Search Filters

Search Results

Found 8 results

510(k) Data Aggregation

    K Number
    K253150

    Validate with FDA (Live)

    Date Cleared
    2025-10-24

    (29 days)

    Product Code
    Regulation Number
    890.5650
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Cryon-X Pro
    The private-label Cryon-X Pro device combines cold, heat, contrast, and compression therapy. It is intended to treat post-surgical and acute injuries to reduce edema, swelling, and pain for which cold and compression are indicated. It is intended to treat post-traumatic and post-surgical medical and/or surgical conditions for which localized thermal therapy (hot or cold) is indicated.

    The device may optionally provide DVT therapy, intended to reduce the risk of the formation of deep venous thrombosis (DVT) by aiding blood flow back to the heart via lower extremity limb compression.

    This device is intended to be used by or on the order of licensed healthcare professionals in rehabilitation facilities, outpatient clinics, athletic training settings, and home settings.

    Cryon-X One
    The private-label Cryon-X One device combines cold and compression therapy. It is intended to treat post-surgical and acute injuries to reduce edema, swelling, and pain for which cold and compression are indicated. It is intended to be used by or on the order of licensed healthcare professionals in rehabilitation facilities/hospitals, outpatient clinics, athletic training settings, and home settings.

    Device Description

    This submission includes two models JKH-152 (private labelled as Cryon-X Pro) and JKH-151 (private labelled as Cryon-X One) as the subject devices. The private-label Cryon-X Pro has all the same features of combining cold, heat, contrast, and compression therapies as its original 510(k) cleared device in K223541, and Cryon-X One has the simplified features of cold and compression, which are the same as those of its original 510(k) cleared device in K223541. Each of the subject devices is an AC powered, software-controlled multimodality device, intended to be used by or on the order of licensed healthcare professionals in rehabilitation facilities/hospitals, outpatient clinics, athletic training settings, and home settings.

    The subject device is a prescriptive device, which is intended to treat post-surgical and acute injuries to reduce edema, swelling, and pain for which cold and compression are indicated. It is intended to treat post traumatic and post-surgical medical and/or surgical conditions for which localized thermal therapy are indicated. It is optionally intended to reduce the risk of the formation of deep venous thrombosis (DVT) by aiding blood flow back to the heart via lower extremity limb compression.

    The subject device and its accessories are clean and non-sterile. The device has a power switch, and then conducts the remaining operations on its touch screen, including the treatment temperature adjustment, air compression adjustment, and treatment time adjustment. The device works by circulating cooled or heated water and air through a treatment wrap that is placed on the treatment body area. The cooled or heated water circulates through the treatment wrap and provides cold or hot therapy, and the air compression inflates and deflates the treatment wrap to compress around the treatment body area. A connecting tube/hose is used to connect the device to the treatment wrap. The disposable or reusable wraps come with a variety of options to cover different body areas, including the universal, back, shoulder, ankle/foot, hip, knee, etc. To avoid any potential adverse skin reactions such as redness, irritation, and cod/hot injury, the sock/clothing should be worn by the patient prior to use.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K240788

    Validate with FDA (Live)

    Date Cleared
    2025-06-04

    (439 days)

    Product Code
    Regulation Number
    890.5300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Apply stationary use of ultrasound to: Generate deep heat within body tissues for the treatment of selected medical conditions such as the relief of pain, the relief of muscle spasms, the treatment of joint contractures, and the local increase in circulation.

    Apply continuous movement of ultrasound for:

    1. Pain.
    2. Pain relief, muscle spasms, and joint contractures.
    3. Relief of pain, muscle spasms, and joint contractures that may be associated with:
      • Adhesive capsulitis,
      • Bursitis with slight calcification,
      • Myositis,
      • Soft tissue injuries, and
      • Shortened tendons due to past injuries and scar tissues.
    4. Relief of pain, muscle spasms, and joint contractures resulting from:
      • Capsular tightness, and
      • Capsular scarring.
    5. Localized increase in blood flow.
    6. Increased range of motion of contracted joint using heat and stretch techniques.
    Device Description

    As a portable and rechargeable prescriptive device in the home or clinical/hospital setting, the subject device is intended to apply ultrasound to the patient's treatment site. It is demonstrated ultrasound can be used to apply therapeutic deep heat for the treatment of medical conditions including relief of pain, muscle spasms, and joint contractures.

    It is intended to be used for 30 minutes per cycle (up to 4 cycles) in the home or clinical/hospital setting by or under the direction of a medical professional to apply ultrasound to the patient's treatment site. The device automatically alerts the patient in case of improper application or performance. All the control components and the rechargeable battery of the device are protectively housed, and the device is connected to one or two applicators of ultrasound.

    The subject device consists of the main device, the applicator, the charger/adapter, the liquid and/or solid ultrasound gel, and the sticky pad. For the best result, use the ultrasound gel and the sticky pad supplied.

    AI/ML Overview

    This looks like a fascinating case study. Based on the provided FDA 510(k) clearance letter and summary, here's an analysis of the acceptance criteria and the study that proves the device meets them.

    It's important to note that this document is for an Ultrasound Stimulator (Ultrasonic Diathermy), which delivers therapeutic ultrasound, not an AI/Software as a Medical Device (SaMD) that typically involves image analysis or diagnostic support. Therefore, many of the typical acceptance criteria and study designs related to diagnostic accuracy, MRMC studies, and nuanced ground truth establishment for AI models (as implied by the original request's detailed points) are not applicable to this type of physical therapy device.

    The "acceptance criteria" here are primarily based on demonstrating substantial equivalence to predicate devices through engineering specifications, performance testing, and safety profiles, rather than clinical efficacy studies in the way an AI diagnostic tool would require.


    Acceptance Criteria and Study Proving Device Meets Criteria

    Given that the device is an Ultrasound Stimulator aiming for 510(k) clearance based on substantial equivalence, the "acceptance criteria" are not framed in terms of diagnostic performance metrics (e.g., sensitivity, specificity, AUC) but rather in terms of demonstrating that its technical characteristics, performance, and safety are equivalent to legally marketed predicate devices. The "study" proving it meets these criteria is a comprehensive set of non-clinical, bench performance tests and comparisons.


    1. Table of Acceptance Criteria and Reported Device Performance

    For an Ultrasound Stimulator, the "acceptance criteria" are primarily a demonstration of equivalence to the predicate devices across various technical and functional parameters, and compliance with relevant safety standards. The "reported device performance" are the results of the non-clinical tests and direct comparisons to the predicate devices.

