510(k) Data Aggregation

    K Number
    K091323

    Validate with FDA (Live)

    Date Cleared
    2009-06-04

    (30 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Modified INTERmedic Diode Laser 980nm System (MULTIDIODE SST 180) is indicated for use in surgical applications requiring the vaporization, excision, ablation, cutting and hemostasis, or coagulation of soft tissue in conjunction with endoscopic equipment for medical specialties including: Urology, Genitourinary (Urology), Thoracic Surgery, Plastic Surgery and Dermatology, Aesthetics including vascular lesions and hair removal, General Surgery, Ophthalmology, Orthopedics, Podiatry, Arthroscopy, Spinal Surgery, Gynecology, Pulmonary Surgery, Neurosurgery, Gastroenterology, Head/Neck/ENT and Radiology, Endovascular coagulation, Oral Surgery and Dental procedures.

    Device Description

    The modified INTERmedic Diode Laser 980mm System (MULTIDIODE SST 180), and the delivery accessories that are used with it, is comprised of the following main components:

    • Laser console with fiber port and RF ID bracket
    • Operating software
    • Detachable footswitch
    • Delivery device accessories (fiber optic, handpieces and scanner)
    AI/ML Overview

    The provided text describes a 510(k) summary for the modified INTERmedic Diode Laser 980nm System (MULTIDIODE SST 180). This document is for a medical device that has been found substantially equivalent to predicate devices, rather than a diagnostic AI/ML device that would typically have the performance criteria and studies described in your request.

    Therefore, the document does not contain the information requested regarding acceptance criteria, device performance tables, sample sizes, expert ground truth adjudication, MRMC studies, standalone performance, or training set details.

    The document focuses on:

    • Device Name: Modified INTERmedic Diode Laser 980nm System (MULTIDIODE SST 180)
    • Indications for Use: Surgical applications requiring vaporization, incision, ablation, cutting, and hemostasis/coagulation of soft tissue across numerous medical specialties (Urology, Thoracic Surgery, Plastic Surgery, Dermatology, General Surgery, Ophthalmology, Orthopedics, Podiatry, Arthroscopy, Spinal Surgery, Gynecology, Pulmonary Surgery, Neurosurgery, Gastroenterology, Head/Neck/ENT, Radiology, Endovascular coagulation, Oral Surgery, Dental procedures, Cardiac Surgery, Vascular Surgery).
    • Predicate Devices: INTERmedic Diode Laser Family (K053540) and Biolitec, Inc. 180W Ceralas D 980nm Diode Laser (Model D180) (K083682).
    • Rationale for Substantial Equivalence: The modified device shares the same indications for use, device operation, technical and functional capabilities with the predicate INTERmedic device, and similar functional capabilities as the Biolitec predicate device.
    • Conclusion: The device was found substantially equivalent to predicate devices.

    This type of submission primarily demonstrates that the new device is as safe and effective as a legally marketed device, and does not require complex performance studies against a ground truth as would be expected for a diagnostic AI/ML algorithm.

    Ask a Question

    Ask a specific question about this device

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The INTERmedic Diode Laser Family (and their delivery accessories used to deliver optical, ultrasound and RF energy) are indicated for use in surgical applications requiring the vaporization, incision, excision, ablation, cutting and hemostasis, or coagulation of soft tissue in coniunction with endoscopic equipment for medical specialties including: Urology, Genitourinary (Urology), Thoracic Surgery, Plastic Surgery and Dermatology, Aesthetics including vascular lesions and hair removal, General Surgery, Ophthalmology, Orthopedics, Podiatry, Arthroscopy, Spinal Surgery, Gynecology, Pulmonary Surgery, Neurosurgery, Gastroenterology, Head/Neck/ENT and Radiology, Endovascular coagulation, Oral Surgery and Dental procedures.

