in the way your prompt describes.

    To directly answer your questions based only on the provided text, the answer is that this information is not present for the HOPKINS Telescopes device.

    If this were an AI/software device and the information was present, here's how I would answer each point:

    1. Table of acceptance criteria and reported device performance: This would be a table detailing metrics like sensitivity, specificity, AUC, or agreement rates, with specific target thresholds (acceptance criteria) and the actual performance achieved in the study.
    2. Sample size and data provenance: This would specify the number of cases/patients in the test set (distinct from training data), whether the data was collected retrospectively or prospectively, and from which countries or institutions.
    3. Number of experts and qualifications: This would state how many experts (e.g., radiologists, pathologists) were involved in establishing ground truth and their relevant experience (e.g., board-certified, years of experience, subspecialty).
    4. Adjudication method: This would describe how discordant readings or interpretations among experts were resolved (e.g., 2+1 means two initial readers, with a third if they disagree; 3+1 means three initial readers, with a fourth for adjudication if necessary).
    5. Multi-reader multi-case (MRMC) comparative effectiveness study: If conducted, this would provide details on how the AI system impacted human reader performance, including statistical metrics like the effect size (e.g., mean increase in AUC, sensitivity, or specificity when AI assistance was used compared to no AI assistance).
    6. Standalone performance: This would report the performance of the algorithm itself, without any human interaction, against the ground truth.
    7. Type of ground truth: This would specify the reference standard used (e.g., expert consensus, pathology results, follow-up clinical outcomes, surgical findings).
    8. Sample size for training set: This would state the number of unique cases/patients used to train the AI model.
    9. Ground truth for training set: This would explain the methodology used to label or establish the ground truth for the data utilized during the training phase of the algorithm.
    Ask a Question

    Ask a specific question about this device

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The KARL STORZ suction and irrigation system (consisting of handle and suction/irrigation tubes, cannulae or needles) is intended for use by qualified surgeons to provide suction and/or irrigation function to help flush or remove blood, fluids, and tissue debris from the operation site during general, gynecologic, urologic, and thoracic laparoscopic surgical procedures.

    Device Description

    The devices in the KARL STORZ Suction and Irrigation System are manually operated, reusable surgical devices consisting of handle, suction/irrigation tubes, cannulae or needles. The handle is intended to be used in combination with suction/irrigation cannulae and tubing and controls the suction and irrigation. Enabled by the handle with stop cock valve, the regulation of the irrigated quantity of fluids via the tubes, or rather the quantity of aspirated fluid (e.g., blood, fluids, and tissue debris) into or out of the patient is performed. Cannula and needle: Enabled by the pointed distal end, both puncture and injection/extraction cannulae and needles, perform puncturing of target organs or tissues and injecting fluids into operating site i.e. extracting blood or fluids from them. The prerequisite for suction is either a vacuum powered suction wall apparatus or a pump, which serve as impulse for the suction. The proximal end of the tube set is connected to the vacuum or pump. Either a pump is required as impulse for irrigation or the natural force of gravity, i.e., by positioning the flush bag at the appropriate height.

    AI/ML Overview

    The provided document is an FDA 510(k) Premarket Notification for medical devices, specifically KARL STORZ suction and irrigation systems. This type of submission focuses on demonstrating substantial equivalence to legally marketed predicate devices, primarily through non-clinical performance data and does not typically include studies involving artificial intelligence (AI) or machine learning (ML) components, nor does it present "acceptance criteria" and "device performance" in the context of an AI/ML model's diagnostic or predictive capabilities.

    Therefore, the requested information regarding AI/ML-related acceptance criteria and study details (such as sample sizes for test and training sets, expert adjudication methods, MRMC studies, standalone performance, ground truth establishment, etc.) cannot be extracted from this document, as it is not relevant to the type of device and submission described.

    The document focuses on:

    • Device Description: Manually operated, reusable surgical devices (handle, suction/irrigation tubes, cannulae, needles).
    • Intended Use: Suction and/or irrigation to flush or remove blood, fluids, and tissue debris during laparoscopic surgical procedures.
    • Basis for Substantial Equivalence: Comparison to predicate KARL STORZ devices (K945059 and K951190) based on technological characteristics and non-clinical performance data.
    • Non-Clinical Performance Data Provided:
      • Biocompatibility testing: Compliance with ISO 10993.
      • Reprocessing Validation: Compliance with standards like ANSI/AAMI/ISO 14937, 11138-3, 11607-1, 11737-1, 17664, 17665-1 regarding cleaning and sterilization.
      • Bench Top Performance Testing: Leak Test, Flow Incoming Test, Flow Outgoing Test.

