Search Filters

Search Results

Found 15 results

510(k) Data Aggregation

    K Number
    K243468

    Validate with FDA (Live)

    Date Cleared
    2025-07-02

    (236 days)

    Product Code
    Regulation Number
    868.5630
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Compressor Nebulizer system is intended to provide air to the pneumatic nebulizer in order to aerosolize medications for inhalation by the patient for respiratory disorders. The nebulizer is driven by the integral air compressor.

    The system is designed for use with pediatric (ages 2 years and above) and adult patients in the home, hospital, and sub-acute settings.

    Device Description

    The compressor Nebulizer is mainly composed of the compressor and nebulizer kit(Optional). The compressor is mainly composed of shell, compressed motor, pump, fuse wire, air filter, power cord and plug (NB-1100,NB-1101,NB-1102,NB-1103 applicable), PCB (only model NB-1102,NB-1103applicable). The device is equipped with a nebulizer kit including Nebulizer Cup, Air Tube, Mouthpiece(Optional), Adult Mask(Optional), Child Mask(Optional) to easily delivery the medical aerosol. The Nebulizer kit is available in three models/configurations as below.

    • NK-101: Nebulizer Cup, Air Tube, Adult Mask and/or Child Mask, Mouthpiece;
    • NK-301: Nebulizer Cup, Air Tube, Adult Mask and/or Child Mask;
    • NK-501: Nebulizer Cup, Air Tube, Mouthpiece
      This device operates on the Venturi principle. According to the principle of Venturi, the compressed air of the motor is used to form a high-speed air flow through the small pipe mouth. The negative pressure generated drives the liquid or other fluid to spray together on the barrier, and under the high-speed impact makes the droplets turn into mist particles to spray from the outlet trachea.
    AI/ML Overview

    The provided FDA 510(k) clearance letter and summary for JOYTECH Healthcare Co., Ltd.'s Compressor Nebulizer (K243468) primarily details the substantial equivalence of the new device to a predicate device, focusing on non-clinical performance testing, specifically particle size characterization.

    It's important to note that this document does not describe a study involving human subjects or artificial intelligence (AI). The "acceptance criteria" and "study that proves the device meets the acceptance criteria" in this context refer to the engineering and performance specifications for a nebulizer, evaluated through laboratory testing against a predicate device and relevant standards. Therefore, many of the requested points related to AI/MRMC studies, expert ground truth, adjudication methods, and training sets for algorithms are not applicable to this specific device clearance.

    Below is an interpretation of the provided information within the framework of your request, focusing on the relevant parts and explicitly stating where information is not available or not applicable.


    Acceptance Criteria and Device Performance for JOYTECH Compressor Nebulizer (K243468)

    The acceptance criteria for this device are established by demonstrating substantial equivalence to a legally marketed predicate device (Omron Compressor Nebulizer Systems NE-C801, K110860) and by meeting relevant performance standards, particularly regarding aerosol particle characterization. The 'study' proving this is primarily the "Particle Size characterization testing" detailed in the submission.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implicitly defined by performing "comparative particle test comparison" against the predicate device. The goal is to show that the proposed device performs similarly or equivalently to the predicate across key aerosolization parameters. The performance data provided is the direct result of this testing.

    Table: Comparative Particle Test Comparison (Summary of Select Parameters)

    Test ItemEndpoint (Acceptance Criteria Implicitly Compared to Predicate)Subject Device (JOYTECH NB-1103) Performance (Range across drugs/interfaces)Predicate Device (OMRON NEC801) Performance (Range across drugs/interfaces)Comparison Result (as per Submitter)
    MMAD ($\mu$m)Particle size distribution suitable for inhalation therapy, comparable to predicate.3.413 - 5.7573.7 - 6.248Equivalent / Similar
    GSDParticle size distribution uniformity, comparable to predicate.1.764 - 2.3811.605 - 2.141Equivalent / Similar
    Respirable Dose ($\mu$g)Amount of drug delivered in respirable range, comparable to predicate.51.307 - 1628.63350.723 - 1239.14Equivalent / Similar
    Respirable Fraction (%)Proportion of aerosolized drug in respirable range, comparable to predicate.39.103 - 67.95831.781 - 63.741Equivalent / Similar
    Total Delivered Dose ($\mu$g)Total amount of drug delivered, comparable to predicate.112.976 - 2562.202146.303 - 2131.073Equivalent / Similar
    Total Delivered Dose Fraction (%)Proportion of initial drug dose delivered, comparable to predicate.21.305 - 49.06327.55 - 40.801Equivalent / Similar
    Fine Particle Fraction (%) (<4.7$\mu$m)Percentage of particles less than 4.7 $\mu$m, comparable to predicate.35.454 - 64.7727.946 - 60.533Equivalent / Similar
    Coarse Particle Fraction (%) (>4.7$\mu$m)Percentage of particles greater than 4.7 $\mu$m, comparable to predicate.35.232 - 64.54739.467 - 72.054Equivalent / Similar
    Ultra-Fine Particle Fraction (%) (<1.0$\mu$m)Percentage of particles less than 1.0 $\mu$m, comparable to predicate.Ranges provided are minimal/absent for Budesonide, otherwise 4.006-5.891Ranges provided are minimal/absent for Budesonide, otherwise 2.456-4.26Equivalent / Similar

    Note: The ranges represent the lowest and highest values across the tested drugs (Albuterol Sulfate, Ipratropium Bromide, Budesonide) and patient interfaces (adult mask, child mask, mouthpiece) for both adult and pediatric conditions. The "Comparison Result" is the submitter's statement from the document.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The tables show mean and standard deviation values for each parameter, implying multiple measurements were taken for each drug and interface combination. However, the exact number of replicates (sample size for each data point in the tables) is not explicitly stated in the provided text. It is typically standard practice in such testing to perform a minimum of triplicate measurements.
    • Data Provenance: The data is generated from laboratory performance testing of the device and its predicate. This is a non-clinical study.
      • Country of Origin: The manufacturer, JOYTECH Healthcare Co., Ltd., is located in Hangzhou City, Zhejiang, China. The predicate manufacturer, Omron Healthcare, Incorporated, is typically based in Japan with a strong presence in the US. The testing itself would have been conducted in a certified lab, likely in China or an associated region.
      • Retrospective or Prospective: This is prospective laboratory testing conducted specifically for this 510(k) submission.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    Not applicable. This study does not involve expert interpretation of medical images or patient data to establish a ground truth. Instead, it involves physical and chemical measurements (particle size and drug delivery) performed on the nebulizer using standardized laboratory methods outlined in FDA guidance and relevant ISO standards (e.g., ISO 18562 series). The "ground truth" is derived from these objective measurements and comparison to the predicate's performance.

    4. Adjudication Method for the Test Set

    Not applicable. As this is laboratory performance testing, there's no "adjudication" in the sense of reconciling human expert opinions. The results are quantitative measurements.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, an MRMC comparative effectiveness study was not done. This device is a medical device for drug delivery, not an AI diagnostic or assistive tool for human readers. Therefore, there is no human-in-the-loop performance to improve with AI assistance.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) was Done

    Not applicable. This device is hardware (a compressor nebulizer), not a software algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's performance testing is objective physical and chemical measurement data obtained through standardized laboratory protocols (e.g., cascade impaction for particle sizing, chemical analysis for drug quantity) rather than expert consensus, pathology, or outcomes data. The acceptability is determined by evaluating if these measurements are comparable to the well-established performance of a predicate device already on the market and align with general principles for effective nebulizer function.

