510(k) Data Aggregation

    K Number
    K230085

    Validate with FDA (Live)

    Device Name
    Lung-CAD
    Date Cleared
    2023-10-03

    (265 days)

    Product Code
    Regulation Number
    892.2070
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Lung-CAD is a computer-assisted detection (CADe) software device that analyzes chest radiograph studies for lung hyperinflation. The device uses a deep learning algorithm to identify regions of interest (ROIs) with lung hyperinflation and produces boxes around the ROIs.

    Lung-CAD is intended for use as a concurrent reading aid for physicians interpreting chest X-rays. The device is not intended for clinical diagnosis of any disease. It does not replace the role of other diagnostic testing in the standard of care for lung parenchymal findings. Lung-CAD is indicated for adults only.

    Device Description

    Lung-CAD is computer-assisted detection (CADe) software designed to increase the accurate detection of lung hyperinflation. Lung-CAD's output is available for physicians interpreting chest radiographs as a concurrent reading aid. The device helps physicians more effectively identify lung hyperinflation. Lung-CAD does not replace the role of the physician or of other diagnostic testing in the standard of care and does not provide a diagnosis for any disease. Lung-CAD uses modern deep learning and computer vision techniques to analyze chest radiographs.

    For each image within a study, Lung-CAD generates a DICOM Presentation State file (output overlay). If any region of interest (ROI) is detected by Lung-CAD in the study, the output overlay for each image includes "Lung hyperinflation". In addition, if ROI(s) are detected in an image, bounding boxes surrounding each detected ROI are included in the output overlay for that image and are labeled with the radiographic findings: "Lung hyperinflation". If no ROI is detected by Lung-CAD in the study, the output overlay for each image will include the text "No Lung-CAD ROI(s)" and no bounding boxes will be included. Regardless of whether an ROI is detected, the overlay includes text identifying the X-ray study as analyzed by Lung-CAD and a customer configurable message containing a link or instructions, for users, to access labeling documents. The Lung-CAD overlay can be toggled on or off by the physician within their Picture Archiving and Communication System (PACS) viewer, allowing for concurrent review of the X-ray study.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided FDA 510(k) summary for Lung-CAD:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implicitly defined by the reported performance metrics that demonstrate substantial equivalence and effectiveness. While explicit "acceptance criteria" are not listed as pass/fail thresholds in this summary, the strong statistical significance and high performance metrics indicate successful validation.

    Performance MetricAcceptance Criteria (Implicit)Reported Device Performance
    Standalone Performance
    Sensitivity (Lung-CAD)High (e.g., above certain threshold)0.898 (95% CI: 0.856, 0.929)
    Specificity (Lung-CAD)High (e.g., above certain threshold)0.894 (95% CI: 0.885, 0.902)
    AUC (Lung-CAD)High (e.g., close to 1.0)0.964 (95% Bootstrap CI: 0.956, 0.972)
    Reader Study (MRMC) Performance
    Reader AUC Improvement (Aided vs. Unaided)Statistically significant improvement0.0632 (95% CI: 0.0632, 0.0633)
    Statistical Significance (Aided vs. Unaided)p-value < 0.05p-value < 0.001

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: 5,000 chest radiograph cases were used for the standalone performance assessment. For the MRMC reader study, clinical readers each evaluated 244 cases.
    • Data Provenance: Not explicitly stated regarding country of origin. The summary mentions cases "representative of the intended use population," which typically implies a diverse patient population relevant to the device's market. The studies were retrospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not explicitly state the number or qualifications of experts used to establish the ground truth for the test set. However, for such studies, it is standard practice to have multiple qualified radiologists involved in establishing ground truth.

    4. Adjudication Method for the Test Set

    The adjudication method for establishing ground truth is not explicitly described in the provided text. For ground truth establishment in medical imaging AI, common methods include consensus reading by multiple experts (e.g., 2+1, where two experts agree, or a third adjudicates disagreement) or a single expert read with secondary review.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Yes, a fully-crossed MRMC retrospective reader study was done.

    • Effect Size of Improvement: The reader AUC improvement for lung hyperinflation was 0.0632 (95% Confidence Interval: 0.0632, 0.0633) when aided by Lung-CAD versus unaided. This represents the absolute improvement in the Area Under the Curve of the Receiver Operating Characteristic curve for readers when assisted by the AI.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, a standalone performance assessment was conducted on 5,000 chest radiograph cases.

    7. The Type of Ground Truth Used

    The document does not explicitly state the specific method used to establish ground truth (e.g., expert consensus, pathology, outcomes data). However, considering the device's function (detecting lung hyperinflation on chest radiographs) and the context of FDA clearance, the ground truth would most likely be based on:
    * Expert Consensus: A panel of highly qualified radiologists providing a definitive determination of lung hyperinflation for each case.
    * Potentially correlated with clinical findings or other diagnostic tests if available and deemed robust enough for ground truth.

