Search Filters

Search Results

Found 5 results

510(k) Data Aggregation

    K Number
    K231216

    Validate with FDA (Live)

    Device Name
    POTENZA
    Date Cleared
    2024-12-10

    (592 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The POTENZA is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis.

    Device Description

    The POTENZA is an RF (radiofrequency), software-controlled electrosurgical device used for electrocoagulation of soft tissue and hemostasis. The POTENZA generates radiofrequency (RF) energy by means of high RF at IMHz or 2MHz. The RF energy is delivered through the skin into the target tissue via a handpiece equipped with an electrode tip. As the RF energy passes through the tissue, it generates an electrothermal reaction which is capable of coagulating the tissue. The POTENZA has two operating modes: monopolar mode and bipolar mode.

    AI/ML Overview

    The provided text is a 510(k) Summary for the POTENZA device, which is an electrosurgical cutting and coagulation device. The document primarily focuses on demonstrating substantial equivalence to a predicate device (also named POTENZA, K201685) rather than presenting a study to prove acceptance criteria for a new device.

    The "acceptance criteria and reported device performance" in this context refer to the performance standards and test results that show the modified device is substantially equivalent to the previously cleared predicate device.

    Here's an analysis of the provided information, framed to address your request as best as the document allows:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" in the traditional sense of a performance study. Instead, it details various tests performed to demonstrate that the changes made to the device (new electrode tips) do not raise new questions of safety or effectiveness and that the device continues to conform to applicable standards. The "reported device performance" are the results of these tests, which consistently show "Pass."

    Test TypeStandard / Acceptance Criteria (Implied)Reported Device Performance
    Biocompatibility Testing (for new electrode tips)
    CytotoxicityISO 10993-05:2009, Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity (Implied acceptance: pass cytotoxicity assessment)Pass
    Sensitization: Guinea Pig Maximization TestISO 10993-10:2021 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization (Implied acceptance: no sensitization)Pass
    Irritation or Intracutaneous ReactivityISO 10993-23: Biological Evaluation of Medical Devices: Test for irritation, 2021 (Implied acceptance: no irritation/reactivity)Pass
    Acute Systemic ToxicityISO 10993- 11: Biological Evaluation of Medical Devices- Part 11: Tests for Systemic Toxicity, 2017 (Implied acceptance: no acute systemic toxicity)Pass
    Material Mediated Pyrogen TestISO 10993- 11: Biological Evaluation of Medical Devices- Part 11: Tests for Systemic Toxicity, 2017 (Implied acceptance: no pyrogenic response)Pass
    In-vitro Hemolysis Test direct contact methodISO 10993-04: Biological Evaluation of Medical Devices - Part 04, Selection of Tests for Interactions with Blood, 2017/ ASTM F 756-17 (Standard Practice for the Assessment of Hemolytic Properties of Materials) (Implied acceptance: no hemolysis)Pass
    In-vitro Hemolysis Test Indirect MethodISO 10993-04: Biological Evaluation of Medical Devices - Part 04, Selection of Tests for Interactions with Blood, 2017/ ASTM F 756-17 (Implied acceptance: no hemolysis)Pass
    Electrical Safety and Electromagnetic Compatibility (EMC)
    Electrical SafetyIEC 60601-2-2:2017 (Implied acceptance: conformance to standard)Conforms
    EMCIEC 60601-1-2:2014 (Implied acceptance: conformance to standard)Conforms
    Bench Testing
    Output accuracy (Monopolar at 1MHz and 2 MHz)Predefined design specifications (Implied acceptance: within specified accuracy range)Conforms
    Output accuracy (Bipolar at 1MHz and 2MHz)Predefined design specifications (Implied acceptance: within specified accuracy range)Conforms
    Frequency: manual and standardPredefined design specifications (Implied acceptance: correct frequency output)Conforms
    Power fluctuation characteristicsPredefined design specifications (Implied acceptance: stable power output)Conforms
    Negative output protectionPredefined design specifications (Implied acceptance: protection functions correctly)Conforms
    Impedance measurement accuracy and rangePredefined design specifications (Implied acceptance: accurate measurement within range)Conforms
    HO count accuracyPredefined design specifications (Implied acceptance: accurate count)Conforms
    Safety test of various warnings / failsafe mechanismsPredefined design specifications (Implied acceptance: warnings/failsafes function correctly)Conforms
    Needle depthPredefined design specifications (Implied acceptance: accurate needle depth)Conforms
    Motor speed levelPredefined design specifications (Implied acceptance: correct motor speed)Conforms
    Ex Vivo testing (Thermal testing)Guidance for Industry and FDA Staff: Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery, August 15, 2014 (Implied acceptance: consistent thermal effects for electrocoagulation/hemostasis)Conforms