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI/machine learning device, so there is no "training set."

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. There is no "training set" or "ground truth" in the context of an AI model for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K250548

    Validate with FDA (Live)

    Date Cleared
    2025-06-26

    (121 days)

    Product Code
    Regulation Number
    870.1130
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Arm-type Fully Automatic Digital Blood Pressure Monitors are intended to measure blood pressure (systolic and diastolic) and pulse rate of adults and adolescents over 12 years of age with circumference ranging from 22cm to 42cm and 40cm-56cm.

    Device Description

    The Arm-type Blood Pressure Monitor (BPM) series is an automatic, non-invasive, blood pressure measurement system for over-the-counter (OTC) use in home and clinical environment. The systolic and diastolic pressures are determined using the Oscillometric method, where the cuff is inflated with an integral controllable piezoelectric pump and deflates via an electric automatic rapid deflation valve. During measurements, an electric pump within the main unit slowly inflates the arm cuff, generating cuff pressure which is monitored and from which pulse waveform data is extracted. This waveform data is analyzed by software algorithms within the microprocessor to determine pulse rate, systolic pressure, and diastolic pressure. The cuff can measure pressure range from 0 to 299mmHg, and the pulse rate range from 30 to 180 beats per minute.

    The DBP-6286B embed bluetooth module to transfer data to APP. With the use of software (including APP) and Bluetooth communication module, the wireless software function and hardware function are solely intended to transfer, store, convert formats, or display medical device data and results (blood pressure and pulse rate readings), without controlling or altering the functions or parameters of any connected medical devices, which is not be intended for active patient monitoring, therefore, based on the FDA guidance titled "Medical Device Data Systems, Medical Image Storage Devices, and Medical Image Communications Devices" (issued on September 28, 2022.), this software function is belong to Non-device-MDDS.

    AI/ML Overview

    Based on the provided FDA 510(k) clearance letter for the JOYTECH Healthcare Co., Ltd. Arm-type Fully Automatic Digital Blood Pressure Monitors (DBP-6286B, DBP-6186), here's a description of the acceptance criteria and the study proving the device meets them:

    Acceptance Criteria and Device Performance for Blood Pressure Monitors

    The acceptance criteria for blood pressure monitors are typically defined by recognized international standards, primarily ISO 81060-2:2018+AMD2020, which pertains to the clinical investigation of intermittent automated measurement type for non-invasive sphygmomanometers. This standard specifies the accuracy requirements for blood pressure measurements.

    The key acceptance criterion is that the accuracy of the blood pressure monitor must be within the acceptable scope specified in ISO 81060-2:2018+AMD2020. While the exact numerical thresholds for group means and standard deviations aren't explicitly stated in this clearance letter, the general requirement is clear.

    The reported device performance confirms that the device meets these criteria.

    Acceptance CriterionReported Device Performance
    Blood Pressure Accuracy (Systolic and Diastolic) within acceptable scope of ISO 81060-2:2018+AMD2020For cuff with arm circumference 22-42cm: The results showed the accuracy of the blood pressure monitor is within acceptable scope specified in ISO 81060-2:2018+AMD2020. This was covered by a representative study on DBP-6279B, which shares the same key factors (principle, measurement type, core algorithm, pressure sensor, MCU, cuff size, and material) affecting clinical accuracy.
    Blood Pressure Accuracy (Systolic and Diastolic) within acceptable scope of ISO 81060-2:2018+AMD2020For cuff with arm circumference 40-56cm: The results showed the accuracy of the blood pressure monitor is within acceptable scope specified in ISO 81060-2:2018+AMD2020. This was demonstrated by a study on DBP-6286B.

    Study Details Proving Device Meets Acceptance Criteria:

    1. A table of acceptance criteria and the reported device performance: (See table above)

    2. Sample sizes used for the test set and the data provenance:

      • Cuff (22-42cm): 91 valid subjects' data were used for analysis.
      • Cuff (40-56cm): 85 valid subjects' data were used for analysis.
      • Data Provenance: The document does not explicitly state the country of origin of the data, but it implies a prospective clinical study ("All the subjects were volunteer to take part in the clinical study, all the subjects completed the clinical study without any AE or side-effect").
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • The document implies the use of human observers for the reference measurements, as it states: "The manual Mercury Sphygmomanometer was used as a reference device." This suggests that the ground truth was established by human operators taking readings from a mercury sphygmomanometer against which the automated device's readings were compared. However, the exact number and qualifications of these "experts" (human observers) are not specified in the provided document.
    4. Adjudication method for the test set:

      • The document states, "Same arm sequential method was adopted during clinical testing." This method involves comparing simultaneous or near-simultaneous measurements from the test device and a reference device on the same arm. While this describes the measurement procedure, it does not describe an adjudication method for discrepancies between readings, as would be typical for expert consensus. Given that the ground truth is a direct measurement from a mercury sphygmomanometer, an adjudication process in the traditional sense (e.g., for image interpretation) is not directly applicable.
    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No. This is a medical device clearance for a blood pressure monitor, not an AI-assisted diagnostic imaging device. Therefore, an MRMC study and analysis of human reader improvement with AI assistance are irrelevant and were not performed.
    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • The study design (clinical validation against a reference standard) inherently evaluates the "standalone" performance of the automated blood pressure monitor's algorithm. The device itself is designed to operate without continuous human intervention during the measurement process, making its performance essentially "algorithm only" in terms of blood pressure determination from the oscillometric data.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The ground truth was established by direct measurement using a manual Mercury Sphygmomanometer. This is considered the clinical gold standard for non-invasive blood pressure measurement in this context.
    8. The sample size for the training set:

      • The document does not specify a separate training set or its sample size. For traditional medical devices like blood pressure monitors, the "training" (development and calibration) often occurs during the device engineering and manufacturing phases, and clinical validation is then performed on a separate, independent test set as described.
    9. How the ground truth for the training set was established:

      • As no specific "training set" is mentioned in the context of a separate clinical dataset, the method for establishing ground truth for such a set is not provided. The device's underlying algorithms would have been developed and refined through engineering and calibration processes, likely using extensive internal testing and validation against standardized pressure sources and reference measurements, rather than a distinct "clinical training set" in the AI/ML sense.
    Ask a Question

    Ask a specific question about this device

    K Number
    K241431

    Validate with FDA (Live)

    Date Cleared
    2025-01-13

    (237 days)

    Product Code
    Regulation Number
    870.1130
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Arm-type Fully Automatic Digital Blood Pressure Monitors are intended to measure blood Pressure (systolic and diastolic) of adults and adolescents over 12 years of age with circumference ranging from 22cm to 24cm.

    Device Description

    The Arm-type Fully Automatic Digital Blood Pressure Monitor (BPM) series is automatic, non-invasive, blood pressure measurement system for over-the-counter (OTC) use at home and clinical environment . The systolic and diastolic pressures are determined using the oscillometric method, where the cuff is inflated with an integral controllable piezoelectric pump and deflates via an electric automatic rapid deflation valve. During inflation measurements, an electric pump within the main unit slowly inflates the arm cuff, generating cuff pressure which is monitored and from which pulse waveform data is extracted. This waveform data is analyzed by software algorithms within the microprocessor to determine systolic pressure, and diastolic pressure. The cuff can measure pressure range from 0 to 299mmHg.

