Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K951014

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1996-10-18

    (592 days)

    Product Code
    Regulation Number
    862.1700
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Isolab's T4 Test Kit is intended for use in determining the total thyroxine (T4) in blood specimens dried on filter paper.

    Device Description

    The Isolab's T4 Test Kit is a sequential, competitive enzyme immunossay with fluorometric detection in which T4 is eluted from dried blood disks and simultaneously forms a complex with the biotin-labeled monoclonal antibody which binds to the streptavidin coated solid phase. After sufficient incubation with biotinylated antibody, a conjugate of T4 with horseradish peroxidase is added to the wells and incubated with the reaction mixture. This conjugate competes with the analyte for the free on the bound monoclonal antibody. Excess of unbound reagent is washed away after a sequential incubation with biotinylated antibody and conjugate, and the enzymatic reaction with a fluorogenic substrate 3-phydroxyphenylpropionic acid (HPPA) [9] takes place. The reaction is stopped by addition of glycine buffer and the fluorescence in each well is measured with Fluoroscan II Neonate at 405 nm (excitation being 320 nm).

    AI/ML Overview

    Here's an analysis of the provided text regarding the Isolab T4 Test Kit's acceptance criteria and studies:

    Device: Isolab T4 Test Kit for determining total thyroxine (T4) in blood specimens dried on filter paper.

    Intended Use: Screening for congenital hypothyroidism (CH) in neonates. The results are considered accurate for screening purposes, but due to hematocrit value differences, they should not be assumed as accurate as serum-based assays.

    Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria/MetricTarget/Expected Value (Implicit)Reported Device Performance
    LinearityRange of linearityNot explicitly stated, but generally as wide as physiologically relevant.0 to 25 ug/dL of T4
    Recovery% RecoveryPreferably close to 100%88% at low T4 levels (6.25 µg/dL spiked), 97% at normal T4 levels (12.5 µg/dL spiked)
    Sensitivity (Detection Limit)Lower limit of detectionAs low as clinically relevant, ideally below the CH cutoff.1.0 ug/dL
    Reproducibility (Precision)Total SD / Total CVLower values indicate better precision.T4-depleted neonate (6.25 µg/dL spiked): - Mean: 5.5 µg/dL - Total SD: 0.57 - Total CV: 10.4% - Within-run SD: 0.43 - Within-run CV: 7.8% Normal neonate (12.5 µg/dL spiked): - Mean: 12.1 µg/dL - Total SD: 1.89 - Total CV: 15.6% - Within-run SD: 1.52 - Within-run CV: 12.6%
    Method Comparison (Cutoff Values)Agreement with established methods for cutoff.Cutoff value generally accepted for screening. For CH, typically based on lower 10%.Isolab: 8.5 µg/dL (State 1), 8.1 µg/dL (State 2) Neometrics: 11.7 µg/dL (State 2) Wallac Oy: 10.2 µg/dL (State 1)
    InterferenceMinimal interference from common substances or therapeutic agents.Interference should not significantly impact results.D-T4: ~100% increase (expected). DL-T, TA3: Slight interference. 5-propyl-2-thiouracil (PTU): Slight interference at tested concentration (likely higher than neonatal therapeutic levels).

    Study Details

    1. Sample Sizes Used for Test Set and Data Provenance:

      • Linearity: Unspecified number of whole blood samples, but each spiked concentration (6.25, 12.5, 25 ug/dL, plus unspiked) was run five times. This suggests a small sample size per concentration. Data provenance not specified (country, retrospective/prospective).
      • Recovery: Not a separate sample set; derived from the same data as Reproducibility.
      • Sensitivity (Detection Limit): Sample size for determining detection limit is not specified.
      • Reproducibility: Two sample pools (T4-depleted neonate and normal neonate) were used. Each pool was run in duplicate, twice a day, for 20 days. This means 80 measurements per pool (2 assays/day * 20 days * 2 duplicates). Data provenance not specified.
      • Comparison of Methods: Not explicitly stated how many samples were tested for the comparison. It mentions "the lower 10% of the values for each of the assays are shown," implying a larger dataset from which these cutoffs were derived. The comparative assays (Neometrics and Wallac Oy) were "performed at two state laboratory facilities," suggesting the data comes from these facilities (country not specified). It is likely retrospective data as the comparison is against existing methods.
      • Interference Testing: A control sample was assayed seven times. Each spiked substance sample was also assayed seven times. Data provenance not specified.
    2. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts:

      • No information is provided about the use of experts to establish ground truth for the test set in the traditional sense (e.g., for diagnostic accuracy).
      • For the "Comparison of Methods," the cut-off values for the comparative assays were "performed at two state laboratory facilities." These facilities would presumably employ qualified laboratory personnel, but their specific qualifications are not detailed.
      • The "ground truth" for linearity, recovery, and reproducibility studies is based on the known spiked concentrations and repeated measurements, not expert consensus on diagnostic outcomes.
    3. Adjudication Method for the Test Set:

      • Not applicable/Not described. The studies focus on analytical performance (linearity, precision, recovery, interference) and comparison of analytical cutoffs, not diagnostic adjudication.
    4. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

      • No. This type of study (comparing human readers with and without AI assistance affecting diagnostic performance) is not applicable to an immunoassay like the T4 Test Kit. The device is a standalone diagnostic test, not an AI-assisted interpretation tool.
    5. Standalone (Algorithm Only) Performance:

      • Yes, all the described studies (linearity, recovery, sensitivity, reproducibility, interference, and method comparison) represent the standalone performance of the Isolab T4 Test Kit as an analytical device. There is no human-in-the-loop component for these evaluations.
    6. Type of Ground Truth Used:

      • Known Spiked Concentrations: For Linearity, Recovery, and Reproducibility, the ground truth is established by adding known amounts of T4 to samples.
      • Analytical Measurement: For Sensitivity (detection limit), the ground truth is the reliably detectable lowest concentration.
      • Reference Method/Established Cutoffs: For the Comparison of Methods, the "ground truth" for comparison involves the established performance and cutoff values from Neometrics Accuscreen T4 and Wallac Oy Delphia Neonatal T4 assays, as interpreted by state laboratory facilities.
      • Known Interferent Presence: For Interference testing, the ground truth is the presence of known interfering substances.
    7. Sample Size for the Training Set:

      • Not applicable/Not explicitly mentioned. The Isolab T4 Test Kit is a chemical immunoassay, not a machine learning or AI algorithm that requires a "training set" in the computational sense. The development of the assay itself would have involved extensive R&D and optimization, but this isn't typically referred to as a "training set" for the final product in regulatory submissions.
    8. How the Ground Truth for the Training Set Was Established:

      • Not applicable, as there's no "training set" for this type of device in the context of AI/ML. The "ground truth" in the development of such assays is based on established biochemical principles, purified standards, and clinical validation against reference methods to ensure the assay accurately measures T4.
    Ask a Question

    Ask a specific question about this device

    K Number
    K950481

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1996-02-01

    (363 days)

    Product Code
    Regulation Number
    862.1310
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K943547

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1996-01-26

    (554 days)

    Product Code
    Regulation Number
    862.1555
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1