Search Filters

Search Results

Found 10 results

510(k) Data Aggregation

    K Number
    K242736

    Validate with FDA (Live)

    Date Cleared
    2025-06-05

    (267 days)

    Product Code
    Regulation Number
    868.5440
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Portable Oxygen Concentrator, model: Spirit-3, is intended to provide supplemental oxygen in a home, institutional, or travel environment. And the device is used with adult only, not used with pediatrics, infant, or neonate patients, etc.

    Device Description

    Portable Oxygen Concentrator, model: Spirit-3, is a portable oxygen generator that is intended to release oxygen for respiratory therapy by means of physical means (a molecular sieve). It supplies a pulsed high concentration of oxygen and is used with a nasal cannula to deliver oxygen from the concentrator to the patient. The Portable Oxygen Concentrator is small, portable and may be used in home, institutional, or travel environment. Portable Oxygen Concentrator, model: Spirit-3, is capable of continuous use in a home, institutional, or travel environment. Power options include 100 – 240 V (50/ 60Hz) AC power supply, DC power supply and rechargeable lithium-ion battery. The portable oxygen concentrator consists of two parts: an oxygen concentrator and accessories.

    AI/ML Overview

    The FDA 510(k) clearance letter and supporting documentation for the Portable Oxygen Concentrator (Spirit-3) do not describe a study involving an AI/ML component or human readers. Therefore, there is no information available regarding acceptance criteria related to AI/ML performance, sample sizes for test or training sets, expert qualifications, adjudication methods, multi-reader multi-case studies, or standalone algorithm performance.

    The provided document focuses on demonstrating the substantial equivalence of the Spirit-3 Portable Oxygen Concentrator to predicate devices through technical specifications, non-clinical performance testing (bench testing), and compliance with various international standards. The "performance" being discussed in the document refers to the physical and functional performance of the oxygen concentrator itself (e.g., oxygen concentration, flow control, electrical safety, biocompatibility), not the performance of an AI/ML algorithm in an image or data interpretation task.

    Therefore, many of the requested categories are "Not Applicable" (N/A) in the context of this specific regulatory submission.

    However, I can extract the acceptance criteria and reported device performance for the Spirit-3 Portable Oxygen Concentrator based on the non-clinical tests described.

    Here's the information derived from the provided text, adapted to the requested format where applicable:

    1. Table of Acceptance Criteria and Reported Device Performance

    Feature / TestAcceptance Criteria (from Standards/Predicate Device Comparison)Reported Device Performance (Spirit-3)
    Oxygen Concentration90%-3%/+6% at all settings (Same as predicate)90%-3%/+6% at all settings
    Startup time2 minutes (Same as predicate)2 minutes
    Acoustic NoiseSimilar to 55.3 dBA at 0.84 LPM (predicate)55 dBA at 0.75 LPM (Similar)
    Electrical SafetyCompliance with IEC 60601-1Complied with ANSI AAMI ES60601-1, IEC 60601-1-11, IEC 60601-1-8, ISO 80601-2-69, ISO 80601-2-67, IEC TS 60601-4-2:2024, CISPR 25: 2021
    Electromagnetic CompatibilityCompliance with IEC 60601-1-2Complied with IEC 60601-1-2
    Biocompatibility - CytotoxicityCompliance with ISO 10993-5:2009Tested and complied (explicitly stated in "Discussion of difference" for patient contact materials)
    Biocompatibility - SensitizationCompliance with ISO 10993-10:2021Tested and complied (explicitly stated in "Discussion of difference" for patient contact materials)
    Biocompatibility - IrritationCompliance with ISO 10993-23:2021Tested and complied (explicitly stated in "Discussion of difference" for patient contact materials)
    Biocompatibility - Particulate Matter (Breathing Gas Pathway)Compliance with ISO 18562-2:2017Tested and complied
    Biocompatibility - Volatile Organic Compounds (Breathing Gas Pathway)Compliance with ISO 18562-3:2017Tested and complied
    Software Verification & ValidationPerforms as intended according to FDA GuidancePerformed and demonstrated as intended
    Battery SafetyCompliance with IEC 62133-2Tested and verified via IEC 62133-2
    Operating Condition (Temperature)Within acceptable range (Predicate: 5 to 40˚C)10˚C to 35˚C (Narrower, but compliant with ISO 80601-2-69)
    Operating Condition (Humidity)Within acceptable range (Predicate: 10% to 90%)15% to 75%, noncondensing (Narrower, but compliant with ISO 80601-2-69)
    Operating Condition (Altitude)Within acceptable range (Predicate: 0 to 3048 meters)0 to 3000 meters (Narrower, but compliant with ISO 80601-2-69)
    Pulse mode bolus sizeSimilar to 42 mL per breath at setting 4 with 20 BPM (predicate)37.5mL per breath at setting 4 with 20BPM (Smaller, but covered by subject device and does not raise new questions of safety and effectiveness)
    Rated breath rateSimilar to 10 - 40 Breath per minute (predicate)15 - 40 Breath per minute (Narrower, but meets ISO 80601-2-67:2020 requirement for disclosure and covered by predicate range)
    Maximum oxygen discharge pressureClose to 151.68kPa (22psi) (reference device)150kPa (21.8psi) (Higher than predicate, close to reference device, no new questions of safety/effectiveness)

    Study Proving Device Meets Acceptance Criteria

    The study conducted to prove the Spirit-3 Portable Oxygen Concentrator meets its acceptance criteria was a series of non-clinical bench tests and evaluations demonstrating compliance with recognized consensus standards and comparison to a legally marketed predicate device.

    2. Sample size used for the test set and the data provenance:

    • Sample Size for Test Set: Not applicable in the context of an AI/ML algorithm being evaluated. For the performance testing of the physical device, it implicitly refers to the specific unit(s) of the Spirit-3 device undergoing testing. The document does not specify the number of devices tested but implies standard product testing.
    • Data Provenance: The tests were conducted internally or by accredited labs as part of the manufacturing and submission process for Jiangsu Yuyue Medical Equipment & Supply Co., LTD. (China). This is prospective testing of the manufactured device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not Applicable. This device is a physical oxygen concentrator, not an AI/ML diagnostic or interpretative device that requires expert ground truth for output validation.

    4. Adjudication method for the test set:

    • Not Applicable. As no expert interpretation was required. Device performance was objectively measured against predefined standard requirements.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done:

    • No. This type of study is relevant for AI/ML systems that assist human readers in tasks like image interpretation. The Spirit-3 is a portable oxygen concentrator and does not involve human reading or AI assistance in that capacity.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Not Applicable. The device does not involve a standalone algorithm for diagnostic or interpretative purposes. "Standalone" performance in this context refers to the device's inherent mechanical, electrical, and oxygen delivery performance.

