510(k) Data Aggregation
(211 days)
The gown is intended to protect the wearer from blood and other body fluids encountered during surgery
This surgical gown is similar in design and composition to the Kimberly Clark gown identified in a pre-market notification #K781682. The Kappler Med-Guard surgical gown is constructed from a tri-layer composite material which is a 100% polypropylene. The specific arrangement is spunbonded polypropylene/meltblown polypropylene film/spunbonded polypropylene. The fabric is treated with an anti-static agent.
The provided text describes a 510(k) summary for the Kappler Med-Guard Surgical Gown, focusing on its substantial equivalence to a predicate device (Kimberly Clark gown #K781682). The submission primarily relies on non-clinical testing to demonstrate performance and clinical testing to address the safety of an anti-static treatment. There is no information provided about AI or machine learning algorithms, therefore the questions related to AI-specific studies are not applicable.
Here's the breakdown of the acceptance criteria and the studies performed, based on the provided text:
Acceptance Criteria and Device Performance for Kappler Med-Guard Surgical Gown
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are implicitly defined by the performance characteristics of the predicate device (Kimberly-Clark surgical gown). The Kappler Med-Guard surgical gown aims to meet or exceed these characteristics to demonstrate substantial equivalence.
| Characteristic | Acceptance Criteria (Predicate Device K-C) | Reported Device Performance (Med-Guard) | Met Acceptance Criteria? |
|---|---|---|---|
| Material Composition | 100% Polypropylene | 100% Polypropylene | Yes |
| Configuration | Tri-laminate | Tri-laminate | Yes |
| Barrier Layer | Meltblown | Meltblown | Yes |
| Composite Weight | 1.5 oz./yd² | 1.8 oz./yd² | Yes (Equal or better) |
| Hydrostatic Head | 20.7 inches | 44.0 cm (approx. 17.3 inches) | No (Lower performance) |
| Bacterial Filtration | 81.3% | 78% | No (Lower performance) |
| Moisture Vapor Transmission Rate | 1794 grs/sqm/24 hr | 1789 g/24 hr m² | Yes (Comparable) |
| Flammability | Class 1 | Class 1 | Yes |
| Sterilization | 100% ETO | 100% ETO | Yes |
| Anti-static treatment | Yes | Yes | Yes |
Note on "Met Acceptance Criteria?": While some Med-Guard performance values (Hydrostatic Head, Bacterial Filtration) are numerically lower than the predicate, the FDA ultimately made a "substantially equivalent" determination. This suggests that these differences were deemed acceptable and did not raise new questions of safety or effectiveness. For example, a lower Hydrostatic Head might indicate less resistance to water, and lower bacterial filtration might mean slightly less protection. However, a slight variation within a functional range might still be considered equivalent for the intended use. The text doesn't provide the specific thresholds for "acceptable" values, only the comparison to the predicate.
2. Sample Size Used for the Test Set and the Data Provenance
Non-clinical testing: The text does not specify the exact sample sizes (number of gowns or material samples) for each non-clinical test (e.g., Hydrostatic Pressure Resistance, Bacterial Filtration Efficiency). It only states that "the individual test data sheets identify the samples as 'Blue Hawk'".
Clinical testing: The text does not specify the sample size (number of subjects or samples) for the clinical tests (Acute Systemic Toxicity, Cytotoxicity, Delayed Contact Sensitization Study).
Data provenance: Not specified, but given the nature of the tests (material properties, biological safety), they would be conducted in a laboratory setting, likely in the US, and are inherently prospective in nature.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts
Not applicable in the context of device performance testing. The "ground truth" here is the empirically measured physical and biological properties of the material and the device, as determined by standardized test methods (e.g., ASTM, MIL, NFPA). These are objective measurements rather than subjective expert opinion.
For the clinical testing, North American Science Associates, Inc. (NAmSA) established the protocol and evaluated the results, concluding no concerns. This suggests experts at NAmSA were involved, but their number and specific qualifications are not detailed.
4. Adjudication Method for the Test Set
Not applicable. This relates to objective measurements from standard tests, not subjective interpretations requiring adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done
No. This is not relevant for a surgical gown. The submission evaluates the physical, barrier, and comfort characteristics, and the safety of anti-static treatment, not diagnostic accuracy requiring human reader interpretation.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done
No. This is not relevant for a surgical gown. There is no algorithm or AI component in this device.
7. The Type of Ground Truth Used
- Non-clinical testing: Objective, empirically measured physical and biological properties of the gown material, determined by adherence to established test standards (e.g., ASTM D 751, SOP/ARO/007F.1/MIL-M-36954C, ASTM E 96, NFPA 702, ASTM D 5733).