    The document explicitly states: "Clinical testing was not required to demonstrate the substantial equivalence to the predicate device. Non-clinical bench testing was sufficient to assess safety and effectiveness and to establish the substantial equivalence of the modifications."

    In summary, this document does not contain information about an AI/ML device or its associated performance studies.

    Ask a Question

    Ask a specific question about this device

    K Number
    K221893

    Validate with FDA (Live)

    Date Cleared
    2023-03-17

    (261 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The KARL STORZ bipolar resectoscopes with HF cable are intended for use by qualified surgeons for cutting, ablation, vaporization, and/or coagulation of tissue during various endoscopic urological procedures.

    The KARL STORZ bipolar resectoscopes with HF cable are intended for use by qualified surgeons for cutting, ablation, vaporization, and/or coagulation of tissue during various endoscopic electrosurgical procedures.

    The KARL STORZ bipolar resectoscopes with HF cable are intended for use by qualified surgeons for cutting, ablation, vaporization, and/or coagulation of tissue during various endoscopic urological and hysteroscopic electrosurgical procedures.

    Indications for Use in UROLOGY: The KARL STORZ bipolar resectoscopes with HF cable are intended for use by qualified surgeons for cutting, ablation, vaporization, and/or coagulation of tissue during various endoscopic urological procedures.

    Indications for Use in GYNECOLOGY: The KARL STORZ bipolar resectoscopes with HF cable are intended for use by qualified surgeons for cutting, ablation, vaporization, and/or coagulation of tissue during various endoscopic hysteroscopic electrosurgical procedures

    Indications for Use in UROLOGY and GYNECOLOGY: The KARL STORZ bipolar resectoscopes with HF cable are intended for use by qualified surgeons for cutting, ablation, vaporization, and/or coagulation of tissue during various endoscopic urological and hysteroscopic electrosurgical procedures.

    Device Description

    Resectoscopes are used for endoscopically controlled ablation of tissue. They are used for examination, diagnosis and/or therapy in conjunction with endoscopic accessories in medical disciplines such as Urology and Gynecology. A resectoscope is a combined cystoscope (urology) or hysteroscope (gynecology) and electrosurgical instruments and consists of a sheath, obturator, working element and HF electrode.

    AI/ML Overview

    This document describes the KARL STORZ Bipolar Resectoscopes with HF cable (K221893). It is important to note that this device is not an AI/ML powered device, therefore, many of the typical acceptance criteria and study details related to AI performance, such as sample sizes for test and training sets, expert qualifications, and adjudication methods, are not applicable.

    Here's an analysis of the provided information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Since this is a traditional medical device and not an AI/ML powered device, the "acceptance criteria" are based on compliance with recognized standards and design specifications, rather than quantitative performance metrics like sensitivity or specificity.

    Acceptance Criterion (Non-Clinical)Reported Device Performance
    Electrical Safety and EMCComplies with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-18
    BiocompatibilityComplies with ISO 10993 (specific parts not detailed, but generally covers biocompatibility)
    Performance Testing
    - Attachment/Detachment force of the electrodeVerified (Met design specifications)
    - System Interlocking TestVerified (Met design specifications)
    - Flow Test (comparative)Verified (Met design specifications)
    - Bending Force TestVerified (Met design specifications)
    - HF Tissue and HF Durability TestVerified (Met design specifications)
    Design SpecificationsAll design specifications were met through additional bench testing. (Specific quantitative specifications are not provided in this summary, but are internally verified by the manufacturer).
    Substantial Equivalence to Predicate DeviceThe conclusions drawn from the nonclinical tests demonstrate that the subject device performs as well as or better than devices currently marketed for the same intended use. (This is the overarching acceptance criterion for 510(k) clearance).

    2. Sample Size Used for the Test Set and Data Provenance

    Not applicable for a non-AI/ML medical device. Testing was primarily bench testing to verify engineering specifications and compliance with standards. There is no "test set" in the context of image or data analysis.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    Not applicable. Ground truth, in the context of AI, refers to expert-labeled data. For this device, "ground truth" would be established by engineering specifications and objective measurements against those specifications.

