Found 22 results

510(k) Data Aggregation

    K Number
    K222026

    Validate with FDA (Live)

    Date Cleared
    2023-07-14

    (368 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Zirconia Base is an abutment placed over Neodent Zirconia Implants in order to provide support for patient-specific prosthetic restorations, such as copings or crowns. It may be used for cement- or screw retained single unit restorations. All digitally designed copings and/or crowns to be used with the Neodent Zirconia Base Abutment System are intended to be sent to Straumann for manufacture at a validated milling center.

    Device Description

    The Zirconia Base abutment is used as two-piece abutment, where the base is premanufactured from zirconia and the top half is created via burn out coping or a digital workflow with CAD/CAM at a validated milling center. The top half and base pieces are cemented together to form the final abutment. This submission aims to introduce the conventional workflow for obtaining prostheses manually designed by the technician using the Zirconia Bases, already cleared under K201491. With that, the introduction of one more restoration material (IPS e.max® Press) for the top half confection is necessary. The conventional workflow, using the IPS e.max® Press as restoration material along with the suggested cement materials is equivalent to the design workflow indicated for reference device (K142890).

    AI/ML Overview

    The provided text describes the regulatory clearance of a dental implant system and focuses on demonstrating substantial equivalence to predicate devices, rather than establishing performance against specific acceptance criteria for an AI/ML device. Therefore, the information requested for an AI/ML device, such as specific acceptance criteria and detailed study designs involving test sets, ground truth establishment, and expert involvement, is not available in the provided document.

    The document primarily discusses:

    • Device: Neodent Implant System - Zirconia Implant System, specifically the Zirconia Base abutment.
    • Purpose of Submission: To introduce a conventional workflow for manually designed prostheses using Zirconia Bases and to include a new restoration material (IPS e.max® Press).
    • Method of Proof: Substantial Equivalence to legally marketed predicate devices, as per FDA 510(k) requirements. This involves comparing indications for use, technological characteristics, materials, and manufacturing processes, supported by performance data (mechanical testing, MR compatibility, sterilization validation, biocompatibility) demonstrating that the new workflow and material do not raise new questions of safety or effectiveness.

    Therefore, I cannot populate the requested tables and information as it pertains to AI/ML device acceptance criteria and study details. The document does not describe an AI/ML device or its evaluation.

    Ask a Question

    Ask a specific question about this device

    K Number
    K223638

    Validate with FDA (Live)

    Date Cleared
    2023-06-23

    (200 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Indications for Use for Helix Short Implant:

    The Neodent Implant System is recommended for surgical procedures on maxilla or mandible bones. It provides support for prosthetic components such as artificial teeth, thus restoring the chewing function. It may be used with single-stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading.

    Indications for Use for Mini Straight, Angled Abutment and screws:

    The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading.

    Indications for Use for Helix Short Attachment:

    The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. It may be used with single stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading.

    Indications for Use for Helix Short Healing Abutment:

    This product is used for the maintenance of soft tissue, during the bone integration stage of Neodent implants to be rehabilitated with the late loading technique. The Healing Abutment may be used in the implant installation surgery or the reopening surgery (second surgical stage).

    Indications for Use for Helix Short Cover Screw:

    This product is used for the maintenance of soft tissue, during the bone integration stage of Neodent implants to be rehabilitated with the late loading technique. The Cover Screw must remain intragingival, preventing the growth of tissue over the implant platform. The Cover Screw takes place in the conical region of the implant interface.

    Indications for Use for Temporary Abutment for Helix Short Implant

    The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. It may be used with single stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. Temporary Abutments are installed over Helix Short Implants and to provide support for prosthetic structures for up to 6 months.

    Indications for Use for Helix Short Titanium Base:

    The Helix Short Implant Titanium base that is placed over Neodent dental implants to provide support for customized prosthetic restorations, such as copings and crowns. It is indicated for single- and multiple-structure restorations, screw- or cement-retained on implants installed in the maxilla or mandible. All digitally-designed copings and/or crowns to be used with the Neodent Titanium Base System must be sent to Straumann for manufacture at a validated milling center.

    Device Description
    • All the subject devices are intended for single use;
    • The subject implants are provided sterile via gamma irradiation and the subject abutments are provided sterile via Ethylene Oxide to an SAL of 1x10-9;
    • The subject implants are manufactured of Commercially Pure Titanium (Grade 4), the subject abutments, prosthetic screws and abutment screws are manufactured of Titanium alloy Ti-6Al-4V ELI (ASTM F136) and the restorations placed onto Helix Short Titanium Bases are made of various top half materials.
    • The subject implants have a wide cone prosthetic interface with an internal hexagon for short implants lengths available from 4.0 to 7.0; cylindrical shape with double threads and conical apex with three helical flutes for diameters from 3.75 to 7.0mm.
    • The implants are provided in two different surfaces: Neoporos: Sand blasted and acid etched, and Acqua: Hydrophilic surface aggregated to the rough surface obtained by double treatment: abrasive blasting and acid subtraction.
    • The Helix Short Implant Titanium Base abutments is used as two-piece abutment, where the base is premanufactured from titanium alloy and the top half is created via burn out coping or a digital workflow with CAD/CAM at a validated milling center. The final top half can be created from Ticon, Coron, Zerion LT, IPS e.max, or PMMA. The top half and base pieces are cemented together to form the final abutment.
    AI/ML Overview

    Here's an analysis of the acceptance criteria and supporting studies for the Neodent Implant System - Helix Short Implant System, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The FDA 510(k) summary does not explicitly present a table of specific numerical acceptance criteria alongside reported performance for each test. Instead, it describes what was tested and states that the results "met the acceptance criteria" or demonstrated "equivalent performance." We can infer the general acceptance criterion from the study descriptions.

