510(k) Data Aggregation
(115 days)
IPDmilled Blanks are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations.
All digitally designed CAD/CAM customizations for the IPDmilled Blanks are to be designed and manufactured according to digital dentistry workflow or to be sent to an IPD validated milling center for manufacture. The workflow system integrates multiple components of the digital dentistry workflow: scan files from intra-oral and lab (desktop) scanners, CAD software, CAM software, milling machine and associated tooling and accessories. Compatible Implant Systems: Dentium Co., Ltd Implantium, Friadent Implant Systems, Neoss ProActive Implant, SPI® Dental Implant, Inicell®, BEGO Semados® S-Line, ANKYLOS® C/X Implant System, MIS V3 Conical Connection Dental Implant System, Conical Connection Implants (MIS® C1), MIS C1 Narrow Platform Conical Connection Implant System, MIS C1 Wide Platform Conical Connection Abutments, PRAMA White Implant Systems, Altatec GmbH CAMLOG®/CONELOG® PROGRESSIVE-LINE Implants, Straumann TLX Implant System, Kontact Dental Implant System, ICX-Implant System, Tapered Pro Conical Implant System, Straumann® Tissue Level, Zimmer Tapered Screw-Vent®, Nobel Biocare® Nobel Active®, Straumann® Bone Level Tapered Implants, Neodent Implant System – GM Line, Osstem Implant System, Xpeed AnyRidge Internal Implant System, 3i Osseotite® Certain® Dental Implants, Astra Tech Implant System (Osseospeed®), OsseoSpeed™ Plus, BioHorizons Tapered Internal Implant System, MIS Internal Hex Dental Implant System (MIS® Seven®), Straumann BLX Ø3.5 mm Implants, Straumann® BLX Implant System, ALTATEC Camlog Screwline Implant System, Replace TiUnite Endosseous Implant.
The purpose of this submission is to expand abutment categories of IPD Dental Implant Abutments, which were previously cleared.
IPDmilled Blanks is a dental implant abutment system composed of dental abutments and screws intended to be placed into dental implants to provide support for dental prosthetic restorations. Abutments provide basis for single or multiple tooth prosthetic restorations. They are available in a variety of connection types to enable compatibility with commercially available dental implants systems. IPDmilled Blanks have a pre-manufactured connection interface that fits directly with a pre-specified dental implant. The customized shape of the abutment is intended to be designed and manufactured according to a digital dentistry workflow or to be sent to an IPD validated milling center for manufacture. IPDmilled Blank is delivered non-sterile and the final restoration, including the screw, is intended to be sterilized at the dental clinic before it is placed in the patient.
The IPDmilled Blanks are blistered together with their specific screw. The screws are intended to attach the prosthesis to the dental implant.
The metallic components of the subject abutments and screws are made of titanium alloy conforming to ISO 5832-3 "Implants for surgery – Metallic materials – Part 3: Wrought titanium 6-aluminium 4-vanadium alloy".
The design and manufacturing of the custom abutment will be conducted using a digital dentistry workflow requiring the use of the following equipment and software:
Intraoral Scanner: 3Shape TRIOS A/S Series
Desktop Scanner: 3Shape E Series.
Design Software: 3Shape Abutment Designer Software, K151455.
AbutmentCAD, K193352.
Milling unit: CORiTEC 350i PRO / CORiTEC 350i Loader PRO with MillBox CAM software.
IPDmilled Blank libraries' have built-in design limitations, and the user isn't allowed to exceed these limitations. The custom abutment design limitation specifications are as follows:
Minimum gingival height: 0.5 mm
Maximum gingival height: 6.0 mm
Minimum wall thickness: 0.4 mm
Minimum post height: 4.0 mm
All IPDmilled Blanks are for straight abutments only.
N/A
Ask a specific question about this device
(99 days)
IPD Dental Implant Abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for single or multiple dental prosthetic restorations.
IPD Dental Implant Abutments is a dental implant abutment system composed of dental abutments, screws, as well as other dental abutment accessories, intended to be placed into dental implants to provide support for dental prosthetic restorations.
Abutments provide basis for single or multiple tooth prosthetic restorations. They are available in a variety of connection types to enable compatibility with commercially available dental implants systems.
IPD Dental Implant Abutments includes the following categories of dental abutment designs:
- Titanium base (Interface) abutments (INC3D);
- Multi-Unit abutments (MUA);
- Overdenture Abutments (PSD);
- Temporary Abutments (PP);
- Healing Abutments (TC).