    Device Description

    The INTERmedic Diode Laser Family is designed to deliver laser power at wavelengths of 810nm and 980nm, depending on model, that can be used for the procedures indicated in the section "Indication for Use" in this summary. The ContrÂge model can be used with or without the RF unit. The different models can be identified by their different names, output frequency and maximum power. Each system is comprised with the following main components: Operating software controlled by a microprocessor, A footswitch (detachable), A laser cabinet/console with fiber port, A safety Interlock.

    AI/ML Overview

    The provided document is a 510(k) summary for the INTERmedic Diode Laser Family. This document is a premarket notification to the FDA to demonstrate that the device is substantially equivalent to a legally marketed predicate device, rather than a study designed to prove the device meets specific acceptance criteria in a clinical setting.

    Therefore, many of the requested elements for describing an acceptance criteria study (such as sample sizes, expert qualifications, adjudication methods, MRMC studies, standalone performance, ground truth types, and training set information) are not applicable and not present in this type of regulatory submission.

    The document primarily focuses on demonstrating substantial equivalence based on technological characteristics and intended use compared to predicate devices.

    However, I can extract information related to the device's technical specifications and intended uses, which serve as the basis for the FDA's determination of substantial equivalence.

    Here's the partial information that can be extracted, framed within the context of the device's substantial equivalence claim:


    1. A table of acceptance criteria and the reported device performance

    In the context of a 510(k) submission, "acceptance criteria" for the new device largely revolve around demonstrating that its performance specifications and indications for use are substantially equivalent to those of the predicate devices. The "reported device performance" is a comparison of these technical specifications and intended uses.

    The tables in the document compare the INTERmedic Diode Laser Family models with various predicate devices. Below is a synthesized table based on these comparisons, highlighting key characteristics that are compared for substantial equivalence.

    Feature / ModelINTERmedic ENDO Laser (980nm)INTERmedic SURGICAL Laser (980nm)INTERmedic PL3D (980nm)INTERmedic ContrÂge (980nm + RF)INTERmedic SLP (810nm)INTERmedic VARIUS / VARIUS 1000 (980nm)INTERmedic ODONT (800-980nm)
    Acceptance Criteria (Predicate Device Specification)Dornier K020339 Medilas: Wavelength 940nm, Peak Power 120W, Pulse Mode 10-100ms, Same Indications for use.BioLitec K050824 Medilas: Wavelength 980nm, Peak Power 100W, Exp. Time 0.01-99.9s, Same Indications for use.BioLitec K050824 Ceralas: Wavelength 980nm, Peak Power 100W, Exp. Time 0.01-99.9s, Same Indications for use.Syneron K052324/K050452/K0507796 Polaris/Aurora: Wavelength 900-980nm, Peak Power Up to 100W, Same Indications for use.Dornier K050824/K021724/K020339 Medilas D Skin Pulse: Wavelength 940nm, Peak Power 120W, Same Indications for use.Dornier K050824/K021724/K020339 Medilas D Skin Pulse: Wavelength 940nm, Peak Power 120W, Same Indications for use.BioLitec K050824 Ceralas: Wavelength 980nm, Peak Power 100W, Exp. Time 0.01-99.9s, Same Indications for use.
    Reported Device Performance (INTERmedic Device)INTERmedic ENDO Laser: Wavelength 980nm, Peak Power 50W, Pulse Mode 5ms to 10s, Same Indications for use (Minimally invasive endovenous treatment).INTERmedic SURGICAL Laser: Wavelength 980nm, Peak Power Up to 120W, Exp. Time 5ms to 10s, Same Indications for use (broad surgical applications).INTERmedic PL3D: Wavelength 980nm, Peak Power Up to 25W, Exp. Time 50ms to cw, Same Indications for use (Neurosurgery).INTERmedic ContrÂge: Wavelength 980nm, Peak Power Up to 40W, RF Freq 0.45MHz, RF Energy 50-150 J/cm², Same Indications for use (Aesthetics, Dermatology, wrinkle reduction).INTERmedic SLP: Wavelength 810nm, Peak Power Up to 90W, Exp. Time 5-500ms (vascular), 50-1200ms (hair removal), Same Indications for use (Dermatology, vascular/hair removal).INTERmedic VARIUS/VARIUS 1000: Wavelength 980nm, Peak Power Up to 50W, Pulse interval 50-900ms, Same Indications for use (Dermatology, vascular, skin treatments, cutaneous surgery).INTERmedic ODONT: Wavelength 800-980nm, Peak Power Up to 50W, Pulse interval 50-900ms, Same Indications for use (Oral Surgery).
    Comparison Rationale"Same intended uses and indications for uses as the listed predicate devices, and have similar technological characteristics." Wavelength and power are within acceptable range or variations explained (e.g., higher power doesn't affect safety)."Same intended uses and indications for uses as the listed predicate devices, and have similar technological characteristics." (Increase of power to 120W does not affect safety and security characteristics and provides shorter treatment time)."Same intended uses and indications for uses as the listed predicate devices, and have similar technological characteristics.""Same intended uses and indications for uses as the listed predicate devices, and have similar technological characteristics.""Same intended uses and indications for uses as the listed predicate devices, and have similar technological characteristics.""Same intended uses and indications for uses as the listed predicate devices, and have similar technological characteristics.""Same intended uses and indications for uses as the listed predicate devices, and have similar technological characteristics."