    8. The Sample Size for the Training Set

    The document does not provide the sample size directly for the training set. It only mentions the test set sizes.

    9. How the Ground Truth for the Training Set was Established

    The document does not describe how the ground truth for the training set was established. However, for deep learning models, particularly in medical imaging, the ground truth for the training set is typically established through a rigorous process involving expert annotation (e.g., by radiologists) based on their clinical expertise, often guided by established diagnostic criteria or reference standards.

    Ask a Question

    Ask a specific question about this device

    K Number
    K223811

    Validate with FDA (Live)

    Device Name
    Lung-CAD
    Date Cleared
    2023-09-13

    (267 days)

    Product Code
    Regulation Number
    892.2070
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Lung-CAD is a computer-assisted detection (CADe) software device that analyzes chest radiograph studies for interstitial thickening. The device uses a deep learning algorithm to identify regions of interstital thickening and produces boxes around the ROIs.

    Lung-CAD is intended for use as a concurrent reading aid for physicians interpreting chest X-rays. The device is not intended for clinical diagnosis of any disease. It does not replace the role of other diagnostic testing in the standard of care for lung parenchymal findings. Lung-CAD is indicated for adults only.

    Device Description

    Lung-CAD is computer-assisted detection (CADe) software designed to increase the accurate detection of interstitial thickening. Lung-CAD's output is available for physicians interpreting chest radiographs as a concurrent reading aid. The device helps physicians more effectively identify interstitial thickening. Lung-CAD does not replace the role of the physician or of other diagnostic testing in the standard of care and does not provide a diagnosis for any disease. Lung-CAD uses modern deep learning and computer vision techniques to analyze chest radiographs.

    For each image within a study, Lung-CAD generates a DICOM Presentation State file (output overlay). If any ROI is detected by Lung-CAD in the study, the output overlay for each image includes "Interstitial thickening". In addition, if ROI(s) are detected in an image, bounding boxes surrounding each detected ROI are included in the output overlay for that image and are labeled with the radiographic finding: "Interstitial thickening". If no ROI is detected by Lung-CAD in the study, the output overlay for each image will include the text "No Lung-CAD ROI(s)" and no bounding boxes will be included. Regardless of whether an ROI is detected, the overlay includes text identifying the X-ray study as analyzed by Lung-CAD and a customer configurable message containing a link or instructions, for users, to access labeling documents. The Lung-CAD overlay can be toggled on or off by the physician within their Picture Archiving and Communication System (PACS) viewer, allowing for concurrent review of the X-ray study.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided text:

    Acceptance Criteria and Reported Device Performance

    CriteriaReported Device Performance (Lung-CAD)
    Standalone Performance
    Sensitivity0.913 (95% Wilson's CI: 0.850-0.951)
    Specificity0.866 (95% Wilson's CI: 0.856-0.875)
    Area Under the Curve (AUC) of ROC curve0.961 (95% Bootstrap CI: 0.948-0.972)
    Clinical Performance (Reader Study)
    Reader AUC improvement (Aided vs. Unaided)0.0797 (95% Confidence Interval: 0.0797, 0.0798); statistically significant (p-value < 0.001)

    Study Details

    2. Sample Size and Data Provenance for Test Set

    • Sample Size: 5,000 chest radiograph cases for the standalone performance assessment.
    • Data Provenance: The text states the cases were "representative of the intended use population" but does not specify country of origin, or if they were retrospective or prospective.

    3. Number of Experts and Qualifications for Ground Truth (Test Set)

    • The document does not explicitly state the number of experts used to establish the ground truth for the standalone or clinical test sets, nor their specific qualifications (e.g., radiologists with X years of experience).

    4. Adjudication Method (Test Set)

    • The document does not explicitly state the adjudication method used for establishing the ground truth for the test set.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • Yes, an MRMC retrospective reader study was performed.
    • Effect Size of Human Readers Improvement with AI vs. without AI assistance: The reader AUC improvement for interstitial thickening when aided by Lung-CAD versus unaided was 0.0797 (95% Confidence Interval: 0.0797, 0.0798). This improvement was statistically significant (p-value < 0.001).

    6. Standalone Performance Study

    • Yes, a standalone performance assessment was conducted.
    • Results:
      • Sensitivity: 0.913 (95% Wilson's CI: 0.850-0.951)
      • Specificity: 0.866 (95% Wilson's CI: 0.856-0.875)
      • AUC: 0.961 (95% Bootstrap CI: 0.948-0.972)

    7. Type of Ground Truth Used

    • The document implies the ground truth for both standalone and clinical studies was based on expert interpretation, as it refers to "Ground Truth Positive" cases and "accuracy of readers." However, it does not explicitly state the method (e.g., expert consensus, pathology, outcome data).