The system also includes the use of the corresponding screws intended to attach the prosthesis to the dental implant. Specifically:
- Ti Screw (TT): Used during restoration fabrication.
- TiN Screw (TTN): Used in finished restorations, with TiN coating.
- TPA Screw (TPA): Used in finished angulated restorations, with TiN coating.
The metallic components of the subject abutments and screws are made of titanium alloy conforming to ISO 5832-3 "Implant for surgery – Metallic materials – Part 3: Wrought titanium 6-aluminium 4-vanadium alloy".
The purpose of this submission is to expand IPD Dental Implant Abutments offerings with:
• New IPD's compatible dental implant systems,
• New angulations available abutment-category specific.
• New in-house TiN coating.
IPD dental implant abutments and screws are compatible with the following commercially available dental implant systems:
(Table 2. Summary of IPD abutments categories with compatibilized OEM Implant/Abutment Systems with specific reference to maximum angulation specifically included in this submission. provided in original text)
Ti Base (Interface) abutments are attached (screw-retained) to the implant/abutment and cemented to the zirconia superstructure.
The Ti Base is a two-piece abutment composed of the titanium component, as the bottom-half, and the zirconia superstructure, as the top-half. It consists of a pre-manufactured prosthetic component in Titanium alloy per ISO 5832-3, as well as the supporting digital library file for FDA-cleared design software (3Shape Abutment Designer™ Software, cleared under K151455) which enables the design of a patient-specific superstructure by the laboratory/clinician and which will be manufactured in FDA-cleared Zirconia (e.g., DD Bio Z, K142987) according to digital dentistry workflow at the point of care, or at a dental laboratory.
The design and fabrication of the zirconia superstructure for Ti Base (Interface) will be conducted using a digital dentistry workflow requiring the use of the following equipment, software and materials:
Scanner: 3D Scanner D850.
Design Software: 3Shape Abutment Designer Software, K151455.
Zirconia Material: DD Bio Z, K142987.
Milling machine/Brand: Dental Concept System Model: DC1 Milling System.
Cement: Multilink® Automix, K123397.
Ti Base (Interface) abutment design parameters for the zirconia superstructure are defined as follows:
Minimum gingival height: 1.5 mm
Minimum wall thickness: 0.43 mm
Minimum post height for single-unit restorations: 4.75 mm (1)
Maximum gingival height: 6.0 mm
Maximum angulation of the final abutment 30° (2)
The resulting final prosthetic restoration is screwed to the dental implant. All subject abutments are single-use and provided non-sterile. Final restoration (which includes the corresponding screw) is intended to be sterilized at the dental clinic before it is placed in the patient.
The provided FDA 510(k) clearance letter pertains to IPD Dental Implant Abutments, a medical device, not an AI/ML-driven software product. Therefore, the information requested regarding acceptance criteria and study data for an AI/ML device (e.g., sample size for test/training sets, expert ground truthing, MRMC studies, standalone performance) is not applicable to this document.
The document describes the device, its intended use, comparison to predicate devices, and the non-clinical performance testing conducted to demonstrate substantial equivalence. These tests are physical and chemical in nature, not related to the performance of an AI/ML algorithm.
Here's a breakdown of why an AI/ML-focused response is not possible, based on the provided text:
- Device Type: The device is "IPD Dental Implant Abutments," which are physical components used in dentistry (titanium alloy abutments, screws, designed for zirconia superstructures). It is not software, a diagnostic imaging tool, or an AI/ML algorithm.
- Purpose of Submission: The submission aims to expand compatibility with new dental implant systems and include new angulations and in-house TiN coating. This is a modification of a physical medical device, not a new AI/ML development.
- Performance Data (Section VII): This section explicitly lists non-clinical performance testing such as:
- Sterilization validation (ISO 17665-1)
- Biocompatibility testing (Cytotoxicity, Sensitization, Irritation per ISO 10993)
- Reverse engineering and dimensional analysis for compatibility
- Validation of the digital workflow and software system (but this refers to the CAD/CAM software used to design the physical abutments, not an AI/ML diagnostic tool)
- Static and dynamic fatigue testing (ISO 14801)
- Modified Surfaces Information
- MRI safety review
Conclusion:
The provided document describes a 510(k) clearance for a physical dental implant component. It does not contain any information about the acceptance criteria or study design for an AI/ML driven medical device. Therefore, a table of acceptance criteria and reported device performance related to AI/ML, sample sizes for test/training sets, details on expert ground truthing, MRMC studies, or standalone performance of an algorithm cannot be extracted from this text.
