510(k) Data Aggregation

    K Number
    K251975

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-09-02

    (68 days)

    Product Code
    Regulation Number
    888.3535
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ACHIEVE PARTIAL KNEE SYSTEM is intended for unicompartmental knee arthroplasty of the medial and/or lateral femorotibial compartments to treat one or more of the following conditions:

    • Moderately disabling joint disease of the knee resulting from painful osteo or post traumatic arthritis.
    • Revision of previous unsuccessful surgical procedures, including prior unicompartmental knee arthroplasty.
    • As an alternative to tibial osteotomy in patients with unicompartmental osteoarthritis.

    Fixation Methods

    Tibial trays and femoral components with engineered porous surfaces are intended for cemented or cementless use.

    Tibial trays and femoral components without engineered porous surfaces are intended for cemented use only.

    Device Description

    The subject ACHIEVE™ Partial Knee System – Porous Coated CoCrMo Femoral Component is a line extension to the ACHIEVE™ Partial Knee System (cleared via K242307). The line extension is primarily focused on the addition of a Porous Coated Cobalt Chromium Molybdenum (CoCrMo) Femoral Component into the system, but also includes some minor modifications to material, manufacturing location, and design of the Femoral Components and Tibial Trays, which were previously cleared in K242307. The Instructions for Use and Surgical Technique were updated to account for the material changes of the subject devices. The indications for use were also updated to clarify that the device is intended for use in the medial and/or lateral compartment of the tibiofemoral joint.

    The ACHIEVE™ PARTIAL KNEE SYSTEM – POROUS COATED COCRMO FEMORAL COMPONENT is a knee joint femorotibial (unicompartmental) prosthesis that is intended for cementless or cemented fixation. The implant system consists of individually packaged implants: a metal tibial tray (titanium alloy), a polyethylene tibial insert, and a metal femoral component (titanium alloy or cobalt-chromium). The tibial trays and femoral components with HAnano Surface™ have an adjunct coating of hydroxyapatite (HA) on the engineered porous surfaces. All tibial inserts are composed of a Cross-linked, Vitamin E Ultra High Molecular Weight Polyethylene (Cross-Linked, VE UHMWPE).

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K242307

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-12-16

    (133 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ACHIEVE PARTIAL KNEE SYSTEM is intended for unicompartmental knee arthroplasty to treat one or more of the following conditions:

    • · Moderately disabling joint disease of the knee resulting from painful osteo or post traumatic arthritis.
    • · Revision of previous unsuccessful surgical procedures, including prior unicompartmental knee arthroplasty.
    • · As an alternative to tibial osteotomy in patients with unicompartmental osteoarthritis.
    Device Description

    The ACHIEVE™ PARTIAL KNEE SYSTEM is a knee joint femorotibial (unicompartmental) prosthesis that is intended for cementless or cemented fixation.
    The implant system consists of individually packaged implants: a metal tibial tray (titanium alloy), a polyethylene tibial insert, and a metal femoral component (titanium alloy or cobalt-chromium). All tibial inserts are composed of a Cross-linked, Vitamin E Ultra High Molecular Weight Polyethylene (Cross-Linked, VE UHMWPE).

    AI/ML Overview

    This document does not contain information about the acceptance criteria and study detailed in the request. The document is a 510(k) premarket notification for a medical device (ACHIEVE™ Partial Knee System) and focuses on demonstrating substantial equivalence to a predicate device through non-clinical testing. It explicitly states that clinical testing was not necessary to demonstrate substantial equivalence.

    Therefore, I cannot provide the requested information regarding acceptance criteria, device performance, sample sizes for test or training sets, data provenance, expert qualifications, ground truth, or MRMC studies, as these aspects are typically associated with clinical trials or performance studies involving AI/software devices.

    The document describes engineering tests and analyses performed on the physical components of the knee system. For instance:

    • Range of Motion (RoM) Evaluation: Acceptance criteria met per ASTM F2083.
    • Femoral Fatigue Testing: Acceptance criteria met per modified ASTM F3210 (10 Mc).
    • Tibial Tray Fatigue Testing: Acceptance criteria met per modified ASTM F3140 (10 Mc).
    • Component Interlock Strength Testing: Acceptance criteria met for static AP and ML shear testing and static tensile pull-off testing.
    • Wear Resistance Evaluation: Wear rate does not represent a new worst-case compared to the predicate device per ISO 14243-3.
    • Biocompatibility Assessments: Devices found to be biocompatible per ISO 10993-1 and FDA Guidance.
    • Porous Structure Characterization: Meets recommendations of Class II Special Controls Guidance Document per ASTM F1044, ASTM F1147, ASTM F1160, ASTM F1978, and ASTM F1854.
    • Shelf-Life Evaluation: Five-year shelf life established per ISO 11607-1 and ISO 11607-2.
    • Sterilization Validation: Sterility Assurance Level (SAL) of 10-6 found per ISO 11137-1 and ISO 11137-2.

    However, these are all engineering benchmarks for the physical orthopedic implant, not performance metrics for a diagnostic AI/software device.