Zirconia Implants:
The Neodent Implant System is intended to be surgically placed in the bone of upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage surgical procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with physiological occlusion loading. Multiple teeth applications can be rigidly splinted. The implants with length of 5 mm (short implants) may be used only with two-stage surgical procedures. The recommended healing time before loading is between 10 to 12 weeks.
Zi Transmucosal Cover Screw and Healing:
The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage surgical procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with physiological occlusal loading. Multiple teeth applications can be rigidly splinted.
Zi Transmucosal Provisional Coping:
The Neodent Implant System is intended for surgical procedures in maxilla or mandible, providing support for prosthetic devices such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage procedures, for single- or multi-unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading.
Zi Transmucosal Abutment Replacement Screw:
The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading.
Zi Transmucosal Universal Base:
The Universal Ceramic Base Zi Transmucosal 5.0 is an abutment placed over Neodent Zi Transmucosal 5.0 Ceramic Implant System in order to provide support for custom-made prosthetic restorations, such as copings or crowns. It may be used for cement or screw-retained single unit restorations. All digitally designed copings and/or crowns to be used with the Neodent Zirconia Base Abutment System are intended to be sent to Straumann for manufacture at a validated milling center.
Zirconia Base for Bridge:
The Zirconia Base for Bridge is an abutment placed over Neodent Zirconia Implants in order to provide support for custom-made prosthetic restorations. It may be used for cement or screw-retained multi-unit restorations. All digitally designed copings and/or crowns to be used with the Neodent Zirconia Base Abutment System are intended to be sent to Straumann for manufacture at a validated milling center.
Zirconia Base C:
The Zirconia Base C is an abutment placed over Neodent Zirconia Implants in order to provide support for customized prosthetic restorations, such as copings or crowns. It may be used for single-unit restorations that are screw- or cement-retained in esthetic areas over implants installed in the maxilla or mandible. All copings and/or crowns digitally designed for use with the Titanium Base C are to be designed using Sirona inLab software or Sirona CEREC Software and manufactured using a Sirona CEREC or inLab MC X or MC XL milling unit.
This premarket notification includes new ceramic devices into Neodent Implant System, which are compatible with Zirconia Implant System. The Zirconia Implants and Abutments proposed on this submission are similar to devices already cleared in previous submissions of Neodent Implant System – Zirconia Implant System, according to predicate devices described above. This submission intends to expand the portfolio with new solutions and diameter, in order to provide more treatment options to the customers.
The Zirconia Implants are manufactured in Zirconia Y-TZP and are available in Bone Level (BL) or Tissue Level (TL or Transmucosal) configurations. The Zirconia Implants (BL) are available in a diameter of 5.0 mm and lengths in a range of 8 to 13 mm. The Zi Transmucosal Implants (TL) are available in a diameter of 5.0 mm and lengths in a range of 5 to 11.5 mm.
The Zi Transmucosal Healing and Cover Screw are temporary abutments manufactured in PEEK and used during the healing phase. They are compatible with the Zi Transmucosal Implants Ø5.0. The Zi Transmucosal Healing Abutment is available in the heights of 2 and 3.5mm.
The Zi Transmucosal Provisional Coping is a temporary abutment made of polycarbonate (PC) and has a double function: used for molding procedures or production of provisional restoration.
The Zi Transmucosal Abutment Replacement Screw is a prosthetic component manufactured in titanium alloy and used to fix the fix the Zi Transmucosal Base to the Zi Transmucosal Implant.
The Zi Transmucosal Universal Base is a two-piece abutment of base and top-half prosthetic structure to provide support for customized single-unit restorations over Zi Transmucosal Implant (TL). The base is manufactured in Zirconia Y-ZTP and used with a patient-specific top-half prosthetic structure. The two-piece abutment has a cementable portion of 4mm and is available with gingival heights of 0.3, 1.0 and 1.5 mm. The top-half prosthetic structure to be used with Zi Transmucosal Universal Base must be designed and milled in a Straumann Validated Milling center, using the following restoration materials and dimensions:
Material: IPS e.max CAD HT, Associated Material 510(k): K132209, Minimum wall thickness: 0.9 mm, Maximum angulation: 30°
Material: IPS e.max CAD LT, Associated Material 510(k): K132209, Minimum wall thickness: 0.9 mm
Material: N!ce, Associated Material 510(k): K171773, Minimum wall thickness: 1.0 mm
Material: IVOCLAR Multilink cement, Associated Material 510(k): K130436, Minimum wall thickness: N/A
Material: Zirconia N!ce® LT, Associated Material 510(k): K222836, Minimum wall thickness: 0.4 mm
Material: Zirconia N!ce® HT, Associated Material 510(k): K222836, Minimum wall thickness: 0.4 mm
Material: Zirconia N!ce® XT, Associated Material 510(k): K222836, Minimum wall thickness: 0.4 mm