Ask a specific question about this device
(92 days)
IPD Dental Implant Abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for single or multiple dental prosthetic restorations.
IPD Dental Implant Abutments is a dental implant abutment system composed of dental abutments and screws intended to be placed into dental implants to provide support for dental prosthetic restorations. Abutments provide basis for single or multiple tooth prosthetic restorations. They are available in a variety of connection types to enable compatibility with commercially available dental implants svstems. IPD Dental Implant Abutments includes the following categories of dental abutment designs: -Titanium base (Interface) abutments; -Multi-unit abutments; -Overdenture abutments (PSD). The system also includes the corresponding Titanium Screws, which were included in previous premarket notifications. The devices are intended to attach the prosthesis to the dental implant. Specifically: - Ti Screw: Used during restoration fabrication. - TiN Screw: Used in finished restorations, with TiN coating. - TPA Screw: Used in finished angulated restorations, with TiN coating. The metallic components of the subject abutments and screws are made of titanium alloy conforming to ISO 5832-3 "Implants for surgery – Metallic materials – Part 3: Wrought titanium 6-aluminium 4-vanadium alloy". The purpose of this submission is to expand IPD Dental Implant Abutments cleared under K240570: . To expand IPD's abutment designs (Multi-unit and Overdenture abutments), as well as to expand the range of angulations available for certain compatible Ti Bases (Interfaces). IPD dental implant abutments and screws are compatible with the following commercially available dental implant systems: (List of compatible implant systems provided in tables). The zirconia superstructures for use with the Ti Base (Interface) are only intended to be designed and manufactured according to digital dentistry workflow system integrates multiple components of the digital dentistry workflow: scan files from intra-oral scanners, CAD software, CAM software, ceramic material, milling machine and associated tooling and accessories. Ti Base (Interface) abutments are attached (screw-retained) to the implant and cemented to the zirconia superstructure. The Ti Base is a two-piece abutment composed of the titanium component, as the bottom-half, and the zirconia superstructure, as the top-half. It consists of a pre-manufactured prosthetic component in Titanium alloy per ISO 5832-3, as well as the supporting digital library file for FDA-cleared design software (3Shape Abutment Designer™ Software, cleared under K151455) which enables the design of a patient-specific superstructure by the laboratory/clinician and which will be manufactured in FDA-cleared Zirconia (e.g., DD Bio Z, K142987) according to digital dentistry workflow at the point of care, or at a dental laboratory. The design and fabrication of the zirconia superstructure for Ti Base (Interface) will be conducted using a digital dentistry workflow requiring the use of the following equipment, software and materials: Scanner: 3D Scanner D850. Design Software: 3Shape Abutment Designer Software, K151455. Zirconia Material: DD Bio Z, K142987. Milling machine/Brand: Dental Concept System Model: DC1 Milling System Cement: Multilink® Automix, K123397. Ti Base (Interface) abutment design parameters for the zirconia superstructure are defined as follows: Minimum gingival height: 1.5 mm Minimum wall thickness: 0.43 mm Minimum post height for single-unit restorations: 4.75 mm (1) Maximum gingival height: 6.0 mm Maximum angulation of the final abutment 30° (2) Note 1: Post height is the length above the abutment collar. Note 2: In Table 2 it has been specifically referred to the angulation depending on the dental implant system and platform. The resulting final prosthetic restoration is screwed to the dental implant. All subject abutments are single-use and provided non-sterile. Final restoration (which includes the corresponding screw) is intended to be sterilized at the dental clinic before it is placed in the patient.
The provided text is a 510(k) summary for a dental implant abutment system, primarily focusing on proving substantial equivalence to previously cleared predicate devices. It does not present a study with acceptance criteria and reported device performance in the manner typically expected for AI/Machine Learning (AI/ML) medical devices, or a multi-reader multi-case (MRMC) comparative effectiveness study.
The "device" in this context is a physical dental implant abutment, not an AI/ML algorithm. Therefore, the questions related to AI/ML specific criteria (such as training sets, ground truth establishment for AI, human reader improvement with AI assistance, or standalone algorithm performance) are not applicable to this submission.
However, I can extract the relevant information regarding the acceptance criteria and the "study" (bench testing in this case) that proves the device meets those criteria, as described in the provided document.
Acceptance Criteria and Device Performance for IPD Dental Implant Abutments
This submission is for a physical medical device, IPD Dental Implant Abutments, and demonstrates substantial equivalence to predicate devices through a combination of design characteristics and non-clinical performance (bench) testing. The "acceptance criteria" are implied by the standards and guidance associated with dental implants, and the "performance" is demonstrated through compliance with these standards.
