Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K241091

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-12-23

    (245 days)

    Product Code
    Regulation Number
    868.5665
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SIMEOX 200 airway clearance device is intended to promote airway clearance and to improve bronchial drainage by enhancing mobilization of bronchial secretions via high frequency oscillatory vibrations and intermittent negative pressure to the airway during exhalation. The device is intended to be prescribed for use in patients capable of independently generating cough.

    SIMEOX 200 is intended for use in the home, hospital, and other healthcare institutions by individuals weighing at least 23 Kg.

    Device Description

    The SIMEOX 200 Airway Clearance Device consists of a portable generator and a single patient use expiratory kit. The patient inhales from the ambient environment, naturally and independent of the device, and then exhales into the device mouthpiece. During the patient's exhalation, the SIMEOX 200 generates a pneumatic vibration throughout the airway. This vibration, applied at 6 Hz or 12 Hz via negative pressure pulses, disseminates throughout the bronchial tree in order to decrease mucus viscosity and increase secretion mobility. During or after therapy, the patient can clear the secretions by coughing independently.

    AI/ML Overview

    This document describes the FDA's 510(k) clearance for the SIMEOX 200 Airway Clearance Device. It primarily focuses on demonstrating substantial equivalence to predicate devices through technical specifications and non-clinical performance testing.

    The provided text does not contain information about acceptance criteria or a study proving the device meets those criteria in the context of human-in-the-loop performance, AI assistance, or expert-based ground truth establishment for diagnostic purposes.

    The document states:

    • "Performance verification testing was conducted on the bench with the SIMEOX 200 Airway Clearance Device to demonstrate the subject device performs to its pre-specified requirement specifications, which align with those of the predicate devices."
    • "The SIMEOX 200 device also went through comprehensive software verification and validation testing to confirm the device operates within specification."
    • "Testing demonstrated the SIMEOX 200 performed as intended and was safe and effective."
    • "A human factors validation study was conducted with simulated first-time use of the product in adult users and dyads of children with caregivers. The study concluded that the device is safe for its intended users."

    This indicates that the "acceptance criteria" and "study that proves the device meets the acceptance criteria" in this context refer to engineering and safety performance specifications, rather than clinical performance metrics typically associated with AI/diagnostic devices (like sensitivity, specificity, or reader improvement).

    Therefore, I cannot populate the requested table and study details as they are not present in the provided text. The document describes a medical device (Airway Clearance Device), not an AI or diagnostic device that would typically involve the type of acceptance criteria and studies you've outlined.

    The acceptance criteria implied are the device's ability to meet its pre-specified parameters (e.g., pressure range, oscillation frequencies) and demonstration of safety and effectiveness through bench testing, software validation, and human factors validation.

    Here's a summary of the information that is available, reframing it in the context of device performance rather than AI/diagnostic performance:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Implied from document)Reported Device Performance (from document)
    Technical Specifications:
    Exhalation Pressure (Negative)0 to -60 cm H2O. Stated as "nearly identical to that of SimplyClear and falls within a subset of the SimplyClear available negative pressure range."
    Inhalation Pressure (Positive)Atmospheric (0 cm H2O). Bench verification testing confirms operation within prespecified operating specifications for negative pressure delivery, consistent with predicate devices in exsufflation-only mode.
    Oscillation FrequenciesFixed Frequencies: 12 Hz (first 8 phases), 6 Hz (final 2 phases). Stated as "within the available frequencies during SimplyClear treatment."
    Power ModesStandard Mode (25%, 50%, 75%, 100% power) and ADAPT Mode (automatically regulated). Stated as "operating conditions are within the range available for the SimplyClear."
    Trigger SettingsManual Trigger or DETECT Trigger.
    Cycle Duration10 breaths (exhalations).
    Duration of TreatmentMax 10 cycles (100 exhalations) or 60 minutes, whichever occurs first.
    Safety and Effectiveness:
    BiocompatibilityExpiratory kit accessories not considered cytotoxic, sensitizing, or irritating; demonstrated biocompatible.
    Human FactorsDevice safe for intended users (adults, children with caregivers) in simulated first-time use.
    Software Verification & ValidationConfirmed device operates within specification.
    Cybersecurity, Wireless, EMC, Electrical,Comprehensive testing conducted.
    Mechanical, Thermal Safety
    Overall PerformanceBench testing confirms device complies with specifications, operates equivalently to predicates, is as safe, as effective, and performs as well as legally marketed predicate devices. "Performed as intended and was safe and effective."

