510(k) Data Aggregation

    K Number
    K232328

    Validate with FDA (Live)

    Date Cleared
    2024-03-01

    (211 days)

    Product Code
    Regulation Number
    880.5440
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Insignis™ Syringe Infusion System is specifically indicated for the subcutaneous infusion of the following human plasma-derived immunoglobulins when used according to the FDA approved biologic labeling: Cuvitru®, Immune Globulin Infusion (Human) 20% (manufactured by Takeda®); Hizentra®, Immune Globulin Subcutaneous (Human) 20% Liquid (manufactured by CSL Behring®); and Xembify®, Immune Globulin Subcutaneous (Human) 20% Liquid (manufactured by Grifols®) in the home, hospital, or ambulatory settings when administered according to the approved biologic or drug product labeling. The system is intended for use by clinicians and lay users. It is intended for use in adults and in adolescents ages 13-21. For adolescents, it is to be operated under the supervision of an adult after instruction is provided by a healthcare provider.

    The Insignis TM Syringe Infusion System with the Insignis Syringe Driver and Intravenous Controller, is specifically indicated for the intravenous infusion of fluids such as 0.9% Sodium Chloride (normal saline) and dextrose solutions, and the following antibiotics when used according to the FDA approved drug product labeling: vancomycin, ertapenem, meropenem, oxacillin, and tobramycin.

    Device Description

    The Insignis™ Syringe Infusion System is a portable mechanical infusion system that operates without electricity. The Insignis system is intended to deliver indicated fluids subcutaneously or intravenously, depending on the prescribed medicine type, route of administration, and dose. The Insignis system incorporates into its design, a means to maintain a constant nominal pressure (13.5 psi), a controlled flowrate within a restrictive tubing circuit and needles to ensure the desired flowrate of medication to the patient is achieved. The spring-loaded Insignis™ Syringe Driver a constant mechanical force to the plunger of a BD 50 mL syringe (ref. #309653), that dispenses fluid through a restricted flowrate tubing and / or needle set. The Insignis system can infuse the rated BD syringe volume of 0 - 50 mL. Through a selection of IHS products total flowrates of 8 mL/hr -250 mL/hr are available: application specific.

    • A. For subcutaneous immunoglobulin (SCIg) administrations, the Insignis system includes the OneSett™ Subcutaneous Administration Set, which consists of a selectable rate flow controller connected in-line to an administration set comprised of 1 - 4 needle set configurations (6mm, 9mm, 12 mm, and 14mm lengths). The OneSett is dedicated to deliver 20% concentrate SCIg drugs at specific flowrates which meet the drug manufacturer's package inserts recommended flowrate limitations. The selectable rate flow controller controls the flowrate delivered to each subcutaneous needle site.

    • B. For intravenous administrations, the Insignis system includes the Intravenous Controller, which consists of a selectable rate flow controller connected in-line to a tubing set. The Intravenous Controller is dedicated to delivering select antibiotics, and saline and dextrose solutions. Again, the selectable rate flow controller allows for selectable and controllable flowrates.
      For administering the prescribed medications, the Insignis system requires that the user fill the syringe with the prescribed medications, connect the infusion set to the syringe, eliminate air from the fluid line, secure the syringe into the Insignis™ Syringe Driver, connect to the infusion set to the patient, and load the syringe driver to begin the infusion.

    The system is comprised of the Insignis™ Syringe Driver, a syringe, and either intravenous or subcutaneous infusion sets. To elaborate:

    • a. The Insignis™ Syringe Driver provides the constant driving force to dispense fluids.
    • b. The syringe is the reservoir that holds the fluids (manufactured by Becton & Dickinson).
    • c. The infusion sets (intravenous or subcutaneous) form the fluid path and delivers the fluid to the patient at controlled flowrates.
    AI/ML Overview

    This document is a 510(k) Summary for the Insignis™ Syringe Infusion System, a medical device. It focuses on demonstrating substantial equivalence to a predicate device, not on proving clinical efficacy via a multi-reader multi-case (MRMC) study or complex AI-driven ground truth establishment. Therefore, many of the requested points regarding AI model performance, expert adjudication, and training/test set details are not applicable as this is a traditional medical device submission, not an AI/ML product submission.