    Acceptance Criteria (Inferred)Reported Device Performance
    Mechanical Testing:
    Dynamic Fatigue Test (per ISO 14801): Equivalent fatigue strength to predicate devices.The tested subject devices exhibit a level of performance equivalent to that reviewed for the predicate devices.
    Torsion Test (static torsional loading): Meet acceptance criteria.The results met the acceptance criteria.
    Insertion Test: Evaluate insertion torque in various bone types.Performed to evaluate the insertion torque of the Helix Short System when it is inserted in bones type I, II, III, and IV. (No explicit "met criteria" statement, but implied by successful evaluation).
    Implant Surface Area Simulation and Pull Out Test: Demonstrate acceptable BIC rate and pull out resistance.Results demonstrated that the subject devices, despite having a slightly smaller surface area than the reference devices, presented a higher BIC rate and pull out resistance upon placement. Thus, the subject devices exhibit a level of benchtop performance equivalent to that reviewed for the reference device.
    MR Compatibility Testing:
    Not configure a new worst case compared to reference devices, allow safe scanning under established parameters.An assessment was made to demonstrate that the subject devices do not configure a new worst case and can be represented by the previously conducted studies reviewed for reference devices, since both have the same raw material and similar dimensions. The subject devices are therefore MR conditional devices and a patient treated with the subject devices can be safely scanned observing the parameters previously established per reference devices.
    Surface Treatment:
    Identical or similar chemical processes to predicate/reference devices to achieve roughness.The surface treatments applied to subject devices are identical to these applied and previously evaluated for primary predicate devices. Although some reference devices have a surface with different nomenclature, the chemical processes applied to the surface are similar to create the roughness surface. Acqua implants are submitted to an additional step in order to increase their hydrophilicity, as already cleared for primary predicate device.
    Sterilization Validation:
    Implants (Gamma Irradiation): Minimum Sterility Assurance Level (SAL) of 1 x 10^-6.A minimum Sterility Assurance Level (SAL) of 1 x 10^-6 has been validated based on ISO 11137-1 and ISO 11137-2.
    Abutments (Ethylene Oxide): Minimum Sterility Assurance Level (SAL) of 1 x 10^-6.A minimum Sterility Assurance Level (SAL) of 1 x 10^-6 has been validated based on ISO 11135-1.
    Endotoxin Testing (LAL):
    Meet pyrogen limit specifications (acceptance criteria for endotoxin levels).Subject devices meet the acceptance criteria for endotoxin levels as well as the primary predicate and reference devices.
    Shelf Life Validation:
    Packaging integrity maintained; established expiration date.The shelf life for devices provided sterile is 5 years, except Helix Short Acqua Implants that have a shelf life of 4 years. The expiration date was determined considering product integrity and packaging tests.
    Biocompatibility:
    No new issues compared to predicate/reference devices.No new issues of biocompatibility are raised for the subject devices when compared to primary predicate and reference devices. Therefore, no additional biocompatibility testing was required.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the numerical sample size for the test sets in the mechanical, MR compatibility, or other in-vitro studies. It refers to "tested subject devices" or "representative samples."

    • Data Provenance: The studies described (mechanical testing, MR compatibility, surface treatment, sterilization, endotoxin, shelf life, biocompatibility) are all benchtop/laboratory studies (in vitro). There is no mention of human subject data, animal data, or clinical retrospective/prospective data for these specific performance criteria. The manufacturer is JJGC Indústria e Comércio de Materiais Dentários S.A. (dba Neodent) based in Curitiba, Paraná, Brazil.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. As these are benchtop/laboratory performance studies, "ground truth" is established by adherence to recognized standards (e.g., ISO 14801, ISO 11137, ISO 11135, ISO 10993) and scientific methods, not by expert consensus in the typical sense for clinical data.

    4. Adjudication Method for the Test Set

    Not applicable. Adjudication methods like 2+1 or 3+1 are typically used for interpreting clinical images or patient outcomes, not for objective benchtop performance data against engineering standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    Not applicable. This device is an endosseous dental implant system, which is a physical medical device, not an AI-powered diagnostic or assistive tool. Therefore, MRMC studies involving human readers and AI assistance are not relevant.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not applicable, as this is a physical medical device, not an algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" for the performance claims is based on:

    • Adherence to International Standards: Performance measured against specifications outlined in standards like ISO 14801 (dynamic fatigue), ISO 11137 and ISO 11135 (sterilization), ISO 10993 series (biocompatibility).
    • Benchtop Test Results: Direct measurements and observations from the mechanical, physical, chemical, and biological laboratory tests performed on the device.
    • Comparison to Predicate Devices: Performance is frequently established by demonstrating equivalence or superiority to already legally marketed predicate devices, which serve as a benchmark.

    8. The Sample Size for the Training Set

    Not applicable. This is a physical medical device. The concepts of "training set" and "validation set" are primarily used for machine learning and AI, not for the regulatory approval of physical implants based on benchtop testing.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable for the same reasons as point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K223662

    Validate with FDA (Live)

    Date Cleared
    2023-04-20

    (135 days)

    Product Code
    Regulation Number
    880.6850
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The HS Surgical Kit Cases are indicated for the organization of surgical and/or prosthetic instruments during sterilization. storage, and transport. Use of this product facilitates the storage and organization of instruments during and after surgical procedures.

    Neodent instrument kit cases are intended to allow sterilization of the medical devices included.

    Neodent instrument kit cases must be wrapped in FDA-approved materials to maintain the sterility of the devices included.

    The kits should be placed in an FDA-approved sterilization wrap for the indicated cycles and sterilized by moist heat (steam) using one of the following cycles:

    Dynamic Air Removal (Pre-Vacuum): exposure at 132°C for 4 minutes, drying for 30 minutes.

    Gravity displacement: exposure at 132°C for 15 minutes, drying for 50 minutes.

    Neodent instrument kit cases are intended for sterilization of non-porous fillers.

