Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K181316

    Validate with FDA (Live)

    Date Cleared
    2018-06-14

    (27 days)

    Product Code
    Regulation Number
    862.1345
    Age Range
    2 - 12
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The POGO Automatic Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips. The POGO Automatic Blood Glucose Monitoring System is intended to be used by a single person and should not be shared.

    The POGO Automatic Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The POGO Automatic Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. The POGO Automatic Blood Glucose Monitoring System is indicated for use in adults and adolescents (13 and up).

    POGO Automatic Test Cartridges are for use with the POGO Automatic Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips.

    Device Description

    The POGO BGMS is a quantitative assay for the detection of glucose in capillary whole blood sampled from the fingertip. The system includes multiple glucose-oxidase-based dry-reagent test strips housed in a cartridge and a photometer to read the glucose-dependent color change from the in-use strip. The meter uses calibration information from a barcode on the disposable cartridge to convert the reflectance information into a plasma-equivalent glucose value. A Bluetooth Low Energy (BLE) module in the meter allows the user to authorize the POGO meter to send data wirelessly, securely, and automatically after each test to specified electronic computing devices.

    The POGO BGMS automates finger lancing, blood sample collection and placement onto the test strip, and calculation of the blood glucose result, and so requires significantly fewer steps than existing BGMSs to obtain a glucose result. The POGO BGMS accomplishes this via a 10-test cartridge where each foil-sealed test includes a hollow lancet, spring, and test strip. The user does not need a separate lancing device since the lancing mechanism is built into individual test cells of the cartridge. The blood sample is drawn into the test through capillary action and the meter, by monitoring the time and degree of blood spreading on the reagent pad, can detect when the test is under-filled and allow the user a limited amount of time to add blood to the test trip to obtain a glucose result rather than an error. Additionally, used tests are retained within the cartridge for added user convenience.

    AI/ML Overview

    The provided document is a 510(k) summary for the POGO Automatic Blood Glucose Monitoring System (POGO BGMS). It describes the device, its intended use, and compares it to a predicate device (an earlier version of the same system).

    However, this document does not contain the detailed information necessary to fully answer your request regarding acceptance criteria and the study that proves the device meets them. Specifically, it lacks:

    • A table of acceptance criteria and reported device performance for glucose measurement accuracy. While it lists general characteristics, it does not provide specific performance metrics like accuracy studies (e.g., ISO 15197 compliance data).
    • Sample sizes used for testing.
    • Ground truth establishment and expert qualifications.
    • Information about MRMC studies, standalone algorithm performance, or training set details.

    The document primarily focuses on demonstrating substantial equivalence to a predicate device based on modifications that are largely non-functional (BLE capability, display changes, speaker, feet). It asserts that the fundamental technological characteristics, blood sampling, blood glucose measurement, and algorithm remain unchanged from the predicate. Therefore, it relies on the predicate's proven performance for establishing substantial equivalence in its core function (glucose measurement).

    Based on the provided text, I can infer the following:

    The acceptance criteria for the POGO Automatic Blood Glucose Monitoring System's glucose measurement performance is that it performs as safely and effectively as the predicate device (K162203). The document explicitly states: "The subject modifications for the POGO BGMS do not affect any of the above attributes that were included in the substantial equivalence section of the predicate device and relied upon for the substantial equivalence finding for the predicate device, K162203." It also mentions "Same fundamental technological characteristics. The following remain unchanged: blood ● sampling and blood glucose measurement, use of the same cartridge (which contains the reagent system used to measure blood glucose), technology and procedure to acquire a blood sample and monitor the color development on the reagent pad and the algorithm to calculate a glucose value from the reaction at the reagent pad."

    Therefore, the "study that proves the device meets the acceptance criteria" for glucose measurement is implicitly the original study data for the predicate device (K162203), as the core measurement technology has not changed. The current submission focuses on demonstrating that the modifications (BLE, display, etc.) do not degrade this established performance.

    Here's what can be extracted and what is missing based on your request:


    Acceptance Criteria for the POGO Automatic Blood Glucose Monitoring System

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria CategorySpecific Criterion (Inferred from non-change)Reported Device Performance (Inferred/Stated)
    Glucose Measurement AccuracyEquivalent to predicate device (K162203) as core technology is unchanged.The document states the blood glucose measurement technology and algorithm are "unchanged." Specific accuracy data (e.g., % within ISO 15197 zones) is not provided in this document for either the subject or predicate.
    Functional ModificationsPerformance specifications established for modified components."Pass" result for all performance tests on modified items.
    Safety TestingCompliance with Mechanical, Environmental, and Electrical safety standards for SMBG devices."Pass" result for all safety tests.
    Cleaning & DisinfectionEfficacy and robustness challenges on raw materials and assembled device."Pass" result for efficacy and robustness challenges.
    Software V&VSoftware V&V requirements met."Pass" result for Software V&V.
    Wireless Connectivity (BLE)Bluetooth wireless proximity and coexistence testing requirements met."Pass" result for BLE testing.
    LabelingReadability."Pass" result for labeling readability.
    Under-filled detectionDetects under-filled reagent pad."Yes," device detects under-filled reagent pad.
    Blood reapplicationAllows blood reapplication for up to 15 seconds."Yes," provides glucose result or error code after blood addition; 15 seconds allowed for reapplication.
    Blood application time90 seconds allowed if blood does not reach reagent pad."Yes," 90 seconds allowed.
    Strip exposure limitsMeter enforces limits on strip exposure (5 min open cell time window)."Yes," device design prevents strip use after 5 minutes.
    Lancet re-usePrevents lancet re-use."No," device design prevents re-use.
    Cal CodingAutomated."Automated."
    Hematocrit Range20% to 60%."20% to 60%."
    Humidity Range10% to 90% RH."10% to 90% RH."
    Measurement Range20 - 500 mg/dL."20 - 500 mg/dL."
    Min. Sample Size0.25 µL."0.25 µL."
    Operating Temperature Range50 to 104° F."50 to 104° F."

    2. Sample size used for the test set and the data provenance:

    • Test Set Sample Size: Not specified for any of the performance tests. The document only states that "Multiple nonclinical tests were identified via risk assessment activities as requirements to demonstrate comparable safety and effectiveness."
    • Data Provenance: Not specified (e.g., country of origin). The document implies the tests were conducted by the manufacturer, Intuitv Medical, Inc. The data is retrospective in the sense that it relies on the predicate's established performance and new testing for the modifications.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable / Not specified. This type of information is usually relevant for AI/ML diagnostic devices where human expert consensus establishes ground truth for image interpretation or diagnosis. For a blood glucose monitoring system, the ground truth for glucose measurements would typically be established by a laboratory reference method (e.g., YSI analyzer), rather than expert consensus.

    4. Adjudication method for the test set:

    • Not applicable / Not specified. See point 3.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This is not relevant for a blood glucose monitoring system, which provides a direct quantitative measurement and does not involve human interpretation of complex medical images or data requiring AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Yes (in principle, for the core function). The device itself performs the glucose measurement and calculation ("algorithm to calculate a glucose value from the reaction at the reagent pad"). The performance refers to the device's accuracy against a known standard. The 510(k) submission states that this core algorithm and measurement technology are unchanged from the predicate device. Therefore, the standalone performance was established for the predicate device.

    7. The type of ground truth used:

    • Laboratory Reference Method. For blood glucose monitoring systems, the ground truth for glucose concentration is typically established by a highly accurate laboratory analyzer (e.g., YSI glucose analyzer) that is well-calibrated and traceable to international standards. This document does not explicitly state the reference method, but it is standard practice for such devices.

    8. The sample size for the training set:

    • Not applicable / Not specified. Blood glucose monitoring systems typically use a chemical reaction and photometric measurement with a pre-defined calibration curve. They are not generally "trained" in the typical machine learning sense with a "training set" of patient data that needs a ground truth established in the same way as an AI algorithm for image recognition. The "algorithm" here refers to the fixed mathematical conversion from optical signal to glucose concentration, derived from chemical principles and calibrated during manufacturing. While calibration data is used, it's not described as a "training set" in this context.

    9. How the ground truth for the training set was established:

    • Not applicable. As above, the device uses a chemical-photometric principle. Calibration curves are established using samples with known glucose concentrations, typically measured by a laboratory reference instrument (the "ground truth").
    Ask a Question

    Ask a specific question about this device

    K Number
    K162203

    Validate with FDA (Live)

    Date Cleared
    2017-04-06

    (244 days)

    Product Code
    Regulation Number
    862.1345
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The POGO Automatic Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips. The POGO Automatic Blood Glucose Monitoring System is intended to be used by a single person and should not be shared.

    The POGO Automatic Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The POGO Automatic Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. The POGO Automatic Blood Glucose Monitoring System is indicated for use in adults and adolescents (13 and up).

    POGO Automatic Test Cartridges are for use with the POGO Automatic Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips.

