510(k) Data Aggregation
(47 days)
The Ivenix LVP Epidural Administration Set with NRFit Connector (SET-0030-1) is intended for the controlled administration of pharmaceutical drugs or other therapies via the epidural pathway. The Ivenix LVP Epidural Administration Set is intended only for use with the Ivenix Infusion System in a hospital or outpatient.
The Ivenix LVP Epidural Administration Set with NRFit connector is comprised of a primary channel inlet tubing with a vented cap spike, a cassette which contains the fluid pumping mechanism and serves as the interface to the Ivenix LVP, and outlet tubing for connection to patient access. The end of the outlet tubing is fitted with a NRFit connector that is compliant to ISO 80369-6:2013.
The Ivenix LVP Epidural Administration Set with NRFit Connector provides a sterile fluid path for the delivery of pharmaceutical drugs and other therapies via the epidural pathway, provides a means for a clinician to aseptically connect to the fluid source and to the patient access site, and interfaces with the LVP to enable measurement and control of the fluid delivery.
The Ivenix LVP Epidural Administration Set with NRFit Connector is intended for use within the Ivenix Infusion System for controlled fluid delivery; it is not recommended for gravity infusion.
The provided text describes the 510(k) premarket notification for the Ivenix LVP Epidural Administration Set with NRFit Connector. It is not an AI/ML medical device, and therefore, many of the requested categories in your prompt are not applicable.
However, I can extract information related to the acceptance criteria and the studies performed to demonstrate substantial equivalence for this specific device, which is a hardware medical device (an administration set).
Here's a breakdown of the applicable information:
1. A table of acceptance criteria and the reported device performance
The document presents performance testing results to demonstrate that the new subject device (with NRFit connector) is substantially equivalent to the predicate device (with luer connector). The "acceptance criteria" can be inferred from the standards and requirements the device was tested against.
| Acceptance Criteria (Inferred) | Reported Device Performance |
|---|---|
| Conformance to ISO 80369-6:2016 for NRFit connector | The NRFit connector was tested against the requirements of ISO-80369-6:2016, using the methods described in ISO-80369-20:2015, by an accredited test laboratory. All requirements were met. |
| Connector bond strength (meets ISO 8536-4:2019 requirements) | The epidural administration set with NRFit met the pull strength requirements of ISO 8536-4:2019. |
| Flow accuracy and shelf life meets functional requirements | The epidural administration set with NRFit connector met all functional requirements of the Ivenix epidural set with luer connector (the predicate device). |
| Microbial Ingress (maintains sterile fluid path) | The sterile fluid path of the administration set was challenged, under simulated worst-case conditions, at the patient access device, with four strains of bacteria (2 gram positive, 2 gram negative). Test results showed no growth of microorganism in any test article, demonstrating the integrity of the NRFit connector to maintain the sterile fluid path under simulated conditions of use. |
| Biocompatibility (equivalent to predicate device material) | The polymer material of the NRFit connector is identical to the standard luer connector of the predicate device in its final finished form in formulation, processing, sterilization, and geometry (with the exception of the geometry necessary to meet the requirements of ISO 80369-6:2016), and no other chemicals have been added. Therefore, no additional biocompatibility testing was conducted. (This implies an "acceptance" that no new testing was needed due to material equivalence). |
2. Sample size used for the test set and the data provenance
- Sample Size: The document does not explicitly state the sample sizes used for each of the bench tests (e.g., how many connectors were tested for bond strength or how many sets for flow accuracy). For microbial ingress, it mentions "any test article" implying multiple articles were tested for each of the four bacteria strains.
- Data Provenance: This was "bench testing" performed by the manufacturer, or by an "accredited test laboratory" on behalf of the manufacturer. This is typical for a hardware device submission. The data is prospective for the purpose of this submission. The country of origin is not specified but implicitly within the regulatory framework of the U.S. FDA.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. This device is a hardware administration set, not an AI/ML algorithm that requires expert "ground truth" for diagnostic or predictive purposes. The "ground truth" here is objective physical and biological performance as measured against established standards.
4. Adjudication method for the test set
Not applicable. This is not an AI/ML algorithm with subjective interpretations requiring adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is a hardware device, not an AI-assisted diagnostic tool.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This is a hardware device.
