Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K251464

    Validate with FDA (Live)

    Date Cleared
    2025-10-03

    (143 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TRI-BEAM PRO, TRI-BEAM, Nd: YAG Laser Unit is intended for use in aesthetic, cosmetic and surgical applications requiring incision, excision, ablation, vaporization of soft tissues for general dermatology, dermatologic and general surgical procedures for coagulation and hemostasis.

    1064 Wavelength in G, TH mode:

    • Tattoo removal: dark ink (black, blue and brown)
    • Removal of Nevus of Ota
    • Removal or lightening of unwanted hair with or without adjuvant preparation.
    • Treatment of Common Nevi
    • Skin resurfacing procedures for the treatment of acne scars and wrinkle
    • Treatment of melasma

    1064nm in GN (PTP-Off) mode:

    • Treatment of wrinkles
    • Treatment of mild to moderate inflammatory acne vulgaris

    532nm Wavelength (nominal delivered energy of 585 nm and 650 nm with optional dye handpieces):

    • Tattoo removal: light ink (red,tan, purple, orange, sky blue,green)
    • Removal of Epidermal benign Pigmented Lesions
    • Removal of Minor Vascular Lesions including but not limited to telangiectasias
    • Treatment of Lentigines;
    • Treatment of Cafe-Au-Lait
    • Treatment of Seborrheic Keratoses
    • Treatment of Post Inflammatory Hyper-Pigmentation
    • Treatment of Becker's Nevi, Freckles and Nevi Spilus
    Device Description

    TRI-BEAM PRO, Nd:YAG Laser unit in the wavelength 1064 nm, 532 nm, 585 nm, and 650 nm is intended for treatment for general dermatology. The TRI-BEAM Series, utilizing Nd:YAG laser optical energy, is based on the principle of selective photothermolysis, enabling targeted treatment of pigmented and vascular lesions, tattoos, wrinkles, acne, and other dermatologic conditions while minimizing damage to surrounding tissue. The system comprises the main unit, a color touch screen, a handpiece, and a foot switch.

    Handpiece Mode

    PTP_ON: Single Pulse
    Laser energy is delivered to the target tissue as a single, continuous pulse.

    PTP_OFF: Dual Pulse
    One pulse is divided into two sub-pulses, delivered sequentially with a short time interval to reduce peak thermal load.

    G Mode
    Beam profile with energy concentrated at the center and gradually tapering toward the periphery, allowing focal energy delivery.

    TH Mode
    Flat-top energy distribution in which the output intensity at the center and periphery is nearly uniform, ensuring consistent tissue exposure.

    GN Mode
    Flat-top energy distribution designed to minimize strong irritation to the skin, even when applied at higher energy levels.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K250065

    Validate with FDA (Live)

    Date Cleared
    2025-07-08

    (179 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This device indicated for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis.

    Device Description

    The DENSITY, DENZA, DENSITY Noir generates radiofrequency (RF) energy by means of high RF at 6.78MHz. The RF energy is delivered through the skin into the target tissue via a handpiece equipped with an electrode tip. As the RF energy passes through the tissue, it generates an electrothermal reaction which is capable of coagulating the tissue. In detail, the main body uses the "reverse thermal gradient" principle to deliver RF energy, and thus, heat is generated and selective coagulation occurs while cooling the epidermis, resulting in denaturation and contraction of collagen fibers.

    The devices are an RF (radiofrequency), software-controlled electrosurgical device used for electrocoagulation of soft tissue and hemostasis.

    It consists of the following components:

    • Electrosurgical Unit - Main body
    • Handpiece
    • electrode tips
    • Neutral electrode pad and neutral electrode pad cable, cleared under K201685
    • Foot switch
    • Power cord
    AI/ML Overview

    This document is a 510(k) clearance letter for an electrosurgical device, not a study report for an AI device. As such, most of the requested information regarding AI device acceptance criteria, study details, and performance metrics (like effect sizes for AI assistance, standalone performance, training sets, etc.) is not present in the provided text.

    The device described, "DENSITY; DENZA; DENSITY Noir," is an electrosurgical cutting and coagulation device. The clearance is based on substantial equivalence to a predicate device (K230663), implying that its performance is expected to be similar to legally marketed devices based on established standards and non-clinical testing.

    Here's a breakdown of the information that can be extracted from the provided text, along with an explanation of why other requested information is not available:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state "acceptance criteria" for performance that would be found in a typical AI device study (e.g., sensitivity, specificity, AUC thresholds). Instead, the acceptance is based on demonstrating substantial equivalence to a predicate device through adherence to recognized standards and non-clinical performance testing.