    Device Description

    The POGO Automatic Blood Glucose Monitoring System (POGO BGMS) is a quantitative assay for the detection of glucose in capillary whole blood sampled from the fingertip. The system includes multiple glucose-oxidase-based dry-reagent test strips housed in a cartridge and a photometer to read the glucose-dependent color change from the in-use strip. The meter uses calibration information from a barcode on the disposable cartridge to convert the reflectance information into a plasma-equivalent glucose value.

    The POGO BGMS automates finger lancing, blood sample collection and placement onto the test strip, and calculation of the blood glucose result, and so requires significantly fewer steps than existing BGMSs to obtain a glucose result. The POGO BGMS accomplishes this via a 10-test cartridge where each foil-sealed test includes a hollow lancet, spring, and test strip. The user does not need a separate lancing device since the lancing mechanism is built into individual test cells of the cartridge. The blood sample is drawn into the test through capillary action and the meter, by monitoring the time and degree of blood spreading on the reagent pad, can detect when the test is under-filled and allow the user a limited amount of time to add blood to the test trip to obtain a glucose result rather than an error. Additionally, used tests are retained within the cartridge for added user convenience. Control solutions are available to confirm correct system performance.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    Acceptance Criteria and Device Performance

    Acceptance Criteria (from FDA Guidance for Self-Monitoring Blood Glucose Test Systems for Over-the-Counter Use, October 2016)Reported Device Performance (POGO BGMS Self-Test Glucose Results)
    Accuracy relative to YSI reference:
    For glucose concentrations < 75 mg/dL:
    All data within ±15 mg/dL bias to the plasma YSI reference results.100% of POGO BGMS self-test glucose results were within ±15 mg/dL of the reference YSI results at a glucose concentration of <75 mg/dL for both Automated and Reapplication results.
    For glucose concentrations ≥ 75 mg/dL:
    At least 95% of results within ±15% of the YSI reference.At glucose concentrations of ≥75 mg/dL, 95.0% of POGO BGMS self-test glucose results were within ±15% of the reference YSI (95.8% Automated and 91.5% Reapplication).
    (Implicitly, better than 95% within ±20% is also typically expected)99.2% of POGO BGMS results were within ±20% of the reference YSI (99.1% Automated and 100% Reapplication).
    Reapplication Feature Accuracy (implied by additional studies):The overall results from each study demonstrated at least 95% of results within 15% (for glucose > 75mg/dL) or 10 mg/dL (for glucose < 75mg/dL) of the YSI when used in Reapplication Mode.

    The document states, "These results met the predetermined accuracy specifications."

    Study Details

    2. Sample sizes used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Clinical Lay User Evaluation (initial study):
      • Sample Size: 285 subjects with diabetes.
      • Data Provenance: The document does not explicitly state the country of origin. It indicates it was a "multi-site" study, suggesting it was prospective clinical data.
    • Additional Reapplication Studies:
      • Sample Size: 215 data points were obtained in total.
      • Data Provenance: The document does not explicitly state the country of origin. It indicates these were "clinical studies" where "lay users tested," suggesting prospective clinical data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    The ground truth was established using a YSI reference. YSI is a laboratory-grade glucose analyzer, not an expert panel. Therefore, no human experts were involved in establishing the ground truth for comparison.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    There was no human adjudication as the ground truth was established by a YSI laboratory reference.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, an MRMC comparative effectiveness study involving human readers and AI assistance was not performed. This device is a blood glucose monitoring system, not an AI-assisted diagnostic imaging tool.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Yes, the studies evaluated the standalone performance of the POGO Automatic Blood Glucose Monitoring System. The results provided are for the device (algorithm and hardware) directly compared to the YSI reference, representing its intrinsic accuracy without human interpretation influencing the numerical glucose value. The "self-test data" implies users operated the device themselves, but the performance metrics are about the device's accuracy.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth used was plasma YSI reference results. YSI (Yellow Springs Instruments) analyzers are commonly used as "gold standard" or laboratory reference methods for glucose measurement in clinical studies due to their high accuracy and precision.

    8. The sample size for the training set

    The document does not provide specific details about a training set for an algorithm. It discusses "nonclinical data" where "IMI obtained 60 POGO BGMS whole blood results via intermittent sampling at nominal glucose levels from 60 to 225 mg/dL. All data were within ±15% bias to the plasma YSI reference results." This sounds more like a verification or early validation step rather than a formal, large-scale training set for a machine learning model. The device's "photometric" detection method and "glucose oxidase" enzyme suggest a more traditional biochemical assay rather than a complex AI algorithm that typically requires a large training dataset in the context of AI/ML devices.

    9. How the ground truth for the training set was established

    As mentioned above, the available information describes a nonclinical dataset where results were compared against "plasma YSI reference results." This indicates the YSI system was also used as the ground truth for this nonclinical testing, similar to the clinical studies.

