Search Filters

Search Results

Found 13 results

510(k) Data Aggregation

    K Number
    K240087

    Validate with FDA (Live)

    Date Cleared
    2024-09-13

    (246 days)

    Product Code
    Regulation Number
    870.2900
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Ivy Biomedical RT Carbon ECG Leads are intended for use in X-ray imaging environments. They connect between ECG electrodes that are placed on the patient by a trained healthcare professional and an ECC trunk cable that connects to the ECG monitoring device. The RT Carbon ECG Leads are reusable and nonsterile. The RT Carbon ECG Leads are intended for adult, pediatric and neonatal patient populations.

    Device Description

    The RT Carbon ECG Leads are intended to connect electrodes placed on the patient to a physiological detection module or monitor (CTM-300, CTM-400, CTM-500, 7000 Series and 3000 Series). The design includes 3 and 4 lead designs with overall lengths of 24", 30" and 36".

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the Ivy Biomedical RT Carbon ECG Leads. This document focuses on demonstrating substantial equivalence to a predicate device through performance testing rather than presenting novel clinical efficacy or diagnostic performance data. Therefore, many of the requested categories related to AI performance studies are not applicable.

    Here's the information extracted and categorized as requested:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly present a table of acceptance criteria alongside reported device performance in the format of specific thresholds and measured values for the RT Carbon ECG Leads. Instead, it states that the device was "thoroughly tested from design requirements through device verification and validation testing as required by 21 CFR 870.2900" and in accordance with EC53. The reported performance is summarized as:

    "The performance data support the safety of the device and hardware verification and validation demonstrate that the RT Carbon ECG Leads should perform as intended in the specified use conditions and is substantially equivalent to the predicate."

    This implies that the device met the acceptance criteria derived from the listed tests.

    Acceptance Criterion (Implicit from tests performed)Reported Device Performance
    Biocompatibility:
    CytotoxicityMet (implies no cytotoxicity)
    SensitizationMet (implies no sensitization)
    IrritationMet (implies no irritation)
    Systemic ToxicityMet (implies no systemic toxicity)
    Pyrogen TestingMet (implies no pyrogenicity)
    Performance Testing (per EC53):
    Patient Leadwire to Trunk Cable InterconnectionMet (implies proper interconnection)
    Flex lifeMet (implies satisfactory durability)
    Tensile strengthMet (implies satisfactory strength)
    Connector mating/unmating cyclesMet (implies satisfactory durability)
    Connector retention forceMet (implies satisfactory retention)
    Contact resistanceMet (implies satisfactory electrical contact)
    Leadwire resistanceMet (implies satisfactory electrical resistance)
    Dielectric withstand voltageMet (implies satisfactory insulation)

    2. Sample size used for the test set and the data provenance

    The document does not specify the sample size for individual tests conducted on the RT Carbon ECG Leads. The testing was described as "conducted at Ivy" (implying in-house lab testing) and does not involve human data, thus the provenance (country of origin, retrospective/prospective) is not applicable in the context of clinical data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This is not an AI/diagnostic device that requires expert ground truth establishment for a test set. The tests are engineering and biological evaluations.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This is not an AI/diagnostic device that requires adjudication of interpretations.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is an ECG lead, not an AI system or diagnostic tool for interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is an ECG lead, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The 'ground truth' for the performance testing is based on established engineering standards (EC53) and biocompatibility standards (ISO 10993-1). Compliance with these standards serves as the benchmark against which the device's performance is measured.

    8. The sample size for the training set

    Not applicable. This is a medical device (ECG lead), not an AI model that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable. This is a medical device (ECG lead), not an AI model.

    Ask a Question

    Ask a specific question about this device

    K Number
    K202138

    Validate with FDA (Live)

    Date Cleared
    2021-05-14

    (287 days)

    Product Code
    Regulation Number
    870.2300
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Ivy Biomedical Model 7600EP/7800EP is a basic cardiac monitor used to provide cardiac trigger pulse outputs used by third-party diagnostic imaging systems that require ECG synchronization, such as nuclear medicine, computed axial (CAT), or positron emission (PET) tomography and other imaging systems requiring similar cardiac cycle specific timing. The Ivy Biomedical Model 7600EP/7800EP monitors can also be used to provide cardiac trigger pulse output used by a third-party ablation and lithotripsy systems.