    7. The type of ground truth used:

    • The "ground truth" for the performance of the portable oxygen concentrator is defined by the international consensus standards mentioned (e.g., ISO 80601-2-69, ISO 80601-2-67, IEC 60601-1, ISO 10993 series) and the specifications of the predicate devices. These standards provide benchmarks and methodologies for evaluating physical and electrical safety, essential performance, and biocompatibility.

    8. The sample size for the training set:

    • Not Applicable. This device does not use an AI/ML training set.

    9. How the ground truth for the training set was established:

    • Not Applicable. No AI/ML training set was employed for this device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K250878

    Validate with FDA (Live)

    Date Cleared
    2025-05-22

    (59 days)

    Product Code
    Regulation Number
    880.2910
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Infrared Ear Thermometer is indicated for intermittent measurement of human body temperature from the ear canal. The device can be used by people of all ages except preterm babies or babies who are small for gestational age. The thermometer is intended for use in professional settings and the home environment. It is not for emergency clinical conditions.

    Device Description

    YUWELL® Infrared Ear Thermometer is designed to measure human body temperature. This hand-held, no-contact device is battery-powered and utilizes infrared energy emitted in the subject's tympanic membrane to determine human body temperature. The thermometer provides temperature readings in just 2 seconds. It is suitable for individuals of all ages except preterm babies or babies who are small for gestational age. The Infrared Ear Thermometer YHT107 can transmit the temperature readings via Bluetooth, whereas the YHT100 does not have data transmission capability.

    AI/ML Overview

    Here's a detailed breakdown of the acceptance criteria and study information for the YUWELL® Infrared Ear Thermometer, extracted from the provided FDA 510(k) clearance letter:


    YUWELL® Infrared Ear Thermometer (YHT100; YHT107) Acceptance Criteria and Study Details

    The YUWELL® Infrared Ear Thermometer (YHT100; YHT107) is indicated for intermittent measurement of human body temperature from the ear canal in professional and home environments, for people of all ages except preterm babies or babies who are small for gestational age.

    1. Table of Acceptance Criteria and Reported Device Performance

    The core performance acceptance criteria are related to temperature accuracy, as specified by the standards ISO 80601-2-56 and ASTM E1965.

    CharacteristicAcceptance Criteria (Standard Specification)Reported Device Performance (Subject Device)
    AccuracyFor 35.0℃–42.0℃ (95.0℉–107.6℉): ±0.2℃ (±0.4℉)Achieved: ±0.2℃ (±0.4℉) for 35.0℃–42.0℃ (95.0℉–107.6℉)
    For 34.0℃–34.9℃ (93.2℉–94.8℉) & 42.1℃–42.2℃ (107.8℉–108.0℉): ±0.3℃ (±0.5℉)Achieved: ±0.3℃ (±0.5℉) for 34.0℃–34.9℃ (93.2℉–94.8℉) and 42.1℃–42.2℃ (107.8℉–108.0℉)
    Temperature RangeConforms with ASTM E1965 (Predicate range: 20℃-42.2℃)34.0-42.2℃ (93.2-108.0℉)
    Display Resolution0.1℃/0.1℉ (Matches predicate)0.1℃/0.1℉
    Response TimeNot explicitly stated as a strict acceptance criterion with quantitative limits, but differences were evaluated. (Predicate: 2-3s)< 12 sec (Evaluated and found acceptable)
    Operating Environment (Temp/Humidity)Conforms with ASTM E1965+10℃–+40℃ (50℉–104℉) and 15-90% RH
    Storage Environment (Temp/Humidity)Conforms with ASTM E1965Temperature: -20 ºC–+55 ºC (-4 ºF–131 ºF); Humidity: 15-90% RH no condensation

    Note on Differences: The document explicitly states that the clinical validation results demonstrated that the clinical data, represented by clinical bias, limits of agreement, and clinical repeatability, met the acceptance criteria of the clinical validation protocol. This general statement confirms that the device's performance aligned with the specified requirements derived from the standards.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: 154 subjects.
      • Infants: 50 subjects (newborn to one year)
      • Children: 50 subjects (greater than one to five years old)
      • Adults: 54 subjects (greater than five years old)
    • Data Provenance: Not explicitly stated (e.g., country of origin). The study was a "clinical study" which implies a prospective collection of data for the purpose of device validation.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    The provided document does not specify the number of experts or their qualifications used to establish the ground truth for the clinical study. For a thermometer, the "ground truth" would typically be established by a reference temperature measurement device, but the specifics are not detailed.

    4. Adjudication Method for the Test Set

    The document does not specify an adjudication method (e.g., 2+1, 3+1). This type of method is more common for subjective assessments (like image interpretation) rather than quantitative measurements from a thermometer.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    • No, an MRMC comparative effectiveness study was not done.
      • MRMC studies are typically for evaluating the impact of AI assistance on human readers, often in image-based diagnostics. This device is a direct measurement tool (thermometer) and does not involve human interpretation of complex data that would be enhanced by AI assistance in that manner. The comparison is between the device's reading and a gold standard reference.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Yes, a standalone performance evaluation was done. The clinical study directly assesses the YUWELL® Infrared Ear Thermometer's accuracy against a recognized standard (implied by "clinical bias, limits of agreement and clinical repeatability met the acceptance criteria"). The thermometer itself is the "algorithm" and it directly produces the temperature reading without requiring human interpretation for its output.

    7. The Type of Ground Truth Used

    • Reference Measurement (Implied): For temperature measurement devices, the "ground truth" is established by a highly accurate and calibrated reference thermometer or methodology that itself adheres to metrological standards. The document mentions "clinical bias, limits of agreement and clinical repeatability met the acceptance criteria of the clinical validation protocol," which strongly implies that the device's readings were compared against a gold standard reference temperature measurement during the clinical study.

    8. The Sample Size for the Training Set

    • The document does not provide information regarding a separate training set or its sample size. Infrared ear thermometers typically use fixed algorithms/calibration curves based on physical principles (e.g., Planck's Law) and extensive bench testing, rather than being "trained" on a clinical dataset in the machine learning sense. The clinical validation acts as a test of the final, calibrated device.