    4. Adjudication Method

    Not applicable. This is relevant for studies involving human interpretation or labeling of data, which is not the case here.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No. This a traditional medical device, not an AI-assisted device, therefore, no such study was conducted to evaluate human reader improvement with AI.

    6. Standalone Performance Study

    Yes, in a sense. The non-clinical performance data (electrical safety, biocompatibility, and various performance tests) represent the "standalone" performance of the device without human interaction being a variable in the performance assessment itself (though human surgeons use the device). The tests verify the device's inherent functional characteristics and safety.

    7. Type of Ground Truth Used

    For non-clinical testing of a traditional medical device, the "ground truth" used would be:

    • Engineering Specifications: Pre-defined technical requirements for the device's dimensions, material properties, electrical characteristics, mechanical strength, and functional performance.
    • Recognized Standards: International (e.g., IEC, ISO) and national standards that set minimum safety and performance requirements for medical devices of this type.
    • Comparative Performance to Predicate Device: The performance of the legally marketed predicate device (K210651), which serves as a benchmark for substantial equivalence.

    8. Sample Size for the Training Set

    Not applicable. This device does not involve machine learning; therefore, there is no "training set."

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As there is no training set for a machine learning model, establishing ground truth for such a set is irrelevant.

    Ask a Question

    Ask a specific question about this device

    K Number
    K221004

    Validate with FDA (Live)

    Date Cleared
    2022-12-20

    (259 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The HOPKINS Telescopes are intended to provide visualization during laparoscopy and general surgery in adults and pediatrics.

    Device Description

    The HOPKINS Telescopes are rigid telescopes that utilize the rod lens technology. At the distal end of the telescope's shaft is the lens and the other end of the shaft is attached to the eyepiece. Throughout the central lumen of the HOPKINS Telescopes, optical glass rods are used to transmit and magnify the image received from the lens. The HOPKINS Telescopes are available with 0°, 25° and 45° direction of view, 5mm diameter and 29cm working lengths.

    AI/ML Overview

    The provided text is a 510(k) summary for the KARL STORZ HOPKINS Telescopes. This document is a premarket notification for a medical device to demonstrate substantial equivalence to a legally marketed predicate device. As such, it focuses on comparing the new device to an existing one based on technical characteristics, rather than establishing acceptance criteria and proving performance through a clinical study for an AI/ML-driven device.

    The document discusses the physical and optical characteristics of the endoscope, reprocessing methods, and non-clinical performance data (e.g., adherence to ISO standards, biocompatibility, thermal safety, reprocessing validation). It explicitly states: "There are no performance standards or special controls developed under Section 514 of the FD&C Act for endoscopes." This indicates that the regulatory pathway for this device does not require the kind of performance validation typically seen for AI/ML devices, which would involve acceptance criteria for metrics like sensitivity, specificity, or AUC, and detailed clinical study designs.

    Therefore, the document does not contain the information required to answer the prompt questions regarding acceptance criteria for an AI/ML device. The HOPKINS Telescopes are a traditional optical medical device, not an AI/ML product.

    To reiterate, the provided text does not contain any information about:

    1. Acceptance criteria or reported device performance for AI/ML metrics.
    2. Sample sizes for test sets or data provenance (as it's not an AI/ML study).
    3. Number and qualifications of experts for ground truth establishment.
    4. Adjudication methods.
    5. MRMC comparative effectiveness studies or effect sizes.
    6. Standalone algorithm performance studies.
    7. Type of ground truth (as it's not a diagnostic AI/ML device).
    8. Sample size for training sets (no AI/ML training is mentioned).
    9. How ground truth for a training set was established.
    Ask a Question

    Ask a specific question about this device

    K Number
    K222504

    Validate with FDA (Live)

    Device Name
    H1
    Date Cleared
    2022-11-10

    (84 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The KARL STORZ H1 camera head, in combination with an appropriately indicated camera control unit (CCU), light source, and monitor, and with an appropriately indicated endoscope, or microscope, is used for real-time visualization in diagnostic and surgical procedures.