    Device Description

    Not Found

    AI/ML Overview

    I understand you're asking for details about the acceptance criteria and study data for a medical device. However, the provided text is an FDA 510(k) clearance letter for the Ivy Biomedical Model 7600EP/7800EP Cardiac Synchronization Monitor.

    This document explicitly states that the device is a "Cardiac Monitor (Including Cardiotachometer And Rate Alarm)" and is intended to "provide cardiac trigger pulse outputs used by third-party diagnostic imaging systems that require ECG synchronization."

    This is not an AI/ML medical device, and the provided text does not contain any information about acceptance criteria, study data, ground truth establishment, or any of the other AI/ML-specific details you are asking for.

    Therefore, I cannot extract the information requested for a study proving an AI/ML device meets acceptance criteria, as the document concerns a traditional medical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K203653

    Validate with FDA (Live)

    Date Cleared
    2021-03-02

    (77 days)

    Product Code
    Regulation Number
    892.1000
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MRI PPG Patient Cable is intended to connect to a parent device and facilitate the generation of continuous noninvasive peripheral pulse cardiac gating on adult, pediatric, and neonatal patients in hospital and/or health care environments.

    Device Description

    Not Found

    AI/ML Overview

    The provided document is a 510(k) clearance letter for the "MRI PPG Patient Cable" and contains "Indications for Use" information. It does not include details on acceptance criteria or a study proving that the device meets such criteria.

    The information provided describes the device's intended use and regulatory classification, but it does not delve into its performance metrics, how those metrics were evaluated, or any studies conducted to validate its performance against specific acceptance criteria.

    Therefore, I cannot provide the requested information based on the given text. The document focuses on regulatory clearance rather than technical performance validation.

    Ask a Question

    Ask a specific question about this device

    K Number
    K192852

    Validate with FDA (Live)

    Date Cleared
    2020-03-28

    (177 days)

    Product Code
    Regulation Number
    892.1000
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MRI ECG Patient Cable is intended to connect electrodes placed on a patient to a host device. The cable is intended for use in up to 3.0T in MRI environments. It is used in a professional healthcare facility environment.

    Device Description

    Not Found

    AI/ML Overview

    I am sorry, but the provided text is a 510(k) clearance letter from the FDA for a medical device (MRI ECG Patient Cable). This letter does not contain any information about acceptance criteria, study details, performance data, or ground truth establishment.

    Therefore, I cannot fulfill your request to describe the acceptance criteria and the study that proves the device meets them, as this information is not present in the given document. The letter only states that the device is substantially equivalent to legally marketed predicate devices and is subject to general controls.

    Ask a Question

    Ask a specific question about this device

    K Number
    K170828

    Validate with FDA (Live)

    Date Cleared
    2017-05-17

    (58 days)

    Product Code
    Regulation Number
    870.2300
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Ivy Biomedical Model 7600, 7800 is a basic cardiac monitor used to provide cardiac trigger pulse outputs used by third-party systems that require ECG synchronization is commonly used in diagnostic imaging modalities (i.e., nuclear medicine, computed axial (CAT), or positron emission (PET) tomography) or other applications requiring such synchronization.

    The Model 7600/7800 is intended for use in neonatal, pediatric and adult patients undergoing diagnostic imaging and related procedures in inpatient and outpatient centers under the supervision of licensed healthcare professionals.

    The product is not intended for use as a life support, home monitoring, or magnetic resonance imaging (MRI) modality.

    Device Description

    The Ivy Biomedical Model 7600, 7800 is a basic cardiac monitor used to provide cardiac trigger pulse outputs used by third-party systems that require ECG synchronization.

    AI/ML Overview

    The provided text does not contain information about acceptance criteria or a study proving that a device meets such criteria. The document is an FDA 510(k) clearance letter for the Ivy Biomedical Systems Cardiac Trigger Monitor, Models 7600 and 7800. It outlines the regulatory classification, indications for use, and general requirements for marketing the device.

    Therefore, I cannot answer the questions regarding acceptance criteria, study details, sample sizes, expert qualifications, adjudication methods, MRMC studies, standalone performance, or ground truth establishment based on the provided input.

