Search Filters

Search Results

Found 6 results

510(k) Data Aggregation

    K Number
    K211753

    Validate with FDA (Live)

    Date Cleared
    2021-10-14

    (129 days)

    Product Code
    Regulation Number
    880.5860
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Sterile Syringe with Safety Needle is intended for use in the aspiration of fluids for medical purpose. After withdrawal of the needle from the body, the attached needle safety shield can be manually activated to cover the needle immediately after use to minimize risk of accidental needle sticks.

    Device Description

    The Sterile Disposable Syringe with Safety Needle is intended for manual and single use only. It consists of a hypodermic needle with a safety sheath attached to the needle hub and a luer slip or luer lock syringe. The subject device is available in a variety of syringe volumes and needle sizes. The safety sheath is manually activated to cover the needle immediately after use to minimize risk of accidental needle sticks.

    AI/ML Overview

    The provided FDA 510(k) summary (K211753) is for a Sterile Syringe with Safety Needle. This document outlines the premarket notification for a Class II medical device, comparing it to a predicate device (K193526) to demonstrate substantial equivalence.

    It's important to note that this document does not describe AI/ML device performance or studies proving such performance. This is a traditional medical device submission, focusing on compliance with established engineering and safety standards for the physical product.

    Therefore, many of the requested elements pertaining to AI/ML device performance, such as sample size for test/training sets, data provenance, expert adjudication, MRMC studies, standalone algorithm performance, and ground truth establishment for AI models, are not applicable to this document.

    The document primarily focuses on:

    • Indications for Use: The device is for aspiration and injection of fluids, with a safety shield to prevent needle sticks.
    • Technological Characteristics: Comparison of physical attributes like syringe volume, needle gauge, length, and material.
    • Performance Standards: Compliance with international standards for syringes, needles, luer connectors, and sterilization.
    • Biocompatibility: Leveraged from a previously cleared device (K163161) due to identical patient contact materials.

    However, I can extract information related to the physical device's "acceptance criteria" and the "study that proves the device meets the acceptance criteria" in the context of this traditional medical device submission.

    Here's the breakdown based on the provided text, focusing on the physical device performance and safety, not AI/ML:


    Acceptance Criteria and Study for Sterile Syringe with Safety Needle (K211753)

    The acceptance criteria for this device are primarily defined by compliance with recognized international and national standards for medical devices, particularly those related to syringes, needles, luer connectors, biocompatibility, and sterility. The studies proving the device meets these criteria are non-clinical bench tests.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Standard/RequirementDevice Performance (How compliance was demonstrated)
    Syringe PerformanceISO 7886-1: 2017Tested and demonstrated conformance to the standard.
    Needle PerformanceISO 7864: 2016Tested and demonstrated conformance to the standard.
    ISO 9626:2016Tested and demonstrated conformance to the standard.
    Luer Connector PerformanceISO 80369-7: 2016Tested and demonstrated conformance to the standard for small-bore connectors.
    ISO 80369-20: 2015Used for common test methods for small-bore connectors.
    Sharps Injury ProtectionISO 23908: 2011Tested and demonstrated conformance (for sharps protection features for single-use hypodermic needles).
    BiocompatibilityISO 10993 standardsLeveraged existing data from predicate device K163161 due to identical patient contact materials and manufacturing processes. Implies no cytotoxicity, no intracutaneous reactivity, no sensitization, no systemic toxicity, no hemolysis, no pyrogenicity. (Specific tests for K163161 not detailed in K211753)
    USP <85> Bacterial Endotoxins TestBacterial endotoxin limit evaluated on proposed device.
    USP <788> Particulate Matter in InjectionsParticulate testing evaluated on proposed device.
    SterilityISO 11135:2014, Annex BEO Sterilized with a SAL of 10^-6 validated via an overkill method.
    Packaging IntegrityASTM F1886/1886M-16Visual inspection performed.
    ASTM F88/F88-15Seal strength tested.
    ASTM F1929-15Dye penetration tested.
    Shipping PerformanceASTM D4169-16Simulated transportation testing performed.
    Shelf-lifeASTM F1980-16Shelf life of 3 years validated using accelerated aging.
    EO ResidualsISO 10993-7:2008EO ECH residue testing performed.
    Labeling21 CFR part 801Complies with regulations.