    8. Sample Size for Training Set

    • The document does not specify the sample size used for the training set.

    9. How Ground Truth for Training Set was Established

    • The document does not specify how the ground truth for the training set was established. It only mentions the use of "Supervised Deep Learning," which implies that the training data was labeled.
    Ask a Question

    Ask a specific question about this device

    K Number
    K213353

    Validate with FDA (Live)

    Device Name
    Aorta-CAD
    Date Cleared
    2022-09-20

    (347 days)

    Product Code
    Regulation Number
    892.2070
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Aorta-CAD is a computer-assisted detection (CADe) software device that analyzes chest radiograph studies for suspicious regions of interest (ROIs). The device uses a deep learning algorithm to identify ROIs and produces boxes around the ROls. The boxes are labeled with one of the following radiographic findings: Aortic calcification or Dilated aorta.

    Aorta-CAD is intended for use as a concurrent reading aid for physicians looking for ROIs with radiographic findings suggestive of Aortic Atherosclerosis or Aortic Ectasia. It does not replace the role of the physician or of other diagnosic testing in the standard of care. Aorta-CAD is indicated for adults only.

    Device Description

    Aorta-CAD is computer-assisted detection (CADe) software designed for physicians to increase the accurate detection of findings on chest radiographs that are suggestive of chronic conditions in the aorta. The ROIs are labeled with one of the following radiographic findings: Aortic calcification or Dilated aorta. Aorta-CAD is intended for use as a concurrent reading aid for physicians looking for suspicious ROIs with radiographic findings suggestive of Aortic Atherosclerosis or Aortic Ectasia. Aorta-CAD's output is available for physicians as a concurrent reading aid and does not replace the role of the physician or of other diagnostic testing in the standard of care for the distinct conditions. Aorta-CAD uses modern deep learning and computer vision techniques to analyze chest radiographs.

    For each image within a study, Aorta-CAD generates a DICOM Presentation State file (output overlay). If any ROI is detected by Aorta-CAD in the study, the output overlay for each image includes which radiographic finding(s) were identified and what chronic condition in the aorta is suggested by these findings, such as "Aortic calcification suggestive of Aortic Atherosclerosis." In addition, if ROI(s) are detected in an image, bounding boxes surrounding each detected ROI are included in the output overlay for that image and are labeled with the radiographic findings, such as "Aortic calcification". If no ROI is detected by Aorta-CAD in the study, the output overlay for each image will include the text "No Aorta-CAD ROI(s)" and no bounding boxes will be included. Regardless of whether an ROI is detected, the overlay includes text identifying the X-ray study as analyzed by Aorta-CAD and a customer configurable message containing a link to our instructions for users to access labeling documents. The Aorta-CAD overlay can be toggled on or off by the physician within their Picture Archiving and Communication System (PACS) viewer, allowing for concurrent review of the X-ray study.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document doesn't explicitly list "acceptance criteria" in a separate section with pass/fail thresholds. Instead, it describes "performance assessment" and "clinical study" results which implicitly serve as the demonstration that the device performs acceptably and is substantially equivalent. The key performance metrics are presented from the standalone and clinical studies.

    Metric (Implicit Acceptance Criteria)Reported Device Performance (Standalone Study)Reported Device Performance (Clinical Study - Aided vs. Unaided)
    Overall Standalone Performance
    Sensitivity0.910 (95% CI: 0.896, 0.922)Not directly comparable (MRMC study focuses on reader improvement, 0.910 refers to algorithm only)
    Specificity0.896 (95% CI: 0.889, 0.902)Not directly comparable (MRMC study focuses on reader improvement, 0.896 refers to algorithm only)
    AUC (Overall)0.974 (95% Bootstrap CI: 0.971, 0.977)Not directly comparable (MRMC study focuses on reader improvement, 0.974 refers to algorithm only)
    Category-Specific Standalone Performance
    Aortic calcification suggestive of Aortic Atherosclerosis (AUC)0.972 (95% Bootstrap CI: 0.967, 0.976)Reader AUC estimates significantly improved (p < 0.001)
    Aortic calcification suggestive of Aortic Atherosclerosis (Sensitivity)0.922 (95% Wilson's CI: 0.908, 0.934)Not directly reported for Aided vs. Unaided readers.
    Aortic calcification suggestive of Aortic Atherosclerosis (Specificity)0.894 (95% Wilson's CI: 0.883, 0.904)Not directly reported for Aided vs. Unaided readers.
    Dilated aorta suggestive of Aortic Ectasia (AUC)0.948 (95% Bootstrap CI: 0.939, 0.957)Reader AUC estimates significantly improved (p < 0.001)
    Dilated aorta suggestive of Aortic Ectasia (Sensitivity)0.830 (95% Wilson's CI: 0.778, 0.872)Not directly reported for Aided vs. Unaided readers.
    Dilated aorta suggestive of Aortic Ectasia (Specificity)0.897 (95% Wilson's CI: 0.888, 0.906)Not directly reported for Aided vs. Unaided readers.
    Impact on Reader PerformanceN/A (Standalone study measures algorithm performance alone)Primary Objective Met: Accuracy of readers aided by Aorta-CAD ("Aided") was superior to the accuracy of readers when unaided by Aorta-CAD ("Unaided") per category as determined by the case-level AUC of the ROC curve.