    Parameter/Acceptance Criteria CategorySpecific Acceptance Criteria (Demonstrated Equivalence/Compliance)Reported Device Performance (Subject Device)
    Intended UseIdentical or substantially equivalent indications to predicate device.Apply stationary/continuous ultrasound for deep heat, pain relief, muscle spasms, joint contractures, increased circulation, range of motion (identical to primary predicate, similar to other predicate).
    Product CodeIdentical to predicate device.IMI, PFW (Identical to predicate ManaSport+, and PFW to sam 2.0)
    Prescription StatusIdentical (Prescription Use).Prescription Use (Identical).
    Power SourceSimilar, safe power delivery.100~240 Vac with 5Vdc Input and rechargeable 3.7Vdc battery (Identical to primary predicate, similar to other predicate).
    Number of Outputs1 or 2 outputs.1 or 2 (Identical to sam 2.0, similar to ManaSport+ which is 1).
    Software/Microprocessor ControlPresence and safe operation.Yes (Identical).
    Safety FeaturesAutomatic No-Load Trip, Automatic Shut Off, User Override Control, Indicator Display (On/Off, Low Battery).All 'Yes' for these features (Identical where predicate information available).
    Timer RangeFunctionally acceptable and safe timer range.30 minutes (per cycle), max 4 cycles (Longer than primary predicate (20min), shorter than predicate (240min), deemed safe by temperature control).
    Compliance with Voluntary StandardsCompliance with relevant IEC/ISO electrical safety, EM compatibility, and biocompatibility standards.Yes (Complies with ANSI AAMI ES60601-1, IEC 60601-1-2, IEC 60601-2-5, ISO 10993-5, ISO 10993-10, IEC 60601-1-6, IEC 60601-1-11).
    BiocompatibilityNon-toxic, non-irritating materials.Yes (Identical).
    SterilityNon-sterile device.Non-Sterile (Identical).
    Housing Construction MaterialCompatible material.ABS (Identical).
    Physical Dimensions/WeightSimilar physical characteristics not affecting safety/effectiveness.Console/Generator: 138×62×22 mm, 0.18 kg (Similar to predicates). Treatment Head: 61 x 45 x 12 mm, 0.047 kg (Similar to predicates).
    Functions and DesignUltrasound generation.Ultrasound (Identical).
    Frequency1.5 MHz or similar.1.5 MHz ± 10% (Identical to primary predicate, similar to other predicate's 3MHz).
    Leakage CurrentBelow maximum limit.<0.3mA (Identical).
    Crystal MaterialPZT or similar.PZT (Identical).
    Technology of Ultrasound GenerationPiezoelectric.Piezoelectric (Identical).
    Output ModeContinuous Wave - 100% duty cycle.Continuous Wave - 100% duty cycle (Identical).
    Power (Output, Temporal Avg., Effective Intensity, Beam Max Intensity)Within specified ranges and tolerances, consistent with therapeutic effect and safety.Max Output Power: 0.60 W ± 20% (Similar to predicates). Temporal Average Power: 0.60 W ± 20% (Identical where available). Max Effective Intensity: 0.15 ± 20% (Identical or similar to predicates, within IEC deviation). Beam Max Intensity: 0.15 ± 20% (Identical or similar to predicates, within IEC deviation).
    Duty Factor100%.100% (Identical).
    Beam TypeCollimated or Divergent.Collimated (Note: Predicate sam 2.0 is Divergent, but both are acceptable).
    Applicator SizeEquivalent treatment area.Area: 3.9 cm² (Identical to primary predicate, similar to other predicate).
    Maximum Patient Contact Surface TemperatureMeets safety standards (e.g., IEC 60601-2-5, section 201.11).Meets IEC 60601-2-5, section 201.11, with an enforced overheating temperature protection to prevent exceeding 43°C. (Identical or safer than predicate 44°C).
    Peak Temperature Rise vs. Time and Tissue DepthDemonstrated safe heating profile over maximum treatment time.8.6°C at 1 cm, 6.1°C at 2 cm, 2.1°C at 3 cm (Max treatment time: 30 min per cycle, 4 cycles). Deemed safe due to 43°C cap. (Similar to predicates' reported temperature rises).
    Beam Non-Uniformity Ratio (BNR)<5:1 or similar.<5 (Identical or similar to predicates).
    Effective Radiating Area (ERA)3.9 cm² or similar.3.9 ± 20% (Identical to primary predicate, similar to other predicate).
    Software VerificationCarried out according to FDA guidance.Verification carried out according to FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices.
    Bench Performance TestingSuccessful completion as per established protocols.Evaluation of power output, total treatment time, electrical safety, EMC, device operation, usability, biocompatibility, thermal safety.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: This is a physical device (Ultrasound Stimulator), not a diagnostic AI. Therefore, the concept of a "test set" in the context of a dataset for an AI model (like images or patient records) does not apply. The "test set" would be the number of devices/units subjected to engineering and performance verification tests. This information is typically not specified in a 510(k) summary; manufacturers test a representative number of units to ensure compliance with standards.
    • Data Provenance (e.g., country of origin of the data, retrospective or prospective): This is not applicable to a physical therapy device like an ultrasound stimulator. The "data" refers to engineering measurements and test results, not patient data. The device is manufactured by JKH Health Co., Ltd. in China, and presumably, the non-clinical tests were conducted at their facilities or accredited test labs.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • Not Applicable. For a device like an Ultrasound Stimulator, there is no "ground truth" to be established by experts in the sense of image interpretation for diagnosis. The "ground truth" for performance is based on defined engineering parameters, physical measurements, and compliance with international standards (e.g., IEC, ISO). These are verified by qualified test engineers and technicians using calibrated equipment, not by clinical experts like radiologists.

    4. Adjudication Method (e.g., 2+1, 3+1, none) for the Test Set

    • Not Applicable. Adjudication methods like 2+1 or 3+1 are used in studies involving human interpretation (e.g., radiologists reviewing images) to establish a consensus "ground truth." This device does not involve human interpretation as part of its function or testing for 510(k) clearance. Compliance is determined by objective measurements against predefined specifications and standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No, an MRMC comparative effectiveness study was not done. This type of study is specifically designed for evaluating diagnostic AI systems (or human performance with and without AI assistance) where human perception and interpretation are key. An Ultrasound Stimulator is a therapeutic device, not a diagnostic one, and its 510(k) clearance path relies on demonstrating substantial equivalence in engineering, performance, and safety, not on improving human diagnostic accuracy.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable in the traditional sense of AI algorithm performance. This is an Ultrasound Stimulator, a physical device. It has software, but the "performance" tested is its ability to generate ultrasound at specific parameters and safely, not to process data or make diagnostic outputs. Its core function is a physical treatment delivery, verified through bench testing, not an "algorithm-only" diagnostic output. The software verification mentioned (according to FDA Guidance for Software Contained in Medical Devices) ensures the software itself is designed, developed, and tested appropriately, but this isn't analogous to testing the diagnostic performance of an AI algorithm.

    7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)

    • Engineering Specifications and International Standards: The "ground truth" for this device is its adherence to predefined engineering specifications and compliance with established international safety and performance standards (e.g., IEC 60601 series, ISO 10993 series). These are objective, measurable criteria. For instance, ground truth for power output is a specific wattage with an allowed tolerance, measured by calibrated equipment, not a clinical diagnosis.

    8. The Sample Size for the Training Set

    • Not Applicable. This device is not an AI model that requires a "training set" of data for machine learning.

    9. How the Ground Truth for the Training Set Was Established

    • Not Applicable. As there is no training set for an AI model, the question of establishing its ground truth does not apply.

    In summary: The provided document is for a Class II Ultrasound Stimulator, a therapeutic device. Its premarket clearance (510(k)) is based on demonstrating substantial equivalence to predicate devices, primarily through comprehensive non-clinical bench testing to verify performance parameters, electrical safety, electromagnetic compatibility, biocompatibility, and thermal safety in accordance with recognized international standards. The evaluation process for such a device does not involve the AI-specific criteria (like test set data provenance, expert adjudication, MRMC studies, or training sets for AI models) that were part of the initial request, as those are typically relevant for diagnostic AI/SaMD submissions.