    2. Sample size used for the test set and the data provenance

    • Sample Size: The document does not specify exact sample sizes for each test (e.g., number of devices tested for electrical safety, number of tissue samples for ex vivo testing). For biocompatibility, it states "four newly added electrode tips" were tested, but not the number of samples for each biological test. For ex-vivo testing, it mentions "three types of tissue Liver, Kidney and Muscle," but not the quantity of each.
    • Data Provenance: The tests were conducted by Jeisys Medical Inc., the manufacturer. The data is retrospective in the sense that it was generated for this regulatory submission. Country of origin for the data is implied to be South Korea, where Jeisys Medical Inc. is based. The tests were laboratory (bench) and ex-vivo animal tissue tests.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This submission relies on performance against recognized standards and technical specifications for electrosurgical devices, rather than human expert-established ground truth for a diagnostic AI algorithm. The device is a physical instrument, not an AI or diagnostic software.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    Not applicable. Adjudication methods are typically used in studies involving human interpretation or subjective assessments, often in diagnostic contexts. The tests described are objective, physical, electrical, and biological performance tests against defined standards.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is an electrosurgical tool, not an AI-assisted diagnostic or interpretation system. Therefore, an MRMC study or evaluation of human reader improvement with AI assistance is irrelevant to this submission.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI algorithm. Its performance is evaluated through bench testing and ex-vivo studies.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the tests performed consists of:

    • Specified limits/ranges within recognized consensus standards (e.g., ISO 10993 series, IEC 60601 series).
    • The device's own predefined design specifications for parameters like output accuracy, frequency, power, impedance, needle depth, motor speed, etc.
    • The expected biological responses for biocompatibility (e.g., no cytotoxicity, no sensitization).
    • Expected thermal effects consistent with electrocoagulation and hemostasis in ex-vivo tissue.

    8. The sample size for the training set

    Not applicable. No AI/machine learning model is described, so there is no training set.

    9. How the ground truth for the training set was established

    Not applicable. As there is no training set mentioned, the establishment of its ground truth is not relevant.

    Ask a Question

    Ask a specific question about this device

    K Number
    K231287

    Validate with FDA (Live)

    Device Name
    POTENZA
    Date Cleared
    2024-11-05

    (551 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The POTENZA is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis.

    Device Description

    The POTENZA is an RF (radiofrequency), software-controlled electrosurgical device used for electrocoagulation of soft tissue and hemostasis. The POTENZA generates radiofrequency (RF) energy by means of high RF at 1MHz or 2MHz. The RF energy is delivered through the skin into the target tissue via a handpiece equipped with an electrode tip. As the RF energy passes through the tissue, it generates an electrothermal reaction which is capable of coagulating the tissue. The POTENZA has two operating modes: monopolar mode and bipolar mode. The NEW POTENZA consists of the following components: Electrosurgical Unit - Main body, handpieces (Predicate Device have Motor, AC handpieces), electrode tips for the handpiece – Add three non-invasive tips connect to S handpiece, Neutral electrode pad and neutral electrode pad cable, cleared under K201685, Handpiece stand, Foot switch, Power cord.