    2. Sample size used for the test set and the data provenance

    • Sample Size: The document does not specify the exact sample sizes used for each individual non-clinical bench test. These are typically defined by the testing standards (e.g., ISO, ASTM).
    • Data Provenance: The tests were conducted to demonstrate compliance with international standards (ISO, ASTM, USP). The manufacturing entity is Jiangxi Hongda Medical Equipment Group Ltd. in China. The testing would have been conducted by the manufacturer or a certified testing laboratory. The data is retrospective in the sense that the testing was completed prior to submission for clearance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not Applicable. This is a traditional medical device (syringe with safety needle), not an AI/ML device relying on expert consensus for ground truth. Compliance is determined by objective physical and chemical testing against established engineering and biological standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable. As above, this is a traditional medical device; "adjudication" in the context of clinical reads or expert consensus for AI models is not relevant here. Test results are compared directly against the quantitative and qualitative requirements of the referenced standards.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable. This document refers to a physical medical device (syringe), not an AI/ML diagnostic or assistive tool. MRMC studies are irrelevant for this type of submission.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable. There is no algorithm or software component described in this submission that would require standalone performance testing.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • Bench Testing Standards and Specifications: The "ground truth" for this device is its adherence to the performance requirements stipulated in the referenced international and national standards (e.g., ISO 7886-1 for syringe performance, ISO 9626 for needle tubing, ISO 80369-7 for luer connectors, ISO 10993 for biocompatibility, etc.). These standards define the acceptable mechanical, physical, chemical, and biological properties of the device.

    8. The sample size for the training set

    • Not Applicable. This is a hardware medical device, not an AI/ML model that requires a training set.

    9. How the ground truth for the training set was established

    • Not Applicable. As above, there is no AI/ML training set.
    Ask a Question

    Ask a specific question about this device

    K Number
    K210578

    Validate with FDA (Live)

    Date Cleared
    2021-07-10

    (134 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Medical Nitrile Examination Gloves (Powder free, Blue) is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.

    Device Description

    The proposed device is Powder Free Nitrile Examination Gloves and has different size specification and color. The proposed device is provided in non-sterile. The proposed device is made of Nitrile. The proposed device has different size specification to the predicate device, but all proposed devices are meet the specifications of ASTM D6319.

    AI/ML Overview

    This document is a 510(k) Summary for Medical Nitrile Examination Gloves. The acceptance criteria and the study proving the device meets these criteria are detailed as follows:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are primarily defined by relevant ASTM and ISO standards, and the device performance is reported as complying with these standards.

    Acceptance Criteria (Standard & Requirement)Reported Device Performance (Compliance/Result)
    Physical Properties (Referencing ASTM D6319)
    Tensile Strength (Before Aging) - 14 MPa, min14 MPa, min (Applicant's device)
    Ultimate Elongation (Before Aging) - 500% min500% min (Applicant's device)
    Tensile Strength (After Aging) - 14 MPa, min14 MPa, min (Applicant's device)
    Ultimate Elongation (After Aging) - 400% min400% min (Applicant's device)
    Dimensions (Referencing ASTM D6319)
    Cuff Thickness, Palm Thickness, Finger Thickness - ≥ 0.05 mmAll proposed devices meet the specifications of ASTM D6319 (Table 1, indicates XS, S, U, M, L, XL, XXL all ≥ 0.05mm for Cuff, Palm, Finger Thickness)
    Width - Specific values with ±10 mm tolerance (e.g., XS: 70±10, S: 80±10)All proposed devices meet the specifications of ASTM D6319 (Table 1, indicates XS: 70±10, S: 80±10, U: 85±10, M: 95±10, L: 110±10, X-L: 120±10, XX-L: 130±10)
    Length - Specific values (e.g., XS/S: ≥ 220 mm, U/M/L/XL/XXL: ≥ 230 mm)All proposed devices meet the specifications of ASTM D6319 (Table 1, indicates XS/S: ≥ 220 mm, U/M/L/XL/XXL: ≥ 230 mm)
    Freedom from Holes (Referencing ASTM D5151)Be free from holes when tested in accordance with ASTM D5151 (Pinhole AQL 2.5)
    Powder Content (Referencing ASTM D6124)Less than 2 mg per glove when tested in accordance with ASTM D6124
    Biocompatibility (Referencing ISO 10993)
    Irritation (ISO 10993-10) - Not an irritantUnder the conditions of the study (ISO 10993-10), not an irritant
    Sensitization (ISO 10993-10) - Not a sensitizerUnder conditions of the study (ISO 10993-10), not a sensitizer
    Acute Systemic Toxicity (ISO 10993-11) - No mortality or systemic toxicityUnder the condition of the study (ISO 10993-11), there is no mortality or evidence of systemic toxicity

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample sizes for each specific test (e.g., number of gloves tested for physical properties, number of animals for biocompatibility tests). However, it implies that the tests were conducted according to the methodologies prescribed by the referenced standards (ASTM D6319, ASTM D5151, ASTM D6124, ISO 10993-10, ISO 10993-11), which would include sample size requirements.