1. Table of Acceptance Criteria and Reported Device Performance:
Since this is a physical device and not an AI/ML algorithm, the "acceptance criteria" are derived from relevant ISO standards and FDA guidance for endosseous dental implant abutments. The "reported device performance" is the demonstration of compliance with these standards through specific non-clinical tests.
| Acceptance Criteria (Implied by Standards/Guidance) | Reported Device Performance (Compliance Demonstrated Via) |
|---|---|
| Biocompatibility: Device materials are non-cytotoxic, non-sensitizing, non-irritating. | Demonstrated via: Cytotoxicity testing (ISO 10993-5), Sensitization (ISO 10993-10), and Irritation (ISO 10993-23) testing. All patient-contacting surfaces were non-cytotoxic. |
| Sterilization Efficacy: Achieves a Sterility Assurance Level (SAL) of 1 x 10^-6. | Demonstrated via: Sterilization validation according to ISO 17665-1. |
| Mechanical Integrity (Static & Dynamic Fatigue): Device withstands expected chewing forces without failure. | Demonstrated via: Static and dynamic fatigue testing of worst-case implant/abutment configurations and combinations in accordance with ISO 14801. |
| Dimensional Compatibility: Abutments are compatible with listed implant systems. | Demonstrated via: Reverse engineering and dimensional analysis of original manufacturer's components (implants, abutments, screws) to confirm compatibility. |
| Digital Workflow Validation: Design and manufacturing of the superstructure are within specified parameters. | Demonstrated via: Validation of the digital workflow and software system. |
| MRI Safety: Safe for use in an MRI environment. | Demonstrated via: Non-clinical worst-case MRI review using scientific rationale and published literature, addressing magnetically induced displacement force and torque. |
2. Sample size used for the test set and the data provenance:
- Sample Size: The document does not specify exact numerical sample sizes for each bench test conducted (e.g., number of abutments tested for fatigue). It generally refers to "worst-case configurations and combinations" being tested. For biocompatibility, it states "all patient-contacting surfaces" were tested.
- Data Provenance: The tests are non-clinical bench tests, not patient data trials. The studies were performed to support a 510(k) submission to the FDA for a device manufactured by IMPLANT PROTESIS DENTAL 2004, S.L., located in Premià de Dalt (Barcelona), Spain. The data is thus internal company testing. It is inherently "prospective" in the sense that the tests were designed and conducted to meet regulatory requirements for this submission.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable in the context of an AI/ML device. For this physical device, "ground truth" is established by relevant international standards (e.g., ISO 14801, ISO 10993, ISO 17665-1) and internal design specifications, verified through engineering bench tests rather than expert interpretation of images/data for AI. Expertise would come from engineers, material scientists, and quality assurance professionals performing and reviewing the tests. The document does not specify the number or specific qualifications of individuals involved in the execution or review of these bench tests, beyond implying adherence to recognized standards.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Not applicable. Adjudication methods like 2+1 or 3+1 are typically for clinical studies or radiology readings where multiple experts independently review data and discrepancies are resolved. For bench testing of a physical device, the "adjudication" is compliance with predetermined test protocols and acceptance criteria outlined in the relevant ISO standards and internal specifications, often involving quantitative measurements.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No MRMC study was done, and it is not applicable for this physical device. This type of study is relevant for AI/ML diagnostic or assistive devices, not for dental implant abutments. The submission explicitly states: "No clinical testing was performed, the determination of substantial equivalence is supported by non-clinical testing."
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not applicable. This question pertains to AI/ML algorithms. The device is a physical dental implant abutment, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- For this physical device, the "ground truth" for performance evaluation is defined by established engineering standards (e.g., ISO 14801 for mechanical fatigue, ISO 10993 for biocompatibility) and validated design specifications, rather than clinical outcomes or expert consensus on clinical data. Compliance with these quantitative and qualitative standards serves as the ground truth.
8. The sample size for the training set:
- Not applicable. This question refers to AI/ML models. The "training" for this device would refer to its design and manufacturing processes, not a data set for machine learning.
9. How the ground truth for the training set was established:
- Not applicable. As above, this pertains to AI/ML models. For a physical medical device, the design is based on engineering principles and regulatory requirements, not a "training set ground truth" in the AI sense. The "ground truth" for design validation comes from international standards, material properties, and biomechanical principles.
Ask a specific question about this device
(99 days)
IPD Dental Implant Abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for single or multiple dental prosthetic restorations.