    2. Sample Size for the Test Set and Data Provenance

    • Standalone Test Set Sample Size: 5,000 chest radiograph cases.
    • Clinical (MRMC) Test Set Sample Size: 244 cases (each reader evaluated these under both aided and unaided conditions).
    • Data Provenance: The document states "cases representative of the intended use population" for the standalone study. For the MRMC study, it also doesn't explicitly state the country of origin but implies a general patient population suitable for chest radiographs. Both studies are explicitly stated as retrospective.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    • Number of Experts: A "panel" of U.S. board-certified radiologists. The exact number is not specified beyond "a panel."
    • Qualifications of Experts: U.S. board-certified radiologists. No specific experience length (e.g., 10 years) is mentioned.

    4. Adjudication Method for the Test Set Ground Truth

    • The ground truth for the MRMC study cases was established by "a panel of U.S. board-certified radiologists who assigned a ground truth binary label indicating the presence of Aortic calcification suggestive of Aortic Atherosclerosis and Dilated aorta suggestive of Aortic Ectasia."
    • This suggests a consensus or majority vote approach, but the specific adjudication method (e.g., 2+1, 3+1, etc.) is not explicitly detailed. It only mentions "a binary label."

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study and Effect Size

    • Was an MRMC study done? Yes, a "fully-crossed multiple reader, multiple case (MRMC) retrospective reader study" was conducted.
    • Effect Size of Human Reader Improvement with AI vs. Without AI Assistance:
      • Reader AUC improvement for Aortic calcification suggestive of Aortic Atherosclerosis: 0.0984 (95% Confidence Interval: 0.0984, 0.0985).
      • Reader AUC improvement for Dilated aorta suggestive of Aortic Ectasia: 0.0885 (95% Confidence Interval: 0.0885, 0.0886).
      • The study demonstrated that "Reader AUC estimates significantly improved for both categories (p-values < 0.001)."

    6. Standalone (Algorithm Only) Performance Study

    • Was a standalone study done? Yes. "Imagen conducted a standalone performance assessment on 5,000 chest radiograph cases..."
    • Performance:
      • Sensitivity: 0.910 (0.896, 0.922)
      • Specificity: 0.896 (0.889, 0.902)
      • Overall AUC: 0.974 (0.971, 0.977)
      • Category-specific AUCs:
        • Aortic calcification: 0.972 (0.967, 0.976)
        • Dilated aorta: 0.948 (0.939, 0.957)
      • Category-specific Sensitivity/Specificity (Table 3 on page 9)

    7. Type of Ground Truth Used

    • The ground truth for both the standalone and MRMC studies was established by expert consensus (a "panel of U.S. board-certified radiologists"). They assigned a "binary label" (presence/absence). There is no mention of pathology or outcomes data being used for ground truth.

    8. Sample Size for the Training Set

    • The document does not provide the sample size used for the training set. It only mentions that the device uses "modern deep learning and computer vision techniques" and "Supervised Deep Learning."

    9. How the Ground Truth for the Training Set Was Established

    • The document does not specify how the ground truth for the training set was established. It only states that the device uses "Supervised Deep Learning," which implies a labeled dataset, but the method of labeling is not described.
    Ask a Question

    Ask a specific question about this device

    K Number
    K210666

    Validate with FDA (Live)

    Device Name
    Chest-CAD
    Date Cleared
    2021-07-20

    (137 days)

    Product Code
    Regulation Number
    892.2070
    Age Range
    All
    Reference & Predicate Devices
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Chest-CAD is a computer-assisted detection (CADe) software device that analyzes chest radiograph studies using machine learning techniques to identify, categorize, and highlight suspicious regions of interest (ROI). Any suspicious ROI identified by Chest-CAD is assigned to one of the following categories: Cardiac, Mediastinum/Hila, Lungs, Pleura, Bones, Soft Tissues, Hardware, or Other. The device is intended for use as a concurrent reading aid for physicians. Chest-CAD is indicated for adults only.

    Device Description

    Chest-CAD is a computer-assisted detection (CADe) software device designed to assist physicians in identifying suspicious regions of interest (ROIs) in adult chest X-rays. Suspicious ROIs identified by Chest-CAD are assigned to one of the following categories: Cardiac, Mediastinum/Hila, Lungs, Pleura, Bones, Soft Tissues, Hardware, or Other. Chest-CAD detects suspicious ROIs by analyzing radiographs using deep learning algorithms for computer vision and provides relevant annotations to assist physicians with their interpretations.