    Regarding the other requested information, it is largely not applicable or not provided for this type of device and submission:

    2. Sample size used for the test set and the data provenance:
    * Test Set: Not explicitly stated in terms of a "test set" for performance metrics like sensitivity/specificity. Performance testing was "on the bench" and included software V&V and human factors simulation.
    * Data Provenance: Not applicable for this type of technical/safety validation. The human factors study involved "simulated first-time use," but details about participant numbers or demographic provenance are not provided.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
    * This is not relevant for this device's 510(k) submission, as it's not a diagnostic device requiring expert interpretation for ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
    * Not applicable.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
    * Not applicable, as this is not an AI-assisted diagnostic device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
    * Not applicable, as this is not an AI/algorithm-driven device in the sense of image analysis or diagnostic prediction. Its "performance" is based on its physical and software operation parameters.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
    * "Ground truth" in this context refers to the pre-defined engineering specifications and safety standards (e.g., pressure measurements, frequency measurements against a calibrated standard, successful execution of software functions, successful completion of simulated tasks in human factors). It's not a clinical or diagnostic ground truth.

    8. The sample size for the training set:
    * Not applicable. This device does not use machine learning that requires a "training set."

    9. How the ground truth for the training set was established:
    * Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K132489

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2014-06-20

    (316 days)

    Product Code
    Regulation Number
    868.5440
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Inogen At Home Oxygen Concentrator is used on a prescriptive basis by patients requiring supplemental oxygen. It supplies a high concentration of oxygen and is used with a nasal cannula to channel oxygen from the concentrator to the patient. The Inogen At Home Oxygen Concentrator may be used in a home or institution.

    Device Description

    The Inogen At Home is a stationary concentrator providing approximately 90% oxygen to patients on a continuous flow basis at a rate of 1.0 liter per minute (LPM) to 5.0 liters per minute (LPM), in 1.0 liter per minute (LPM) increments. Nasal cannula channels oxygen from device to patient; both concentrator and nasal cannula are non-sterile. The device is capable of continuous use in the home or in an institutional setting. The Inogen At Home can be powered by mains AC power found in the home or in an institutional setting. The Inogen At Home includes a detachable AC power cord to make the electrical connection to the AC mains. The Inogen At Home utilizes Pressure Swing Adsorption (PSA) technology to produce oxygen. Oxygen is separated from ambient air under pressure by molecular characteristics and affinity for adsorbent material. The molecular sieve preferentially adsorbs nitrogen, allowing enriched oxygen to collect in an accumulator; the process then swings to low pressure to desorb the nitrogen, exhausting it back into the environment. A series of sieve beds, manifold with precision valves, sensors, and embedded software are used collectively to govern system functionality.

    AI/ML Overview

    The provided document describes the Inogen At Home Oxygen Concentrator, a medical device for supplemental oxygen delivery. It includes details about its purpose, technological characteristics, and conformance to relevant standards. However, the document does not contain a study that validates the device against specific acceptance criteria in the format requested.

    Here's an analysis based on the information provided in the document, highlighting what is missing for a complete answer to your request:

    The document clearly states the device's conformance to several international and national standards (EN ISO 8359:2012, ASTM F 1464:2005, IEC 60601-1). It also lists "performance testing" that was done, but does not provide specific acceptance criteria for these tests or the reported performance data in a quantitative table.

    Here's a breakdown of what can and cannot be answered based solely on the provided text:

    1. A table of acceptance criteria and the reported device performance

    • Acceptance Criteria Mentioned (Implicitly by conformance to standards or general tests):

      • Oxygen Concentration: Approximately 90% (This is a design spec, not explicitly an "acceptance criterion" from a study).
      • Flow Rate: 1.0 to 5.0 LPM in 1.0 LPM increments (Design spec).
      • Electrical Safety/Compatibility: Conformance to IEC 60601-1.
      • Software Functionality: Tested.
      • Fire Mitigation: Tested.
      • Noise, Temperature, and Altitude Ranges: Tested.
      • Outlet Pressure: Tested.
      • Flow Indicator Accuracy: Tested.
      • Particulate Matter, Ozone, Volatile Organic Compounds, Carbon Dioxide, Carbon Monoxide: Biocompatibility testing included evaluation of these.
      • Overall safety requirements per EN ISO 8359:2012 and ASTM F 1464:2005.
    • Reported Device Performance: The document states that "Performance testing included the assessment of electrical safety/ compatibility, software functionality, fire mitigation and noise, temperature and altitude ranges, outlet pressure, flow indicator accuracy, and oxygen concentration." It also mentions "Biocompatibility testing included the evaluation of particulate matter, Ozone, Volatile Organic Compounds, Carbon Dioxide, and Carbon Monoxide." However, no specific quantitative results or performance data against numerical acceptance criteria are provided in the text. For example, it doesn't say "oxygen concentration met specific criteria of X% +/- Y%" or "noise levels were Z dB below threshold."

    Therefore, a table cannot be fully constructed with specific quantitative acceptance criteria or reported performance from this document.


    Regarding the study that proves the device meets the acceptance criteria:

    The document describes "Non-clinical Performance Data" and "Performance testing" as well as "Biocompatibility testing." These are the studies mentioned. However, the details required for points 2-9 are not present in the provided text.