- Clinical testing: Results from standardized biological assays (Acute Systemic Toxicity, Cytotoxicity, Delayed Contact Sensitization Study) to assess biocompatibility and safety related to the anti-static treatment.
8. The Sample Size for the Training Set
Not applicable. There is no "training set" as this device does not involve machine learning or AI.
9. How the Ground Truth for the Training Set was Established
Not applicable. There is no "training set" as this device does not involve machine learning or AI.
Ask a specific question about this device
(58 days)
The TE Gown is intended to provide the wearer an increased level of blood and other body fluid protection during certain surgical procedures that result in excessive blood exposures by providing total encapsulation.
The COOL-ZONE System is either a gown or hood that are similar to the Kappler Pro/Vent Surgical Gown which was the subject of Premarket Notification #K932694. The fabric in the COOL-20NE gown and hood is identical to the fabric in the Pro/Vent Surgical Gown. The Pro/Vent TE Gown and Hood are constructed from the same fabric as the predicate device. The design of the gown differs in that the gament provides totally encapsulated protection to the wearer. Total encapsulation means total coverage of the entire upper body, thus providing greater protection than conventional surgical apparel.. The TE Gown utilizes a mylar face shield for visibility. A zipper is located in the back of the garment. The TE Hood provides the same level of face and head protection, but does not extend down as far on the body. It will typically be worn with another surgical gown.
This document is a 510(k) Summary for a medical device (surgical gown/hood) and does not describe an AI/ML powered device. Therefore, most of the requested information regarding AI-specific evaluation criteria, such as "Multi Reader Multi Case (MRMC) comparative effectiveness study," "standalone (i.e., algorithm only) performance," "sample size for the training set," or "adjudication method for the test set," is not applicable.
The document focuses on demonstrating substantial equivalence to a predicate device based on material composition, design, and non-clinical performance characteristics.
Here's an attempt to extract relevant information given the limitations:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are implicitly defined by the properties of the predicate device (Kappler Pro/Vent Surgical Gown). The device ("PRO/VENT TE") is considered to meet acceptance criteria if its characteristics are "identical" or equivalent to the predicate.
| CHARACTERISTIC | Acceptance Criteria (Predicate: PRO/VENT GOWN) | Reported Device Performance (PRO/VENT TE) |
|---|---|---|
| Material Composition | 100% Polypropylene | 100% Polypropylene |
| Fabric Configuration | Trilaminate | Trilaminate |
| Barrier Layer | Microporous Film | Microporous Film |
| Composite Weight | 2.1 oz/square yard | 2.1 oz/square yard |
| Hydrostatic Head | 43 inches | 43 inches |
| Bacterial Filtration | 99.9% | 99.9% |
| Moisture Vapor Transmission Rate | 1385 g/sqm/24hr | 1385 g/sqm/24hr |
| Flammability | Class 1 | Class 1 |
| Synthetic Blood/Viral Resistance | Pass | Pass |
| Sterility Assurance Level (SAL) | 10⁻⁶ (typical for medical devices) | 10⁻⁶ |
| Ethylene Oxide Residues | < 25 ppm (FDA limits) | < 25 ppm |
| Ethylene Chlorohydrin Residues | < 50 ppm (FDA limits) | < 50 ppm |
| Ethylene Glycol Residues | < 250 ppm (FDA limits) | < 250 ppm |
2. Sample size used for the test set and the data provenance
- Sample Size for Test Set: Not explicitly stated. The performance data is presented as generic characteristics of the material rather than results from a specific "test set" in the context of individual units. It's likely these are properties of the material itself.
- Data Provenance: Not specified. It represents the inherent properties of the material and product design as characterized by laboratory tests.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. This is a physical product, and "ground truth" is established through standardized material testing, not expert consensus on interpretations.
4. Adjudication method for the test set
Not applicable. Material properties are measured objectively.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is not an AI/ML powered device.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
Not applicable. This is not an AI/ML powered device.
7. The type of ground truth used
The "ground truth" for the performance characteristics is derived from standardized material testing and laboratory measurements (e.g., ASTM or ISO standards for hydrostatic head, bacterial filtration, flammability, synthetic blood resistance) and sterilization validation protocols (e.g., overkill method for ETO sterilization and residue testing).
8. The sample size for the training set
Not applicable. This is not an AI/ML powered device, so there is no "training set."
9. How the ground truth for the training set was established
Not applicable. There is no training set for this type of device.