    AI/ML Overview

    This 510(k) summary describes a device (POTENZA) that is substantially equivalent to a previously cleared predicate device. The key changes are the addition of a new S handpiece and three new non-invasive electrode tips. The provided text primarily focuses on demonstrating the safety and effectiveness of these additions through non-clinical testing rather than complex AI performance studies. As such, many of the requested points regarding AI acceptance criteria, MRMC studies, and large-scale data sets are not directly applicable or available in this specific document.

    However, based on the provided text, we can describe the acceptance criteria and study that proves the device meets them as they relate to the changes in the device.

    1. Table of Acceptance Criteria and Reported Device Performance

    For the specific changes (new S handpiece and non-invasive tips), the acceptance criteria are generally focused on safety and functionality demonstrating equivalence to the predicate device.

    Acceptance Criteria CategorySpecific Acceptance Criteria (Inferred from text)Reported Device Performance (as stated or inferred)
    BiocompatibilityAll four newly added electrode tips must pass biocompatibility tests (Cytotoxicity, Sensitization, Skin Irritation) as per ISO 10993 standards for devices with limited skin contact (<24 hours).Pass: All specified tests (Cytotoxicity, Sensitization: Guinea Pig Maximization Test, Skin Irritation Test) were passed for the non-invasive electrode tips as per ISO 10993-05:2009 and ISO 10993-10:2010.
    Electrical Safety & EMCThe POTENZA system (including new components) must conform to recognized consensus standards for electrical safety and electromagnetic compatibility (EMC), specifically IEC 60601-2-2:2017 and IEC 60601-1-2:2014.Conformity: Electrical safety, EMC, device-related electrical safety for high frequency, and usability were conducted according to IEC 60601-2-2:2017 and IEC 60601-1-2:2014. Implies successful conformance.
    Bench Testing / PerformanceThe device (with new handpiece and tips) must operate safely and within predefined design specifications, including accurate output power (monopolar/bipolar, 1MHz/2MHz), correct frequency, stable power fluctuation, negative output protection, accurate impedance measurement, HO count accuracy, motor speed, and safe failure mechanisms.Conformity: Bench testing was conducted to assure safe operation and conformance to design specifications. Tested parameters included (and implicitly passed): Output accuracy (Monopolar/Bipolar at 1MHz and 2MHz), Frequency, Power fluctuation, Negative output protection, Impedance measurement accuracy/range, HO count accuracy, Safety test of various warnings/failsafe mechanisms, Motor speed level.
    Ex Vivo TestingMust demonstrate appropriate electrocoagulation and hemostasis characteristics on various tissue types (Liver, Kidney, Muscle) under GLP, and thermal testing in accordance with FDA guidance.Conformity: Ex Vivo testing was conducted on three types of tissue (Liver, Kidney and Muscle) under GLP. Thermal testing was done in accordance with FDA's "Guidance for Industry and FDA Staff: Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery", August 15, 2014. Implies successful outcomes for these tests.
    Impact on Safety & EffectivenessThe addition of new S handpiece with three electrode tips must not introduce new or different questions of safety or effectiveness.No New Questions: Verification and validation activities confirmed that the additions do not introduce new or different questions of safety or effectiveness. The new tip is explicitly stated as being "safer than the invasive type" due to being non-invasive.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document does not specify numerical sample sizes for most performance tests (e.g., number of units tested for electrical safety, number of times functions were tested). For biocompatibility, it refers to standard ISO test protocols, which imply specific sample sizes for each test type (e.g., Guinea Pig Maximization Test typically involves a certain number of animals). For Ex Vivo testing, it mentions "three types of tissue (Liver, Kidney and Muscle)," but not the quantity of each.
    • Data Provenance: The tests are described as non-clinical (bench and ex vivo) and conducted by Jeisys. There is no information regarding country of origin for data samples (as it's not a clinical study involving patients), or if the testing was retrospective or prospective. It is implied to be prospective testing specifically for this 510(k) submission.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This information is not provided in the document as it pertains to AI/clinical ground truth establishment, which is not the focus of this 510(k) submission for an electrosurgical device modification. The "ground truth" here is based on engineering specifications and recognized consensus standards (e.g., ISO, IEC).