7. The type of ground truth used
The "ground truth" for this device is based on:
- International standards (ISO 80369-6:2016, ISO 80369-20:2015, ISO 8536-4:2019).
- Functional requirements for existing predicate devices.
- Objective measurements of physical properties (e.g., pull strength, flow accuracy, sterility maintenance).
8. The sample size for the training set
Not applicable. This is a hardware device, not an AI/ML algorithm that requires a training set.
9. How the ground truth for the training set was established
Not applicable.
Ask a specific question about this device
(190 days)
The Ivenix Infusion System is indicated for use in a hospital and in outpatient care environments for the controlled administration of fluids through clinically accepted routes of administration: intra-arterial, epidural, and subcutaneous, to adults, pediatric and neonate patients. Administered fluids may be pharmaceutical drugs, red blood cells, platelets, plasma, and other mixtures required for patient therapy.
The Ivenix Large Volume Pump (LVP) is indicated for use only with the Ivenix sterile, single use, disposable administration sets, including:
Primary dual inlet, single outlet, 1 Y-site Primary single inlet, single outlet Blood Set, dual inlet, single outlet, 1 Y-site, for administration of red blood cells, platelets and plasma Microbore single inlet, single outlet, for administration of epidural fluids Microbore dual inlet, single outlet, 1 Y-site
Infusion Management System (IMS):
The IMS provides information to the clinician regarding the use of the Ivenix Large Volume Pump (LVP) by way of a drug library, LVP configurations, and by providing remote information regarding LVP status. It also provides information on pump usage data reports to various functions within an institution.
The Ivenix Infusion System (IIS) consists of three primary components:
-
- Single channel large volume pump (LVP),
-
- Single use, sterile, disposable fluid administration sets, and
-
- Infusion Management System (IMS) server applications.
The Large Volume Infusion Pump (LVP) is built around a closed-loop, pneumatically driven fluid delivery mechanism. Negative pressure is used to draw liquid from the source container into an intermediate pumping chamber (IPC). Positive pressure is then used to push the liquid out of the IPC and to the patient. The pneumatic drive allows for direct measurement of volume changes in the IPC.
The pump used the ideal gas equation, P1V1 = P2V2, to accurately calculate volume of fluid delivered. Using a known reference volume, pressure measurements are taken at various intervals during the fill and empty cycles to allow the system to determine flow rate to the patient. The LVP then adjusts drive pressure and/or fluidic resistance through a variable fluidic resistor to bring the actual flow rate to match the target flow rate.
The LVP System is a dual-inlet, single-outlet delivery system that can selectively draw from up to two fluid sources and deliver either of those fluids to the patient through a single outlet. Active delivery from two inlets provides for managed intermittent and "piggy-back" infusions. The system does not support concurrent delivery of the same time. The LVP can use 5 ml to 60 ml syringes, manufactured by Becton Dickinson, B. Braun, or Covidien, directly connected to the secondary inlet of the primary and microbore administration sets as a fluid source.
The system delivers fluid through a single-use, dual-inlet, single-outlet, disposable administration set. There are five varieties of administration sets:
-
- Primary dual inlet, single outlet, 1 Y-site
-
- Primary single inlet, single outlet
- Blood Set, dual inlet, single outlet. 1 Y-site, for administration of red blood cells, platelets and plasma 3.
- Microbore single inlet, single outlet, for administration of epidural fluids 4.
-
- Microbore dual inlet, single outlet, 1 Y-site
The Infusion Management System (IMS) allows the pump to share information with other pumps and applications provided by Ivenix, such as an infusate library, or other systems that reside within the hospital. In addition, the IMS provides other features, such as remote viewing dashboards, clinical analytics, and advisories that provide information about the infusion delivery both at, and away from, the point of care.
The medical device described in the provided text is an infusion pump, and the document is a 510(k) summary for its premarket notification. Infusion pumps are typically subject to performance testing against established standards, rather than clinical studies with "acceptance criteria" and "ground truth" as might be seen for diagnostic AI devices.
Based on the provided information, I can answer your questions, specifically pointing out where the concepts you've asked about (like "acceptance criteria" for accuracy, "ground truth" for training, etc.) are addressed in the context of an infusion pump's regulatory submission.