    Ask a Question

    Ask a specific question about this device

    K Number
    K240986

    Validate with FDA (Live)

    Device Name
    Cold Compression
    Date Cleared
    2024-05-01

    (21 days)

    Product Code
    Regulation Number
    890.5650
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Cold Compression is intended to treat post-surgical and acute injuries to reduce edema, swelling, and pain for which cold and compression are indicated.
    Cold Compression is intended to be used by or on the order of licensed healthcare professionals in rehabilitation facilities/hospitals, outpatient clinics, athletic training settings, and home settings.

    Device Description

    The subject device is an AC powered, software-controlled multimodality device, and is intended to be used by or on the order of licensed healthcare professionals in rehabilitation facilities/hospitals, outpatient clinics, athletic training settings, and home settings.
    The subject device features cold therapy and compression therapy. Cold Compression is intended to treat post-surgical and acute injuries to reduce edema, swelling, and pain for which cold and compression are indicated.
    The device has a power/time button, temperature adjustment button, pressure setting button; Patients can adjust the temperature, compression pressure, and duration of treatment according to their needs.
    The device has a bucket body and a lid. The ice and water can be added by opening the lid, and the loading level of the ice and water can be indicated/seen.
    The subject device and its packaging/clothing are clean and non-sterile. A connecting tube is used to connect the subject device to the wraps come with a variety of reusable cold treatment inflators for cold and compression treatment of the universal, back, shoulder, ankle/foot, hip, and knee.

    AI/ML Overview

    The provided text is a 510(k) Summary for a medical device called "Cold Compression". This document aims to demonstrate that a new device is substantially equivalent to a legally marketed predicate device, rather than proving the device meets specific acceptance criteria based on a study of its performance in a clinical or analytical setting.

    Therefore, the input document does not contain the information requested regarding:

    • A table of acceptance criteria and reported device performance.
    • Sample size used for a test set and data provenance.
    • Number of experts and their qualifications for establishing ground truth.
    • Adjudication method for a test set.
    • Multi-Reader Multi-Case (MRMC) comparative effectiveness study.
    • Standalone (algorithm only) performance study.
    • Type of ground truth used.
    • Sample size for the training set.
    • How ground truth for the training set was established.

    The document states that "Clinical testing was not needed in support of this 510(k) application." (Section 8) and focuses on demonstrating substantial equivalence through comparison with predicate devices and non-clinical tests for electrical safety and electromagnetic compatibility.

    Instead, the document provides:

    Non-Clinical Tests Performed:

    • Tests: Non-clinical tests were performed to validate the design and assure conformance with voluntary design standards related to medical device electrical safety and electromagnetic compatibility.
    • Standards:
      • IEC 60601-1 "Medical Electrical Equipment Part 1: General requirements for basic safety and essential performance (IEC 60601-1)".
      • IEC 60601-1-2 "Medical Electrical Equipment Part 1-2: General Requirements for Basic Safety and Essential Performance - Collateral standard: Electromagnetic Compatibility - Requirements and Tests".

    Comparison Table (Table 1):
    This table compares the subject device ("Cold Compression") with two predicate devices (K223541 and K183702) across various technical characteristics and specifications to establish substantial equivalence. While it doesn't present "acceptance criteria" in the traditional sense of a performance study, it details the characteristics where the device's performance parameters are deemed acceptable due to their similarity to a legally marketed device.

    CharacteristicSubject Device PerformancePredicate Device Performance (relevant to equivalence)Equivalence Conclusion
    Product CodesIRP and ILOPrimary Predicate: IRP, ILO, and JOW Predicate: IRP and ILOThe subject device has the product codes covered by the primary predicate device, and also identical to the predicate device.
    Indications for UseTo treat post-surgical and acute injuries to reduce edema, swelling, and pain for which cold and compression are indicated. Used by or on order of licensed healthcare professionals in rehabilitation facilities/hospitals, outpatient clinics, athletic training settings, and home settings.Primary Predicate: Combines cold, heat, contrast, and compression therapy. Intended to treat post-surgical and acute injuries to reduce edema, swelling, and pain. Also provides DVT therapy. Used by or on order of licensed healthcare professionals in rehabilitation facilities, outpatient clinics, athletic training settings, and home settings. Predicate: To treat post-surgical and acute injuries to reduce edema, swelling, and pain where cold and compression are indicated. Used by or on order of licensed healthcare professionals in hospitals, outpatient clinics, athletic training settings, or home settings.The subject device has the Indications for Use covered by the primary predicate device, and also substantially equivalent to the predicate device.
    Intended UsersHealth Care Professionals and lay users (under prescription), with adult use; children and elderly under direct medical professional supervision.Primary Predicate and Predicate: Identical.Identical
    Prescription or OTCPrescriptionPrimary Predicate and Predicate: PrescriptionIdentical
    Power Source(s)100-240 VAC, 50/60 HzPrimary Predicate and Predicate: 100-240 VAC, 50/60 HzIdentical
    Compliance with Voluntary StandardsYesPrimary Predicate and Predicate: YesIdentical
    Safety (Electrical, Mechanical, Chemical, Thermal, Radiation)YesPrimary Predicate and Predicate: YesIdentical
    TherapyCold and Compression. Work together or independently.Primary Predicate: Cold, Heat, Contrast and Compression. Predicate: Cold and Compression. Work together or independently.The subject device has the therapy covered by the primary predicate device, and also identical to the predicate device. The difference between the subject device and the primary predicate device is the former does not provide heat or contrast therapy.
    Cold Therapy Temperature45°F - 55°FPrimary Predicate: Default: 50°F; Custom: 41°F-55°F Predicate: 45°F - 60°FEquivalent Note 1: The temperature range of cold therapy for the subject device is within that of the primary predicate device and the predicate device. Adjustable by healthcare professionals. Insignificant difference.
    Heat TherapyNot AvailablePrimary Predicate: Default: 105°F; Custom: 105°F-109°F Predicate: Not AvailableIdentical to the predicate device. The difference from the primary predicate device (lack of heat/contrast therapy) is noted but does not raise new safety/effectiveness issues for this specific device.
    Compression Setting0 - 60 mmHgPrimary Predicate: Alternate mode: 15 -75 mm Hg; Continuous mode: 15 -75 mm Hg Predicate: Low (0-25 mmHg), Regular (0-50 mmHg)Equivalent Note 2: The pressure range for the subject device is within that of the primary predicate device and slightly higher than the predicate device. Adjustable by healthcare professionals. Insignificant difference.
    Reservoir Fluid Capacity620 mLPrimary Predicate: 350 mL Predicate: Not AvailableDifferent, but it will not raise any new issue of safety or effectiveness.
    Recommended CoolantTap WaterPrimary Predicate: 90% Distilled Water, 10% Isopropyl Alcohol Predicate: Tap WaterIdentical to the predicate device. The difference from the primary predicate is noted but does not raise new safety/effectiveness issues.
    Water Cooling SourceIcePrimary Predicate: Vapor compression Predicate: IceIdentical to the predicate device. The difference from the primary predicate is noted but does not raise new safety/effectiveness issues.
    User InterfaceKeypad with indicator lights or LCD screenPrimary Predicate: Touch Screen Predicate: Keypad with indicator lightsDifferent from the primary predicate device and similar to the predicate device. It will not raise any new issue of safety or effectiveness.
    DimensionsL235xW235xH280mmPrimary Predicate: L295xW285xH295 mm Predicate: Not AvailableDifferent. The difference of dimensions does not change the product performance or parameters, which will not raise any new issue of the safety or effectiveness.
    Weight Approx.2.4kgPrimary Predicate: 8.2kg Predicate: Not AvailableDifferent. The difference of weight does not change the product performance or parameters, which will not raise any new issue of the safety or effectiveness.
    Types of GarmentsVarious anatomical thermal garments for: Back, Elbow/Universal, Shoulder, Knee, Ankle, Hip.Primary Predicate: Various anatomical thermal garments for: Back, Elbow, Shoulder, Knee, Ankle, Hip. DVT Garments: Calf and Foot. Predicate: Various anatomical thermal garments for: Back, Universal, Shoulder, Knee, Foot/Ankle, Hip.The subject device has the garments covered by the primary predicate device, and also substantially equivalent to the predicate device.
    Biocompatibility of Patient Contacting MaterialBiocompatiblePrimary Predicate and Predicate: BiocompatibleIdentical
    Sterile/Non-SterileNon-sterilePrimary Predicate and Predicate: Non-sterileIdentical
    Cleaning Disinfection Validation of LabelingYesPrimary Predicate and Predicate: YesIdentical
    Operational and Environmental ConditionsNormal working ambient temperature: 540°C; humidity: 15%90%; Store and transport ambient temperature: -2570°C; humidity: 1090%; Atmospheric pressure: 70~106kPaPrimary Predicate: Normal working ambient temperature: 1627°C; humidity: Below 60%; Store and transport ambient temperature: 150°C; humidity: Below 60%; Atmospheric pressure: 70106kPa Predicate: Normal working ambient temperature: 540°C; humidity: 15%90%; Store and transport ambient temperature: -2570°C; humidity: 1090%; Atmospheric pressure: 70106kPaIdentical to the predicate device. The difference between the subject device and the primary predicate device is the former does not provide heat or contrast therapy. (Note: The operational and environmental conditions are listed as identical to the second predicate device, not the primary one, despite the "difference between the subject device and the primary predicate device" wording).
    Ask a Question