    The data provenance is not explicitly stated in terms of country of origin or retrospective/prospective nature of the data. However, the manufacturer is Jiangxi Hongda Medical Equipment Group Ltd. from China, and the testing results seem to be specific to their manufactured gloves. These appear to be prospective studies conducted specifically for this 510(k) submission to demonstrate compliance.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    This information is not applicable to this type of device submission. The "ground truth" for medical gloves is established through objective physical, chemical, and biological testing against recognized international standards (ASTM and ISO), not through expert consensus or interpretation of images.

    4. Adjudication Method for the Test Set

    This information is not applicable as the evaluations are based on objective laboratory testing against predefined criteria in standards, not subjective assessments requiring expert adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable. This is a submission for medical gloves, a physical device, and does not involve AI software or human readers for diagnostic interpretation.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable. This is a submission for medical gloves, a physical device, and does not involve an algorithm.

    7. The Type of Ground Truth Used

    The ground truth used for this device is based on objective measurements and laboratory testing results compared against the specifications defined in recognized international standards:

    • ASTM D6319-19 (Standard Specification for Nitrile Examination Gloves for Medical Application)
    • ASTM D5151-19 (Standard Test Method for Detection of Holes in Medical Gloves)
    • ASTM D6124-17 (Standard Test Method for Residual Powder on Medical Gloves)
    • ISO 10993-10:2010 (Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization)
    • ISO 10993-11:2017 (Biological evaluation of medical devices - Part 11: Tests for systemic toxicity)

    8. The Sample Size for the Training Set

    This information is not applicable. The device is a physical product (medical gloves), not an AI algorithm, and therefore does not have a training set.

    9. How the Ground Truth for the Training Set was Established

    This information is not applicable as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K210821

    Validate with FDA (Live)

    Date Cleared
    2021-07-09

    (112 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Medical PVC Examination Gloves is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.

    Device Description

    The proposed device, Medical PVC Examination Gloves are disposable devices intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.

    The proposed device is Powder Free Vinyl Patient Examination Gloves. The proposed device is clear, non-colored. The design of proposed device is addressing the standards as ASTM D6124, ASTM D5151, and ASTM D5250.

    The proposed device is not provided as sterilized

    The proposed device is made of vinyl chloride.

    AI/ML Overview

    The provided text describes the 510(k) submission for "Medical PVC Examination Gloves" and does not contain information about a study proving the device meets acceptance criteria in the context of an AI/ML medical device. The document explicitly states the device is a physical product (gloves) and outlines its physical and performance specifications, as well as biocompatibility tests.

    Therefore, I cannot provide the requested information for acceptance criteria and a study from an AI/ML perspective because the provided document is not about an AI/ML device.

    However, I can extract the acceptance criteria and reported device performance for the Medical PVC Examination Gloves based on the provided document.

    Acceptance Criteria and Reported Device Performance (for Medical PVC Examination Gloves):

    Acceptance CriteriaProposed Device Performance
    Physical Dimensions (ASTM D5250)
    Length (mm)
    XS230 min
    S230 min
    M230 min
    L230 min
    XL230 min
    Width (mm)
    XS80 $\pm$ 5
    S85 $\pm$ 5
    M95 $\pm$ 5
    L105 $\pm$ 5
    XL115 $\pm$ 5
    Thickness (mm)
    Finger0.08 min
    Palm0.08 min
    Cuff0.08 min
    Performance and Physical Specifications (ASTM D5250)
    Before Aging
    Tensile Strength11 MPa, min
    Ultimate Elongation300 % min
    After Aging
    Tensile Strength11 MPa, min
    Ultimate Elongation300 % min
    Pinhole AQL2.5
    Freedom from Holes (ASTM D5151)Be free from holes
    Powder Content (ASTM D6124)Less than 2 mg per glove
    Biocompatibility (ISO 10993-10)
    IrritationNot an irritant
    SensitizationNot a sensitizer
    Acute Systemic Toxicity (ISO 10993-11)No mortality or evidence of systemic toxicity