    Meanwhile, some models whith bluetooth function can be used as a stand-alone unit to finish the blood pressure measurement or in conjunction with the APP through embed a 2.4GHz BLE module that allow users to connect with nearby BT receiving terminal. Once measurement is over, the LCD of the device displays results. And the device will start to transmit data to the pair-up terminal automatically. This app is only intended to display measurement results from the blood pressure monitor device, which does not provide any diagnostic or measurement functions, and does not interpret or analyze the data for medical decision making. Unlimited readings can be stored in the app for archiving and review by the user.

    There are total 7 arm-type blood pressure monitor models we submitted: DBP-62D0L, DBP-62D0B, DBP-61D0, DBP-61D0L, DBP-6293L, DBP-6193 and they are both bodyworn medical devices.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided FDA 510(k) summary for the JOYTECH Arm-type Fully Automatic Digital Blood Pressure Monitor:

    Acceptance Criteria and Reported Device Performance

    The primary acceptance criteria for a non-invasive blood pressure monitor are its accuracy in measuring systolic and diastolic blood pressure. The relevant standard cited is ISO 81060-2:2018+AMD2020 (Non-invasive sphygmomanometers — Part 2: Clinical investigation of intermittent automated measurement type).

    The document states: "The results showed the accuracy of the blood pressure monitor is within acceptable scope specified in ISO 81060-2:2018+AMD2020." While specific numerical performance values (e.g., mean difference and standard deviation for systolic and diastolic pressure) are not explicitly provided in this summary, the statement indicates that the device met the statistical requirements defined by this international standard for accuracy.

    Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Based on ISO 81060-2:2018+AMD2020)Reported Device Performance
    Accuracy of Systolic Blood Pressure MeasurementWithin acceptable scope specified in ISO 81060-2:2018+AMD2020
    Accuracy of Diastolic Blood Pressure MeasurementWithin acceptable scope specified in ISO 81060-2:2018+AMD2020

    Details of the Study Proving Device Meets Acceptance Criteria:

    1. Sample Size and Data Provenance:

      • Test Set Sample Size: A total of 90 subjects were included in the clinical validation study.
      • Data Provenance: The document does not explicitly state the country of origin of the data, but the manufacturer is based in China. The study appears to be prospective as it describes recruitment of subjects, their participation, and the measurement process.
    2. Number of Experts and Qualifications:

      • The document mentions "The manual Mercury Sphygmomanometer was used as a reference device." This implies measurements were taken by trained observers (experts) using this reference device. However, the exact number of experts or their specific qualifications (e.g., "radiologist with 10 years of experience" is not applicable here) are not specified in the provided text. For blood pressure clinical validation, these would typically be trained clinicians or technicians.
    3. Adjudication Method for the Test Set:

      • The study used a "Same arm sequential method." This means the test device measurements and reference device measurements were taken sequentially on the same arm.
      • The document does not describe a specific "adjudication method" in the sense of multiple experts reviewing and reaching consensus on an interpretation (as would be common in image-based AI studies). Instead, the comparison is made between the device readings and the reference standard readings.
    4. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

      • No, an MRMC comparative effectiveness study was not done. This type of study is typically performed for AI devices that assist human readers in interpreting medical images (e.g., AI for chest X-rays assisting radiologists). The device in question is a standalone blood pressure monitor, not an AI-assisted diagnostic tool that requires human interpretation of complex data.
    5. Standalone Performance:

      • Yes, a standalone performance study was done. The clinical validation detailed here assesses the accuracy of the blood pressure monitor itself (the algorithm and hardware) against a recognized gold standard (manual mercury sphygmomanometer). The device operates independently to provide blood pressure measurements.
    6. Type of Ground Truth Used:

      • The ground truth used was manual Mercury Sphygmomanometer measurements. This is a widely accepted reference standard for validating automated blood pressure monitors, considered an expert consensus/reference standard method in this context.
    7. Sample Size for the Training Set:

      • The document does not provide information regarding the sample size of a training set. This is typical for a traditional medical device like a blood pressure monitor where the "algorithm" is often based on well-established non-AI methods (oscillometric principles) and validated through clinical testing rather than trained on large datasets like deep learning models. If there are software algorithms involved, they are likely more deterministic or based on classical signal processing rather than machine learning that requires a separate "training set."
    8. How Ground Truth for Training Set Was Established:

      • As no "training set" in the context of typical AI/machine learning is mentioned, this information is not applicable and therefore not provided in the document. The device's underlying measurement principles (oscillometric method) use known physical principles, not a data-driven training process that requires a labeled ground truth for learning.
    Ask a Question

    Ask a specific question about this device

    K Number
    K241322

    Validate with FDA (Live)

    Date Cleared
    2024-12-19

    (223 days)

    Product Code
    Regulation Number
    884.5160
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Electric Breast Pump (Model LD-208L, LD-3010L, LD-3010, LD-2010) is a powered breast pump to be used by lactating women to express and collect milk from their breasts. The Electric Breast Pump is intended for a single user.

    Device Description

    The Electric Breast Pump (LD-208L, LD-3010L, LD-3010, LD-3010, LD-2010) is an electrically powered breast pump to be used in a home environment by a single user. The device is provided non-sterile and can be used on one breast (single pumping) or on both breasts at the same time (double pumping).

    The subject devices are capable of expression, stimulation, and the hybrid 'bionio' and 'variable frequency' modes. Stimulation mode is associated with six suction levels, and expression, bionic, and variable frequency modes are associated with nine suction levels. The subject devices feature various combinations of on/off buttons, level up/down buttons, mode selection buttons, suction level/time indicators, battery visual indicator display, and mode displays. The LD-208L, LD-3010L, and LD-2010L pumps are powered by internal, nonreplaceable, rechargeable lithium-ion batteries which are charged using the included AC power supply and cable. The LD-2010 and LD-3010 are powered by alkaline batteries or AC power.

    The breast pumps use cyclic negative pressure (suction) to mimic the suckling patterns of a feeding infant. A DC motor drives a membrane vacuum pump to generate the suction required to stimulate and express breast milk. The timing of this pattern is dependent upon the suction/speed settings selected by the user and is pre-programmed in the devices. The devices are capable of producing peak suction levels between -40 and -290 mmHg at speeds between 20 and 120 cycles per minute. The subject device ensures backflow protection between the breast shield and the electronic components via a physical barrier (silicone diaphragm) mechanism.

    All other components (i.e., motor unit) of the subject device are not in contact with the breast. All milk contacting components are compliant with 21 CFR 177.

    AI/ML Overview

    The provided document is a 510(k) summary for an Electric Breast Pump (LD-208L, LD-3010L, LD-2010L, LD-3010, LD-2010). This type of regulatory submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving independent effectiveness or advanced performance metrics typically associated with AI/ML devices. Therefore, much of the information requested in your prompt (e.g., sample size for test sets, expert ground truth, MRMC studies, standalone performance with effect sizes) is not applicable or generally required for a 510(k) clearance of a mechanical device like a breast pump.

    Here's the information that can be extracted or inferred from the provided text:

    1. A table of acceptance criteria and the reported device performance

    For a breast pump, the "acceptance criteria" and "device performance" are primarily based on meeting safety standards, functional specifications, and demonstrating equivalence to a predicate. The document highlights various tests performed to ensure this.