    For each image within a study, Chest-CAD generates a DICOM Presentation State file (output overlay). If any suspicious ROI is detected by Chest-CAD in the study, the output overlay for all images includes the text "ROI(s) Detected:" followed by a list of the category/categories for which suspicious ROI(s) were found, such as "Lungs, Bones". In addition, if suspicious ROI(s) are detected in the image, bounding boxes surrounding each detected suspicious ROI are included in the output overlay. If no suspicious ROI is detected by Chest-CAD in the study, the output overlay for each image will include the text "No ROI(s) Detected" and no bounding boxes will be included. Regardless of whether a suspicious ROI is detected, the overlay includes text identifying the X-ray study as analyzed by Chest-CAD and a customer configurable message containing a link to or instructions for users to access labeling. The Chest-CAD overlay can be toggled on or off by the physician within their Picture Archiving and Communication System (PACS) viewer, allowing for concurrent review of the X-ray study.

    AI/ML Overview

    Acceptance Criteria and Device Performance for Chest-CAD (K210666)

    The Chest-CAD device is a computer-assisted detection (CADe) software that analyzes chest radiographs to identify, categorize, and highlight suspicious regions of interest (ROI). It is intended as a concurrent reading aid for physicians and is indicated for adults only.

    1. Acceptance Criteria and Reported Device Performance

    The acceptance criteria for Chest-CAD were established through both a standalone performance assessment and a multi-reader, multi-case (MRMC) comparative effectiveness study. No explicit numerical acceptance criteria were stated as targets for the standalone performance metrics (sensitivity, specificity, AUC). Instead, the performance demonstrated in the standalone test formed a basis for comparison and validation. For the MRMC study, the primary objective was to demonstrate superiority of aided reading over unaided reading in terms of AUC.

    Table 1: Acceptance Criteria and Reported Device Performance

    MetricAcceptance Criteria (Internal/Implicit)Standalone Device PerformanceMRMC Study Performance (Aided vs. Unaided)
    Standalone Performance
    Overall SensitivityHigh sensitivity is expected0.908 (95% CI: 0.905, 0.911)N/A
    Overall SpecificityHigh specificity is expected0.887 (95% CI: 0.885, 0.889)N/A
    Overall AUCHigh AUC is expected0.976 (95% CI: 0.975, 0.976)N/A
    Clinical Performance (MRMC Study)Aided vs. Unaided
    Reader AUCAided AUC > Unaided AUCN/AAided: 0.894 (95% CI: 0.879, 0.909) Unaided: 0.836 (95% CI: 0.816, 0.856)
    Reader SensitivityAided Sensitivity > Unaided SensitivityN/AAided: 0.856 (95% CI: 0.850, 0.862) Unaided: 0.757 (95% CI: 0.750, 0.764)
    Reader SpecificityAided Specificity > Unaided SpecificityN/AAided: 0.870 (95% CI: 0.866, 0.873) Unaided: 0.843 (95% CI: 0.839, 0.847)

    2. Sample Sizes and Data Provenance

    Standalone Test Set:

    • Sample Size: 20,000 chest radiograph cases.
    • Data Provenance: From 12 hospitals, outpatient centers, and specialty centers in the United States. The study was retrospective.

    MRMC Study Test Set:

    • Sample Size: 238 cases.
    • Data Provenance: From 9 hospitals, outpatient centers, and specialty centers in the United States. The study was retrospective.

    3. Number and Qualifications of Experts for Ground Truth (Test Set)

    • Standalone Test Set: Not explicitly stated for the standalone test set. The document indicates "Suspicious ROIs were manually annotated and categorized by board-certified radiologists before the images were used for benchmarking." This implies that the ground truth for standalone testing was established by expert radiologists, but the number and specific qualifications (years of experience) are not detailed.
    • MRMC Study Test Set: "Each case was previously evaluated by a panel of U.S. board-certified radiologists who assigned a ground truth binary label indicating the presence or absence of a suspicious ROI for each Chest-CAD category." The exact number of experts in the panel is not specified, nor is their specific level of experience (e.g., 10 years).

    4. Adjudication Method (Test Set)

    • The document implies a consensus-based adjudication method for the ground truth of the MRMC study cases, stating that "a panel of U.S. board-certified radiologists who assigned a ground truth binary label..." This strongly suggests that multiple radiologists reviewed the cases and reached a consensus for the ground truth. It does not explicitly state a 2+1 or 3+1 method but refers to a "panel" establishing the ground truth.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • Yes, a fully-crossed MRMC retrospective reader study was conducted.
    • Effect Size of Human Reader Improvement with AI vs. Without AI Assistance:
      • Reader AUC: Improved from 0.836 (Unaided) to 0.894 (Aided). Effect size = 0.058 increase.
      • Reader Sensitivity: Improved from 0.757 (Unaided) to 0.856 (Aided). Effect size = 0.099 increase.
      • Reader Specificity: Improved from 0.843 (Unaided) to 0.870 (Aided). Effect size = 0.027 increase.
        The study concluded that the accuracy of readers was superior when aided by Chest-CAD.