Material: PMMA N!ce, Associated Material 510(k): K071548, Minimum wall thickness: 0.7 mm
Material: Panavia—Kuraray Cement, Associated Material 510(k): K150704, Minimum wall thickness: N/A, Maximum angulation: N/A
The Zi Base for Bridge is a two-piece abutment of base and top-half prosthetic structure to provide support for customized multi-unit restorations over Zirconia Implants (BL). The base is manufactured in Zirconia Y-ZTP and used with a patient-specific top-half prosthetic structure. The two-piece abutment has a cementable portion of 4mm and is available with gingival heights of 1.5, 2.5 and 3.5 mm. The top-half prosthetic structure to be used with Zi Base for Bridge must be designed and milled in a Straumann Validated Milling center, using the following restoration materials and dimensions:
Material: Zirconia N!ce® LT, Associated Material 510(k): K222836, Minimum wall thickness: 0.4 mm, Maximum angulation: 30°
Material: Zirconia N!ce® HT, Associated Material 510(k): K222836, Minimum wall thickness: 0.4 mm
Material: Zirconia N!ce® XT, Associated Material 510(k): K222836, Minimum wall thickness: 0.4 mm
Material: PMMA N!ce, Associated Material 510(k): K071548, Minimum wall thickness: 0.7 mm
Material: Panavia—Kuraray Cement, Associated Material 510(k): K150704, Minimum wall thickness: N/A, Maximum angulation: N/A
The Zi Base C is a two-piece abutment of base and top-half prosthetic structure to provide support for customized single-unit restorations over Zirconia Implants (BL). The base is manufactured in Zirconia Y-ZTP and used with a patient-specific top-half prosthetic structure. The two-piece abutment has a cementable portion of 4mm and is available with gingival heights of 1.5, 2.5, 3.5 and 4.5 mm. The top-half prosthetic structure to be used with Zi Base C must be designed and milled in a Sirona InLab Validated Workflow, using the following restoration materials and dimensions:
Material: IPS e.max CAD, Associated Material 510(k): K132209, Minimum wall thickness: 0.9 mm, Maximum angulation: 20°
Material: IVOCLAR Multilink cement, Associated Material 510(k): K130436, Minimum wall thickness: N/A, Maximum angulation: N/A
All these abutments have an internal connection with the implants (ZiLock) and the prosthetic platform is identical for all subject devices described in this submission. They are intended for single use and provided sterile via Ethylene Oxide method, along with undergoing moist heat sterilization after end-user customization.
The provided FDA 510(k) clearance letter and its associated summary for the Neodent Implant System - Zirconia Implant System contain extensive information about the device, its intended use, and comparisons to predicate devices. However, it does not include specific acceptance criteria with numerical thresholds directly stated within the tables, nor does it detail a study that directly proves the device meets such criteria in terms of performance metrics like sensitivity, specificity, or image quality assessments.
Instead, the submission focuses on demonstrating substantial equivalence to predicate devices through various tests, implying that if the new device performs similarly to or better than previously cleared devices, it meets the necessary standards. The performance testing section describes the types of tests conducted (e.g., dynamic fatigue, torsion, insertion, pull-out, and software validation), but it does not present clear quantitative acceptance criteria or the specific performance results in a comparative table format.
Therefore, many of the requested fields cannot be directly extracted from the provided text as they pertain more to the performance evaluation of AI/software in interpreting medical images, which is not the primary focus of this dental implant submission.
Here's an attempt to answer the questions based on the available information, noting where information is not explicitly provided in the document:
Acceptance Criteria and Device Performance Study for Neodent Implant System - Zirconia Implant System
The FDA 510(k) summary for the Neodent Implant System - Zirconia Implant System focuses on demonstrating substantial equivalence to predicate devices through a combination of bench testing, software validation, MRI compatibility, biocompatibility, and sterilization validation. It does not present specific quantitative acceptance criteria or performance metrics directly from a comparative study in the way one might expect for an AI/software-based medical device (e.g., sensitivity, specificity thresholds). Instead, the "acceptance criteria" are implicitly met by demonstrating that the proposed devices perform at a level substantially equivalent to legally marketed predicate devices under standardized testing conditions.
1. A table of acceptance criteria and the reported device performance
As mentioned, explicit numerical acceptance criteria and reported device performance in a comparative table (e.g., for diagnostic accuracy) are not provided in this 510(k) summary. The summary indicates that tests were conducted according to relevant ISO standards and FDA guidance, and the results demonstrated that the subject devices exhibit a level of performance substantial equivalent to the predicate and reference devices.
Below is a conceptual table based on the types of tests mentioned, noting that specific numerical acceptance criteria and performance data are not detailed in the provided text.