    The document primarily details the results of non-clinical performance testing (bench testing) to verify the device meets its intended use and design requirements. The "acceptance criteria" can be inferred from the performance data presented, particularly the acceptable flow rate ranges and deviations.

    Here's an analysis of the provided text in relation to your request:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document provides tables of flow rate performance for different configurations and medications. The "acceptance criteria" are implied by the ranges and percentage deviations provided for each labeled flowrate. The reported device performance is these very ranges and deviations. There isn't a separate, explicit table of acceptance criteria before the results, but the results themselves define the product's performance specifications.

    Intravenous Controller (for 0.9% NaCl, Vancomycin, Ertapenem, Meropenem, Oxacillin, Tobramycin):

    Controller Dial Labeled Flowrate (mL/hr)Reported Flowrate Performance (mL/hr)Reported Flowrate Percent Deviation
    KVO* (8ml/hr)3.09 - 13.4-61.4%, +67.5%
    3230.7 - 37.6-4.06%, +17.5%
    6563.5 - 70-2.36%, +7.69%
    130125 - 143-3.85%, +10%
    192187 – 201-2.6%, +4.69%
    255248 - 264-2.75%, +3.53%
    265191 - 355-27.9%, +34%

    *KVO for saline and dextrose only.

    Subcutaneous IgG Infusion - Hizentra® for Primary Immunodeficiency (PI):
    Throughout the infusion, the measured flow rate may vary ±36% from the flow rate labeled on the dial.

    OneSett™ Subcutaneous Administration SetDial Setting - Nominal Flowrate per Infusion Site (mL/hr/site)
    10 (mL/hr/site)
    1 Needle6.45 - 13.6 (±36%)
    2 Needles6.45 - 13.6 (±36%)
    3 Needles6.45 - 13.6 (±36%)
    4 Needles6.45 - 13.6 (±36%)

    *Subsequent infusion after initial infusion.
    = Outside of drug's prescribing information recommended minimum - maximum flowrate.

    Subcutaneous IgG Infusion - Hizentra® for Chronic Inflammatory Demyelinating Polyneuropathy (CIDP):
    Throughout the infusion, the measured flow rate may vary ±36% from the flow rate labeled on the dial.

    OneSett™ Subcutaneous Administration SetDial Setting - Nominal Flowrate per Infusion Site (mL/hr/site)
    10 (mL/hr/site)
    1 Needle6.45 – 13.6 (±36%)
    (Similarly for 2, 3, 4 Needles)(Same as above)

    *Subsequent infusion after initial infusion.
    = Outside of drug's prescribing information recommended minimum - maximum flowrate.

    Subcutaneous IgG Infusion - Xembify® for Primary Immunodeficiency (PI):
    Throughout the infusion, the measured flow rate may vary -48.4%, +9% from the flow rate labeled on the dial.

    OneSett™ Subcutaneous Administration SetDial Setting - Nominal Flowrate per Infusion Site (mL/hr/site)
    10 (mL/hr/site)
    1 Needle5.16 - 10.9 (-48.4%, +9%)
    (Similarly for 2, 3, 4 Needles)(Same as above)

    = Outside of drug's prescribing information recommended minimum - maximum flowrate.

    Subcutaneous IgG Infusion - Cuvitru® for Primary Immunodeficiency (PI):
    Throughout the infusion, the measured flow rate may vary from the flow rate labeled on the dial.

    OneSett™ Subcutaneous Administration SetDial Setting - Nominal Flowrate per Infusion Site (mL/hr/site)
    10 - 15 (mL/hr/site)
    1 NeedleDo not set the flowrate dial below 20 mL/hr/site.
    (Similarly for 2, 3, 4 Needles)(Same as above)

    *Subsequent infusion after first two infusions.
    *Subsequent infusion after first two information recommended minimum - maximum flowrate.