    6. Standalone (Algorithm Only) Performance Study

    • Yes, a standalone performance assessment was conducted.
    • Reported Performance:
      • Overall Sensitivity: 0.908 (95% CI: 0.905, 0.911)
      • Overall Specificity: 0.887 (95% CI: 0.885, 0.889)
      • Overall AUC: 0.976 (95% CI: 0.975, 0.976)
      • AUC by Category: Ranged from 0.921 (Mediastinum/Hila) to 0.994 (Hardware).
      • Sensitivity by Category: Ranged from 0.854 (Bones) to 0.967 (Hardware).
      • Specificity by Category: Ranged from 0.830 (Mediastinum/Hila) to 0.960 (Hardware).

    7. Type of Ground Truth Used (Test Set)

    • Expert Consensus: For both the standalone testing and the MRMC study, the ground truth was established by board-certified radiologists who manually annotated and assigned binary labels for suspicious ROIs.

    8. Sample Size for the Training Set

    • The document does not explicitly state the sample size for the training set. It refers to "deep learning algorithms for computer vision" but does not detail the dataset used for training or validation of these algorithms.

    9. How the Ground Truth for the Training Set was Established

    • The document does not explicitly state how the ground truth for the training set was established. It only mentions that the device uses "machine learning techniques" and "deep learning algorithms." Typically, for such devices, the training set ground truth would also be established by expert radiologists, usually through a process similar to or more extensive than that used for the test sets. However, this is not detailed in the provided text.
    Ask a Question

    Ask a specific question about this device

    K Number
    K193417

    Validate with FDA (Live)

    Date Cleared
    2020-07-30

    (234 days)

    Product Code
    Regulation Number
    892.2090
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    FractureDetect (FX) is a computer-assisted detection and diagnosis (CAD) software device to assist clinicians in detecting fractures during the review of radiographs of the musculoskeletal system. FX is indicated for adults only.

    FX is indicated for radiographs of the following industry-standard radiographic views and study types.

    Study Type(Anatomic Areaof Interest⁺)Radiographic View(s)Supported*
    AnkleFrontal, Lateral, Oblique
    ClavicleFrontal
    ElbowFrontal, Lateral
    FemurFrontal, Lateral
    ForearmFrontal, Lateral
    HipFrontal, Frog Leg Lateral
    HumerusFrontal, Lateral
    KneeFrontal, Lateral
    PelvisFrontal
    ShoulderFrontal, Lateral, Axillary
    Tibia / FibulaFrontal, Lateral
    WristFrontal, Lateral, Oblique

    *For the purposes of this table, "Frontal" is considered inclusive of both posteroanterior (PA) and anteroposterior (AP) views.

    +Definitions of anatomic area of interest and radiographic views are consistent with the American College of Radiology (ACR) standards and guidelines.

    Device Description

    FractureDetect (FX) is a computer-assisted detection and diagnosis (CAD) software device designed to assist clinicians in detecting fractures during the review of commonly acquired adult radiographs. FX does this by analyzing radiographs and providing relevant annotations, assisting clinicians in the detection of fractures within their diagnostic process at the point of care. FX was developed using robust scientific principles and industry-standard deep learning algorithms for computer vision.

    FX creates, as its output, a DICOM overlay with annotations indicating the presence or absence of fractures. If any fracture is detected by FX, the output overlay is composed to include the text annotation "Fracture: DETECTED" and to include one or more bounding boxes surrounding any fracture site(s). If no fracture is detected by FX, the output overlay is composed to include the text annotation "Fracture: NOT DETECTED" and no bounding box is included. Whether or not a fracture is detected, the overlay includes a text annotation identifying the radiograph as analyzed by FX and instructions for users to access labeling. The FX overlay can be toggled on or off by the clinicians within their PACS viewer, allowing for uninhibited concurrent review of the original radiograph.

