510(k) Data Aggregation

    K Number
    K222266

    Validate with FDA (Live)

    Date Cleared
    2022-09-26

    (60 days)

    Product Code
    Regulation Number
    878.4040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Surgical Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. The medical surgical mask is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This a single use, disposable device(s), provided non-sterile.

    Device Description

    The proposed surgical masks are blue color, three-layer, flat-pleaded masks with nose piece, ear loops or ties, which are composed of inner layer, middle layer and outer layer, with or without anti-fog foam strips and eye shield. The outer layers contains blue colorant and the inner layer contains white colorant, which are made of polypropylene.

    Below are the configuration of each model.

    GFM 96 - three-layer, flat-pleated with ear loops

    GFM 91 - three-layer, flat-pleated with ear loops and anti-fog foam strip attached to nose clip

    GFM 90 - three-layer, flat-pleated with ear loops, anti-fog foam strip attached to nose clip and eye shield

    GFM 88 - three-layer, flat-pleated with ties

    GFM 81 - three-layer, flat-pleated with ties and anti-fog foam strip attached to nose clip

    GFM 80 - three-layer, flat-pleated with ties, anti-fog foam strip attached to nose clip and eye shield

    The inner layer and outer layer of the mask are made of spunbond polypropylene, the middle layer is made of melt-blown polypropylene. The ear loops or ties of the subject mask are held in place over the users' mouth and nose by two ear loops or ties welded to the face mask. The ear loop is made with polyester and spandex, while the ties are made of spunbond polypropylene. The nose piece in the layers of face mask is to allow the user to fit the mask around their nose, which is made of PP and metal.

    The masks may also contain an eye shield made from a polyethylene terephthalate film (PET) with foam strip. The eye shield is adhered to the top edge of the mask to cover the upper part of the eye to prevent potential exposure to blood and body fluids.

    The surgical masks are sold non-sterile and are intended to be single use, disposable devices.

    The masks are designed and manufactured in accordance with ASTM F2100-19 Standard Specification for Performance of Materials Used in Medical Face Masks.

    AI/ML Overview

    This document is a 510(k) Premarket Notification for a Surgical Mask, specifically models GFM 96, GFM 91, GFM 90, GFM 88, GFM 81, and GFM 80, manufactured by Jingzhou Haixin Green Cross Medical Products Co., Ltd.

    The purpose of this submission is to demonstrate that the new device is substantially equivalent to a legally marketed predicate device (K210147). The "acceptance criteria" and "study that proves the device meets the acceptance criteria" in this context refer to the performance standards and non-clinical testing conducted to show the device meets specific requirements for surgical masks and is equivalent to the predicate.

    Here's a breakdown of the requested information based on the provided document:


    1. Table of Acceptance Criteria and Reported Device Performance

    Test MethodologyPurposeAcceptance Criteria for Level 3 BarrierReported Device Performance
    Bacterial Filtration Efficiency (ASTM F2101)Measure bacterial filtration efficiency≥98%Passed (Lot 1: ≥98%, Lot 2: ≥98%, Lot 3: ≥98%)
    Differential Pressure (mmH2O/cm²) (EN 14683:2019 Annex C)Determine breathability of the mask<6.0 mmH2O/cm²Passed (Lot 1: <6.0, Lot 2: <6.0, Lot 3: <6.0)
    Sub-micron Particulate Filtration Efficiency (ASTM F2299-17)Measure initial particle filtration efficiency≥98%Passed (Lot 1: ≥98%, Lot 2: ≥98%, Lot 3: ≥98%)
    Resistance to Penetration by Synthetic Blood (ASTM F1862-17)Evaluate resistance to penetration by impact of synthetic blood29 out of 32 pass at 160 mmHgPassed (Lot 1: 32 out of 32 pass, Lot 2: 32 out of 32 pass, Lot 3: 32 out of 32 pass)
    Flammability (16 CFR Part 1610-2008)Response of materials to heat and flameClass IPassed (Lot 1: Class 1, Lot 2: Class 1, Lot 3: Class 1)
    Biocompatibility Evaluation:
    Cytotoxicity (ISO 10993-5)Non-cytotoxicPassed (Non-cytotoxic)
    Sensitization (ISO 10993-10)Non-sensitizingPassed (Non-sensitizing)
    Irritation (ISO 10993-10)Negligibly irritatingPassed (Negligibly irritating)

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: For the performance tests (Bacterial Filtration, Differential Pressure, Particulate Filtration, Synthetic Blood Penetration, Flammability), 3 non-consecutive lots were tested, using a sample size of 32 devices per lot.
    • Data Provenance: The document does not explicitly state the country of origin of the test data (e.g., where the independent labs were located). It mentions the applicant's address as "JINGZHOU HUBEI, CHINA." The study is a non-clinical study, which typically means lab testing, not patient-derived data, so the terms "retrospective" or "prospective" are not applicable in their usual clinical sense. The tests were performed to demonstrate compliance with recognized standards.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • This is a non-clinical device performance and safety study, not a study involving medical imaging or clinical diagnosis where experts establish ground truth.
    • The "ground truth" for these tests is defined by the established ASTM, EN, ISO, and CFR standards themselves (e.g., a specific filtration efficiency percentage, a specific pressure drop value, etc.). The tests are quantitative measurements against these predefined objective criteria.
    • The document does not mention the number or qualifications of experts involved in "establishing ground truth" in the way one would for diagnostic AI. The expertise would lie in the accredited labs and personnel performing the standardized tests.