    Acceptance Criteria (Implied by Substantial Equivalence and Standards Adherence):

    Criterion TypeDescriptionReported Device Performance/Verification Method
    Safety & Performance StandardsCompliance with recognized national and international safety and performance standards for electrosurgical devices.Evaluated in accordance with: - IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012, IEC 60601-1:2005/AMD2:2020 (Basic safety and essential performance) - IEC 60601-1-2:2014 (Electromagnetic disturbances) - IEC 60601-2-2 (HF surgical equipment particular requirements) - IEC 60601-1-6:2013 (Usability)
    SoftwareSoftware verified and validated to a moderate level of concern, adhering to software lifecycle processes.Evaluated according to IEC 62304:2006.
    Usability/Human FactorsApplication of usability engineering to medical devices.Evaluated in accordance with IEC 62366:2008 based on Human Factor Engineering.
    BiocompatibilityMaterials in contact with patients are biocompatible.Documented in reference to ISO 10993-1:2009, ISO 10993-5:2009, and 10993-10:2010.
    Bench Testing PerformanceDevice operates safely and within predefined design specifications, demonstrating intended function for electrocoagulation and hemostasis.Bench testing conducted to assure safe operation and adherence to design specifications.
    Ex Vivo Thermal TestingPerformance in tissue coagulation and hemostasis, particularly regarding thermal effects.Ex Vivo testing conducted on three types of tissue under GLP Thermal testing in accordance with FDA guidance "Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery".
    Risk ManagementRisks identified, analyzed, and controlled.Risk analysis conducted based on ISO 14971:2019.
    Substantial EquivalenceDevice is as safe and effective as a legally marketed predicate device, with no new questions of safety or effectiveness raised by differences.Compared to predicate device K230663. All key parameters (output energy type, user interface, operating frequency, max power, electrode tip types, temperature range, impedance, coolant solution, communication) were found to be substantially equivalent. Minor differences in electrode tip styles and number of tips were deemed not to affect performance or safety.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided as this letter describes non-clinical testing for an electrosurgical device, not a clinical study or AI performance evaluation. The "test set" here refers to the materials and conditions used for bench and ex-vivo testing. The document states:

    • "Ex Vivo testing conducted on three types of tissue". No specific sample size (e.g., number of tissue samples) or provenance is mentioned beyond "under GLP Thermal testing."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable/not provided. Ground truth for an AI device is typically established by expert interpretation of medical images or data. For an electrosurgical device, performance is evaluated against engineering specifications, recognized safety standards, and physiological responses during ex-vivo testing, not by expert interpretation generating a "ground truth" in the same sense as an AI diagnostic.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable/not provided. Adjudication methods are used in studies involving human interpretation or labeling, especially for AI algorithm ground truth establishment. This is not relevant for the type of non-clinical testing described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable/not provided. An MRMC study is relevant for evaluating the impact of AI assistance on human performance, typically in diagnostic imaging. The device in question is a standalone electrosurgical instrument and does not involve AI assistance for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable/not provided. This device is a manual electrosurgical instrument, not an AI algorithm. Therefore, "standalone algorithm performance" is not relevant.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the non-clinical testing:

    • Bench testing: Ground truth is defined by the device's technical specifications and engineering design (e.g., power output within tolerance, correct operating frequency).
    • Ex Vivo testing: Ground truth would be the expected physiological effect (e.g., desired thermal coagulation) confirmed through histological examination or other scientific methods in the tissue samples. This is implied by "Ex Vivo testing conducted... under GLP Thermal testing."

    8. The sample size for the training set

    This information is not applicable/not provided. This device is an electrosurgical instrument, not an AI system that requires a "training set."

    9. How the ground truth for the training set was established

    This information is not applicable/not provided for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K192545

    Validate with FDA (Live)

    Device Name
    Potenza
    Date Cleared
    2020-02-20

    (157 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The POTENZA is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis.

    Device Description

    The POTENZA is an RF (radiofrequency), software-controlled electrosurgical device used for electrocoagulation of soft tissue and hemostasis. The POTENZA consists of the following components: Electrosurgical Unit - Main body, Two different handpieces ( motor and AC), Six different electrode tips for the motor handpiece and three needle tips for the AC handpiece; provided as single use, sterile products, Neutral electrode pad and neutral electrode pad cable, cleared under K092761, Handpiece stand, Foot switch, Power cord. The POTENZA generates radiofrequency (RF) energy by means of high RF at 1MHz or 2MHz. The RF energy is delivered through the skin into the target tissue via a handpiece equipped with an electrode tip. As the RF energy passes through the tissue, it generates an electrothermal reaction which is capable of coagulating the tissue. The POTENZA has two operating modes: monopolar mode and bipolar mode.