    Acceptance Criteria (Measured Performance)Reported Device Performance
    Biocompatibility: Non-cytotoxic, non-irritating, non-sensitizingUser-contacting materials shown to be non-cytotoxic, non-irritating, and non-sensitizing.
    Electrical Safety: Compliance with IEC 60601-1, IEC 62133-2, IEC 60086-5, IEC 60601-1-11Testing conducted in accordance with specified IEC standards.
    Electromagnetic Compatibility: Compliance with IEC 60601-1-2Testing conducted in accordance with IEC 60601-1-2.
    Software: Basic Documentation level as per FDA guidanceSoftware evaluated at the Basic Documentation level.
    Vacuum Level Verification: Meet mode/cycle specificationsDevices meet mode/cycle specifications.
    Backflow Protection: No liquid backflow into tubingLiquid does not backflow into the tubing.
    Use Life: Maintains specifications throughout proposed use lifeDevice maintains its specifications throughout its proposed use life.
    Battery Performance: Battery functional during stated use-lifeBattery remains functional during its stated use-life.
    Battery Status Indicator: Functional during stated battery lifeBattery status indicator remains functional during its stated battery life.
    Peak Suction Levels: Between -40 and -290 mmHgDevices are capable of producing peak suction levels between -40 and -290 mmHg.
    Cycles per Minute: Between 20 and 120 cycles per minuteDevices are capable of producing speeds between 20 and 120 cycles per minute.

    2. Sample size used for the test set and the data provenance

    The document does not specify sample sizes for the performance tests (e.g., how many units were tested for vacuum levels, how many use-life cycles). Data provenance details (country of origin, retrospective/prospective) are not provided as these are primarily engineering performance tests rather than clinical studies on patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. The ground truth for these performance tests is based on engineering specifications and adherence to international standards, not expert clinical consensus.

    4. Adjudication method for the test set

    Not applicable. This is not a study requiring adjudication of clinical findings.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is an electric breast pump, a mechanical medical device, and does not involve AI or human readers for diagnostic interpretation.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This device does not use algorithms for diagnostic or interpretive performance separate from its mechanical function.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth for the performance testing is based on:

    • Engineering specifications: Pre-defined ranges for suction levels, cycles per minute, and other functional parameters.
    • International standards: Compliance with standards like ISO 10993 for biocompatibility, IEC 60601 for electrical safety, and IEC 60601-1-2 for EMC.
    • Design requirements verification: Ensuring the device meets its intended design functions (e.g., backflow protection, battery function).

    8. The sample size for the training set

    Not applicable. There is no "training set" in the context of this 510(k) submission for a non-AI/ML device.

    9. How the ground truth for the training set was established

    Not applicable. (See #8)

    Ask a Question

    Ask a specific question about this device

    K Number
    K231303

    Validate with FDA (Live)

    Date Cleared
    2024-01-26

    (266 days)

    Product Code
    Regulation Number
    870.1130
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Arm-type Fully Automatic Digital Blood Pressure Monitors are intended to measure (systolic and diastolic) and pulse rate of adults and adolescents over 12 years of age with circumference ranging from 22cm to 42cm.

    Device Description

    The Arm-type Fully Automatic Digital Blood Pressure Monitor (BPM) series, noninvasive, blood pressure measurement system for over-the-counter (OTC) use in home and clinical environment. The systolic and diastolic pressures are determined using the oscillometric method, where the cuff is inflated with an integral controllable piezoelectric pump and deflates via an electric automatic rapid deflation valve. During measurements, the electric pump within the main unit slowly inflates the arm cuff, generating cuff pressure which is monitored and from which pulse waveform data is extracted. This waveform data is analyzed by software algorithms within the microprocessor to determine pulse rate, systolic pressure, and diastolic pressure. The measure pressure range from 0 to 299mmHg, and the pulse rate range from 30 to 180 beats/min.

    The pulse rate measurement compares the longest and the shortest time intervals of detected pulse waves to the mean time interval and displays a warning signal with the reading to indicate the detection of irregular heartbeat when the difference of the time intervals is over 25%.

    Meanwhile, this blood pressure monitor device can be used as a stand-alone unit to finish the blood pressure measurement or in conjunction with the APP through embed a 2.4GHz BLE module that allow users to connect with nearby BT receiving terminal. Once measurement is over, the LCD of the device displays results. And the device will start to transmit data to the pair-up terminal automatically. This app is only intended to display measurement results from the blood pressure monitor device, which does not provide any diagnostic or measurement functions, and does not interpret or analyze the data for medical decision making. Unlimited readings can be stored in the app for archiving and review by the user.

    AI/ML Overview

    The provided text describes the 510(k) submission for the "Arm-type Fully Automatic Digital Blood Pressure Monitor, Model: BM 92". The main study used to demonstrate the device meets acceptance criteria is a clinical investigation following the ISO 81060-2:2018+AMD2020 standard, which focuses on the clinical investigation of intermittent automated measurement type non-invasive sphygmomanometers.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implicitly defined by compliance with the ISO 81060-2:2018+AMD2020 standard, which specifies accuracy requirements for blood pressure monitors. The document states:
    "The results showed the accuracy of the blood pressure monitor is within acceptable scope specified in ISO 81060-2:2018+AMD2020."

    While specific numerical acceptance criteria (e.g., mean difference and standard deviation between device and reference measurements) are not explicitly stated in the document, the adherence to this ISO standard implies meeting its criteria. The general accuracy specifications are given in the comparison table:

    Criteria CategoryAcceptance Criteria (Implied by ISO 81060-2:2018+AMD2020)Reported Device Performance (Summary)
    Blood Pressure AccuracyWithin acceptable scope specified in ISO 81060-2:2018+AMD2020."The results showed the accuracy of the blood pressure monitor is within acceptable scope specified in ISO 81060-2:2018+AMD2020."
    Static Pressure Accuracy± 3mmHg± 3mmHg
    Pulse Accuracy± 5%± 5%

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: "A total of 90 patients participated in the clinical study (2 were excluded because they were part of other clinical trials)." So, the effective sample size for the test set was 88 patients.
    • Data Provenance: The document does not explicitly state the country of origin. However, the manufacturer is JOYTECH Healthcare Co., Ltd. located in Hangzhou, Zhejiang, China. Given this, it's highly probable the study was conducted in China. The study was a prospective clinical investigation.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document states: "A manual Mercury Sphygmomanometer was used as a reference device." This implies that the ground truth was established by human readers taking measurements using a mercury sphygmomanometer. However, the number of experts and their qualifications are not specified in the provided text. The ISO 81060-2 standard typically requires trained observers for reference measurements, but the details are omitted here.

    4. Adjudication Method for the Test Set

    The text mentions: "Same arm sequential method was adopted during the clinical testing. A manual Mercury Sphygmomanometer was used as a reference device." This suggests a comparative method, but it does not describe an adjudication method for discrepancies, which is usually relevant for subjective interpretations (e.g., image reading). For blood pressure measurements, the reference device readings are typically considered the objective ground truth; thus, a formal "adjudication" between multiple observers might not be explicitly detailed if the standard protocol for reference measurement was followed.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is typically performed for AI-assisted image interpretation devices to assess the impact of AI on human reader performance. The device described is an automated blood pressure monitor, not an AI interpretation system. Its function is to directly measure and display blood pressure, not to assist human readers in interpreting complex data.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Yes, in essence, the device's accuracy was tested in a standalone manner. The device "determines" systolic and diastolic pressures using its internal algorithms (oscillometric method). The clinical study directly evaluated the accuracy of these measurements against a reference standard (manual mercury sphygmomanometer). The performance cited ("The results showed the accuracy of the blood pressure monitor is within acceptable scope specified in ISO 81060-2:2018+AMD2020") refers to the device's direct measurement capability.