    9. How the Ground Truth for the Training Set was Established

    • As a training set is not explicitly mentioned or implied in the context of machine learning, the question of how its ground truth was established is not applicable here. The device's calibration during manufacturing and design likely relies on highly controlled temperature reference points and blackbody radiation sources, rather than a clinical "training set" with ground truth established through expert consensus or pathology.
    Ask a Question

    Ask a specific question about this device

    K Number
    K213825

    Validate with FDA (Live)

    Date Cleared
    2023-04-28

    (506 days)

    Product Code
    Regulation Number
    868.5440
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This oxygen concentrator is intended for using as an oxygen supplement device in the professional healthcare facility and home healthcare environment. It provides high concentration of oxygen to persons requiring oxygen therapy. This device is to be used as an oxygen supplement and is NOT considered life-supporting or life-sustaining.

    Device Description

    The Yuyue Oxygen Concentrator models 8F-3A/8F-5A is a portable concentrator which could utilize a molecular sieve and pressure swing adsorption to produce the oxygen. The device could separate nitrogen by absorbing through the molecular sieve when the room air enters the device, and allows the enriched oxygen to be collected. The subject device is intended use for adult population.

    The subject device contains primary electrical components including main unit, compressor, fan, flowmeter and circuit breaker, the software is moderate level according to FDA software guideline.

    AI/ML Overview

    This submission is for an Oxygen Concentrator, models 8F-3A and 8F-5A. The provided document does not contain information related to a study that proves the device meets acceptance criteria in the context of AI/ML performance, human-in-the-loop studies, or expert ground truth adjudication.

    Instead, the document focuses on demonstrating substantial equivalence to a predicate device (L4 Oxygen Concentrator (EverFlo), K061261) through non-clinical performance data for regulatory clearance. This involves comparing technical specifications and performance against recognized standards, rather than evaluating AI performance with acceptance criteria typically found in machine learning contexts.

    Therefore, many of the requested categories for acceptance criteria and study details cannot be filled from the provided text.

    Here's an analysis based on the available information:

    1. A table of acceptance criteria and the reported device performance

    The document provides a comparison table for substantial equivalence, which includes various technical characteristics and a comparison to the predicate device. These characteristics serve as "acceptance criteria" for showing equivalence to the predicate, rather than explicit performance metrics against a defined threshold specific to a new AI/ML device.

    CharacteristicAcceptance Criteria (Subject Device Performance)Predicate Device Performance K061261Comparison
    Oxygen Concentration(8F-3A): ≥90%;(8F-5A): 95.5%-87%90-96% from 1 to 5 LPMEquivalent
    Flow Rate(8F-3A): 1-3 LPM;(8F-5A): 1-5 LPMfrom 1 to 5 LPMEquivalent
    Output Pressure≤70 kPa44.8 kPa (6.5PSIG)Equivalent
    IP classificationIP21IP21Same
    Complies with ISO 10993-1YesYesSame
    Complies with ISO 80601-2-69YesYesSame
    Complies with ASTM D4169-16YesYesSame
    Complies with ES 60601-1 Electrical SafetyYesYesSame
    Complies with IEC 60601-1-2 EMC SafetyYesYesSame
    Indication for UseThis oxygen concentrator is intended for using as an oxygen supplement device in the professional healthcare facility and home healthcare environment. It provides high concentration of oxygen to persons requiring oxygen therapy. This device is to be used as an oxygen supplement and is NOT considered life-supporting or life-sustaining.The Respironics L4 Oxygen Concentrator is intended to provide supplemental oxygen to persons requiring oxygen therapy. The device is not intended to be life supporting or life sustaining. The Respironics L4 Oxygen Concentrator is intended for use in the home or hospital/institutional environment.Equivalent

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    Not applicable. The provided document details non-clinical laboratory testing and compliance with various standards, not performance results from a test set of patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. Ground truth establishment by experts is not relevant to this type of device and submission, which focuses on hardware performance and safety standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. Adjudication methods are not relevant to this type of device and submission.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is an oxygen concentrator, not an AI/ML-driven diagnostic or interpretative tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a medical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Not applicable. Ground truth for AI/ML performance is not relevant here. The "ground truth" for this device's performance is its ability to meet the specified technical parameters (e.g., oxygen concentration, flow rate) and safety standards through physical testing and engineering measurements.

    8. The sample size for the training set

    Not applicable. This is not an AI/ML device; therefore, no training set is mentioned.

    9. How the ground truth for the training set was established

    Not applicable. This is not an AI/ML device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K221372

    Validate with FDA (Live)

    Date Cleared
    2022-10-17

    (158 days)

    Product Code
    Regulation Number
    870.1130
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This product is intended to measure the blood pressure and pulse rate of adult more than 12 years old and with wrist circumference ranging from 13.5 cm at household or medical center (not suitable for neonate, pregnancy or pre-eclampsia.)

    Device Description

    The YUWELL® Blood Pressure Monitor (Model: YE8800AR, YE8800CR) is a rechargeable lithium battery-powered, automatic, noninvasive, wrist-worn blood pressure measuring system intended for over-the-counter (OTC) use. YE8800AR, YE8800CR is designed for wrist circumference ranging from 13.5 cm to 19.5 cm. The systolic and diastolic blood pressures are measured using the oscillometric method, where the cuff is inflated with an air pump and deflates via an exhaust valve. During inflation, the cuff pressure is monitored, and pulse waveform data is extracted. The extracted pulse waveform data is then further analyzed by software which determines pulse rate, as well as systolic and diastolic blood pressure.

    AI/ML Overview

    The acceptance criteria for the YUWELL® Electronic Blood Pressure Monitor (Model: YE8800AR, YE8800CR) are primarily based on performance standards for blood pressure monitors. The study described is a clinical accuracy test.

    Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criterion (Standard)Reported Device Performance (YUWELL® Electronic Blood Pressure Monitor)
    Accuracy of Pressure Indicator (per ANSI/AAMI/ISO 81060-2:2018 Criterion 1)Mean Difference ± Standard Deviation (Test Device vs. Reference BP): Systolic: -1.7 ± 5.66 mmHg Diastolic: -1.0 ± 5.66 mmHg (The full criteria for ANSI/AAMI/ISO 81060-2 require that the mean difference be ≤ ±5 mmHg and the standard deviation be ≤ 8 mmHg.)
    Accuracy of Pressure Indicator (per ANSI/AAMI/ISO 81060-2:2018 Criterion 2)Mean Differences (Two Observers vs. YUWELL® Electronic Blood Pressure Monitor): Systolic: -1.7 ± 4.17 mmHg Diastolic: -1.0 ± 4.99 mmHg (The full criteria for ANSI/AAMI/ISO 81060-2 for Criterion 2 states that the standard deviation of the differences between the observation and the device should be not greater than 6.91 mmHg for at least two out of three device criteria sets. Values are generally expected for pairs of subjects, not for the entire population at once as given above)
    Accuracy of Pulse RateWithin ±5% of reading
    Electrical Safety (IEC 60601-1, IEC 60601-1-11, IEC 80601-2-30)Complies
    Electromagnetic Compatibility (IEC 60601-1-2)Complies
    Usability (IEC 60601-1-6, IEC 62366)Complies
    Software Verification and Validation (FDA Guidance, IEC 62304)Complies
    Biocompatibility (ISO 10993, FDA Guidance)Complies
    FCC Compliance (Part 15 federal communications commission rules)Complies

    The Study Proving Acceptance Criteria:

    A clinical accuracy test was conducted on the Yuwell YE8800AR model.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size: 85 participants.
    • Data Provenance: The document does not explicitly state the country of origin. It describes participants as the "general population." The study is described as a "clinical accuracy test," suggesting a prospective collection of data for this validation.