    AI/ML Overview

    This looks like a 510(k) summary for a medical device. Based on the provided text, the device in question is the "POTENZA," an RF electrosurgical device. However, the document does NOT contain information about acceptance criteria or a study proving the device meets those criteria in the way you've outlined for AI/CADe devices (i.e., sensitivity, specificity, reader studies, etc.).

    This document describes the regulatory process for establishing substantial equivalence to a predicate device, which primarily relies on demonstrating that the new device has similar technological characteristics and performs comparably to devices already on the market. The performance data presented here are primarily engineering bench tests, electrical safety, biocompatibility, and software validation. These are standard tests for electrosurgical devices to ensure safety and functionality, not to evaluate diagnostic performance or AI algorithm effectiveness.

    Therefore, many of the specific questions you've asked, particularly those related to AI/CADe system validation (sample sizes for test sets, data provenance, ground truth establishment, expert adjudication, MRMC studies, standalone performance), are not applicable to the information provided in this 510(k) summary for an electrosurgical device.

    Here's a breakdown of what can be extracted, and why other parts are not present:


    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state "acceptance criteria" in the format of a clinical performance metric (like sensitivity or specificity) with a target value. Instead, the acceptance criteria are implicitly met by passing various engineering and biocompatibility tests.

    Test TypeStandard / ObjectiveReported Device Performance
    Biocompatibility (for electrode tips)FDA's "Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and testing within a risk management process", June 6, 2016. Specific ISO-10993 and ASTM F756 standards for Cytotoxicity, Sensitization, Irritation, Acute Systemic Toxicity, Hemolysis, and Pyrogenicity.All tests (Cytotoxicity, Sensitization, Irritation or Intracutaneous Reactivity, Acute Systemic Toxicity, Hemolysis, USP Rabbit Pyrogen) Pass.
    Electrical Safety & Electromagnetic Compatibility (EMC)IEC 60601-1:2005 (Third Edition) + CORR. 1:2006 + CORR. 2:2007 + A1:2-12 (General requirements for basic safety and essential performance) IEC 60101-2-2:2017 (High-frequency surgical equipment) IEC 60601-1-2:2014 (EMC) IEC 60601-1-6:2010, AMDI:2013 (Usability)System was tested according to these standards, implying "satisfactory" or "compliant" outcome for these safety and functional aspects (explicit "pass" not stated for each, but inferred).
    Software Verification and ValidationFDA's "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices", May 11, 2005, commensurate with a moderate level of concern.Conducted, implying "satisfactory" or "compliant" outcome.
    Bench Testing (Output Accuracy, Frequency, Power Fluctuation, etc.)To assure the POTENZA operates safely and within predefined design specifications. Includes thermal testing per FDA guidance.Conducted, implying "satisfactory" or "compliant" outcome for all listed parameters: Output accuracy (Monopolar/Bipolar, 1MHz/2MHz), Frequency, Power fluctuation, Negative output protection, Impedance measurement, HO count accuracy, Safety test of warnings/failsafe, Needle depth, Motor speed level, Thermal testing.

    Regarding the specific questions about AI/CADe validation:

    1. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective).

      • Not applicable. This device is an electrosurgical unit, not an AI/CADe system that analyzes data like images. The "test set" here refers to physical components and the device itself undergoing engineering and biological safety tests.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience).

      • Not applicable. Ground truth, in the context of AI/CADe, involves expert interpretation of data. For this electrosurgical device, "ground truth" relates to objective physical and biological properties measured by standardized tests (e.g., cytotoxicity assays, electrical measurements).
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set.

      • Not applicable. Adjudication is for resolving discrepancies in expert interpretations of data. This is not performed for physical device testing.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance.

      • Not applicable. This is not an AI-assisted diagnostic device.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done.

      • Not applicable. This is not an AI algorithm. Its performance is evaluated as a standalone electrosurgical unit through bench and safety tests.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc).

      • The "ground truth" for the tests performed are highly standardized scientific and engineering measurements. For example, for biocompatibility, the ground truth is whether specific biological responses (e.g., cell death, irritation) occur above a predefined threshold according to the specified ISO standards. For electrical safety, the ground truth is whether the device parameters (e.g., leakage current, output power) fall within the limits set by IEC standards.
    7. The sample size for the training set.

      • Not applicable. There is no AI training set for this electrosurgical device.
    8. How the ground truth for the training set was established.

      • Not applicable. There is no AI training set or associated ground truth.

    In summary, the provided document is a 510(k) summary for an electrosurgical device, detailing its regulatory pathway to market based on substantial equivalence. It focuses on safety and fundamental performance characteristics through engineering and biocompatibility testing, not on the diagnostic accuracy or AI performance metrics you've requested.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1