    7. The Type of Ground Truth Used

    The ground truth used was comparative measurements against a manual Mercury Sphygmomanometer. The text states: "A manual Mercury Sphygmomanometer was used as a reference device." This is a standard method for establishing ground truth for automated blood pressure monitors.

    8. The Sample Size for the Training Set

    The document does not specify the sample size for a training set. This is common for traditional medical devices like blood pressure monitors, where "training" in the machine learning sense isn't explicitly detailed. The device likely relies on established oscillometric principles and algorithms developed over time, not a specific "training set" of patient data in the context of deep learning models.

    9. How the Ground Truth for the Training Set Was Established

    As no specific "training set" for an AI model is mentioned or implied, the question of how its ground truth was established is not applicable in the context of the provided document. The device utilizes an "oscillometric method" and "software algorithms within the microprocessor," which are based on known physiological principles and traditional signal processing rather than data-driven machine learning training requiring a designated ground-truth dataset.

    Ask a Question

    Ask a specific question about this device

    K Number
    K213189

    Validate with FDA (Live)

    Date Cleared
    2022-03-01

    (153 days)

    Product Code
    Regulation Number
    870.1130
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Wrist-type Fully Automatic Digital Blood Pressure Monitors are intended to measure blood pressure (systolic and diastolic) and pulse rate of adults and adolescents over 12 years of age.

    Device Description

    The Wrist-type Fully Automatic Digital Blood Pressure Monitor is consist of sphygmomanometer main body and cuff. The mainbody contains ABS housing, LCD, keys, measurement control module, pneumatic control module, power supply module. The Wrist-type Fully Automatic Digital Blood Pressure Monitor is automatic, non-invasive, blood pressure measurement system for over-the-counter (OTC) use in home and clinical environment. The device with an inflatable cuff wrapping around the patient's wrist and the cuff is inflated automatically by an internal pump in the device. The systolic and diastolic blood pressures are determined by oscillometric method and silicon integrate pressure sensor technology. The deflation rate is controlled by a preset mechanical valve at a constant rate. The pressure of the cuff is completely released automatically at the end of the measurement. At the same time, the measurements are displayed on the LCD display for three minutes. The cuff can measure pressure range from 0 to 300mmHg, and the pulse rate range from 30 to 180 beats/min.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are derived from the ISO 81060-2:2018+AMD2020 standard for clinical investigation of automated noninvasive sphygmomanometers.

    Acceptance Criteria (from ISO 81060-2:2018+AMD2020)Reported Device Performance (Model DBP-2208, Cuff 13.5-21.2cm)Reported Device Performance (Model DBP-2242, Cuff 12.5-23cm)
    Method 1:
    Average difference for Systolic Blood Pressure ≤ ±5 mmHg0.84 mmHg0.87 mmHg
    Standard deviation for Systolic Blood Pressure ≤ 8 mmHg5.25 mmHg5.22 mmHg
    Average difference for Diastolic Blood Pressure ≤ ±5 mmHg0.21 mmHg0.20 mmHg
    Standard deviation for Diastolic Blood Pressure ≤ 8 mmHg4.92 mmHg4.93 mmHg
    Method 2:
    Average difference for Systolic Blood Pressure ≤ ±5 mmHg0.84 mmHg0.87 mmHg
    Standard deviation for Systolic Blood Pressure ≤ Maximum permissible standard deviation (based on mean difference, from Table 1 in document)4.52 mmHg (<6.95mmHg)4.48 mmHg (<6.887mmHg)
    Average difference for Diastolic Blood Pressure ≤ ±5 mmHg0.21 mmHg0.20 mmHg
    Standard deviation for Diastolic Blood Pressure ≤ Maximum permissible standard deviation (based on mean difference, from Table 1 in document)4.36 mmHg (<6.879 mmHg)4.38 mmHg (<6.95 mmHg)

    Conclusion: Both tested models (DBP-2208 and DBP-2242) met all specified acceptance criteria according to both Method 1 and Method 2 of ISO 81060-2:2018+AMD2020.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: 85 patients for each clinical investigation (one for cuff 13.5-21.2cm using DBP-2208 as representative, and one for cuff 12.5-23cm using DBP-2242 as representative). Each group comprised 43 females and 42 males.
    • Data Provenance: The document does not explicitly state the country of origin. It describes clinical studies conducted on "patients" and "subjects" indicating prospective data collection. "All the subjects were volunteer to take part in the clinical study, all the subjects completed the clinical study without any AE or side-effect."

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document states that a "manual Mercury Sphygmomanometer was used as a reference device" to establish the ground truth. It does not explicitly mention the number of experts or their specific qualifications (e.g., "radiologist with 10 years of experience"). However, for a clinical study involving a mercury sphygmomanometer as the reference, it is implied that trained medical professionals would perform these measurements. The standard ISO 81060-2 typically requires measurements by qualified observers.

    4. Adjudication Method for the Test Set

    The adjudication method described is the "Same wrist sequential method." This means the device under test and the reference device (manual Mercury Sphygmomanometer) measurements were taken sequentially on the same wrist. The document does not specify a separate "adjudication" process involving multiple experts reviewing discrepancies, but rather a direct comparison to the reference device.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The study focuses on the accuracy of the device itself compared to a reference standard, not on how human readers' performance might improve with AI assistance. This is a standalone device accuracy study.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    Yes, this was a standalone performance study. The "Wrist-type Fully Automatic Digital Blood Pressure Monitor" is an automated device designed to measure blood pressure. The clinical investigations directly assessed the device's accuracy against a manual reference without human intervention in the measurement process (other than proper placement and initiation).

    7. Type of Ground Truth Used

    The ground truth used was established by a manual Mercury Sphygmomanometer. This is considered a gold standard for blood pressure measurement in clinical studies.

    8. Sample Size for the Training Set

    The document does not provide any information about a training set. This is a validation study for a medical device's accuracy, not an AI/machine learning algorithm that typically requires a training set. The device determines blood pressure using an oscillometric method based on pre-programmed algorithms and hardware.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as no training set for an AI/machine learning algorithm is mentioned in the document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K200649

    Validate with FDA (Live)

    Date Cleared
    2020-08-03

    (144 days)

    Product Code
    Regulation Number
    870.1130
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Fully Automatic Blood Pressure Monitors are intended to measure blood pressure (systolic and diastolic) and pulse rate of adults and adolescents age 12 through 21 years of age.

    Device Description

    The Arm-Type series use an inflatable cuff which is wrapped around the patient's upper arm. The cuff is inflated automatically by an internal pump in the device. The systolic and diastolic blood pressures are determined by oscillometric method and silicon integrate pressure sensor technology. The deflation rate is controlled by a preset mechanical valve at a constant rate. The pressure of the cuff is completely released automatically at the end of the measurement. At the same time, the measurements are displayed on the LCD display for three minute. There is a maximum pressure safety setting at 300 mmHg. The device will not inflate the cuff higher than 300 mmHg. An irregular heartbeats rhythm is defined as a rhythm that varies by 25% less or more than the average rhythm detected while the monitor is measuring the systolic and diastolic blood pressure. AND, these subject device can be used as a stand-alone unit to finish the blood pressure measurement or in conjunction with the "JOYTECH healthcare" APP through the embed a 2.4GHz BLE module that allow users to connect with nearby BT receiving terminal. Once measurement is over, the LCD of the device displays results. And the device will start to transmit data to the pair-up terminal automatically. This app is only intended to display trend graphs of measured systolic and diastolic blood pressure, and pulse rate, does not provide any diagnostic or measurement functions, and does not interpret or analyze the data for medical decision making. Unlimited readings can be stored in the app for archiving and review by the user.