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not Applicable. This document is a 510(k) summary, which does not typically include data from clinical test sets in the manner of a clinical trial. The "test" for substantial equivalence relies on comparing the device's design, specifications, and intended uses against legally marketed predicate devices.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not Applicable. There is no "test set" or ground truth established by experts in this type of regulatory submission. The FDA reviews the provided technical and descriptive information for substantial equivalence.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable. No test set or adjudication process as described in clinical studies is reported.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable. This device is a laser surgical instrument, not an AI-assisted diagnostic or therapeutic tool for which MRMC studies would be relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable. This device is a physical medical instrument, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Not Applicable. The "ground truth" here is the established safety and effectiveness of the predicate devices as legally marketed, and the demonstration that the new device shares similar technological characteristics and intended uses.

    8. The sample size for the training set

    • Not Applicable. This document does not describe a machine learning model or a training set.

    9. How the ground truth for the training set was established

    • Not Applicable. This document does not describe a machine learning model or a training set.

    Summary of the Study that Proves the Device Meets Acceptance Criteria (Substantial Equivalence):

    The "study" described in the 510(k) summary is a comparative analysis demonstrating substantial equivalence to predicate devices already cleared by the FDA.

    • Rationale for Substantial Equivalence: The INTERmedic Diode Laser Family, including its accessories, is claimed to have:
      • The same intended uses and indications for use as the listed predicate devices.
      • Similar technological characteristics to the predicate devices, specifically mentioning "treatment wavelengths, laser media, mode of operation, power output, exposure duration, treatment intervals, laser energy delivery control."
    • Methodology: The submission provides detailed specification tables comparing each INTERmedic model (ENDO, SURGICAL, PL3D, ContrÂge, SLP, VARIUS/VARIUS 1000, ODONT) against specific predicate devices (e.g., PhotoMedex Laser Pro, Syneron Lasers, Ceralas D, Dornier's Diode Laser Family). The comparison highlights parameters such as:
      • Wavelength (nm)
      • Peak Power (W)
      • Pulse Mode/Exposure time
      • Laser Medium
      • Aiming Beam characteristics
      • Power Requirements
      • Operation Mode (Continuous or Pulsed)
      • Physical Dimensions and Weight
      • Indications for use (emphasized as being the same or functionally equivalent).
    • Conclusion of the Comparison: INTERmedic Arfran S.A. believes that "no significant differences exist between the INTERmedic Diode Laser Family and the predicate devices" based on overall performance and technological characteristics, and that "They all work in the same range of wavelength and have the same indications for use." The document explicitly states that the device's introduction provides "no new clinical indications, which have previously demonstrated clinical effectiveness."

    This type of submission relies on the established safety and efficacy of the predicate devices. The "proof" is the detailed comparison showing that the new device does not raise new questions of safety or effectiveness.