    Ask a Question

    Ask a specific question about this device

    K Number
    K110987

    Validate with FDA (Live)

    Date Cleared
    2011-07-22

    (105 days)

    Product Code
    Regulation Number
    870.2300
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Ivy Biomedical Model 7600 Cardiac Trigger Monitor is a simple to use instrument for monitoring the ECG and Heart Rate. It is designed for use in the ICU, CCU and operating room conditions. It can sound an alarm when the HR falls outside of preset limits. It provides an output pulse, synchronized to the R-wave for use in applications requiring precision R-wave synchronization. The Model 7600 is limited to use by trained and qualified medical professionals. This product is not intended for home care monitoring.

    Device Description

    The Ivy Biomedical Model 7600 Cardiac Trigger Monitor is a simple to use instrument for monitoring the ECG and Heart Rate. It is designed for use in the ICU, CCU and operating room conditions. It can sound an alarm when the HR falls outside of preset limits. It provides an output pulse, synchronized to the R-wave for use in applications requiring precision R-wave synchronization.

    AI/ML Overview

    Based on the provided text, the document is a 510(k) clearance letter for the Ivy Biomedical Model 7600 Cardiac Trigger Monitor. It confirms the device's substantial equivalence to a predicate device but does not contain information regarding detailed acceptance criteria, performance studies, or technical specifications of the device's performance.

    Therefore, I cannot provide the requested information about acceptance criteria, study details, sample sizes, expert qualifications, or ground truth establishment. The document focuses on regulatory clearance, not the technical validation of the device's performance against specific metrics.

    Ask a Question

    Ask a specific question about this device

    K Number
    K091787

    Validate with FDA (Live)

    Date Cleared
    2009-10-09

    (114 days)

    Product Code
    Regulation Number
    870.2700
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Model 6000 is intended to provide continuous noninvasive monitoring of SpO2 and EtCO2 indications. The parameters are derived from both the measurements of arterial oxygen hemoglobin saturation and pulse rates as well as relative expired and inspired CO2 levels. The recorded data may be displayed locally and also transmitted wirelessly to any device supporting the 802.11 communication protocol.

    This device is intended for use on adult, geriatric, and neonatal patients in bedside or mobile applications within health care institutions and/or health care environments only under the direction of a physician or qualified clinician.

    Device Description

    The device is a two parameter patient monitor. Both EtCO2 and SpO2 parameters are derived from commercially available modules that are incorporated internally. The device displays the waveforms and numerical values associated with each parameter on a color LCD display, and transmits the information to a network based computing device over the 802.11a - g IEEE wireless protocol. The device is intended for use on adult, geriatric, pediatric, and neonatal patients only under the direction of a physician or qualified clinician.

    AI/ML Overview

    The provided text describes a 510(k) summary for the Ivy Biomedical Systems Model 6000 Two Parameter Bedside Monitor. It details the device's characteristics, intended use, and substantial equivalence to a predicate device. However, it does not include specific information about acceptance criteria for performance metrics (e.g., accuracy, precision for SpO2 or EtCO2) nor a dedicated study section with the detailed information requested in the prompt (sample sizes, ground truth establishment, expert qualifications, MRMC studies, standalone performance, etc.).

    The document states: "The device was verified and validated according to the product specifications. The test criteria consist of standardized levels and internal product requirements. Tests performed on the device include environmental and mechanical stress testing, electromagnetic immunity and emissions testing, and medical device safety testing. Software on the device was verified and validated according to the functionality of the operations of the device. The test results confirm that the device is in accordance with its specifications."

    This summary indicates that performance and safety testing were conducted internally, aligning with various industrial and medical device standards (UL 60601-1, IEC 60601-1-2, etc.). However, it does not explicitly state the quantitative acceptance criteria for SpO2 and EtCO2 accuracy, nor does it present the results of such performance tests in a format that allows for the construction of the requested table or detailed study description.

    Therefore, for many of the requested points, the information is not available in the provided text.