    AI/ML Overview

    The provided text is a 510(k) Summary for a blood pressure monitor and does not contain detailed information about a specific study proving the device meets acceptance criteria. Instead, it states that clinical testing was not required because the device is considered equivalent to a previously cleared predicate device (Joytech DBP-1307, K173024) due to having the same core algorithm and cuff circumference.

    However, the summary does list the standards that were utilized for testing to demonstrate safety and effectiveness. The most relevant standard for performance claims is ISO 81060-2:2013, which covers the clinical validation of automated non-invasive sphygmomanometers. The document states that the clinical validation was performed according to ISO 81060-2:2013 for the predicate device.

    Therefore, the information I can extract regarding the acceptance criteria and a study demonstrating compliance is limited. I will present the information available based on the provided text, primarily focusing on what ISO 81060-2:2013 generally entails for blood pressure monitors, as the detailed study specific to this device is not present but is referenced through the predicate.

    Here's the breakdown of the information requested, based on the provided text and general knowledge of ISO 81060-2:2013:

    1. Table of acceptance criteria and the reported device performance

    Since the document states "clinical testing was not required to establish equivalency of the device" because it is based on the predicate device's validation per ISO 81060-2:2013, the acceptance criteria and performance would be those defined by this standard. The document does provide performance specifications for the device, which are implicitly stated to meet the requirements of the standards listed, including ISO 81060-2:2013 via the predicate.

    Acceptance Criteria (from ISO 81060-2:2013 for clinical validation)Reported Device Performance (from this device's specifications)
    Systolic Blood Pressure (SBP) & Diastolic Blood Pressure (DBP) Accuracy (mean difference): Mean difference between device measurement and auscultatory reference method for SBP and DBP should be ≤ ±5 mmHg.Static Pressure: ±3 mmHg
    Systolic Blood Pressure (SBP) & Diastolic Blood Pressure (DBP) Accuracy (standard deviation): Standard deviation of the differences should be ≤ 8 mmHg.Not explicitly stated with comparison to auscultatory reference, but implied to meet standard through static pressure accuracy.
    Pulse Rate Accuracy: Accuracy within a specified range (typically ±5% or ±5 beats/min)Pulse: ±5%
    Maximum Cuff Pressure: Safety limit for inflation pressure300 mmHg

    2. Sample size used for the test set and the data provenance

    The document explicitly states that "clinical testing was not required" for the new device because it is considered equivalent to the predicate device K173024 (Joytech DBP-1307), which had undergone clinical validation.

    • Test Set Sample Size: Not provided in the document for the new device. For the predicate device, a clinical validation study adhering to ISO 81060-2:2013 would typically require a minimum of 85 subjects with specific distributions of age, sex, and blood pressure ranges.
    • Data Provenance: Not provided for the new device. For the predicate device's clinical validation, it would have been prospective clinical data. The country of origin is not specified but is typically mentioned in clinical study reports.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information pertains to the clinical validation of the predicate device (DBP-1307, K173024) under ISO 81060-2:2013.

    • Number of Experts: For ISO 81060-2:2013, ground truth is typically established by three trained observers taking sequential auscultatory measurements according to a defined protocol.
    • Qualifications of Experts: The standard requires trained observers who are blinded to the device readings and are experienced in taking auscultatory blood pressure measurements. Specific medical qualifications (e.g., cardiologist, certified BP measurement technician) are usually detailed in the study protocol. The document does not specify their exact qualifications but implies they meet the standard's requirements.

    4. Adjudication method for the test set

    For the clinical validation of the predicate device under ISO 81060-2:2013:

    • Adjudication Method: The standard specifies a method where the three observers conduct measurements. The ground truth (reference blood pressure) is derived from the average of the two closest readings among the three observers, or if there's a significant disagreement, specific rules are followed for exclusion or further measurement. This is a form of consensus-based ground truth.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • MRMC Study: No, an MRMC comparative effectiveness study is not applicable to this device. This is a standalone blood pressure monitor, not an AI-assisted diagnostic imaging device requiring human reader interpretation. The device provides automatic blood pressure readings.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Standalone Performance: Yes, the fundamental performance of the blood pressure measurement algorithm is evaluated in standalone mode as part of the clinical validation process (for the predicate device). The device itself performs the measurement and displays the result without direct human interpretation of waveforms for diagnosis. While a human initiates the measurement and reads the display, the accuracy assessment is of the device's algorithmic output against a reference.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • Type of Ground Truth: For the clinical validation of the predicate device (following ISO 81060-2:2013), the ground truth for blood pressure measurements is established by expert auscultatory measurements (Korotkoff sounds) by trained observers, as stated in point 3. This is a form of expert consensus derived from a gold standard manual method.

    8. The sample size for the training set

    • Training Set Sample Size: Not applicable/Not provided. The document describes a medical device leveraging an established oscillometric method for blood pressure measurement, not a machine learning or AI model that requires a distinct training set in the conventional sense. The "core algorithm" of the device is validated against the predicate and established standards, rather than "trained" on a dataset in the way an AI model would be.

    9. How the ground truth for the training set was established

    • Ground Truth for Training Set: Not applicable. As mentioned, this is not an AI/ML device that uses a distinct training set. The underlying oscillometric algorithm's principles are based on known physiological responses and signal processing, not on labeled training data. The validation of its performance (for the predicate device) uses auscultatory reference measurements as ground truth.
    Ask a Question

    Ask a specific question about this device

    K Number
    K200599

    Validate with FDA (Live)

    Date Cleared
    2020-07-08

    (121 days)

    Product Code
    Regulation Number
    880.2910
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Digital Thermometer DMT-4756 is intended to measure the human body temperature in regular mode orally, rectally or under the arm, and the device is reusable for clinical or home use on people of all ages, including children under 8 with adult supervision.

    Device Description

    The digital thermometer DMT-4756 is hand held device which can measure human body's temperature orally, axillary (under the arm), and rectally. The results can be displayed on LCD. The flexible tip is foldable,when the flexible tip folded, DMT-4756 is used to measure human body's temperature rectally, while the flexible tip unfolded, it is used to measure temperature orally and axillary (under the arm). The digital thermometer have several functions,such as beep alarm,unit switchable, low battery detection,memories,backlight,auto power off functions. DMT-4756 is a predictive digital thermometer. Users only need 10 seconds for predictive quick read and 30 seconds for final temperature readings.