    The combined weight of the HS Surgical Kit case and associated instruments is 302.88 grams.

    The weight of the empty case is approximately 214.85 grams. Neodent instrument kit cases should not be stacked during sterilization.

    Device Description

    The subject device kit cases are reusable rigid containers, comprising a case bottom (or base), a removable inner tray, and tray lid (lid). The subject device kits are to be used to organize and protect instruments and accessories that are to be sterilized by the healthcare provider. The subject kit cases are manufactured in autoclavable polymer. The design of the subject devices include grommets manufactured from medical grade silicone that retain the instruments within the tray. They also presents markings that guide instrument use during procedures. The subject device kit cases are provided nonsterile to the end-user.

    AI/ML Overview

    This is a medical device submission for the Neodent Implant System - Helix Short Surgical Kit Cases. The submission focuses on sterilization wrap containers (Class II, product code KCT).

    Here's an analysis of the acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance CriteriaReported Device Performance
    Manual Cleaning Validation:
    * Visual Inspection: No Visible SoilPassed
    * Hemoglobin Test: <2.2 µg/cm²Passed
    * Protein Test: <6.4 µg/cm²Passed
    Sterilization Validation (ANSI/AAMI/ISO 17665-1, ANSI/AAMI/ISO 17665-2):
    * All Biological Indicators must be incubated for at least 7 days at 55-60°C.Passed
    * All positive controls for SAL testing must show characteristic growth of the indicator organism.Passed
    Life Cycle (Simulate Usage) Testing (Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling Guidance):
    * The tested samples must withstand 100 cycles of use (cleaning, sterilization and functional tests) without compromising their functionalities.Passed
    Biocompatibility (ANSI/AAMI/ISO 10993-5 - Cytotoxicity):
    * Less than 30% cell proliferation inhibitionPassed
    Sterilization Cycles:
    * Gravity displacement: Sterilization temperature: 132 °C; Sterilization time: 15 minutes; Drying time: 50 minutes.The device demonstrated equivalency to the predicate device's cleared cycle parameters, which included these parameters (or similar shorter drying times for the predicate). The subject device's longer drying time (50 min) is within safe parameters.
    * Dynamic Air Removal (Pre-Vacuum): Sterilization temperature: 132 °C; Sterilization time: 4 minutes; Drying time: 30 minutes.The device demonstrated equivalency to the predicate device's cleared cycle parameters, which included these parameters (or similar shorter drying times for the predicate). The subject device's longer drying time (30 min) is within safe parameters.
    Sterile Barrier: Sterilization wrap, FDA-cleared for indicated method and cycles.Declared "Identical" to the predicate, which uses a sterilization pouch. This implies the device is designed to be used with an FDA-cleared wrap/pouch.

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not explicitly state the sample size for the individual tests (manual cleaning, sterilization, life cycle, cytotoxicity). It refers to "tested samples" for the life cycle testing, but no specific number is provided.

    The data provenance is not explicitly mentioned in terms of country of origin or whether studies were retrospective or prospective, beyond the fact that these are non-clinical performance data (laboratory testing).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    This information is not provided in the document. The studies are non-clinical (laboratory tests), and ground truth would have been established by the validated methodologies and acceptance criteria themselves, not by expert consensus on individual "cases."

    4. Adjudication Method for the Test Set:

    This information is not applicable as the tests are objective laboratory performance validations against predefined criteria, not subjective interpretations requiring "adjudication."

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    An MRMC comparative effectiveness study was not performed, nor is it applicable for this type of device (sterilization kit case). This device is not an AI-assisted diagnostic tool that would involve human readers.

    6. Standalone Performance Study:

    The studies conducted are standalone performance studies in the sense that they evaluate the device itself (e.g., its ability to withstand cleaning, facilitate sterilization, and maintain biocompatibility) against established standards and internal criteria, without direct human interaction as a primary variable. The document refers to them as "Non-Clinical Performance Data."

    7. Type of Ground Truth Used:

    The ground truth for the non-clinical performance studies is based on:

    • Established scientific and regulatory standards/guidance: e.g., ANSI/AAMI/ISO 17665-1, ANSI/AAMI/ISO 17665-2 for sterilization, ANSI/AAMI/ISO 10993-5 (Cytotoxicity), and FDA guidance for reprocessing medical devices.
    • Validated laboratory methodologies: Custom test methods for cleaning and specific incubation/growth criteria for biological indicators in sterilization.
    • Objective measurement: Visual inspection (for soil), quantifiable chemical tests (hemoglobin, protein), and cell proliferation inhibition measurements are objective.

    Essentially, the "ground truth" is defined by the objective pass/fail criteria derived from these standards and methods.

    8. Sample Size for the Training Set:

    This device is not an AI/machine learning product; therefore, the concept of a "training set" is not applicable.

    9. How the Ground Truth for the Training Set Was Established:

    As there is no training set, this information is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K220251

    Validate with FDA (Live)

    Date Cleared
    2022-09-06

    (218 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. It may be used with single stage procedures, for single- or multiple-unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. The Implant Narrow GM is indicated to replace upper lateral incisors, lower incisors or for retaining overdenture prostheses.

    Prosthetic component to be installed in the Narrow GM Implant, providing support for prosthetic structures. Indicated for temporary restorations with unitary structures, screwed, on implants installa or mandible. Temporary Abutments Narrow GM have a maximum duration of usage of 180 days.

    Titanium Base Abutment is a titanium base placed onto Neodent dental implants to provide support for customized prosthetic restorations as copings and crowns. It is indicated for cemented or screw mounted single-cemented applications on implants installed on the maxilla or mandible. All digitally designed copings and/or crowns to be used with the Titanium Base Abutment System are intended to be sent to Straumann for manufacture at a validated milling center.