1. A table of acceptance criteria and the reported device performance
For an infusion pump, "acceptance criteria" primarily relate to engineering and performance specifications outlined in recognized standards and internal validation. The document compares the subject device (Ivenix Infusion System) to a predicate device (Baxter SIGMA Spectrum Infusion Pump) on various technological characteristics, which act as de-facto performance criteria for achieving substantial equivalence.
| Acceptance Criteria (Characteristic) | Reported Device Performance (Ivenix Infusion System - IIS) | Discussion/Comparison to Predicate |
|---|---|---|
| Infusion Accuracy | +/- 5% @ 0.5 - 1000 ml/hr. (Subject) | Predicate: +/- 5% @ 2 – 999 ml/hr, +/- 0.1 ml/hr for rates < 2 ml/hr. IIS flow rate accuracy is consistent at all flow rates, while the predicate has wider tolerance at rates < 2 ml/hr. Considered substantially equivalent. |
| Maximum Infusion Pressure | 931 mm Hg (Subject) | Predicate: 1488 mm Hg. Differences due to pumping mechanisms; both greater than max occlusion pressure alarm thresholds. No new safety/effectiveness questions. |
| Programmable Flow Rate Range | 0.5 – 1000 ml/hr. (Subject) | Predicate: 0.5 - 999 ml/hr. Essentially the same; 1 ml/hr difference at high end not clinically significant. |
| Time to Occlusion Alarm | Approximately 9 seconds to 30 minutes, depending on flow rate, pressure threshold setting, and distal tubing type. (Subject) | Predicate: Approximately 50 seconds to 3 hours. IIS generally has more sensitive detection and raises alarm faster. No new safety/effectiveness questions. |
| Post-occlusion Bolus | Less than 0.5 mL at maximum occlusion detection pressure and all rates. (Subject) | Predicate: Less than 0.8 mL at maximum occlusion detection pressure and 25 mL/hr. IIS is constant at 0.5mL at any flow rate. |
| Alarms & Alerts | Air-in-line, Battery low, Battery depleted, Bolus complete, KVO, Infusion complete, Infusion not started, Restricted flow, Infusion paused, Occlusion (upstream or downstream), Pump problem, Tubing set problem, Tubing set removed, Dose limits exceeded. | Alarms validated based on intended use and unique characteristics. Alarm priorities similar. No new safety/effectiveness questions. |
| Device Service Life | 5 years (Subject) | Predicate: Not specified. IIS evaluated through performance testing to meet essential performance throughout service life. |
| Ingress Protection | IPX3 Spraying water (Subject) | Predicate: IPX0 – None. IIS has higher protection, verified through performance testing. Does not raise new safety/effectiveness questions. |
| Upstream Air Management | Air elimination feature. (Subject) | Predicate: Air Trap allows 1 mL of air in either A or B line before infuser sounds cassette alarm. Both attempt to purge or trap excess air. |
| Cannula/Connector Type | 2-piece luer lock. (Subject) | Predicate: Luer connector with a locking collar. Both use standard luer connections. |
| Syringe Compatibility | 5 ml to 60 ml syringes (Becton Dickinson, B. Braun, Covidien) (Subject) | Predicate: 3 ml to 60 ml syringes. IIS minimum syringe size is 5 ml. Essentially the same. |
| General Safety (Elec., EMC, Wireless) | Complies with AAMI/ANSI ES 60601-1, IEC 60601-1-2, FDA Guidance "Radio Frequency Wireless Technology," ANSI C63.27-2017, AIM 7351731 Rev. 2.00 RFID Immunity. | Standard safety and performance criteria for medical devices. Demonstrated compliance. |
| Biocompatibility | Complies with ISO 10993-1, testing for cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity, hemocompatibility. USP <661>, AAMI ST72, USP 39 <788>, ISO 8536-4. | Standard biocompatibility criteria for patient-contacting materials. Demonstrated compliance. |
| Microbial Ingress | No growth of microorganism showed under simulated worst-case conditions. (Subject) | Standard for sterile fluid path. Demonstrated integrity of sterile fluid path. |
| Blood Products Compatibility | No appreciable degradation of packed red blood cells, platelets, plasma observed when pumped at various flow rates. (Subject) | Demonstrated compatibility with various blood products. |
| Software & Cybersecurity | Complies with FDA's Guidance for "Software Contained in Medical Devices," IEC 62304, FDA Guidance "General Principles of Software Validation," FDA Guidance "Cybersecurity in Medical Device." | Standard for software validation and cybersecurity. Demonstrated compliance for a "Major" level of concern software. |
| Human Factors | Complies with FDA Guidance "Applying Human Factors and Usability Engineering to Medical Devices." Studies conducted with intended user population/environment/scenarios. | Standard for human factors validation. Demonstrated validation of device per intended use. |
2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)
For the performance testing of an infusion pump, "sample size" refers to the number of devices or components (e.g., administration sets, syringes) tested under various conditions to ensure they meet their specifications. The document details that various tests were performed, but does not specify quantitative sample sizes (e.g., N=X devices tested) or the provenance of this test data (e.g., conducted internally, by a third-party lab). These are typically highly controlled engineering tests, not based on real-world patient data in the way an AI diagnostic device would be.
