Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K214017

    Validate with FDA (Live)

    Date Cleared
    2022-07-14

    (204 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Examination gloves-Type A (Latex gloves) is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

    Device Description

    The proposed device, The Examination gloves-Type A ( Latex gloves ) is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. The proposed device is made of natural rubber latex and meet all the current specifications listed under the ASTM Specification D3578-19, Standard Specification for Rubber Examination Gloves. The principal operation and mechanism of this device is to prevent contamination between patient and examiner and this principle is achieved through testing of barrier, physical properties and other testing stated in the performance data. This device is provided non-sterile and the shelf life is 5 years.

    AI/ML Overview

    This document is a 510(k) summary for Examination gloves-Type A (Latex gloves). It details the device's characteristics, intended use, and a comparison to a predicate device, along with non-clinical performance test data to demonstrate substantial equivalence.

    Here's an analysis of the provided information, focusing on acceptance criteria and study details:

    1. A table of acceptance criteria and the reported device performance

    Test ItemTest StandardAcceptance CriteriaReported Device PerformanceConclusion
    DimensionASTM D3578-19Length: Min 230 mm for all sizes. Width: Small: 80 ± 10 mm, Medium: 95 ± 10 mm, Large: 111 ± 10 mm. Thickness: Palm: 0.08 mm min, Finger: 0.08 mm min.Length: S: 240-244mm, M: 241-245mm, L: 242-245mm. Width: S: 84-86mm, M: 96-98mm, L: 105-108mm. Thickness: Palm: 0.11-0.13mm, Finger: 0.13-0.14mm.Pass
    Physical Properties - Tensile StrengthASTM D3578-19Before Ageing: 18 MPa Minimal for all sizes. After Ageing: 14 MPa Minimal for all sizes.Before Ageing: 20.6 MPa Min. After Ageing: 18.9 MPa Min.Pass
    Physical Properties - Ultimate ElongationASTM D3578-19Before Ageing: 650% Min for all sizes. After Ageing: 500% Min for all sizes.Before Ageing: 683% Min. After Ageing: 623% Min.Pass
    Watertight testASTM D5151-19Sample size: 200 pcs, Inspection level: GI, AQL 1.5, Acceptance Number 7, Rejection Number 8.Sample size: 200 pcs, Inspection level: GI, AQL 1.5, Result: 0.Pass
    Residual powderASTM D6124-06 (Reapproved 2017)2 mg per glove or less.0.02 mg per glove.Pass
    Protein contentASTM D5712-15200 µg/dm² Max for all sizes.75 µg/g.Pass
    Skin SensitizationISO 10993-10Under the conditions of the study not a sensitizer.Non-sensitizing.Pass
    IrritationISO 10993-10Under the condition of study not an irritant.Non-irritating.Pass
    Systemic toxicityISO 10993-11:2017Under the conditions of study the device extracts do not pose a systemic toxicity concern.Non-systemic toxicity.Pass

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    • Sample Size for Watertight test: 200 pieces (gloves).
    • Sample Size for other tests: The document specifies "Minimal for all sizes" or "for all sizes" without providing explicit sample counts for each specific physical and chemical test.
    • Data Provenance: The document does not explicitly state the country of origin of the data or whether the studies were retrospective or prospective. All submissions are from China.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This device is a physical medical device (examination gloves), not an AI/software device that requires expert adjudication for ground truth related to image interpretation or diagnosis. The "ground truth" for these tests is established by standardized methods and measurements as defined by ASTM and ISO standards, not by expert consensus. Therefore, this question is not applicable in the context of these non-clinical performance tests.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Adjudication methods like 2+1 or 3+1 are typically used for establishing ground truth in AI/imaging studies where there might be inter-reader variability. For the physical and chemical tests conducted on examination gloves, the determination of pass/fail is based on objective measurements against predefined acceptance criteria from international standards. There is no mention of an adjudication method as it is not relevant to this type of testing.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This question is not applicable. The device in question is examination gloves, not an AI-assisted diagnostic tool. Therefore, MRMC studies or assessments of human reader improvement with AI assistance are irrelevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is not applicable. The device is a physical product (examination gloves), not an algorithm or AI system.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for the performance tests on these examination gloves is based on:

    • Standardized measurement and testing protocols: Adherence to established international standards (ASTM and ISO).
    • Objective physical and chemical properties: Measurements of dimensions, tensile strength, elongation, residual powder, protein content, and biological reactivity (skin sensitization, irritation, systemic toxicity).
    • Defined Pass/Fail criteria: Each test has specific numerical or qualitative acceptance criteria derived from the respective standard.

    8. The sample size for the training set

    This question is not applicable. The device is a physical product, not a machine learning model, so there is no "training set."

    9. How the ground truth for the training set was established

    This question is not applicable for the same reason as in point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K211953

    Validate with FDA (Live)

    Date Cleared
    2022-05-25

    (336 days)

    Product Code
    Regulation Number
    878.4460
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Disposable Sterilized Latex Surgical Glove is a device made of nature rubber intended to be worn by operating room personnel to protect a surgical wound from contamination.

    Device Description

    The proposed device, Disposable Sterilized Latex Surgical Gloves is a sterilized and disposable medical glove intended to be worn by operating room personnel to protect a surgical wound from contamination. The proposed device is made of natural rubber latex, per standard ASTM D3577-09(15), the rubber surgical gloves classification is: "Type 1-gloves compounded primarily from nature rubber latex." The gloves are powder-free and available in white in sizes 6, 6.5, 7, 7.5, 8, and 8.5. The proposed device is provided EO sterilized to achieve the sterility Assurance Level (SAL) of 106. The disposable sterilized latex surgical glove shelf life is 3 years.

    AI/ML Overview

    This document is a 510(k) summary for Disposable Sterilized Latex Surgical Gloves, demonstrating substantial equivalence to a predicate device (K171550). It focuses on non-clinical performance data and does not involve AI or clinical studies as typically defined in medical imaging or diagnostic device contexts. Therefore, many of the requested points are not applicable.

    Here's an analysis of the provided information in the context of the requested questions:

    Acceptance Criteria and Device Performance

    This product is a medical device (Disposable Sterilized Latex Surgical Gloves), not an AI algorithm. The "acceptance criteria" and "reported device performance" are based on physical, chemical, and biological properties of the gloves, compared against recognized ASTM and ISO standards.

    Table of Acceptance Criteria and Reported Device Performance:

    Test Method (Standard)Test PurposeAcceptance CriteriaResult
    Dimension
    ASTM D3577-09(Reapproved 2015)Determine length of gloves265mm, minPass
    ASTM D3577-09(Reapproved 2015)Determine width of glovesSize 6: 76 ±6 mm; Size 6.5: 83 ± 6 mm; Size 7: 89 ± 6 mm; Size 7.5: 95 ± 6 mm; Size 8: 102 ± 6 mm; Size 8.5: 108 ± 6 mmPass
    ASTM D3577-09(Reapproved 2015)Determine thickness of glovesPalm: 0.10mm, min; Finger: 0.10mm, min; Cuff: 0.10mm, minPass
    Physical Properties
    ASTM D3577-09(Reapproved 2015)Determine physical properties (Tensile strength)Before Ageing Tensile Strength: 24Mpa Minimal for all sizes; After Ageing Tensile Strength: 18Mpa Minimal for all sizesPass
    ASTM D3577-09(Reapproved 2015)Determine physical properties (Ultimate Elongation)Before Ageing Ultimate Elongation: 750% Min for all sizes; After Ageing Ultimate Elongation: 560% Min for all sizesPass
    Watertight test
    ASTM D5151-19Determine watertightness of glovesSample size: 200 pcs; Inspection level: GI; AQL 1.5; Criteria: ≤8 piecesPass
    Residual powder
    ASTM D6124-06 (Reapproved 2017)Determine residual powder in gloves2 mg per glove or lessPass
    Protein content
    ASTM D5712 - 15Determine extractable protein in gloves200 µg/dm² Max for all sizesPass
    Biocompatibility
    ISO 10993-10Evaluate potential of a test article to cause skin sensitizationUnder the conditions of the study, not a sensitizerPass
    ISO 10993-10Evaluate potential skin irritation caused by test article contact with skin surface of rabbitsUnder the condition of study, not an irritantPass
    ISO 10993-11:2017Evaluate whether leachables extracted from the test article would cause acute systemic toxicity following injection into miceUnder the conditions of study, the device extracts do not pose a systemic toxicity concernPass
    USP 43-NF 38<151>Evaluate risks of febrile reaction in the rabbit to administration by injectionUnder the condition of study, not a pyrogenPass
    USP 43 <85>Bacterial Endotoxin Test<20 EU/ pair of glovesPass
    Sterilization Residuals
    ISO 10993-7:2008Determine if Ethylene Oxide residues of test article are within requirements≤10ug/cm²Pass
    ISO 10993-7:2008Determine if Ethylene Chlorohydrin residues of test article are within requirements≤5mg/cm²Pass

    1. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set:
      • For the Watertight test (ASTM D5151-19), the sample size was 200 pieces.
      • Other tests (e.g., dimensions, physical properties, residual powder, protein content, biocompatibility, sterilization residuals) refer to specific ASTM/ISO standards. While the standards themselves define sampling plans, the document does not explicitly state the exact sample size used for each of these specific tests beyond the watertight test.
    • Data Provenance: The document does not specify the country of origin of the data (other than the manufacturer being in China) or whether the studies were retrospective or prospective. Given the nature of mechanical/chemical testing of medical devices, these are typically prospective laboratory tests conducted to specific standard protocols.

    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not Applicable. This is a medical device (surgical gloves) undergoing non-clinical performance and biocompatibility testing, not an AI algorithm requiring expert ground truthing of images or clinical data. The "ground truth" here is defined by physical, chemical, and biological measurement standards (ASTM, ISO, USP).

    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable. As above, this is non-clinical device testing, not a study involving human interpretation or consensus. The "adjudication" is based on meeting pre-defined numerical or qualitative criteria from recognized standards.

    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable. This document pertains to the clearance of a physical medical device (surgical gloves), not an AI algorithm. Therefore, no MRMC study was conducted.

    5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    • Not Applicable. This document pertains to the clearance of a physical medical device (surgical gloves), not an AI algorithm.

    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • For the non-clinical tests, the "ground truth" is defined by the established criteria within the cited ASTM, ISO, and USP standards. For example:
      • Dimensions are compared to specified minimums/ranges.
      • Tensile strength is compared to specified minimal MPa.
      • Watertightness is determined by meeting an AQL (Acceptance Quality Limit) of 1.5, meaning a certain percentage of defects is acceptable.
      • Biocompatibility (sensitization, irritation, systemic toxicity, pyrogenicity, endotoxin) is determined by standardized biological test methods with defined pass/fail criteria (e.g., "not a sensitizer," "not an irritant," "no systemic toxicity concern," "<20 EU/pair of gloves").
      • Protein and residual chemical content are compared to specified maximum limits.

    7. The sample size for the training set

    • Not Applicable. This is not an AI algorithm; there is no "training set."

    8. How the ground truth for the training set was established

    • Not Applicable. As there is no training set, this question is not relevant.
    Ask a Question

    Ask a specific question about this device

    K Number
    K212290

    Validate with FDA (Live)

    Date Cleared
    2021-10-28

    (99 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Examination gloves-Type C (Nitrile gloves) is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

    Device Description

    The proposed device is one kind of patient examination glove made from nitrile rubber compounds. The typical characters of these examination gloves are blue color, single-layer, powder-free, non-sterilized, single-use, variable size (see table 2.). These examination gloves can be worn on the examiner's hands or fingers during a medical examination. In addition, it is for over-the-counter use and also for adult use only.