    2. Sample size used for the test set and the data provenance:

    • Sample Size: The document states, "Flow rate was measured by visually assessing the change in volume in a graduated cylinder every 30 seconds." It does not specify the number of individual tests, number of devices, or number of runs for each configuration. It only describes the method of measurement and the resulting performance ranges.
    • Data Provenance: The data is from "bench testing conditions," which generally implies a controlled laboratory environment. The country of origin is not explicitly stated, but the submission is to the U.S. FDA by a U.S.-based company (Innovative Health Sciences, LLC, Chester, NY), implying the testing was likely conducted in the US. The data is prospective, generated specifically for this submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not Applicable. This is a mechanical/fluid dynamics device, not an AI/ML diagnostic device requiring expert interpretation of medical images or data for ground truth. The "ground truth" for flow rate performance is established through direct, objective physical measurements using calibrated equipment (graduated cylinder, timing device, etc.), as described in the "Flowrate Performance" section: "Flow rate was measured by visually assessing the change in volume in a graduated cylinder every 30 seconds, when using the Insignis™ Syringe Infusion System as intended."

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not Applicable. There is no expert adjudication for this type of performance testing.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not Applicable. This is not an AI-assisted diagnostic device. No MRMC study was conducted or is relevant for this type of submission.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • Not Applicable. This is a mechanical infusion system, not an algorithm. Bench testing evaluates the device's standalone physical performance parameters.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The ground truth for flow rate performance is based on direct, objective physical measurements (volume over time) under controlled bench testing conditions, as described.

    8. The sample size for the training set:

    • Not Applicable. This is not an AI/ML device that requires a training set.

    9. How the ground truth for the training set was established:

    • Not Applicable. No training set was used.
    Ask a Question

    Ask a specific question about this device

    K Number
    K202279

    Validate with FDA (Live)

    Date Cleared
    2021-08-13

    (367 days)

    Product Code
    Regulation Number
    880.5440
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Insignis® Subcutaneous Needle Sets are intended for the delivery of medication to the subcutaneous tissue for use not to exceed 24 hours.

    Device Description

    The Insignis® Subcutaneous Needle Sets are sterile, single use, packaged subcutaneous needle sets. Each Subcutaneous Needle Set has a luer lock at one end and a 90-degree 26G (0.46mm) needle mounted to a closing wing-stabilizer (or butterfly wings) at the distal end of one or more needles. The needles are available in four different lengths: 6mm, 9mm, 12mm, and 14mm. Each set has a slide clamp to stop the flow immediately as well as a butterfly wing that can be closed around the needle upon completion of the infusion. The Insignis® Subcutaneous Needle Sets is available as a single set, as well as a two, three and four needles sets.

    The Insignis® Subcutaneous Needle Sets is designed to be used for less than 24 hours

    AI/ML Overview

    This document is a 510(k) Premarket Notification from the FDA regarding the "Insignis Subcutaneous Needle Sets." It is a regulatory document to demonstrate that a new device is substantially equivalent to a legally marketed predicate device, rather than a clinical study evaluating the performance of an AI-powered diagnostic device.

    Therefore, the information requested in your prompt regarding acceptance criteria, study design, sample sizes, expert ground truth adjudication, MRMC studies, standalone algorithm performance, and training set details for an AI/diagnostic device is not present in this document.

    This document describes a medical device (subcutaneous needle sets) and its equivalence to a predicate device based on physical characteristics, intended use, and conformance to various industry standards for manufacturing, materials, sterility, and biocompatibility.

    To directly answer your request based only on the provided text, and noting the mismatch between the prompt's focus on AI/diagnostic acceptance criteria and the document's content:


    Based on the provided FDA 510(k) document (K202279) for Insignis Subcutaneous Needle Sets:

    The document describes the acceptance criteria and proof of device performance in the context of demonstrating substantial equivalence to a predicate device, as required for a 510(k) submission, not for an AI-powered diagnostic device. The performance claimed is related to the physical and functional attributes of the needle sets for medication delivery.