    Device Description

    All the subject devices are intended for single use;
    The subject implants are provided sterile via gamma irradiation and the subject abutments are provided sterile via Ethylene Oxide to an SAL of 1x10 -;
    The subject implants are manufactured of Commercially Pure Titanium (Grade 4) and the subject abutments are manufactured of Titanium alloy Ti-6Al-4V ELI (ASTM F136),
    The subject implants present a Grand Morse prosthetic interface with internal indexer; cylindrical shape with double threads; cervical diameter equal to the implant body diameter (2.9 mm)
    The implants are provided in two different surfaces: Neoporos: Sand blasted and acid etched, and Acqua: Hydrophilic surface aggregated to the rough surface obtained by double treatment: abrasive blasting and acid subtraction.

    AI/ML Overview

    The provided text does not contain information about acceptance criteria and a study proving a device meets these criteria in the context of an AI/ML medical device. Instead, it is a 510(k) summary for a "Neodent Implant System - GM Narrow Implant System," which is a dental implant system, not an AI/ML device.

    The information provided pertains to demonstrating substantial equivalence of a new dental implant system to legally marketed predicate devices, focusing on:

    • Indications for Use: Comparing the intended purposes of the new implant system and its components (implants, conventional abutments, temporary abutments, titanium bases) to existing predicate devices.
    • Technological Characteristics: Comparing material, design (e.g., interface, shape, threads, gingival height), sterilization methods, and reusability.
    • Performance Data: Mentioning biocompatibility testing (ISO 10993 standards), mechanical testing (Dynamic fatigue per ISO 14801), and sterilization validation (ISO 11137 for gamma irradiation, ISO 11135 for Ethylene Oxide).

    Therefore, I cannot fulfill the request to describe the acceptance criteria and the study that proves the device meets the acceptance criteria for an AI/ML device based on the provided text. The text is relevant for a traditional medical device (dental implant system) and its substantial equivalence determination.

    Ask a Question

    Ask a specific question about this device

    K Number
    K203542

    Validate with FDA (Live)

    Date Cleared
    2021-08-02

    (242 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Mini Conical Abutments are indicated for use with Zygomatic Implants, in cases of severe jaw resorption, in order to restore patient aesthetics and chewing function. It may be used with single-stage procedures, for multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading.

    Device Description

    Mini Abutments 60°

    • Intended for single use;
    • Provided sterile via ethylene oxide gas;
    • Manufactured of titanium alloy (Ti6Al4V-ELI) per ASTM F136;
    • Available in different gingival heights;
    • Screw-retained to the implant;
    • Provided with an anti-rotational implant-to-abutment interface compatible with GM Zygomatic Implants;
    • Provided with coronal geometries in rotational (non-indexed) versions to support multi-unit restorations
    • . The subject devices are compatible with Impression copings, Provisional cylinders, Protective cylinders and Copings of the Neodent GM Line previously cleared per market.
    AI/ML Overview

    The provided document is a 510(k) Summary for the Neodent Implant System - Mini Abutment 60°. It describes the device, its indications for use, and a comparison to predicate devices, focusing on demonstrating substantial equivalence through non-clinical performance data.

    Based on the provided text, there is no information about a clinical study involving human subjects or AI performance. The summary explicitly states "SUMMARY OF NON-CLINICAL PERFORMANCE DATA". Therefore, most of the requested information regarding acceptance criteria, study type, sample sizes, experts, ground truth, and MRMC studies for AI devices is not available in this document.

    Here's what can be extracted:

    Acceptance Criteria and Reported Device Performance

    The document describes non-clinical acceptance criteria related to mechanical properties, sterilization, and biocompatibility.

    Acceptance Criteria CategorySpecific Test/EvaluationAcceptance CriteriaReported Device Performance/Conclusion
    Mechanical PropertiesDynamic fatigue test (per ISO 14801 and FDA guidance)Performance suitable for intended useDemonstrated performance suitable for intended use
    Mechanical PropertiesTorsion testing (for screws)Results met the acceptance criteriaResults met the acceptance criteria
    SterilizationEthylene Oxide sterilization (per ISO 11135)Minimum Sterility Assurance Level (SAL) of 1 x 10^-6Validated to a minimum SAL of 1 x 10^-6
    SterilizationEthylene oxide residuals (per ISO 10993-7)Within accepted limitsResiduals are within accepted limits
    BiocompatibilityCytotoxicity testing (per ISO 10993-5)(Not explicitly stated, but implied positive)Performed; reference to previous biocompatibility testing
    BiocompatibilityChemical characterization (per ISO 10993-18)(Not explicitly stated, but implied positive)Performed; reference to previous biocompatibility testing
    Shelf-lifeProduct and package stability (per ASTM F1980)(Not explicitly stated, but implied positive)Determined to be 5 years
    MRI SafetyMR CompatibilityNo new worst-case constructs, safe scanningSubject devices do not result in new worst-case constructs for MR compatibility; can be safely scanned observing established parameters

    Information not available in the document:

    1. Sample size used for the test set and the data provenance: Not applicable. The document describes non-clinical (laboratory) testing, not a test set of data from clinical subjects.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for clinical data is not relevant to these non-clinical tests.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: No MRMC study was done, as this is not an AI device.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: No standalone algorithm performance was done, as this is not an AI device.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not applicable, as this is not an AI device and no clinical ground truth was established for the non-clinical tests. The "ground truth" for these tests are objective measurements against established standards (e.g., ISO 14801, ISO 11135).
    7. The sample size for the training set: Not applicable, as this is not an AI device.
    8. How the ground truth for the training set was established: Not applicable, as this is not an AI device.

    In summary, this document pertains to a traditional medical device (dental abutment) and relies on non-clinical engineering and materials testing to demonstrate substantial equivalence, not AI algorithm performance or clinical trial data.