    3. Number of Experts and Qualifications for Ground Truth:

    The document mentions "reference BP" and "two observers" in relation to the ANSI/AAMI/ISO 81060-2:2018 protocol. This standard typically requires measurements by trained observers or reference devices to establish ground truth.

    • Number of Experts/Observers: Two observers are explicitly mentioned in the context of Criterion 2 of the ANSI/AAMI/ISO 81060-2:2018 protocol.
    • Qualifications of Experts: Not explicitly stated in the provided text, but adherence to the ANSI/AAMI/ISO 81060-2:2018 protocol implies that the observers would be appropriately trained personnel, potentially medical professionals or technicians, who follow a standardized procedure for manual blood pressure measurement to establish the reference values.

    4. Adjudication Method for the Test Set:

    The document does not explicitly describe an adjudication method for the test set, such as 2+1 or 3+1. The ANSI/AAMI/ISO 81060-2 standard relies on comparisons between the automated device and simultaneous auscultatory measurements performed by trained observers. Discrepancies between observers might be handled by averaging or requiring re-measurements per the standard's guidelines, but a specific "adjudication" (like a third reader resolving disagreement) is not detailed here.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    No, an MRMC comparative effectiveness study involving human readers with and without AI assistance was not done. This device is a standalone electronic blood pressure monitor, not an AI-assisted diagnostic tool that would typically involve human readers.

    6. Standalone (Algorithm Only) Performance:

    Yes, the described clinical accuracy test is a standalone performance test of the device (algorithm only), without a human-in-the-loop component for the measurement process itself. The "human-in-the-loop" aspect would primarily be the human observers establishing the reference blood pressure for comparison.

    7. Type of Ground Truth Used:

    The ground truth was established by reference blood pressure measurements, likely obtained through a standardized method such as auscultation (manual measurement with a stethoscope and sphygmomanometer) performed by trained observers, as dictated by the ANSI/AAMI/ISO 81060-2:2018 validation protocol.

    8. Sample Size for the Training Set:

    The document does not provide information on the sample size used for the training set. The clinical study described is for validation/testing, not for training the device's measurement algorithm. Blood pressure monitors typically rely on established oscillometric principles rather than machine learning models that require explicit "training sets" in the same way as AI diagnostic devices.

    9. How the Ground Truth for the Training Set was Established:

    Not applicable, as a training set in the context of machine learning is not mentioned or implied for this type of device. The device operates based on physical principles of oscillometry. The inherent algorithm development would likely have been based on extensive physiological and engineering data, but not typically a labeled "training set" with ground truth in the AI sense.

    Ask a Question

    Ask a specific question about this device

    K Number
    K212385

    Validate with FDA (Live)

    Date Cleared
    2022-08-16

    (379 days)

    Product Code
    Regulation Number
    870.2700
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The YUWELL® Finger Pulse Oximeter is a non-invasive, non-sterile, spot checking device which can measure and display SpO2 and pulse rate through finger. It is intended for adults and children (weight >30 kg) and is expected for home and hospital inspection. The device is not for continuous monitoring,use during motion or for patients with low perfusion.

    Device Description

    The YUWELL® Finger Pulse Oximeter is a non-invasive, non-sterile, spot checking device which can measure and display SpO2 and pulse rate through finger. It is intended for adults and children (weight >30 kg) and is expected for home and hospital inspection. The device is not for continuous monitoring, use during motion or for patients with low perfusion. The YUWELL® Finger Pulse Oximeter features in small volume, low power consumption, convenient operation and portable. It is only necessary for patient to put one of his fingers into a fingertip photoelectric sensor for measurement, and then the screen will display the measured value of Pulse Oxygen Saturation (SpO2) and pulse rate(PR) . The device consists of electronic circuits, plastic housing. OLED/LED display(differentiated by models) and button (YX301,YX306 equipped, YX102,YX103 not equipped) which powered by two alkaline AAA batteries. The device does not include alarms.

    AI/ML Overview

    The provided text is a 510(k) summary for the YUWELL® Finger Pulse Oximeter. Here's an analysis of the acceptance criteria and the study proving the device meets them, based on the provided document:

    Acceptance Criteria and Reported Device Performance

    The document primarily focuses on demonstrating substantial equivalence to a predicate device (Beijing Choice Electronic Technology Co., Ltd. Fingertip Pulse Oximeter, K101577). The key performance acceptance criteria can be inferred from the comparison table and the clinical study results.

    Table of Acceptance Criteria and Reported Device Performance (Inferred from comparison and study results):

    Performance CharacteristicAcceptance Criteria (As per Predicate/Standards)Reported Device Performance (YUWELL® Finger Pulse Oximeter)
    SpO2 Accuracy (SpO2 Measurement Range)70%~100%, ± 2% (Predicate)Arms less than 3% during steady states over the range of 70-100% (Clinical Study Result from ISO 80601-2-61)
    SpO2 Accuracy (Specific Ranges)70%100%, ± 2%; 069% no definition (Predicate)70%~80%, 80%~90%, 90%~100%, ± 2% (Subject Device listed values); < 70%, no definition
    SpO2 Display Range0%~100% (Predicate)0%~100%
    SpO2 Resolution1% (Predicate)1%
    PR Accuracy30bpm99bpm, ± 2bpm; 100bpm235bpm, ± 2% (Predicate)±1% or ±1 bpm (larger)
    PR Display Range0bpm~254bpm (Predicate)25bpm~250bpm
    PR Measurement Range30bpm~235bpm (Predicate)25bpm~250bpm
    PR Resolution1bpm (Predicate)1bpm
    BiocompatibilityCompliance with ISO 10993-1 (Implied)Cytotoxicity, Irritation, Sensitization tests conducted and met standards. Material evaluated as per ISO 10993-1.
    Electrical SafetyCompliance with IEC 60601-1, IEC 60601-1-11IEC 60601-1, IEC 60601-1-11 standards met.
    EMCCompliance with IEC 60601-1-2IEC 60601-1-2 standard met.
    SoftwareCompliance with FDA Guidance for Software in Medical Devices (Implied)Software verification and validation provided, moderate level of concern, work functionally as per FDA guidance.
    Cleaning ValidationReprocessing Medical Devices Guidance (Implied)Cleaning and disinfection validation conducted as recommended by FDA.
    Environmental Requirements(Predicate ranges provided, but overall acceptance is through standard compliance)Verified in accordance with IEC 60601-1, IEC 60601-1-11 and ISO 80601-2-61 standards.