    Here's an attempt to map the prompt's questions to the available information, while explicitly stating where the information is not applicable or present for an AI/diagnostic device:

    1. Table of acceptance criteria and the reported device performance:

    The acceptance criteria are implicitly met by conformance to recognized industry standards and comparison to a predicate device. The "performance" is demonstrating that the device functions as intended and is safe and effective when used as described.

    Acceptance Criteria (Implicitly from Standards & Equivalence)Reported Device Performance (Summary from Document)
    Intended Use Equivalence:Insignis Subcutaneous Needle Sets are intended for the delivery of medication to the subcutaneous tissue for use not to exceed 24 hours. This is "Same" as the predicate device (RMS Subcutaneous Needle Set, K102512).
    Biocompatibility:The device meets ISO 10993-1 and specific parts (5, 6, 10, 11) for biocompatibility (cytotoxicity, local effect after implantation, irritation, skin sensitization, systemic toxicity). Classified as Externally communicating device, Blood Path indirect, Limited Contact (<24 hours).
    Sterility:Achieves SAL 10-6. Sterilization method is E-beam, which is different from the predicate's Gamma but deemed to provide an identical level of sterility and undifferentiated outcome. Conforms to ISO 11137 series.
    Physical/Functional Performance:Conforms to various ISO standards for needle tubing (ISO 9626), Luer fittings (ISO 594-2), sharps injury protection (ISO 23908), hypodermic needles (ISO 7864), and infusion equipment (ISO 8536-9). Specific tests mentioned include: Simulated clinical Use, PKP testing with various fluids, Butterfly opening Force test, Butterfly Low memory Test, Fluid flow rates, Pull force testing.
    Material Safety & Equivalence:Materials (e.g., TPE tubing vs. PVC, PP luer vs. PVC) are stated to allow for biocompatibility and comparable performance. Stainless steel needles are identical to predicate.
    Shelf Life:Validated for 1 year shelf life. (Predicate is 3 years, but within their respective shelf lives, no performance difference is stated). Conforms to ISO 11607-1 for packaging integrity and sterile barrier.
    Single Use:Confirmed as single use, same as predicate.
    Regulatory Compliance:Device is found substantially equivalent to the predicate, implying compliance with general controls provisions of the FD&C Act.

    2. Sample size used for the test set and the data provenance:

    • Sample Size for Test Set: Not explicitly stated as a "test set" in the context of an AI/diagnostic algorithm testing. Performance data is derived from product testing against various ISO standards. The sample size for each specific test (e.g., fluid flow rates, pull force testing, biocompatibility) is not detailed in this summary.
    • Data Provenance: The data originates from performance testing conducted by the manufacturer (Innovative Health Sciences LLC) and its consultants to demonstrate conformance with international standards and substantial equivalence. The country of origin for the data (where the tests were performed) is not specified, but the applicant is based in New York, USA. The studies would be considered prospective as they are conducted specifically for this regulatory submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    • Not Applicable. This document is for a physical medical device (needle sets), not an AI-powered diagnostic device requiring expert interpretation for ground truth establishment. Performance is evaluated against objective engineering and biocompatibility standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not Applicable. There is no expert adjudication process described or needed for this type of device's performance evaluation.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not Applicable. This is not an AI-powered diagnostic device, and no MRMC study was performed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not Applicable. This is not an AI-powered diagnostic device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • Ground Truth: The "ground truth" for this device's performance is established by conformance to established international standards (ISO series) for medical devices. This includes verifiable physical, chemical, and biological properties, as well as functional performance metrics (e.g., flow rates, material strength, sterility levels) measured objectively by laboratory testing. There is no qualitative "ground truth" derived from human interpretation or patient outcomes data described in this 510(k) summary beyond the standards themselves.

    8. The sample size for the training set:

    • Not Applicable. This document describes a physical medical device, not an AI model requiring a training set.

    9. How the ground truth for the training set was established:

    • Not Applicable. This document describes a physical medical device, not an AI model requiring a training set.