    Regarding the other requested points for an AI/ML study, the document does not provide the information:

    1. Sample sizes used for the test set and the data provenance: Not applicable, as this is a physical medical device (gloves).
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable.
    3. Adjudication method for the test set: Not applicable.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and the effect size: Not applicable.
    5. If a standalone performance (i.e. algorithm only, without human-in-the-loop performance) was done: Not applicable.
    6. The type of ground truth used: Not applicable. The "ground truth" for these gloves is established by laboratory testing against physical and chemical standards.
    7. The sample size for the training set: Not applicable.
    8. How the ground truth for the training set was established: Not applicable.

    The study described in this 510(k) summary is a series of bench tests conducted to verify that the proposed device (Medical PVC Examination Gloves) met design specifications and was substantially equivalent to the predicate device. These tests align with the listed ASTM and ISO standards for physical properties, performance, and biocompatibility of examination gloves. The document states: "The test results demonstrated that the proposed device complies with the following standards: ASTM D5250-19, ASTM D5151-19, ASTM D6124-17, ISO 10993-10:2010, ISO 10993-11:2017."

    Ask a Question

    Ask a specific question about this device

    K Number
    K163160

    Validate with FDA (Live)

    Date Cleared
    2017-04-20

    (161 days)

    Product Code
    Regulation Number
    880.5440
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is intended to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into the vein.

    Device Description

    The proposed device is intended to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into the action of gravity. The proposed device is provided with/without injection needle. It is comprised of protective cap of spike, air vent, air filter, drip chamber, fluid filter, flexible tube, injection site, roller clamp, luer lock connector, protector cap of luer lock connector. The devices are provided sterile and single use.

    AI/ML Overview

    The provided text is a 510(k) summary for a Sterile Single-use Infusion Set. It describes the non-clinical tests performed to demonstrate substantial equivalence to a predicate device, rather than a clinical study of AI performance. Therefore, many of the requested categories regarding AI-specific studies (e.g., sample size for test set, number of experts, MRMC study, standalone performance, training set details) are not applicable to this document.

    However, I can extract the acceptance criteria and reported device performance from the non-clinical test conclusions.

    Acceptance Criteria and Device Performance for Sterile Single-use Infusion Set

    Acceptance CriteriaReported Device Performance
    Infusion Set Performance
    Particulate contaminationContamination index limit is less than 90 (Comply with ISO 8536-4)
    LeakageNo leakage
    Tensile strengthWithstand a static tensile force of not less than 15N for 15s
    Closure piercing deviceComply the dimension of ISO 8536
    Air-inlet deviceFlow of fluid is reduced less than 20% of that from a freely ventilated container
    TubingLength is not less than 1500mm
    Fluid filterRetention of latex particles is not less than 80%
    Drip chamber and drip tube20 drops of distilled water delivered by the drip are equivalence to a volume of (1±0.1)ml(1±0.1)g
    Flow rateDeliver not less than 1000ml
    Injection siteNo leakage
    ConstructionThe clamps can resist the flow of fluid and air at an applied pressure of 50 kPa when closed.
    Needle Performance
    CleanlinessFree from particles and extraneous matter (Comply with ISO 7864 and ISO 9626)
    Needle length38mm (Specific to the proposed device; predicate "Unknown")
    Bond of hub and needle tubeNot broken by the minimum force
    StiffnessDeflection is less than 0.50mm
    Dimensions of needle tubing0.8mm (Specific to the proposed device; predicate "Unknown")
    Resistance to breakageThe tubing is not break
    Resistance to corrosionNo evidence of corrosion
    Biocompatibility
    CytotoxicityNo Cytotoxicity (Conform with ISO 10993 requirements)
    Intracutaneous ReactivityNo Intracutaneous Reactivity (Conform with ISO 10993 requirements)
    Skin SensitizationNo Skin Sensitization (Conform with ISO 10993 requirements)
    Acute Systemic ToxicityNo Acute Systemic Toxicity (Conform with ISO 10993 requirements)
    HemolysisNo Hemolysis (Conform with ISO 10993 requirements)
    PyrogenNo pyrogen (Conform with ISO 10993 requirements)
    Sterile Barrier Packaging Testing
    Seal strength, Internal pressure, Dye PenetrationTested as per ASTM F88/F88-09, ASTM F1140/F1140M-13, ASTM F 1929-12 respectively. (Implicitly met based on SE conclusion)
    Sterilization and Shelf Life Testing
    EO residue, ECH residueTested as per ISO 10993-7:2008 (Implicitly met based on SE conclusion)
    Bacteria Endotoxin LimitTested as per USP 36-NF 31 <85> (Implicitly met based on SE conclusion)
    Shelf Life EvaluationPhysical, Mechanical, Chemical, Package and Sterility Tests performed on accelerated aging samples to verify the claimed shelf life of the device. (Implicitly met based on SE conclusion)
    Transportation TestingPerformed on the final product to verify its package integrity during transportation. (Implicitly met based on SE conclusion)