    Device Description

    The camera head consists of an anodized aluminum, stainless steel enclosure containing a CMOS (Complementary metal-oxide-semiconductor) NTSC image sensor that converts light into electrons, the transistors in each pixel then amplify and move the charge using the more traditional wires forming conventional output signals. The camera head is intended to be attached to the proximal end of the endoscopes and is connected via cable to the compatible CCU for power and operational functions. The camera head is designed to be compatible for use with all standard KARL STORZ Endoscopes, Fiberscopes and Microscope for endoscopic observation in general endoscopic procedures.

    AI/ML Overview

    The provided text describes a 510(k) submission for the KARL STORZ H1 Camera Head, which is an endoscope camera. The document focuses on demonstrating substantial equivalence to a predicate device through non-clinical performance testing.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based solely on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document lists several performance tests conducted for the subject device to ensure it meets its design specifications and is substantially equivalent to the predicate device. However, it does not explicitly state numerical acceptance criteria for each test or a direct comparison of the subject device's performance against those criteria in a table format. It generally states that the device "met all its design specification" and "performs as well as the predicate device."

    Implied Acceptance Criteria Categories (from "Performance Testing" section):

    • Color Reproduction and Color Contrast Enhancement
    • Illumination
    • Detection Uniformity
    • Depth of field
    • Spatial Resolution
    • Dynamic Range & Detection Linearity
    • Distortion
    • Latency
    • Signal-to-Noise Ratio (SNR) & Sensitivity
    • Field of View

    Reported Device Performance:

    The document states, "The bench testing performed verified and validated that the H1 Camera Head has met all its design specification and is substantially equivalent to its predicate devices." It also concludes, "The conclusions drawn from the nonclinical tests demonstrate that the subject devices, the H1 Camera Head performs as well as the predicate device."

    Without specific numerical acceptance criteria and performance data, a detailed table cannot be created.

    2. Sample size used for the test set and the data provenance

    The document explicitly states: "Clinical testing was not required to demonstrate the substantial equivalence to the predicate devices. Non-clinical bench testing was sufficient to establish the substantial equivalence of the modifications."

    Therefore, there was no clinical test set in terms of human subjects or medical images. The testing was entirely non-clinical bench testing. The sample size for this non-clinical testing is not specified. The data provenance is also not applicable in the sense of country of origin for retrospective or prospective data, as it was bench testing.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    As no clinical testing was performed and the ground truth was established through non-clinical bench testing against design specifications, no human experts were used to establish ground truth in the traditional sense of medical image interpretation.

    4. Adjudication method for the test set

    Since no clinical test set involving human experts was used for ground truth establishment, no adjudication method (like 2+1 or 3+1) was employed.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC study was done. The device is a camera head, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is not directly applicable. The device is a camera head, not an algorithm. The "standalone" performance was essentially the non-clinical bench testing of the device itself (hardware performance), which was done without human-in-the-loop in a clinical diagnostic context.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth for the non-clinical testing was based on design specifications and objective physical measurements during bench testing, ensuring the camera met its intended performance characteristics (e.g., spatial resolution measurements, color accuracy, brightness control functionality). This is not expert consensus, pathology, or outcomes data.

    8. The sample size for the training set

    This question is not applicable as the H1 Camera Head is a hardware device and not an AI/machine learning algorithm requiring a training set.

    9. How the ground truth for the training set was established

    This question is not applicable as the H1 Camera Head is a hardware device and not an AI/machine learning algorithm requiring a training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K213194

    Validate with FDA (Live)

    Date Cleared
    2022-06-23

    (267 days)

    Product Code
    Regulation Number
    874.4720
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The HD Mediastinoscope is intended to aid the surgeon in viewing the mediastinum and facilitate the introduction and removal of surgical instruments during ENT endoscopic surgical procedures.

    Device Description

    The HD Mediastinoscope is a reusable videoendoscope intended to be used in conjunction with the X-link module of the Image1 S Camera Control Unit. The HD Mediastinoscope is comprised of four main components: CMOS sensor at the distal end of the endoscope, an oval insertion portion (spatula), handle, and internal LED light source. The spatula can accommodate up to 2 x 5 mm instruments.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the HD Mediastinoscope:

    Due to the nature of the device (a physical surgical instrument rather than an AI/software device), the provided document does not contain information typically found in acceptance criteria and studies for AI/ML devices, such as sensitivity, specificity, or AUC.

    Based on the provided text, the "acceptance criteria" appear to be related to the device's technical specifications and safety/effectiveness in comparison to predicate devices, rather than specific performance metrics like those for an AI algorithm.