| Acceptance Criteria Category (Implicit) | Standard/Guidance | Reported Device Performance (Summary) |
|---|---|---|
| Dynamic Fatigue Strength | ISO 14801, FDA Guidance (Class II Special Controls) | Demonstrated a level of performance substantial equivalent to predicate and reference devices in identical conditions. |
| Torsion Strength | Not specified | Adequate torsion strength in accordance with recommended IFU installation torque. |
| Insertion Torque | Not specified | Evaluated insertion torque in sawbones material (Bone type I, II, III, IV). (Specific values not given). |
| Implant Surface Area & Pull-Out Strength | Not specified | Greater surface area compared to reference devices; higher resistance values in Pull Out Test. |
| Software Validation (Sirona Digital Workflow) | Not specified | Accuracy requirement was met; critical design parameters (min wall thickness, max angulation) respected and monitored. |
| MRI Compatibility | K182620, FDA Guidance | MR conditional labeling from K182620 is applicable; safe for scanning under previously established parameters. |
| Biocompatibility | ISO 10993-1, ISO 10993-18, ISO 10993-5, FDA Guidance | Subject devices are equivalent in material and manufacturing processes to predicates; no new issues raised; no additional testing required. |
| Sterilization Validation (Ethylene Oxide) | ISO 11135:2014 | Validated to a Sterility Assurance Level (SAL) of 1x10⁻⁶; residuals below max allowable limits per ISO 10993-7. |
| Sterilization Validation (Moist Heat) | ISO 17665-1 | Validated using parameters described in IFU. |
| Endotoxin Test | ANSI/AAMI ST72:2011, ISO 11737-3, US Pharmacopeia chapter 85 | Results <0.05 EU/device. |
| Shelf Life | Not specified | Determined to be 5 years; packaging identical to predicates. |
2. Sample size used for the test set and the data provenance
The document details various bench tests (dynamic fatigue, torsion, insertion, pull-out) and validation studies (software, MRI, biocompatibility, sterilization, endotoxin, shelf-life). However, specific numerical sample sizes ("n=") for these individual tests are not provided in the summary. For example, for dynamic fatigue testing, it states "the results demonstrated that...". It doesn't specify how many implants or setups were tested.
- Data Provenance (Country of Origin): The submitter is JJGC Indústria e Comércio de Materiais Dentários S.A. (dba Neodent) based in Curitiba, Paraná, Brazil. The tests are generally conducted to international ISO standards and FDA guidance, but the location where the tests were performed is not explicitly stated for each test beyond the company's Brazilian location.
- Retrospective or Prospective: Not applicable in the context of bench testing and validation studies as described. These are laboratory-based tests rather than clinical studies on patient data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This information is not applicable and not provided in the context of this 510(k) submission. The device is a dental implant system (hardware), not an AI/software device that interprets medical images requiring expert-established ground truth for a test set. The validation processes involve standardized engineering and biological tests as opposed to expert review of clinical cases.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This information is not applicable and not provided. As explained above, the submission pertains to a hardware device, not an AI/software system requiring human adjudication for ground truth establishment.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This information is not applicable and not provided. This type of study is relevant for AI-assisted diagnostic tools, not for dental implant hardware.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This information is not applicable and not provided. This concept is specific to AI/software performance, not a physical dental implant system. The "software validation" mentioned is for the compatibility of milling units (CAD/CAM workflow), ensuring design parameters are respected, not for an AI algorithm interpreting dental images.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
For the bench tests, the "ground truth" is defined by compliance with established engineering standards (e.g., ISO 14801 for dynamic fatigue) and specifications (e.g., material properties, dimensions). For biological safety, it's compliance with ISO 10993. For the CAD/CAM software, it's the accurate adherence to design parameters. This isn't "ground truth" in the diagnostic sense, but rather adherence to predefined engineering and biocompatibility specifications.
- Clinical Literature Review: A clinical literature review was conducted to support the safety of Zi Transmucosal Implant less than 7mm of length, referencing animal studies and histomorphometry evaluation. This aligns with outcomes data or pathological/histological evidence from preclinical studies, suggesting "similar osseointegration outcomes" and "ability to promote bone formation around them in a manner equivalent to titanium implants."
8. The sample size for the training set
This information is not applicable and not provided. This concept is specific to machine learning/AI models, which are not the subject of this 510(k) submission.
9. How the ground truth for the training set was established
This information is not applicable and not provided. As above, this is relevant for AI/ML training data, not for a physical dental implant system.
Ask a specific question about this device
(87 days)
Zirconia Guided Surgery Kit Cases are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. Zirconia Guided Surgery Kit Cases are intended to allow sterilization of the enclosed medical devices. Zirconia Guided Surgery Kit Cases requre the use of FDA-cleared wrap to maintain the sterility of the enclosed devices. The kits are to be enclosed in a sterlization wrap that is FDA-cleared for the indicated cycles, and moist heat (steam) sterilized using one of the following cycles: Dynamic Air Removal (pre-vacuum) – Exposure at 132°C for 4 minutes, 20-minute dry time. Gravity displacement – Exposure at 132°C for 15 minute dry time. Ziroonia Guided Surgery Kit Cases are intended for sterilization of non-porous loads. The Zirconia Guided Surgery Kit Case and the associated instruments is 342.2 g. The weight of the empty Kit is 236.8 grams. Zirconia Guided Surgery Kit Case are recommended not to be stacked during sterilization.
Nuvo CF Guided Surgery Kit Cases are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. Nuvo CF Guided Surgery Kit Cases are intended to allow sterilization of the enclosed medical devices. Nuvo CF Guided Surgery Kit Cases require the use of FDA-cleared wrap to maintain the sterility of the enclosed devices. The kits are to be enclosed in a sterilization wrap that is FDA-cleared for the indicated cycles, and moist heat (steam) sterilized using one of the following cycles: Fractionated vacuum): exposure at 132°C for 4 minutes, drying for 20 minutes. Gravity displacement: exposure at 132°C for 15 minutes, drying for 30 minutes. Nuvo CF Guided Surgery Kit Cases are intended for sterilization of non-porous loads. The combined weight of the Nuvo CF Guided Surgery Kit Cases and the associated instruments is approximately 693 g. The weight of the empty case is 487 grams. Nuvo CF Guided Surgery Kit Cases are not recommended to be stacked during sterilization.