    4. Adjudication Method for the Test Set

    This information is not applicable/provided as it relates to expert review of data (e.g., medical images for AI), which is not detailed for this type of device submission. Performance is based on objective measurements against engineering specifications and standard test protocols.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, an MRMC comparative effectiveness study was not done. The document explicitly states: "No clinical performance testing was performed." MRMC studies are typically for evaluating the impact of AI on human reader performance, which is not relevant here as there is no AI component or clinical reader assessment being evaluated.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    Not applicable. This device is an electrosurgical unit, not an algorithm. There is no standalone algorithm performance described.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's performance validation is based on:

    • Engineering Specifications: Predefined design parameters for power output, frequency, impedance, motor speed, etc.
    • Recognized Consensus Standards: International standards like ISO 10993 (biocompatibility) and IEC 60601 series (electrical safety, EMC).
    • FDA Guidance Documents: Bench and thermal testing were conducted in accordance with FDA guidance for electrosurgical devices.

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI/machine learning device. Therefore, there is no "training set."

    9. How the Ground Truth for the Training Set was Established

    Not applicable. As there is no training set, there is no ground truth establishment for it.

    Ask a Question

    Ask a specific question about this device

    K Number
    K230663

    Validate with FDA (Live)

    Device Name
    Density
    Date Cleared
    2023-08-14

    (157 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    DENSITY indicated for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis.

    Device Description

    The DENSITY generates radiofrequency (RF) energy by means of high RF at 6.78MHz. The RF energy is delivered through the skin into the target tissue via a handpiece equipped with an electrode tip. As the RF energy passes through the tissue, it generates an electrothermal reaction which is capable of coagulating the tissue. In detail, the main body uses the "reverse thermal gradient" principle to deliver RF energy, and thus, heat is generated and selective coaqulation occurs while cooling the epidermis, resulting in denaturation and contraction of collagen fibers.

    The DENSITY is an RF (radiofrequency), software-controlled electrosurgical device used for electrocoagulation of soft tissue and hemostasis.

    The DENSITY consists of the following components:
    •Electrosurgical Unit - Main body
    •Handpiece
    •6 different electrode tips
    •Neutral electrode pad and neutral electrode pad cable, cleared under K201685
    •Foot switch
    •Power cord

    AI/ML Overview

    This document is a 510(k) Premarket Notification for the medical device "DENSITY", an electrosurgical unit for dermatologic and general surgical procedures. It seeks substantial equivalence to a predicate device, the Thermage FLX System.

    Based on the provided text, the device DENSITY is an electrosurgical cutting and coagulation device. The document does not describe an AI/ML medical device or provide information on acceptance criteria and studies related to AI/ML performance metrics such as accuracy, precision, recall, or human reader improvement with AI assistance. Instead, it focuses on demonstrating substantial equivalence based on technical characteristics and safety standards for an electrosurgical device.

    Therefore, many of the requested items related to AI/ML device performance and testing (e.g., acceptance criteria for AI, sample sizes for AI test sets, expert ground truth, MRMC studies, standalone AI performance) are not applicable or cannot be extracted from this document.

    However, I can extract the information pertinent to the device's substantial equivalence and non-clinical performance data.

    Here's a breakdown of the available information:

    1. A table of acceptance criteria and the reported device performance:

    The document does not specify "acceptance criteria" in terms of performance metrics like sensitivity or specificity for a diagnostic/AI device. Instead, "acceptance" appears to be based on demonstrating substantial equivalence to a predicate device through conformity to recognized standards and comparable technical specifications.