    Here's an analysis of the study based on your requested items:

    1. A table of acceptance criteria and the reported device performance

      • See table above. The performance is reported as meeting the specified standards (e.g., ISO, ASTM, USP) or specific numerical/qualitative criteria.
    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

      • The document does not specify sample sizes for individual non-clinical tests.
      • The tests were conducted by a Chinese manufacturer (Jiangxi Hongda Medical Equipment Group, Ltd. in Nanchang City, Jiangxi Province, China). The data provenance is China.
      • These are prospective tests conducted on the manufactured device to demonstrate compliance with standards.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

      • This document describes non-clinical engineering and biological tests, not an AI or diagnostic study involving human expert interpretation for ground truth. Therefore, this question is not applicable. The "ground truth" for these tests is defined by the technical specifications of international and national standards.
    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

      • Not applicable, as this is not a study involving human adjudication of results for a diagnostic or AI algorithm.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

      • No MRMC or AI-related study was done. This is a 510(k) submission for a physical medical device (infusion set), not an AI algorithm. Therefore, this question is not applicable.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

      • No standalone AI algorithm performance study was done. This is not an AI device. Therefore, this question is not applicable.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

      • The "ground truth" for the non-clinical tests is based on established international and national standards (e.g., ISO 8536-4, ISO 9626, ISO 7864, ISO 594-1, ISO 594-2, ASTM F88, ASTM F1140, ASTM F1929, ISO 10993-7, USP 36-NF 31). These standards define the acceptable performance limits and methodologies for testing.
    8. The sample size for the training set

      • Not applicable. This is not an AI device or a machine learning study.
    9. How the ground truth for the training set was established

      • Not applicable. This is not an AI device or a machine learning study.
    Ask a Question

    Ask a specific question about this device

    K Number
    K163162

    Validate with FDA (Live)

    Date Cleared
    2017-03-20

    (130 days)

    Product Code
    Regulation Number
    880.5860
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Sterile Hypodermic Safety Syringes for Single Use is intended to be used for medical purpose to inject fluids into or withdraw fluids from the body. After injection, the anti-needlestick feature is manually activated to aid in the prevention of accidental needle stick injuries.

    Device Description

    The proposed devices are for single use only, which are comprised of syringes and needles with various specifications. The proposed devices are available in a variety combination of needle sizes and syringe volumes. All specifications of proposed devices subject to the same design. The proposed devices are intended for manual use only.

    AI/ML Overview

    The provided text describes the 510(k) summary for the "Sterile Hypodermic Safety Syringes for Single Use". This document focuses on demonstrating substantial equivalence to a predicate device, rather than defining and proving acceptance criteria for a novel AI device or a device requiring extensive clinical trials to establish effectiveness.

    Therefore, many of the requested elements for an AI or complex medical device study (such as MRMC studies, training set details, expert ground truth adjudication) are not applicable or described in this document. The study described is primarily non-clinical verification and a simulated clinical study for the safety mechanism.

    Here's a breakdown of the information that is available in the document, framed as closely as possible to your request, and identifying where information is not present:


    Acceptance Criteria and Device Performance for Sterile Hypodermic Safety Syringes

    The device in question is a "Sterile Hypodermic Safety Syringes for Single Use". The study presented is a 510(k) submission to demonstrate substantial equivalence to a legally marketed predicate device (K072739), not to establish novel acceptance criteria for a new type of functionality (like AI performance).

    The "acceptance criteria" are derived from recognized international standards and FDA guidance for this class of medical device (syringes). The "reported device performance" refers to the results of testing performed against these standards.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document lists various tests conducted with reference to specific ISO and ASTM standards. The reported device performance is generally stated as "met all design specifications" or "demonstrated substantial equivalence" with the predicate device.