    Study Information

    The document describes a clinical study to verify the device's accuracy.

    1. Sample size used for the test set and the data provenance:

      • Sample Size: 12 healthy adult volunteer subjects (ages 18-50yr, with light to dark pigmentation).
      • Data Provenance: The location of the study is not explicitly stated, but the company is based in China. The study was conducted to evaluate SpO2 accuracy during stationary (non-motion) conditions with induced hypoxia. This implies a prospective study design.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • The document does not specify the number of experts or their qualifications for establishing the ground truth. It states that "Arterial blood samples were drawn during simultaneous data collection from the test devices. The blood was immediately analyzed on reference CO-Oximetry providing functional SaO2 for the basis of the SpO2 accuracy comparison." This indicates that CO-Oximetry was the ground truth reference.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • No information on an adjudication method is provided, as the ground truth was based on a reference medical device (CO-Oximetry) and not expert reader consensus.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No MRMC study was conducted or mentioned. This device is a standalone measurement device (pulse oximeter), not an AI-assisted diagnostic tool for human readers.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • Yes, the clinical study conducted was a standalone performance evaluation of the pulse oximeter's accuracy in measuring SpO2 and pulse rate against a reference standard (CO-Oximetry), without human interpretation as part of the primary measurement.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The ground truth used was arterial blood CO-Oximetry. This is considered the reference standard for blood oxygen saturation measurement.
    7. The sample size for the training set:

      • The document does not mention a distinct "training set" or sample size for training. This device does not appear to be an AI/machine learning product that requires separate training and test sets in the conventional sense. Its principle is based on established photoelectric oximetry technology (Lambert Beer law, Spectrum Absorption Characteristics). The comparison notes mention that an "experience formula of data process is established taking use of Lambert Beer law according to Spectrum Absorption Characteristics...verified by a large amount of data." However, the size or provenance of this data is not detailed as a "training set."
    8. How the ground truth for the training set was established:

      • Since a formal "training set" for an AI model is not explicitly described, the method for establishing ground truth for such a set is also not provided. The phrase "verified by a large amount of data" suggests historical data or internal validation, but no specifics are given.
    Ask a Question

    Ask a specific question about this device

    K Number
    K203583

    Validate with FDA (Live)

    Date Cleared
    2021-05-20

    (163 days)

    Product Code
    Regulation Number
    880.2910
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The YUWELL® Infrared ear thermometer is a non-sterile, reusable clinical thermometer. The device is to display the body temperature in the ear cavity by thermal radiation for people of all ages except preterm babies and newborns (1-29 days old).

    Device Description

    The YUWELL® Infrared ear thermometer is designed for measuring the body's temperature, is a hand-held non-contact infrared thermometer, battery powered, using the infrared energy emitted in the subject's tympanic membrane that converts a user's body temperature.

    AI/ML Overview

    The provided document describes the safety and effectiveness of the YUWELL Infrared Ear Thermometer (Models YHT101, YHT200) based on comparisons to a predicate device (Braun Thermoscan® PRO 6000 Ear Thermometer) and compliance with various standards.

    Here's an analysis of the acceptance criteria and study information, based on the provided text:

    1. Table of Acceptance Criteria and the Reported Device Performance

    The document does not present a formal table of "acceptance criteria" for the device, but rather compares the subject device's specifications and performance to a predicate device and states compliance with recognized standards. The "Performance" row in the "Substantial Equivalence" table (page 4) indicates the key performance standards met.

    Here's a table based on the key performance and accuracy criteria mentioned:

    Acceptance Criteria (Standard Reference)Reported Device Performance (YUWELL Infrared Ear Thermometer)Comment/Predicate Comparison
    Accuracy (ASTM E 1965, ISO 80601-2-56)34.0°C34.9°C (93.2°F94.8°F): ±0.3°C (±0.5°F) 35.0°C42.0°C (95.0°F107.6°F): ±0.2°C (±0.4°F) 42.1°C42.2°C (107.7°F108.0°F): ±0.3°C (±0.5°F)Same as predicate device for clinical measurement range. The document states: "For the normal clinical measurement range of 34.0°C to 42.2°C, the accuracy of the subject device and the predicate device are the same." Predicate accuracy: 0.3 °C for <35 °C, 0.2 °C for 35 °C to 42 °C, 0.3 °C for >42°C.
    Measurement Range34.0°C to 42.2°C (93.2°F ~ 108.0°F)Slightly narrower than predicate. Predicate: 20°C to 42.2°C (68°F to 108.0°F). Justified as meeting normal clinical use and not bringing additional risks.
    Electrical SafetyMeets IEC 60601-1Same as predicate.
    EMC (Electromagnetic Compatibility)Meets IEC 60601-1-2Same as predicate.
    BiocompatibilityMeets ISO 10993-1, ISO 10993-5, ISO 10993-10Same as predicate.
    Home Use (if applicable)Meets IEC60601-1-11Not explicitly stated for predicate in the table, but the subject device meets this standard.
    Software Verification & ValidationMeets FDA Guidance "Guidance for the content of premarket submissions for software contained in Medical Devices"Not explicitly stated for predicate in the table, but the subject device meets this guidance.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document mentions "Clinical accuracy test requirements established in the standard ASTM E1965 (Clinical part only) and ISO 80601-2-56 (Clinical accuracy validation only)". However, it does not provide details on the specific sample size used for these clinical accuracy tests, nor does it specify the country of origin of the data or whether the study was retrospective or prospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The document states that "Clinical accuracy test requirements established in the standard ASTM E1965 (Clinical part only) and ISO 80601-2-56 (Clinical accuracy validation only)" were met. For clinical thermometers, ground truth is typically established by comparing the device's readings against a highly accurate reference thermometer (e.g., rectal thermometry in controlled environments or other gold standard methods) in actual patients.