    Ask a specific question about this device

    K Number
    K240011

    Validate with FDA (Live)

    Date Cleared
    2024-02-23

    (52 days)

    Product Code
    Regulation Number
    870.5800
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Lymphedema Compression is comprised of a gradient compression sleeve and a portable intermittent pump to provide graduated compression in both sustained and intermittent settings for use in both the hospital and outpatient setting. It is pre-set to the default setting of 50 mmHg that cannot be adjusted or can be adjusted by the physician to a pressure within the specified range. It is intended for use in:

    • Treatment of lymphedema
    • Treatment of chronic venous insufficiency
    • Treatment and promotion of healing of stasis dermatitis and venous stasis ulcers
    • Reducing venous leg ulcer healing time
    • Reducing edema due to venous stasis
    • Enhancing venous return
      The device is intended for home, and hospital use.
    Device Description

    As a portable and rechargeable prescriptive device, the Lymphedema Compression is intended to be used in the home or clinical/hospital setting by or under the direction of a medical professional. It applies pressure to treat lymphedema and other edematous conditions.
    Supplied clean and non-sterile, the Lymphedema Compression utilizes the pneumatically controlled cuff and is actuated by an electronically controlled air pump unit. All pump, battery and control components are protectively housed in a plastic case of the control unit, which is attached to an inflatable cuff/sleeve. The Lymphedema Compression contains an ON/OFF button, a SET button, and a display for user interface. In addition, there is also a port on the control unit for connecting the battery charger/AC adapter plug. The cuff component consists of multiple air chambers/bladders encased inside a soft nylon fabric or equivalent biocompatible fabric for increased patient comfort and biocompatibility compliance.
    In operation, the user simply press the ON/OFF button to turn on the control unit, to which a cuff containing air chambers/bladders is connected. And the control unit then inflates the cuff to the default pre-determined pressure (50 mmHg) that cannot be adjusted to a pressure within the specified range via the SET button. Once the cuff pressure of the multiple air chambers/bladders reaches the pre-determined or adjusted level, the pump is turned off, and the cuff deflates to the ambient pressure through a valve inside the control unit. After the rest period, the cycle of inflation and deflation repeats until the device is turned off automatically or manually.

    AI/ML Overview

    The provided FDA 510(k) summary (K240011) describes the "Lymphedema Compression" device and its substantial equivalence to a predicate device, ManaFlow (K200353).

    1. A table of acceptance criteria and the reported device performance

    The provided document doesn't explicitly list "acceptance criteria" in the typical sense of numerical thresholds for performance metrics. Instead, the substantial equivalence relies on demonstrating that the subject device's characteristics and performance are identical to or similar to the predicate device. The primary "acceptance criterion" is functional and technical equivalence to the predicate.

    Here's a table based on the comparison provided in the 510(k) summary:

    Feature/CriterionAcceptance Criteria (Based on Predicate Device: ManaFlow K200353)Reported Subject Device Performance (Lymphedema Compression)Equivalence
    Intended UseTo provide graduated compression in both sustained and intermittent settings for: Treatment of lymphedema, chronic venous insufficiency, stasis dermatitis/venous stasis ulcers (healing & reducing healing time), reducing edema due to venous stasis, enhancing venous return. For home and hospital use.Identical to Predicate: To provide graduated compression in both sustained and intermittent settings for: Treatment of lymphedema, chronic venous insufficiency, stasis dermatitis/venous stasis ulcers (healing & reducing healing time), reducing edema due to venous stasis, enhancing venous return. For home and hospital use.Identical
    Pressure SettingPre-set to 50 mmHg (cannot be adjusted) or adjustable (20-80 mmHg) by physician.Pre-set to 50 mmHg (cannot be adjusted) or adjustable (20-80 mmHg) by physician.Identical
    Prescription/OTCPrescriptionPrescriptionIdentical
    Power Source5V DC supply (100-240 VAC input) and 3.7V rechargeable battery5V DC supply (100-240 VAC input) and 3.7V rechargeable batteryIdentical
    Battery Charge TimeApprox. 4 hours from depleted stateApprox. 4 hours from depleted stateIdentical
    Power Supply (Input/Output)Input: 100-240 Vac, 50-60 Hz; Output: 5 Vdc @ 2 AmpInput: 100-240 Vac, 50-60 Hz; Output: 5 Vdc @ 2 AmpIdentical
    Internal Rechargable BatteriesYesYesIdentical
    Compliance with Voluntary StandardsYes (Electrical Safety, Mechanical Safety, Chemical Safety, Thermal Safety, Radiation Safety)Yes (ANSI AAMI ES60601-1, IEC 60601-1-2)Identical
    Functions and DesignAids venous return by cyclic, intermittent, pneumatic pressure application (inflation-deflation) to compress extremities.Aids venous return by cyclic, intermittent, pneumatic pressure application (inflation-deflation) to compress extremities.Identical
    ContraindicationsMUST NOT be used for: Suspected DVT, ischemic vascular disease, severe arteriosclerosis, pulmonary edema, severe congestive heart failure, thrombophlebitis, active infection; on legs interfering with: vein ligation, gangrene, dermatitis, open wounds, recent skin graft, massive edema, extreme deformity; on any neuropathy; on insentitive extremities; where increased venous/lymphatic return is undesirable.Identical to Predicate.Identical
    Target Population / UsersPatients needing venous return and lymphedema treatmentPatients needing venous return and lymphedema treatmentIdentical
    Where UsedHome, Hospital, Surgery Center, Altitude travel, areas of limited mobilityHome, Hospital, Surgery Center, Altitude travel, areas of limited mobilityIdentical
    ApplicationNon-invasive / externalNon-invasive / externalIdentical
    PortabilityPortable, ambulantPortable, ambulantIdentical
    Basis of operationAids venous return by cyclic, intermittent, pneumatic pressure application (inflation-deflation) to compress extremities.Aids venous return by cyclic, intermittent, pneumatic pressure application (inflation-deflation) to compress extremities.Identical
    Anatomical SiteLeg (for predicate)ExtremitiesIdentical or Similar
    System managementMicroprocessorMicroprocessorIdentical
    Pressure SourceMicro pump controlled by microprocessorMicro pump controlled by microprocessorIdentical
    Operating ModesPreset and adjustable modesPreset and adjustable modesIdentical
    Working PressurePreset at 50 mmHg and adjustable from 20-80 mmHgPreset at 50 mmHg and adjustable from 20-80 mmHgIdentical
    Cycle TimeInflation in sequence (foot-up), then deflation; repeats.Inflation in sequence (foot-up), then deflation; repeats.Identical
    System diagnosticsAudible and visual alarms for system faultsAudible and visual alarms for system faultsIdentical
    Air deliveryVia flexible plastic tube(s) connected directly to air bladder.Via flexible plastic tube(s) connected directly to air bladder.Identical
    SterilityClean / non-sterileClean / non-sterileIdentical
    Leg cuff usageSingle Patient UseSingle Patient UseIdentical
    Material UsedAir bladder chambers in soft, non-latex medical fabric or equivalent for comfort/biocompatibility.Air bladder chambers in soft, non-latex medical fabric or equivalent for comfort/biocompatibility.Identical
    FasteningSnap and screwSnap and screwIdentical
    BiocompatibilityBiocompatibleBiocompatibleIdentical
    SoftwareModerateModerateIdentical
    Operating Temperature+10 ℃ (50 ºF) to +40 ºC (104 ºF)+10 ℃ (50 ºF) to +40 ºC (104 ºF)Identical
    Humidity30%-75%30%-75%Identical
    Cleaning and DisinfectingSpecific instructions for pump unit and cuffs; air dry.Specific instructions for pump unit and cuffs; air dry.Identical
    DisposalElectromechanical device with PCBs and rechargeable batteries; not for landfill; consult local requirements/recycle center.Electromechanical device with PCBs and rechargeable batteries; not for landfill; consult local requirements/recycle center.Identical

    2. Sample size used for the test set and the data provenance

    The document explicitly states: "Clinical testing was not needed in support of this 510(k) application."

    Therefore, there was no clinical test set of patient data, no sample size of patients, and no data provenance relevant to clinical performance evaluation. The "test set" for the device's technical specifications consists of the individual components and the assembled device that underwent non-clinical bench testing.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. Since no clinical testing was performed, there was no ground truth established by experts on patient cases. The evaluation of electrical safety, EMC, and physical/performance requirements relies on adherence to published voluntary standards and internal company specifications, not expert consensus on clinical outcomes.

    4. Adjudication method for the test set

    Not applicable. As no clinical testing or human reader study was conducted, there was no need for an adjudication method.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC study was conducted. This device is a pneumatic compression device, not an AI-powered diagnostic or assistive tool for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This is a physical medical device (pneumatic compression) and does not involve an algorithm with standalone performance in the context of diagnostic interpretation. Its "performance" is based on mechanical and electrical parameters (e.g., pressure output, cycle time, safety features).

    7. The type of ground truth used

    The "ground truth" for this submission is adherence to recognized voluntary standards for medical device safety and performance (e.g., ANSI AAMI ES60601-1, IEC 60601-1-2) and the technical specifications and functional characteristics of the legally marketed predicate device. Bench tests validated that the subject device met its own design specifications, which were shown to be identical or very similar to the predicate.

    8. The sample size for the training set

    Not applicable. This device is not an AI/ML algorithm that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable. (See #8)

    Ask a Question

    Ask a specific question about this device

    K Number
    K223541

    Validate with FDA (Live)

    Date Cleared
    2023-10-27

    (336 days)

    Product Code
    Regulation Number
    890.5650
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Cold/Hot Compression combines cold, heat, contrast, and compression therapy. It is intended to treat post-surgical and acute injuries to reduce edema, swelling, and pain for which cold and compression are indicated. It is intended to treat post traumatic and post-surgical medical and/or surgical conditions for which localized thermal therapy (hot or cold) are indicated.

    Cold/Hot Compression also provide DVT therapy. It is intended to reduce the risk of the formation of deep venous thrombosis (DVT) by aiding blood flow back to the heart via lower extremity limb compression.

    ColdHot Compression is intended to be used by, or on the order of, licensed health care professionals in rehabilitation facilities, outpatient clinics, athletic training settings, and home settings.

    Device Description

    The subject device is a prescriptive device, which combines cold/hot therapy and air compression. It is intended to treat post-surgical and acute injuries to reduce edema, swelling, and pain for which cold and compression are indicated. It is intended to treat post traumatic and post-surgical medical and/or surgical conditions for which localized thermal therapy (hot or cold) are indicated. It is intended to reduce the risk of the formation of deep venous thrombosis (DVT) by aiding blood flow back to the heart via lower extremity limb compression.

    The subject device comes with the thermal-therapy wraps and the DVT wraps. The thermal-therapy wrap is nylon fabric on one side and velvet cloth on the other side; the DVT wrap is velvet on both sides. To avoid any potential adverse skin reactions such as redness, irritation, and cold/hot injury, the sock/clothing should be worn by the patient prior to use.

    The subject device includes a main device and an optional DVT device, and has a limited shelf life of 3 years, based on the charge retention characteristics of the device's battery.

    AI/ML Overview

    This document is a 510(k) summary, which focuses on demonstrating substantial equivalence to a predicate device rather than presenting a detailed clinical study with acceptance criteria and performance metrics in the typical sense.

    Therefore, the information requested might not be fully available or explicitly stated in the provided text as it would be in a clinical study report.

    Based on the provided text, here's what can be extracted and inferred:

    1. Table of acceptance criteria and the reported device performance

    The concept of "acceptance criteria" here is primarily tied to demonstrating equivalence to the predicate device. The performance is assessed by comparing technical specifications.