    Acceptance Criteria and Reported Device Performance

    The "acceptance criteria" are implied through the comparison with existing predicate and reference devices and compliance with recognized standards. The "performance" is demonstrated through non-clinical testing, particularly a cadaver test study.

    Acceptance Criteria (Implied)Reported Device Performance
    Substantial equivalence to predicate device (K954910)Demonstrated through technical characteristics, intended use, indications for use.
    Compliance with electrical safety and EMC standardsANSI/AAMI ES:60601-1:2005+A1:2012, IEC 60601-1-2:2014, IEC 62471:2006
    Compliance with biocompatibility standardsISO 10993-10:2010, ISO 10993-11:2006/(R) 2010
    Compliance with reprocessing validation standardsAAMI TIR 12:2010, AAMI TIR 30:2011, AAMI/ANSI/ISO 11737-1:2006/(R)2011, ASTM E1837-96:2014
    Change in spatula design does not raise new questions of safety and effectivenessDemonstrated through a cadaver test study.

    Details of the Study (Non-AI/ML Device)

    Since this is a traditional medical device (mediastinoscope) and not an AI/ML device, many of the requested points related to AI/ML studies (sample size for test set, data provenance, number of experts for ground truth, adjudication method, MRMC study, standalone performance, training set size, ground truth for training set) are not applicable or not specified in this 510(k) summary.

    However, based on the information provided:

    1. Sample size used for the test set and the data provenance:

      • The primary non-clinical performance test mentioned is a cadaver test study.
      • The sample size (number of cadavers or tests performed) is not specified in the document.
      • The data provenance (country of origin, retrospective/prospective) is not specified.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • The document mentions "a cadaver test study was conducted to demonstrate that the change in spatula design...". It does not specify the use of "experts" to establish a "ground truth" in the way an AI/ML study would (e.g., radiologists labeling images). Instead, the study likely involved technical assessment and observation by qualified personnel (e.g., surgeons, engineers).
      • Not explicitly stated in terms of "ground truth experts" for an AI/ML context.
    3. Adjudication method for the test set:

      • Not applicable / Not specified for this type of device and study.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No, an MRMC study was not conducted. This is not an AI-assisted device.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Not applicable. This is a physical surgical instrument, not an algorithm.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

      • For the cadaver study, the "ground truth" would likely be the direct observation and technical assessment of the device's performance in a cadaveric model (e.g., ease of introduction, visualization, ability to facilitate instrument use, assessment of tissue damage, etc.) by the study investigators. This is not a "ground truth" in the sense of a diagnostic label for an AI.
    7. The sample size for the training set:

      • Not applicable. This device does not use a training set as it is not an AI/ML product.
    8. How the ground truth for the training set was established:

      • Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K221174

    Validate with FDA (Live)

    Device Name
    Telepack +
    Date Cleared
    2022-05-13

    (18 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TELE PACK + is an all-in-one Imaging System, which comprises a light source for illumination, Camera Control Unit (CCU) for use with compatible camera heads or video endoscopes for image processing, as well as a monitor for image display, intended for the visualization of endoscopic and microscopic procedures.

    Device Description

    The Telepack + is a portable and compact all-in-one imaging system that includes a 18.5 inch screen display, a camera control unit and internal LED light source, that is intended to be connected to a compatible device (camera head or videoendoscope) for the purpose of visualization and documentation of endoscopic and microscopic procedures as well as stroboscopy. The Telepack + includes a LED illumination light source to illuminate the intended area and a 18.5 inch monitor for display. It also allows the users to redefine the functions that take place when a button is pressed. The Telepack + is a non-patient contacting and require only wipe down as needed.

    AI/ML Overview

    The KARL STORZ TELEPACK+ is an all-in-one imaging system for endoscopic and microscopic procedures. This submission (K221174) is a modification of a previously cleared device (K193235). The key modifications include:

    • Swapping of two camera types (X-link and Office Link) via touchscreen.
    • Addition of capacitors to the Office Link front end.
    • Mechanical changes: added isolation to the microphone, an additional cover for the power supply, and a mechanical metal part as a protective earth.
    • Addition of video playback and image review functionality.
    • Additional cybersecurity features.
    • Additional camera and videoscope compatibility.