    AI/ML Overview

    Here's a detailed breakdown of the acceptance criteria and the study proving the device meets them, based on the provided text:

    Acceptance Criteria and Device Performance

    Acceptance CriteriaReported Device Performance
    Standalone Performance
    Overall Sensitivity0.951 (95% Wilson's CI: 0.940, 0.960)
    Overall Specificity0.893 (95% Wilson's CI: 0.886, 0.898)
    Overall Area Under the Curve (AUC)0.982 (95% Bootstrap CI: 0.9790, 0.9850)
    AUC per Study Type: Ankle0.983 (0.972, 0.991)
    AUC per Study Type: Clavicle0.962 (0.948, 0.975)
    AUC per Study Type: Elbow0.964 (0.940, 0.982)
    AUC per Study Type: Femur0.989 (0.983, 0.994)
    AUC per Study Type: Forearm0.987 (0.977, 0.995)
    AUC per Study Type: Hip0.982 (0.962, 0.995)
    AUC per Study Type: Humerus0.983 (0.974, 0.991)
    AUC per Study Type: Knee0.996 (0.993, 0.998)
    AUC per Study Type: Pelvis0.982 (0.973, 0.989)
    AUC per Study Type: Shoulder0.962 (0.938, 0.982)
    AUC per Study Type: Tibia / Fibula0.994 (0.991, 0.997)
    AUC per Study Type: Wrist0.992 (0.988, 0.996)
    MRMC Comparative Effectiveness (Reader Performance with AI vs. without AI)
    Reader AUC (FX-Aided) vs. (FX-Unaided)Improved from 0.912 to 0.952, a difference of 0.0406 (95% CI: 0.0127, 0.0685) (p=.0043)
    Reader Sensitivity (FX-Aided) vs. (FX-Unaided)Improved from 0.819 (95% Wilson's CI: 0.794, 0.842) to 0.900 (95% Wilson's CI: 0.880, 0.917)
    Reader Specificity (FX-Aided) vs. (FX-Unaided)Improved from 0.890 (95% Wilson's CI: 0.879, 0.900) to 0.918 (95% Wilson's CI: 0.908, 0.927)

    Study Details

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size:
      • Standalone Study: 11,970 radiographs.
      • MRMC Reader Study: 175 cases.
    • Data Provenance: Not explicitly stated, but the experts establishing ground truth are specified as U.S. board-certified, suggesting the data is likely from the U.S. There is no indication whether the data was retrospective or prospective, but for an FDA submission of this nature, historical retrospective data is common.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • Number of Experts: A panel of three experts was used for the MRMC study's ground truth.
    • Qualifications: "U.S. board-certified orthopedic surgeons or U.S. board-certified radiologists." Specific years of experience are not mentioned.

    4. Adjudication Method for the Test Set

    • Adjudication Method: A "panel of three" experts assigned a ground truth binary label (presence or absence of fracture). This implies a consensus-based adjudication. While not explicitly stated (e.g., 2-out-of-3, or further adjudication if there was disagreement), the phrasing suggests a collective agreement to establish the "ground truth." This is analogous to a 3-expert consensus, where the majority rules.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • Was an MRMC study done? Yes.
    • Effect Size (Improvement with AI vs. without AI assistance):
      • Readers' AUC significantly improved by 0.0406 (from 0.912 to 0.952).
      • Readers' sensitivity improved by 0.081 (from 0.819 to 0.900).
      • Readers' specificity improved by 0.028 (from 0.890 to 0.918).

    6. Standalone (Algorithm Only) Performance Study

    • Was a standalone study done? Yes.
    • Performance:
      • Sensitivity: 0.951
      • Specificity: 0.893
      • Overall AUC: 0.982
      • High accuracy across study types and potential confounders (image brightness, x-ray manufacturers).

    7. Type of Ground Truth Used

    • Standalone Study: The ground truth for the standalone study is not explicitly detailed but given the MRMC study, it's highly probable it also leveraged expert consensus, similar to the MRMC setup, for fracture detection.
    • MRMC Study: Expert Consensus by a panel of three U.S. board-certified orthopedic surgeons or U.S. board-certified radiologists.

    8. Sample Size for the Training Set

    • The document does not explicitly state the sample size for the training set. It only mentions "robust scientific principles and industry-standard deep learning algorithms for computer vision" were used for development.

    9. How the Ground Truth for the Training Set Was Established

    • The document does not explicitly describe how the ground truth for the training set was established. It only mentions "Supervised Deep Learning" as the methodology, which implies labeled data was used for training, but the process of obtaining these labels is not detailed.
    Ask a Question

    Ask a specific question about this device

    K Number
    DEN180005

    Validate with FDA (Live)

    Device Name
    OsteoDetect
    Date Cleared
    2018-05-24

    (108 days)

    Product Code
    Regulation Number
    892.2090
    Type
    Direct
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    OsteoDetect analyzes wrist radiographs using machine learning techniques to identify and highlight distal radius fractures during the review of posterior-anterior (PA) and lateral (LAT) radiographs of adult wrists.

    Device Description

    OsteoDetect is a software device designed to assist clinicians in detecting distal radius fractures during the review of posterior-anterior (PA) and lateral (LAT) radiographs of adult wrists. The software uses deep learning techniques to analyze wrist radiographs (PA and LAT views) for distal radius fracture in adult patients.