    Ask a Question

    Ask a specific question about this device

    K Number
    K203382

    Validate with FDA (Live)

    Date Cleared
    2021-03-15

    (118 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. The GM Easy Pack System is indicated to be used on Neodent implants to provide temporary support for prosthesis structure for up to 6 months.

    Device Description
    • Intended for single use;
    • The subject GM Helix Implants are available in diameters of 3.5, 3.75, 4.0 and 4.3 mm and in heights of 8, 10, 11.5 and 13 mm;
    • The subject GM Smart Abutments are available in diameters of 3.5 and 4.5 mm and gingival heights of 1.5, 2.5 and 3.5 mm. The GM Smart Abutments can be customized respecting a minimum post height of 4 mm. The GM Smart Abutments are being presented for the first time for FDA's evaluation;
    • . All digitally designed copings and/or crowns to be used with the GM Smart Abutments are intended to be sent to Straumann for manufacture at a validated milling center;
    • . The subject GM Healing Abutments are available in diameters of 3.5 and 4.5 mm and gingival heights of 1.5, 2.5 and 3.5 mm;
    • . Implant and Cover Screw delivered sterile via Gamma Radiation;
    • Sterile Abutments provided sterile via Ethylene Oxide;
    • The subject implant is manufactured of unalloyed titanium grade 4, according to ASTM F67;
    • . The GM Cover Screw, GM Smart Abutment, GM Healing Abutment and the Removable Screw are manufactured in titanium alloy Ti6Al4V-ELI according to ASTM F136;
    • The GM Helix Implants, Acqua and Neoporos, the GM Cover Screw and the Healings subject of this submission are exactly the same devices already cleared to market per K163194;
    • . The GM Smart abutments, subject of this submission, are new abutments being introduced in this premarket notification;
    • . The implant to abutment interface is a Grand Morse (GM) connection.
    AI/ML Overview

    The provided document is a 510(k) premarket notification for a dental implant system. It does not describe a study involving an AI/Machine Learning (AI/ML) device, nor does it present acceptance criteria or performance data for such a device.

    The document discusses the substantial equivalence of the Neodent Implant System - Easy Pack to previously cleared predicate devices (K163194, K191191, K182620, K193592). The data presented is related to the physical and biological characteristics of the dental implants and abutments.

    Therefore, I cannot extract the information required in your request regarding acceptance criteria and performance data for an AI/ML device, as the document does not pertain to such a device. All the acceptance criteria mentioned and the "results" describe testing for traditional medical device characteristics such as sterilization, shelf life, biocompatibility, pyrogenicity, MRI compatibility, and mechanical properties (torsion, insertion) of the implants and abutments.

    Specifically, points 2 through 9 of your request are not applicable to the content of this document, as they relate to AI/ML device studies (e.g., sample size for test/training sets, data provenance, expert ground truth, MRMC studies, standalone performance).

    Ask a Question

    Ask a specific question about this device

    K Number
    K203618

    Validate with FDA (Live)

    Date Cleared
    2021-03-09

    (88 days)

    Product Code
    Regulation Number
    880.6850
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Neodent Instruments Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. Neodent Instruments Kits are intended to allow sterilization of the enclosed medical devices. Neodent Instruments Kits require the use of FDA-cleared wrap to maintain the sterility of enclosed devices. The kits are to be enclose in a sterilization wrap that is FDA-cleared for the indicated cycles, and moist heat (steam) sterilized using one of the following cycles:

    Fractionated vacuum (pre-vacuum) - Exposure at 132 °C for 4 minutes, 20 minutes dry time.

    Gravity displacement - Exposure at 132 °C for 15 minutes, 20-minutes dry-time.

    Neodent Instruments Kits are intended for sterilization of non-porous loads.

    The combined weight of GM EasyGuide Surgical Kit Case Narrow/Regular Diam Implants and the associated instruments is 310,18 g. The weight of the empty Kit Case is 263,63 g

    Neodent GM EasyGuide Kit Cases should not be stacked during sterilization

    Indications for Use for GM EasyGuide Surgical Kit Case Regular/Large Diam Implants:

    Neodent Instruments Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. Neodent Instruments Kits are intended to allow sterilization of the enclosed medical devices. Neodent Instruments Kits require the use of FDA-cleared wrap to maintain the sterility of enclosed devices.

    The kits are to be enclosed in a sterilization wrap that is FDA-cleared for the indicated cycles, and moist heat (steam) sterilized using one of the following cycles:

    Fractionated vacuum (pre-vacuum) - Exposure at 132 °C for 4 minutes, 20 minutes dry-time.

    Gravity displacement - Exposure at 132 °C for 15 minutes, 45-minutes dry-time.

    Neodent Instruments Kits are intended for sterilization of non-porous loads.

    The combined weight of the GM EasyGuide Surgical Kit Case Regular/Large Diam Implants and the associated instruments is 346,45 g. The weight of the empty Kit Case is 264,12 g.

    Neodent GM EasyGuide Kit Cases should not be stacked during sterilization.

    Device Description

    The subject device kit cases are reusable rigid containers, comprising a case bottom (or base), a removable inner tray, and tray lid (lid). The subject device kits are to be used to organize and protect instruments and accessories that are to be sterilized by the healthcare provider. The subject kit cases are manufactured in autoclavable polymer. The design of the subject devices includes grommets manufactured from medical grade silicone that retain the instruments within the tray. The subject device kit cases are provided nonsterile to the end-user.