    FeaturePredicate Device Performance / Range (Therm-X)Subject Device Performance / Range (Cold/Hot Compression)Equivalence Note
    Cold Therapy (Default)34°F, 45°F, 55°F50°FEquivalent (Note 1)
    Cold Therapy (Custom)34°F - 55°F41°F-55°FEquivalent (Note 1)
    Cold Therapy (Lowest)34°F or 40°F41°FEquivalent (Note 1)
    Heat Therapy (Default)105°F, 107°F, 110°F105°FEquivalent (Note 2)
    Heat Therapy (Custom)105°F - 110°F105°F-109°FEquivalent (Note 2)
    Heat Therapy (Highest)110°F109°FEquivalent (Note 2)
    Edema Pressure (Lowest)5 mm Hg or 20 mmHg15 mm HgEquivalent (Note 3)
    Edema Pressure (Highest)70 mmHg75 mm HgEquivalent (Note 3)
    DVT Pressure (Calf)50-70 mmHg55 mmHgEquivalent (Note 4)
    DVT Pressure (Foot)90 - 130 mmHg130 mmHgEquivalent (Note 4)
    Cycle Length (Longest)40 or 60 minutes60 minutesEquivalent (Note 5)
    Contrast Therapy (Heat)105°F105°FEquivalent (Note 6)
    Contrast Therapy (Cold)38°F49°FEquivalent (Note 6)
    Contrast Cycle (Heat)3-10 minutes10 minutesEquivalent (Note 7)
    Contrast Cycle (Cold)3-10 minutes20 minutesEquivalent (Note 7)
    Reservoir Fluid Capacity650 mL350 mLEquivalent (Note 8)

    Notes on Equivalence (as explained in the document):

    • Note 1 (Cold Therapy): The subject device's cold therapy temperatures are within or similar to the range of the predicate/reference devices, and are adjustable by professionals. Differences are considered insignificant and do not raise new safety/effectiveness concerns.
    • Note 2 (Heat Therapy): Similar to cold therapy, the heat therapy temperatures are within or similar to the predicate/reference devices and adjustable. Differences are insignificant.
    • Note 3 (Edema Pressure): The subject device's lowest pressure is within the predicate's range and similar to the reference. The highest pressure is slightly higher but still adjustable by professionals. Differences are insignificant.
    • Note 4 (DVT Pressure): The subject device's DVT pressures for calf and foot fall within or are slightly higher than the predicate/reference devices. Differences are insignificant.
    • Note 5 (Cycle Length): The subject device's longest cycle length is similar to or longer than the predicate's, though shorter than the reference. Cycle length is adjustable. Differences are insignificant.
    • Note 6 (Contrast Therapy Temp): The subject device's contrast therapy temperatures are either similar to the predicate or identical to the reference. No new safety/effectiveness issues.
    • Note 7 (Contrast Cycle Length): The contrast therapy cycle length is slightly different from the predicate but identical to the reference. No new safety/effectiveness issues.
    • Note 8 (Reservoir Fluid Capacity): The subject device's capacity is smaller than the predicate's but larger than the reference's. This difference does not raise new safety/effectiveness issues.

    2. Sample size used for the test set and the data provenance

    The document explicitly states: "Clinical testing was not needed in support of this 510(k) application."

    Therefore, there is no test set in the sense of patient data from a clinical study. The "testing" involved non-clinical bench tests and comparison of specifications to predicate devices. The data provenance is from internal JKH Health Co., Ltd. testing and comparison to publicly available predicate and reference device information.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. As no clinical testing was performed, there was no "test set" requiring expert ground truth establishment for a patient-based diagnostic or therapeutic outcome.

    4. Adjudication method for the test set

    Not applicable. There was no clinical test set requiring adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a physical medical device (Cold/Hot Compression unit), not an AI-powered diagnostic or assistive tool for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an algorithm-only device.

    7. The type of ground truth used

    The "ground truth" for the non-clinical tests would be the measured physical parameters of the device (e.g., actual temperature output, actual pressure levels, cycle times) against the device's design specifications and the specifications of the predicate devices. For safety standards, the ground truth is adherence to the requirements of the standards (e.g., ANSI AAMI ES60601-1, IEC 60601-1-2).

    8. The sample size for the training set

    Not applicable. There is no mention of a "training set" as this is not an AI/machine learning device. The design was likely based on engineering principles and comparison to existing devices.

    9. How the ground truth for the training set was established

    Not applicable. No training set was used.

    Ask a Question

    Ask a specific question about this device

    K Number
    K211050

    Validate with FDA (Live)

    Date Cleared
    2022-11-28

    (599 days)

    Product Code
    Regulation Number
    880.2910
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Infrared Thermometer is intended temperature measurement at home or hospital for people of all ages without contact to human body.

    Device Description

    Not Found

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the JKH Health Co., Ltd. Infrared Thermometer:

    Based on the provided FDA 510(k) clearance letter and Indications for Use, the document does not contain detailed information about the specific acceptance criteria for performance, the study design, sample sizes, expert qualifications, or ground truth establishment that would typically be found in a clinical study report or detailed submission summary.

    The document primarily focuses on the regulatory clearance process, confirming that the device (Infrared Thermometer, Models: JKH-300, JKH-301, JKH-302, JKH-303, JKH-304) is substantially equivalent to legally marketed predicate devices. It states its intended use for non-contact temperature measurement for people of all ages at home or in a hospital setting.

    Therefore, many of the requested details cannot be extracted from this particular document. I will highlight what can be inferred or is directly stated, and mark what is not present.


    1. A table of acceptance criteria and the reported device performance

    Acceptance CriteriaReported Device Performance
    Not specified in documentNot specified in document
    (Typically would include accuracy, precision, bias, measurement range, e.g., "Accuracy +/- 0.2°C at 37.0°C")(e.g., "Accuracy was measured as +/- 0.15°C for 95% of readings")

    2. Sample sized used for the test set and the data provenance

    • Sample size: Not specified.
    • Data provenance: Not specified (e.g., country of origin, retrospective or prospective).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Number of experts: Not specified.
    • Qualifications of experts: Not specified.

    4. Adjudication method for the test set

    • Adjudication method: Not specified.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • MRMC study: Not applicable. This device is a standalone thermometer, not an AI-assisted diagnostic tool that would typically involve human readers interpreting results.
    • Effect size of human improvement with AI: Not applicable.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Standalone performance: Yes, this is an inference given the device's nature as an infrared thermometer. Its performance would be evaluated as a standalone device without human interpretation of its core function (temperature measurement). However, the specific details of this testing (e.g., how accuracy was validated against a reference standard) are not provided in this document.

    7. The type of ground truth used

    • Type of ground truth: Not explicitly stated. For a thermometer, the ground truth would typically be a highly accurate reference thermometer measurement (e.g., a laboratory-grade precision thermometer, or an invasive core body temperature measurement in a clinical setting for calibration/validation).

    8. The sample size for the training set

    • Sample size for training set: Not applicable and not specified. Infrared thermometers typically do not involve "training sets" in the machine learning sense. Their calibration and performance are based on physical principles and manufacturing processes, demonstrated through verification and validation testing.

    9. How the ground truth for the training set was established

    • Ground truth for training set: Not applicable and not specified. See point 8.