    Here's a table summarizing the comparison with the predicate device (Thermage FLX System K170758), which serves as the basis for demonstrating equivalence in performance and safety for this type of device:

    Feature/Acceptance Criteria (Implied Comparability)Proposed Device (DENSITY) Performance/CharacteristicPredicate Device (Thermage FLX System K170758) Performance/Characteristic
    Indications for Use"DENSITY indicated for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis.""The radiofrequency-energy only delivery components of the Thermage FLX System are indicated for use in: • Dermatologic and general surgical procedures for electrocoagulation and hemostasis;"
    Output Energy TypeRadio FrequencyRadio Frequency
    User InterfaceColor Touch PanelColor Touch Panel
    Operating Frequency6.78MHz6.78MHz
    Max Power400 W400W
    Electrode Tip Types3 types (0.25 cm², 4 cm²)4 types (0.25 cm², 3.0 cm², 4 cm², 16cm²)
    Coolant SolutionCryogenCryogen
    Temperature Range65~75°C65~75°C
    RF Time50 ~ 800 ms50 ~ 1000 ms
    Impedance75 - 400 Ω75 - 400 Ω
    Communication to GeneratorLCD touch screenLCD touch screen
    Style of Electrode TipMonopolar, BipolarMonopolar
    Number of Active Electrodes per Applicator- I-Tip(Around eye) I, F-Tip(Face): 1ea- I-Tip II(Around eye), F-Tip II(Face): 2ea- B-Tip(Body), B-Tip(Body)II: 4ea- EYE Tip 0.25: 1ea- TOTAL Tip 3.0: 1ea- NEW TOTAL TIP 4.0: 1ea- BODY TIP 16.0: 4ea
    Coolant Control ParametersType of gas: 1234ze; Method of gas control: Handpiece solenoid control operation; Mouth valve specifications: 12V; Solenoid valve specifications: 1.8V 24.8 BARType of gas: 1234ze; Method of gas control: Handpiece solenoid control operation; Mouth valve specifications: 12V; Solenoid valve specifications: 1.8V 24.8 BAR

    Key takeaway on performance: The document states: "There are no significant differences between DENSITY and the predicate devices (K170758) that would adversely affect the use of the product. It is substantially equivalent to this device in design, function, and technical characteristics." "Any minor differences in the human interface and accessories design do not raise any new types of safety and effectiveness issues, as verified by performance testing."

    2. Sample sizes used for the test set and the data provenance:

    • Test Set Sample Size: "No clinical performance testing was performed." The evaluation relies on non-clinical performance data and comparison to the predicate device's established safety and effectiveness.
    • Data Provenance: Not applicable as no clinical test set data was used.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable as no clinical test set requiring expert ground truth was performed for this 510(k) submission.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable as no clinical test set requiring adjudication was performed.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This is not an AI/ML device, and no MRMC study was conducted.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is not an AI/ML device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Not applicable as no clinical ground truth was established for this submission. The "ground truth" for demonstrating substantial equivalence is based on established technical standards and the performance of the predicate device.

    8. The sample size for the training set:

    • Not applicable as this is not an AI/ML device.

    9. How the ground truth for the training set was established:

    • Not applicable as this is not an AI/ML device.

    Summary of Device Acceptance (for this traditional medical device):

    The acceptance of the DENSITY device for marketing is based on demonstrating substantial equivalence to an already legally marketed predicate device (Thermage FLX System K170758). This is achieved through:

    • Non-clinical performance data:
      • Evaluation against recognized consensus standards for electrical medical equipment such as IEC 60601-1 (basic safety and essential performance), IEC 60601-1-2 (electromagnetic disturbances), IEC 60601-2-2 (high frequency surgical equipment), IEC 60601-1-6 (usability).
      • Software verification and validation in accordance withIEC 62304 (moderate level of concern) and IEC 62366 (usability engineering/Human Factor Engineering).
      • Biocompatibility testing documented according to ISO 10993 standards.
      • Risk analysis conducted based on ISO 14971.
    • Comparison of technical characteristics: As detailed in the table above, demonstrating analogous specifications (e.g., operating frequency, max power, temperature range, impedance, coolant type) to the predicate device.

