510(k) Data Aggregation

    K Number
    K243449

    Validate with FDA (Live)

    Device Name
    INFINITT DPS
    Date Cleared
    2025-12-09

    (397 days)

    Product Code
    Regulation Number
    864.3700
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    INFINITT DPS is a software device intended for viewing and management of whole slide digital images derived from scanned surgical pathology slides prepared from formalin-fixed, paraffin-embedded (FFPE) tissue. It serves as an aid for pathologists to review and interpret these digital images for the purpose of pathology primary diagnosis.

    INFINITT DPS is intended for use with Hamamatsu NanoZoomer S360MD scanner and Barco MDPC-8127 display.

    It is the responsibility of the pathologist to implement appropriate procedures and safeguards that assure the integrity and accuracy of image interpretation when utilizing the INFINITT DPS. The system is not intended for use with frozen sections, cytology, or non-FFPE hematopathology specimens.

    Device Description

    INFINITT DPS software provides a more efficient and convenient alternative to reviewing slides with physical microscopes. The primary users of INFINITT DPS are pathologists in histopathology, as well as technologists who handle specimens. Users can access whole slide images directly through a web browser, with no need to install additional software on the client PC.

    INFINITT DPS, version 1.1, is a web-based software device that enables qualified pathologists to review digital pathology images (WSIs) of formalin-fixed, paraffin-embedded (FFPE) tissue specimens for pathology primary diagnosis. The device operates through a supported web browser without installing any application locally.

    The Operation and Workflow of INFINITT DPS are described below:

    • Data Input
      • The device accepts WSI files in the NDPI file format generated by a Hamamatsu NanoZoomer S360MD Slide Scanner. The WSI files are internally converted into the Digital Imaging and Communications in Medicine (DICOM) file format using the built-in INFINITT DPS Acquisition module.
    • Data Management and Storage
      • The converted images are stored on the INFINITT DPS server. The user is highly recommended to back up the original NDPI files. The device can be integrated with an external LIS (Laboratory Information System)/EMR (Electronic Medical Record) system to automate case registration and retrieval.
    • Image Access and Navigation
      • Users can log in to the device using the INFINITT DPS WebViewer running on a supported web browser (i.e., Chrome, Edge, or Firefox). The stored WSIs can be searched, selected, and opened directly within the INFINITT DPS WebViewer.
    • Image Review and Interpretation
      • INFINITT DPS WebViewer provides the following interactive features:
        • Zoom, pan, and rotate for image navigation
        • Measurement tools for length and area analysis

    The interoperable components of INFINITT DPS and other system specifications are provided in tables 1 – 3 below.:

    Table 1. Interoperable Components for Use with INFINITT DPS

    ComponentsManufacturerModel
    ScannerHamamatsuNanoZoomer S360MD Slide scanner
    DisplayBarcoMDPC-8127

    Table 2. Computer Environment / System for Use with INFINITT DPS

    ComponentRequirement
    Operating SystemWindows 11
    Memory32GB RAM
    ProcessorIntel Core i7-11800H
    Supported BrowsersGoogle Chrome version 90.0 and above
    Microsoft Edge version 90.0 and above
    Mozilla Firefox version 80.0 and above

    Table 3. Server System Requirements

    ComponentWeb ServerDatabase Server
    CPU2.8GHz 12-core or higher2.2GHz 8-core or higher
    Memory64GB or more16GB or more
    Storage2TB or more4TB or more
    Software PlatformsMicrosoft IIS (Internet Information Services) 10 or higherNot Found
    Operating SystemsMicrosoft Windows Server 2016 or higherNot Found
    DatabaseN/AOracle Database 19c or higher
    Network1Gbps1Gbps
    AI/ML Overview

    N/A