    The device's non-clinical performance data and bench testing were deemed sufficient to demonstrate substantial equivalence to its predicate device. This implies that the device meets its design specifications and is as safe and effective as the legally marketed predicate devices.

    Here's the breakdown of the acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria / Performance MetricReported Device Performance
    Electrical Safety and EMC
    IEC 60601-1 (General medical electrical equipment)Passed
    IEC 60601-1-2 (Electromagnetic compatibility)Passed
    IEC 60601-2-18 (Endoscopic equipment)Passed
    IEC 62471 (Photobiological safety)Passed
    Software Verification and Validation
    Guidance for the Content of Premarket SubmissionsMet (Level of concern: Moderate)
    Performance Testing
    Minimum IlluminationMet design specifications
    Spatial ResolutionMet design specifications
    Color PerformanceMet design specifications
    LatencyMet design specifications
    White BalanceMet design specifications
    AE (Automatic Exposure) Step ResponseMet design specifications
    Head Button FunctionalityMet design specifications
    Protection against electrical shockClass I protection, Type BF protection (stroboscopy & camera), Type CF protection (light)
    Moisture protectionDrip-water protection per IPX1

    2. Sample size used for the test set and the data provenance

    The document does not specify a separate "test set" and its sample size in the context of a clinical study, as no clinical performance data was required. The "test set" for demonstrating substantial equivalence primarily involved bench testing and electrical, EMC, and software validation tests. The data provenance is internal to KARL STORZ Endoscopy America, Inc., as these were laboratory and engineering tests conducted on the device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. For this type of device (an imaging system, not an AI diagnostic tool), "ground truth" established by medical experts for a test set is typically not germane to the substantial equivalence pathway, especially when no clinical data is required. The ground truth for engineering and performance measurements is established by standard measurement techniques and reference equipment.

    4. Adjudication method for the test set

    Not applicable. Since no clinical study or expert review for diagnostic accuracy was performed or required, an adjudication method for a test set is not relevant.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. The TELEPACK+ is an imaging system and not an AI-assisted diagnostic device. Therefore, an MRMC study comparing human readers with and without AI assistance was not conducted.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. The TELEPACK+ is a medical imaging system, not a standalone algorithm. Its performance is evaluated on its ability to produce high-quality images and manage endoscopic/microscopic procedures, not on an algorithm's diagnostic capabilities.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth for the performance testing (e.g., minimum illumination, spatial resolution, color performance) was established through objective engineering measurements using calibrated equipment and standard test methods against the device's design specifications. For electrical safety and EMC, the "ground truth" is compliance with recognized international standards (e.g., IEC 60601 series).

    8. The sample size for the training set

    Not applicable. The TELEPACK+ relies on traditional imaging processing and control technologies, not machine learning or artificial intelligence that would require a "training set" in the context of algorithm development.

    9. How the ground truth for the training set was established

    Not applicable. As no training set for an AI algorithm was used, this question is not relevant.

    Ask a Question

    Ask a specific question about this device

    K Number
    K212476

    Validate with FDA (Live)

    Date Cleared
    2021-09-03

    (28 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Flexible HD Cysto-Urethroscope System is used to provide visualization and operative access during diagnostic and therapeutic endoscopic procedures of urinary tract including the urethra, bladder, ureters, and kidneys.

    Device Description

    The Flexible HD Cysto-Urethroscope System (Part Number: 11272VH(U)-TL) is intended to be used with the IMAGE1 S CCU consisting of Connect / Connect II and X-Link modules (cleared in K201135). Identical to the predicates, the scope cannot be operated on its own because it produces image data output in the form of video signals which require external processing and display. When the scope is used with the compatible CCU, it becomes the Flexible HD Cysto-Urethroscope System, which can then be operated as intended and indicated.

    The Flexible HD Cysto-Urethroscope System consists of the following key components:
    11272VH-TL / 11272VHU-TL – Cysto-Urethroscopes with positive (VH-TL) / contra-positive (VHU-TL) deflection control and permanent T-Luer affixed to working channel
    TC200US / TC201US / TC301US – Connect/Connect II/Link camera control unit (CCU) modules for image processing

    The subject device is a lower cost alternative to the primary predicate and incorporates the following differences. Housing Plug: Due to lack of clinical need, the subject device replaces the suction channel of the primary predicate device with a permanent stainless steel housing plug. Fixed LUER: The subject device uses a removable T-Luer whereas the primary predicate supported a removable T-Luer or Stopcock.