    AI/ML Overview

    Here's a summary of the acceptance criteria and study information for the Neodent EasyGuide Kit Cases, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Test NamePerformance Test / Acceptance CriteriaReported Device Performance
    Manual Cleaning Validation
    Visual InspectionNo Visible SoilNo visible soil
    Hemoglobin Test (against Blood Soil - BLSO)< 2.2 µg/cm²< 0.012 µg/cm²
    Protein Test (against Blood Soil - BLSO)< 6.4 µg/cm²< 0.028 µg/cm²
    Sterilization Validation
    Sterility Assurance Level (SAL)All Biological Indicators must be incubated for at least 7 days at 55-60°C. All positive controls for SAL testing must show characteristic growth of the indicator organism.SAL 10⁻⁶ approved
    Drying Time (Pre-vacuum 132°C, 4 min exposure)(Implied to achieve sufficient dryness to maintain sterility)Approved (20 minutes dry time)
    Drying Time (Gravity 132°C, 15 min exposure)(Implied to achieve sufficient dryness to maintain sterility)Approved (20 minutes dry time for model 110.313, 45 minutes for model 110.314)
    Life Cycle (Simulated Usage) TestingThe tested samples must withstand 100 cycles of use (cleaning, sterilization, and functional tests) without compromising their functionalities.All evaluated cassettes were approved after 100 cycles of use, respecting the established acceptance criteria. Functional and visual analysis showed perfect assembly, no difficulty in instrument insertion, and no failure of locking mechanisms. No damage or breakdown observed that could compromise functionality.
    Biocompatibility (Cytotoxicity)Less than 30% cell proliferation inhibitionIn the presence of the test extract proliferation of L929 cell culture was not affected compared to untreated reagent control cultures, indicating substances were not released in cytotoxic concentrations. The results of controls confirmed sensitivity and accuracy.

    2. Sample Sizes and Data Provenance

    • Test Set Sample Size: The document does not explicitly state the numerical sample size for each test. For the cleaning validation, it refers to "tests" and "All the acceptance criteria were met," implying multiple runs. For sterilization validation, it mentions "All Biological Indicators," suggesting multiple BIs were used per cycle. For life cycle testing, it states "All evaluated cassettes," suggesting multiple devices were tested. For cytotoxicity, "L929 cell culture" refers to a cell line, not human subjects.
    • Data Provenance: The document generally indicates that the testing was performed by JJGC Indústria e Comércio de Materiais Dentários S.A. (Neodent). The country of origin for the data is implied to be Brazil, where the sponsor is located. The studies are non-clinical performance data, likely prospective in nature as they describe validation processes for the device.

    3. Number of Experts and Qualifications for Ground Truth

    This information is not provided in the document. The studies described are technical performance validations (cleaning, sterilization, life cycle, biocompatibility), which rely on objective measurements and established scientific standards rather than expert consensus on diagnostic interpretations.

    4. Adjudication Method for the Test Set

    This information is not applicable as the studies are focused on objective device performance rather than interpretations that would require adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The device is an instrument kit case for sterilization, not an AI-powered diagnostic or assistive tool for human readers.

    6. Standalone Performance (Algorithm Only)

    This information is not applicable. The device is a physical medical device (sterilization kit case), not a software algorithm.

    7. Type of Ground Truth Used

    The ground truth used for these studies is based on:

    • Objective Measurements/Standards: For cleaning validation (hemoglobin, protein levels), sterilization (Sterility Assurance Level - SAL, dryness), life cycle (functional integrity after cycles), and biocompatibility (cell proliferation).
    • Established Industry Standards: The tests were performed according to recognized standards like AAMI TIR30:2011, ANSI/AAMI/ISO 17665-1, ANSI/AAMI/ISO 17665-2, "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling Guidance for Industry and Food and Drug Administration Staff," and ANSI/AAMI/ISO 10993-5 (Cytotoxicity).

    8. Sample Size for the Training Set

    This information is not applicable. The device is a physical medical device, not an AI model that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K201491

    Validate with FDA (Live)

    Date Cleared
    2020-12-22

    (201 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Indications for Use for Zirconia Implants, Cover Screw and Healing Abutment:
    The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. It may be used with singlestage or two-stage surgical procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with physiological occlusal loading. Multiple teeth applications can be rigidly splinted.

    Indications for Use for Zirconia Bases:
    The Zirconia Base is an abutment placed over Neodent Zirconia Implants in order to provide support for patient-specific prosthetic restorations, such as copings or crowns. It may be used for cement- or screw retained single unit restorations. All digitally designed copings and/or crowns to be used with the Neodent Zirconia Base Abutment System are intended to be sent to Straumann for manufacture at a validated milling center.

    Device Description
    • Intended for single use;
    • Zirconia implants provided sterile via Ethylene Oxide; Zirconia Base abutments are provided sterile but require end user sterilization via moist heat following cementation of the restoration and before use
    • All devices of Zirconia Implant System are manufactured of Y-TZP (Yttrium-stabilized zirconium dioxide), except for the Zirconia Implant Cover Screw and the Basal Screw, which is manufactured of titanium alloy according to ASTM F136 standard;
    • ZiLock prosthetic interface with internal indexer;
    • Apically tapered implant with trapezoidal thread profile;
    • Implant provided with sand-blasted, acid etched surface finish to facilitate osseointegration.
    • The final finished Zirconia Base abutments are two-piece abutments composed of a zirconia base bottom-half bonded to a CAD-CAM zirconia top-half.
    AI/ML Overview

    The provided text describes a medical device submission (K201491) for the Neodent Implant System - Zirconia Implant System to the FDA. This document focuses on demonstrating substantial equivalence to predicate devices, rather than presenting a standalone study with specific acceptance criteria and performance results for a new AI/software device.

    Therefore, the requested information about acceptance criteria, detailed device performance, sample sizes for test/training sets, ground truth establishment, expert adjudication, and MRMC studies cannot be extracted from this document as it pertains to a different type of medical device submission (dental implants and their components) and regulatory pathway (510(k) for substantial equivalence).

    The document details the comparison of the subject device (Neodent Implant System - Zirconia Implant System) to predicate and reference devices in terms of indications for use, technological characteristics, and performance data from various tests. However, these are evaluations against established standards and comparisons to existing devices, not metrics of an AI's performance.