    The document explicitly states: "No clinical performance testing was performed." This means the device's safety and effectiveness for its intended use were demonstrated through the non-clinical testing and the comparison to a legally marketed predicate, rather than through clinical trials with human subjects.

    Ask a Question

    Ask a specific question about this device

    K Number
    K183284

    Validate with FDA (Live)

    Date Cleared
    2020-01-15

    (415 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    INTRAcel RF Microneedle System is indicated for use in dermatological and general surgical procedures for electrocoagulation, hemostasis and the percutaneous fractional treatment of the skin in monopolar and bipolar mode.

    Device Description

    The INTRAcel RF Microneedle System comprises of two primary components; an RF Generator with User Interface Software and a disposable Electrode Insertion Device with an integrated cable.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and supporting studies for the INTRAcel RF Microneedle System:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria / Test DescriptionReported Device Performance
    Electrical Safety & EMCIEC 60601-1:2012 (Medical electrical equipment - General requirements for basic safety and essential performance)Met all acceptance criteria in accordance with IEC 60601-1:2012.
    IEC 60601-1-6:2010, AMD1:2013 (Medical electrical equipment - General requirements for basic safety and essential performance - Collateral standard: Usability)Met all acceptance criteria in accordance with IEC 60601-1-6:2010, AMD1:2013.
    IEC 60601-2-2:2009 (Medical electrical equipment - Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories)Met all acceptance criteria in accordance with IEC 60601-2-2:2009.
    Performance TestingPerformance Test Report for ESU (Electrosurgical Unit)Met the acceptance criteria.
    INTRAcel Tip Count TestMet the acceptance criteria.
    Verification Performance of INTRAcel TipMet the acceptance criteria.
    Biocompatibility TestingCytotoxicityMicroneedle electrodes are biocompatible.
    SensitizationMicroneedle electrodes are biocompatible.
    Intracutaneous ReactivityMicroneedle electrodes are biocompatible.
    Sterilization & Shelf LifeEthylene Oxide Sterilization ResidualsMet the acceptance criteria.
    Accelerated aging testMet the acceptance criteria.
    Software Verification & ValidationSoftware tested against established Software Design Specifications to assure performance as intended; Device Hazard Analysis completed, risk control implemented; acceptance criteria of each module and interaction of processes met.The software for this device was considered a "moderate" level of concern. All acceptance criteria for software development, verification, and validation were met, and the device performs as intended, supporting substantial equivalence and safe operation.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state a "test set" in the context of clinical images or patient data that would require sample sizes and data provenance like country of origin or retrospective/prospective nature. The studies described are primarily non-clinical, focusing on device performance, electrical safety, biocompatibility, and software validation.

    • Animal Performance Testing (Wound Healing): The "study was conducted to see the wound healing response following FRM treatment for both human and porcine skin." While human skin is mentioned, the detailed procedural description (biopsy time points, H&E, HSP47, RT-PCR) is for "Micro-pig covering 10 weeks." The specific number of human or porcine subjects is not provided. The study appears to be prospective as it involves active treatment and observation over time.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    This information is not provided in the document. The studies detailed are technical (electrical, biocompatibility, software) or animal studies for wound healing, which do not typically involve human expert adjudication for ground truth establishing in the sense of medical image interpretation.