    AI/ML Overview

    The FDA 510(k) summary for the KARL STORZ Flexible HD Cysto-Urethroscope System (K212476) describes the acceptance criteria and supporting studies as follows:

    This submission is a Special 510(k), indicating a modification to a previously cleared device. Therefore, the focus is on demonstrating that the modifications do not raise new questions of safety and effectiveness compared to the predicate devices.

    1. Table of Acceptance Criteria and Reported Device Performance

    For this Special 510(k), the acceptance criteria are largely based on adherence to recognized consensus standards and demonstrating that the modified device maintains the same technological characteristics and performance as the predicate devices, or that any differences do not impact safety or effectiveness.

    Acceptance CriterionReported Device Performance / Supporting Data
    Intended Use / Indications for UseIdentical to predicate devices: "The Flexible HD Cysto-Urethroscope System is used to provide visualization and operative access during diagnostic and therapeutic endoscopic procedures of urinary tract including the urethra, bladder, ureters, and kidneys." Minor modifications do not change the intended use or indications for use.
    Physical Characteristics (e.g., Insertion Shaft Diameter, Length, Working Channel Diameter, Deflection)Same as primary predicate (K191357) for: - Type of scope (Flexible) - Insertion Shaft Diameter (5.5 mm) - Insertion Shaft Length (37 cm) - Working Channel Diameter (2.3 mm) - Deflection (Up: 210°, Down: 140°)
    Optical Characteristics (e.g., Type of Imager, Field of View, Direction of View, Depth of Field, On-axis Resolution, Light Source)Same as primary predicate (K191357) for: - Type of Imager (CMOS) - Field of View (100°) - Direction of View (0°) - Depth of Field (3-50 mm) - On-axis Resolution (40 lp/mm @ 3 mm, 2.5 lp/mm @ 50 mm) - Light Source (Internal LED)
    Material / Design Characteristics (Differences from Predicate)Suction Port / Channel: Proposed device has "No" (due to permanent stainless steel housing plug), while primary predicate has "Yes". This difference is described as due to "lack of clinical need" for the suction channel. Luer Type: Proposed device has "Permanent T-Luer", primary predicate has "Bayonet with removable T-Luer or Stopcock assembly". This is a minor modification.
    Cleaning, Disinfection and Sterilization MethodsSame as primary predicate for Manual Cleaning, and High-Level Disinfection (Revital-Ox RESERT). Sterilization: Proposed device supports STERRAD 100NX (FLEX and DUO cycles), STERRAD NX (Advanced Cycle), V-PRO maX (Flexible Cycle), V-PRO 60. The primary predicate supported these with additional support for SSIE (Standard Cycle). The proposed device's methods are a recognized subset.
    Electrical Safety and EMCCompliance with recognized consensus standards: - IEC 60601-1:2005/(R)2012, Ed. 3.1 - IEC 60601-1-2:2014, 4th Ed. - IEC 60601-2-18:2009, Ed. 3.0
    Optical/Mechanical PerformanceCompliance with recognized consensus standards: - IEC 62471:2006 - ISO 8600
    BiocompatibilityCompliance with recognized consensus standards: - ISO 10993-1:2009/(R) 2013 - ISO 10993-5:2009/(R) 2014 - ISO 10993-10:2010 - ISO 10993-11:2006/(R) 2010
    ReprocessingCompliance with recognized consensus standards: - AAMI TIR 12:2010 - ISO 15883-5:2005 - AAMI TIR 30:2011 - AAMI/ANSI/ISO 11737-1:2006/ (R)2011 - ASTM E1837-96:2014
    Substantial Equivalence to effectivenessSupported by comparison of images and standard image quality characteristics, including the performance testing listed. The differences do not raise new or different questions of safety and effectiveness.

    2. Sample size used for the test set and the data provenance:

    The document does not detail specific "test set" sample sizes or data provenance in the context of clinical or image-based studies for the current 510(k). This submission relies on a comparison to predicate devices and adherence to established performance standards.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    Not applicable. The submission focuses on engineering specifications and adherence to standards, not a clinical study requiring expert ground truth for image interpretation or diagnosis.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    Not applicable. No clinical or image-based test set with adjudication is described.