    AI/ML Overview

    This document is a 510(k) summary for the Joytech Healthcare Co., Ltd. Digital Thermometer, Model DMT-4756. It outlines the device's characteristics, its comparison to a predicate device, and the testing conducted to demonstrate its substantial equivalence.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Feature/TestAcceptance Criteria (from predicate or standard)Reported Device Performance (DMT-4756)
    Accuracy (Temperature)±0.1°C between 35.5°C to 42.0°C (±0.2°F, 95.9°F-107.6°F) ±0.2°C under 35.5°C or over 42.0°C (±0.4°F under 95.9°F or over 107.6°F)Identical to predicate device. Clinical performance test protocol and data analysis conducted in accordance with ISO 80601-2-56. "The test report showed the clinical performance of the subject device complied with the requirement of ISO 80601-2-56."
    Measurement Range32.0°C~43.9°C (89.6°F-111.0°F)Identical to predicate device.
    Display Resolution0.1 °C/0.1 °FIdentical to predicate device.
    Measurement TimePredicate: Approximate 5 ~ 10sDMT-4756: 10s Predictive read, 30s Final read. "Performance test including clinical test shows the subject device complies with performance standard."
    Predictive ModePredicate: OptionalDMT-4756: Yes (has this function). "Performance test including clinical test shows the subject device complies with performance standard."
    Construction (Foldable Probe)Predicate: Flexible Digital ThermometerDMT-4756: The probe is foldable into the body of the thermometer. "Performance testing shows the subject device complies with standard ISO 80601-2-56."
    BiocompatibilityComply with ISO 10993-5 and ISO 10993-10Complied with ISO 10993-5 and ISO 10993-10. "Biocompatibility test and performance bench test met the requirements in the standards."
    Electrical SafetyComplied with IEC 60601-1 / AAMI / ANSI ES60601-1:2005/(R)2012 And A1:2012,C1:2009/(R)2012 And A2:2010/(R)2012, ISO 80601-2-56:2017, ASTM E1112:00(Reapproved 2011)Complied with IEC 60601-1 / AAMI / ANSI ES60601-1:2005/(R)2012 And A1:2012,C1:2009/(R)2012 And A2:2010/(R)2012, ISO 80601-2-56:2017, ASTM E1112:00(Reapproved 2011). "Performance testing was conducted to validate and verify that Digital Thermometer, DMT-4756 met all requirements of related international standards, including electrical safety... Results of these tests demonstrate compliance to the requirements of the below consensus standards."
    Electromagnetic Compatibility (EMC)Complied with IEC 60601-1-2:2014Complied with IEC 60601-1-2:2014. "Performance testing was conducted to validate and verify that Digital Thermometer, DMT-4756 met all requirements of related international standards, including ... EMC... Results of these tests demonstrate compliance to the requirements of the below consensus standards."
    Home-used medical equipment requirementsComplied with IEC 60601-1-11:2015Complied with IEC 60601-1-11:2015. "Performance testing was conducted to validate and verify that Digital Thermometer, DMT-4756 met all requirements of related international standards... Results of these tests demonstrate compliance to the requirements of the below consensus standards."
    Software Validation"Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices," dated May 11, 2005"The software/firmware verification and validation was provided in accordance with the 'Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices,' dated May 11, 2005."

    2. Sample size used for the test set and the data provenance:

    • Sample Size: 450 subjects for the clinical tests. These were divided into three groups:
      • Infants: newborn to one year
      • Children: greater than one to five years
      • Adults: greater than five years old
    • Data Provenance: The document does not explicitly state the country of origin or whether the data was retrospective or prospective. It only mentions "The clinical tests were conducted on the DMT-4756 for each measurement site." Without further information, it's not possible to determine the geographical origin of the subjects. The description of the test as "clinical tests were conducted" implies a prospective study.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • The document does not provide information on the number of experts used or their qualifications for establishing ground truth. The clinical performance was evaluated against the requirements of ISO 80601-2-56, which are standards for clinical thermometers.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • The document does not describe any adjudication method. The clinical performance was evaluated against a standard (ISO 80601-2-56) rather than against expert consensus on individual cases.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • This information is not applicable. The device is a digital thermometer, not an AI-assisted diagnostic tool that would involve human readers.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • This information is not applicable in the context of a digital thermometer. The device itself is standalone in its function of measuring temperature. The clinical tests evaluate the device's accuracy in direct measurement.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For the clinical tests, the ground truth is implicitly established by the performance standard ISO 80601-2-56. This standard defines the acceptable accuracy limits for clinical body temperature measurement devices. The study aimed to show that the device's readings fell within these established accuracy criteria.

    8. The sample size for the training set:

    • This information is not applicable. This is a traditional medical device (digital thermometer) and does not involve machine learning or AI models that require a separate "training set" in the conventional sense. The device's calibration and internal algorithms are developed through engineering and testing, not by training on a dataset of patient readings.

    9. How the ground truth for the training set was established:

    • This information is not applicable, as there is no "training set" for an AI model in this context.
    Ask a Question

    Ask a specific question about this device

    K Number
    K190886

    Validate with FDA (Live)

    Date Cleared
    2019-07-29

    (116 days)

    Product Code
    Regulation Number
    870.1130
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Measure blood pressure (systolic and diastolic) and pulse rate of adults and adolescents age 12 through 21 years of age.

    Device Description

    The Arm-Type Fully Automatic Digital Blood Pressure Monitor (Models: DBP-1204,DBP-1307,DBP-1332,DBP-1305,DBP-1334,DBP-1231,DBP-1314,DBP-1209,DBP-1303,DBP-1335) uses an inflatable cuff which is wrapped around the patient's upper arm. The cuff circumference is limited to 22cm-48cm (be comprised of 3 cuffs, 22cm-36cm,22cm-42cm,32cm-48cm,select one. And the model DBP-1335 is comprised of 4 cuffs,22cm-36cm,22cm-42cm,32cm-48cm,33-43cm, select one.) which is inflated automatically by an internal pump in the device. The systolic and diastolic blood pressures are determined by oscillometric method and silicon integrate pressure sensor technology. The deflation rate is controlled by a preset mechanical valve at a constant rate. The pressure of the cuff is completely released automatically at the end of the measurement. At the same time, the measurements are displayed on the LCD display for three minute. There is a maximum pressure safety setting at 300 mmHg. The device will not inflate the cuff higher than 300 mmHg.

    AI/ML Overview

    The provided document is a 510(k) Summary for a blood pressure monitor, not an AI/ML device. Therefore, many of the requested elements for acceptance criteria and study design for an AI/ML device (such as the number of experts used to establish ground truth, adjudication methods, MRMC studies, effect size of human readers with AI assistance, sample size for training set, and how ground truth for the training set was established) are not applicable and are not present in this document.

    However, I can extract the relevant information regarding the clinical validation for the device's accuracy.

    1. A table of acceptance criteria and the reported device performance

    The acceptance criteria are derived from ISO 81060-2:2013, which specifies accuracy requirements for non-invasive sphygmomanometers. While the document states that the device complies with these requirements, it does not provide a table with specific numerical thresholds for acceptance criteria and the exact reported performance of the device against these thresholds. It only states that the clinical tests "comply with the accuracy requirements of ISO 81060-2 Second edition 2013-05-01."

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not explicitly state the specific sample size used for the clinical validation. It mentions that clinical tests were performed for cuff sizes 22-42cm and 32-48cm. It does not provide information on the country of origin or whether the study was retrospective or prospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable, as this is a blood pressure monitor, not an AI/ML diagnostic device requiring expert interpretation of images. The "ground truth" for blood pressure accuracy in such studies is typically established by trained observers using standardized auscultatory methods simultaneously with the device under test. The document does not specify the number or qualifications of these observers.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable in the context of a blood pressure monitor accuracy study. The ISO 81060-2 standard outlines procedures for simultaneous measurements by trained observers and the device.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI/ML diagnostic device with human-in-the-loop performance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, a standalone performance assessment was done for the blood pressure monitor. Its accuracy was evaluated against the requirements of ISO 81060-2.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The ground truth for blood pressure measurements is established by trained observers using the auscultatory method (manual sphygmomanometry) as per the protocols outlined in standards like ISO 81060-2.