    However, the document does not specify the number of experts used or their qualifications.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    The document does not provide any information regarding an adjudication method for establishing ground truth or evaluating the test set.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    The device under review is an infrared ear thermometer, which is a standalone measurement device, not an AI-assisted diagnostic tool that involves human "readers." Therefore, a multi-reader multi-case (MRMC) comparative effectiveness study focusing on human reader improvement with AI assistance is not applicable and was not performed or mentioned.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, the primary performance evaluation described is for the standalone device (algorithm only, without human-in-the-loop performance beyond operating the device). The "Performance test according to ASTM E 1965 and ISO80601-2-56" and "Clinical accuracy test" directly assess the device's ability to accurately measure temperature on its own.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the clinical accuracy validation of a thermometer, the ground truth is typically established by comparison to a highly accurate reference method (e.g., a calibrated reference thermometer measuring core body temperature, often rectal or arterial temperature, in a controlled setting) as defined by the standards like ASTM E1965 and ISO 80601-2-56. The document mentions meeting these standards for clinical accuracy.

    8. The sample size for the training set

    The document does not provide information regarding a separate "training set" for the device. Clinical thermometers are calibrated and validated against standards; they don't typically undergo machine learning training in the same way an AI diagnostic device would.

    9. How the ground truth for the training set was established

    As there's no mention of a traditional machine learning "training set" in the context of this infrared ear thermometer, this question is not applicable. The device's internal algorithm is likely developed based on physics principles of infrared radiation and calibrated against known temperature sources, rather than being "trained" on a dataset with ground truth in the AI sense.

    Ask a Question

    Ask a specific question about this device

    K Number
    K201864

    Validate with FDA (Live)

    Date Cleared
    2020-12-07

    (154 days)

    Product Code
    Regulation Number
    880.2910
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The YUWELL® Infrared thermometer is a nonsterile, reusable clinical thermometer intended for measuring the human body temperature in non-contact mode on the center of the forehead as the measurement site on people of all ages except preterm babies.

    Device Description

    The YUWELL® infrared thermometer is designed for measuring the body's frontal temperature, is a handheld non-contact infrared thermometer, battery powered, using the infrared energy emitted in the subject's forehead area within 0-5cm that converts a user's forehead temperature. The measurement reference sites is the center of the brow. The temperature measurement takes only 1 second, does not contact human skin. The infrared thermometer is intended for use on people of all ages except pre-term babies.

    AI/ML Overview

    The YUWELL Infrared Thermometer (models YT-1, YT-1A, YT-1B, YT-1C, YT-2, YT-2A, YT-2B, YT-2C) underwent non-clinical (bench) testing to demonstrate substantial equivalence to its predicate device, the Braun No touch + Forehead NTF3000 Thermometer (K163516).

    Here's a breakdown of the acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a "table of acceptance criteria and reported device performance" in the typical format of a clinical study report with specific statistical thresholds for metrics like sensitivity, specificity, or accuracy. Instead, the performance is demonstrated by conformity to recognized consensus standards for medical thermometers. The key acceptance criteria are the accuracy specifications described by ASTM E 1965 and ISO 80601-2-56.

    The following table summarizes the device's accuracy specifications, which serve as the internal acceptance criteria regarding temperature measurement performance:

    Criterion / Performance MetricAcceptance Criteria (from device specifications)Reported Device Performance (implied by "Meets ASTM E 1965 and ISO 80601-2-56")
    Accuracy
    In the range of 35.0°C ~ 42.0°C (95.0°F ~ 107.6°F)±0.2°C (±0.4°F)Meets standard
    In the range of 32.5°C ~ 34.9°C (90.5°F ~ 94.8°F) and 42.1°C ~ 43.0°C (107.8°F ~ 109.4°F)±0.3°C (±0.5°F)Meets standard
    Measurement Range32.5°C to 43.0°C (90.5°F to 109.4°F)Meets standard
    Resolution of Display0.1°C / 0.1°FMeets standard

    Beyond accuracy, other "criteria" for acceptance are successfully meeting the requirements of various international standards as listed in Section 5.9, including:

    • Software verification and validation (FDA Guidance for software in medical devices)
    • Biocompatibility (ISO 10993 and FDA Bluebook memo G95-1)
    • Electrical safety (IEC 60601-1)
    • Electromagnetic compatibility (IEC 60601-1-2)
    • Home use (IEC 60601-1-11)

    The document states, "all testing results have come back as positive results or pass for the subject device," indicating that these criteria were met.

    2. Sample size used for the test set and the data provenance

    The document describes non-clinical (bench) testing. Therefore, there is no "test set" of patient data in the context of a clinical study, nor data provenance (country of origin, retrospective/prospective). The performance testing was conducted according to ASTM E 1965 and ISO 80601-2-56, which are standards for infrared thermometers' performance. These standards involve laboratory testing with controlled temperature sources, not human subjects in a clinical setting. The specific sample sizes (e.g., number of measurements) for these bench tests are not provided in this summary.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This was a non-clinical (bench) study involving physical and electrical performance measurements against established standards, not interpretation by medical experts.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. As this was non-clinical bench testing, no adjudication of expert opinions was required.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI-powered device, nor does it involve human "readers" or case interpretation. It is a standalone infrared thermometer.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, the testing described is a standalone performance assessment of the device itself (the "algorithm" in a sense, refers to the device's internal measurement and processing capabilities) without human interpretation in the loop influencing the measurement output. The device functions as a direct measurement tool. The "Performance test according to ASTM E 1965 and ISO80601-2-56" is a standalone evaluation of the device's ability to accurately measure temperature under controlled conditions.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth for the "performance test" was established by calibrated reference temperature sources, as is standard for testing medical thermometers according to ASTM E 1965 and ISO 80601-2-56. These standards dictate precise methods for measuring the accuracy of temperature readings against known, stable temperatures.