    Test CategoryAcceptance Criteria (Standard Reference)Reported Device Performance
    Physical, Mechanical, and Chemical Tests (Needle)
    MaterialsClause 3 of ISO 9626:1991/AMD-1:2001Met specifications
    Surface finishClause 4 of ISO 9626:1991/AMD-1:2001Met specifications
    CleanlinessClause 5 of ISO 9626:1991/AMD-1:2001Met specifications
    Acidity and alkalinityClause 6 of ISO 9626:1991/AMD-1:2001Met specifications
    Size designationClause 7 of ISO 9626:1991/AMD-1:2001Met specifications
    DimensionsClause 8 of ISO 9626:1991/AMD-1:2001Met specifications
    StiffnessClause 9 of ISO 9626:1991/AMD-1:2001Met specifications
    Resistance to breakageClause 10 of ISO 9626:1991/AMD-1:2001Met specifications
    Resistance to corrosionClause 11 of ISO 9626:1991/AMD-1:2001Met specifications
    Physical, Mechanical, and Chemical Tests (Syringe)
    CleanlinessClause 5 of ISO 7886:1993Met specifications
    Acidity or alkalinityClause 6 of ISO 7886:1993Met specifications
    Extractable metalsClause 7 of ISO 7886:1993Met specifications
    LubricantClause 8 of ISO 7886:1993Met specifications
    Tolerance on graduated capacityClause 9 of ISO 7886:1993Met specifications
    Graduated scaleClause 10 of ISO 7886:1993Met specifications
    BarrelClause 11 of ISO 7886:1993Met specifications
    Piston/plunger assemblyClause 12 of ISO 7886:1993Met specifications
    NozzleClause 13 of ISO 7886:1993Met specifications
    PerformanceClause 14 of ISO 7886:1993Met specifications
    Luer Connectors (Needle Hub)
    Dimension GaugingClause 4.1 of ISO 594-2:1998Met specifications
    Liquid leakageClause 4.2 of ISO 594-2:1998Met specifications
    Air leakageClause 4.3 of ISO 594-2:1998Met specifications
    Separation forceClause 4.4 of ISO 594-2:1998Met specifications
    Unscrewing torqueClause 4.5 of ISO 594-2:1998Met specifications
    Ease of assemblyClause 4.6 of ISO 594-2:1998Met specifications
    Resistance to overridingClause 4.7 of ISO 594-2:1998Met specifications
    Stress crackingClause 3 of ISO 594-1:1986, Clause 4.1-4.5 of ISO 594-1:1986Met specifications
    Sterile Barrier Packaging
    Seal strengthASTM F88/F88-09Met specifications
    Internal pressureASTM F1140/F1140M-13Met specifications
    Sterilization and Shelf Life
    EO residueISO 10993-7:2008Met specifications
    ECH residueISO 10993-7:2008Met specifications
    Sterility testISO 11737-2: 2009Met specifications
    Bacteria Endotoxin LimitUSP 36-NF 31<85>Met specifications
    Shelf Life EvaluationPhysical, Mechanical, Chemical, Package & Sterility Tests on real-time aging samplesVerified claimed shelf life
    BiocompatibilityISO 10993 standardsNo cytotoxicity, no intracutaneous reactivity, no skin sensitization, no systemic toxicity, no hemolysis, no pyrogen
    Simulated Clinical StudyFDA Guidance: Medical Device with Sharps Injury Prevention Feature, August 9, 2005Met pre-established criteria
    Safety Feature TestComparison to predicate deviceDid not show a significant difference from predicate device

    2. Sample size used for the test set and the data provenance

    The document specifies that "Non clinical tests were conducted to verify that the proposed device met all design specifications". Specific sample sizes for each test are not provided in this summary. The data provenance is from JiangXi HongDa Medical Equipment Group Ltd. in Nanchang City, Jiangxi Province, China, indicating the testing was performed by the manufacturer, likely in China. The studies are non-clinical, meaning they involve laboratory or bench testing of the device itself, rather than human subjects.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This point is not applicable. For a device like a safety syringe, "ground truth" as requested (e.g., related to disease diagnosis or outcome based on expert consensus) is not established by medical experts for the performance tests listed. The "ground truth" for the non-clinical tests is the performance against the specified engineering and material standards.