    Ask a Question

    Ask a specific question about this device

    K Number
    K152493

    Validate with FDA (Live)

    Date Cleared
    2016-04-25

    (237 days)

    Product Code
    Regulation Number
    862.1345
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The POGO Automatic Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips. The POGO Automatic Blood Glucose Monitoring System is intended to be used by a single person and should not be shared.

    The POGO Automatic Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The POGO Automatic Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. The POGO Automatic Blood Glucose Monitoring System is indicated for use in adults and adolescents (13 and up).

    POGO Automatic Test Cartridges are for use with the POGO Automatic Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips.

    POGO Control Solutions are used with the POGO Automatic Meter to indicate appropriate user technique and to indicate that the POGO Automatic Test Cartridge and POGO Automatic Meter are functioning properly.

    Device Description

    The POGO Automatic Blood Glucose Monitoring System (POGO BGMS) is a quantitative assay for the detection of glucose in capillary whole blood sampled from the fingertip. The system includes multiple glucose-oxidase-based dry-reagent test strips housed in a cartridge and a photometer to read the glucose-dependent color change from the in-use strip. The meter uses calibration information from a barcode on the disposable cartridge to convert the reflectance information into a plasma-equivalent glucose value.

    The POGO BGMS automates finger lancing, blood sample collection and placement onto the test strip, and calculation of the blood glucose result, and so requires significantly fewer steps than existing BGMSs to obtain a glucose result. The POGO BGMS accomplishes this via a 10-test cartridge where each foil-sealed test includes a hollow lancet, spring, and test strip. The user does not need a separate lancing device since the lancing mechanism is built into individual test cells of the cartridge. Additionally, used tests are retained within the cartridge for added user convenience. Control solutions are available to confirm correct system performance.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Pogo Automatic Blood Glucose Monitoring System, based on the provided text:

    Acceptance Criteria and Device Performance

    Acceptance Criteria CategoryAcceptance Criteria (from "The protocol acceptance criteria were met")Reported Device Performance (as described in the text)
    RepeatabilityMet the protocol acceptance criteria.%CVs ranged from 2.3% to 3.4% across five whole blood samples (41.8 mg/dL to 306.7 mg/dL) tested in replicates of 10 with 10 meters.
    Intermediate PrecisionMet the protocol acceptance criteria.%CVs were 6.5% (2.1 SD), 3.1%, and 3.2% for three levels of commercial controls (approx. 30, 100, and 300 mg/dL) tested in duplicate with 10 meters over 10 days.
    LinearityMet the protocol acceptance criteria.Linear throughout the claimed dynamic range, based on testing a whole blood sample spiked with nine glucose levels (17 mg/dL to 529 mg/dL) in replicates of 10.
    Hematocrit ToleranceMet the protocol acceptance criteria.Accurate throughout the claimed 20-60% hematocrit range, based on studies with four glucose levels (60, 120, 250, and 400 mg/dL) and seven different hematocrits.
    Altitude ToleranceMet the protocol acceptance criteria.Provides accurate results at altitudes up to the claimed 10,000 feet, based on studies at sea level, 5,000 feet, and 10,000 feet with blood at nominal glucose levels of 60, 120, and 300 mg/dL.
    Temperature ToleranceMet the protocol acceptance criteria.Provides accurate results across the tested 4°C to 40°C range and the claimed temperature range of 10°C to 40°C, based on testing at five temperatures with nominal glucose levels of 60, 120, and 300 mg/dL.
    Humidity ToleranceMet the protocol acceptance criteria.Provides accurate results when tested across the claimed humidity range of 10% to 90% RH, based on testing at 10%, 50%, and 90% relative humidity with nominal glucose levels of 60, 120, and 300 mg/dL.
    Temperature and Humidity CombinationsMet the protocol acceptance criteria.Provides accurate results at extreme combinations of temperature and humidity (25°C/45% RH, and 10℃/40℃ crossed with 10%/90% RH) with three nominal glucose levels of 60, 120, and 300 mg/dL.
    Sample Volume ToleranceMet the protocol acceptance criteria.Accurate when tested with a sample volume between 250 and 1000 nL. At < 250 nL, provided no result or an accurate glucose result. Demonstrated accuracy across a wide range of sample volumes (from less than claimed minimum to volumes beyond normal usage) at nominal glucose levels of 60, 120, and 300 mg/dL.
    Interfering SubstancesMet the protocol acceptance criteria.No interference beyond acceptance criteria when challenged at two glucose concentrations (80 mg/dL and 320 mg/dL) with therapeutic and toxic levels of bilirubin, cholesterol, hemoglobin, triglycerides, uric acid, common medications, and alternate sugars.
    Test Range LimitsMet the protocol acceptance criteria.Displayed "LO" for all tested glucose levels below 20 mg/dL and "HI" for all tested glucose levels above 500 mg/dL, with no numeric results, when challenged with blood spiked to 10, 550, and 1100 mg/dL. Correct responses for samples outside claimed quantitation range (20-500 mg/dL).
    Open Test ToleranceMet the protocol acceptance criteria for the 6.5 minute time point.Provides accurate results at up to the claimed 5 minutes of open time prior to the start of a test, based on testing nominal blood glucose levels of 80 and 320 mg/dL.
    Open Cartridge ToleranceMet the protocol acceptance criteria.Provides accurate results for the claimed open cartridge duration of 45 days, based on testing cartridges after up to 45 days of exposure in environmental chamber (4°C-30°C and 10-90% humidity) using venous whole blood spiked to 80 mg/dL and 320 mg/dL.
    Clinical Accuracy (<75 mg/dL)100% of self-test results within ±15 mg/dL of reference.User Self-Testing: 100% of POGO BGMS self-test glucose results were within ±15 mg/dL of the reference YSI results. HCP Assisted Testing: 94.4% of POGO BGMS glucose results were within ±15 mg/dL of the reference YSI results.
    Clinical Accuracy (≥75 mg/dL)95% of self-test results within 15% of reference or 20% of reference.User Self-Testing: 95.8% of POGO BGMS self-test glucose results were within 15% of the reference YSI and 99.1% of POGO BGMS results were within ±20% of the reference YSI. HCP Assisted Testing: 96.1% of POGO BGMS accuracy verification results were within 15% of the reference YSI and 98.5% of POGO BGMS results were within ±20% of the reference YSI.
    Ease of Use/User UnderstandingNot explicitly stated as "acceptance criteria," but implied through positive user feedback thresholds.92.3% of patients found the POGO BGMS easy to use. 96.5% of subjects were capable of using the POGO BGMS (evaluated by HCP). 91.9% of subjects found instructions easy to understand. At least 72% of subjects "agreed" or "strongly agreed" it was easier than their current BGMS, and >80% felt it was more convenient. Majority (51%) agreed they'd better adhere to testing, and >81% would recommend it.