    Summary of what's provided for the JKH Health Co., Ltd. Infrared Thermometer:

    • Device Name: Infrared Thermometer, Model: JKH-300, JKH-301, JKH-302, JKH-303, JKH-304
    • Regulation Number/Name: 21 CFR 880.2910, Clinical Electronic Thermometer
    • Regulatory Class: Class II
    • Product Code: FLL
    • Intended Use: Temperature measurement at home or hospital for people of all ages without contact to the human body.
    • Type of Use: Over-The-Counter Use
    • Regulatory Status: Cleared via 510(k) (K211050) as substantially equivalent to predicate devices.

    What is NOT provided in this FDA 510(k) clearance letter (and would typically be found in a detailed test report or submission summary):

    • Specific numerical acceptance criteria for accuracy, precision, or other performance characteristics.
    • Detailed results of the device's performance against these criteria.
    • Information about the clinical or laboratory studies conducted (e.g., number of subjects, age groups, measurement sites, comparison methods, statistical analysis).
    • Any information regarding training data, ground truth establishment, or expert involvement, as these are more pertinent to AI/ML-driven devices.
    Ask a Question

    Ask a specific question about this device

    K Number
    K202839

    Validate with FDA (Live)

    Date Cleared
    2021-01-21

    (118 days)

    Product Code
    Regulation Number
    870.5800
    Age Range
    21 - 65
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Compression Therapy Device is intended to be an easy to use portable by a physician, for use in the home or clinical setting to help prevent the onset of DVT in patients by stimulating blood flow in the extremities (stimulating muscle contractions). This device can be used to:

    • · Aid in the prevention of DVT;
    • · Enhance blood circulation;
    • · Diminish post-operative pain and swelling;
    • · Reduce wound healing time;
    • · Aid in the treatment and healing of stasis dermatitis, venous stasis ulcers, arterial and diabetic leg ulcers, chronic venous insufficiency and reduction of edema in the lower limbs.
      The unit can also be used as an aid in the prophylaxis for DVT by persons expecting to be stationary for long periods of time.
    Device Description

    Compression Therapy Device

    AI/ML Overview

    Based on the provided text, the document is an FDA 510(k) clearance letter for a "Compression Therapy Device." This letter does not contain the type of detailed study information (acceptance criteria, specific performance metrics, sample sizes, ground truth establishment, expert qualifications, etc.) that would be required to answer your request.

    The letter confirms that the device is substantially equivalent to legally marketed predicate devices and lists its indications for use. It discusses regulatory compliance, labeling, and other administrative aspects. It does not provide a technical report of device performance or the study design and results validating its functionality against specific acceptance criteria.

    Therefore, I cannot provide the requested information based on the given text.

    To answer your request, you would need a different type of document, such as a summary of safety and effectiveness data (SSED), a clinical study report, or a comprehensive technical submission to the FDA.

    Ask a Question

    Ask a specific question about this device

    K Number
    K162517

    Validate with FDA (Live)

    Date Cleared
    2017-04-14

    (217 days)

    Product Code
    Regulation Number
    882.5890
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    TENS (Modes 1, 2, 4, 5, 6, 8)

    To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, arm, and leg, due to strain from exercise or normal household and work activities.

    It is also intended for symptomatic relief and management of chronic, intractable pain and relief of pain associated with arthritis.

    PMS (also called EMS, Modes 1, 3, 7)

    To stimulate healthy muscles in order to improve and facilitate muscle performance. To be used for the improvement of muscle tone and firmness, and for strengthening muscles in the arms, abdomen, legs, and buttocks. Not intended for use in any therapy or for the treatment of any medical conditions or diseases.

    It is also intended to temporarily increase local blood circulation in the healthy muscles of lower extremities.

    Heating Mode Temporary relief of minor aches and pains.

    Device Description

    Electronic Pulse Stimulator delivers electric pulses generated to the user's body areas such as the back neck and foot through the electrodes. The portable and compact device has multiple modes of different pulse frequencies, covering TENS and PMS that is also called Electrical Muscle Stimulation (EMS). It includes operating elements of ON/OFF button, intensity increase button, intensity decrease button, mode selection button, and/or timer selection button, and could be attached to electrodes. In addition, the device may also provide heat/temperature. While used in the heating mode, the device is coupled with electronically controlled electrodes to provide automatic thermal heat to the skin with the maximum temperature of 43 °C.

    The device could be easily operated through its buttons to manually realize its functions, such as turning on/off, increasing/decreasing intensity, changing mode/timer, and providing heat/temperature if needed. The optional wireless control via a remote or Bluetooth APP could provide a secondary operation way to the user, who could be able to wirelessly realize the functions mentioned above.

    The electrodes cleared include the electrode pads and electrode garments, which could be packaged separately and/or together with the subject device.

    AI/ML Overview

    This document is a 510(k) summary for an Electronic Pulse Stimulator (device K162517). It describes the device, its intended use, and demonstrates substantial equivalence to predicate devices (K153520 and K070299). Since this is a 510(k) summary for a Transcutaneous Electrical Nerve Stimulator (TENS) and Powered Muscle Stimulation (PMS) device, the acceptance criteria and study information will focus on its electrical and heating parameters rather than clinical performance metrics typically associated with AI/ML devices (like sensitivity, specificity, AUC).

    Here's a breakdown of the requested information based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" in the traditional sense of a predetermined numerical threshold for a clinical outcome (e.g., "sensitivity must be >90%"). Instead, it demonstrates "substantial equivalence" based on various technical specifications and intended use. The "reported device performance" is essentially a comparison of these technical specifications between the subject device and the predicate devices.

    The table below summarizes the comparison from Table 1 in the document for the subject device (PL-029K13, as it includes the heating mode) and its primary predicate (PL-029K6, K153520) for electrical stimulation, and the heating predicate (ezFit Digital Heating TENS, K070299) for the heating function. The ranges and specific values are presented as performance data.