    5. If a multi-reader, multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This device is an endoscope system for visualization and operative access; it does not involve AI assistance for human readers in diagnostic interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Not applicable. This device is not an algorithm-only device. It is a medical instrument.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    Not applicable in the context of diagnostic performance. The "ground truth" for this submission are the established performance parameters of the predicate device and the requirements of recognized consensus standards for medical devices.

    8. The sample size for the training set:

    Not applicable. As a medical device, and specifically a modification through a Special 510(k), there is no "training set" in the context of machine learning.

    9. How the ground truth for the training set was established:

    Not applicable. See point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K202272

    Validate with FDA (Live)

    Date Cleared
    2021-05-11

    (273 days)

    Product Code
    Regulation Number
    874.4720
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The HD Mediastinoscope is intended to aid the surgeon in viewing the mediastinum and facilitate the introduction and removal of surgical instruments during ENT endoscopic surgical procedures.

    Device Description

    The components subject of this submission are: HD Mediastinoscope (Part Number: 10973HD), and the IMAGE1 S CCU. The CCU consists of the IMAGE1 S Connect Module (Model Number: TC200US) and IMAGE1 S X-Link (Model Number: TC301US).

    AI/ML Overview

    This document describes the FDA's 510(k) clearance for the Karl Storz HD Mediastinoscope (K202272). The device is intended to aid surgeons in viewing the mediastinum and facilitating the introduction and removal of surgical instruments during ENT endoscopic procedures.

    The information provided focuses on the substantial equivalence determination based on non-clinical performance data, rather than a study proving the device meets AI-specific acceptance criteria. Therefore, most of the requested information regarding AI model performance, ground truth establishment, expert review, and clinical study details is not present in the provided text.

    Here is a breakdown of what can be extracted and what is not available based on the request:

    1. A table of acceptance criteria and the reported device performance

    The document does not specify quantitative acceptance criteria in a table format for performance. Instead, it states compliance with recognized standards and successful completion of various bench tests.

    Test CategoryAcceptance Criteria (Implied by Compliance)Reported Device Performance
    Electrical Safety & EMCCompliance with ANSI/AAMI ES:60601-1:2005+A1:2012 and IEC 60601-1-2:2014In compliance with the stated standards.
    BiocompatibilityCompliance with ISO 10993-1, ISO 10993-5:2009/(R) 2014, ISO 10993-10:2010, ISO 10993-11:2006/(R) 2010, and FDA GuidanceEvaluation performed, tests conducted based on contact type and duration.
    Reprocessing ValidationCompliance with AAMI TIR 12:2010, AAMI TIR 30:2011, AAMI/ANSI/ISO 11737-1:2006/(R)2011, ASTM E1837-96:2014Validation activities performed for cleaning and sterilization; in compliance with standards.
    Optical PerformanceNot explicitly quantified, but successful completion of tests is implied for substantial equivalence.
    - Color Reproduction & Contrast(Implied successful performance for medical imaging)Tested.
    - Illumination Detection Uniformity(Implied successful performance)Tested.
    - Instantaneous Dynamic Range (IDR) & Detection Linearity(Implied successful performance)Tested.
    - Spatial Resolution & Depth of Field(Implied successful performance)Tested.
    - Distortion & Field of View(Implied successful performance)Tested.

    2. Sample size used for the test set and the data provenance

    The document explicitly states: "Clinical testing was not required to demonstrate the substantial equivalence to the predicate devices. Non-clinical bench testing was sufficient to establish the substantial equivalence of the modifications." This means there was no clinical test set in the traditional sense for assessing device performance in a human population. All testing was non-clinical bench testing. Therefore, sample size and data provenance (country, retrospective/prospective) related to a clinical test set are not applicable here.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable, as no clinical test set or ground truth in the context of diagnostic interpretation (e.g., for an AI model) was established. The ground truth for bench testing would be metrology standards and physical measurements.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable, as no clinical test set involving human readers or interpretation was performed.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a mediastinoscope (a physical viewing instrument), not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an algorithm-only device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the non-clinical bench testing, the ground truth would be based on established engineering and optical metrology standards, and potentially manufacturer specifications for the device's components. For biocompatibility and reprocessing, the ground truth is compliance with the listed ISO and AAMI standards.

    8. The sample size for the training set

    Not applicable, as this is not an AI/ML device that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable, as this is not an AI/ML device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 3