    Here is a summary of what can and cannot be extracted from the provided text based on your prompt:


    Acceptance Criteria and Study for Model 6000 Two Parameter Bedside Monitor

    1. Table of Acceptance Criteria and Reported Device Performance:

    Performance MetricAcceptance CriteriaReported Device Performance
    SpO2 AccuracyNot specifiedNot specified
    EtCO2 AccuracyNot specifiedNot specified
    Pulse Rate AccuracyNot specifiedNot specified
    Respiratory Rate AccuracyNot specifiedNot specified
    Environmental & Mechanical Stress TestingStandardized levels and internal product requirementsConfirmed compliance with specifications
    Electromagnetic Immunity & Emissions TestingStandardized levels and internal product requirementsConfirmed compliance with specifications
    Medical Device Safety TestingStandardized levels and internal product requirementsConfirmed compliance with specifications
    Software Verification & ValidationFunctionality of operationsConfirmed compliance with specifications

    Explanation: The document broadly states that "The device was verified and validated according to the product specifications. The test criteria consist of standardized levels and internal product requirements. The test results confirm that the device is in accordance with its specifications." However, it does not provide specific numerical acceptance criteria or reported performance values for parameters like SpO2 or EtCO2 accuracy which would typically be included in a detailed performance study for such a device.

    2. Sample size used for the test set and the data provenance:

    • Sample Size: Not specified.
    • Data Provenance: Not specified (e.g., country of origin, retrospective/prospective). The document implies internal testing and compliance with various international standards (UL, IEC, ISO, CAN/CSA, CISPR), but no details on clinical data collection are provided.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable/Not specified. This device is a monitor measuring physiological parameters, not an diagnostic imaging or interpretation device that would typically rely on human expert consensus for ground truth on a test set. The ground truth for such devices usually involves highly accurate reference measurement equipment. However, the document does not detail how the ground truth for its performance validation (e.g., SpO2 or EtCO2 accuracy) was established.

    4. Adjudication method for the test set:

    • Not applicable/Not specified. Given the nature of the device (physiological monitor), an adjudication method in the context of expert review is not typically relevant as it would be for AI-driven interpretation tasks.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable/Not specified. The Model 6000 is a patient monitor, not an AI-assisted diagnostic tool that interprets data for human readers. Therefore, an MRMC study comparing human readers with and without AI assistance is not relevant to this type of device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • The device is a standalone monitor. Its performance would inherently be evaluated in a standalone manner, measuring its accuracy against reference standards. However, the details of such standalone performance testing (e.g., specific accuracy metrics, methodologies) are not provided in this 510(k) summary beyond the general statement of "verification and validation."

    7. The type of ground truth used:

    • Not specified. For physiological monitors, ground truth typically involves highly accurate, gold-standard reference devices (e.g., a CO-oximeter for SpO2, a calibrated gas analyzer for EtCO2). The document mentions that "EtCO2 and SpO2 parameters are derived from commercially available modules that are incorporated internally," implying these modules themselves may have established accuracy, but it doesn't detail the validation of the integrated system against an independent ground truth.

    8. The sample size for the training set:

    • Not applicable/Not specified. This device is a physiological monitor, not a machine learning model that requires a "training set." Its functionality is based on established engineering principles and commercially available modules, rather than learned patterns from a large dataset.

    9. How the ground truth for the training set was established:

    • Not applicable. As there's no mention of a training set for a machine learning model, this question is not relevant to the described device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K091985

    Validate with FDA (Live)

    Date Cleared
    2009-07-24

    (22 days)

    Product Code
    Regulation Number
    870.2300
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Universal Physiological Measurement Module is intended for use on patients requiring Rwave synchronization and timed imaging studies. Applications include: gamma cameras, TMR/PMR, multislice CT, PET/CT, and SPECT/CT.

    The UPMM is responsible for detecting the patient's heart waveform and producing a synchronized pulse that is used to trigger a device.

    Device Description

    Universal Physiological Measurement Module (UPMM)

    AI/ML Overview

    This document is a 510(k) clearance letter from the FDA for a device called "Universal Physiological Measurement Module (UPMM)". It is not a study report or clinical trial documentation. Therefore, it does not contain the detailed information required to describe the acceptance criteria and the study that proves the device meets those criteria.

    The document states that the FDA reviewed a "Section 510(k) premarket notification" and determined the device is substantially equivalent to legally marketed predicate devices. This means the device met the regulatory requirements for 510(k) clearance at the time, which typically involves demonstrating that the new device is as safe and effective as an already legally marketed device. However, the clearance letter itself does not provide:

    • A table of acceptance criteria or reported device performance.
    • Details about sample sizes for test sets, data provenance, number of experts, or adjudication methods for any studies.
    • Information on comparative effectiveness studies (MRMC) or standalone performance studies.
    • Information on ground truth establishment for test or training sets, nor the sample size of any training sets.