- Administration Set Compatibility: "Verification of the pump essential performance was completed with the indicated administration sets"
- Syringe Compatibility: "Performance testing was conducted on 5 ml to 60 ml syringes manufactured by B. Braun, Becton Dickinson, and Covidien at flow rates from 0.5 ml/hr to 20 ml/hr." The specific number of syringes or tests is not given.
- Microbial Ingress: "The sterile fluid path of the administration set was challenged, under simulated worst case conditions, at the needless access ports, the spike connection to the fluid source, and the luer connection to the patient access device, with four strains of bacteria..."
- Blood products compatibility: "Packed red blood cells, platelets, and plasma were pumped through the Ivenix device at 14 ml/hr, 350 ml/hr, and 500 ml/hr flow rates." The number of samples or trials for each condition is not stated.
- Human Factors: "The human factors studies were conducted with the intended user population, use environment and use scenarios to simulate clinical conditions." The number of participants is not specified.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)
This question is not applicable in the context of an infusion pump's performance testing for regulatory clearance. Infusion pump performance is verified against technical specifications and recognized standards, not "ground truth" established by human experts in the diagnostic sense. For example, flow rate accuracy is measured by calibrated instruments, not by an expert's assessment.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
This concept is not applicable. Adjudication methods like 2+1 or 3+1 are used in medical imaging studies to establish accurate labels (ground truth) for complex diagnostic tasks. For an infusion pump, performance is measured objectively against predefined engineering standards.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This question is not applicable. MRMC studies are used to evaluate human reader performance with and without AI assistance for diagnostic tasks. This device is an infusion pump, not a diagnostic AI system, and therefore, such a study would not be performed for its clearance.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
The device is an infusion pump with an "Infusion Management System (IMS)" that includes a drug library and remote monitoring features. The performance tests ("Device performance", "Software & Cybersecurity") implicitly cover the "standalone" performance of the pump's mechanical and software components in delivering fluids accurately. The "Human Factors" study assesses the human-in-the-loop performance, but this is about usability and safety, not diagnostic accuracy improvement.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
As discussed, the concept of "ground truth" for an infusion pump's performance is different from that of a diagnostic algorithm. The "ground truth" or reference for the infusion pump's performance testing consists of:
- Established engineering specifications: E.g., flow rate accuracy within +/- 5%.
- Validated measurement methods and equipment: Use of calibrated sensors and experimental setups to objectively measure parameters like flow rate, pressure, and bolus volume.
- Recognized industry standards: Compliance with standards like IEC 60601 (electrical safety), ISO 10993 (biocompatibility), IEC 62304 (software life cycle), etc.
8. The sample size for the training set
This question is not applicable since the device is an infusion pump, which is not an AI/ML algorithm that requires a "training set" in the conventional sense. The "training" for such a device involves its design, engineering, and manufacturing processes to meet predefined specifications.
9. How the ground truth for the training set was established
Not applicable, as it's not an AI/ML device with a "training set." The "ground truth" for its development is based on engineering principles, clinical requirements for fluid delivery, and regulatory standards.