The Zirconia and Nuvo CF Guided Surgery Kit Cases subject to this submission are reusable rigid containers, composed by a case bottom, a removable inner tray and a lid, which are made of autoclavable polymer. They are indicated to hold various dental instruments (not subject of this submission) in order to organize, sterilize and protect them. Sterilization via moist heat (steam) is performed by the healthcare provider.
This document is a 510(k) Summary for the Zirconia and Nuvo CF Guided Surgery Kit Cases. It describes the non-clinical performance data used to evaluate the device.
Here's an analysis of the acceptance criteria and the study that proves the device meets them:
1. Table of Acceptance Criteria and Reported Device Performance:
| Standard or Test Method | Purpose of the Testing | Acceptance Criteria | Reported Device Performance |
|---|---|---|---|
| ANSI/AAMI/ISO 17665-1ANSI/AAMI/ISO 17665-2 | Steam sterilization validation, including sterilant penetration and drying time | All Biological Indicators must be incubated for at least 7 days at 55-60°C. All positive controls for SAL testing must show characteristic growth of the indicator organism. | Passed |
| AAMI TIR 30:2011 | Manual cleaning validation• Test Soil: Blood Soil (BLSO)• Cleaning Method: Manual• Residuals Tested: Hemoglobin and Protein | • Visual Inspection: No visible soil• Hemoglobin Test: < 2.2 µg/cm²• Protein Test: < 6.4 µg/cm² | Passed |
| Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling Guidance for Industry and Food and Drug Administration Staff (Life cycle testing) | Life cycle (simulate usage) testing | The tested samples must withstand 100 cycles of use (cleaning, sterilization and functional tests) without compromising their functionalities. | Passed |
| ANSI/AAMI/ISO 10993-5 | Biocompatibility of subject devices via cytotoxicity testing | Less than 30% cell proliferation inhibition | Passed |
2. Sample Size Used for the Test Set and Data Provenance:
The document does not explicitly state the sample sizes for each test set (e.g., number of kit cases tested for sterilization, cleaning, or life cycle). It mentions "representative samples" for biocompatibility.
The data provenance is not explicitly stated in terms of country of origin but is from the manufacturer (JJGC Indústria e Comércio de Materiais Dentários S.A. dba Neodent, Brazil, and Straumann USA, LLC). The studies appear to be prospective, specifically designed to validate the new devices.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications:
This document describes non-clinical performance data involving laboratory testing against established standards. Therefore, it does not involve "experts" in the sense of clinicians establishing ground truth for diagnostic accuracy. The ground truth is defined by the technical specifications and performance requirements established by the listed ISO and AAMI standards.
4. Adjudication Method for the Test Set:
Not applicable, as this is laboratory testing against objective criteria defined by standards, not subjective evaluation by human experts.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
Not applicable. This is not a study assessing the effectiveness of an AI or diagnostic device with human reader interaction. It relates to the sterilization and reprocessing of medical device cases.
6. Standalone Performance Study:
Yes, the studies described are standalone performance studies of the device itself (the Zirconia and Nuvo CF Guided Surgery Kit Cases). The tests evaluate characteristics like sterilization effectiveness, cleaning efficacy, durability over multiple cycles, and biocompatibility, without human-in-the-loop performance being a factor.
7. Type of Ground Truth Used:
The ground truth used is based on established international and national medical device standards (ISO 17665-1, ISO 17665-2, AAMI TIR 30:2011, ISO 10993-5) and regulatory guidance documents (FDA Guidance document for ISO 10993-1). These standards define objective criteria for successful performance (e.g., specific microbial reduction, acceptable residual levels, physical integrity over cycles, cytotoxicity limits).
8. Sample Size for the Training Set:
Not applicable. These are non-clinical validation studies for physical devices, not an AI or machine learning model that requires a training set.
9. How the Ground Truth for the Training Set Was Established:
Not applicable, as there is no training set involved. The "ground truth" for the device's performance is established by its compliance with the specified performance standards in the validation testing.
Ask a specific question about this device
(149 days)
The Custom Abutment Ti with Screw is a customized prosthetic abutment, manufactured in titanium alloy, placed onto dental implants to provide support for customized prosthetic restorations. All abutments are only intended to be digitally designed and manufactured using specifics CAD/CAM software according to digital dentistry workflow. Custom Abutments Ti with Screw are indicated for screw-retained single restorations or cemented-retained single or multiple restorations. All digitally designed abutments for use with the Custom Abutment Ti with Screw are intended to be sent to Straumann for manufacturing at a validated milling center.
The Custom Abutment with Angled Screw Channel is a customized prosthetic abutment, manufactured in titanium alloy, placed onto dental implants to provide support for customized prosthetic restorations. All abutments are only intended to be digitally designed and manufactured using specifics CAD/CAM software according to digital dentistry workflow. Custom Abutments with Angled Screw Channel are indicated for screw-retained single restorations or cemented-retained single or multiple restorations. All digitally designed abutments for use with the Custom Abutment Ti with Angled Screw Channel are intended to be sent to Straumann for manufacturing at a validated milling center.
The Custom Abutments subject to this submission are similar to the devices already cleared in previous Neodent and Nuvo Implant System submissions, per the predicate and reference devices described above. This submission intends to expand the portfolio of digital abutments for the Grand Morse (GM), Narrow Grand Morse (NGM) and Helix Short (HS) lines, to provide more treatment options for customers. These abutments are composed of a unique body with two regions: the upper region is the customized portion, while the end region presents the prosthetic interface that fits with the implant, which does not allow customization.