    FeaturePredicate Device (K153520) Performance (PL-029K6)Predicate Device (K070299) Performance (ezFit Digital Heating TENS)Subject Device (K162517) Performance (PL-029K13)Acceptance Criteria/Basis for Equivalence
    TENS Mode
    Intended Use (TENS)Temporary relief of pain (sore/aching muscles, chronic intractable pain, arthritis)Symptomatic relief and management of chronic intractable pain.Temporary relief of pain (sore/aching muscles, chronic intractable pain, arthritis)Substantially Equivalent to K153520 for TENS.
    Max Output Voltage (500Ω)Mode 2: 39.2; Mode 3: 64; Mode 4: 56.8; Mode 5: 33.2; Mode 6: 31.6UndisclosedMode 2: 31.2; Mode 3: 46.0; Mode 4: 42.0; Mode 5: 27.6; Mode 6: 27.6; Mode 7: 40.8; Mode 8: 23.2Similar technical characteristics topredicate. The 7th and 8th modes of the subject device are similar to the 6 initial modes of the predicate.
    Max Output Voltage (2kΩ)Mode 2: 82.4; Mode 3: 84; Mode 4: 79.2; Mode 5: 70.4; Mode 6: 67.2UndisclosedMode 2: 68.0; Mode 3: 90.4; Mode 4: 68.8; Mode 5: 60.0; Mode 6: 60.0; Mode 7: 84.0; Mode 8: 50.4Similar technical characteristics topredicate.
    Max Output Voltage (10kΩ)Mode 2: 129; Mode 3: 120; Mode 4: 84.8; Mode 5: 121; Mode 6: 124UndisclosedMode 2: 118; Mode 3: 124; Mode 4: 78.4; Mode 5: 115; Mode 6: 115; Mode 7: 124; Mode 8: 99.2Similar technical characteristics topredicate.
    Max Output Current (500Ω)Mode 2: 78.4; Mode 3: 128; Mode 4: 113.6; Mode 5: 64.4; Mode 6: 63.2UndisclosedMode 2: 62.4; Mode 3: 92.0; Mode 4: 84.0; Mode 5: 55.2; Mode 6: 55.2; Mode 7: 81.6; Mode 8: 46.4Similar technical characteristics topredicate.
    Max Output Current (2kΩ)Mode 2: 41.2; Mode 3: 42; Mode 4: 39.6; Mode 5: 35.2; Mode 6: 33.6UndisclosedMode 2: 34.0; Mode 3: 45.2; Mode 4: 34.4; Mode 5: 30.0; Mode 6: 30.0; Mode 7: 42.0; Mode 8: 25.2Similar technical characteristics topredicate.
    Max Output Current (10kΩ)Mode 2: 12.9; Mode 3: 12; Mode 4: 8.5; Mode 5: 12.1; Mode 6: 12.4UndisclosedMode 2: 11.8; Mode 3: 12.4; Mode 4: 7.84; Mode 5: 11.5; Mode 6: 11.5; Mode 7: 12.4; Mode 8: 9.92Similar technical characteristics topredicate.
    Pulse Period (mSec)10-833Undisclosed5.6-806Similar technical characteristics topredicate.
    Frequency (Hz)Mode 2: 69.4; Mode 3: 13.0-52.1; Mode 4: 1.2; Mode 5: 96.2; Mode 6: 96.2UndisclosedMode 2: 73.5; Mode 3: 13.7-59.5; Mode 4: 1.24; Mode 5: 104.1; Mode 6: 104.1; Mode 7: 20.8; Mode 8: 178.5Similar technical characteristics topredicate.
    Max Phase Charge (500Ω)Mode 2: 15.1; Mode 3: 25.6; Mode 4: 18.2; Mode 5: 12.8; Mode 6: 10.1UndisclosedMode 2: 11.5; Mode 3: 16.9; Mode 4: 15.5; Mode 5: 10.2; Mode 6: 10.2; Mode 7: 15.0; Mode 8: 8.54Similar technical characteristics topredicate.
    Max Current Density (500Ω)Mode 2: 2.18; Mode 3: 3.56; Mode 4: 3.16; Mode 5: 1.84; Mode 6: 1.76UndisclosedMode 2: 2.23; Mode 3: 3.29; Mode 4: 3.00; Mode 5: 1.97; Mode 6: 1.97; Mode 7: 2.91; Mode 8: 1.66Similar technical characteristics topredicate.
    Max Average Power Density (500Ω)Mode 4: 0.03; Mode 2: 1.14; Mode 3: 0.64-2.56; Mode 5: 1.13; Mode 6: 0.85UndisclosedMode 2: 0.92; Mode 3: 0.32-1.37; Mode 4: 0.04; Mode 5: 1.04; Mode 6: 1.04; Mode 7: 0.46; Mode 8: 1.26Similar technical characteristics topredicate.
    PMS Mode
    Intended Use (PMS)Stimulate healthy muscles (improve/facilitate muscle performance, muscle tone/firmness, strengthening, increase local blood circulation)Not applicable (K070299 only TENS and heating)Stimulate healthy muscles (improve/facilitate muscle performance, muscle tone/firmness, strengthening, increase local blood circulation)Substantially Equivalent to K153520 for PMS.
    Heating Mode
    Intended Use (Heating)Not applicable (K153520 only TENS and PMS)Temporary relief of minor aches and pains and muscle spasms.Temporary relief of minor aches and pains.Substantially Equivalent to K070299. The technology is similar, and the indications are comparable.
    Heating SettingN/AAdjustable (36-42 °C)Low and HighSimilar technology (delivery of heat).
    Maximum Temperature SettingN/A42 °C43 °CSimilar maximum temperature setting, within acceptable safety margins for human skin.
    Skin TemperatureN/AN/A (implied safe operation across range)Reached 41 °C in 20 min and remained stable.Demonstrates safe temperature limits and stable operation.
    Prescription/OTCOTCPrescriptionOTCTransition from prescription to OTC for heating is supported by substantial equivalence in technology and safety. The electrical stimulation aspects are OTC as well (similar to K153520).

    For non-clinical tests, the "acceptance criteria" were compliance with specific voluntary standards:

    • IEC 60601-1 (medical electrical equipment safety and essential performance)
    • IEC 60601-1-2 (electromagnetic compatibility)
    • IEC 60601-2-10 (particular requirements for nerve and muscle stimulators)
    • FDA Guidance for Premarket Submissions for Software Contained in Medical Devices.

    The reported device performance for these non-clinical tests is that the device "conformed" and "met the requirement of safety" for the standards and guidance, respectively.

    2. Sample Size Used for the Test Set and the Data Provenance

    The document explicitly states: "The subject device does not conduct, nor rely upon, clinical tests to determine substantial equivalence. Non-clinical tests were performed on the subject device..."

    Therefore, there is no clinical test set, human subject data, or data provenance (country of origin, retrospective/prospective) described in this 510(k) summary for the purpose of demonstrating substantial equivalence. The evaluation relies on technical comparisons to predicate devices and adherence to engineering and safety standards.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    As no clinical test set was used, there was no ground truth established by experts in a clinical context for this submission. The "ground truth" for the non-clinical tests would be the specifications and requirements of the regulatory standards themselves, which are established by expert committees in their respective fields (e.g., electrical engineering, medical device safety).

    4. Adjudication Method for the Test Set

    Since no clinical test set with human subject outcomes was used for demonstrating substantial equivalence, no adjudication method (like 2+1 or 3+1 consensus) was applied.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done

    No, a Multi Reader Multi Case (MRMC) comparative effectiveness study was not done. This type of study is typically performed to evaluate the impact of an AI/ML device on human reader performance, which is not applicable to an Electronic Pulse Stimulator submission focused on electrical and heating parameters and substantial equivalence to existing devices.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was Done

    No, a standalone performance study in the context of an AI/ML algorithm was not done. The "performance" in this submission refers to the technical specifications of the device (e.g., output voltage, current, temperature) and its compliance with safety standards.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc)

    For the technical specifications, the "ground truth" is established by engineering measurements against reference standards (e.g., voltmeters, ammeters, thermometers calibrated to international standards) to verify the device's output. For compliance with voluntary standards (IEC 60601-1, -1-2, -2-10), the "ground truth" is the requirements outlined in those international standards. For software, it's compliance with FDA guidance for software in medical devices.

    8. The Sample Size for the Training Set

    Since this is not an AI/ML device, there is no training set sample size described. The device's operation is based on fixed electrical circuits and heating elements, not learned models.

    9. How the Ground Truth for the Training Set Was Established

    As there is no training set, this question is not applicable.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1