    4. Adjudication Method for the Test Set

    This information is not applicable/provided for the types of studies performed (technical bench testing, software validation, animal studies). There's no indication of a test set requiring adjudication in the context of expert consensus.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study and Effect Size

    A MRMC comparative effectiveness study was not performed or reported in this document. The device is an electrocoagulation and microneedling system; therefore, its performance is assessed directly through physical and electrical properties, biological safety, and tissue response, rather than as an aid to human interpretation like an AI diagnostic tool.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    This concept is not applicable to the INTRAcel RF Microneedle System in the way it's usually applied to AI diagnostic algorithms. The device itself is an active treatment device, not a diagnostic algorithm. The software validation mentioned is for the control and functionality of the device, implying a "standalone" software performance in terms of controlling the hardware as intended, rather than interpreting medical data.

    7. Type of Ground Truth Used

    • For Electrical Safety & EMC: Ground truth would be defined by the specifications and pass/fail criteria of the referenced international standards (IEC 60601 series).
    • For Performance Testing: Ground truth would be defined by the device's design specifications and expected functional outputs (e.g., specific power output, tip count accuracy).
    • For Biocompatibility Testing: Ground truth is established by the results of standardized biological tests (cytotoxicity, sensitization, intracutaneous reactivity) against acceptance criteria for biocompatible materials.
    • For Sterilization & Shelf Life: Ground truth is established by the ability to achieve sterility (e.g., meeting sterility assurance levels) and maintain device integrity/functionality over time via accelerated aging.
    • For Software Verification & Validation: Ground truth is the established "Software Design Specifications" — the software's ability to operate the device as programmed and safely (per the "moderate" level of concern).
    • For Animal Performance Testing (Wound Healing): Ground truth is established through histological examination (H&E, HSP47 stain) and molecular analysis (RT-PCR for mRNA changes in collagen, HSPs, MMPs) of biopsied tissue, documenting physiological wound healing responses at different time points and energy levels.

    8. Sample Size for the Training Set

    This information is not provided and is specifically irrelevant as the document describes a physical medical device, not an AI/ML algorithm that requires a training set of data. The software mentioned handles device control, not predictive modeling or classification.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable as there is no "training set" in the context of an AI/ML model for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K162169

    Validate with FDA (Live)

    Date Cleared
    2016-10-27

    (85 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    It is indicated for incision, ablation, vaporization and coagulation of body soft tissues in medical specialties including aesthetic (dermatology and plastic surgery), podiatry, otolaryngology(ENT), gynaecology, neurosurgery, orthopaedics, general and thorasic surgery (including open and endoscopic), dental and oral surgery and genitourinary surgery. The use with the scanning unit is indicated for ablative skin resurfacing.

    Device Description

    The Jeisys EdgeOne CO2 Laser is comprised of the following major components:

    1. The main console unit
    2. Delivery handpieces (2 surgical and one fractional)
    3. Footswitch.
    4. Accessories
    AI/ML Overview

    This document is a 510(k) premarket notification for the Jeisys EdgeOne CO2 Laser and does not contain information about acceptance criteria or a study proving device performance in the context of an AI/ML medical device.

    The document details the device's substantial equivalence to predicate devices based on:

    • Indications for Use: The Jeisys EdgeOne CO2 Laser is indicated for incision, ablation, vaporization, and coagulation of body soft tissues in various medical specialties, and for ablative skin resurfacing with a scanning unit. Its indications are identical to the secondary predicate (K133895) and largely overlap with the primary predicate (K100590).
    • Technical Characteristics: A table compares the Jeisys EdgeOne CO2 Laser's technical specifications (e.g., laser type, wavelength, output power, pulse duration, scan area, spot size, operational modes) to those of the predicate devices. The specifications are largely identical to the primary predicate and comparable to the secondary predicate.
    • Safety and Effectiveness Information: The document states that the device's compliance with several IEC standards (IEC 60601-1, IEC 60601-2-22, IEC 60601-1-2) demonstrates its safety.

    Therefore, I cannot provide the requested information regarding acceptance criteria and a study proving device performance as none of that information is present in the provided text. The document focuses solely on establishing substantial equivalence to legally marketed predicate devices for a conventional medical device (a CO2 laser), not an AI/ML-driven one.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1