    AI/ML Overview

    1. Table of Acceptance Criteria and Reported Device Performance

    Standalone Performance

    Performance MetricAcceptance Criteria (Implicit)Reported Device Performance (Estimate)95% Confidence Interval
    AUC of ROCHigh0.965(0.953, 0.976)
    SensitivityHigh0.921(0.886, 0.946)
    SpecificityHigh0.902(0.877, 0.922)
    PPVHigh0.813(0.769, 0.850)
    NPVHigh0.961(0.943, 0.973)
    Localization Accuracy (average pixel distance)Small33.52 pixelsNot provided for average distance itself, but standard deviation of 30.03 pixels.
    Generalizability (AUC for all subgroups)High≥ 0.926 (lowest subgroup - post-surgical radiographs)Not explicitly provided for all, but individual subgroup CIs available in text.

    MRMC (Reader Study) Performance - Aided vs. Unaided Reads

    Performance MetricAcceptance Criteria (Implicit: Superiority of Aided)Reported Device Performance (OD-Aided)Reported Device Performance (OD-Unaided)95% Confidence Interval (OD-Aided)95% Confidence Interval (OD-Unaided)p-value for difference
    AUC of ROCAUC_aided - AUC_unaided > 00.8890.840Not explicitly given for AUCs themselves, but difference CI: (0.019, 0.080)Not explicitly given for AUCs themselves, but difference CI: (0.019, 0.080)0.0056
    SensitivitySuperior Aided0.8030.747(0.785, 0.819)(0.728, 0.765)Not explicitly given for individual metrics, but non-overlapping CIs imply significance.
    SpecificitySuperior Aided0.9140.889(0.903, 0.924)(0.876, 0.900)Not explicitly given for individual metrics, but non-overlapping CIs imply significance.
    PPVSuperior Aided0.8830.844(0.868, 0.896)(0.826, 0.859)Not explicitly given for individual metrics, but non-overlapping CIs imply significance.
    NPVSuperior Aided0.8530.814(0.839, 0.865)(0.800, 0.828)Not explicitly given for individual metrics, but non-overlapping CIs imply significance.

    2. Sample Size and Data Provenance for Test Set

    Standalone Performance Test Set:

    • Sample Size: 1000 images (500 PA, 500 LAT)
    • Data Provenance: Retrospective. Randomly sampled from an existing validation database of consecutively collected images from patients receiving wrist radiographs at the (b) (4) from November 1, 2016 to April 30, 2017. The study population included images from the US.

    MRMC (Reader Study) Test Set:

    • Sample Size: 200 cases.
    • Data Provenance: Retrospective. Randomly sampled from the same validation database used for the standalone performance study. The data includes cases from the US.

    3. Number of Experts and Qualifications for Ground Truth

    Standalone Performance Test Set and MRMC (Reader Study) Test Set:

    • Number of Experts: Three.
    • Qualifications: U.S. board-certified orthopedic hand surgeons.

    4. Adjudication Method for Test Set

    Standalone Performance Test Set:

    • Adjudication Method (Binary Fracture Presence/Absence): Majority opinion of at least 2 of the 3 clinicians.
    • Adjudication Method (Localization - Bounding Box): The union of the bounding box of each clinician identifying the fracture.

    MRMC (Reader Study) Test Set:

    • Adjudication Method: Majority opinion of three U.S. board-certified orthopedic hand surgeons. (Note: this was defined on a per-case basis, considering PA, LAT, and oblique images if available).

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • Was an MRMC study done? Yes.
    • Effect Size (Improvement of Human Readers with AI vs. without AI assistance):
      • The least squares mean difference between the AUC for OsteoDetect-aided and OsteoDetect-unaided reads is 0.049 (95% CI, (0.019, 0.080)). This indicates a statistically significant improvement in diagnostic accuracy (AUC) of 4.9 percentage points when readers were aided by OsteoDetect.
      • Sensitivity: Improved from 0.747 (unaided) to 0.803 (aided), an improvement of 0.056.
      • Specificity: Improved from 0.889 (unaided) to 0.914 (aided), an improvement of 0.025.

    6. Standalone (Algorithm Only) Performance Study

    • Was a standalone study done? Yes.

    7. Type of Ground Truth Used

    Standalone Performance Test Set:

    • Type of Ground Truth: Expert consensus (majority opinion of three U.S. board-certified orthopedic hand surgeons).

    MRMC (Reader Study) Test Set:

    • Type of Ground Truth: Expert consensus (majority opinion of three U.S. board-certified orthopedic hand surgeons).

    8. Sample Size for Training Set

    The document does not explicitly state the sample size for the training set. It mentions "randomly withheld subset of the model's training data" for setting the operating point, implying a training set existed, but its size is not provided.

    9. How Ground Truth for Training Set Was Established

    The document does not explicitly state how the ground truth for the training set was established. It only refers to a "randomly withheld subset of the model's training data" during the operating point setting.

    Ask a Question

    Ask a specific question about this device