    Here's what can be extracted, though it doesn't directly answer the detailed questions about AI performance criteria:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't provide a table of acceptance criteria for software performance in the typical sense (e.g., sensitivity, specificity, AUC). Instead, it presents a comparison of technological characteristics and states that the "tested subject devices exhibit a level of performance equivalent to that reviewed for the predicate devices" in mechanical testing.

    Here's a summary of the performance data presented, which implies meeting an "equivalent" performance to predicates rather than specific numerical acceptance criteria for a new software.

    Performance AreaReported Device Performance
    BiocompatibilityRepresentative samples of each device subjected to tests per ISO 10993-1, ISO 10993-5, ISO 10993-12, and ISO 10993-18. (Implied acceptance is successful completion according to these standards).
    Mechanical TestingDynamic Fatigue Test (per ISO 14801): "exhibit a level of performance equivalent to that reviewed for the predicate devices."
    Wear Assessment (Titanium/Ceramic contacting pieces): "concluded of comparable behavior of the subject device to the reference devices in terms of wear on the implant-to-abutment connection."
    Torsion Test: Performed to evaluate static torsional loading. (Implied acceptance is meeting design specifications and being comparable to predicates).
    Insertion Test: Performed to evaluate insertion torque in sawbones materials (bone type I, II, III, IV). (Implied acceptance is meeting design specifications and being comparable to predicates).
    SterilizationAchieved a "Sterility Assurance Level of 1x10⁻⁶" via Ethylene Oxide per ISO 11135-1. Residuals according to ISO 10993-7. Not represented as "pyrogen free."
    MRI TestingDone according to documents presented in K182620 (a previously cleared device for MRI compatibility). (Implied acceptance is being MRI compatible as per the reference device).
    Shelf Life5 years. Packaging assessment per ISO 11607-1, with maintenance of sterile barrier and sealing integrity confirmed by Dye Penetration, Sealing Strength, and Bubble Test after Accelerated Aging.

    No information is available regarding points 2-9 as they are specific to AI/software device studies to establish performance against a ground truth. This document is a 510(k) summary for a physical dental implant system and components, establishing substantial equivalence to existing devices.

    Ask a Question

    Ask a specific question about this device

    K Number
    K202515

    Validate with FDA (Live)

    Date Cleared
    2020-11-24

    (84 days)

    Product Code
    Regulation Number
    880.6850
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Indications for Use for Drill Stop Pre-Mounted Kit and Drill Stop Kit: Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. Instrument Kits are intended to allow sterilization of the enclosed medical devices. Instrument Kits require the use of FDA-cleared wrap to maintain the sterility of the enclosed devices. The Kits are to be enclosed in a sterilization wrap that is FDA-cleared for the indicated cycles, and moist heat using one of the following cycles: Fractionated vacuum (pre-vacuum) - Exposure at 132 °C for 4 minute dry time Gravity displacement -Exposure at 132 °C for 15 minutes, 30-minute dry time. Instrument Kits are intended for sterilization of nonporous loads. The combined weight of the Kit and the associated instruments is 194.90 grams. The weight of the 189.12 grams. Instrument Kits are recommended not to be stacked during sterilization.

    Device Description

    The subject device kit cases are reusable rigid containers, comprising a case bottom (or base), a removable inner tray, and tray lid (lid). The subject device kits are to be used to organize and protect instruments and accessories that are to be sterilized by the healthcare provider. The subject kit cases are manufactured in polymer. The design of the subject devices include grommets manufactured from polyphenylsulfone that retain the instruments within the tray. The subject device kit cases are provided nonsterile to the end-user.

    AI/ML Overview

    The provided text describes the acceptance criteria and the study conducted for the NUVO Implant System - NUVO Instrument Kit Cases, a device intended for sterilizing other medical devices.

    Here's the breakdown of the information requested:

    1. A table of acceptance criteria and the reported device performance

    Acceptance CriteriaReported Device Performance
    Manual cleaning validation:
    Visual Inspection: No Visible SoilPassed
    Hemoglobin Test: < 2.2 µg/cm²Passed
    Protein Test: < 6.4 µg/cm²Passed
    Sterilization validation (ANSI/AAMI/ISO 17665-1, ANSI/AAMI/ISO 17665-2):
    All Biological Indicators must be incubated for at least 7 days at 55-60°C.Passed
    Positive controls for SAL testing must show characteristic growth of the indicator organism.Passed
    Life cycle (simulate usage) testing (Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling Guidance):
    The tested samples must withstand 100 cycles of use (cleaning, sterilization and functional testes) without compromising their functionalities.Passed
    Cytotoxicity testing (ANSI/AAMI/ISO 10993-5):
    Less than 30% cell proliferation inhibitionPassed

    2. Sample sized used for the test set and the data provenance

    The document does not explicitly state the sample sizes for each specific test or the data provenance (e.g., country of origin, retrospective/prospective). It only mentions "The tested samples" for life cycle testing and "All Biological Indicators" for sterilization validation, without specifying the number.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This type of information (number and qualifications of experts for ground truth establishment) is not applicable to the non-clinical performance data presented in the document. The tests performed are objective laboratory tests with defined acceptance criteria, not subjective interpretations requiring expert consensus.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Adjudication methods are not applicable to the types of non-clinical, objective tests described in the document.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance

    An MRMC comparative effectiveness study was not done. This device is an instrument kit case for sterilization, not an AI-powered diagnostic tool. Therefore, the concept of human readers improving with AI assistance is not relevant.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This is not applicable as the device is not an algorithm or AI system. The tests evaluate the physical and functional performance of the instrument kit case itself.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    The "ground truth" for the tests performed is based on established scientific and regulatory standards and objective measurements. For example:

    • Manual cleaning validation: Objective measurements of residual hemoglobin and protein, and visual inspection against a "no visible soil" criterion.
    • Sterilization validation: Growth or non-growth of biological indicators, indicating successful sterilization (achieving a sterility assurance level (SAL) of 10⁻⁶).
    • Life cycle testing: Ability to withstand a specified number of cycles (100) without functional degradation.
    • Cytotoxicity testing: Measured cell proliferation inhibition against a set threshold.