They are intended for single use and provided non-sterilization recommended before installation in the mouth.
All proposed abutments are delivered to the final user already customized in a validated milling center, with a straight channel solution. For the GM line, the abutments could also be milled with a new angled channel solution for the screw access. All subject devices must be used with their corresponding prosthetic interface.
The provided text describes the Neodent Implant System - Custom Abutments, a medical device. This product is not an AI/ML device, and therefore does not have acceptance criteria or a study that proves the device meets those criteria in the way requested in the prompt.
The document discusses the substantial equivalence of the Neodent Implant System - Custom Abutments to legally marketed predicate devices, which is the basis for its 510(k) clearance by the FDA. The performance testing section describes bench tests (dynamic fatigue and torsion tests), sterilization validation, MR compatibility testing, and biocompatibility testing. These tests evaluate the physical, chemical, and biological properties of the device, comparing them to established standards and predicate devices to demonstrate equivalence, not to meet a set of acceptance criteria for an AI/ML algorithm's performance.
Because the device is not an AI/ML device, the specific information requested about acceptance criteria for AI performance, sample sizes for test/training sets, data provenance, expert ground truth establishment, adjudication methods, MRMC studies, standalone performance, and how ground truth for training was established is not applicable and is not present in the provided text.
Ask a specific question about this device
(176 days)
The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage surgical procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with physiological occlusal loading. Multiple teeth applications can be rigidly splinted. Healing,
The Zirconia Base is an abutment placed over Neodent Zirconia Implants in order to provide support for custom-made prosthetic restorations, such as copings or crowns. It may be used for cement- or screw-retained single unit restorations. All digitally designed copings and/or crowns to be used with the Neodent Zirconia Base Abutment System are intended to be sent to Straumann for manufacture at a validated milling center.
The CR Abutment is an abutment placed over Neodent Zirconia Implants in order to provide support for prosthetic restorations, such as copings or crowns. It may be used for single unit restorations that are cement-retained in esthetical areas over implants installed in maxilla or mandible.
The PEEK CR Abutment is indicated to be used on Neodent Implants to provide support for prosthetic structures for up to 6 months. They can be used in single- or two-stage procedures and they are intended to be placed out of occlusion.
The Zirconia Abutments subject of this submission are similar to devices already approved in previous submissions of Zirconia Implant System, according to predicate devices described above. This submission intends to expand the portfolio of Zirconia Abutments with new dimensions of gingival height for Zirconia Healing Abutments, Zirconia Bases, CR Abutments for Zirconia and PEEK CR Abutments for Zirconia, to provide more treatment options to the customers. All these abutments have an internal connection with the implants (ZiLock) and the prosthetic platform is identical for all subject devices described in this submission. They are intended for single use and provided sterile via Ethylene Oxide method. The new transmucosal heights are 3.5mm and 4.5mm.
The restorative materials Polycon ae, N!ce and IPS e.max CAD LT were not presented in previous submissions of Zirconia Implant System and are being included in the scope of Zirconia Bases indication.
The provided document is a 510(k) summary for the Neodent Implant System - Zirconia Implant System. This type of FDA submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving a device meets specific acceptance criteria through a clinical study with defined performance metrics, sample sizes, and expert adjudication as might be seen for novel AI/software as a medical device (SaMD) products.
Therefore, the document does not contain the information requested regarding acceptance criteria related to a study proving performance, sample sizes for test sets, data provenance, number or qualifications of experts for ground truth, adjudication methods, MRMC studies, standalone algorithm performance, or ground truth establishment methods for training sets.
The performance testing section (pages 14-15) describes:
- Bench Testing:
- Dynamic fatigue testing: Conducted according to FDA guidance and ISO 14801. The results demonstrated the subject devices do not introduce a new worst case compared to predicate and reference devices.
- Torsion tests: Performed to evaluate screw strength, and results proved adequate torsion strength according to installation torque.
- Biocompatibility Testing:
- Assessment performed according to ISO 10993-1 and FDA guidance. No new issues were raised and no additional testing was required because the subject devices are identical in material and manufacturing processes to predicate and reference devices.
- Sterilization Validation and Packaging:
- Ethylene Oxide (EO) sterilization validated to SAL of 10-6 per ISO 11135:2014. EO residuals verified per ISO 10993-7. The packaging and sterilization are identical to predicate devices.
- Steam sterilization validated per ISO 17665-1, using parameters identical to predicate devices.
- Pyrogen limit specifications met via LAL Endotoxin Analysis (testing limit 20 EU/device) for blood contacting and implanted devices.
In essence, the study presented is a non-inferiority or equivalence-based assessment against previously cleared devices, focusing on engineering and biocompatibility characteristics, not clinical performance metrics or AI algorithm validation metrics.
Therefore, I cannot provide the requested table and details because the information is not present in the provided text. The document describes a traditional medical device (dental implants and abutments) clearance, not a software or AI device that would have the specific performance study criteria outlined in your request.