    AI/ML Overview

    This document summarizes the acceptance criteria and performance data for Examination gloves - Type C (Nitrile gloves), as presented in the provided FDA 510(k) Premarket Notification.


    1. Table of Acceptance Criteria and Reported Device Performance

    Standard / TestTest DescriptionAcceptance CriteriaReported Device Performance
    ASTM D5151-19Freedom from holesNo water leakage (0 pinhole)0 pinhole found
    ASTM D6319-19Physical DimensionMulti acceptance criteria refer to ASTM D6319 Clause 7.4- TABLE 2.Meet requirements. (Specific values in Table 4)
    ASTM D6319-19Tensile strength and ElongationMulti acceptance criteria refer to ASTM D6319 Clause 7.5- TABLE 3.Meet requirements. (Specific values in Table 5)
    ASTM D6124-06 (Reapproved 2017)Residual powder$\leq$2.0 mg/glove0.10 mg/glove (before aging), 0.12 mg/glove (after aging)
    ISO 10993-10:2010Skin sensitization and irritationNo irritation and skin sensitizationNo irritation and skin sensitization
    ISO 10993-11:2017Systemic toxicityNo acute systemic toxicityNo acute systemic toxicity

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the specific sample sizes for each test within the "Non-Clinical Testing" section. However, the tests are conducted in conformance with recognized international and national standards (ASTM, ISO), which typically specify appropriate sample sizes and methodologies.

    The provenance of the data is not explicitly stated in terms of country of origin, but it is derived from non-clinical laboratory testing performed by the manufacturer, Jiangxi Kemei Medical Apparatus & Instruments Group Co., Ltd., located in Nanchang City, Jiangxi Province, China. The data is retrospective, as it represents results from tests already performed on the finished device to demonstrate compliance.


    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    This information is not applicable. The device in question is a physical medical device (examination gloves), and the "ground truth" for its performance is established through objective, standardized laboratory tests (e.g., water leakage for pinholes, mechanical strength tests, chemical analysis for residual powder, and biological safety assessments). These are not typically assessed by human experts establishing a "ground truth" in the way, for example, a radiologist would for an AI image analysis system.


    4. Adjudication Method for the Test Set

    This information is not applicable. The tests performed are objective, quantitative measurements or qualitative observations (e.g., "no irritation") based on standardized protocols rather than subjective interpretations requiring adjudication among multiple human reviewers.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study was not done. This type of study is typically conducted for diagnostic or screening devices where human readers (e.g., radiologists) interpret images or data, and their performance is compared with and without AI assistance. This document pertains to non-AI-powered examination gloves.


    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable. The device is a physical product (examination gloves), not an algorithm or software requiring a standalone performance evaluation.


    7. The Type of Ground Truth Used

    The ground truth for the performance of the examination gloves is established through:

    • Standardized Test Methods: Adherence to established industry standards like ASTM (e.g., D5151-19 for holes, D6319-19 for physical properties, D6124-06 for residual powder) and ISO (e.g., 10993-10 for irritation/sensitization, 10993-11 for systemic toxicity).
    • Objective Measurements: Quantitative measurements (e.g., length, width, thickness, tensile strength, elongation, residual powder amount) with defined acceptable ranges.
    • Biological Assessments: Standardized biological evaluations to determine irritation, sensitization, and systemic toxicity, with "no irritation," "no sensitization," and "no acute systemic toxicity" as the ground truth for biocompatibility.

    8. The Sample Size for the Training Set

    This information is not applicable. This device is an examination glove, not an AI or machine learning model that requires a training set.


    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1