    8. The sample size for the training set

    Not applicable. This is not an AI/ML device that requires a training set in the conventional sense. The device's algorithms for blood pressure measurement are based on established oscillometric principles, not trained on a dataset.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set in the context of an AI/ML model for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K190873

    Validate with FDA (Live)

    Date Cleared
    2019-07-26

    (114 days)

    Product Code
    Regulation Number
    880.2910
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Infrared Ear/Forehead Thermometer DET-218 is intended for the intermittent measurement of human body temperature by people of all ages. The devices are reusable for home use only.

    Device Description

    The Infrared Ear/Forehead Thermometer is a hand-held, battery powered device designed to measure human body temperature from the auditory canal or forehead. Its operation is based on measuring the natural thermal radiation from the tympanic membrane or forehead.

    The Infrared Ear/Forehead Thermometer use a thermopile sensor with integrated thermistor for the target reading, a thermistor mounted in the head of the thermometer for ambient temperature readings, a thermopile collect, the infrared energy emitted from the forehead.The results can be displayed on LCD. And, the results measured by DET-218 can also be transmitted to mobile device (i.e. iPhone) with Bluetooth and the application APP installed on a mobile device using the iOS system or the Android system. The thermometer is reusable for home use on people of all ages with adult supervision. The device can be used with as a stand-alone device or in conjunction with the App on a compatible BLE enabled smartphone.

    The purpose of data transfer to a mobile device is convenient for users to view the measured data through the APP. The APP operating system is IOS 9.0+ system , Android 5.0+ system.

    The name of APP is "Healthforyou",it adopts AES+RSA combination to encrypt and decrypt interface parameters and the data transmission between APP and server adopts HTTP+SSL mode to ensure the security of information.

    The App only visualizes the Measurements from the device via. bluetooth to smart phone. The APP receives and records data, and displays measurement on smart phone in correspondence with DET-218 Measurement site (Ear/Forehead) set. The App has the following functions:

    l. data recording;

    II. curve display;

    III. alarm functions.

    The light bar of DET-218 turns on green light when the measurement is < 38.0 ℃ (100.4 °F ). The alarm function refers to the light bar of the device DET-218 turns on red light when the measurement is ≥38.0 ℃ (100.4 下 ), the APP will give an alarm when the measurement received simultaneously from DET-218 displays≥38.0℃.

    AI/ML Overview

    Here's an analysis of the provided text to extract the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the infrared ear/forehead thermometerDET -218 are established by relevant international standards, particularly ASTM E1965-98 (2016) for clinical performance. The reported device performance is stated to comply with these requirements.

    Acceptance Criteria CategorySpecific Standard/RequirementReported Device Performance
    Accuracy (Ear/Forehead mode)ASTM E1965-98 (2016) defines accuracy requirements for infrared thermometers.±0.2°C (0.4°F) during 35.5°C ~ 42.0°C (95.9°F ~ 107.6°F) at 15°C ~ 35°C (59.0°F ~ 95.0°F) operating temperature range.±0.3°C (0.5°F) for other measuring and operating temperature range.
    Electrical Safety and PerformanceAAMI / ANSI ES60601-1:2005/(R)2012 And A1:2012,C1:2009/(R)2012 And A2:2010/(R)2012 Medical Electrical EquipmentISO 80601-2-56:2017 Medical electrical equipment Part 2-56 Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurementASTM E 1965-98 Standard Specification for Infrared Thermometers for Intermittent Determination of Patient TemperatureResults of these tests demonstrate compliance to the requirements of the standards.
    Home-Used Medical Equipment & EnvironmentalIEC 60601-1-11:2015 General requirements for basic safety and essential performance -- Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environmentCompliance demonstrated.
    Electromagnetic Compatibility (EMC)IEC 60601-1-2:2014 Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and testsEN300328:Electromagnetic compatibility and Radio spectrum Matters (ERM); Wideband transmission systems; Data transmission equipment operating in the 2,4 GHz ISM band and using wide band modulation techniques;ETSI EN 301489-1: Electromagnetic compatibility and Radio spectrum Matters(ERM); ElectroMagnetic Compatibility (EMC)standard for radio equipment and services;Part 1:Common technical requirements;ETSI EN 301489-17: Electromagnetic compatibility and Radio spectrum Matters (ERM);ElectroMagnetic Compatibility (EMC) standard for radio equipment; Part 17: Specific conditions or Broadband Data Transmission Systems;Compliance demonstrated; FCC, EN301489-1, EN301489-17, EN 300328 and wireless coexistence test passed for the Bluetooth transmission.
    BiocompatibilityISO 10993-5:2009 Biological evaluation of medical devices -- Part 5: Tests for in vitro cytotoxicityISO10993-10:2010 Biological evaluation of medical devices -- Part 10: Tests for irritation and skin sensitizationCompliance demonstrated.
    Software/Firmware Verification and ValidationGuidance for the Content of Premarket Submissions for Software Contained in Medical Devices, dated May 11, 2005.Software Validation Passed (for memory set changes and lack of voice function).
    Response TimeImplied functional requirement (not explicitly given as an acceptance criterion from a standard, but noted as a difference from the predicate).Approx. 1s (shorter than predicate, and all tests passed, indicating this difference "does not raise new performance questions").

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Clinical Performance Test: 150 subjects.
    • Data Provenance: The document does not explicitly state the country of origin for the clinical test data. It only mentions that "clinical tests were conducted on the DET-218." It is reasonable to assume it was a prospective study, as it was conducted to evaluate the device's performance against a standard. The study included three groups based on age:
      • Infants: newborn to one year
      • Children: greater than one to five years
      • Adults: greater than five years old

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not provide information on the number of experts used or their qualifications for establishing ground truth in the clinical performance test. The ground truth method is described below.

    4. Adjudication Method for the Test Set

    The document does not describe an adjudication method. The clinical performance test protocol and data analysis were conducted as per the requirements of ASTM E1965-98 (2016). This standard typically outlines the methods for comparison against a reference thermometer, implicitly defining the ground truth measurement process rather than an adjudication among multiple readings of the device under test.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The device is a thermometer, and its performance is evaluated against reference measurements rather than subjective human interpretation of cases.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, a standalone performance evaluation was done. The entire summary focuses on the device (Infrared Ear/Forehead Thermometer DET-218) and its conformance to standards. The "clinical tests" described are for the device itself, not for an algorithm where human input is part of the performance metric. The device also has an associated app, but the app's function is for data visualization, recording, and alarm, not for independent diagnostic or measurement capabilities requiring a human in the loop for its core function.

    7. The Type of Ground Truth Used

    The ground truth for the clinical performance test was established by a reference measurement method as defined by ASTM E1965-98 (2016). This typically involves comparing the device's readings against highly accurate reference thermometers or established clinical measurement techniques to determine accuracy. It is not expert consensus, pathology, or outcomes data in the usual sense for diagnostic imaging or similar devices.

    8. The Sample Size for the Training Set

    The document describes a clinical performance test, which is a test set evaluation. It does not mention a "training set" or any machine learning model training process for the core measurement functionality of the thermometer. The device's operation is based on physical principles (measuring thermal radiation), not a data-trained algorithm for its primary measurement.

    9. How the Ground Truth for the Training Set Was Established

    As no training set is mentioned for the primary measurement function of the device, this question is not applicable based on the provided text.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 2