Ask a specific question about this device
(110 days)
Indications for Use for Zygoma-S GM Implant:
Zygomatic implants are indicated for intraoral surgical procedures in the zygoma region in cases of severe maxilla bone resorption, to restore the patient's chewing function and aesthetics. Zygomatic Implants may be used in one or two-stage procedures, multiple unit restorations, and immediate loading when there is primary stability and adequate occlusal load.
Indications for Use for GM Mini Conical Abutment 52° and 45°:
Product indicated for surgical procedures in zygomatic bones, making possible the rehabilitation with screw-retained abutments over the implant, thus restoring the chewing function. It may be used in one- or two-stage procedures, multiple unit restorations, and immediate loading when there is primary stability and adequate occlusal load. Multiple rehabilitations may be splinted rigidly.
Indications for Use for Coping for Removable Prosthesis:
The product, when used with non-zygomatic implants, is intended to be surgically placed in the bone of the upper or lower jaw, to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. It may be used with single stage or two-stage procedures, for multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with approprate occlusal loading. Multiple tooth applications may be rigidly splinted. When used with zygomatic implants, the product is indicated for surgical procedure only in upper jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. It may be used in two-stage procedures (delayed loading protocol) and for multiple unit restorations. Multiple rehabilitations may be splinted rigidly.
The subject devices implants are single use devices, provided sterile by Gamma Radiation, made of commercially pure Titanium grade 4 (ASTM F67). The Zygomatic Implant is a long implant with a nominal diameter of 3.50 and 3.75mm, conical apex with rounded tip, 3 helical chambers, trapezoidal thread and GM prosthetic interface. The cervical portion in both models has Ø 4.3 mm. The length varies from 30 to 55 mm. Zygomatic Implants are intended to provide support for a fixed prosthesis in patients with severe atrophy in the maxillary region. They are extra-long to allow bone anchorage in the zygomatic bone. The subject implants consider the Zygoma Anatomy-Guided approach (ZAGA) for planning the surgical steps, being indicated for extrasinus and extra maxillary installation technique (ZAGA-4).
The subject abutments are single use devices, provided sterile by Ethylene Oxide, made of Titanium alloy (Tl6Al4V-ELI). They are intermediate prosthetic components angled at 52 and 45 degrees with gingiva heights 1.5 and 2.5 mm, to be installed on the implant, offering a structure to support the screw-retained multiple prosthesis. The abutments present GM prosthetic interface and are available in the anti-rotational form in the implant to abutment interface, and rotational form in the abutment to prosthesis interface, presenting different transmucosal heights.
The subject coping is single use device, provided sterile by Ethylene Oxide, made of titanium alloy (Ti6Al4V-ELI). It presents a NEO torque interface for torque application. The subject device has diameter 4.80 mm and total height 4.60 mm. They are indicated for removable prostheses with fitting over Mini Conical Abutment installed in the maxilla or mandible and can be used in partial removable prostheses or over dentures. They are compatible with all previously cleared GM Mini Conical abutments.
The provided document is a 510(k) premarket notification for a medical device (Neodent Implant System - GM Zygomatic Implant System). It focuses on demonstrating substantial equivalence to predicate devices through comparisons of indications for use, technological characteristics, and performance data, primarily bench testing, sterilization validation, shelf life validation, and biocompatibility.
Crucially, this document does not contain information about an AI/ML algorithm or a study validating its performance, nor does it detail acceptance criteria for software or AI/ML components. The product being cleared is a set of dental implants and abutments. Therefore, I cannot extract the acceptance criteria and study details for an AI-powered device as requested.
The document discusses performance data in the context of the physical dental implants and abutments, which includes:
- Performance Testing – Bench: Dynamic fatigue testing (ISO 14801), insertion tests, torsion tests, and microscopic analysis of the implant body.
- Surface treatment: Discussion of the NeoPoros surface.
- MR Compatibility testing: Assessment of safety in MRI environments.
- Sterilization validation: Conformance to ISO standards for Gamma Irradiation (implants) and Ethylene Oxide (abutments).
- Shelf Life validation: Based on packaging integrity and physical stability.
- Biocompatibility: Testing per ISO 10993 standards.
- Clinical data: A review of published literature on zygomatic implants, including survival rates and observed complications (not a specific clinical study conducted for this device clearance).
If you are looking for information regarding acceptance criteria and study data for an AI/ML medical device, this document does not provide it, as it pertains to a different type of medical device.
Ask a specific question about this device
(106 days)
Indications for Use for Neodent Complement Kit Cases:
Neodent Instrument Kit Cases are indicated for organization of surgical and/or prosthetic instruments during sterilization, storage and transport. The use of this product facilitates storage and organization of instruments during and after surgical procedures. Neodent Instrument Kit Cases are intended to allow sterilization of the enclosed medical devices. Neodent Instrument Kit Cases require the use of FDA cleared wrap to maintain the sterility of the enclosed devices. The kits are to be enclosed in a sterilization wrap that is FDA-cleared for the indicated cycles, and moist heat using one of the following cycles:
Dynamic Air Removal (pre-vacuum)- Exposure at 132 ℃ for 4 minutes, 30-minute dry time.
Gravity displacement - Exposure at 132 °C for 15 minutes, 20-minute dry time.
Neodent Instrument Kit Cases are intended for sterilization of non-porous loads. The combined weight of the Complement Case and the associated instruments is 75,7 g. The weight of the empty Kit Case is 53.62 grams.
Neodent Instrument Kit Cases should not be stacked during sterilization.