    4. Adjudication Method for the Test Set

    • Not applicable. This is a non-clinical, quantitative performance testing of a physical device against specified standards. Results are direct measurements, not subject to human adjudication or consensus in the way clinical interpretations are.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This submission is for a surgical mask, a physical medical device, not an AI or diagnostic imaging device that assists human readers. No MRMC study was conducted.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • Not applicable. This is a physical device, not an algorithm. The "standalone" performance here refers to the intrinsic physical properties and performance of the mask as measured by the specified tests.

    7. The Type of Ground Truth Used

    • The "ground truth" for these non-clinical tests are the quantitative performance specifications defined by internationally recognized standards (ASTM F2101, EN 14683, ASTM F2299, ASTM F1862, 16 CFR Part 1610 for performance, and ISO 10993-5, ISO 10993-10 for biocompatibility). There is no "expert consensus," "pathology," or "outcomes data" ground truth in this context.

    8. The Sample Size for the Training Set

    • Not applicable. This is a physical medical device. The "training set" concept is relevant for machine learning/AI models. For device manufacturing, the "training" involves process control and quality assurance to meet specifications, not data-driven model training.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable for the reasons stated above (physical device, not an AI model). Quality control and manufacturing processes ensure the device is produced to meet the established performance specifications.
    Ask a Question

    Ask a specific question about this device

    K Number
    K123787

    Validate with FDA (Live)

    Date Cleared
    2013-07-17

    (219 days)

    Product Code
    Regulation Number
    878.4040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The GreenCross Surgical Masks are indicated as a protective nose and mouth covering for health care workers and patients involved in medical and surgical procedures. The masks are indicated in any procedure or situations where there is a risk of microorganism, body fluid, and particulate aerosol transfer.

    Device Description

    GreenCross surgical masks, Type: Earloop or Tie-on, are pleated 3-ply mask. Inner and outer layers are made of 100% spun-bond polypropylene. Middle player is made of 100% meltblown polypropylene filter media. Ear-loops are made of soft latex free elastic loops. All of the materials used in the construction of the GreenCross surgical masks are being used in currently marketed devices, except the nose piece used plastic wire instead of aluminum wire.

    AI/ML Overview

    K123787: GreenCross Surgical Mask Acceptance Criteria and Performance Study Analysis

    This submission describes the non-clinical performance testing of the GreenCross Surgical Mask to establish substantial equivalence to a predicate device (K101000).

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for each test are implicitly set by achieving comparable performance to the predicate device, or by meeting the "relevant requirements in the test standards." The reported device performance is directly compared to the predicate device's performance.

    TestAcceptance Criteria (Implicit)GreenCross Surgical Mask PerformancePredicate Device (K101000) Performance
    Fluid Resistance (ASTM F1862)Comparable to predicate; meets relevant standard requirements29 out of 32 pass at 120 mm Hg31 out of 32 pass
    Particle Filtration Efficiency (ASTM F2299)Comparable to predicate; meets relevant standard requirements99.1%99.8%
    Bacteria Filtration Efficiency (ASTM F2101)Comparable to predicate; meets relevant standard requirements99.8%99.9%
    Differential Pressure (MIL-M-36945C)Comparable to predicate; meets relevant standard requirements2.6-2.9 (mm Water/cm²)3.7-4.0 (mm Water/cm²)
    Flammability (16CFR 1610)Class 1 None FlammableClass 1 None FlammableClass 1 None Flammable
    Biocompatibility (ISO 10993-5, -10)BiocompatibleBiocompatibleBiocompatible

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the specific sample sizes used for each individual non-clinical test (e.g., number of masks tested for BFE, PFE). For the Fluid Resistance test (ASTM F1862), the result "29 out of 32 pass" indicates a sample size of 32 masks for that particular test.

    The data provenance is from non-clinical tests conducted by the manufacturer, Jingzhou Haixin Green Cross Medical Products Co., Ltd. The country of origin for the testing is not explicitly stated, but the manufacturer is based in Wuhan City, Hubei, P. R, China. These are prospective test results generated specifically for this submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    This section is not applicable. The ground truth for this device is based on objective, standardized laboratory performance tests designed to measure physical properties and filtration capabilities. It does not involve human expert interpretation or subjective evaluation.

    4. Adjudication Method for the Test Set

    This section is not applicable. As the ground truth is established through standardized laboratory tests, there is no need for expert adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    An MRMC study was not performed. This submission is for a surgical mask, and its performance is evaluated through non-clinical bench testing, not by comparing human reader performance.

    6. Standalone (Algorithm Only) Performance

    This section is not applicable. As this is a physical medical device (surgical mask), there is no algorithm or AI involved.

    7. Type of Ground Truth Used

    The ground truth used for evaluating the GreenCross Surgical Mask is based on objective, standardized laboratory performance testing as outlined by international and national standards (e.g., ASTM F2101, ASTM F2299, ASTM F1862, MIL-M-36945C, 16CFR 1610, ISO 10993). These standards define specific methodologies and criteria for measuring the performance attributes of surgical masks.

    8. Sample Size for the Training Set

    This section is not applicable. This is a physical device, not an AI/ML algorithm, so there is no "training set."

    9. How the Ground Truth for the Training Set Was Established

    This section is not applicable. There is no training set for a physical surgical mask.