    4. Adjudication method for the test set

    This point is not applicable. No adjudication by experts is described for these non-clinical performance and safety tests. The results are quantitative measurements against predefined standards.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This point is not applicable. The device is a sterile hypodermic safety syringe, not an AI-assisted diagnostic or therapeutic device. Therefore, no MRMC study, human reader improvement, or AI component is involved.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This point is not applicable as the device is not an algorithm or AI system. Its performance relates to its physical and mechanical properties and its safety mechanism.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The ground truth used for the non-clinical tests is based on the objective measurements and results obtained when testing the device against the specified physical, mechanical, chemical, and biological criteria defined in the referenced ISO and ASTM international standards, as well as FDA guidance. For the simulated clinical study and safety feature test, the "ground truth" refers to meeting the pre-established criteria specified in the FDA guidance and demonstrating functional equivalence to the predicate device, respectively.

    8. The sample size for the training set

    This point is not applicable. This device does not involve a "training set" in the context of machine learning or AI.

    9. How the ground truth for the training set was established

    This point is not applicable as there is no "training set" for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K163161

    Validate with FDA (Live)

    Date Cleared
    2017-03-20

    (130 days)

    Product Code
    Regulation Number
    880.5860
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Sterile Single-use Syringe with Needle is intent fluids into or withdraw fluids from the body.

    Device Description

    The proposed devices are for single use only, which are comprised of syringes and needles with various specifications. The syringe consists of barrel, plunger, piston, and the needle consists of needle tube, needle hub, needle cap. The proposed devices shall be operated manually, are not intended for use with syringe pumps. All specifications of proposed devices subject to the same design and the difference between each specification is syringe volume and needle size. The proposed devices are available in a variety combination of needle size and syringe volume.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a "Sterile Single-use Syringe with Needle." This document outlines the device's technical specifications and the testing conducted to demonstrate its substantial equivalence to a predicate device.

    Crucially, this document is for a medical device (syringe with needle), not an AI/ML software device. Therefore, the concepts of acceptance criteria for AI model performance, test/training sets for algorithms, ground truth established by experts, MRMC studies, or standalone algorithm performance are not applicable to this submission.

    The "acceptance criteria" discussed in this document refer to adherence to established international and national standards for physical, mechanical, chemical, and biological properties of syringes, needles, and packaging. The "study that proves the device meets the acceptance criteria" refers to the non-clinical bench testing and biocompatibility assessments conducted according to these standards.

    Here's an interpretation of the information provided within the context of a medical device submission, recognizing it does not fit the AI/ML framework:


    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are defined by various clauses within a range of ISO and ASTM standards. The reported device performance is that the "non-clinical tests were conducted to verify that the proposed devices met all design specifications and are Substantially Equivalent (SE) to the predicate device." The document implies that the device passed these tests, thereby meeting the acceptance criteria set by these standards.