IPD Dental Implant Abutments is a dental implant abutment system composed of dental abutments and screws intended to be placed into dental implants to provide support for dental prosthetic restorations. Abutments provide basis for single or multiple tooth prosthetic restorations. They are available in a variety of connection types to enable compatibility with commercially available dental implants systems. IPD Dental Implant Abutments includes the following categories of dental abutment designs: Healing abutments; Temporary abutments; Cementing titanium abutments; Titanium base (interface) abutments; The system also includes the corresponding Titanium Screws intended to attach the prosthesis to the dental implant.
This is a Premarket Notification (510(k)) summary for the IPD Dental Implant Abutments. This document does not describe a study proving the device meets acceptance criteria as would be found in a clinical trial for a novel AI device with specific performance metrics. Instead, it demonstrates substantial equivalence to predicate devices based on bench testing and engineering analysis.
Here's an breakdown of the information provided, tailored to your request, but emphasizing that this is not a typical AI/ML performance study:
1. Table of Acceptance Criteria and Reported Device Performance
Since this is a 510(k) for dental abutments, the "acceptance criteria" are not reported as specific performance metrics (like sensitivity/specificity for an AI model), but rather as compliance with recognized standards and demonstration of substantial equivalence to predicate devices. The "reported device performance" refers to successful completion of various non-clinical tests.
| Acceptance Criteria (Demonstrated Equivalence/Compliance) | Reported Device Performance (Test Results) |
|---|---|
| Same intended use as predicate device | Intended use is identical |
| Same operating principle as predicate device | Operating principle is identical |
| Similar design and same device categories | Very similar design, same categories |
| Same materials and surface coatings | Identical materials and surface coatings |
| Same sterilization processes | Identical sterilization processes |
| Sterilization validation (SAL of 1 x 10^-6) | Met ISO 17665-1 requirements |
| Biocompatibility: Cytotoxicity | Non-cytotoxic (ISO 10993-5 met) |
| Biocompatibility: Sensitization (TiN coated) | Met ISO 10993-10 requirements |
| Biocompatibility: Irritation (TiN coated) | Met ISO 10993-23 requirements |
| Mechanical compatibility (with OEM implants) | Confirmed by reverse engineering/dimensional analysis |
| Validation of digital workflow and software system | Ensured design/manufacturing within parameters |
| Static and dynamic fatigue testing | Met ISO 14801 requirements (worst-case configurations) |
| MRI safety (magnetically induced displacement force/torque) | Reviewed using scientific rationale and published literature, parameters addressed per FDA Guidance |
2. Sample size used for the test set and the data provenance
- Sample Size for Test Set: Not applicable in the context of human data or AI model test sets. The "test set" here refers to the physical samples of the IPD Dental Implant Abutments (and their components, potentially in various configurations) that were subjected to the specified bench and biocompatibility tests. The exact number of physical abutments, screws, or zirconia superstructures tested for each specific criterion is not detailed in this summary but would be found in the full test reports.
- Data Provenance: Not applicable in the context of country of origin for patient data. The tests were performed on the manufactured device components. The materials are specified (e.g., Titanium alloy conforming to ISO 5832-3).
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. "Ground truth" in this context would generally refer to expert-derived labels for clinical data, which is not part of this 510(k) submission. The "truth" for the performance tests is established by adherence to recognized international standards (e.g., ISO, FDA Guidance) and objective measurements from engineering or biological tests.
4. Adjudication method for the test set
Not applicable. Adjudication methods (like 2+1 or 3+1) are used for resolving disagreements among human readers or expert labelers of clinical data, which is not relevant here. The evaluation criteria for the bench tests are objective and defined by established standards.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No. This device is a dental implant abutment, not an AI software. Therefore, MRMC studies and AI assistance effect sizes are not relevant to this submission.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
No. This is a physical medical device (dental abutments), not a standalone algorithm.
7. The type of ground truth used
The "ground truth" for this device's performance is established by:
- Compliance with international standards (e.g., ISO 14801 for fatigue testing, ISO 10993 series for biocompatibility, ISO 17665-1 for sterilization).
- Engineering specifications and dimensional analysis to ensure compatibility with other specified dental implant systems.
- Validated digital dentistry workflow parameters (e.g., minimum gingival height, wall thickness, post height, angulation for zirconia superstructures), which are based on established dental prosthetic principles.
8. The sample size for the training set
Not applicable. This is not an AI/ML device that requires a training set.
9. How the ground truth for the training set was established
Not applicable. No training set is used for this physical device.