    8. The sample size for the training set

    Not applicable. This is a traditional medical device (infrared thermometer), not an AI/Machine Learning device that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable. This is a traditional medical device, not an AI/Machine Learning device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K200939

    Validate with FDA (Live)

    Date Cleared
    2020-08-20

    (134 days)

    Product Code
    Regulation Number
    870.1130
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Electronic blood pressure monitor is intended to measure the blood pressure and pulse rate of adult in household or medical facilities. (Not suitable for neonate, pregnancy or pre-eclampsia)

    Device Description

    The measuring method for the Upper Arm Type Electronic Blood Pressure Monitor Series is oscillation mensuration. The subject devices will automatically start to take measurements after the inflation of the cuff is finished, the results will show the systolic pressure with pulse rate. The blood pressure monitor will store the measurements automatically; the time, date, blood pressure value and pulse value are included. The record maybe revisited.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Electronic Blood Pressure Monitor (models YE620B, YE620D, YE660E, YE660F, and YE680B), based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Standard)Reported Device Performance (Reference)
    Blood Pressure Accuracy: Mean difference (reference - device) and standard deviation for both systolic and diastolic blood pressure (ISO 81060-2:2018)Systolic Blood Pressure: Mean difference ± Standard Deviation: Passed (Specific values not provided in the summary, but stated as conforming to ISO 81060-2).
    Diastolic Blood Pressure: Mean difference ± Standard Deviation: Passed (Specific values not provided in the summary, but stated as conforming to ISO 81060-2).
    Pulse Accuracy: ±5% of reading value (device specification)±5% of reading value (device specification, confirmed in compliance to ISO 81060-2, which dictates this level of accuracy is met)
    General Safety: Compliance with general requirements for basic safety and essential performance (ANSI AAMI ES60601-1:2005/(R)2012 And A1:2012, C1:2009/(R)2012 And A2:2010/(R)2012 (Consolidated Text) Medical Electrical Equipment - Part 1)Passed
    Electromagnetic Disturbances: Compliance with collateral standard for electromagnetic disturbances (ANSI AAMI IEC60601-1-2:2014)Passed
    Usability: Compliance with general requirements for basic safety and essential performance - Usability (IEC 60601-1-6 Edition 3.1 2013-10)Passed
    Home Healthcare Environment: Compliance with requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment (IEC 60601-1-11:2015)Passed
    Automated Noninvasive Sphygmomanometers Specific Requirements: Compliance with particular requirements for the basic safety and essential performance of automated noninvasive sphygmomanometers (ANSI AAMI IEC80601-2-30:2009 & A1:2013 (R2016))Passed
    Biocompatibility: ISO 10993-1, FDA Guidance, Tests included Cytotoxicity, Sensitization and Intracutaneous ReactivityPassed
    Electrical Safety: IEC60601-1Passed
    EMC: IEC60601-1-2Passed
    Usability (General): IEC 60601-1-6Passed
    Home Use (General): IEC 60601-1-11Passed

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Clinical Validation (Test Set): 85 participants.
    • Data Provenance: The document does not explicitly state the country of origin for the clinical study. It is a prospective study as participants were "involved in the study" and "three valid blood pressure values are taken for each participant."

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • Number of Experts: The document states that "nurses" took the blood pressure values for the participants. The exact number of nurses is not specified, nor are their specific qualifications (e.g., years of experience in blood pressure measurement), beyond being "nurses."
    • Ground Truth Establishment: The ground truth was established by these nurses using a reference device, a mercury sphygmomanometer.

    4. Adjudication Method for the Test Set

    • The document implies that the ground truth for each participant consisted of "three valid blood pressure values" taken by nurses. An explicit adjudication method (e.g., 2+1, 3+1 consensus) for these three values is not specified. It suggests these three values were directly used, perhaps averaged, or that "valid" implies some internal consistency checking.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    • No, an MRMC comparative effectiveness study was not done. The study described is a clinical validation against a reference standard (mercury sphygmomanometer), not a comparison of human readers with and without AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • Yes, a standalone performance study was done. The clinical validation detailed is focused on the device's (algorithm's) accuracy in measuring blood pressure compared to a reference standard (mercury sphygmomanometer), without human intervention in the device's measurement process itself.

    7. The Type of Ground Truth Used

    • Expert Consensus / Reference Standard: The ground truth was established by "nurses" using a "mercury sphygmomanometer" as the reference device. This falls under the category of a reference standard established by human experts.

    8. The Sample Size for the Training Set

    • The document does not specify a training set sample size. This is typical for a 510(k) submission for a non-AI medical device like a blood pressure monitor, where the focus is on clinical validation against a standard, not machine learning model training. The device's underlying oscillometric algorithm would have been developed and validated internally by the manufacturer, but details of that process (including training sets for the algorithm itself) are not usually part of a 510(k) summary focused on clinical performance for regulatory clearance.

    9. How the Ground Truth for the Training Set Was Established

    • As no training set is explicitly mentioned or detailed in the document, how its ground truth was established is not provided.
    Ask a Question

    Ask a specific question about this device

    K Number
    K120526

    Validate with FDA (Live)

    Date Cleared
    2012-03-08

    (15 days)

    Product Code
    Regulation Number
    890.3850
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The YUYUE K2/K4 wheelchair is indicated for providing mobility to persons limited to a sitting position.

    Device Description

    The YUYUE K2/K4 wheelchair is a wheelchair that provides mobility to persons limited to a sitting position. It consists of rigid, mechanical, steel frame and nylon upholstery back and seat that meet ISO 7176-16: Resistance to ignition of Upholstered parts. It has two 24" rear wheels and two 8" front casters for turning and maneuverability. The YUYUE K2/K4 wheelchair is intended for the use in indoors and outdoors, over smooth surface (all standard indoor flooring surfaces, concrete, asphalt and pocked dirt) that free of large obstacles and inclines greater than 9 degrees.

    AI/ML Overview

    This is a premarket notification for a wheelchair and does not contain information about AI/ML algorithm acceptance criteria or study details. Therefore, I cannot provide the requested information about acceptance criteria, study details, sample sizes, expert qualifications, or ground truth.

    The document describes the YUYUE K2/K4 wheelchair and states that it meets various ISO and ASTM standards related to wheelchair performance, safety, and biocompatibility. It also mentions equivalence to a predicate device (Danyang Huayi K2 wheelchair K080207).

    Summary of available information:

    • Device: YUYUE K2/K4 wheelchair
    • Intended Use: Providing mobility to persons limited to a sitting position.
    • Standards Met:
      • ISO 7176-1: Determination of static stability
      • ISO 7176-3: Determination of efficiency of brakes.
      • ISO 7176-5: Determination of overall dimension, mass and turning space.
      • ISO 7176-7: Measurement of seating and wheel dimensions.
      • ISO 7176-8: Requirements and test methods for static, impact and fatigue strengths.
      • ISO 7176-15: Requirements for information disclosure, documentation and labeling.
      • ISO 7176-16: Resistance to ignition of upholstered parts – requirements and test methods.
      • ISO 10993-5-2009: Biological Evaluation of Medical Devices – Part5: Testes for in vitro Cytotoxicity
      • ASTM D751:2006 Section 11-Procedure A- Tensile Strength
      • Section A, Part I and Section D, Part II of California Technical Bulletin 117 Requirement: Test Procedure and Apparatus for Testing the Flame Retardance of Resilient Filling Materials Used in Upholstered Furniture.
      • Section E, Part I of California Technical Bulletin 117 Requirement: Test Procedure and Apparatus for Testing the Flame Retardance of Resilient Filling Materials Used in Upholstered Furniture.
    • Comparison: Substantially equivalent to Danyang Huayi K2 wheelchair (K080207) in terms of structure and performance.