    8. The sample size for the training set

    This is not applicable as the device is not an AI/ML model which requires a training set. The tests are for the physical device.

    9. How the ground truth for the training set was established

    This is not applicable as the device is not an AI/ML model and therefore does not have a training set or associated ground truth.

    Ask a Question

    Ask a specific question about this device

    K Number
    K201225

    Validate with FDA (Live)

    Date Cleared
    2020-09-04

    (121 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading.

    Device Description
    • Intended for single use;
    • Provided sterile via gamma irradiation;
    • Manufactured of commercially pure titanium (Ti) grade 4 per ASTM F67;
    • Morse Taper prosthetic interface with internal hexagonal indexer, exclusively from the Grand Morse (GM) line;
    • Body center and apex with conical format;
    • Trapezoidal-profile double threads;
    • Rounded apex end;
    • Available in Neoporos and Acqua surfaces;
    • Ø7.0 mm and 8, 10, 11.5, and 13 mm lengths;
    • The subject devices are compatible with all the abutments of the Neodent Implant System that present the GM implant to abutment interface, previously cleared per K163194, K173902, K180536, K191191, and K1922229.
    AI/ML Overview

    This document is a 510(k) Premarket Notification from the U.S. Food and Drug Administration (FDA) for the Neodent Implant System - GM Helix Implants 7.0. It aims to demonstrate substantial equivalence to previously cleared predicate devices.

    Based on the provided text, the device is a dental implant. The concept of acceptance criteria and performance related to the AI/ML device is not applicable here. The document describes different types of testing for device safety and effectiveness.

    Here's a breakdown of the requested information based on the document:

    1. A table of acceptance criteria and the reported device performance

    Please note that this document is for a medical device (dental implant), not an AI/ML device. Therefore, the "acceptance criteria" and "device performance" are related to mechanical, biological, and sterilization aspects, not AI performance metrics like accuracy, sensitivity, or specificity.

    Test CategoryAcceptance Criteria (Implied/Standard)Reported Device Performance
    BiocompatibilityConformance to ISO 10993 standards (e.g., ISO 10993-1, ISO 10993-5, ISO 10993-12, ISO 10993-18).Biological Safety Assessment guided by ISO 10993-1 performed. Cytotoxicity testing performed per ISO 10993-5. Chemical characterization performed per ISO 10993-18. Made of commercially pure titanium grade 4 (ASTM F67), same as predicate devices. The subject devices do not present a new worst case for biocompatibility; predicate device testing is relied upon.
    Mechanical TestingEquivalence to predicate devices in dynamic fatigue strength, as per ISO 14801 and FDA Guidance.Dynamic fatigue test per ISO 14801 was performed. The tested subject devices exhibit a level of performance equivalent to that reviewed for the predicate devices. Insertion torque testing was also performed.
    SterilizationSterility Assurance Level (SAL) of 10^-6, validated by ISO 11137-1 and ISO 11137-2. Endotoxin limit of 20 EU/device.Sterilized by gamma irradiation at a nominal dose of 25 kGy. Validation by VDma22 method, according to ISO 11137-1 and ISO 11137-2, achieved a SAL of 10^-6. Routine endotoxin testing performed monthly on representative samples with a test limit of 20 EU/device.
    Shelf LifeMaintenance of sterile barrier after storage for the validated shelf life.Validated for a shelf life of 5 years by accelerated aging according to ASTM F1980. Packaging validated according to ISO 11607-1 and ISTA 2A. Sealing strength (ASTM F88), Dye Penetration (ASTM F1929), and Bubble Test (ASTM F2096) were conducted to verify sterile barrier maintenance.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not specify exact sample sizes for each test mentioned (biocompatibility, mechanical testing, sterilization validation, shelf life validation). It refers to "representative samples" for endotoxin testing and indicates that tests were "performed" (e.g., dynamic fatigue test per ISO 14801). The "data provenance" is not explicitly stated in terms of country of origin or retrospective/prospective studies for these specific tests. However, the sponsor is JJGC Indústria e Comércio de Materiais Dentários S.A. (dba Neodent) located in Curitiba, Parana, Brazil. The regulatory contact is in Andover, Massachusetts, USA.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This question is not applicable as the document describes a medical device, not an AI/ML diagnostic system requiring expert interpretation for ground truth establishment. The document relies on established international standards (ISO, ASTM, USP) and FDA guidance for evaluating the physical, chemical, and biological properties of the dental implant. Expert roles would be in the design, testing, and regulatory review processes, but not in establishing "ground truth" in the diagnostic sense.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This concept is relevant for AI/ML diagnostic studies where multiple readers interpret cases and their discrepancies are resolved. The document focuses on laboratory and material testing of a physical device.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a 510(k) submission for a physical medical device (dental implant), not an AI-assisted diagnostic tool. Therefore, no MRMC study or AI-related comparative effectiveness analysis with human readers would have been performed or reported here.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This relates to AI/ML device performance. The described device is a physical dental implant.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Not applicable. For this type of physical device, "ground truth" is established by adherence to recognized standards and validated test methods (e.g., ISO 10993 for biocompatibility, ISO 14801 for mechanical fatigue, ISO 11137 for sterilization) rather than expert consensus on diagnostic images or pathology. The "truth" is whether the device meets the physical and biological specifications under test conditions.

    8. The sample size for the training set

    Not applicable. The concept of a "training set" refers to data used to train an AI/ML model. This document describes the validation of a physical medical device.

    9. How the ground truth for the training set was established

    Not applicable (as above).