    Study Details

    1. Sample size used for the test set and the data provenance:

      • Test Set Sample Size (Clinical): 287 subjects with diabetes.
      • Data Provenance: Subjects were enrolled across US clinical sites. The study was prospective as subjects performed tests, were observed, and completed questionnaires.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Number of Experts: Not explicitly stated as a specific number. The text mentions "health care professionals (HCPs)" observed subjects and "HCPs also performed assisted fingerstick tests." It doesn't specify how many or their exact qualifications beyond "health care professionals."
      • Qualifications: "Health care professionals (HCPs)." No specific details like years of experience or specialization (e.g., endocrinologist, nurse) are provided.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • The text does not describe an adjudication method for the test set results. The accuracy of the POGO BGMS results was compared directly against a reference instrument (YSI).
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This is a blood glucose monitoring system, not an imaging AI device. The study focused on the performance of the device itself and user experience.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • Yes, a standalone performance was done for the analytical studies (Repeatability, Intermediate Precision, Linearity, Hematocrit Tolerance, etc.). These tests evaluate the device's inherent analytic capabilities under controlled conditions.
      • The clinical accuracy study also includes a "user self-testing" arm, which represents the device's performance when used by the intended lay user (human-in-the-loop).
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • Clinical Accuracy Ground Truth: YSI reference instrument. For blood glucose monitoring systems, a YSI (Yellow Springs Instrument) glucose analyzer is considered a highly accurate laboratory reference method.
    7. The sample size for the training set:

      • The provided document describes a 510(k) summary for a medical device (blood glucose monitor). For such devices, "training sets" in the context of machine learning (AI) are not typically applicable. The device relies on chemical reactions and optical measurement, calibrated during manufacturing and design verification. The "training" for the device's performance is embedded in its design, chemical formulation, and manufacturing process.
      • Therefore, a specific "training set sample size" as one would discuss for an AI algorithm is not applicable here.
    8. How the ground truth for the training set was established:

      • As noted above, a "training set" in the AI sense is not applicable. The device's internal calibration and performance parameters would have been established during its development and manufacturing using high-precision reference methods (like YSI) and extensive analytical studies to ensure accuracy and precision across its operational range. These internal calibration processes are part of the device's engineering and quality control, not a "training set" for an algorithm.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1