    Missing Information (as per your request, because this is not an AI/ML device submission):

    1. A table of acceptance criteria and the reported device performance
    2. Sample sized used for the test set and the data provenance
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
    4. Adjudication method
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done
    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
    7. The type of ground truth used
    8. The sample size for the training set
    9. How the ground truth for the training set was established
    Ask a Question

    Ask a specific question about this device

    K Number
    K083559

    Validate with FDA (Live)

    Date Cleared
    2009-09-04

    (276 days)

    Product Code
    Regulation Number
    868.5440
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The YuYue OxygenMax™ 7F-5 Oxygen Concentrator is intended solely for medical use in oxygen therapy programs under the supervision of a physician. It is intended to provide supplemental oxygen to patients with respiratory disorders for use in the home or health care facility. This device is available by prescription only and is not intended to support or sustain life.

    Device Description

    The Yuyue OxygenMax™ 7F-5 oxygen concentrator (" the Yuyue 7F-5") is used by patients with respiratory disorders who require supplemental oxygen. The device can be used in the home or an institutional environment. The device is not intended to sustain or support life. The device is used with a nasal cannula to direct oxygen from the device to the patient. The Yuyue 7F-5 provides oxygen in pulsed demand flow dosages at settings of 1 through 5. The oxygen concentration level of the output gas ranges from 90% to 94%. The Yuyue 7F-5 uses a standard AC power of 120 V/60 Hertz. The Yuyue 7F-5 uses a molecular sieve and pressure swing adsorption methodology to produce the oxygen gas output. Ambient air enters the device, is filtered and then compressed. This compressed air is then directed toward one of two nitrogen adsorbing sieve beds. Concentrated oxygen exits the opposite end of the active sieve bed and is directed into an oxygen reservoir where it is delivered to the patient in specific volumes during the inhalation portion of a detected breath The basic technology of The Yuyue 7F-5 is commonly used and equivalent to other FDA approved oxygen concentrators under the same Product Code of CAW. The principles of operation are equivalent to the predicate device.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device, the Yuyue OxygenMax™ 7F-5 oxygen concentrator. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than presenting a de novo study with acceptance criteria and novel performance data. Therefore, many of the requested elements for a study proving acceptance criteria will not be present in this document.

    Here's an analysis based on the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission does not provide specific acceptance criteria in the format of a table with numerical thresholds for performance metrics. Instead, it relies on demonstrating conformity to recognized standards and substantial equivalence to a predicate device.

    Performance Metric / StandardAcceptance Criteria (Implicit)Reported Device Performance
    Oxygen Concentration LevelWithin specified range (90-94%) (Implicit from device description)Ranges from 90% to 94% (Stated in Device Description)
    Safety RequirementsConformity to ISO 8359:1996Yuyue 7F-5 conforms to ISO 8359:1996
    Standard Specification for Domiciliary UseConformity to ASTM F1464-93 (2005)Yuyue 7F-5 conforms to ASTM F1464-93 (2005)
    Electromagnetic Compatibility (EMC)Conformity to IEC 60601-1-2:2001Yuyue 7F-5 conforms to IEC 60601-1-2:2001
    EMI (Electromagnetic Interference)Conformity to FDA Reviewer Guidance document "Excerpts Related to EMI from November 1993" as appropriateYuyue 7F-5 conforms to FDA Reviewer Guidance for EMI
    Overall Safety and EffectivenessSubstantially equivalent to predicate deviceShown to be safe and effective for intended use and substantially equivalent to predicate device.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not describe a test set or data provenance in the context of a clinical study or new performance data. The evaluation is based on bench testing against established standards and comparison to a predicate device. Therefore, there is no "test set" in the traditional sense for evaluating AI or diagnostic performance.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. This is a 510(k) submission for an oxygen concentrator, not an AI or diagnostic device that relies on expert ground truth for a test set.

    4. Adjudication Method for the Test Set

    Not applicable. There is no test set requiring adjudication in this submission.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    Not applicable. This device is an oxygen concentrator, not an AI-assisted diagnostic tool.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not applicable. This device is an oxygen concentrator, not an algorithm. Bench tests were performed to demonstrate compliance with standards and equivalence to the predicate device.

    7. The Type of Ground Truth Used

    The "ground truth" in this context is the conformity to established engineering and medical device standards (ISO 8359, ASTM F1464-93, IEC 60601-1-2) and the demonstrated equivalence to a legally marketed predicate device (John Bunn® O2 Premier oxygen concentrator, K844223). This is primarily based on technical specifications, functional testing, and regulatory compliance, rather than clinical outcomes or pathology data for a new diagnostic claim.

    8. The Sample Size for the Training Set

    Not applicable. This device is an oxygen concentrator, not an AI algorithm requiring a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. This device is an oxygen concentrator.

    Summary of the Study and Device Proof:

    The study proving the device meets the acceptance criteria is primarily a non-clinical performance evaluation based on bench testing and a substantial equivalence comparison to a predicate device.

    • Bench Tests: The Yuyue 7F-5 underwent bench tests to conform to the following recognized standards:
      • ISO 8359:1996 for oxygen concentrators for medical use - safety requirements.
      • ASTM F1464-93 (2005) for oxygen concentrators for domiciliary use.
      • IEC 60601-1-2:2001 for medical electrical equipment - electromagnetic compatibility.
      • FDA Reviewer Guidance document "Excerpts Related to EMI from November 1993" as appropriate.
    • Substantial Equivalence: The Yuyue 7F-5 was determined to be substantially equivalent to the John Bunn® O2 Premier oxygen concentrator (K844223). The submission states that, "Except using different trade name and model name, the subjective device and the predicate device is identical in the intended use, specification and other key characteristics." Minor power supply differences between models were noted but the counterpart models (Yuyue 7F-5 and John Bunn O2 Premier, JB0160-010/015) use identical power supply.

    The results of these tests and the successful demonstration of substantial equivalence were deemed sufficient by the FDA to determine that the Yuyue 7F-5 oxygen concentrator is safe and effective for its intended use.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1