Indications for Use for Nuvo Complement Kit Cases:
Nuvo™ Instrument Kit Cases are indicated for organization of surgical and/or prosthetic instruments during sterilization, storage and transport. The use of this product facilitates storage and organization of instruments during and after surgical procedures. Nuvo™ Instrument Kit Cases are intended to allow sterilization of the enclosed medical devices. Nuvo™ Instrument Kit Cases require the use of FDA cleared wrap to maintain the sterility of the enclosed devices. The kits are to be enclosed in a sterilization wrap that is FDA-cleared for the indicated cycles, and moist heat using one of the following cycles:
Dynamic Air Removal (pre-vacuum)- Exposure at 132 ℃ for 4 minutes, 30-minute dry time.
Gravity displacement - Exposure at 132 °C for 15 minutes, 20-minute dry time.
Nuvo™ Instrument Kit Cases are intended for sterilization of non-porous loads. The combined weight of the Complement Case and the associated instruments is 75.7 g. The weight of the empty Kit Case is 52.72 grams.
Nuvo™ Instrument Kit Cases should not be stacked during sterilization.
The subject devices Complement Kit Cases consist of a storage case for Neodent and Nuvo instruments set. They are made of autoclavable polymer, the case has silicone holders to store and hold each instrument securely during sanitization and surgical procedures. The case can be equipped according to the procedure.
The provided text is a 510(k) summary for a medical device (Complement Kit Cases), focusing on the device's substantial equivalence to a predicate device. It details non-clinical performance data.
Here's the information extracted and organized according to your request:
1. Table of Acceptance Criteria and Reported Device Performance
| Standard or Test Method | Purpose of the Testing | Acceptance Criteria | Results |
|---|---|---|---|
| Custom (Manual cleaning validation_ Test Soil: Blood Soil (BLSO) Cleaning Method: Manual_ Residuals Tested: Hemoglobin and Protein) | Manual cleaning validation | Visual Inspection: No Visible SoilHemoglobin Test: <2.2 µg/cm²Protein Test: <6.4 µg/cm² | Passed |
| ANSI/AAMI/ISO 17665-1ANSI/AAMI/ISO 17665-2 | Sterilization validation, including sterilant penetration and drying time | All Biological Indicators must be incubated for at least 7 days at 55-60°C.All positive controls for SAL testing must show characteristic growth of the indicator organism. | Passed |
| Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling Guidance for Industry and Food and Drug Administration Staff | Life cycle (simulate usage) testing | The tested samples must withstand 100 cycles of use (cleaning, sterilization and functional tests) without compromising their functionalities | Passed |
| ANSI/AAMI/ISO 10993-5 (Cytotoxicity) | Cytotoxicity testing | Less than 30% cell proliferation inhibition | Passed |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Test Set: The document mentions "Six simulated use cycles" for manual cleaning validation and refers to "the tested worst case devices" for life cycle validation. It does not explicitly state the number of individual devices or test runs beyond these general descriptions for all tests.
- Data Provenance: The document does not specify the country of origin for the data or whether the studies were retrospective or prospective. These are non-clinical (laboratory) tests, not clinical studies involving patient data.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
This document describes non-clinical performance testing for a medical device (sterilization kit cases), not a diagnostic AI/ML device that requires expert-established ground truth from images or clinical data. Therefore, this section is not applicable. The "ground truth" here is based on validated laboratory methods and established material/sterilization standards.
4. Adjudication Method for the Test Set
Not applicable. This is not a study involving human readers or subjective interpretations requiring adjudication. Performance is measured against physical and biological criteria.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
No. This is a non-clinical performance study for a physical medical device (sterilization kit cases), not an AI/ML clinical study. Therefore, no MRMC study was performed, and no effect size for human readers improving with AI assistance is provided.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done
Not applicable. This is not an AI/ML device; it is a physical medical device. The performance evaluated is of the device itself (e.g., its ability to withstand sterilization cycles, its biocompatibility).
7. The Type of Ground Truth Used
The "ground truth" in these non-clinical performance tests is based on:
- Established Standards: ANSI/AAMI/ISO 17665-1, ANSI/AAMI/ISO 17665-2, AAMI TIR30:2011, ISO 10993-1, ISO 10993-5, and FDA Guidance documents.
- Physical and Biological Parameters: Visual inspection for soil, quantifiable levels of hemoglobin and protein (for cleaning), sterility assurance level (SAL) of 1 x 10⁻⁶ (for sterilization), structural integrity after 100 cycles, and cell proliferation inhibition percentages (for biocompatibility).
These are objective, measurable criteria defined by regulatory and industry standards.
8. The Sample Size for the Training Set
Not applicable. This is not a machine learning or AI device that requires a training set.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as no training set was used.
Ask a specific question about this device
(107 days)
Helix NGM Compact Surgical Kit Cases are indicated for organization of surgical and/or prosthetic instruments during sterilization, storage and transport. The use of this product facilitates storage and organization of instruments during and after surgical procedures.
Neodent Instrument Kit Cases are intended to allow sterilization of the enclosed medical devices. Neodent Instrument Kit Cases require the use of FDA cleared wrap to maintain the sterility of the enclosed devices.