    Missing Information:

    • 2. Sample sized used for the test set and the data provenance: Not mentioned.
    • 3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable for this type of device testing. "Ground truth" in this context would be objective measurements from calibrated equipment, not expert consensus interpretation.
    • 4. Adjudication method for the test set: Not applicable.
    • 5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI-assisted diagnostic device.
    • 6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is a physical oxygen concentrator, not an algorithm.
    • 7. The type of ground truth used: For performance tests, the ground truth would be objective measurements from calibrated sensors and analytical equipment (e.g., oxygen analyzers, flow meters, pressure gauges, sound level meters, gas chromatographs for impurities). For biocompatibility, it would be analytical results against established safe limits. This is implied by the testing, but not explicitly detailed.
    • 8. The sample size for the training set: Not applicable. This is a hardware device, not an AI algorithm requiring a training set.
    • 9. How the ground truth for the training set was established: Not applicable.

    Summary:

    The provided text serves as a 510(k) summary for a medical device (oxygen concentrator). It broadly discusses performance and biocompatibility testing and conformance to standards to demonstrate "substantial equivalence" to predicate devices. However, it does not detail a specific quantitative study with acceptance criteria and results in the manner typically expected for AI/diagnostic devices as implied by your questions, which are more tailored to software-based systems or image analysis. The "acceptance criteria" here are largely defined by compliance with the cited medical device standards (EN ISO 8359, ASTM F 1464, IEC 60601-1) and internal design specifications.

    Ask a Question

    Ask a specific question about this device

    K Number
    K032818

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2004-05-13

    (246 days)

    Product Code
    Regulation Number
    868.5440
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Inogen One Oxygen Concentrator is used on a prescriptive basis by patients requiring supplemental oxygen. It supplies a high concentration of oxygen and is used with a nasal cannula to channel oxygen from the concentrator to the patient. The Inogen One Oxygen Concentrator may be used in a home, institution, vehicles and various mobile environments.

    Device Description

    The Inogen One Oxygen Concentrator is used on a prescriptive basis by patients requiring supplemental oxygen. Patients may include but are not restricted to those with chronic obstructive pulmonary disease (COPD). The device is not intended to be life sustaining or to be life supporting. It is used with a nasal cannula to channel oxygen from the device to the patient. The concentrator and the nasal cannula are non-sterile.

    The Inogen One Oxygen Concentrator provides approximately 90% oxygen to the patient on a demand flow basis at an "equivalent" rate of 1.0 liters per minute to 5.0 liters to minute in increments of 0.5 liters per minute.

    The Inogen One Oxygen Concentrator is capable of continuous use in a home, institution, vehicles and various mobile environments. Power options include 110 -- 220 VAC, 12 -- 14 VDC or rechargeable batteries.

    The Inogen One Oxygen Concentrator uses molecular sieve / pressure swing adsorption technology. Ambient air is drawn thru particle filters by a compressor and forced thru molecular sicve beds, which adsorb nitrogen and allow oxygen to pass. The airflow is then changed and nitrogen is desorbed from the molecular sieve, allowing it to adsorb again during the next cycle. Oxygen is collected in an accumulator reservoir. Waste nitrogen is exhausted back into the room. A series of sieve beds, a manifold with precision valves, sensors and embedded software to control the cycle are used to make the system function.

    Oxygen is delivered to the patient on a demand flow basis in precise amounts during the inhalation part of the breathing cycle. This conserver technology eliminates waste of unused oxygen at other times in the breathing cycle when it is not needed. Inogen One Oxygen Concentrator senses the beginning of the inhalation cycle and releases a specified dose of oxygen enriched gas from the accumulator reservoir, thru a final filter, into the connected nasal cannula and onto the patient.

    The design of the Inogen One Oxygen Concentrator has focused on maximizing subsystem efficiencies and miniaturizing components to enable continuous duty use and to provide minimal weight and battery operation for mobile use.

    The basic technology of the Inogen One Oxygen Concentrator is equivalent to other approved oxygen concentrators. The principles of operation are equivalent to the predicate device noted in the submission.

    AI/ML Overview

    The provided text is a 510(k) summary for the Inogen One Oxygen Concentrator. It describes the device, its intended use, and claims substantial equivalence to a predicate device. However, it does not contain any information about a study proving the device meets acceptance criteria, nor does it explicitly list acceptance criteria in a table format with reported device performance.

    Therefore, I cannot fulfill the request to build a table of acceptance criteria and reported device performance or provide details about a study that proves the device meets those criteria, as that information is not present in the given text.

    The document only states that:

    • "Benchtop performance testing has demonstrated that the Inogen One Oxygen Concentrator is equivalent to the AirSep LifeStyle Oxygen Concentrator."

    This is a statement of equivalence based on testing, but the specifics of that testing, the acceptance criteria used, or the results are not detailed in this summary.

    In summary, the requested information is absent from the provided text.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1