    In summary, based on the provided text, it is not possible to answer your questions as the document is an FDA clearance letter, not a scientific study report.

    Ask a Question

    Ask a specific question about this device

    K Number
    K083854

    Validate with FDA (Live)

    Date Cleared
    2009-02-13

    (51 days)

    Product Code
    Regulation Number
    870.2300
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Cardiac Trigger Module model CTM-300 is intended for use on patients requiring Rwave synchronization and timed imaging studies. Applications include: gamma cameras, TMR/PMR, multislice CT, PET/CT and SPECT/CT.

    The CTM-300 is responsible for detecting the patient's heart waveform and producing a synchronized pulse that is used to trigger a device.

    Device Description

    The Cardiac Trigger Module model CTM-300 is responsible for detecting the patient's heart waveform and producing a synchronized pulse that is used to trigger a device.

    AI/ML Overview

    This document is a 510(k) clearance letter from the FDA for a medical device called the "Cardiac Trigger, Model CTM300." It does not include detailed information about acceptance criteria or specific study results to prove device performance.

    Therefore, I cannot fulfill the request to describe the acceptance criteria and the study that proves the device meets them based only on the provided text. The document primarily focuses on regulatory approval based on "substantial equivalence" to a predicate device, rather than providing a performance study report.

    The information you are asking for (acceptance criteria table, sample sizes, expert qualifications, adjudication methods, MRMC studies, standalone performance, ground truth types, training set details) is typically found in the 510(k) summary or the full 510(k) submission, which is not provided here.

    Ask a Question

    Ask a specific question about this device

    K Number
    K023616

    Validate with FDA (Live)

    Date Cleared
    2002-11-04

    (6 days)

    Product Code
    Regulation Number
    870.2300
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Ivy Biomedical Model 3000 Patient Monitor is a simple to use instrument for monitoring the ECG and Heart Rate. It is designed for use in the ICU, CCU and operating room conditions. It can sound an alarm when the HR falls outside of preset limits. It provides an output pulse, synchronized to the R-wave for use in applications requiring precision R-wave synchronization. The Model 3000 is limited to use by trained and qualified medical professionals. This product is not intended for home care monitoring.

    Device Description

    Not Found

    AI/ML Overview

    The provided text is an FDA 510(k) clearance letter for the Ivy Biomedical Model 3000 Patient Monitor. It does not contain the detailed information required to answer your request regarding acceptance criteria, study details, sample sizes, expert qualifications, or ground truth establishment.

    A 510(k) summary typically includes a brief description of the device, its intended use, a comparison to a predicate device, and a summary of performance data (often in the form of a statement of substantial equivalence). However, it generally does not provide the granular technical details or study methodologies you're asking for.

    The letter explicitly states: "We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices..." This heavily implies that the clearance was based on demonstrating equivalence to an existing device, rather than a novel study with specific acceptance criteria that would then be detailed.

    Therefore, I cannot extract the following information from the provided text:

    • Table of acceptance criteria and reported device performance: Not present.
    • Sample size used for the test set and data provenance: Not present.
    • Number of experts used to establish ground truth and qualifications: Not present.
    • Adjudication method: Not present.
    • Multi-reader multi-case (MRMC) comparative effectiveness study, effect size: Not present.
    • Standalone (algorithm only) performance: Not explicitly discussed, as this is a patient monitor, not an AI algorithm in the context you're suggesting.
    • Type of ground truth used: Not present.
    • Sample size for the training set: Not present (and likely not applicable in the context of this type of device and clearance).
    • How the ground truth for the training set was established: Not present.

    The only information that can be extracted related to the device and its use is from the "Indications for Use Statement" (Attachment 3):

    • Device Name: Ivy Biomedical Model 3000
    • Indications for Use: "The Ivy Biomedical Model 3000 Patient Monitor is a simple to use instrument for monitoring the ECG and Heart Rate. It is designed for use in the ICU, CCU and operating room conditions. It can sound an alarm when the HR falls outside of preset limits. It provides an output pulse, synchronized to the R-wave for use in applications requiring precision R-wave synchronization. The Model 3000 is limited to use by trained and qualified medical professionals. This product is not intended for home care monitoring."
    Ask a Question

    Ask a specific question about this device

    Page 1 of 2