The kits are to be enclosed in a sterilization wrap that is FDA-cleared for the indicated cycles, and moist heat (steam) sterilized using one of the following cycles:
Fractionated vacuum (pre-vacuum)- Exposure at 132 °C for 4 minutes, 20-minute dry time.
Gravity displacement - Exposure at 132 °C for 15 minutes, 20-minute dry time.
Neodent Instrument Kit Cases are intended for sterilization of non-porous loads.
The combined weight of the GM Surgical Kit Case and the associated instruments is 268.0 g. The weight of the empty Kit Case is 208.15 grams. Neodent Instrument Kit Cases should not be stacked during sterilization.
The subject device kit cases are reusable rigid containers, comprising a case bottom (or base), a removable inner tray, and tray lid (lid). The subject device kits are to be used to organize and protect instruments and accessories that are to be sterilized by the healthcare provider. The subject kit cases are manufactured in autoclavable polymer. The design of the subject devices include grommets manufactured from medical grade silicone that retain the instruments within the tray. They also presents markings that guide instrument use during procedures. The subject device kit cases are provided nonsterile to the end-user.
The provided document is a 510(k) Premarket Notification for a medical device (Neodent Implant System - Helix NGM Compact Surgical Kit Cases). This type of submission aims to demonstrate substantial equivalence to a legally marketed predicate device, rather than proving the device's efficacy through a clinical study with primary and secondary endpoints often seen in drug trials or some Class III device PMAs.
Therefore, the "acceptance criteria" here refer to the performance standards the device must meet to demonstrate substantial equivalence to its predicate for its stated indications for use, primarily regarding sterilization and reusability. The "study that proves the device meets the acceptance criteria" refers to the non-clinical performance testing conducted.
Here's the breakdown based on your request, using the information available in the provided FDA document:
1. Table of Acceptance Criteria and Reported Device Performance
| Standard or Test Method | Purpose of the Testing | Acceptance Criteria | Reported Device Performance |
|---|---|---|---|
| Custom | Manual cleaning validation (Test Soil: Blood Soil (BLSO) Cleaning Method: Manual Residuals Tested: Hemoglobin and Protein) | Visual Inspection: No Visible SoilHemoglobin Test: <2.2 µg/cm²Protein Test: <6.4 µg/cm² | Passed |
| ANSI/AAMI/ISO 17665-1ANSI/AAMI/ISO 17665-2 | Sterilization validation, including sterilant penetration and drying time | All Biological Indicators must be incubated for at least 7 days at 55-60°C. All positive controls for SAL testing must show characteristic growth of the indicator organism. | Passed |
| Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling Guidance for Industry and Food and Drug Administration Staff | Life cycle (simulate usage) testing | The tested samples must withstand 100 cycles of use (cleaning, sterilization, and functional tests) without compromising their functionalities. | Passed |
| ANSI/AAMI/ISO 10993-5 (Cytotoxicity) | Cytotoxicity testing | Less than 30% cell proliferation inhibition | Passed |
2. Sample Size Used for the Test Set and Data Provenance
The document does not specify the precise sample sizes used for each of the non-clinical tests (e.g., how many units were tested for cleaning validation, sterilization, life cycle, or biocompatibility). It only states that the testing was performed and "Passed" the acceptance criteria.
Data Provenance: The testing appears to be non-clinical performance testing, likely conducted in a lab setting by the manufacturer or a third-party lab on units of the device. The document does not specify the country of origin of the data beyond the manufacturer's location in Brazil. This is an application for a medical device, and the data is prospective testing to support the 510(k) submission.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts
This type of submission (510(k) for sterilization trays) relies on objective engineering and microbiological testing standards, rather than expert human interpretation of images or clinical outcomes that would require a "ground truth" established by experts in the typical sense of a diagnostic AI study. The "ground truth" for these tests are the established scientific and engineering principles outlined in the referenced ISO, AAMI standards, and FDA guidance documents, which define what constitutes effective cleaning, sterilization, durability, and biocompatibility.
4. Adjudication Method for the Test Set
Not applicable for this type of non-clinical, objective performance testing. Adjudication methods (like 2+1, 3+1) are typically used in clinical studies involving human interpretation or subjective assessments to establish a consensus "ground truth."
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done
No, an MRMC comparative effectiveness study was not done. This device is a sterilization and instrument organization kit, not a diagnostic imaging AI algorithm.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done
Not applicable. This is not an AI algorithm. The performance evaluation is of the physical device's ability to be cleaned, sterilized, and resist wear over time.
7. The Type of Ground Truth Used
The "ground truth" for this device's performance testing is based on:
- Objective Measurement and Standards: This includes quantifiable measures for residual proteins/hemoglobin after cleaning, demonstration of sterility assurance level (SAL) of 10⁻⁶ using biological indicators, and quantifiable measures of cell proliferation inhibition for biocompatibility.
- Functional Longevity: The ability of the device to withstand 100 cleaning and sterilization cycles without compromising functionality.
- Compliance with Recognized Standards: The tests are performed according to established industry and regulatory standards (e.g., ANSI/AAMI/ISO 17665, AAMI/ANSI/ISO 10993-5, and FDA guidance on reprocessing). These standards define the acceptable performance metrics.
8. The Sample Size for the Training Set
Not applicable. This is not an AI/machine learning device. There is no training set.
9. How the Ground Truth for the Training Set Was Established
Not applicable. As noted above, this is not an AI/machine learning device.