    Acceptance Criteria (Standard Clause)Reported Device Performance
    ISO 9626:1991/AMD-1:2001 (Needle)
    Clause 3: MaterialsTests performed, device met design specifications.
    Clause 4: Surface finishTests performed, device met design specifications.
    Clause 5: CleanlinessTests performed, device met design specifications.
    Clause 6: Limits for acidity and alkalinityTests performed, device met design specifications.
    Clause 7: Size designationTests performed, device met design specifications.
    Clause 8: DimensionsTests performed, device met design specifications.
    Clause 9: StiffnessTests performed, device met design specifications.
    Clause 10: Resistance to breakageTests performed, device met design specifications.
    Clause 11: Resistance to corrosionTests performed, device met design specifications.
    ISO 7864:1993 (Needle)
    Clause 4: CleanlinessTests performed, device met design specifications.
    Clause 5: Limits for acidity or alkalinityTests performed, device met design specifications.
    Clause 6: Limits for extractable metalsTests performed, device met design specifications.
    Clause 7: Size designationTests performed, device met design specifications.
    Clause 8: Colour codingTests performed, device met design specifications (hub color complies with ISO 6009).
    Clause 9: Needle hubTests performed, device met design specifications.
    Clause 10: SheathTests performed, device met design specifications.
    Clause 11: Needle tubeTests performed, device met design specifications.
    Clause 12: Needle pointTests performed, device met design specifications.
    Clause 13: PerformanceTests performed, device met design specifications.
    ISO 7886:1993 (Syringe)
    Clause 5: CleanlinessTests performed, device met design specifications.
    Clause 6: Limits for acidity or alkalinityTests performed, device met design specifications.
    Clause 7: Limits for extractable metalsTests performed, device met design specifications.
    Clause 8: LubricantTests performed, device met design specifications.
    Clause 9: Tolerance on graduated capacityTests performed, device met design specifications.
    Clause 10: Graduated scaleTests performed, device met design specifications.
    Clause 11: BarrelTests performed, device met design specifications.
    Clause 12: Piston/plunger assemblyTests performed, device met design specifications.
    Clause 13: NozzleTests performed, device met design specifications.
    Clause 14: PerformanceTests performed, device met design specifications.
    ISO 594-1:1986 (Conical Fittings)
    Clause 3: DimensionTests performed, device met design specifications.
    Clause 4.1: GaugingTests performed, device met design specifications.
    Clause 4.2: Liquid leakageTests performed, device met design specifications.
    Clause 4.3: Air leakageTests performed, device met design specifications.
    Clause 4.4: Separation forceTests performed, device met design specifications.
    Clause 4.5: Stress crackingTests performed, device met design specifications.
    ISO 594-2:1998 (Luer-Lock Fittings)
    Clause 4.1: GaugingTests performed, device met design specifications.
    Clause 4.2: LeakageTests performed, device met design specifications.
    Clause 4.3: Separation forceTests performed, device met design specifications.
    Clause 4.4: Unscrewing torqueTests performed, device met design specifications.
    Clause 4.5: Ease of assemblyTests performed, device met design specifications.
    Clause 4.6: Resistance to overridingTests performed, device met design specifications.
    Clause 4.7: Stress crackingTests performed, device met design specifications.
    Sterile Barrier Packaging Testing
    ASTM F88/F88-09: Seal strengthTests performed, device met design specifications.
    ASTM F1140/F1140M-13: Internal pressureTests performed, device met design specifications.
    Sterilization and Shelf Life Testing
    ISO 10993-7:2008: EO residueTests performed, device met design specifications.
    ISO 10993-7:2008: ECH residueTests performed, device met design specifications.
    USP 36-NF 31<85>: Bacteria Endotoxin LimitTests performed, device met design specifications.
    Shelf Life Evaluation (Physical, Mechanical, Chemical, Package, Sterility)Tests performed on real-time aging samples to verify claimed shelf life.
    Biocompatibility Testing (ISO 10993-1)
    Cytotoxicity, Irritation, Sensitization, Systemic Toxicity, Hemolysis, PyrogenTests performed, device met design specifications.

    Since this document is not about an AI/ML device, the remaining points are addressed as "Not Applicable" for AI/ML concepts, or reinterpreted in the context of a traditional medical device submission if relevant.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Test Set: The document does not specify exact sample sizes for each physical, mechanical, chemical, sterility, or biocompatibility test. It indicates "tests were performed on real time aging samples" for shelf life, implying prospective testing for that aspect. Other tests are generally conducted on representative samples of the manufactured device.
    • Data Provenance: The manufacturing company is JiangXi HongDa Medical Equipment Group Ltd. in China. The testing was likely performed in China, but the document does not explicitly state the country of origin for the testing data. The testing is performed on the proposed device itself, not on human data.
    • Retrospective/Prospective: Shelf-life testing is prospective (real-time aging). Other tests are typically performed on newly manufactured devices, which could be considered prospective testing for that specific batch.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not Applicable (for AI/ML ground truth): Ground truth in this context refers to the verifiable performance of the physical device against established engineering and biological standards. There are no "experts" establishing ground truth in the AI/ML sense (e.g., clinicians labeling images). The standards themselves define the ground truth, and the testing methodology is designed to objectively measure compliance. The "experts" would be the engineers and technicians performing the tests, qualified in areas of materials science, mechanical testing, chemistry, microbiology, and biocompatibility.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable (for AI/ML adjudication): This concept does not apply to the physical testing of a syringe and needle. Test results are objectively measured against defined limits in the standards.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable: This is a physical medical device, not an AI-assisted diagnostic tool. No MRMC study was performed.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    • Not Applicable: This is a physical medical device, not an algorithm.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    • Ground Truth (for physical device): The "ground truth" for this device is its adherence to the predefined physical, mechanical, chemical, sterility, and biocompatibility specifications as outlined in the referenced ISO and ASTM standards. This is determined by quantitative measurements and qualitative assessments described in these standards, rather than expert consensus on a diagnostic outcome or pathology.

    8. The sample size for the training set

    • Not Applicable: This is a physical medical device, not an AI/ML algorithm. There is no "training set."

    9. How the ground truth for the training set was established

    • Not Applicable: This is a physical medical device. There is no "training set" or ground truth established for training.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1